GENERAL LABORATORY REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

Similar documents
ICE CREAM TOPPING CABINETS REFRIGERATOR or FREEZER Installation, Operation and Maintenance Instructions

REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

UNDERCOUNTER LABORATORY REFRIGERATORS and FREEZERS Installation, Operation and Maintenance Instructions

User Manual. Commercial School Milk Cooler User s Manual. Commercial School Milk Coolers: MC34, MC49, MC58

BSD SERIES REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

REFRIGERATION SYSTEMS INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

BACK BAR REFRIGERATION Installation, Operation and Maintenance Instructions

BOTTOM MOUNT REFRIGERATORS & FREEZERS Installation, Operation and Maintenance Instructions

BOTTOM MOUNT REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

Beverage Air IMPORTANT

DIRECT DRAW BEER COOLERS Installation, Operation and Maintenance Instructions

CHEF BASE CABINETS Installation, Operation and Maintenance Instructions

POLAR TEMP FARM MORTALITY UNIT OPERATION MANUAL

Installation, Operation, and Maintenance Information

User Manual. Commercial Reach-In Refrigerators: RF-1D, RF-2D, RF-3D. Commercial Reach-In Freezers: FZ-1D, FZ-2D

IMPORTANT PLEASE RETAIN FOR YOUR RECORDS

CHEF BASE CABINETS Installation, Operation and Maintenance Instructions

Installation and Operations Manual for HORIZONTAL DISPLAY MERCHANDISERS

Operation Manual SCT14B and SCT18B. Inspection. 3 General Description. 3 General Requirements. 3 Standard Features.

Table of Contents. Specifications... page 2. Installation... page 3. Customizing... page 4. Reversing door swing... page 5

Installation & Operations Manual

InstructIon Manual KrEs EQuIPMEnt stands

POLAR TEMP XPORTABLE ICE MERCHANDISER SETUP MANUAL

SCIENTIFIC WARMING CABINET Installation, Operation and Maintenance Instructions

Installation & Operations Manual for. BMG-HGP Model Bottom Mounted Full-Length Swing Glass Door Refrigerator Merchandisers. 4/17 Rev.

EQUIPMENT CARE AND CLEANING

COMMERCIAL REFRIGERATOR AND FREEZER USER S MANUAL

SERIES NAV S Display Merchandisers

Installation & Operations Manual for R-290 SERIES. BMG-HGP Model Bottom Mounted Full-Length Swing Glass Door Refrigerator Merchandisers 2/

Refrigerators. Technical Manual and Replacement Parts List FORM NO REV H MODEL SKTTR7FW MODEL SKC1220W

Glass Door Merchandisers GDC10, GDC12F, GDC15, GDC23, GDC24F, GDC40, GDC40F, GDS47, GDC69

UNDERCOUNTER REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

BLG-HGP Model Bottom Mounted Full-Length Swing Glass Door Freezer Merchandisers

INSTALLATION and OPERATION INSTRUCTIONS

BMG-HGP & BMG-SLP Model Bottom Mounted Full-Length Glass Door Refrigerator Merchandisers

Installation & Operations Manual. Master-Bilt Products 908 Highway 15 North New Albany, MS Phone: (800) REV: LN PN

T-Series Air Conditioner T15 Model

Installation & Operations Manual Master-Bilt Products 908 Highway 15 North New Albany, MS Phone: (800)

Installation & Operations Manual Master-Bilt Products 908 Highway 15 North New Albany, MS Phone: (800)

PROAIR Air Conditioner. CR23 Model INSTRUCTION MANUAL nvent Rev. D P/N

Installation and Operation Instructions

User Manual. Commercial Refrigerator And Freezer User s Manual

INSTALLATION and OPERATION INSTRUCTIONS

Drop-In Refrigeration System

SPECTRACOOL Air Conditioner. N21 Model INSTRUCTION MANUAL nvent Rev. G P/N

UNIT MODEL/ POWER SUPPLY VOLTAGE VARIATION AMBIENT AIR ON CONDENSER

T-SERIES Air Conditioner. T20 Model INSTRUCTION MANUAL nvent Rev. C P/N

-30 C Undercounter Plasma Freezer

REFRIGERATED PREP TABLES Installation, Operation and Maintenance Instructions

RCM-77. Instruction Manual. G-Series Cooler. Manual is for the following model: RCM-77-N23EB. U.S. Patent No. 8,215,125 RECHARGEABLE COLD MERCHANDISER

REFRIGERATOR SAFETY. Your safety and the safety of others are very important.

