INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

Similar documents
INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

MST MODEL S1 AIRLINE "CO" MONITORING AND LOW PRESSURE WARNING SYSTEM MANUAL

RAM 50 Series Operation Manual

Three Phase Simplex. Installation (937) Installation Instructions and Operation/Troubleshooting Manual. Installation of Floats.

RAM 50 Series Operation Manual

WARNING CMS 1 CARBON MONOXIDE MONITOR O. M

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

Carbon Monoxide Monitor and Retrofit Carbon Monoxide Monitor Kit W-2808/37027

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

OPERATING INSTRUCTIONS AND REPLACEMENT PARTS

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

Deluxe Discharge Alarm

WARNING CMS 1 CARBON MONOXIDE MONITOR O. M

PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA USA [215] / FAX [215]

Interscan Corporation. Instruction Manual Series Compact Portable Analyzer

Air Cleaning Equipment, Inc. 303 N. Main St. Broadway, NC iers.com

Models: CO-91IS and CO-91ISLA

RK/ RK Carbon Monoxide Detector Operator s Manual

Single Phase Simplex SXL21=3, SXL24=3, SXH21=3, and SXH24=3

Propane Fume Detection System

WA A Aqua Sub Bottom-load Water Cooler

MGC Dock User s Manual

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Before Putting Into Use. Important. Tips for Energy Savings

Industrial Vacuums, Inc

Document No

IMPORTANT SAFETY INFORMATION:

RK Carbon Monoxide Transmitter Operator s Manual

Carbon Monoxide Sample-Draw Detector Operator s Manual

3 963/963E Benchtop Air Ionizer Instructions

Getz Equipment Innovators 450 lb Dual Portable Dry Chemical Fill System

Propane Fume Detection System With Solenoid Valve Control

Portable Air Conditioner

Installation & Operation Manual Ice Cream Freezers

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02

Model 405 AIRGARD Fume Hood Monitor

OWNERS MANUAL. CARBON MONOXIDE DETECTOR MODELS Panel Mount & Remote Detectors (R) Guardian Avionics 1951 E. AIRPORT DRIVE TUCSON, AZ.

Carbon Dioxide Sample-Draw Detector Operator s Manual

Propane Fume Detection System With Solenoid Valve Control

RK/ RK Oxygen Detector Operator s Manual

Macurco HD-11 Hydrogen Gas Detector User Instructions. Important: Keep these User Instructions for reference

RK-02 Multi Point Detector Operator s Manual

OI-2400-DOCK Multi Gas Docking Station. Operation Manual trevision 2.2w

OWNER'S MANUAL RSL MyBinding.com 5500 NE Moore Court Hillsboro, OR Toll Free: Local:

Owner s Manual Refrigerated Compressed Air Dryers Models F-3528, F-3529, F-3530, F-3531 & F-3532

OWNERS MANUAL MODEL 451 (-101 & -201)

Models LBW-420-LEL (24 VDC powered) Ammonia Leak Detector

Owner s Manual. Filterless UV Air Purifier. This Manual Includes:

MGC Dock User s Manual

Prime chiller. Inline Chiller. Instructions for Models #2645,2646

Combustible Gas Sample-Draw Detector Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

FUME HOOD MONITOR ALNOR AIRGARD 200

Instruction Manual. Nacho Cheese Dispenser and Warmer

Owner s Manual Refrigerated Compressed Air Dryers Models F-200, 250, 300 & F350

Rev. None AUGUST 2012 MDM SERIES MODULAR MEMBRANE AIR DRYERS OPERATOR MANUAL MDM1 MDM4 MDM9 MDM13 MDM24 MDM49 MDM66 MDM106 MDM138

User Manual. Wine Cellar

RK Carbon Monoxide Transmitter Operator s Manual

GAS MONITOR Model Model

Macurco HD-11 Hydrogen Gas Detector

Operator s Manual. The Quietek

RK/ RK Oxygen Detector Operator s Manual

breeze easytm model # F100-1W

CO Guardian LLC Document: E. AIRPORT DRIVE Date: 11/15/05 OWNERS MANUAL

Drink Mixer Operation Manual

H-16.5 Record Cleaning Machine Setup and Instruction Manual

IMPORTANT SAFETY INFORMATION:

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

GAS-Mate. Combustible Gas Leak Detector

Model 1850F Dehumidifier Installation and Operating Manual

PRACTICAL USER S GUIDE FOR THE DIMPLEX ELECTRIC FIREPLACE WITH REMOTE CONTROL

IMPORTANT SAFETY INFORMATION:

IMPORTANT SAFETY INFORMATION:

RK-05 Carbon Monoxide Detector Operator s Manual

VP17A BUNN OPERATING & SERVICE MANUAL

User s Manual For Model Aero-551 (Rev. Orig.)

