NATIONAL INSTITUTES of HEALTH FUME HOOD TESTING PROTOCOL. For CONSTANT VOLUME FUME HOODS

Similar documents
MD and Its Impact on Lab Facilities

City of Duluth Festival Center 04/25

UNIVERSITY OF ALASKA FAIRBANKS DIVISION 25- INTEGRATED AUTOMATION DESIGN STANDARDS SECTION INTEGRATED CONTROL OF FACILITY LAB FUME HOODS

SECTION PLUMBING/HVAC FINAL TESTING, ADJUSTING AND BALANCING

SECTION AIR DUCT ACCESSORIES

CHEMICAL FUME HOOD PROGRAM

Union County Vocational - Technical Schools Scotch Plains, New Jersey

HVAC Controls Upgrades: Requirement Details ( )

CLARK COUNTY FIRE DEPARTMENT Fire Prevention Bureau

Design & Construction Standards, Revised January

ME 410 MECHANICAL ENGINEERING SYSTEMS LABORATORY MASS & ENERGY BALANCES IN PSYCHROMETRIC PROCESSES EXPERIMENT 3

Temperature Distribution of Rotary Heat Recovery Units

TESTING, ADJUSTING AND BALANCING

SECTION FUNCTIONAL PERFORMANCE TESTING (FPT) PROCEDURES - CONSTRUCTION MANAGER EDITION

NORTHWESTERN UNIVERSITY PROJECT NAME JOB # ISSUED: 03/29/2017

CEE:3371 Principles of Hydraulics and Hydrology Project #2 Flow Measurement with a Weir

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION

HVAC Controls Upgrades: Requirement Details ( )

ICSV14. Cairns Australia 9-12 July, Noise reduction of handheld vacuum cleaners according to geometric optimization of air passages

15900 HVAC INSTRUMENTATION AND CONTROLS

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC

THE EVALUATION OF MULTI-ZONE AIR FLOW PATTERN AND VENTILATION RATES WITH TRACER GAS METHODS IN APARTMENT HOUSE

Impact of heat load distribution and strength on airflow pattern in rooms with exposed chilled beams

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC

D. Reference Standards (Utilize latest editions available):

ME 410 MECHA ICAL E GI EERI G SYSTEMS LABORATORY

High Plume Blowers Suggested Specification Section (Master Format 1996) Suggested Specification Section (Master Format 2004)

Museum Archive Dehumidification in Hot & Humid Climates

SECTION SECURITY TESTING AND COMMISSIONING

Chapter 4 - Preparation of Stormwater Site Plans

Ventilation System Description. September 8, 2000 HETA

SECTION SPECIAL EXHAUST SYSTEMS. A. Laboratory Ventilation and Hazardous Exhaust Systems. 3. Chemical Fume Hoods and Storage Cabinets

CHAPTER 61G15-32 RESPONSIBILITY RULES OF PROFESSIONAL ENGINEERS CONCERNING THE DESIGN OF FIRE PROTECTION SYSTEMS

1.1 DESCRIPTION A. The purpose of this section is to specify Division 23 responsibilities in the commissioning (Cx) process.

SECTION AIR COILS

C. Manufactured double wall chimneys for fuel fired equipment. D. NFPA 54 - National Fuel Gas Code; National Fire Protection Association; 2009.

Appendix A.4.1. Safe Fume Hood User Guide

Secuflow laboratory fume cupboards

D. The work addressed in this section consists of a fire protection system, which may include coordination with one or more of the following:

L a b o r a t o r y E q u i p m e n t C a t a l o g

Secuflow laboratory fume cupboards

AIR DISTRIBUTION. C. Section Building Automation and Control System Guidelines. D. Section Fire Alarm System & Detection Systems

Thermal Comfort with Convective and Radiant Cooling Systems

Modeling of Ceiling Fan Based on Velocity Measurement for CFD Simulation of Airflow in Large Room

ENSC 388: Engineering Thermodynamics and Heat Transfer

Update all extract references to NFPA documents (including NFPA 72) in Chapter 3 and related annex material to the latest editions.