G-7s. Instruction Manual. G-Series Cooler COUNTERTOP COOLER. Part No.11IPA

GCG-10. Instruction Manual. G-Series Cooler. Manual is for the following models: GCG-10-N33EB G-10-N33EB UPRIGHT COOLER

T-Series Air Conditioner T20 Model

RPI Industries, Inc.

Operation, Installation and Instruction Manual for Backbar, Keg & Bottle Coolers And Backbar, Glass & Plate Chillers

User Manual

Instruction Manual. Chest Freezer. Model: ACFM050ADB INSTALLATION INSTRUCTIONS. Before Using Your Freezer Remove the exterior and interior packing.

Condensing Unit Installation and Operating Instructions

User Manual. Commercial Chest Freezer User s Manual 360CF5, 360CF8, 360CF14, 360CF18

PROAIR Air Conditioner. CR43 Model INSTRUCTION MANUAL nvent Rev. H P/N

GCG-9. Instruction Manual. G-Series Cooler. Manual is for the following models: GCG-9-N13EB G-9-N13EB GCG-9-B13EB UPRIGHT COOLER

T-SERIES Air Conditioner. T50 Model INSTRUCTION MANUAL nvent Rev. F P/N

Technical Manual and Replacement Parts List. f Real Freezers

Worktop and Undercounters

INSTRUCTION MANUAL (UNIT APPEARANCE MAY VARY FROM IMAGE) BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

Installation & Operations Manual. Master-Bilt Products 908 Highway 15 North New Albany, MS Phone: (800) /17 Rev.

Ui REFRIGERATOR SPEC SHEET

PROAIR Air Conditioner. CR29 Model INSTRUCTION MANUAL nvent Rev. I P/N

Glass Door Undercounter and Countertop Freezers

HOUSEHOLD FREEZER 048-GM-48303/048-GM-48304/048-GM-48305

T-Series Air Conditioner T53 Model

Outdoor and Indoor Electric Refrigerator Model COAR

GS-1. Instruction Manual. G-Series Cooler. Manual is for the following models: GS-1-N14NF GS-1-N14NG GS-1-N14NI COUNTERTOP COOLER

Installation and User Manual for Model UF424 and UN324

INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS for HOAM and VOAM Open Air Merchandisers

Installation and User's Manual for Refrigerator Model SCR33

Glass Door Merchandiser Refrigerator and Freezer

Instruction Manual. 4.6 Cu. Ft. Refrigerator/Freezer. Wards.com Wards.com The solution for your cold-storage needs!

Solid Door Top Mount Reach-In Freezer

T-SERIES Air Conditioner. T43 Model INSTRUCTION MANUAL nvent Rev. I P/N

G-10f/GCG-10f UPRIGHT COOLER

UNDERCOUNTER LABORATORY FRONT BREATHING REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions

!"#$%&'(&) !(*$%"'&!$() !"#$

Chef Base Refrigerators

Compact Refrigerator. User's Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE USING YOUR REFRIGERATOR AND KEEP IT FOR FUTURE REFERENCE.

Preparing the Cabinet

REF5-BB Undercounter Blood Bank Refrigerator

T-SERIES Air Conditioner. T29 Model INSTRUCTION MANUAL nvent Rev. I P/N

Owner s Manual Unit Specifications...3 Rating Information...3. Built Right First... To Last! Warning!

