Preparing the 6850 Series II GC for 597x MSD Installation

Similar documents
TCD-to-FID Series Connection on a 6820 GC, Accessory G4319A

Thermal Conductivity Detector on a 6820 GC, Accessory G4313A

Installing a Fast Oven Upgrade Kit

Agilent 7890 Series Gas Chromatograph

Keysight 85025/85026 Detector Exchange Program - All Serial Numbers

Agilent Electronic Crimping and Decapping Tools Operating Guide. Agilent Technologies

Replacing a Gun Thermostat

Detailed Cleaning Instructions

Installation. Leveling

Installation Instructions

Dryer. User manual DV22K6800** DV22K A-00_EN (US)_ indd :15:41

Advanced Test Equipment Rentals ATEC (2832) Installation Note

Heater with Air Conditioning. E-Series Ford Aeromaster

Installation Instructions

Retrofit Instructions CLEARVIEW DAY COVER CAUTION: Please read this manual completely before attempting to install, operate or service this equipment

Installation Instructions

OPERATOR S MANUAL. BASIC PROJECT CENTERS Some instructions and drawings may not apply to your specific unit. SAFETY SERVICE PARTS CAPACITIES

Heat Exchanger Block Replacement Instructions

Installation and Operation Manual CLEARVIEW DAY COVER CAUTION: To view a video scan the QR code above

OR Replacement Instructions Heat Exchanger Assembly, ASME

Agilent 7696A Sample Prep WorkBench

Installation Instructions

Installation Instructions

Heating and ventilation system,

INSTALLATION. Glass Panel Doors (select models) CAUTION

Installation Instructions

Installer's Guide. WARNING: HAZARDOUS VOLTAGE - DISCONNECT POWER and DISCHARGE YCC-IG EB23D1-12

INSTALLATION INSTRUCTIONS TXV Coils for Manufactured Housing EMA

Installation Instructions

5800 Refrigerator Compressor Capacitor/Relay Kit

Oreck Edge - Upright Tune-Up & Service Guide 02/26/2010

Installation Instructions

Menumaster. Commercial Microwave Oven Owner's Manual GCSAND7RR

TECHNICAL INSTRUCTIONS

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

Thermal Stripper, TSAB-40 User Manual

Installation Instructions

ACQUITY UPLC HT Column Heater Instructions

DM60-xxx Antennas. 1. Assembly. Installation Instruction for CMAX Antennas

WARNING Risk of fire and electrical shock. DISCONNECT POWER AT ELECTRICAL PANEL before installing or servicing this product.

Service Manual for X9.5 and Older Models

Installation Instructions

Thermal Stripper, TSFB-125 User Manual

ISLAND MOUNT RANGE HOOD. This manual is made with 100 % recycled paper. Electronic version of this manual is available at:

Automatic Dryers. Installation. Keep These Instructions for Future Reference. (If this machine changes ownership, this manual must accompany machine.

569, 570, 571, 572 Series

CryoFit. Handling of Flow Cell and Capillaries. User Guide. Version 002

Installation Instructions

Cooktop Chimney Ventilation Hoods

PUREPOWER SERIES CENTRAL VACUUM POWER UNITS PP500, PP600 & PP650

Wide-Bore Heat Exchanger Installation Kit ( )

Standard Operating Procedure Gas Chromatograph Mass Spectrometry (GC/MS) - Agilent 7890B GC & 240 Ion Trap MS

WALL MOUNT RANGE HOOD. This manual is made with 100 % recycled paper. Electronic version of this manual is available at:

Installation Instructions Part No , Part No Part No

1320 Detectors. 1 of 30. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Dishwasher Installation Instructions DW 24XT/DW 24XV

Installation Instructions

GC DETECTORS Reduction Gas Detector - RGD. Overview

PME. Series Enclosures. Battery Enclosure (BE) Field Upgrade Installation Instructions. Effective: January, Alpha Technologies

TECHNICAL INSTRUCTIONS

Installation Instructions

INTRODUCTION. NOTE: Read the entire instruction manual before starting the installation. FIRE, EXPLOSION, ELECTRICAL SHOCK HAZARD

Heater for BRP Commander. Caution: Before using this product, read this manual and follow all safety instructions. Owner s Manual

Page 1 of 18. Part# /5/2013

hp Dust Collector With Vacuum Attachment

HWX GAS BOILERS USER S INFORMATION MANUAL

D1/D2 Direct Vent Pilot Replacement Model #: D INSTALLER: Leave these instructions with consumer. CONSUMER: Retain for future reference.