DESIGN AND CONSTRUCTION STANDARDS SPECIFICATION GUIDELINES

SECTION TESTING, ADJUSTING AND BALANCING

A. Hydraulic calculations shall be prepared in accordance with Chapter 23 of NFPA 13 with the following exceptions:

SECTION DIGITAL, ADDRESSABLE FIRE-ALARM SYSTEM

Michigan State University Construction Standards COMMISSIONING OF FIRE ALARM SYSTEMS PAGE

Recommended location under the current MasterFormat 2004 is Section Elevator Door Smoke Containment System.

A. Standards: Except where modified or exceeded by the requirements of this Specification, conform to the following standards:

CONTENTS AUTOMATIC TRANSFER SWITCHES

STACK EFFECT IN LIGHT WELL OF HIGH RISE APARTMENT BUILDING

CLARK COUNTY FIRE DEPARTMENT Fire Prevention Bureau

B. Use UT Austin specifications and equipment schedule format for HVAC equipment where available.

Recommended location under the current MasterFormat 2004 is Section Elevator Door Smoke Containment System.

The Current (2017) State of the Art for Air Leakage in Ductwork

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

FIRE ALARM APPLICATION CITY OF MADISON HEIGHTS

International Fire Code 2006 Requirements for Construction Plan Reviews of Commercial and Multi-Family Structures

Procedure for the Approval of New Fire Detection and Alarm Technologies

research highlight Assessment of Suite Compartmentalization and Depressurization in New High-Rise Residential Buildings

Recommended location under the current MasterFormat 2004 is Section Elevator Door Smoke Containment System.

CHAM Case Study Numerical Simulation of a Air-to-air Cross-flow Heat Exchanger

Linear Heat Detection Engineering Specifications

RAMAPO COLLEGE OF NEW JERSEY CCA0803 PHASE II HOUSING REHABILITATION 2012

Development of the San Ace Airflow Tester - A Measuring Device for System Impedance and Operating Airflow of Equipment

NFPA 72 Code Changes vs 2013

UC DAVIS Campus Design Guide DIVISION 11 - EQUIPMENT

Design & Construction Standards, Revised January

SECTION : FIRE ALARM & DETECTION SYSTEMS

PHYSICAL FACILITIES Consultant s Handbook Division 01 General Special Requirements for Laboratories. Page 1 of 6

COMMERCIAL KITCHEN SUPPRESSION SYSTEMS INFORMATION PACKET

March 2014 AHRI Presents Research Findings at Low-GWP AREP Conference

5.07 Air Replenishment Systems

Air Distribution and Ventilation Effectiveness in a room with Floor/Ceiling Heating and Mixing/Displacement Ventilation

Minimum Standards for Engineers Practicing Fire Protection Engineering in the State of Oklahoma September 14, 2016

AMCA Publication

Development of a Psychrometric Test Chamber. Michael J. Swedish. Associate Professor Mechanical Engineering Department Milwaukee School of Engineering

CHECKLIST FOR PHASE II DRAINAGE REPORT

GUIDE SPECIFICATION ADVANCED RADIANT SYSTEMS KMI GoGas RANGE HIGH EFFICIENCY, HIGH INTENSITY INFRARED HEATERS

ETSI TS V1.2.1 ( )

LABORATORY EQUIPMENT. A. Section Accessibility for People with Disabilities. B. Section Plumbing Basic Materials and Methods

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC

Lab Design shall comply with the UC Lab Safety Design Manual located at: LABORATORY FUME HOODS

S a n F r a n c i s c o F i r e D e p a r t m e n t B u r e a u o f F i r e P r e v e n t i o n & I n v e s t i g a t i o n 1 of 8 P a g e

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

Bold items are particular to the City of Euless

Public Comment No. 23-NFPA [ Section No ] Statement of Problem and Substantiation for Public Comment Related Item

A. Section Common Motor Requirements for HVAC Equipment: Fan motors.

GUIDE SPECIFICATION ADVANCED RADIANT SYSTEMS DU/DUH DUAL OUTPUT RANGE INFRARED TUBE HEATERS

Your competent partner for measuring and services.