YOUR PORTABLE AIR PURIFIER

GETZ MANUFACTURING PART # CYLINDER DRYER

User Manual. Juice Extractor MODEL: WJE2BSLA

Combustible Gas/Carbon Monoxide Sample-Draw Detector Operator s Manual

QAS-101 SERIES SINGLE CHANNEL CONTROLLERS

GASGUARD NH 3 2% Ammonia Sensor OPERATING & INSTALLATION MANUAL

PM100/PM100E PLANTER MONITOR. Operator s Manual

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

Installer Friendly Series In-Site. Installation Instructions and Operation/Troubleshooting Manual. Installation. Mounting the Control Panel

Blue Air. Commercial Refrigeration Inc. Installation & Operation Manual Ice Cream Freezers


B.I.C.A Built-In Coffee Appliance

Indoor/Outdoor Electric Patio Heater Model No: HP Operating Instructions HP Soleus Air International

CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE

Introduction... 1 System Overview... 1 System Diagram... 2

Table of Contents Welcome and Thank You Note:

Because you re not like everyone else. MISTING FAN MF24B OWNER S MANUAL

INSTRUCTION MANUAL FOR RIKEN PORTABLE TOXIC GAS MONITOR

REFRIGERATED DROP-INS (2-6)FT-DI Installation and Operating Manual

PRACTICAL USER S GUIDE FOR THE ELECTRIC STOVE

DWC-063RL ROOM AIR CONDITIONER USE & CARE MANUAL MODEL #: DWC-055RL S/N : WINDOW TYPE ROOM AIR CONDITIONER

Transcription:

INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model ABM - 715 M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: 800-831-1525 Mazeppa, MN 55956 Fax : (507)843-4953

Table of Contents Page Number General................... 2 Indicator Lamps............... 3 Run Calibration Switch............ 6 Armed/Disarm Switch............ 6 Electrical Requirements............ 7 Operation - Set up for operation......... 7 - Warm up period.......... 8 Operating Mode.............. 9 Shut Down procedure............ 9 Maintenance................ 10 LCD Analog Display............. 12 Calibration test - Calibration procedure......... 12 - Calibration test........... 14 Troubleshooting guide............ 14 Replacement Parts listing........... 16 Warranty Statement............. 17-1 -

ABM - 715 ABM - 715 CARBON MONOXIDE MONITOR This manual will cover the set-up, operation, maintenance, troubleshooting, and replacement parts listing for the ABM-715 respiratory airline monitor. The ABM - 715 monitor is provided by Martech Services Company, as a stand alone monitor, or as a component of a complete Quality Air Breathing System. This monitor is to be operated as a life saving device. Such a unit requires proper care, testing, calibration, and maintenance. Please read this manual prior to operating and using this instrument. It could save a life! - 2 -

GENERAL: The instrument is housed in a NEMA IV non-metallic case. The case is corrosion resistant, positively pressured by the compressor supply line, and sealed except for a bleed hole to exhaust the compressor's air. The instrument is powered by a standard 110 volt electric source. The instrument comes complete with a case mounted electronic horn assembly. The monitor continuously samples the breathing air from the breathing air source. During use, the monitor takes a continuous sample of the filtered air, which is monitored by a broad band sensor. If the sample air supply fails for any reason such as air blockage, loss of pressure, or excessive high pressure, an intermittent alarm horn will sound and the yellow, STANDBY lamp will illuminate. In the event carbon monoxide is detected in levels of 10 ppm or above, and/or a wide range of toxic gases, a continuous alarm horn will sound and the 'ALARM" red lamp will illuminate. - 3 -

