CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Similar documents
CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

CLEANING INSTRUCTIONS

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

Aesculap Implant Systems

Instructions for Use Spinal Punches IFU

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Aesculap Implant Systems

Cleaning & Sterilization

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

RECOMMENDATIONS FOR CLEANING DEVICES

STERITITE INSTRUCTIONS FOR USE

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

FLASH PAK STERILIZATION CONTAINER SYSTEM

Operating / User Manual

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Recommendations for cleaning, decontamination and sterilisation. devices

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

I.F.U Re/Processing Reusable Medical Devices

2015 Myco Industries, Inc Fax:

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY

Instructions for Use: Fiberscope and LED Light Source

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

When you sterilize an instrument set, you know what the standards are.

RAP-hip Hip stem extraction set

Reprocessing Instructions for Reusable Surgical Instruments English

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG

Cleaning and Sterilization for Instruments & Posts

Important information

Cleaning and Sterilisation for Burs, accessories and filling material

Processing Instructions. Cleaning so easy.

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Reusable Laryngoscope Blades and Handles

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

THE BASICS OF DECONTAMINATION

SteriTite - The Container System of Choice SteriTite

SteriTite - The Container System of Choice SteriTite

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

01694-REVC_081315_ENGLISH

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

ComPact Air Drive II USER S MANUAL

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Classification of Instruments -

Safety. Rinse Kit for Multi-Pro 1200 and 1250 Turf Sprayers Model No Safety and Instructional Decals. Installation Instructions

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Additionally, consider the legal provisions valid for your country as well as to the hygienic instructions of the doctor s practice or hospital.

Processing instructions for the Ligamys suture forceps

Instructions for Cleaning, Sterilization, Inspection and Maintenance. of Trauma & Extremities Medical Devices

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

Stainless Steel Pump for use with food containers

Stainless Steel Pump. Thank You MODELS INCLUDE: CP-200 CP-F FP FP-V FP-200 FP-200V FP-1½ FP-DI FPV-DI TP-V TP-200V

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

The GAK-800 Heated Exercise Gas Analyzer Accessory Kit

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT

Nilfisk Inc Winnetka Avenue North Minneapolis, MN REV.03( ) VF80189

SteriTite - The Container System of Choice SteriTite Instructions for Use

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

STERIS Instrument Cleaning Chemistries

User s Manual. Small Electric Drive. Compact and powerful system for use in hand and foot surgery.

Elaborating of a Reprocessing Instruction According to ISO Klaus Roth

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope,

idrive Ultra Powered Stapling System

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Reprocessing Instructions for Reusable Surgical Instruments English

Syrup Server. Thank You MODEL: SY MODEL: SY for purchasing this Syrup Server to heat and neatly dispense sticky syrups and sauces.

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

PERFORM Operating Document

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

Stainless Steel Pump to dispense portions up to 2oz MODELS INCLUDE:

EL400 OPERATION MANUAL

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED

Air-driven Universal Power Tool System for Traumatology, Endoprosthetics, and Spine. Compact Air Drive II. Instructions for Use

PINNACLE MONITOR FOR AUTOMATED ENZYMATIC CLEANING (AEC) PROCESS

Instrument Cleaning Chemistries Automated Washing Process

Washer Test Kits User Guide

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

OWNER S MANUAL AND INSTALLATION GUIDE PLEASE READ THIS MANUAL CAREFULLY BEFORE ATTEMPTING INSTALLATION

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Instructions for Use and Reprocessing Instructions

Thermal Steel Care & Maintenance Guide

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

Device Reprocessing Checklist Scoring Matrix 1=low 5=high

3/4 HP Model Trimmer Operating Instructions

Transcription:

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling edges. Anytime instruments are to be cleaned (either contaminated or potentially contaminated) Personal Protective Equipment should be used. 1.2. Instruments should remain moistened prior to cleaning, do not allow instruments to dry. 1.3. Do not use metal brushes or scouring pads during manual cleaning process. 1.4. Use cleaning agents with low foaming surfactants for manual cleaning to enable visualization of the instruments in the cleaning solution. The cleaning agents selected must be easily rinsed from the instrument. 1.5. Enzymatic-neutral ph cleaning solutions are recommended for cleaning reusable instruments. 1.6. Some of the Conventus instruments are aluminum with an anodized surface. These instruments must not come into contact with strong acidic or alkaline cleaners and disinfectants, or solutions that contain iodine or chlorine. 1.7. For exposed features (e.g., site preparation instrument blades), saturate the crevices with copious amounts of cleaning solution to flush out any soil. Scrub the surface with a soft bristled brush to remove all visible soil from the surface and crevices. Rotate the instruments while scrubbing to ensure all surfaces and crevices are clean. 1.8. Mineral oil or silicone lubricants should not be used on instruments. 2. INSTRUMENT PREPARATION Instrument trays/lids and index boxes must be cleaned separately from instruments. Non-sterile, single use implants (e.g., plates, screws, and washers) may remain in their respective caddy for reprocessing. Screw caddy lids should be closed and secure during automated cleaning. Remove major contaminants in the operating room before returning the instruments to the tray. It is preferable to remove contaminants using a dry method. If contaminants are removed using a wet method, place the instruments in a prepared solution directly after they are used. The instruments must be disassembled and open as much as possible. The entire product (including grooves, holes, lumens, etc.) must be sufficiently flushed and covered with solution. 3. INSTRUMENT CLEANING Remove each instrument from the Instrument Set. Disconnect any mating parts (e.g., drill/driver bits). Remove any pins or wires that may be inside cannulated instruments. Do not remove or tighten any screws or knobs. Instruments with specific instructions beyond the general cleaning is noted in the illustrations below. Clean each instrument after each use according to the following instructions: 1. Rinse each instrument under running water for two (2) minutes to remove visible soil. Scrub each instrument, including any cannulation, with an appropriately sized soft-bristled brush while rinsing. If applicable, actuate any moving parts (e.g., threads, knobs or pivot points) and scrub with a soft-bristled brush while rinsing. 2. For cannulated instruments, flush the internal lumens by connecting a pressurized water line to an open port until water flows freely out the opposite end. Once water flows freely, flush for an additional two (2) minutes. 3. Prepare a fresh enzymatic-neutral ph detergent (such as Enzol Enzymatic Detergent) solution per the detergent manufacturer s recommendations and place the solution in an ultrasonic cleaner. 4. Immerse the instrument in the detergent solution and sonicate for twenty (20) minutes. Where applicable, leave instruments in an open configuration. 5. Scrub instrument, including any cannulation, with a soft-bristled brush to remove any remaining debris if necessary. If applicable, actuate any moving parts (e.g., threads, knobs or pivot points) and scrub with a soft-bristled brush while rinsing. 6. Rinse the instrument again with distilled or reverse osmosis water for two (2) minutes. If applicable, actuate any moving parts (e.g., threads, knobs or pivot points) and flush all surfaces during rinse. 6883 Rev 13 Page 2 of 8

7. For cannulated instruments, flush the internal lumens again with distilled or reverse osmosis water by connecting a pressurized water line to an open port until water flows freely out the opposite end. Once water flows freely, flush for an additional two (2) minutes. 8. Dry the instrument with a clean, soft lint free cloth or clean compressed air. 9. Automated washing may be used after manual washing is complete. 10. Examine the instrument under normal lighting for visible soil. If present, repeat cleaning. 11. Visually inspect under normal lighting for corrosion, damage, and function. Discard if corroded, damaged, or does not function as intended. 12. Movable parts, such as threads and hinges, should be lubricated prior to sterilization with a non-silicone based surgical instrument lubricant as needed to ensure smooth operation. Unique instructions are provide in the illustrations below. Closed Configuration Open Configuration Leave in this position for cleaning and sterilization Do not tighten knob, leave small gap here Site Preparation Instrument Shown in Open and Closed Configuration Lubricate Lubricate Delivery Handle 4. AUTOMATED CLEANING/DISINFECTION INSTRUCTIONS Automated washing systems are not recommended as the only cleaning method for these surgical instruments. As previously described, an automated cleaning system may be used after manual cleaning is complete. All surgical instruments and trays should be thoroughly inspected prior to sterilization to ensure the cleaning was effective. 5. STERILIZATION INSTRUCTIONS The tray and instruments should be inspected to ensure they were thoroughly cleaned prior to sterilization. If any soil is present, they should be re-cleaned. As the manufacturer, Conventus recommends sterilizing in the provided instrument tray wrapped in standard FDA cleared medical grade steam sterilization wrap or within an Aesculap rigid sterilization container, Model (JK446). Do not stack trays during sterilization. The following sterilization parameters have been validated by Conventus. Cycle Minimum Minimum Minimum Temperature Exposure Time Drying Time Pre-vacuum Autoclave 132 C 4 Minutes 30 Minutes Pre-vacuum Autoclave 135 C 3 Minutes 30 Minutes 6883 Rev 13 Page 3 of 8