Displacement Ventilation in Classrooms

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

Verification Report SolarVenti SV14 Danish Technological Institute Refrigeration and Heat Pump Technology J.no Test no.

Sound Performance Rating of Outdoor Unitary Equipment

Safely Using a Centrifuge Near a Biological Safety Cabinet in Your Cell Culture Laboratory

SECTION DUCTWORK

Transcription:

NATIONAL INSTITUTES of HEALTH FUME HOOD TESTING PROTOCOL For CONSTANT VOLUME FUME HOODS PART I. GENERAL Test identified below was created by Farhad Memarzadeh of the National Institutes of Health in 1997 and further revised by Memarzadeh and Brightbill in 1999. 1.01 DESCRIPTION OF WORK A. The work of this section consists of testing the performance of constant volume chemical fume hoods. B. On Site Testing of Fume Hoods shall be work of this Section C. This section specifies procedures that are common to the scope of Section 11800 Laboratory Hoods 1.02 SUBMITTALS A. TESTING EQUIPMENT AND FACILITIES 1.03 CERTIFICATIONS 1. Submit 3 copies of specification sheets on all equipment proposed for on site testing specified in Part III of this section. 2. Submit 3 copies of sample test reports for approval. Contractor shall demonstrate the ability to perform necessary calculations on site the day of the test for the on site testing requirements. 3. Submit 3 copies of layouts of the testing facility A. Submit test data attesting that each type of hood to be provided has been tested in accordance with the Modified ANSI/ASHRAE 110-1999 and meets the requirements of this specification. 1.04 QUALITY ASSURANCE A. Testing agency for On Site testing shall be independent of the fume hood control system manufacturer. B. Independent testing agency shall be approved by the NIH project officer. Testing agency shall submit Statement of Qualifications demonstrating experience relating to fume hood testing. As a minimum testing agency shall: TESTING CONSTANT VOLUME FUME HOODS By; F. Memarzadeh Page 1 of 5

1. Have a registered Professional Engineer or Industrial Hygienist executing the testing 2. Demonstrate prior execution of indicated tests and submit a sample of the test report 1.05 REFERENCE DOCUMENTS A. American National Standard Institute B. ANSI/ASHRAE 110-1999 Standard for testing performance of Fume hoods C. National Fire Protection Association NFPA 45 - Fire Protecting for Laboratories Using Chemicals D. ANSI/ASHRAE III Testing and Balancing E. ANSI/AIHA Z9.5 Laboratory Ventilation Standard PART II. PRODUCTS 2.01 TEST AND MEASUREMENT EQUIPMENT A. Anemometers: 1. Accuracy: ±5% of reading 2. Internal Time Constant: <= 100ms B. Tracer Gas Ejector in accordance with ANSI/ASHRAE 110 C. Tracer Gas (SF 6 ) Sensor: 1. Sensitivity: 0.01 to 100 ppm 2. Accuracy: a) Above 0.1 ppm: ±10% of reading b) Below 0.1 ppm: ±25% of reading D. Data Acquisition System: minimum 6 channel system capable so simultaneous sampling at 10 hz or greater PART III. EXECUTION 3.01 FUME HOOD CONTAINMENT TESTING (ON SITE) A. General: Laboratory areas and Constant volume fume hoods shall be tested as installed to assess the level of containment. Testing shall be conducted as outlined below for 50% of the hoods provided in the project. B. Testing shall be conducted in accordance with ASHRAE 110 - Method of Testing Performance of Laboratory Fume Hoods with the following modifications. This is primarily a test of the hood and laboratory configuration. 1. Hoods will be tested with simulated apparatus. This apparatus will consist of: two each 3.8 L round paint cans, one 300mm by 300mm by 300mm cardboard box, three TESTING CONSTANT VOLUME FUME HOODS By; F. Memarzadeh Page 2 of 5

