PQS performance specification

Similar documents
PQS performance specification

PQS performance specification

PQS Independent type-testing protocol

PQS performance specification

TITLE: Ice-lined Refrigerator or Combined Refrigerator-Icepack Freezer, Compression Specification reference:

PQS performance specification

PQS performance specification

PQS performance specification

PQS performance specification

PQS performance specification

PQS Independent type-testing protocol

PQS Quality Assurance protocol. TITLE: Cold Room/Freezer Room:

PQS Independent type-testing protocol

No previous protocol.

PQS Independent type-testing protocol

` PQS Independent type-testing protocol

PQS Independent type-testing protocol

PQS Independent type-testing protocol

Record of Amendments. Version Date Amendment Section/page affected

Guidance on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment (RoHS) Regulations 2012

This is a preview - click here to buy the full publication. IEC Quality Assessment System for Electronic Components (IECQ System)

Guidance Note No 21/1. CE Marking Windows and doorsets to BS EN 16034

CO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 RECOMMENDATION FOR USE

Installation Manual Premier Elite IXP-W

safety requirements are to comply with this compulsory specification twelve (12) months after the publication of this notice.

IECRE OPERATIONAL DOCUMENT

7.3 Design and Development Design and development planning Design and development inputs Design and development outputs 7.3.

INTERNATIONAL STANDARD

Premier External TD. Installation Manual INS352-3

EVC-P CONVENTIONAL PHOTOELECTRIC SMOKE DETECTOR. instruction manual. evolution EVC-P Conventional Photoelectric Smoke Detector

Installation Manual. Premier IXP-W INS538

Tinytag Plus LAN Data Logger with Temperature & Relative Humidity Probe (-25 to +85 C/0 to 100% RH) Standalone. Tinytag Connect

Achieve Success with European Medical Device Commercialization. By Russ King

RADAR REFLECTOR STANDARD FOR CERTIFICATION. DET NORSKE VERITAS Veritasveien 1, N-1322 Høvik, Norway Tel.: Fax:

Implementation of the Construction Products Regulation (CPR) in harmonized standards - Template for Annex ZA -

Shipboard incinerators with capacities up to 4000 kw

Update all extract references to NFPA documents (including NFPA 72) in Chapter 3 and related annex material to the latest editions.

INTERNATIONAL STANDARD

Guidance Note No 19/1. CE Marking windows and doorsets

INTERNATIONAL STANDARD

DEPARTMENT OF TRADE AND INDUSTRY DEPARTEMENT VAN HANDEL EN NYWERHEID. NATIONAL REGULATOR FOR COMPULSORY SPECIFICATIONS ACT (Act 5 of 2008)

INTERNATIONAL STANDARD

WÄRTS1LÄ. EC-Declaration of Conformitv. In accordance with the Marine Equipment Directive 2014/90/EU

Test Report. Appliance EMF Testing

Infrared Thermometer

Part 2-55: Medical electrical equipment. Particular requirements for the basic safety and essential performance of respiratory gas monitors

CE Marking and Technical Standardisation

MARINE GYRO COMPASS STANDARD FOR CERTIFICATION. DET NORSKE VERITAS Veritasveien 1, N-1322 Høvik, Norway Tel.: Fax:

AUGUST 2016 CONTENTS:

RULEBOOK ON ELECTROMAGNETIC COMPATIBILITY

ISO INTERNATIONAL STANDARD. Fire detection and alarm systems Part 16: Sound system control and indicating equipment

Technical Assessment Guideline Applicable for the Kiwa KQ Certification Scheme:

LogTag TICT - is0 Tag

INTERNATIONAL STANDARD

SUPPLEMENTARY REGULATIONS SPECIFIC TO CERTIFICATION OF PERSONAL PROTECTIVE EQUIPMENT (PPE)

On Electromagnetic Compatibility

EL765. Fire Extinguishers [EL /2/ ]

CONFORMITY ASSESSMENT GUIDELINES. Version 1.0

KFP-CF Series Operation Manual

2. Safety instructions

SOUND RECEPTION SYSTEM

ISO 3999 INTERNATIONAL STANDARD. Radiation protection Apparatus for industrial gamma radiography Specifications for performance, design and tests

MT4 MAX/MT4 MAX + Infrared Thermometer

Chain saws (electric mains powered and battery powered)

Wöhler CM 220. Carbon Monoxide Meter. The Measure of Technology. Contents. 1. Important information Specification... 13

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family

Portable Gas Detector RP110. User Guide. Explosive gas detection SEMICONDUCTOR sensor technology Intrinsically safe. Conformity EMC EN50270.

