MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

Similar documents
MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

MST MODEL S1 AIRLINE "CO" MONITORING AND LOW PRESSURE WARNING SYSTEM MANUAL

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE

Carbon Monoxide Monitor and Retrofit Carbon Monoxide Monitor Kit W-2808/37027

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

RK-02 Multi Point Detector Operator s Manual

RK/ RK Oxygen Detector Operator s Manual

CT-7 Series Toxic Gas Transmitter Operator s Manual

RK Hydrogen Sulfide Transmitter Operator s Manual

RK Multi Point Detector Operator s Manual

RK/ RK Carbon Monoxide Detector Operator s Manual

Carbon Dioxide Sample-Draw Detector Operator s Manual

GAS MONITOR Model Model

RK CO 2 Transmitter Operator s Manual

RK-05 Combustible Gas Transmitter Operator s Manual

OPERATING INSTRUCTIONS AND REPLACEMENT PARTS

Models TX-KE and TX-KP Toxic Gas Transmitters

RK Combustible Gas Transmitter Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

MODEL LP3 INSTALLATION AND SERVICE INSTRUCTIONS FOR STREAMLINE LP3 STROBE LIGHTS

SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE -

Combustible Gas Sample-Draw Detector Operator s Manual

RK-03 Hydrogen Detector Operator s Manual

Carbon Monoxide Sample-Draw Detector Operator s Manual

RK-05 Carbon Monoxide Detector Operator s Manual

RK-05 Carbon Monoxide Transmitter Operator s Manual

Models: CO-91IS and CO-91ISLA

Tox-COMB/ANA Combustible Gas Transmitter Operator s Manual

RKA-LEL Sample Draw Combustible Gas Detector

RK Carbon Monoxide Transmitter Operator s Manual

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

GLD-30 Gas Leak Detector

Beacon 800 Gas Monitor Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual

RK/ RK Oxygen Detector Operator s Manual

Model OEM-2 INSTRUCTIONS FOR USE

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

RK-CH4-4 Combustible Gas Transmitter Operator s Manual

TECHNICAL MANUAL TM... Fryer, Deep-Fat, Electric w/solid State Controls 440 Volt, 60 HZ, 3 Phase NSN:

Model Gas Alarm Panel APPLICABILITY & EFFECTIVITY. This manual provides instructions for the following Sierra Monitor products:

INSTALLATION. Glass Panel Doors (select models) CAUTION

ph COOLING TOWER CONTROLLER

Model AV1LED Audio-Visual LED Light Installation and Maintenance Manual

H-10PM. Refrigerant Leak Detector

RK/ RK Combustible Gas Detector Operator s Manual

RK Transmitter Technical Notice

Combustible Gas/Carbon Monoxide Sample-Draw Detector Operator s Manual

Automatic Humidity Controller. Model 512 (Revision C) Operating Manual

Rev B, 9/2/2009. Kodiak Chiller Overview

SEC 2000 Millenium Infrared Gas Detector

Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual

R100 Oil-Less Refrigerant Recovery Unit

RKA-06 Sample-Draw Detector

V SERIES HDR GAS RANGES

REF Series Undercounter Refrigerator

RK / RK Methane Detector Operator s Manual

QAS-101 SERIES SINGLE CHANNEL CONTROLLERS

Dual Zone Wine Cellar

BFN. Static Eliminators. Instruction Manual. BFN 802 Ionizing Blower

Instruction Manual CHLORALERT TM Series T17CA4000 Chlorine Gas Leak Detector CAPITAL CONTROLS

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

V-220 / V-320 Heater Kits Installation Instructions TK IM (Rev. 0, 06/17)

Oil and Water Retort Instruction Manual

Ceiling Mount Air Purification System

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

Warranty Registration

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

NO LONGER IN PRODUCTION

Interscan Corporation. Instruction Manual Series Compact Portable Analyzer

MAYFIELD CEILING FAN LISTED E ITEM # MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE.

Guardian CR2000 Overhead Ionizing Air Blower. User s Manual

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

BFN. Static Eliminators. Instruction Manual. BFN 803 Ionizing Blower

Installation Guide. For Models: E-MLV36BNK4LK1 E-MLV36ESP4LK1 E-MLV36WW4LK1 READ THESE INSTRUCTIONS AND AND SAVE THEM FOR FUTURE USE

Los Angeles Abrasion Machine HM-70A & HM-70AF

Incutrol/2 Regulator

Beacon 110 Gas Monitor Operator s Manual

G-7s. Instruction Manual. G-Series Cooler COUNTERTOP COOLER. Part No.11IPA

RKA-03 Sample-Draw Detector Head

RLD-H10P Refrigerant Leak Detector

Your electrician must furnish sufficient voltage levels, and current for proper operation of your machine. We recommend this machine be on a

Controllers. Instruction Manual WARNING

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

Instruction Manual WARNING

FIELD-INSTALLED MOTION ACTIVATED LED DIMMER CONTROL

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

Victoreen Primalert 35 Area Radiation Monitor

Eldex Column Heater. Operator s Manual

KARMEL KING Instruction Manual Models: 2620 & 2621 (240 Volt) 2630 & 2631 (208 Volt) Part No Revised: February 2009

Installation Instructions

ELECTRIC FLAT PANEL FIREPLACE HEATER

PRE INSTALLATION PROCEDURES

Caliber Series LED Edgelit Exit Sign EXPORT AC Only or Emergency Operation READ AND FOLLOW ALL SAFETY INSTRUCTIONS

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model 120VC Single Relay (100/25 PPM) (200/35 PPM)

Control the Elements. 220V / 1PH / 50Hz

Transcription:

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL IMPORTANT WARNING WHEN THE OXYGEN MONITOR IS CORRECTLY INSTALLED AND MAINTAINED, IT MONITORS THE LEVEL OF OXYGEN IN ITS SURROUNDINGS 07/11 Rev. 4

MST MODEL 2002 OXYGEN MONITOR SPECIFICATIONS: RANGE: 0.0-25.0% O 2 POWER REQUIREMENTS: INTERNAL RELAYS (RL1 & RL2) DRY CONTACT-TYPE MAXIMUM CURRENT RATING: 120/240 VAC-50/60 Hz, 1 PH, OR 12 VDC; 2 A 5A@12 VDC/120 VAC SENSOR TYPE: ELECTROCHEMICAL. SENSOR LINEARITY AT 68F: +/-2.5% OF READING SENSOR TEMPERATURE COEFFICIENT: +/-10% OF READING PER SIGNAL OVER RANGE 32-122F OPERATING TEMPERATURE (AMBIENT): 32-104F (0-40C) STORAGE TEMPERATURE (AMBIENT): 14-122F (-10-50C) PHYSICAL DIMENSIONS: WIDTH: 7-5/8" (194 mm) HEIGHT: 7-5/8" (194 mm) DEPTH: 4-1/8" (105 mm) WEIGHT: 2.3 LBS. (1 kg) 2

MST MODEL 2002 OXYGEN MONITOR 3

MST MODEL 2002 OXYGEN MONITOR I. INTRODUCTION The MST, Inc. Model 2002 OXYGEN (O 2 ) Monitor is a continuous O 2 monitor utilized to monitor AREA locations such as parking garages, maintenance rooms, and the like. WARNING: THE MST MODEL 2002 OXYGEN MONITOR IS NOT APPROVED FOR USE IN HAZARDOUS OR EXPLOSIVE ATMOSPHERES. PERSONAL INJURY, PROPERTY DAMAGE OR DEATH COULD RESULT FROM IMPROPER USE OF THIS DEVICE. The Model 2002 Monitor comes with internal dry contact-type relays that can be utilized to switch external devices such as solenoid valves, alarms, etc. As shipped from the factory,1/2 of each of the internal relays is powered with 12 VDC to provide power to the audible and visual alarms located on the front of the unit. Should you wish to utilize both sides of the internal relays for other applications, please consult factory prior to making any changes. The internal relays are preset at the factory to activate when the Oxygen in the sample either exceeds 23 ½ % (red lamp), or drops below 19 ½ % (audible alarm). These can be adjusted, if required. Please consult this manual. II. DETAILED DESCRIPTION INSIDE CLEAR COVER DISPLAY: Indicates the percent (%) of O 2 present in the air sample. FUSE: Main Power Fuse,.2 A 3AG (TYPE 312). POWER LAMP: Green LED Lamp indicates power supplied to unit. SPAN: SPAN Adjustment for calibration. BOTTOM OF UNIT KNOCK OUTS: For direct wiring in various applications. POWER CORD: 3-Prong Grounded cord for power connection. AUDIBLE ALARM: Piezo buzzer type alarm - energized when low O 2 alarm level activated. SENSOR SAMPLING OPENING: Where actual O 2 Sensor is located. HIGH OXYGEN WARNING LAMP: Illuminates when O 2 level exceeds maximum permissible O 2 level. 4

MST MODEL 2002 OXYGEN MONITOR 5

III. OPERATION PRELIMINARY 1. Connect the Power Cord to a suitable SWITCHED receptacle according to Local, State and National Regulations. Follow proper electrical codes. 2. Observe that the green Power lamp illuminates and that the display is operating showing (3) digits such as xx.x. 3. With the included calibration screwdriver (P/N 80044), carefully adjust the SPAN Adjustment Screw, noting the reading on the display before turning and slowly turn the adjustment clockwise to increase the reading, counterclockwise to decrease the reading. At the alarm level setting you will hear the relay(s) activate and, depending on whether the reading is low (below 19 1/2% factory preset audible alarm) or high (above 23 ½ % factory preset-red warning lamp), you will observe the actual alarming device. Verify the setting is correct. To adjust the alarm level setting(s), consult this manual. 4. Perform the calibration procedure as outlined in this manual. IV. NORMAL OPERATION During normal operation of MST s Model 2002 O 2 Monitor, a continuous air sample is taken by the sensor. The sensor outputs a current proportional to the amount of O 2 that is present in the air sample and this current is converted to % Oxygen and read on the display. When O 2 levels above or below the alarm set point is achieved, the internal relay(s) are activated, and the appropriate alarm device is activated, and/or some other suitable device that is connected to its own power supply (relays operated as dry contacts). The internal relays have the capability of operating in a Normally Open or Normally Closed position. When powering the monitor s internal devices, the unit is factory wired so the low oxygen setting (19 1/2% factory preset-rl2, controlling SK2) is connected to the normally open terminal, and the high oxygen setting (23 ½% factory preset- RL1, controlling SK1) is connected to the normally closed terminal. 6

V. MONITOR CALIBRATION CALIBRATION FREQUENCY The MST Model 2002 O 2 Monitor should be calibrated PRIOR TO USE and rechecked after the first (2) weeks. The monitor should then be calibrated on a MONTHLY basis if it is used on a daily or continuous basis. The monitor should be calibrated on a PRIOR TO USE basis if it is only used occasionally (non-continuous or non-daily basis). CALIBRATION PROCEDURE (MST Calibration Kit P/N 8003107 required) Before calibrating the instrument, connect the power cord to the appropriate power source and verify that the green lamp illuminates and the display is operating properly. Allow the instrument to warm up for a period of five (5) minutes or longer prior to the following calibration procedure. Connect one end of the supplied PVC Calibration tubing (3/16" ID) to the barbed fitting on the gas regulator. Connect the gas regulator to the calibration gas bottle and hold the open end of the Calibration tubing almost directly on the sensor. Allow approximately 1-2 minutes for the sensor to stabilize before making any adjustments as noted below. NOTE: When using the 8003107 Small Calibration Kit, unscrew regulator immediately following any necessary adjustments to conserve your calibration gas. CALIBRATION STEPS: 1. Remove the protective plastic cap (if supplied) from the ZERO AIR (21% O 2 ) cylinder outlet. 2. Open the clear cover on the Model 2002 CO Monitor by depressing the latch on the right side and opening the cover by pulling up on the right side of the cover and swinging it to the left (cover is hinged on left side of monitor). 3. Screw the regulator valve onto the ZERO AIR (21% O 2 ) cylinder outlet, making sure not to cross thread the regulator on the cylinder outlet port, and tighten regulator firmly. Hold the calibration tubing almost directly on the sensor. Allow approximately 1-2 minutes for the sensor to stabilize before making any adjustments. 4. Observe the display on the instrument. The reading should move to 20.9% after approximately (1) minute. If the display does not read 20.9%, adjust the SPAN potentiometer such that display now reads between 20.9 and 21.0%. 5. When the above steps have been completed, un-screw the regulator valve from the Zero Air (21% O 2 ) cylinder and replace plastic protective cap (if supplied). 7

Place all calibration kit components back into protective plastic case to prevent damage or loss of components. WARNING: ONLY QUALIFIED AND TRAINED ELECTRICAL TECHNICIANS SHOULD PERFORM ANY REPAIRS OR ADJUSTMENTS INTERNAL TO THE MONITOR. PROPERTY DAMAGE, PERSONAL INJURY OR DEATH COULD RESULT. VI. INTERNAL RELAY ADJUSTMENTS To change/set the internal relays to activate when the OXYGEN level reaches a specified point, refer to the following procedure (Note: You must have 21% O 2 available to properly perform this procedure). 1. Disconnect power source from the monitor and open front clear cover as described in the Calibration Procedure noted above. 2. With the proper screwdriver, remove the four (4) small phillips-head screws located on the front cover plate to expose the main circuit board. Carefully lift off the cover by grasping the fuse holder diameter and gently pulling upward. DO NOT USE EXCESSIVE FORCE AND DO NOT PULL COVER FURTHER THAN WIRING WILL ALLOW. DAMAGE TO MONITOR COULD OCCUR. 3. Move faceplate cover slightly to the left of the monitor s plastic housing, being careful not to damage any wiring or components. 4. Locate the two (2) small potentiometers situated in the UPPER RIGHT HAND corner of the main circuit board. The one on the left adjusts RL1(Low O 2 alarm) and the one on the right adjusts RL2 (High O 2 alarm). 5. Making sure that no components are interfering or contacting any of the internal main circuit board components, connect appropriate power source and observe that monitor green power lamp illuminates and display is operating properly. 6. To adjust the high-level relay set point, slowly adjust the SPAN calibration potentiometer to force the display to read the high-level alarm setting you wish the internal relay to be activated at. Then, slowly adjust RL1 potentiometer counterclockwise until the High Oxygen Warning Lamp turns on (if not already on), then turn the potentimeter clockwise until lamp turns off, then very slowly counterclockwise until the lamp just turns on. Check setting by turning the SPAN adjustment potentiometer counter-clockwise (below high level setting) and then clockwise, noting when the high level alarm activates. 7. To adjust the low-level relay set point, slowly adjust the SPAN calibration potentiometer to force the display to read the low-level alarm setting you wish the internal relay to be activated at. Then, slowly adjust RL2 potentiometer clockwise until the Low Oxygen Audible Alarm activates (if not already activated), then turn the potentiometer counter-clockwise until the Low Oxygen Audible Alarm deactivates, then very slowly clockwise until the alarm just activates. Check setting by turning the SPAN adjustment potentimeter clockwise (above low alarm level setting) and then counter-clockwise, noting when the Low Oxygen Audible Alarm activates. 8

