Installation & Calibration Guide

Similar documents
AFA 500 FUME HOOD ALARMS

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

LABCONCO 1000 GUARDIAN DIGITAL MONITOR Operating and Instruction Manual

Laboratory Airflow Monitors & Controls TYPE C STANDARD & NIGHT SETBACK LED & DIGITAL - VAV SYSTEMS

AFA1000/1 MK3 Airflow Monitor

Critical Environments. Model FHC50/FHM10. Fume Hood Controller/Monitor. Operation and Service Manual. P/N , Revision B October 2010

FUME HOOD MONITOR ALNOR AIRGARD 335

RPM1600 Series Room Pressure Monitors

Manual Supplement. Model Number: 8680-N2. Communications Protocol. Contents of this manual supplement include:

AirGard. Air Flow Monitors

AIRFLOW SPECIFICATIONS

HVAC Controls Laboratory. Model EVERWATCH Face Velocity Monitor. Operation and Service Manual. P/N , Revision E October 2002

S E R M A N U A L. Alarm. Module. Rooster Monitor100

EVERWATCH FACE VELOCITY MONITOR MODEL 8610

AirGard 385 Fume Hood Monitor

CRITICAL ENVIRONMENT MONITOR ALNOR AIRGARD 350 CEM

Smoothflow Student Fume Cupboard Installation Instructions

Model 405 AIRGARD Fume Hood Monitor

PRESSURA ROOM PRESSURE CONTROLLER MODEL RPC30

AD-1272 Advanced Thermal Dispersion Probe Airflow Measuring System

KONTROL INVIKTA Double

PRESSURA ROOM PRESSURE CONTROLLER MODEL RPC30

Manual Supplement. Product/System Title: Static Pressure Controller. Contents of this manual supplement include:

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

APM110 LOW AIR VELOCITY ALARM

Model 8682-KET Sequenced Offset (Flow Tracking) and Static Pressure Controller with Dual Exhaust Modulation for Temperature Control

CONTENTS USING THE LIGHTS KEY...31 NETWORK ALARMS...32 FAULT AND ERROR MESSAGES...33

SUREFLOW ADAPTIVE OFFSET CONTROLLER MODELS BAC

ANALOX SENSOR TECHNOLOGY LTD ANALOX DD1101 INSTALLATION & OPERATION MANUAL

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

THERMAX Efficiency Monitor

ALARM FOR SAFETY CABINETS SPECIFICATIONS

Flopurge TS. Operation Manual

Residential Gas Condensing Boiler Greenstar ZBR16/21/28/35/42-3A... ZWB28/35/42-3A...

TV2 Room Pressure Monitor Quick Start Guide

Manual Vane Anemometer PCE-VA 20

TX7100. Intelligent Addressable Optical Smoke Detector Installation and Operation Manual. TANDA UK Technology Copyright 2015, All right reserved.

Service Manual Models: 1.0, 1.3, and 1.5

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

Follett Performance Plus

FLOW CONTROLLER TYPE S/601

Warranty Registration

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

Alaris Syringe Module v9

MVE TS Controller (Touch Screen)

FVC-2000 plus Fume Hood Controller. MODEL FVC-2600 Installation, Operation & Maintenance Manual

CONTENTS INSTALLATION 21 USING THE SIGNAL KEY 10 SPECIFICATION 23

MacDonald Steel Limited

RAE Systems RAE Systems RAE Systems Presented by David Lucha RAE Systems Protectio n Through Detection

OPERATION MANUAL. StableTemp Economy Mechanical Convection Oven. Model : ,-90,-91,-92,-93,-94

Fume Hood Operating Display Panel

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English

FUME HOOD MONITOR ALNOR AIRGARD 200

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

Dual Axis Inclinometer - Active

USER MANUAL ADVANCED BIO SAFETY CABINET (ABS) CLASS I. TM010-O&M-001 Revision 3 Page 1 of 61

Tranquility Digital Quick Start Guide

Document No

York 25-Ton VAV Rooftop Unit

GLD-30 Gas Leak Detector

Danfoss Gas Sensor DGS MODBUS

Cascade Sequencer. Control Installation and Operation Instructions. Infinite Energy Boiler Manual. Also read and follow: CS-IOM-1