CHEST FREEZER INSTRUCTION MANUAL. Model No.: EWCF5WBX EWCF7WBX

Bar Fridge USER MANUAL MB46W

COMMERCIAL REFRIGERATOR & FREEZER SERVICE MANUAL (SCL/SCLM)

RCM-77. Instruction Manual. G-Series Cooler. U.S. Patent No. 8,215,125 RECHARGE COLD MERCHANDISER

INSTALLATION AND OPERATION INSTRUCTIONS STANDARD BACK BAR COOLERS, SELF CONTAINED MODEL NOS.

INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

KellllRlore REFRIGERATOR. Safety Instructions Installation Operation Care and Maintenance Troubleshooting Parts List

Before Putting Into Use. Important. Tips for Energy Savings

G-10s. Instruction Manual. G-Series Cooler UPRIGHT COOLER. Part No.11IPA

TCW 2000 Ice liner refrigerator and freezer PIJ

Transcription:

GENERAL LABORATORY REFRIGERATORS AND FREEZERS Installation, Operation and Maintenance Instructions INSPECTION When the equipment is received, all items should be carefully checked against the bill of lading to insure all crates and cartons have been received. All units should be inspected for concealed damage by uncrating the units immediately. If any damage is found, it should be reported to the carrier at once, and a claim should be filed with the carrier. This equipment has been inspected and tested in the manufacturing facility and has been crated in accordance with transportation rules and guidelines. Manufacturer is not responsible for freight loss or damage. INSTALLATION GENERAL After the unit crate and crate base have been removed, locate the casters or legs found inside the cabinet. Attach the casters or legs to the unit base by screwing them into the same threaded fittings that were used to secure the crate base to the cabinet. Insure that the legs or casters are screwed completely into the base. If for some reason the doors are not squared up on the cabinet, the doors can be adjusted. Opening the door(s) and loosening the screws that hold both the top and bottom hinges to the cabinet can accomplish this. After adjusting the door so that it is aligned correctly, tighten the screws to securely hold the hinges in place. VAPORIZER INSTALLATION The General Laboratory forced air models require the installation of the condensate vaporizer, which is shipped inside the unit compartment. The vaporizer should be mounted to the cabinet bottom. Reference View 1. In lieu of the vaporizer, these units can be plumbed to a floor drain following local plumbing codes. The cold wall units do not use a vaporizer. Customer installation is not required. 1 4/06 119503

Note: Disconnect power prior to installation. 1. Connect the power cord to the vaporizer and place it in the vaporizer bracket. 2. Attach the vaporizer bracket to the back of the unit using the four screws as shown, making sure the plastic drain tube is in the vaporizer pan. LOCATION QuickTime and a Photo - JPEG decompressor are needed to see this picture. The refrigeration system located at the top of the cabinet requires free air access for proper operation. Allow a minimum four-inch clearance on the top, rear, and sides of the cabinet. The cabinet should also be leveled when it is placed in its permanent location. View 1 ELECTRICAL Check the proposed outlet to be used to insure that the voltage, phase, and current carrying capacity of the circuit from the electrical panel correspond to the requirements of the cabinet. Never use an extension cord to wire any unit. On permanently connected units, those not furnished with a plug-in service cord, all inter-wiring between the electrical panel and the unit must be done in accordance with the National Electric Code and all state and local codes. Refer to the serial tag for all pertinent electrical information. Observe all Warning Labels. Disconnect power supply to eliminate injury from electrical shock or moving parts when servicing equipment. GENERAL OPERATION FORCED AIR UNITS The General Laboratory forced air cabinets employ a unit cooler evaporator located inside the cabinet as the heat-removing source. Through the refrigeration process, heat is captured in the evaporator, transferred to the condensing unit on top of the cabinet, and expelled to the surrounding outside air. It is extremely important to allow a four-inch clearance on the top, rear, and sides of the unit for the refrigeration process to function properly. Cold control adjustment To increase or decrease the internal cabinet temperature, rotate the cold control, located on the unit cooler, to the desired temperature. 2 4/06 119503