ABM - 715 INSTRUMENT CONTROLS AND INDICATORS: Annunciator Lights and Audible Alarm Horn (see figure #4) The panel has three lamps that provide the following information to the user of this monitor; Alarm light indicates the following of three possible situations; OFF - No audible horn, Safe condition. RED - Audible horn sounding continuously, un-safe condition. This means the air quality is above the allowable exposure limit. Do not use the supplied air system unless this light is OFF. YELLOW - No audible horn, un-safe condition. This means a possible failure in the sensor circuit. Do not use the supplied air system unless this light is OFF. - 4 -

Air Quality light indicates the following of two possible situations; GREEN - No audible horn. Safe condition. This means the air quality is within the permissible exposure levels. Do not use the supplied air system unless this light is ON and GREEN. RED - Audible horn sounding continuously. Un-safe condition. This means the air quality is above the allowable exposure limits. Standby light indicates one of three possible conditions; OFF - No audible horn. Safe condition. This indicates there is correct flow and pressure to the sensor. YELLOW - No audible horn. Warm-up condition. During approximately the first minute of operation, the monitor goes through a warm-up period to allow time for the sensor to stabilize. The audible alarm is de-activated during this time. YELLOW - Audible horn sounds every 10 seconds. This indicates a test sample failure: This alarm is caused by one of three conditions; low flow, high flow, or the unit is in the calibration mode. Low flow is typically due to low pressure setting at the regulator or a plugged or restricted sample line. High flow is typically caused from the regulator being set to high. Do not use the supplied air system unless this light is OFF. Audible horn status light (see figure #4) GREEN - This light indicates whether the audible horn is activated (green) or de-activated (off). - 5 -

LCD Analog Display (see figure #4) - The LCD analog display shows the level of Carbon Monoxide (CO) in parts per million (ppm). It is not unusual for the display to be slightly high for a brief period while the sensor stabilizes after the warm up mode. Run/Calibration Switch; (see figure #4) - Position the toggle switch in the "RUN" position for all functions other than the actual calibration process. If the switch is not in the "RUN" position, the monitor will indicate a yellow, "STANDBY" lamp, with an intermittent horn sound, indicating that a test sample is not reaching the sensor. By correctly positioning the switch, to "RUN", the alarm condition will terminate, after one minute. Armed/Disarm Switch; (see figure #4) - The switch arms and disarms the audible alarm horn. The switch is tripped by inserting a small probe, such as a straightened paper clip, through the small access opening in the faceplate. A small hole will be found on the front panel directly below the "M" in ABM-715. It may take one or two pushes to trip the switch. The audible horn is enabled when the "Armed" lamp is illuminated green. The audible horn is disabled when the "Armed" lamp is unlit. It is recommended that the audible horn be enabled at all times, other than during a recalibration procedure. - 6 -

Electrical Requirements - 120 Volt AC power The monitor is supplied with a 120 volt AC power cord, and can be plugged into a 120 volt grounded electrical outlet. - 12 Volt DC power It is possible to convert the monitor to operate on 12 volt DC. Please contact Martech Services Company for assistance and instructions that will explain this conversion. - External Alarms In addition to the existing case mounted horn, an external alarm may also be added. If for any reason, you wish to add additional, external audible or visual alarms, please contact Martech Services Company for assistance and instructions that will explain this conversion. OPERATION: WARNING Do not use the monitor if it has not be calibrated and tested, as outlined in the calibration section of this manual. Using a monitor that has not been calibrated and tested could result in levels of carbon monoxide to pass above the permitted standards. Set Up for Operation Position the Run/Calibration toggle switch to the "RUN" position. Adjust the sample regulator to allow the flowmeter ball to remain within the flow range indicated on the panel front. - 7 -

Plug the monitor into a grounded 120 Volt AC power supply. If the monitor has been converted to operate on 12 Volt DC, be certain the positive and negative leads are correctly attached. The instrument is not fitted with a on/off switch. Plugging into the proper power supply will begin the warm up process. Warm Up Period With electrical power applied and the sample air flow set as indicated on the flow range (approximate.5 to.8 scfh) the monitor will go through a warm-up period of approximately two minutes. During the warm up period the audible horn is disabled and the "STANDBY" lamp will illuminate yellow. At the end of the warm up period the audible horn is activated, if so selected, and all of the indicator lamps will indicate their respective condition, as previously outlined in the manual. WARNING DO NOT use the air to the hood or mask during the warm up period. During the warm up period the alarms are disabled. OPERATING MODE Following the warm up period, with the toggle switch in the "RUN" position, and the test sample flowing to the instrument in the flow range, the monitor is in full operating mode. - 8 -