8. Disconnect power source and carefully replace front faceplate, making sure wires are not caught under edge of faceplate and nothing is interfering with internal components or wiring. 9. Reconnect power source and retest both alarm settings by turning the SPAN calibration potentiometer clockwise to high oxygen set point and counter-clockwise to low oxygen alarm set point to verify correct settings. 10. Reconnect power source and retest the setting per #7 above. 11. Re-calibrate the monitor using ZERO (21% O 2 ) air as outlined in the Calibration Procedure noted above. VII. INTERNAL WIRING Except for sensor replacement, this device contains no user-serviceable parts, however a trained electrical technician may, at their option, attach various control and alarm devices to the terminal strip located under the sensor access door. Electrical knockouts are provided at the lower end of the monitor for purposes of connecting various devices to this monitor. Refer to the following schematic for proper location and connections to terminal strips. The two (2) banks of two (2) terminal strips on the lower right hand side on the main circuit board are used to provide dry contact relay connections. The relays in the monitor are DPDT-type relays and one half (½) of each relay is utilized by the factory to power the internal alarm devices (audible and visual alarms, powered by main circuit board). The remaining side of each relay are dry contacttype only. The bank on the farthest right is activated by the alarm setting made by the RL1 adjustment potentiometer. The next bank left of that is activated by the alarm setting by the RL2 adjustment potentiometer. Proper wiring practices shall be followed in all instances regarding connections to these terminal strips. 9

VIII. MAINTENANCE Except for fuse replacement, sensor replacement and calibration, this device does not require any maintenance unless subjected to abuse or neglect. Make sure the air sample supply is free from any oils, water or solvent to prevent the sensor from premature failure. Keep all solvents away from plastic housing at all times. SENSOR REPLACEMENT 1. Disconnect the monitor from the power source. 2. Loosen the two (2) screws from the front access cover to allow the cover to be removed. 3. The sensor is a circular device located inside a gray housing and secured in place by the cover when assembled. The sensor has a red and a black wire soldered to tabs on the sensor. 4. Snip the red and black wire close to the actual sensor tab, noting the correct location of each wire on the sensor (+ or -, + is red, - is black). Remove and discard the old sensor, keeping in mind that the sensor contains a small amount of sulphuric acid. 5. Strip and tin the two wires and solder them neatly to the two tabs on the new sensor. DO NOT USE EXCESSIVE HEAT OR SENSOR DAMAGE MAY OCCUR AND THIS WILL VOID YOUR WARRANTY. 6. Slide new sensor and wiring back down into the gray housing, arranging such that the two tabs on the sensor face the recess in the gray housing. 7. Replace the access cover, arranging all internal wires such that no wire is caught when the cover is secured and the wires are not interfering with the sensor end and sealing foam. Secure cover with two (2) screws. WARNING: SULFURIC ACID IS POISONOUS AND CAN CAUSE SEVERE BURNS. DO NOT ALLOW ACID TO CONTACT SKIN OR EYES. IF EYES ARE EXPOSED TO ACID, FLUSH THOROUGHLY AND SEEK IMMEDIATE MEDICAL ATTENTION. ALWAYS WASH HANDS THROUGHLY AFTER HANDLING DETECTOR CELL. 8. Calibrate instrument according to the above Calibration Procedure. 10

REPLACEMENT PARTS (Please contact your local distributor to purchase these parts.) REPLACEMENT O 2 SENSOR: 80589 REPLACEMENT POWER LAMP 80055-G SMALL CALIBRATION KIT 8003107 REPLACEMENT FUSE 80569 REPLACEMENT HIGH O 2 LEVEL LAMP (Lens unscrews to access lamp) 80109 Consult Factory for any other replacement parts. MST INC. SHALL NOT BE LIABLE FOR ANY INJURY, LOSS OR DAMAGE, (DIRECT OR CONSEQUENTIAL), ARISING OUT OF THE USE OF OR THE INABILITY TO USE THIS PRODUCT, BEYOND THE REPLACEMENT OF DEFECTIVE MATERIALS OR WORKMANSHIP. USERS OF SUPPLIED AIR RESPIRATORS SHOULD EVALUATE THEIR OWN PARTICULAR APPLICATION AND PERFORM THEIR OWN TESTS FOR AIR QUALITY TO DETERMINE THE SUITABILITY FOR USE OF THIS PRODUCT. For further information, or questions about service or maintenance care of this unit, contact your local distributor or MST, Inc. at (800) 542-6646. 11