Models Goldfisch Fat Extraction Apparatus

Similar documents
Labconco Small Stoppering Chamber

Open Kjeldahl Apparatus INSTRUCTION MANUAL

Part No Revised Nov Lil Max. Instruction Manual Model #2389, 2389EX

36 IN. ELECTIC FIREPLACE FMI MODEL # LEF36 Mounting Base Feet and Wall Mounting Bracket Included

RAPID DIGESTORS. Model Rapid Digestor-25 without digestion tubes Model Rapid Digestor - 4 without digestion tubes INSTRUCTION MANUAL

Hooded Kjeldahl Apparatus

!"" #$% "!&' ( ( ) *

WET/DRY VACUUM. QUEST for Continuous Improvement Windsor s Quality Management System is Certified ISO MODEL: T1. Operating Instructions (ENG)


Air Pump Up to 800 gallons

e Bath Fan with Light User s Guide

Operating Instructions (ENG) MODEL: AM115V. Read instructions before operating the machine. AM /24/00

Instruction Manual. Fun Pop 4oz Popper

Reacti-Therm I and III Heating Modules

MACHO POP Instruction Manual Model 2554EX, 2554BCX

User s Manual and Operating Instructions

Part No Revised April Pop Maxx. Instruction Manual Model #2552, #2552C, #2552EX, #2552XJ

User s Manual & Operating Instructions

Pressurized Filter Up to 1250 gallons

OWNERS MANUAL For IN ROOM BREWING SYSTEM MODEL REFILLABLE CAPSULES K901. Includes: Installation Use & Care Servicing Instructions

RapidStill II MODEL INSTRUCTION MANUAL. Product designs are subject to change without notice REV. B / ECO A631

16 Stand Fan with Remote Control Model No: FS3-40R-30

e Heater/Exhaust Fan/Light User s Guide

20 High Velocity Fan with 135 Tilt Head

Instruction Manual. Dogeroo, Super Dogeroo, and Mini Dogeroo

40E ELECTRIC FAN HEATER

e Bath Fan with Light User s Guide

INSTALLATION USE & CARE MANUAL ALL WEATHER SL-SERIES QUARTZ TUBE ELECTRIC INFRARED RADIANT HEATER

MANUFACTURING NUMBERS: CORN CAROUSEL. Model CC-19 Series P/N /99. Owner s Manual

User s Manual and Operating Instructions

e Bath Fan with Light User s Guide

6 Oz. Twin Pop. Popcorn Machine Instruction Manual Model #2666EX. Cincinnati, OH USA. Part No EX Revised June 1996

DISCARD DECANTER IF: FAILURE TO COMPLY RISKS INJURY FUNNEL CONTENTS ARE HOT ILLUSTRATED PARTS CATALOG

Instruction Manual. Double Candy Apple Cooker

InstructIon Manual KrEs EQuIPMEnt stands

Reacti-Therm I and III Heating/Stirring Modules

Hot Dog Hutch. Models HDH-3, HDH-3DR, & HDH-4 owner s manual. Manufacturing Numbers:

MRS-6 Maxi-Guard II. Model: C , 12, 13, 14 OPERATION SERVICE PARTS CARE. Revised 2/02

2454 RETRO POP Popcorn Machine Instruction Manual Model #2454

Pro Pop 12/14 oz. Unimaxx Model 2214ST Instruction Manual (2214ST shown, see next page for other models covered)

DISCARD DECANTER IF: FUNNEL CONTENTS ARE HOT FAILURE TO COMPLY RISKS INJURY

!"" #$% "!&' ( ( ) *

MODEL: 1407, , 1408, & INSTALLATION AND INSTRUCTION MANUAL

Instruction Manual. Cheese Warmer with Pump

Model: PS-2 ProSense II Item # OPERATION SERVICE PARTS CARE

Instruction Manual. Super 88

Citation Models #2001, 2001ST, 2001BN Instruction Manual

Parts and Service Manual BURNISHER MODEL: MR , MR , MR , MR

Owner s Manual WARMER DRAWER. (Without Water Tray) Models WD-20, WD21A, & WD-35A. WD-20 Shown MANUFACTURING NUMBERS: P/N Rev.