Mazda North American Operations Irvine, CA

Manual No: Revision: D. Sump Sensor. Installation Guide

Installation Instructions for Med Touch Master Alarms - Wireless

Chromalox. Installation, Operation. Portable Industrial Unit Blower Air Heater Types SDRA and SDRA-RG RENEWAL PARTS IDENTIFICATION

WARNING: ELECTRICAL GROUNDING INSTRUCTIONS

Installation Instructions

STOP! Read this notice. STOP! Read before proceeding. These instructions are for gas conversion of the Evergreen boiler only.

Installation Instructions

Installation Instructions Built-In Dishwasher

569, 570, 571, 572 Series

Installation Instructions. For the 18 Built-In Dishwasher and Front Color Panels

Manual No: Revision: E. Groundwater Sensor. Installation Guide

Veritas Appliance Hardware Service Procedure

BQ-260/260L. Important Information BOOK BINDER

HG 675 CX 60 HG 675 CN 60 HG 675 CW 60

AD XT. Air Dryer Service Manual. Innovation with Integrity. Version 003 NMR

ADD FLUID 300 NDNL 300 CT 300 OC 700 NDNL 725 NDNL 950 NDNL

WIDE AREA VACUUM MODEL PF2030 WARNING: OPERATOR MUST READ AND UNDERSTAND THIS MANUAL COMPLETELY BEFORE OPERATING THIS EQUIPMENT.

CONSTANT SUCTION UNIT

Installation Instructions

DRAIN WATER TEMPERING INSTALL

22 The Flame Ionization Detector

ENGLISH (Original instructions) INSTRUCTION MANUAL. Power Mixer UT2204 DOUBLE INSULATION. IMPORTANT: Read Before Using.

Installation Instructions

Veritas Appliance Hardware Service Procedure

OPERATOR S MANUAL STANDARD BALL BEARING CHESTS & ROLLER CABINETS

Installation Instructions

Installation Instructions

Installation Instructions

INSTALLATION INSTRUCTION

Installation Instructions

Gas Fired Humidifier. SKG Series Combustion Field Adjustment Instructions. SKG Combustion Field Adjustment Instructions

Installation Electric Dryers Instructions 01

Transcription:

Preparing the 6850 Series II GC for 597x MSD Installation Agilent 6850 Series II Network GC System This installation sheet documents the procedures needed to prepare the 6850 Series II GC for 597x MSD installation. This procedure is intended for Agilent-trained service personnel only. NOTE This kit is not compatible with GCs configured for cryo cooling. This kit contains: Description Quantity 250 µm Graphite ferrule 1 MSD heated transfer line cable 1 MS Ship kit * 1 MS Detector plate 1 Cable ties 2 M4 Machine screw, Torx T-20 2 Insulation detector blank 1 5973N-6850A Site Prep Manual 1 * The MS Ship kit contains additional items that will be needed to complete the setup. Agilent Technologies

Required tools Overview T-20 Torx screwdriver Flathead screwdriver 9/32-inch open-end wrench 7-mm nut driver or open-end wrench CAUTION Before starting, review the safety information located at the end of this document. 1 Prepare the GC. 2 Remove the knockout plugs. NOTE There are three possible procedures for step 3. The procedure you perform depends upon the GC s current configuration. Please read these instructions carefully to ensure you perform the appropriate procedure for the given configuration. 3 Prepare the GC for installation of the the heated transfer line cable. 4 Route the heated transfer line cable. 5 Restore the GC. 6 Complete installation to the 597x MSD. 2

Prepare the GC WARNING The first two steps of this procedure are very important. Working on a GC without first switching off and disconnecting power and waiting for the GC to cool can cause possible burns and exposure to dangerous voltages. To prepare the GC: 1 Turn off the GC and disconnect its power cord. 2 Wait for the GC and all heated zones to cool. 3 Turn off gas supplies and disconnect lines to the inlet purge/split vents. 4 Remove the lid top cover. Remove inlet vent tubing Lid top cover, without valve box accessory Remove T-20 Torx screws (8 places) Remove detector cover 3