each 150mm by 150mm by 300mm cardboard boxes. These items will be positioned from 150mm to 250mm behind the sash, randomly distributed, and supported off the work surface by 50mm by 50mm blocks. 2. The test gas will have a 6 LPM flow rate. 3. The test will be conducted at the center position for the manikin only. 4. Each test duration will be 5 minutes. 5. Acceptable test results will be 0.05 PPM or better. 6. At the conclusion of each 5-minute test there will be three rapid walk-by at 300mm behind the manikin. Each walk-by will be spaced 30 seconds apart. If there is a rise in test gas concentration, it cannot exceed 0.10 ppm and must return to 0.05 ppm within 15 seconds. 7. There will be a minimum of three and a maximum of five people in the test room during the test procedure. 8. Representatives of the NIH will witness the tests. C. Face velocity testing shall be conducted to determine flow rates and turbulence at the face of the hood. 1. Face Velocity Parameters shall include: a) Measured Face Velocity (FVm expressed in m/s): Face velocity measured in the plane of the sash at three locations at any point in time. Samples for each sensor shall be recorded simultaneously at no less than 10 Hz. The sensors shall basically be point sensors located in the middle of 1 by 1 grid sectors. Move the three sensors to other sectors for subsequent testing periods. Averages shall be calculated for any point in time to assess overall measured face velocity, however individual sensor samples shall be used in calculating TI for each sensor. These face velocity measurements shall be made with sash(es) open and closed. Open shall mean the sashes positioned to their maximum design position. (To sash stop typically 18 to 22 ). Closed shall mean open to 6 for a vertical rising only sash and center panels open 6 on the horizontally sliding panels with the vertically rising sash closed. b) Steady State Face Velocity (SSFV): The average of all sampled face velocities for a 5 second period. Two SSFVs will be determined for both measured face velocity and calculated face velocity; one before the event (SSFVb) and one after (SSFVa). The SSFVa will start two seconds after the end of TSS. The second suffix of m for measured and c for calculated shall be used to indicated the type of assessment c) Turbulence Intensity (TI expressed in m/s): Calculated root mean square of the fluctuating face velocity determined using FVm, calculated as follows: sqrt[sum 1-n{(FVm 1 -SSFV)^2+...+(FVm n -SSFV)^2}/n] This value shall be calculated for each of the steady state conditions preceding and following each event. This shall be correlated to a "Box Leakage Factor" using the following graph of the installation using the Methodology for Optimization of Laboratory Fume Hood Containment" (MOLHC) by NIH Office of the Director, Farhad Memarzadeh principal investigator. While this value does not TESTING CONSTANT VOLUME FUME HOODS By; F. Memarzadeh Page 3 of 5

have a pass/fail requirement, it is the fundamental indicator of containment and therefore shall be clearly reported. Projected Sash Leakage Sash Leakage Factor (%) 0.01 0.009 0.008 0.007 0.006 0.005 0.004 0.003 0.002 0.001 0 0 0.05 0.1 0.15 0.2 0.25 Sash Turbulence Intensity (m/s) D. Hood Static Pressure: Take traverse readings to measure exhaust rate and measure the hood static pressure two straight-line duct diameters downstream from the point of connection between the hood and the exhaust line. The readings shall be taken with a face velocity of.51 m/s ±.05 m/s at the full open sash position. (Open sash typically is 18 to 22 ). TESTING CONSTANT VOLUME FUME HOODS By; F. Memarzadeh Page 4 of 5

E. Performance Parameter Requirements 1. Tracer Gas Leakage per ANSI/ASHRAE 110: 4.00 AM 0.05 2. Average Face Velocity a) Sash(es) open:.51 m/s ±.05m/s (Open to sash stop typically 18 to 22 ). b) Sash(es) Closed: <1.53 m/s ±.15 m/s (exhaust rate shall not change) 3. SSFV Deviation Across Face: <15% of average face velocity 4. Pressure Drop Through the Hood: <= 13 mm water gage F. Test Execution: Testing agency shall be equipped to execute the testing and assess all performance parameters on site the day of the test. Data acquisition of required parameters shall be simultaneous. G. Test Documentation: All testing, calculated, and recorded parameters shall be presented in a report that shows the recorded parameters graphically and tabulates and summarizes all the results. Performance of the hood, the hood controls, and the laboratory in general shall be described and summarized. TESTING CONSTANT VOLUME FUME HOODS By; F. Memarzadeh Page 5 of 5