Operator s Manual. Histology Bath

User s Manual and Operating Instructions

POLISHERS. MODELS SeaRay 133 OPERATING & MAINTENANCE

DUST FREE CARBON Whole House Air Purifier

Operation Manual and Parts Directory

#2389, 2389EX, 2389JX

OWNER S MANUAL. Installation Operation Maintenance. Saturn Equipment


OWNERS MANUAL MODEL EXTERMINATOR VS900 C/R BLACK VAPOR CLEANING SYSTEM SAFETY INSTRUCTIONS OPERATIONS CARE AND MAINTENANCE TROUBLESHOOTING WARRANTY

Model 6021 Electrically Heated Rain and Snow Gauge

MIGHTY PRO 1/4 HP CHILLER

USER MANUAL Stock Pot Range

FLOOR MACHINE EDGER MINI SCRUBBER MODEL E-1

Service Parts, Kits and Accessories

Wax Base Heater & Dispenser

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

DUAL VOLTAGE REFRIGERATORS 220/240 VOLTS AC AND 12/24 VOLTS DC INSTALLATION AND OWNER S MANUAL

Instruction Manual. For All Corning Hot Plates, Stirrers, Stirrer/Hot Plates with Digital Displays, and External Temperature Controller (6795PR)

IMPORTANT INFORMATION - PLEASE READ CAREFULLY

OPERATIONS MAINTENANCE MANUAL

Part No EX Revised: January Super Pop Maxx. Instruction Manual Model 2600EX

ELECTRIC FIREPLACE HEATER WITH SINGLE GLASS DOOR

CLIMAGUARD Air-to-Air Outdoor Heat Exchangers INSTRUCTION MANUAL. Rev. H 2015 Pentair Equipment Protection P/N

M Series Fan Filter Units

RPI Industries, Inc.

40E ELECTRIC FAN HEATER

User s Manual and Operating Instructions

SSS Intrepid 16 FMS. Manual

Service Manual Models CPH2 & CPH3

GULPER 16SV 16 GALLON WET/DRY VACUUM WITH FRONT MOUNT SQUEEGEE

ELECTRIC FIREPLACE OWNER S MANUAL

OWNER S MANUAL FOR MODELS DXA (Air-Automatic) DXM (Manual/Non-Air) DXH (Hydraulic Automatic) DXE (Electrical Mechanical Automatic)

P-60 and 60 Special. Popcorn Machine Instruction Manual Model #1866, #2085, #2086 and #2656. Cincinnati, OH USA

User Manual / Installation Instructions

KNHX32 SERIES NEMA 4/4X AIR-TO-AIR HEAT EXCHANGERS

Instruction Manual. Cheddar Easy All-In-One Cheese Corn Shop

5-Gallon Pail Mixer OWNERS MANUAL M-2000 SERIES MIXERS. Contents: Page: Safety, General & Receiving... 2 Operation... 3 Parts List...

Oil and Water Retort Instruction Manual

BUNN H5E H5M H5X BUNN-O-MATIC CORPORATION POST OFFICE BOX 3227 SPRINGFIELD, ILLINOIS PHONE: (217) FAX: (217)

Self Serve Poppers. Instruction Manual All Export Models. Cincinnati, OH USA. Part No EX Revised June 1996

Owner s Manual. Terpene Trap CF 1000

Professional Warming Drawer and Warming Drawer Cabinet

Instruction Leaflet. Read instructions before operating. Retain for future reference.