Always be certain that the AIR QUALITY lamp is illuminated "Green", and no audible horn sounding, before using the air to supply a hood or mask. SHUT DOWN If the monitor is attached to a compressed air system that is shut down on a regular basis, then the power to the monitor should be removed also. However, if the monitor is attached to a compressed air system that is powered 24 hours per day, there is no need to shut the monitor off. CAUTION If the compressed air source is shut down without removing the power supply to the monitor, the monitor will go into a low flow alarm, sounding an intermittent alarm horn, and displaying a yellow "STANDBY" lamp. MAINTENANCE Cleaning of monitor case and faceplate - Due to the occasional opening of the case cover to periodically calibrate the monitor, this will allow external contaminates to come in contact with the faceplate. Care should be given to prevent contaminants from getting into the instrument itself. - Clean the exterior of the case and the faceplate, with a cloth soaked in a solution of warm water and a mild detergent. Do not clean with solvent of other harsh cleaners. - 9 -

CALIBRATION and TESTING WARNING Be certain you are using the correct calibration test gas for this monitor when performing the calibration procedure. Using the wrong test gas, may cause the monitor to be out of calibration. Calibration; The monitor must be in operation for at least 2 HOURS before performing a calibration. DO NOT calibrate the monitor until it has gone through the warm up period and has been in operation with the sample air flowing through it for at least 2 HOURS. - 10 -

Set up of calibration equipment; (see figure #7) -Dampen the sponge by putting a few drops of water down the hose barb on the top of the clear cylinder. - Confirm that the calibration connector valve is closed. The valve is closed when turned fully clockwise. Use only slight pressure on this valve. - Thread the calibration connector onto the top of the test gas cylinder. Prepare the monitor; (see figure #1) - Open the case cover to access the faceplate by loosing the two screws. - Connect the calibration connector to the "Test Gas port", by aligning the tabs and inserting the tube end connector into the port and turning slightly clockwise to lock into position. - Place the RUN/CAL switch towards "CAL". The "STANDBY" light will immediately turn yellow, and within a few seconds the intermittent audible horn will sound. You can disarm the audible horn as outlined previously in this manual. This will silence the audible horn during the calibration process. CALIBRATION PROCEDURE - Slowly open the calibration connector valve to introduce the test gas. The test gas is entering the unit when the flow meter ball rises. If the flow meter ball does not rise with the calibration valve fully open, then the test gas cylinder is empty. - Adjust the calibration connector valve until the flow meter ball floats between.5 and.8 scfm. The valve is sensitive, and you may need to make several small adjustments to keep the ball in the proper range. - 11 -

- Allow the test gas to flow through the monitor until the LCD analog display stabilizes. (This will take about one minute) - If the LCD analog display stabilizes at 10 ppm, (see figure #8) the calibration process is complete. You may return the monitor to the normal operating mode as outlined previously in this manual - If the display stabilizes at any value other than 10 ppm, you will need to make a calibration adjustment, as outlined in the next section of this manual. (see figure #8) ** DO NOT CALIBRATE THE MONITOR UNLESS IT HAS GONE THOUGH A STABILIZING PERIOD ** CALIBRATION ADJUSTMENT PROCEDURE With the test gas flowing, make up or down adjustments to the LCD analog display by adjusting the calibration control. Use a small screwdriver and turn the calibration control clockwise to increase or counter clockwise to decrease the display reading. During this procedure, ensure the "AIR QUALITY and 'ALARM" lamps have turned RED. - 12 -

When the display shows 10 PPM (center of display) and the "AIR QUALITY" and "ALARM" lamps are RED the instrument is calibrated. RETURNING TO OPERATION MODE AFTER CALIBRATION - Close the calibration connector valve. - Remove the calibration tube end connector from the monitor, by turning the connector counterclockwise and then pulling straight out from the faceplate. - Place the calibration toggle switch to the "RUN" position. The flow meter ball should rise to.5 to.8 scfh. If not, adjust the sample air regulator to keep the ball in the flow range. The "STANDBY" lamp should turn off after approximately one minute. - Make certain the "STATUS INDICATOR" lamp is illuminated green, by activating this feature, as previously outlined in this manual. - Close the monitor case cover and tighten the two screws - Remove the calibration connector from the test gas cylinder. NOTE: The test gas cylinder has a positive seal, and the calibration connector does not. Therefore, the test gas cylinder will leak if the calibration connector is left attached to the cylinder over time. CALIBRATION TEST This procedure may be used, to confirm that the monitor is properly calibrated and responding properly. This test does not include any adjustments of the monitor. - To perform this test, follow the procedure for calibration of this monitor, without making any adjustments to the calibration adjustment control. - 13 -