Remove the knockout plugs WARNING The insulation around the oven is made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend the following safety procedures: ventilate your work area; wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation in a sealed plastic bag; wash your hands with mild soap and cold water after handling the insulation. 1 Locate the two circular caps (one large, one small) on the left side of the GC. The smaller knockout provides access for the MSD heater transfer line wires. The larger knockout provides the MSD with access to the oven. 2 Using a flathead screwdriver, pop out the two circular caps. 3 Carefully remove the insulation from the oven knockout. Make sure that no insulation remains. 4 Clean up and properly dispose of all the insulation that was just removed. Clean up the insulation on both the outside and inside of the GC. Prepare the GC for installation of the the heated transfer line cable Depending upon the current configuration, the customer can choose to install the heated transfer line cable into the detector heater connection (recommended) or the auxillary connection. The customer should choose ONE of the following options based on the GC s current configuration: Detector installed; no auxillary installed Use the auxillary connection. No detector installed; auxillary installed Use the detector connection. Detector installed; auxillary installed Completely remove ONE of the devices. Agilent recommends using the detector connection for the heated transfer cable, but this should be left up to the customer. To use the auxillary connection, remove the valve box and replace any needed inlet parts. To use the detector connection 1 Remove the detector, flow module, and detector electronics board. CAUTION Board components can be damaged by static electricity; use a properly grounded static control wrist strap before gaining access to the detector electronics area and when handling the boards and any of the cables connecting to the boards. 4

1 Install the circular insulation blank into the detector opening in the lid. See the figure below. Install insulation blank 2 Use two 18-mm long M4 screws to install the detector blanking plate over the insulation blank, as shown in the figure below. Detector plate Screws 5

Route the heated transfer line cable 1 Depending on which heated zone you intend to use (detector or auxiliary), plug the connection end of the heated transfer line cable into the appropriate heated zone connector. Auxillary connection Detector flow module Detector connection 2 Route the other end of the heated transfer line cable toward the back of the GC. If there is no detector flow module present, lower the cable down through the top of the lid where the detector flow module would be. Skip to step 3 below. If a detector flow module is present, perform steps a and b below. a b 6850 with split/splitless inlet and µecd shown Loosen the screws on the detector flow module and move it forward far enough to lower the heated transfer line cable through the top of the lid. Replace the detector flow module back to its original position and retighten the screws. Be careful not to pinch any cable or wiring. WARNING When routing the heated transfer line cable down through the top of the lid, be careful around the sharp edges of the opening. The sharp edges may cause damage to the heated transfer line cable, or worse, injury to the user. 6

3 Open the service lid. Locate the counterbalance cam in the left rear corner under the lid. Loosen the screw on the right side of the cam. This allows the stop plate to drop down. Lid in service position Stop plate (raised) Counterbalance cam 4 Raise the lid until it is stopped by the safety cable. 5 Raise the stop plate and tighten the screw to lock the lid in the upright service position. WARNING The lid is heavy. Always lock the lid when it is in the service position. 6 Remove the existing column (if present). 7

CAUTION 7 Loosen the two bottom screws on the detector electronics cover. Board components can be damaged by static electricity; use a properly grounded static control wrist strap before gaining access to the detector electronics area and when handling the boards and any of the cables connecting to the boards. Remove screw Detector electronics cover Loosen screws 8 Remove the top screw on the cover, then lift and remove the detector electronics cover. 9 Locate the heated transfer line cable that was sent down through the top cover and route it with the main wiring harness. Use cable ties to secure cable to wiring harness, if necessary. Be sure to position the cable so that it will not get caught in the cam or lid when opening and closing the lid. 8

10 Remove the sheet metal electronics cover. Loosen the two nuts and remove the two screws. Remove screws Loosen nuts 11 Route the heated transfer line cable over the oven exhaust duct and out through the small knockout located on the left side of the GC. CAUTION When routing the heated transfer line cable, make sure that the cable is free of snags and that it is not resting on any heated components. Restore the GC 12 Reinstall the sheet metal electronics cover and detector electronics cover. 1 Lower the stop plate on the lid cam. Pull the lid forward until the cam follower rests on the curved surface of the cam. Raise the stop plate behind the cam follower and tighten the screw. 2 Close the lid. Be sure that the heated transfer line cable and the folded gas lines do not interfere with it. 3 Install the lid top cover. 9

Upon completion of these steps, the GC should look like the one pictured below. Lid in closed position Exposed knockouts Heated transfer line cable Complete installation to the 597x MSD The 6850 Series II GC is now ready for installation to the 597x MSD. Follow the instructions found in G2570A 6850 GC/MSD System Setup and Installation, part number G3170-90001. 10

11

Agilent Technologies Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. G3352-90010 First edition, July 2005 Printed in USA Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.