VACUUM OVEN. Model: VPX9-2. Installation and Operation Manual

ALL WEATHER SL-SERIES QUARTZ TUBE ELECTRIC INFRARED RADIANT HEATER INSTALLATION USE & CARE MANUAL

Electronic Service Manuals

Instruction Manual. Part No Revised September 2004

Hot Dog Roller Grills

RPI Industries, Inc.

Inline Heater for Solvent or Gas. Installation and Operation Manual

BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

Transcription:

Models 3500100 Goldfisch Fat Extraction Apparatus

Copyright 2017 Labconco Corporation. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change. Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals. The warranty for the Goldfisch Fat Extraction Apparatus will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record. Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3 rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this warranty based on a service agent s travel, working hours, the site s entry restrictions and unobstructed access to serviceable components of the product. Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied. Returned or Damaged Goods Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier. Labconco Corporation and its dealers are not responsible for shipping damages. The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery. Limitation of Liability The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user s compliance with such regulations. Contacting Labconco Corporation If you have questions that are not addressed in this manual, or if you need technical assistance, contact Labconco s Customer Service Department or Labconco s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time. Part # 3518100, Rev. L ECO L123

TABLE OF CONTENTS CHAPTER 1: INTRODUCTION 1 Components Shipped 1 General Description 2 Component Identification 2 CHAPTER 2: INSTALLATION 4 Preparation 4 Location 4 Utility Connections 4 Initial Start-Up and Check Out 4 Gasket Assembly 5 CHAPTER 3: NORMAL OPERATION & ROUTINE MAINTENANCE 6 Summary of Operation 6 Procedures 6 Maintenance 8 APPENDIX A: REPLACEMENT PARTS 9 APPENDIX B: DIMENSIONS 11 APPENDIX C: SPECIFICATIONS 12 APPENDIX D: WIRING DIAGRAM 13

Chapter 1: Introduction Components Shipped Carefully check the contents of the carton for damage that might have occurred in transit. Do not discard the carton or packing material until all components have been checked against the following component list and the equipment has been installed and tested. As shipped, the carton should contain the following: Part Number Description 35001 Goldfisch Fat Extraction Apparatus 35051 Beakers, 100 cc (12) 35049 Sample Tubes (6) 35048 Reclaiming Tubes (6) 35050 Extraction Thimbles (6) 35052 Condenser Gaskets (6) 35191 Wave Washer Beaker Seal (6) 35054 Ring, Beaker Retainer (6) 35185 Assembly, Heater Cover (6) 35181 Instruction Manual (this document) Product Service 1-800-522-7658 1

Chapter 1: Introduction General Description The Labconco Goldfisch Fat Extraction Apparatus is designed for single or multiple fat extractions according to the methods prescribed in the Association of Official Analytical Chemists (AOAC) or other technical methods involved with the use of a continuous extractor. These units operate on a continuous type extraction principle; however, the refluxing time normally required by the Soxhlet Extraction Method is usually reduced through the design principles incorporated in the Goldfisch Extractors. The following instructions outline the features, installation and maintenance of the Goldfisch Extractor. This manual should be kept handy so that questions of use or service can be answered. Component Identification Refer to Figure 1 for the specific location of these components. 1. Manifold Assembly. This assembly is located at the top of the apparatus and provides an even flow and distribution of water to the condenser heads. Water exits through the water control valve, located on the upper left hand portion of the unit. And flows through the manifold assembly, depositing cool water in the condenser head system. 2. Pressure Release Valve. These valves are located at the top of the manifold above each of the condensers. It provides an escape for excess gaseous pressure which may build up in the condenser. The valves automatically reset themselves to assure a closed system of extraction. 3. Condenser Head System. These cylindrical shaped condensers located above each heater provide an essential part of the extraction process. The solvent vapor rises from the boiling solvent and reaches a point inside the condenser head coming into contact with metal cone shaped surface which is cooled by water. The vapors condense and drip through the sample held in the ceramic thimble, extracting the fat from the sample. 4. Heater System. Each 100 watt adjustable spring loaded heater provides an even distribution of heat through the use of coiled heater wires embedded in a ceramic material. The variable heat control switch located at the front of the unit provides a variety of settings. 2 Product Service 1-800-522-7658