TROUBLESHOOTING WARNING Shorting electrical components could result in serious electrical shocks, or could damage the monitor. Intermittent Alarm: - A pressure switch mounted on the circuit board monitors pressure of the test sample delivered to the sensor. If pressure at the switch (which is monitored by flow through the flow meter) drops below minimum requirement, or rises above maximum, the monitor will initiate with an intermittent audible alarm. - Before doing pressure tests make sure the RUN/CAL toggle switch is in the "RUN" position. - Make sure the compressed air source is between 55 psi and 145 psi. If the test sample is higher than 55 psi, the restriction is internal. Proceed as follows; - Check the flow meter ball to be certain it is within the range as indicated on the faceplate. In some cases the range may be slightly higher or lower than the indicated range. Slowly increase or decrease pressure on the sample air regulator. If the regulator pressure was outside the normal limits, the audible alarm will turn off as soon as the correct pressure is reached. - If conditions continue, check the internal monitor air lines for breaks, kinks, or disconnection. If a line has come loose, reconnecting it will re-establish function and the monitor will then operate properly. - 14 -

TROUBLESHOOTING (continued) No illumination at the lamps or audible alarms; - This condition indicates a loss of electrical power to the monitor. Make certain the power cord is connected to the appropriate power source. - Make certain the fuse, located on the outside, left of the case is in good condition - Make certain the power supply is in fact, on. Instrument will not calibrate; - If the display does not stabilize, or respond to calibration adjustments, then the RUN/CAL toggle switch is not set to "CAL" - Check flow meter. Calibration connector valve may be off, or calibration bottle may be empty. - Sensor has failed. Replace with new sensor. - 15 -

REPLACEMENT PARTS Item # Part Number Description 1 74715 AMB-715 Monitor - complete 2 74710 Calibration connector 3 85910 Test gas - 10 ppm CO (not pictured) 4 74714 Sample air regulator - complete 5 74716 Flow meter 6 85958 Sensor - Broad Band (not pictured) 7 85985 Calibration Kit, includes test gas - 16 -

WARRANTY STATEMENT: Martech Services Company, warrants each new ABM-715 offered by them, to be free from defective material and workmanship as to its merchantability and fitness for the purpose intended, for a period of one (1) year, from the date of sale to the original purchaser, and agrees to remedy any such defect or to furnish a new part (at the company's option) in exchange for any part of any unit of its manufacture which under normal use, and service disclosed such defect; provided the product is delivered by the purchaser to Martech Services Company's factory, intact, for our examination, provided that such examination discloses, in our judgment, that it is thus defective. The warranty does not extend to any systems which have been subjected to misuse, neglect, accident, unauthorized modification, or to use in violation of instructions furnished by us, nor does it extend to systems which have been repaired or altered outside of our factory. THIS WARRANTY IS IN LIEU OF ALL WARRANTIES EXPRESSED OR IMPLIED AND NO REPRESENTATIVE OR PERSON IS AUTHORIZED TO ASSUME FOR US ANY OTHER LIABILITY IN CONNECTION WITH THE SALE OF OUR SYSTEMS. ALL IMPLIED WARRANTIES ARE LIMITED TO THE DURATION OF THIS WRITTEN WARRANTY. IN NO EVENT IS MARTECH SERVICES COMPANY, LIABLE FOR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES ARISING FROM ANY BREACH OF WARRANTY OR THE ABM - 715 MONITOR. - 17 -

RECORDS: Date of Purchase: Purchased from: Serial Number: CONTACT INFORMATION: Dealer Name: Dealer Address: Dealer Phone: MANUFACTURER INFORMATION: Martech Services Company P.O. Box 7079 Mazeppa, MN 55956 U.S.A. Office: 507-843-4700 Toll Free: 800-831-1525 Fax: 507-843-4953 Website: www.breathingsystems.com - 18 -

NOTES: