SAMPLE MANUAL - INSTALLATION - FOR REFERENCE ONLY. ChemScan mini Analyzer. Parameter. Method #### Installation, Operation & Maintenance Manual

Similar documents
ChemScan mini Analyzer for Low Ammonia Monitoring Installation, Operation and Maintenance Manual

ph COOLING TOWER CONTROLLER

Installation & Operation Manual. AromaFlo. mr.steam. Feel Good Inc.

M770 ph Controller Owner s Manual

Series 5000 Hydrazine Analyzer Model 60002

CL17 Chlorine Analyzer

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

ORP/MILLIVOLT CONTROLLER

Duct and Rough Service Carbon Monoxide Sensor

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

ECO1ZN3P, ECO2ZN4P ECO

INSTALLATION & OPERATION MANUAL

D3000. Installation & Setup Guide

Beacon 800 Gas Monitor Operator s Manual

LIQUIDATOR Electronic Ionization System

Owner s Manual RD432-0 Chemical Controller

CONTENTS CONSIDERATIONS. General Plumbing Component Connection ILLUSTRATION. Control System CONFIGURATION. Voltage Verification CONNECTION

Date SECTION MEASUREMENT AND CONTROL INSTRUMENTATION Project Name Page 1

AquaLynx 200 ADX. Monitoring and Control System. Background. Operation

Pioneer-R16 Gas Monitor Operator s Manual

Instruction Manual PCS-400-XY. Pump Controller. Y=2: 4-20 ma Y=6: 0-10 VDC X=5: 1-5 VDC. 4 Pump Control Relays 1 Alarm Relay 1 Signal Fail

Ultraviolet Monitoring System. Part Numbers UV/MS-1 V3 UV/MS-2 V3. Installation And Operations Manual

Rev B, 9/2/2009. Kodiak Chiller Overview

DIGITAL METERS Large Display Temperature Meters Instruction Manual

INSTALLATION GUIDE AND OWNER S MANUAL. AccuMix II ELECTRIC INSTANTANEOUS WATER HEATER WITH ASSE 1070 APPROVED MIXING VALVE

PZ1, PZ3 & PZ6 Installation Guide

EASY-WAVE Trash & Linen Chute System Installation

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

CLARITY II TURBIDIMETER

PZ2 Series Manual. PROZONE INTERNATIONAL, Inc. Ozone Water Purification Systems. INSTALLATION GUIDE and OPERATION MANUAL. Prozone PZ2 Series

Installation & Operation Manual Models:

Power Supply Display Module PSDM. Installation and Operation Manual

Analog Room Pressure Monitor RPC Series

Control Panel For Water Wash Hoods AM2

EXPERTS IN WATER CHEMISTRY SINCE Iron Analyzer User Manual. Revision 3.04

Carbon Monoxide Transmitter

INSTALLATION GUIDE AND OWNER S MANUAL

Installation. Built-in Full Height Wine Cellar VCWB301

AS950 REFRIGERATED SAMPLERS

Installation, Operation and Maintenance

Ozone Generator Systems

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

M790 ph/orp Controller. Owner s Manual

Table of Contents. Installation, Operation, and Maintenance Manual V100 Series

OPERATING AND MAINTENANCE MANUAL FOR PLATE HEAT EXCHANGER INDIRECT FIRED WATER HEATER. Electric Heater Company Base Model "BWXP"

Next Generation Corona Discharge Installation & Operation Manual

Installation Instructions

Freeze Protection Thermostat TRACON MODEL GPT 3 Installation and Operation Manual

Internet Version for Reference Only INDUCED DRAFT COMMERCIAL WATER HEATERS SUPPLEMENT INSTRUCTIONS TO PART #

FIL-SPL Flow-through heater for central heating USER MANUAL

DISHWASHER INSTALLATION GUIDE SPECIFICATIONS, INSTALLATION, AND MORE

Viking Installation Guide

INSTALLATION INSTRUCTIONS

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

BSR Addressable Fire Detection control panel. Installation operation manual

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

WASTEWATER Online UV%Transmission Analyzer

Rev Pulse Modulating Anti- Sweat Control (PMAC II) Installation and Operation Manual

SERVICE REQUIREMENTS

Updated: September 12, 2018 DIVISION INTEGRATED AUTOMATION. INTEGRATED AUTOMATION Page 1 of INTEGRATED AUTOMATION

OPTISENS TUR 2000 Technical Datasheet

ENSPECO Recovery/ Recycle/ Evacuate/ Recharge

Installation. S19-320A (6' cable) S19-320D (12' cable) Flow Switch Alarm System. Table of Contents

DISCONTINUED. Installation. S19-320A (6' cable) S19-320D (12' cable) Flow Switch Alarm System. Table of Contents

Table of Contents. Installation, Operation, and Maintenance Manual ELBPxx

MODEL EUR -5A AUTOMATIC SNOW/ICE MELTING SYSTEM CONTROL PANEL

Certificate of Compliance

Improve Performance and Reduce Maintenance Costs. Model Q45D. Dissolved Oxygen Monitor...

ROYCE SUSPENDED SOLIDS SYSTEM MODEL 7011A AND SENSORS MODELS 72A, 72P, 73B, 74A, 75A & 76A

SECTION 1 (Page 2) Introduction, standard accessories and mounting specifications.

for Drinking Water Systems

SECTION Fluorescent Dissolved Oxygen Measuring System

Installation Manual CARBONATOR With Plain-Water Booster

MO n : 12JMC rév A

Installation Instructions If you have questions, call or visit our website at:

Unfired Steam Generators (Series RB) Installation, Operation and Maintenance

Daikin ENVi Thermostat installation overview

Rules for Safe Operation

EXPERT TRI-STAR. Temperature controller. User s Manual

Dual Column Dehydrating Breather Manual

ADC-T2000. Smart Thermostat v1.5

Water Technologies. Strantrol System 3i. Technical Guide

rev3 INSTALLATION & OPERATION MANUAL OIL CIRCULATING HEATING SYSTEM MODEL OSM

AccuBrine automated brine maker Technical Specifications

IF79 CAUTION CONTENTS YOUR THERMOSTAT REPLACES 1 PREPARATIONS. Installation Instructions for. Heating & Air Conditioning

Model Q45P/R. ph/orp Monitor...

Wallace & Tiernan Gas Feed Systems V2000 Chlorinator

Ozone Generator Systems

1. Quick start-up guide

Condensate Monitoring OM0232

INSTALLATION & OPERATION Rev 1.00A DISTRIBUTED EXCLUSIVELY BY CMA DISHMACHINES KNOTT AVENUE GARDEN GROVE, CALIFORNIA 92841

PRE INSTALLATION PROCEDURES

SAFETY MANUAL. X2200 UV, X9800 IR, X5200 UVIR SIL 2 Certified Flame Detectors

INSTALLATION INSTRUCTIONS

OPERATING MANUAL. Reverse Osmosis Equipment Model: Frame Mount Vertical FMV-1 through FMV-6. Made in U.S.A.

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea

SS1095 Ice Maker.

REFRIGERATED CONDIMENT RAIL

Installation and Use Manual

MEAS CAL ALARM WASH CONF APT4000 TC

Transcription:

Parameter Method ####, Operation & Maintenance Manual Rev 3/26/2018

Contact Information Document Information Customer Name: Plant Name/Location: Date: Analyzer Serial #: ASA, Inc. (Applied Spectrometry Associates) 2325 Parklawn Drive, Waukesha WI 53186 PH: 262-717-9500 FX: 262-717-9530 ChemScan mini Analyzer ASAanalytics.com 2 ChemScan mini Analyzer O&M Manual

Table of Contents Table of Contents Technical Specification...6 Analyzer Description...8 Safety...10 1.... 11 1.1 Shipping... 11 1.2 Storage... 11 1.3 Unpacking... 11 1.4 Location...12 1.4.1 Sampling...12 1.4.2 Analyzer...12 1.5 Sample Delivery...13 1.6 Mounting...14 1.6.1 Analyzer...14 1.6.2 Reagent Shelf...14 1.6.3 Cleaning Solution...14 1.7 Electrical...16 1.7.1 Power...16 1.7.2 Communication...16 1.8 Analyzer Connections...18 1.8.1 Cleaning Solution...18 1.8.2 Sample Inlet...18 1.8.3 Sample Drain...18 1.8.4 Reagent(s)...19 2. Startup...21 2.1 Turn on the Analyzer...21 2.2 Verify Sample Flow...22 2.3 Select Parameter/Range...23 2.4 Perform Instrument Zero...24 2.5 Prime Reagent Injector(s)...25 2.6 Start Online Operation...26 2.7 Verify 4-20 ma Operation (optional)...26 2.8 Additional Configuration Variables...26 ChemScan mini Analyzer O&M Manual 3

3. Operation...27 3.1 Online Operation...27 3.1.1 Read Interval...27 3.1.2 Method Description...28 3.1.3 Analysis Cycle...28 3.1.4 Clean Cycle...30 3.2 User Interface...31 3.2.1 Basic Menu Navigation...31 3.2.2 Menu Structure...32 3.2.3 1) Online...33 3.2.4 2) Sample...33 3.2.5 3) Clean...33 3.2.6 4) Setup...34 3.2.7 Factory Default Settings...35 3.2.8 Technician Menu...36 3.3 Additional Functions...39 3.3.1 Internal Data Log...39 3.3.2 Inputs...40 3.3.3 Outputs...40 3.3.4 Testing Outputs and Inputs...41 3.3.5 4-20 ma Communication...42 3.3.6 Alarms...43 3.3.7 External Pump Control...44 4. Maintenance...45 4.1 Maintenance Schedule...45 4.2 Internal Maintenance Preparation...46 4.2.1 Interrupting Online Operation...46 4.2.2 Closing Sample Line Valve...46 4.2.3 Empty the Analyzer...47 4.3 Return to Online Operation...47 4.4 Reagents...48 4.5 Cleaning Solution...51 4.6 Calibration Adjustment...52 4.6.1 Zero-Based Offset Adjustment...52 4.6.2 Mid-Range Offset Adjustment...53 4.6.3 Full Slope and Offset Adjustment...53 Table of Contents 4 ChemScan mini Analyzer O&M Manual

Table of Contents 4.7 Cleaning the Flow-cell...54 4.8 Inlet Strainers...56 4.9 Cleaning/Air Pump Head Replacement...57 4.10 Cleaning/Air Pump Replacement...58 4.11 Sample Valve Replacement...59 4.12 Reagent Injector Replacement...60 4.13 Manually Prime Reagent Injectors... 61 5. Troubleshooting... 63 5.1 Alarm and Maintenance Messages...63 5.2 Diagnostic Tests...64 5.2.1 Sample Volume Test...64 5.2.2 Flow-cell Window Fouling Test...64 5.2.3 Reagent Injector Volume Test...65 5.2.4 Reagent Injector Back Flow Test...66 5.3 Electrical Diagram...67 5.4 Troubleshooting Tables...68 5.4.1 Readings Constantly at Maximum of Range...68 5.4.2 Readings at Zero...68 5.4.3 Inaccurate or Unstable Readings...69 5.4.4 Unstable Instrument Zero...69 5.5 Replaceable Parts... 70 5.5.1 Parts Diagram...70 5.5.2 Parts List...71 ChemScan mini Analyzer O&M Manual 5

Analyzer Description ChemScan mini Analyzer The ChemScan mini Analyzer is a photometric instrument, designed to provide reliable and accurate analysis of water and wastewater. To reduce maintenance requirements, the sample path maintains an internal diameter of 1/8 (3mm) or greater. The analyzer consists of two enclosures. The upper enclosure contains a power supply, main circuit board, control panel and electrical connections. The lower enclosure contains the sample control manifold, reagent injector(s), flow-cell and optical components. Voltage in the lower enclosure is 24 volts. A water-tight cord-grip provides a seal between the two enclosures. To measure the parameter, the analyzer uses an automated process similar to a typical test kit. It performs an initial measurement of the sample water and stores it as a blank. Then a reagent is added to the sample and mixed. This mixture is allowed to react for a period of time. A second optical measurement is made and the previous blank measurement is subtracted. The parameter concentration is calculated and displayed on the LCD screen, and the 4-20mA communication signal is updated. Emitter (Light Source) Sample Detector Analyzer Description Reading Blank Adding Reagent(s) and Mixing Reading Concentration 8 ChemScan mini Analyzer O&M Manual

Analyzer Description Flow-cell & Optrode Sample Drain Sample Inlet Air Pump Power Switch Main Circuit Board Control Panel Reagent Injector(s) Cleaning Pump Manifold Valve Cleaning Solution Reagent(s) Actual Bottle(s) may differ ChemScan mini Analyzer O&M Manual 9

Safety ChemScan mini Analyzer To safely operate or maintain this analyzer, all instructions in this manual must be read and fully understood by a trained/qualified technician. Failure to follow safety procedures could result in serious injury or equipment damage. The following symbols will be observed throughout this manual. UV Protective Eye wear Safety Gloves Electrical Shock Hazard Corrosive Hazard UV Radiation Inside Rayonnement ultraviolet à l intérieur Safety 10 ChemScan mini Analyzer O&M Manual

1. For safety guidance, refer to the safety section at the beginning of this manual. 1.1 Shipping This ChemScan mini Analyzer is shipped in a double-wall, corrugated carton suitable for protecting the analyzer from damage during normal shipping and handling. We recommend retaining this carton to provide optimum protection for return shipping if needed. 1.2 Storage If the analyzer will not be installed immediately, store the analyzer, reagents, and accessories in their original cartons, in a temperature controlled environment, protected from physical and moisture damage. 1.3 Unpacking When unpacking the analyzer, inspect the equipment for any damage that may have occurred in the shipping process. If damage is found, document and report to ASA Analytics. ChemScan mini Analyzer O&M Manual 11

1.4 Location 1.4.1 Sampling ChemScan mini Analyzer For best instrument performance, select a sampling point representative of the process. The sample location should be a sufficient distance away from chemical additions to ensure the process is well mixed and chemical reactions are complete. Sample line should draw from center of process stream to minimize sediment and air bubbles. 1.4.2 Analyzer Install the analyzer indoors on a vertical surface near a 120 VAC outlet. The environment should be non-hazardous with moderate temperature cycles. NOTE: Extreme temperatures may cause degradation of the reagents and cleaning solution. (See Technical Specification) 12 ChemScan mini Analyzer O&M Manual

1.5 Sample Delivery The analyzer requires a pressurized sample between 2 and 10 psi (13.8-69 kpa). Continuous bypass flow is recommended to ensure a fresh and representative sample. NOTE: The analyzer only draws sample during an internal flush cycle. Install a bypass tee near the analyzer to minimize dead volume. If an external pump is required to pressurize the sample, refer to Section 3.3.7 External Pump Control. Option 1: A basic bypass flow system consists of a tee fitting and two manual ball valves. When the analyzer s sample valve opens, the tee will allow sample to flow into the analyzer. The two ball valves are used to control the sample flow rates. A pressure regulator may be required if pressure is too high. Option 2: ASA offers a Sample Extraction Accessory that provides continuous bypass flow and coarse filtering all in one component. This fast loop assembly is mounted on two SS U-Channel rails and is equipped with a pressure gauge. This is recommended if large stringy solids are present, such as algae (<150 mg/l TSS). PRESSURIZED SAMPLE STREAM PRESSURIZED SAMPLE STREAM SAMPLE TO ANALYZER (2-10 PSI) RETURN TO PROCESS RETURN TO PROCESS SAMPLE TO ANALYZER (2-10 PSI) ChemScan mini Analyzer O&M Manual 13

1.6 Mounting 1.6.1 Analyzer ChemScan mini Analyzer The analyzer should be mounted securely to an indoor wall using appropriate fasteners. NOTE: Provide a minimum clearance of 30 [762mm] in front of analyzer operator access. 1.6.2 Reagent Shelf Mount the reagent shelf on the vertical surface adjacent to the analyzer. Keep the reagent shelf mounting holes inside the mounting region shown on the following page. NOTE: The recommended position of reagent racks ensures the tubes will reach the reagents. 1.6.3 Cleaning Solution Provide space for the cleaning solution container on the floor below the analyzer. Component Analyzer (Enclosure) Reagent Rack Cleaning Container WxHxD 9x22.5x6 [229x572x152] 7.5x10.5x6.8 [190x267x173] 7x12.5x11.5 [178x318x292] for 14 ChemScan mini Analyzer O&M Manual

23.8 [605] MOUNTING HOLES 10.2 [260] 0.31 [ 8] MOUNTING REGION 3.5 [89] REAGENT RACK 3 3.5 [89] 6.0 [152] 12.0 [305] 7.75 [197] CLEANING SOLUTION CONTAINER 6.0 [152] 4.5 [114] 7.4 [188] 7.0 [178] 66.0 [1676] MOUNTING HEIGHT ABOVE FLOOR MOUNTING REGION 4.0 [102] REAGENT RACK 1 16.0 [406] NUMBER OF RACKS MAY DIFFER UNITS: Inch [mm] 0.25 [ 6] 4.0 [102] REAGENT RACK 2 ChemScan mini Analyzer O&M Manual 15

1.7 Electrical 1.7.1 Power ChemScan mini Analyzer Power Requirements: 100 to 240 VAC, 50/60 Hz, 70 VA Maximum NOTE: Power switch also serves as a circuit breaker. Option 1: AC Outlet The analyzer is equipped with a 6 foot 120 VAC power cord. Plug the power cord into an appropriate outlet. Provide access to unplug the analyzer. The plug serves as the primary disconnect. NOTE: DO NOT use an extension cord. Option 2: Conduit The power cord can be replaced by approved watertight conduit. Conduit and wiring must be installed by a qualified person according to local and national codes. The primary disconnect must be provided by others. A switch or circuit breaker clearly marked for this purpose should be located adjacent to the analyzer. WARNING: Conduit connected to the fiberglass enclosure will not serve as an electrical ground. Bonding must be provided between the labeled ground terminal and electrical ground. AVERTISSMENT: Le conducteur branché au boîtier en fibre de verre ne sert pas de mise à la terre- mise de masse. Il faut connecter un fil entre la borne de mise à la terre et la borne électrique. 1.7.2 Communication Remove the hole plug and install the 4-20 ma output wires using watertight conduit. Connect the wires according to the diagram on the following page. The analyzer sources current into the fully isolated analog loop. NOTE: A secondary 4-20 ma Auxiliary Channel (AUX) is unused. The dry-contact alarm relays are fused at 5 amps. Usage is limited to 24 VAC. 16 ChemScan mini Analyzer O&M Manual

Communication Serial Port AC Power Cord AC Power Communication See Section 3.3 for details Auxiliary I/O See Section 3.3 for details L N G 1 2 3 1 Alarm 1 Common 2 NC 3 NO 4 Alarm 2 Common 5 NC 6 NO 7 Shield 8 4-20 ma + 9 Return 4-20 ma - 10 AUX 4-20 ma + 1 Out Common 2 On-Line 3 Busy Reading 4 Drawing Sample 5 Maint. Required 6 Spare 7 Spare 8 External Flow 9 External Run/Start 10 Input Common OUTPUTS INPUTS ChemScan mini Analyzer O&M Manual 17

1.8 Analyzer Connections ChemScan mini Analyzer NOTE: Do not over-tighten the plumbing connections. Excessive tightening can result in damage to the fittings or other components. FITTING NOTE: Ensure connections are firm and vacuum-tight. 1.8.1 Cleaning Solution For mixing instructions, refer to Section 4.5 Cleaning Solution. 1. Screw the tube-fitted cap onto the cleaning solution bottle. 2. Attach the red tube to analyzer. 3. Verify the tube is at the bottom of the bottle. 1.8.2 Sample Inlet 1. Mount the tube fitting nut onto 3/8 sample tube. 2. Screw the nut onto the sample inlet assembly 1.8.3 Sample Drain Run drain tube so it drains completely empty following a sample flush. The drain tubing must have a free fall to an open drain. Low spots in drain tubing where liquid completely fills the tubing may cause erroneous readings. DRAIN SAMPLE INLET CLEANING SOLUTION NUT TUBE 0.70" MAX 18 ChemScan mini Analyzer O&M Manual

1.8.4 Reagent(s) For reagent mixing instructions, refer to Section 4.4 Reagents. 1. Place reagent container(s) in the reagent rack. 2. Screw the tube-fitted cap onto the bottle. 3. Verify the reagent tube(s) reaches the bottom of the bottle(s). 4. Verify all of the reagent tube fittings are finger tight in the analyzer s lower enclosure. Note: Refer to the MSDS for safe handling of the reagents. (Actual bottle may differ) ChemScan mini Analyzer O&M Manual 19

*This page is intentionally left blank* ChemScan mini Analyzer 20 ChemScan mini Analyzer O&M Manual