ZP2 Series Operation Manual

(Publication of titles and references of harmonised standards under Union harmonisation legislation) (Text with EEA relevance) (2016/C 054/02)

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD

EMC Regulatory Requirements: What you need to know for the EU, EAEU & Australia. Dr Alex Martin

This document is a preview generated by EVS

BLUE ANGEL The German Ecolabel Stationary air conditioners DE-UZ 204 Basic Award Criteria Edition August 2016 Version 1

Refrigerator/Freezer Thermometer

Lawnmowers (pedestrian controlled, battery powered)

INTERNATIONAL STANDARD

CE marking & your legal obligations

JBCE input for the RoHS FAQ document

EU-Type Examination Certificate

INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

INTERNATIONAL STANDARD

SIMPLIFIED VOYAGE DATA RECORDER (S-VDR)

ISO INTERNATIONAL STANDARD

TECHNICAL REPORT IEC TR General testing procedures for medical electrical equipment. Edition

Official Journal of the European Union

Conformity assessment procedures for the Alarm Components Scheme

Declaration of Conformity Declaration of Performance Notes

Circuit Breakers JANUARY The electronic pdf version of this document found through is the officially binding version

TECHNICAL SPECIFICATION

62 MAX/62 MAX + Infrared Thermometer

E/ECE/324/Rev.1/Add.27/Amend.5 E/ECE/TRANS/505/Rev.1/Add.27/Amend.5

CE marking & your legal obligations

Procedure for the Approval of New Fire Detection and Alarm Technologies

Operating Instructions

Infrared Thermometer. Getting Started

INTERNATIONAL STANDARD

Important Safety Notice

38-SDMS-03 SPECIFICATIONS FOR LV DIGITAL PANEL METERS

Transcription:

PQS performance specification WHO/PQS/E006/TH06.2 Original: English Distribution: General TITLE: Integrated electronic thermometer, with or without alarm function, for vaccine refrigerators and freezers Specification reference: E006/TH06.2 Product verification protocol: E006/TH06-VP.2 Issue date: 6 July 2010 Date of last revision: 30 November 2006 Contents: 1. Scope:... 2 2. Normative references:... 2 3. Terms and definitions:... 3 4. Requirements:... 3 4.2 Performance:... 3 4.2.1 Operating temperature range:... 3 4.2.2 Accuracy:... 3 4.2.3 Resolution:... 3 4.2.4 Power source:... 4 4.2.5 Sensor:... 4 4.2.6 Product response time:... 4 4.2.7 Calibration:... 4 4.2.8 Unit of measurement:... 4 4.2.9 Logging interval:... 4 4.2.10 Mode of operation:... 4 4.2.11 Alarm:... 4 4.2.12 Alarm settings:... 4 4.2.13 Casing:... 5 4.2.14 IP rating:... 5 4.2.15 Battery:... 5 4.2.16 Electromagnetic compatibility:... 5 4.3 Environmental requirements:... 5 4.3.1 Ambient temperature range during transport and storage:... 5 4.3.2 Ambient humidity range during transport, storage and use:... 5 4.3.3 Resistance to electrical storms:... 5 4.4 Physical characteristics:... 5 4.4.1 Overall dimensions:... 5 4.4.2 Weight:... 5 4.5 Interface requirements:... 5 4.6 Human factors:... 5 4.6.1 User interface:... 5 General requirements:... 5 4.6.2 Activation:... 6 4.6.3 De-activation:... 6 4.7 Materials:... 6 4.7.1 Ozone depleting chemicals... 6 4.7.2 Other restricted materials:... 6 4.8 Warranty:... 6 WHO/PQS/E006/TH06.2 Page 1 of 9 06 July 2010

