GC DETECTORS Reduction Gas Detector - RGD. Overview

Similar documents
POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview

FPD - Flame Photometric Detector

FIXED GAS ANALYZER (FGA)

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

GC Injectors. GC Injectors

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Aromatics and Oxygenates in Gasoline June2018

TYPES AND APPROPRIATE INJECTORS

Chromatography. Research Supplies. 2014/2015 Catalog

INJECTORS Purge & Trap General Operating Procedures continued

White paper on the Dielectric Barrier Discharge Detectors. Introduction:

Preparing the 6850 Series II GC for 597x MSD Installation

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

NEW Restek Electronic Leak Detector

Chemistry Instrumental Analysis Lecture 4. Chem 4631

Chemtron Science Laboratories Pvt. Ltd.

Operate the Mutli-Gas Monitor PID

Selecting the right generator for your GC system. By Ed Connor, GC Product Specialist, Peak Scientific Instruments Ltd

Agilent 8355 Sulfur and 8255 Nitrogen Chemiluminescence Detectors

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

TENDER NOTICE NO. 07/2012 INVITATION FOR BIDS / NIT

VM-3000 Mercury Vapor Monitor

Restek s Electronic Leak Detector

Purge and Trap Troubleshooting Made Easy?

Mercury Vapor Monitor VM-3000

AIR CONDITIONING. Carrier Corporation 2002 Cat. No

GG-CO-EXP CARBON MONOXIDE SENSOR. Installation and Operation Manual

Primer on Flame Ionization Detectors

Agilent 8355 S Sulfur and 8255 S Nitrogen Chemiluminescence Detectors

Thermo Scientific TVA2020 Introduction

VM-3000 Mercury Vapor Monitor

BST-MG07 Multi-gas Detecting Alarm Manual Instruction

TCD-to-FID Series Connection on a 6820 GC, Accessory G4319A

SEC-MALS Noise Assessment and Cleaning Guide

22 The Flame Ionization Detector

Thermo Scientific. TRACE GC Ultra. Gas Chromatograph Maintenance and Troubleshooting Manual. PN , Revision March 2011

Nexera-i Specifications Prominence-i Specifications

A CHEMICAL SENSOR BASED SCHEME TO EVALUATE PACKAGING CONTAINMENT EFFECTIVENESS PART 1: DEVELOPMENT AND OPTIMIZATION

Analyze IT. Process Gas Chromatograph. PGC2000 Series

Recent Improvements in Process GC Instrumentation

Series 8900 GC-PID Gas Chromatograph Installation and User s Manual

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL

FS 231: Final Exam (5-6-05) Part A (Closed Book): 60 points

Process Gas Chromatograph

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

Air Server & Multi Channel Samplers (MCS06 & MCS08)

detectors UltiMate 3000 Variable Wavelength Detectors

Technical specifications of FT-IR Spectrometer

An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

HISTORY OF VARIAN CAPILLARY INJECTORS

Monitoring Flammable Vapors and Gases in Industrial Processes

SECTION 2 SAFETY, TOOLS & EQUIPMENT, SHOP PRACTICES UNIT 8 SYSTEM EVACUATION

ACQUITY UPC 2 Photodiode Array Detector Overview and Maintenance Guide

Emission Monitoring System

GG-LEL2 COMBUSTIBLE GAS SENSOR. Installation and Operation Manual

Mechanism of audible sound generation during flash lamp operation. Tetsuya Kitagawa, Mitsuru Ikeuchi, Yukihiro Morimoto, Fumihiko Oda

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. Clarus 590/690 Gas Chromatography Consumables and Supplies Reference Guide

IAQ_100 C Package

Process Gas Chromatograph

GASGUARD NH 3. Ammonia Sensor OPERATING & INSTALLATION MANUAL

Ultraviolet Systems UV-C 50 INSTALLATION INSTRUCTIONS & PRODUCT MANUAL

1. Desorption Tube Conditioning System

GG-NO2 NITROGEN DIOXIDE GAS SENSOR. Installation and Operation Manual

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

CEM Istanbul - Turkey. True trace emission monitoring system

GASGUARD SO2 Sulfur Dioxide Sensor OPERATING & INSTALLATION MANUAL

GG-NH3 AMMONIA GAS SENSOR. Installation and Operation Manual

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Process Gas Chromatograph

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL

UV Lamp and Quartz Sleeve Replacement/Maintenance

Training Fees 4,000 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch.

Compact Portable QEPAS Multi-gas Sensor

ACQUITY UPLC HT Column Heater Instructions

GAS DETECTOR ST400EC PART.NO

USER MANUAL MINITRACER PORTABLE SNIFFER

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Air Pavilion Quiz. Name: Class:

MICHELL GROUP. Advanced Industrial Hygrometry and Process Instrumentation. Company Profile, 2011

Gas Chromatography (GC-342)

IGD.CO.UK

INSTALLATION & MAINTENANCE

SPECTRAL INSTRUMENTATION

ProStar 325 UV-Vis Detector

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

3SP_H2S_50 Package

Series 9000 Total Hydrocarbon Analyzer

Key Performance. CompactGC 4.0

E11 Tests of trace gas concentration multicomponent. monitoring system

Thermo Scientific Dionex UltiMate 3000 Fluorescence Detectors

Agilent GC Sampler 80/120 SPME Option

GAS CHROMATOGRAPHY PURDUE UNIVERSITY INSTRUMENT VAN PROJECT. DEMYSTIFICATION ACTIVITY FOR THE GAS CHROMATOGRAPH (Revised: ) INTRODUCTION

INDIAN INSTITUTE OF SCIENCE EDUCATION AND RESEARCH PUNE

IMPROVING THE QUANTITATIVE REPRODUCIBILITY OF AN ATMOSPHERIC PRESSURE CHEMICAL IONISATION SOURCE FOR GC/MS AND GC MS/MS ANALYSIS

Sieger Sensepoint. Easy to Use Factory set operation Simple to replace Minimal training required

SLT-Series Installation and maintenance instructions

1998 Expedition/Navigator Workshop Manual

1. Table of Contents VisiJet Materials Overview Disclaimer Ispropyl Alcohol (IPA) and Polypropylene Glycol

Instruction Manual E4 ChamberIR Heater

Transcription:

The SRI Reduction Gas Detector is the most sensitive detector available for hydrogen, carbon monoxide, and other reducing gases. Low ppb detection limits are achievable for hydrogen. Detection limits in the ppt range are achievable for carbon monoxide. Overview The RGD comes with a 10-port gas sampling valve and 1mL sample loop for injecting samples. 10-port gas sampling valve in heated oven Exit tube The RGD exit tube is not factory equipped with a mercury vapor filter, since it was designed for trace detection. Depending on the concentration of the reducing gases in the sample, you may want to add a filter for safety. The amount of mercury vapor produced is proportional to the concentration of reducing gases in the sample.

Theory of Operation The SRI RGD detector uses a mercuric oxide (HgO) reaction tube and a mercury lamp in a heated UV detector cell. The reaction tube is heated to 260-300 o C. Located immediately downstream of the reaction tube, the UV detector cell is heated to 170 o C. The UV detector cell is equipped with a mercury lamp and a UV photodiode. When a reducing gas such as carbon monoxide elutes from the GC column, it reacts with the HgO to form gaseous mercury vapor, which is then swept into the UV cell. The gaseous mercury absorbs the UV light from the mercury lamp as it flows through the cell. The change in transmittance is converted by the data system into an absorbance output (1.00 volt per absorbance unit), which is linearly proportional to the amount of reducing gas. RGD reactor tube Quartz wool HgO Frit CO + HgO = CO 2 + Hg (vapor) H 2 + HgO = H 2 O + Hg (vapor) Photo diode detector Photo diode lead to amplifier Detector cell exit tube Optical path with heater clam shell Detector cell RGD jet with HgO reactor tube inside Heater blocks Column Sample flow Mercury lamp Leads to lamp power terminal

