Consumer Letter (Accu-Chek Insight E-7) Urgent field safety notice. Location, Date

Similar documents
(Place Logo from Lab here)

1. User features of the GSM dialer

SAFETY CODES COUNCIL ORDER. BEFORE THE FIRE TECHNICAL COUNCIL On June 21, 2012

Mobeye CM2410 GSM fire alarm communicator

URGENT FIELD SAFETY NOTICE

Alarm Fatigue in the ICU. Lynn Maguire MS, RN, PMC, NE-BC

User manual. Epi-Care mobile. Danish Care Technology ApS Energivej 3, DK-4180 Sorø version 1.15 Phone: Page 1 of 17

USER MANUAL Mobeye ThermoGuard CM2200

PORTAL USER MANUAL. Mobeye WaterGuard-FS. Float sensor CM2300FS. SW version 5.n

CORAL MANTA User Guide

Follett Performance Plus

URGENT - Field Safety Notice Medical Device Correction

MT-S20 USER S MANUAL. (Motor Alarm &Tracker) (Version 1.4)

Wireless Call Point ( )

LETTERS FOR EUROPEAN UNION MEDDEV RT021

Affordable wireless monitoring without compromise

ir-range 8G /03.15 Changes reserved.

Chapter 5 Inspection and maintenance

ZITON RADIO LOOP MODULE

Note: before using the system read this manual carefully, store it and, in case, transfer it to a new user. prima

Affordable wireless monitoring without compromise

OPERATION MANUAL. Daikin Altherma indoor unit EKHVMRD50ABV1 EKHVMRD80ABV1 EKHVMYD50ABV1 EKHVMYD80ABV1

GE Healthcare Healthcare Systems 9900 Innovation Drive Wauwatosa, WI USA

GRADESHIFT UDL INSTRUCTION MANUAL 2.5 APPENDIX

Quick Start Guide. 1 Charging. Function: Keeps the SoloProtect ID battery fully charged.

v.a.c.simplicity TM Quick Reference Guide Power Connection Canister Battery Charging Light Control Panel

LIFETONE HL TM. Bedside Fire Alarm & Clock. User s Manual MODEL HLAC XI. RECOMMENDED LOCATIONS FOR SMOKE ALARMS ENGLISH SPANISH

ALERiTY Wide-Area Mass Notification. Mass notification. Providing industry-leading intelligibility

10 & 12 Ventilator. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference

<< COMPANY LETTERHEAD >>

Safety Instructions. Be sure to observe the following instructions when using the device.

GSM RFID VOICE Alarm System

Forehead Infrared Thermometer

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010

Operating Instructions Version 1.2

THE EDDY IQ METER PRODUCT MANUAL

THE MANUFACTURER RESERVES THE RIGHT TO ALTER THE SPECIFICATIONS OF THE PRODUCT WITHOUT PRIOR NOTIFICATION

User s Operation Instructions & Important Warranty Information

The Leaders in Security and Life Safety Communications

INSTRUCTION MANUAL PC255O

Asmall Primary School

Instructions. Consumer Info-Line

Connected smoke detectors bring improved safety and service for facility managers

WIRELESS ALARM SYSTEM WITH TELEPHONE AUTO DIALER

PUMPS COMPATIBLE WITH ANY TYPE OF WATER PUMPING SYSTEM. Instructions must remain with installation.

Sip Vacuum Cleaner 1400/35 Wet & Dry

ABB Ability System 800xA Alarm Management

Quick set-up guide. [Simple plug and play system]

CO SENSOR CARBON MONOXIDE DETECTOR

XD-125 OWNER S MANUAL COMMERCIAL DEHUMIDIFIER

Alarm System Performance Metrics

Nurse Call Systems & Emergency Call Systems. Overview of ANSI/UL 1069 and ANSI/UL 2560

QUICK SCAN COUNTERFEIT DETECTOR

Operating Guide. English Version 3.3

FIRE ALARM AUDIBILITY IN EXISTING RESIDENTIAL OCCUPANCIES. 1 of 7 11/12/ :26 PM

CQUBE EVO OPERATION MANUAL EVO_OM_VKI_EN. Part No.: _06

PowerMaxComplete. All-in-one Professional Grade Wireless Security

PLEASE READ ALL INSTRUCTIONS PRIOR TO USE

Operating Manual of Safilo-Bio

Packaged Heat Pumps. Owner s Guide to Operating and Maintaining Your Heat Pump

User Operation Guide Document: VI49.2

LevelSET S GSM Functionalities Instructions for Commissioning and Use

Smart Wellness. Yoshio Sekiguchi Sr. General manager, Micro Devices H.Q. OMRON Corporation

OGB Poly Care 2 Incubator Service Manual

24 & 30 Drum Fan. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference

Detectores de armas metals SuperWand GARRETT manual ingles

Alarm Management. Objectives

Applicable To Employees of the Gundersen St. Joseph s Hospital laboratories and Gundersen Tri-County Hospital laboratories.

ONLY WORKING SMOKE ALARMS SAVE LIVES

SYMS II MANUAL SEALER

United States Engineered Quartz (E-Quartz) Market By Type, By End Use-Sector, By Region: Opportunities & Forecast ( )

Technical Bulletin. Concealed extended flues

Alarms Updated 03/26/2018

USER MANUAL AND TECHNICAL MANUAL

English. Meet Your Neato Robot

P163BBA Radio Enuresis Alarm Monitor

PORTABLE AIR CONDITIONER (LOCAL)

IEC Electroacoustic EMERGENCY WARNING SYSTEM. for a maximum of system security 79

Local air conditioner

User s Operation Instructions & Important Warranty Information. High-efficiency wall-hung boilers Baxi Avanta Heat Only 15h - 18h - 24h - 30h

RFT CARES Android App User Guide

IoT use cases for smart building. IoT is changing the way facility management, security and construction companies are operating.

Interactive Technologies Inc North 2nd Street North St. Paul, MN Technical Manuals Online! -

USER S MANUAL FOR HIGH PERFORMANCE HANDHELD METAL DETECTOR MD HH UHS

COMPILATION OF COMMUNITY PROCEDURES ON INSPECTIONS AND EXCHANGE OF INFORMATION

My Visonic SMS App User Guide

Solutions for Smarter Home Security

Alarm System SECURE AS 302

GENERATOR SELF-TEST VERIFIER (GSV300) INSTALLATION INSTRUCTIONS

Computer Room Guard Model VM Manual and Installation Instructions

Product Recall Notification TX-4200 Wireless Personal Panic Device

Operator's Manual. PCP October 2017

Alarm Management Implementation on a Plant. Standards Certification Education & Training Publishing Conferences & Exhibits

Voice Board. Installation and Programming Guide. Runner 4/8,PowerWave 4/8/16 &, Elite64. Add-on Board For Storing Recorded Voice Messages

Datasheet CALYPSO II R. Interlinkable smoke alarm device

INSTRUCTION MANUAL PC255O

SAFETY & SECURITY AUDIT TOOL

Forehead Thermometer Instruction Manual

Improving Alarm Management Utilizing Technology. NJ/DVHIMSS Annual Fall Event. Empowering the New Healthcare Era

TS0 Series Portable Bulb and Capillary

Transcription:

Consumer Letter (Accu-Chek Insight E-7) Urgent field safety notice <Enter address here> Location, Date Important information on the Accu-Chek Insight system to increase awareness for E-7 alarm messages potentially impacting the acoustic and vibration alarms Dear [please personalize], We would like to inform you about feedback we have received from a small number of people using the Accu-Chek Insight insulin pump system. Customers reported that their pump displayed an E-7 error message. This error message could not be resolved by a battery change and a pump self-test as advised as the first step in the manual. In such an instant, the alarm functions of the pump might not be fully available. The Accu-Chek Insight insulin pump system has three built-in alarm functions, an acoustic alarm, a vibration alarm, and visual notifications that you receive on the pump and handheld display. If the above-described issue occurs, the acoustic or vibration alarm functions might potentially be impacted and no alarm signal can be heard or felt. Missing an alarm e.g. while sleeping can potentially lead to an insulin under-delivery which could result in a hyperglycemic event. We have identified the root cause of this issue and have started to implement the appropriate corrections. To secure a continued optimal user experience for you, we would like to draw your attention to the important E-7 error messages and re-emphasize its correct interpretation and handling as it is described in the product manual. Recommended actions: Please ensure that all alarm functions on your pump are always activated In case your pump is dropped or otherwise has been exposed to mechanical impact, shock or damage, please always check your pump s functionality (insulin delivery and alerts) as described in Chapter 9 of your product manual, and as a first step reinsert the battery to initiate a selftest If you are unsure whether the acoustic alarm function of your pump is working properly, you can always provoke an acoustic signal by pressing a button. (Please make sure that the acoustic button feedback is enabled on your pump) As a routine, please check the pump and handheld displays regularly during the day and before you go to sleep for any notifications In case you receive an E-7 error message on your pump, please follow the important instructions provided in the pump manual: Tel.

If you continue to see this message, please call our Accu-Chek Customer Care line at + xx xxxx xxxxx to get your pump replaced. This issue is only affecting the Accu-Chek Insight system. Due to their different design and electronic components all other Accu-Chek pumps are not impacted by this issue. The competent authority, healthcare professionals and distributors have been informed about this field action. Thank you in advance for your understanding and co-operation. If you have any further questions, please do not hesitate to contact our Accu-Chek Customer Care line at xxx-xxx-xxx or your local Accu- Chek sales representative at any time. Kind regards, 2/2

Distributor Letter (Accu-Chek Insight E-7) Urgent field safety notice <Enter address here> Location, Date Important information on the Accu-Chek Insight system to increase awareness for E-7 alarm messages potentially impacting the acoustic and vibration alarms Dear [please personalize], We would like to inform you about feedback we have received from a small number of people using the Accu-Chek Insight insulin pump system. Customers reported that their pump displayed an E-7 error message. This error message could not be resolved by a battery change and a pump self-test as advised as the first step in the manual. In such an instant, the alarm functions of the pump might not be fully available. The Accu-Chek Insight insulin pump system has three built-in alarm functions, an acoustic alarm, a vibration alarm, and visual notifications that you receive on the pump and handheld display. If the above-described issue occurs, the acoustic or vibration alarm functions might potentially be impacted and no alarm signal can be heard or felt. Missing an alarm e.g. while sleeping can potentially lead to an insulin under-delivery which could result in a hyperglycemic event. We have identified the root cause of this issue and have started to implement the appropriate corrections. To secure a continued optimal user experience for your customers, we would like to draw your attention to the important E-7 error messages and re-emphasize its correct interpretation and handling as it is described in the product manual. Recommended actions: Please ask your customers to ensure that all alarm functions on the pump are always activated In case the pump is dropped or otherwise has been exposed to mechanical impact, shock or damage, please ask your customers to always check the pump s functionality (insulin delivery and alerts) as described in Chapter 9 of the product manual, and as a first step reinsert the battery to initiate a self-test If your customer is unsure whether the acoustic alarm function of the pump is working properly, users can always provoke an acoustic signal by pressing a button. (Please ask your customers to make sure that the acoustic button feedback is enabled on the pump) As a routine, please ask your customers to check the pump and handheld displays regularly during the day and before they go to sleep for any notifications In case your customers receive an E-7 error message on the pump, please ask them to follow the important instructions provided in the pump manual: Tel.

If your customers continue to see this message, please instruct them to call our Accu-Chek Customer Care line at + xx xxxx xxxxx to get the pump replaced. This issue is only affecting the Accu-Chek Insight system. Due to their different design and electronic components all other Accu-Chek pumps are not impacted by this issue. The competent authority, users of the Accu-Chek Insight system and healthcare professionals have been informed about this field action. Thank you in advance for your understanding and co-operation. If you have any further questions, please do not hesitate to contact our Accu-Chek Customer Care line at xxx-xxx-xxx or your local Accu- Chek sales representative at any time. Kind regards, 2/2

Healthcare Professional Letter (Accu-Chek Insight E-7) Urgent field safety notice <Enter address here> Location, Date Important information on the Accu-Chek Insight system to increase awareness for E-7 alarm messages potentially impacting the acoustic and vibration alarms Dear [please personalize], We would like to inform you about feedback we have received from a small number of people using the Accu-Chek Insight insulin pump system. Customers reported that their pump displayed an E-7 error message. This error message could not be resolved by a battery change and a pump self-test as advised as the first step in the manual. In such an instant, the alarm functions of the pump might not be fully available. The Accu-Chek Insight insulin pump system has three built-in alarm functions, an acoustic alarm, a vibration alarm, and visual notifications that you receive on the pump and handheld display. If the above-described issue occurs, the acoustic or vibration alarm functions might potentially be impacted and no alarm signal can be heard or felt. Missing an alarm e.g. while sleeping can potentially lead to an insulin under-delivery which could result in a hyperglycemic event. We have identified the root cause of this issue and have started to implement the appropriate corrections. To secure a continued optimal user experience for your patients, we would like to draw your attention to the important E-7 error messages and re-emphasize its correct interpretation and handling as it is described in the product manual. Recommended actions: Please ask your patients to ensure that all alarm functions on the pump are always activated In case the pump is dropped or otherwise has been exposed to mechanical impact, shock or damage, please ask your patients to always check the pump s functionality (insulin delivery and alerts) as described in Chapter 9 of the product manual, and as a first step reinsert the battery to initiate a self-test If your patient is unsure whether the acoustic alarm function of the pump is working properly, users can always provoke an acoustic signal by pressing a button. (Please ask your patients to make sure that the acoustic button feedback is enabled on the pump) As a routine, please ask your patients to check the pump and handheld displays regularly during the day and before they go to sleep for any notifications In case your patients receive an E-7 error message on the pump, please ask them to follow the important instructions provided in the pump manual: Tel.

If your patients continue to see this message, please instruct them to call our Accu-Chek Customer Care line at + xx xxxx xxxxx to get the pump replaced. This issue is only affecting the Accu-Chek Insight system. Due to their different design and electronic components all other Accu-Chek pumps are not impacted by this issue. The competent authority, users of the Accu-Chek Insight system and distributors have been informed about this field action. Thank you in advance for your understanding and co-operation. If you have any further questions, please do not hesitate to contact our Accu-Chek Customer Care line at xxx-xxx-xxx or your local Accu- Chek sales representative at any time. Kind regards, 2/2