4.9 Servicing provision:... 6 4.10 Disposal and recycling:... 6 4.11 Instructions:... 7 4.12 Training:... 7 4.13 Verification:... 7 5. Packaging:... 7 6. On-site installation:... 7 7. Product dossier:... 7 8. On-site maintenance:... 8 9. Change notification:... 8 10. Defect reporting:... 8 Revision history:... 9 1. Scope: This specification describes the performance requirements for an integrated electronic thermometer, with or without alarm function, for vaccine refrigerators and freezers. This product is intended for use as the primary means for monitoring storage conditions in vaccine refrigerators and freezers in intermediate stores and health facilities. It can be offered in one or more of the following configurations: Type A: Permanently integrated device, supplied by the manufacturer of an electric refrigerator or freezer with a mains power supply. Type B: Permanently integrated device, supplied by the manufacturer of a solar refrigerator or freezer powered by a battery pack charged from a photovoltaic array. Type C: Permanently integrated device, supplied by the manufacturer of a solar direct drive refrigerator or freezer or a gas or kerosene-fueled absorption refrigerator or freezer. Cross-reference to other type-testing protocols: This specification should be read in conjunction with the relevant performance specifications from the E003/ series for all vaccine refrigerators or freezers that incorporate an integrated thermometer of the type described herein. 2. Normative references: EMAS: European Union Eco-Management and Audit Scheme. EN 12830:1999: Temperature recorders for the transport, storage and distribution of chilled, frozen, deep-frozen/quick-frozen food and ice cream. Tests, performance and suitability. European Union Directive 2002/96/EC: Waste Electrical and Electronic Equipment. IEC 60529: Consolidated Edition 2.1 (incl. am1): Degrees of protection provided by enclosures (IP Code). ISO 9001: Quality Management Systems Requirements. ISO 14001: 2004: Environmental management systems Requirements with guidance for use. ISO/IEC 17025: 2005: General requirements for the competence of testing and calibration laboratories. WHO/PQS/E006/TH06.2 Page 2 of 9 06 July 2010

3. Terms and definitions: EPROM: Electrically erasable, programmable, read-only memory. In writing: means communication by letter, fax or email. LCD: Liquid Crystal Display. LED: Light-emitting diode. Legal Manufacturer: The natural or legal person with responsibility for the design, manufacture, packaging and labeling of a product or device before it is placed on the market under his own name, regardless of whether these operations are carried out by that person himself or on his behalf by a third party. Montreal Protocol: Montreal Protocol on Substances that Deplete the Ozone Layer NIST: United States National Institute of Standards and Technology. Reseller: A commercial entity, licensed to act on behalf of a Legal Manufacturer, and which carries product liability and warranty responsibilities no less onerous than those carried by the Legal Manufacturer. 4. Requirements: 4.1 General: Electronic thermometer for monitoring storage conditions in vaccine refrigerators and freezers. These devices are supplied by the refrigerator or freezer manufacturer as a permanently integrated component of the product. The following configurations are covered by this specification: Type A1: Electronic thermometer fitted to a mains powered refrigerator or freezer. Type A2: As Type A1, but with a high and low alarm function. Type B1: Electronic thermometer fitted to a photovoltaic refrigerator or freezer with a solar charged battery pack. This type is NOT suitable for direct drive photovoltaic refrigerators as described in PQS specification E003/RF05, nor is it suitable for gas or kerosene fueled absorption refrigerators. Type B2: As Type B1, but with a high and low alarm function. Type C: Electronic thermometer powered by a photovoltaic cell which forms part of the device. This type draws no power from the product in which it is installed and is therefore suitable for use with gas and kerosene fueled absorption refrigerators and freezers, and with direct drive photovoltaic refrigerators as described in PQS specification E003/RF05 and E003/RF06. 4.2 Performance: 4.2.1 Operating temperature range: Upper limit: +50 C with over-range protection as clause 4.3.1. Lower limit: -30 C with under-range protection as clause 4.3.1. 4.2.2 Accuracy: Type A and Type B devices: ±0.5 C or better within the range -30 C to +20 C. Type C devices: ±1.0 C or better within the range -30 C to +20 C. 4.2.3 Resolution: WHO/PQS/E006/TH06.2 Page 3 of 9 06 July 2010

±0.1 C or better within the range -30 C to +20 C 4.2.4 Power source: Type A: Power is taken from the 110/240 volt 50/60 Hz mains supply. The device must be fitted with a rechargeable back-up battery in the event of power failure. The battery must be capable of powering the device for a minimum period of 72 hours in the absence of mains power. Type B: Power is taken from the main solar-charged battery pack or from the ancillary battery of a direct drive solar refrigerator with ancillary battery (PQS specification E003/RF06). An additional rechargeable back-up battery is not required. Type C: Power is supplied by a photovoltaic cell which forms part of the thermometer. The photovoltaic cell must be able to operate the device at an ambient lighting level of 100 lux, or when a torch is shone onto the device. A response delay of up to 30 seconds is acceptable when a torch is used to activate the device when it is dormant. 4.2.5 Sensor: One or more electronic sensors permanently located in the refrigerator or freezer cabinet to give a true worst-case reading of the load temperature. The worst-case reading in refrigerators is to be the lowest temperature measured during testing and the worst-case reading in freezers is to be the highest temperature measured during testing. The reading must refer to a point lying within the zone allocated for the storage of vaccines. 4.2.6 Product response time: T90 20 minutes maximum in accordance with EN12830:1999. 4.2.7 Calibration: Each product is to be covered by a Certificate of Traceability and Calibration. The traceability declaration is to confirm that the measurement standards and instruments used during calibration of the product are traceable to an ISO/IEC 17025 accredited testing laboratory, to NIST, or to another internationally recognized standards agency. The certificate must be accompanied by a copy of the reference instrument calibration certificate. 4.2.8 Unit of measurement: Temperatures must be recorded in degrees Centigrade. 4.2.9 Logging interval: Devices which measure the storage temperature at discreet intervals are to sample at an interval not greater than 10 minutes. 4.2.10 Mode of operation: The device is to display the most recently logged temperature inside the refrigerator or freezer cabinet. If the temperature in the cabinet exceeds the high alarm setting or drops below the low alarm setting the device is to signal the relevant alarm condition(s) until the alarm is reset by the user. 4.2.11 Alarm: Type A2 and Type B2 devices must include a high breach and low breach visual alarm event indicator. An audible alarm is optional. 4.2.12 Alarm settings: Upper and lower alarm settings in Type A2 and Type B2 devices must be factory programmed into the device as follows: Version for vaccine refrigerators: a) Low alarm setting: Exposure to a single temperature event of -0.5 C or below for 60 minutes. WHO/PQS/E006/TH06.2 Page 4 of 9 06 July 2010

b) High alarm setting: Exposure to a single temperature event of +10 C or above for 10 hours. Version for vaccine freezers: a) High alarm setting: Exposure to a single temperature event above -15 C for 60 minutes. A low alarm is not required. Once an alarm event has occurred, the alarm event indicator is to remain activated until cancelled by the user, even if the temperature inside the cabinet has subsequently returned to within acceptable limits. 4.2.13 Casing: Non-corrodible plastics or metal case. 4.2.14 IP rating: Protection of the product not less than IEC 60529: IP54. 4.2.15 Battery: For Type A devices only: As defined in clause 4.2.4. Rechargeable batteries must be easily replaceable without need for tools. 4.2.16 Electromagnetic compatibility: Operation of the device must be unaffected in the normal electromagnetic compatibility environment in which it is intended to work, taking into account disturbance generated by adjacent apparatus which is compliant with relevant ISO, EN, or other internationally recognized standards. Information required to ensure uninterrupted use of the device must be contained in the user instructions. 4.3 Environmental requirements: 4.3.1 Ambient temperature range during transport and storage: -30 C to +55 C with device inactivated. 4.3.2 Ambient humidity range during transport, storage and use: 5 to 95% RH, non-condensing. 4.3.3 Resistance to electrical storms: The functionality of the device must not be affected by intense electrical storm activity. 4.4 Physical characteristics: 4.4.1 Overall dimensions: Not critical. 4.4.2 Weight: Not critical. 4.5 Interface requirements: The device must be capable of being incorporated into the control panel of a refrigerator or freezer in a position to be decided by the refrigerator/freezer manufacturer. 4.6 Human factors: 4.6.1 User interface: General requirements: The digital temperature read-out and optional visual alarm is to be integrated into the control panel on the refrigerator or freezer cabinet. WHO/PQS/E006/TH06.2 Page 5 of 9 06 July 2010