Refrigerators During the operation of refrigerator units, frost will periodically form on the coil surface. Each time the compressor cycles "off", the evaporator fans will continue to run, which will keep the internal temperature uniform and at the same time remove any frost build up on the coil. The water produced will collect in the unit cooler drain pan and travel down the drain tube to the condensate vaporizer. Freezers After shutting the door on freezer models, a short amount of time must be allowed before the door can be reopened. This is due to the tight seal maintained between the door and the cabinet. Waiting a few moments for the pressure to equalize permits the door to be opened easily. A positive defrost is required to remove frost from the coil in freezer models. This is accomplished by energizing heaters during the defrost cycle that are positioned on the coil surface. The time clock is factory set to allow four defrosts per day. Note: Adjustment of the correct time of day will be required to maintain these defrost periods. QuickTime and a Photo - JPEG decompressor are needed to see this picture. As the preset defrost time is reached, the time clock automatically terminates the refrigeration process by turning off the condensing unit and unit cooler fan motors, and energizes the defrost heaters. As the coil temperature increases, the frost begins to melt producing water that runs down the coil to the unit cooler drain pan and exits through the drain tube to the vaporizer. After all the frost has been removed and the coil temperature reaches approximately 50 F, the defrost is terminated through the action of the defrost termination control located on the unit cooler, and the refrigeration process resumes. In order to insure that any excess water remaining on the coil is not sprayed into the cabinet interior, and all heat generated by the defrost is removed, the unit cooler fans will not operate until the Figure coil 1 temperature reaches approximately 25 F. Time clock adjustments (Freezers) Setting the correct time of day: To set the correct time of day simply rotate the small inner dial counterclockwise until the correct time of day on the large dial is opposite the "time" indicator. In referring to Figure 1, the correct time of day shown is 7:30 a.m. Number of defrosts per day: The timer is factory set to defrost the evaporator four times a day. If more defrosts are required, remove a knurled slotted screw from the holder, insert the screw into the time slot on the large dial where a defrost is desired, and tighten. Insure that an equal time exists between each defrost pin. Defrost length adjustment: On the small, upper dial there is a pointer that is used to set a 100% fail-safe feature. The fail-safe of the timer, refer to Figure 1, is factory set at 30 minutes. With this feature, the defrost will terminate if a system malfunction occurs during defrost. No adjustment of this device should ever be necessary. Lengthening the fail-safe time will not lengthen the defrost cycle. 3 4/06 119503

Figure 1 shows the time switch set to defrost at 2:00 a.m., 8:00 a.m., 2:00 p.m., and at 8:00 p.m. The fail-safe pointer indicates that the system will revert to cooling thirty minutes after a normal defrost cycle began should a malfunction occur. WARNING: The fail-safe feature is a safety mechanism and should never be used to control the length of the defrost period. This feature is provided to protect the contents of the freezer from damage should the system fail to revert to the cooling cycle because of mechanical difficulties. Note: The time switch shown in the illustration is factory set to provide four defrosts per day. This setting should provide adequate defrosting for even the most heavily used system. If experience shows that the four set defrost periods are more than necessary, one or more of the knurled setscrews may be removed. AT LEAST ONE OF THE KNURLED SET SCREWS MUST BE LOCATED ON THE LARGE LOWER DIAL AT ALL TIMES. COLD WALL UNITS The cold wall models employ evaporator tubing embedded in the sidewalls, back, and top of the unit. During the refrigeration process, heat is removed through the evaporator tubing and expelled through the condensing unit. Because of the cold wall design, it will be necessary to occasionally manually defrost the freezer models. To do this, unplug the unit and open the door. Wipe the cabinet dry when defrosting is complete. With cold wall refrigerators, condensation may form on the interior of the cabinet and should be periodically wiped dry. Cold Control Adjustment The temperature control knob is located on the top of the cabinet behind the facade. To make a small increase or decrease in the internal cabinet temperature, rotate the control knob as needed. Enzyme Freezer Only The Enzyme Freezer is equipped with a programmable controller to control the settings. The controller, which is located on the facade of the unit, is factory set. Please see the separate instructions that are included on the operation of this controller. GENERAL MAINTENANCE PERIODIC CLEANING Beginning with the initial installation, the interior surfaces of the cabinet should be periodically wiped down with a solution of warm water and baking soda. This solution will remove any odors from spillage that has occurred. The exterior of the cabinet should also be cleaned frequently with a commercial grade of glass cleaner. Caution: Never use an abrasive or alkaline solution to clean these units. Wipe dry after cleaning. Important: Monthly cleaning of the condenser coil located on top of the cabinet should be done. This cleaning will aid the heat transfer characteristics of the refrigeration system and increase its efficiency. Dust, dirt, and lint will tend to accumulate on the fins of the condensing unit. This obstruction will affect the flow of air through the condenser, thereby lowering the efficiency of the system. A wire brush or a brush with stiff bristles can be used to loosen these particles that are attached to the fins so that they may be removed with a vacuum cleaner. Failure to keep the condenser coil clean and clear of obstructions could result in temperature loss and damage to the compressor. 4 4/06 119503