Other configurations and sterilization parameters may also be suitable, but the user must validate any deviation from these instructions provided by Conventus. 6. INSTRUMENT SET LAYOUT The main set consists of multiple tray levels, driver index, wire index, plate caddy, and screw caddy. Each bracket is designed to hold a specific instrument denoted by the item s model number in the illustration below. For items that do not have a dedicated bracket, place on the pin mat in the bottom tray. Place instruments in set according to the following layouts: TOP TRAY MIDDLE TRAY LID SCREW CADDY BOTTOM TRAY CASE WIRE INDEX DRIVER INDEX PLATE CADDY 6883 Rev 13 Page 4 of 8

TOP TRAY 5700-1 8mm Drill 4662-1 Removal Ext 6290-1 (Qty 12) 14.5 K-Wire 894-1 Screw Forceps 4601-1 Plunger 4511-1 Delivery Handle 1078-2 Depth Gage 4975-1 Rotation 5127-1 Depth Gage 6345-1 Removal Connector 6935-1 Knob 6931-1 (Qty 3) Handle SP-200 Site Prep MIDDLE TRAY 4576-1 K-Wire TP 6315-1 8mm TP 4499-1 Drill TP 5019-1 Reduction Jig 5542-1 Ratchet Driver 5347-1/6650-1 Distal Screw Jig 5710-4 Pointer, S 5710-5 Pointer, M 5710-6 Pointer, L 4964-1 Frag Screw Jig 6883 Rev 13 Page 5 of 8

BOTTOM TRAY 6532-1 Wire Bushing 6537-1 25 Bushing 6537-2 30 Bushing 6537-3 35 Bushing 7018-1/7019-1 Jig/Center 7020-1 Knob 7017-1/7021-1/6798-1 Jig Body/Handle/Screw 6763-1 Perc Bushing 6764-1/6765-1 Obturators Under 7450-1 6461-1 Wire Bushing 6456-1 8mm Obturator 7450-1 Distal Drill Guide 6464-1 Drill Bushing 6490-1 Access Jig 6463-1 Screw Bushing 6528-1 Access Jig WIRE INDEX 7062-3 7062-4 6497-3 6497-4 6536-1 6536-2 1169-2 (Qty 12) 1169-5 (Qty 12) 3294-1 (Qty 12) DRIVER INDEX 6781-1 3663-1 5707-1 6709-1 4754-1 7036-1 7036-2 4866-1 6883 Rev 13 Page 6 of 8

PLATE CADDY SCREW CADDY CH35XX (Qty 3/Size) 3.5mm Cannulated Headed Screws 6038-1 (Qty 6) Washers CL35XX (Qty 2/Size) 3.5mm Cannulated Locking Screws 6486-1 Large Plate 7276-1 Jig Screws 6883 Rev 13 Page 7 of 8

10200 73 rd Avenue North, Suite 122 Maple Grove, MN 55369 Customer Service: 855-41UNION (418-6466) Fax: 763-315-4980 www.conventusortho.com 6883 Rev 13 Page 8 of 8