The display is to be legible by a person with normal vision (with or without glasses) standing in front of the cabinet, both in bright sunlight and in tungsten/fluorescent lighting at 100 lux on the working plane. Type A1 and Type A2 devices: There must be a replace battery warning display which is triggered when 10% or 90 days of battery life remains (whichever period is the longer). Type A2 and Type B2 devices: The visual alarm must unambiguously show both high and low temperature breaches. The optional audible alarm must achieve a minimum sound intensity of 100dB(A) at a distance of one metre from the sounder. The pattern of the signal is to be an intermittent pulse. There must be a switch on the control panel to reset the alarm. It must not be possible to confuse this with the any other control on the refrigerator or freezer. 4.6.2 Activation: Type A1 and Type A2: The device is to be activated by connection of the refrigerator or freezer to a mains electricity supply. In the event of mains failure or loss of photovoltaic power it is to remain fully activated until the supply is restored or the charge on the backup battery expires. Type B1 and Type B2: The device is to be activated by connection of the refrigerator or freezer to a photovoltaic electricity supply. Backup power in low sun conditions is supplied by the main battery pack. Type C: The device is to be activated by the ambient lighting, whether natural or artificial, in the room in which the product is installed. In health facilities without electricity, nighttime activation of a dormant device will be achieved using a torch. 4.6.3 De-activation: It should not be possible to de-activate the product inadvertently. 4.7 Materials: 4.7.1 Ozone depleting chemicals: During manufacture and assembly of the printed circuit boards and final assembly of the product do not use any substance included in Annex A, B or C of the Montreal Protocol. 4.7.2 Other restricted materials: The product and its constituent components, including batteries, must not contain lead, mercury, cadmium, hexavalent chromium, polybrominated biphenyls (PBB) or polybrominated biphenyl ethers (PBDE). 4.8 Warranty: The product is to be covered by the warranty offered by the manufacturer of the appliance into which the device is integrated. 4.9 Servicing provision: The product is to be maintenance-free apart from routine battery replacement. 4.10 Disposal and recycling: WHO/PQS/E006/TH06.2 Page 6 of 9 06 July 2010

The manufacturer is to provide information to the buyer on the hazardous materials contained within the system and suggestions for resource recovery/recycling and/or environmentally safe disposal. For the European Union WEEE compliance in accordance with European Union Directive 2002/96/EC is mandatory. 4.11 Instructions: User instructions are to be incorporated into the instruction manual issued by the manufacturer of the appliance into which the device is integrated. 4.12 Training: No requirement. 4.13 Verification: In accordance with PQS Verification Protocol E006/TH06-VP.2. 5. Packaging: Materials used for packaging the finished product are to be free of ozonedepleting compounds as defined in the Montreal Protocol. 6. On-site installation: Not applicable. 7. Product dossier: General note: This product will be offered as an integrated component of a refrigerator or freezer complying with one of the E003/ series performance specifications. The product dossier requirements below are to be read as a supplement to the relevant E003/ performance specification. The legal manufacturer or reseller is to provide WHO with a pre-qualification dossier containing the following: Dossier examination fee in US dollars. General information about the legal manufacturer of the integrated product, including name and address. Unique identification reference for the product type. Full specifications of the product being offered, covering all the requirements set out in this document, including details of product marking and traceability. Certified photocopy of Certificate of Traceability and Calibration traceable to an ISO/IEC 17025 accredited testing laboratory, to NIST, or to another internationally recognized standards agency. Certified photocopies of all type-approvals obtained for the product, including CE marking and the like. Certified photocopies of the legal manufacturer s ISO 9001 quality system certification. Where relevant, certified photocopies of the legal manufacturer s ISO 14001 certification, EMAS registration or registration with an equivalent environmental audit scheme. Conformity with an environmental audit scheme is not mandatory; however preference will be given to manufacturers who are able to demonstrate compliance with good environmental practice. WHO/PQS/E006/TH06.2 Page 7 of 9 06 July 2010

Where available, laboratory test report(s) proving conformity with the product specifications. Sample refrigerator or freezer incorporating the product as defined in the relevant E003/ series performance specification. 8. On-site maintenance: Not applicable. 9. Change notification: The legal manufacturer or reseller is required to advise WHO in writing of any changes which adversely affect the performance of the product after PQS prequalification has taken place. 10. Defect reporting: The legal manufacturer or reseller is required to advise WHO and the UN purchasing agencies in writing in the event of safety-related product recalls, component defects and other similar events. WHO/PQS/E006/TH06.2 Page 8 of 9 06 July 2010

Revision history: Date Change summary Reason for change Approved 21.09.2006 1. cross-reference clause added. 4.2.16 added. 4.3.1: upper limit changed to +55 C, storage added. 4.3.2: storage added. 4.6.1: 5 th bullet point changed. New clause 4.7.2. 4.7.3 and 4.7.4 deleted. 5: CFC changed to ozonedepleting. Corrections. Consistency with other specifications during final review. EU RoHS Directive material restrictions incorporated. UK (30 November 2006 - PQS secretariat) 06.07.2010 - General revision to include newly defined Type A, Type A1, Type B, Type B1 and Type C devices. - Reference to maximum/minimum thermometers removed. 2. Normative references updated. Comments received from solar refrigerator manufacturers. UK (July 2010) WHO/PQS/E006/TH06.2 Page 9 of 9 06 July 2010