All moving parts have been permanently lubricated and will generally require no maintenance. MAINTENANCE SERVICE AND ANALYSIS GUIDE REFRIGERATION SYSTEMS - ALL MODELS MALFUNCTION POSSIBLE CAUSE SOLUTION Compressor will not start - 1. Service cord unplugged 1. Plug in service cord no hum 2. Fuse blown or removed 2. Replace fuse 3. Overload tripped 3. Determine reasons and correct 4. Control stuck open 4. Repair or replace 5. Wiring incorrect 5. Check wiring against the diagram Compressor will not start - 1. Improperly wired 1. Check wiring against the diagram hums but trips on overload 2. Low voltage to unit 2. Determine reason and correct protector 3. Starting capacitor defective 3. Determine reason and replace 4. Relay failing to close 4. Determine reason, correct or replace Compressor starts and runs, 1. Low voltage to unit 1. Determine reason and correct but short cycles on overload 2. Overload defective 2. Check current, replace overload protector protector 3. Excessive head pressure 3. Check ventilation or restriction in refrigeration system 4. Compressor hot-return gas hot 4. Check refrigerant charge, fix leak if necessary Compressor operates long 1. Short of refrigerant 1. Fix leak, add refrigerant or continuously 2. Control contact stuck 2. Repair or replace 3. Evaporator coil iced 3. Determine cause, defrost manually 4. Restriction in refrigeration 4. Determine location and remove restriction system 5. Dirty condenser 5. Clean condenser Compressor runs fine, but 1. Overload protector 1. Check wiring diagram short cycles 2. Cold control 2. Differential too close - widen 3. Overcharge 3. Reduce charge 4. Air in system 4. Purge and recharge 5. Undercharge 5. Fix leak, add refrigerant Starting capacitor open, 1. Relay contacts stuck 1. Clean contacts or replace relay shorted or blown 2. Low voltage to unit 2. Determine reason and correct 3. Improper relay 3. Replace Relay defective or burned out 1. Incorrect relay 1. Check and replace 2. Voltage too high or too low 2. Determine reason and correct Refrigerated space too warm 1. Control setting too high 1. Reset control 2. Refrigerant overcharge 2. Purge refrigerant 3. Dirty condenser 3. Clean condenser 4. Evaporator coil iced 4. Determine reason and defrost 5. Not operating 5. Determine reason, replace if necessary 6. Air flow to condenser or 6. Remove obstruction for free air flow evaporator blocked Standard temperature system 1. Control setting is too low 1. Reset the control freezes the product 2. Control points stuck 2. Replace the control Objectionable noise 1. Fan blade hitting fan shroud 1. Reform or cut away small section of shroud 2. Tubing rattle 2. Locate and reform 3. Vibrating fan blade 3. Replace fan blade 4. Condenser fan motor rattles 4. Check motor bracket mounting, tighten 5. General vibration 5. Compressor suspension bolts not loosened on applicable models - loosen them 6. Worn fan motor bearings 6. Replace fan motor Pan Area 1. No cooling 1. Make sure switch is in the "on" position 2. Too cold 2. Adjust temperature control - see instructions 5 4/06 119503

under pan area 3. Too warm 3. Adjust temperature control - see instructions under pan area 6 4/06 119503