50ppb Expected Performance Sample: 100mLs room air in 1000mL tedlar bag = 50ppb The following two chromatograms were produced with an SRI RGD equipped GC. Room air samples were diluted with nitrogen in tedlar bags, then injected by gas sampling valve onto the Molecular Sieve packed column. Results: Component Ret.Time Area HYDROGEN 2.433 100.8250 OXYGEN 3.500 796.4090 25ppb Sample: 50mLs room air in 1000mL tedlar bag = 25ppb Common Analytical Parameters: Valve injection; 1mL sample loop Column: 2-meter Molecular Sieve 13X Carrier: helium at 5mLs/minute RGD gain: HIGH Reactor temperature: 290 o C Detector temperature: 170 o C Valve temperature: 60 o C Results: Component Ret.Time Area HYDROGEN 2.450 55.2755 OXYGEN 3.516 571.6580 This chromatogram shows the noise level of the RGD detector, which is approximately 500µV from peak to peak. 500µV

General Operating Procedure 1. When your RGD GC arrives, the photo diode amplifier leads will have been disconnected for transport. Locate the insulated electrical lead with two protruding wires: 1 black, 1 yellow. These wires plug into the back end of the RGD assembly, where the photo diode is located. The black is the ground wire; plug it onto the upper prong. Plug the yellow wire onto the lower prong. photo diode amplifier leads yellow black 2. Turn ON the main GC power (the switch is located on the left-hand side of the GC). 3. Connect helium or nitrogen carrier gas to the fittings on the left-hand side of the GC. The pressure for the required flow is printed on the right-hand side of the GC. The required flow for the RGD is 5mLs/minute. This flow gives the best response. 4. Turn ON the detector lamp by flipping the switch on the GC s front control panel. Detector heat is factory set at 170 o C, which is hot enough to keep the mercury vapor from condensing. The reactor is factory set at 290 o C (a higher temperature gives greater sensitivity). 5. The mercuric oxide needs time to equilibrate when it is first installed, or when the reactor tube is changed; 24 hours works well. Leave the reactor and detector temperatures on the above settings continuously with the carrier gas flowing for day to day operation. 6. The first analytical run of the day should be discarded, because the purged reactor will generate unusually high area counts which will not be reproducible once the detector equilibrates. The reactor can also be poisoned by very high sample concentrations. Purge time with clean carrier gas is usually the best cure.

Maintenance GC DETECTORS Reaction Tube Packing and Replacement SRI provides you with one replacement reactor tube in your accessories kit. However, you may also remove the mercuric oxide (HgO) and re-pack the reactor tube, for which you will need fresh HgO and quartz wool. HgO powder is available from Sigma-Aldrich: part #203793-2G: www.sigmaaldrich.com, or call: 1-800- 325-5832. Quartz wool is available from Restek: catalog #20999: www.restekcorp.com, or call 1-800-356-1688. To remove the reactor tube, loosen the fitting that secures it to the column and RGD jet, and pull the reactor tube out. Loosen this fitting Remove the quartz wool and HgO with a long sharp object, and dispose of them properly. Wipe any HgO off the sharp object with a disposable wipe or tissue and discard. Re-pack the reactor tube with 1/8 of HgO and add fresh quartz wool. The HgO must cover the face of the frit entirely to force the sample through the HgO and into the UV cell. 2 Graphite ferrule Column Quartz wool 1/8 HgO Frit The graphite ferrule seals the end of the reaction tube around the column. Change the graphite ferrule when you change the reaction tube. WARNING! Always avoid contact with the HgO. Wear protective gloves and eyecovers, and take care not to inhale the powder.