Product waste disposal - Protection of the environment:

UNIVERSITY OF ALASKA FAIRBANKS DIVISION 25- INTEGRATED AUTOMATION DESIGN STANDARDS SECTION INTEGRATED CONTROL OF FACILITY LAB FUME HOODS

URS-1 SERVICE AND PARTS MANUAL SERIAL NUMBERS: UP TO

AT-G-SENSE Gas Sensing System

User Manual. Digi-Sense TC9600 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

PRESSURA ROOM PRESSURE MONITOR MODELS 8630-SM 8630-PM

FCCP 200: Fume-cupboard indicator and monitor

SUREFLOW ROOM PRESSURE CONTROLLER MODELS BAC

Installation, service and operating instruction Cetetherm Mini ECO Heating & domestic hot water substation for apartments and single family houses

INSTALLATION INSTRUCTIONS

LXG-300 Series. Models with LXG-300 Series Instruments

Flostop TS D7E and A8E. Operation Manual

Installation and Operation Manual

Autoclave Operations Manual

Application Manual Loop Water Controller

HERCULES 6 GRAPHICS SYSTEM

TV2 Clean Room Monitor

Fume Hood Safety TYPES OF FUME HOODS

IPCM INTELLIGENT PRESSURE CONTROL MODULE INSTALLATION AND OPERATIONS MANUAL FOR CONSTANT PRESSURE CONTROL APPLICATIONS VERSION 1.0

DI220 Using the Optional Docking Station The maintenance and operating functions of the Micro IV can be automated with the Micro IV Docking Station.

Residential Gas Condensing Boiler Greenstar combi 100 p / 151 p ZWB28-3A... ZWB42-3A...

MODEL 4300-PG CENTRAL ACCESS PANEL FOR EC-GOLD TRANSMITTERS. User Manual

TSI Model 8630 PRESSURA Room Pressure Monitor Modbus Communications Application Note LC-106

Static Pressure Alarm For Hazardous Location Fume Hoods

INSTALLATION, OPERATION & MAINTENANCE MANUAL BERG TEMPERATURE CONTROL UNIT

Digital Marine Exhaust Temperature Alarm

CT2211 Aerosol Leak Detection System

Installation, Operation and Maintenance Manual. RQ Series. Mold Temperature Control Units 0.75 to 7.5 hp and 0 to 24 kw

Deltapilot S FMB70. Functional Safety Manual. Level and Pressure Measurement with Output Signal ma

Quick start guide. English

NIPPY junior+ Quick Reference Guide

Platinum Touch/Rotem One Touch

Rev Pulse Modulating and Anti- Sweat Control (PMAC II Solo) Installation and Operation Manual

Instruction Manual. Please read this manual and the Safety Instructions carefully before using your refrigerator.

WiFi Hints & Tips. Contents. WiFi Hints and Tips 1. Page No. Section No. Title

CONTENTS Installation Precautions... 1 Chapter 1 Product Introduction... 2 Chapter 2 Technical Specifications... 3

Transcription:

Airflow Monitor AFA1000/2 MK3 Startup The AFA1000 must be field-calibrated once the room air supply and exhaust is balanced. When the unit is powered up, the following sequence of events occurs: 1. The alarm performs a self-test of its functions, LEDs and audible alarm (approximately 3 seconds) 2. At the end of the delay, the unit will do one of two things: a. If the monitor has been calibrated, the unit enters normal operating mode (displays airflow velocity, LED s and audible alarm enabled). b. If the unit has not been calibrated, the monitor will display Requires set up, press Enter to continue Calibration 1. Press Enter from the Requires set up screen or if the monitor is in the Run screen Press and Hold the Enter button for 5 seconds until the Main Menu is displayed. 2. Using the + / - buttons select SET UP, then select CALIBRATION, then enter the password (factory default password is 0-0-0-0, press Enter to continue. 3. Open the sash to the normal operating height and measure the face velocity using a calibrated instrument. Using the + / -buttons enter the measured face velocity then press Enter, the monitor will then sample the airflow for 5 seconds. 4. If the airflow sample is unstable the monitor will display Deviations too High, follow the instructions to repeat the sample or quit the calibration. 5. If the sample is accepted, lower the sash by half and measure the face velocity using a calibrated instrument. Using the + / - buttons enter the measured face velocity and press Enter, the monitor will then sample the airflow for 5 seconds. 6. If the calibration is successful the monitor will return the Main Menu, select RUN to go to normal operating mode. 7. If the airflow sample is unstable the monitor will display Deviations too High, follow the instructions to repeat the sample or quit the calibration. continued over...

8. The monitor will display Increase higher airflow sample if the second sample value is too close to the first value entered, close the sash a little and repeat the higher sample. The minimum difference between the samples that the monitor will accept is 0.3m/sec (60fpm). 9. The monitor will display Sensor diff too low if the monitor doesn t detect any difference in the sensor output between the 2 airflow samples, check that the sensor hose is connected and repeat the calibration. Calibration Tips 1. Ensure that the sensor hose is connected correctly at the side wall and rear of the sensor. 2. Ensure that the extract fan is running and the Fume Cupboard is balanced before calibrating the monitor. 3. Take extra time (at least 15 to 20 seconds) to wait for the airflow to settle before capturing the airflow samples. This will minimize the chance of a calibration error due to turbulence or fluctuations. 4. Avoid movement in front of the Fume Cupboard whilst calibrating the monitor. 5. The Low and High air samples must be at least 0.3m/sec (60 fpm) apart to calibrate the monitor. This is to avoid inaccuracy in the calibration sue to insufficient difference between the samples. The minimum difference can be changed in the Low High diff parameter in the Cal Config menu. 6. Do not use fully open and fully closed sash positions for the calibration points, the recommended positions are normal operating height (e.g. 500mm 20 ) for the first sample and approximately half the sash opening (e.g. 250mm 10 ) for the second sample. 7. If the Fume Cupboard is under separate VAV control then a different set up may be required to achieve the airflow difference required for calibration. Normal extract for the low sample and Purge for the high sample may be used. Troubleshooting Problem No LED s, Screen blank Sensor Error message Audible alarm not working Check for Check the power supply is securely plugged into the monitor Check the power supply is securely plugged into the mains power socket Check that there is 15VDC power supply on the monitor terminals (broken wire or snagged cable sheath) Check the sensor cable is securely plugged into the sensor Check the sensor cable is securely plugged into the monitor Replace the sensor If Sensor Error message still displayed replace the sensor cable If Sensor Error message still displayed replace the monitor Check for screen messages - if SETBACK is displayed the monitor has been remotely muted. Check if alarm has been disabled in Cal Config menu

Connection Details

Monitor Panel Dimensions Monitor Panel Cutout Dimensions (NOT shown to scale) 1 1 (25.4mm) (25.4mm) 5.2 (132mm) 3 (76.2mm) 4.37 (111mm) 2 fixing holes 2.5 mm diameter for 2 #4 selftapping screws, provided. 3.19 (81mm) 2 (50.8mm) SM6 Sensor Dimensions (rear view) SM6 Panel Cut Out Dimensions 0.47 (12mm) 0.47 (12mm) 0.157 (4mm) Tapped Fixing Hole 1.18 (30mm) Dia. Sensor Hole for Vent Tube 0.157 (4mm) Tapped Fixing Hole 2.91 (74mm) 0.67 (17mm) 0.39 (10mm) 0.79 (20mm) 1 (25mm) 1 (25mm) 2.91 (74mm) 0.47 (12mm) Dia. Hole for SM6 cable 1.06 (27mm) 1.06 (27mm)

Airflow Sensor Installation Diagram For complete manual and product information, log on to www.tel-uk.com