Chapter 1: Introduction Figure 1 Product Service 1-800-522-7658 3

Chapter 2: Installation Preparation The Goldfisch Extractor is shipped complete and fully assembled. The operating parts shipped with the unit will be installed after the unit has been located in the laboratory area. Location Choose a location in the laboratory which has a convenient 115 volt A.C. outlet, an unrestricted 3/8" NPT cold water supply and drain facilities of at least 3/4" NPT. Caution: The Goldfisch Extractor should be operated inside an explosion proof hood with proper airflow to handle the extremely flammable solvents that are used during the Goldfisch Extractor s normal operation. Utility Connections (1) Once the unit has been placed in its final location, connect the cold water supply to the 3/8" NPT control valve water fitting on the upper left. (2) Connect the 3/8" NPT control valve on the upper right to an open drain of sufficient size (3/4" IPS or larger) to accommodate the cooling water. Cooling water should be a minimum of.75 gal/min or 2.8L/min for flow (55 F (13 C) or less is preferable). (3) Plug the unit into an electrical outlet. Initial Start-Up and Check Out Turn main power and variable heat control switches on. Make sure all heating elements heat properly. Wait two to three minutes. 4 Product Service 1-800-522-7658

Chapter 2: Installation Gasket Assembly Remove contents of the operating parts package which should be taped to the unit. Assemble the gasket, wave washer, retaining ring and glassware according to assembly drawing instructions as shown in Appendix A. Hand tighten the retaining rings to achieve vapor seal. Turn water on and make sure there is an even flow leaving the unit. Product Service 1-800-522-7658 5

Chapter 3: Normal Operation & Routine Maintenance Summary of Operation The Labconco extractor reduces extraction time from 16 hours to about four hours through its efficient refluxing system. The operation is carried out by using a Pyrex beaker as the solvent chamber. The procedure involves placing the samples between the boiling solvent and a cold surface. As boiling continues, the solvent vaporizes, condenses on the cold surface and washes back through the samples to the boiling solvent below. Procedures The following section provides a general outline of steps necessary to operate a routine extraction on the Labconco Goldfisch Fat Extractor. Actual test methods, sample preparation, solvent choices, etc. are determined by official methods of the A.O.A.C. and should be referred to in all cases. 1 (1) Turn on cold water supply. Check visual flow at the drain area. Water flow will be adjusted later after system is operating to obtain normal reflux rate. Heater and water flow rates may be adjusted to vary reflux rates according to the operator s preference. (2) Prepare sample(s) per standard method for the product under examination. See assembly drawing in Appendix A. (3) Place required quantity of sample in the extraction thimbles. Filter paper or fritted glass thimbles may be substituted according to operator preference. (4) Place thimble containing sample into Pyrex sample tube. (5) Insert sample tube assembly into clip from the front. Push upward until bulb of sample container meets holding clip. 1 For more information, see Official Methods of Analysis of the Association of Official Analytical Chemists. 6 Product Service 1-800-522-7658

Chapter 3: Normal Operation & Routine Maintenance (6) Place solvent of proper type and quantity into beaker. A quantity of 25 to 35 ml of solvent is normally sufficient for an estimate 3 to 4 hours of refluxing. (7) Assemble beaker to retainer ring with the wave washer inserted and twist to lock securely in place. After beaker is locked in place, pull down elevation control knob and twist ¼ turn to the left. Raise control knob to catch position and again twist ¼ turn to the left. Slowly allow heater to rise against bottom of flask. DO NOT let go of knob until heater is in position against beaker. (8) Select heat rate required using panel control switch(es). Control is provided for full range of power with the HIGH position providing both heaters with approximately 100 watts each. Control position 5 provides approximately 35 watts each. (9) Turn the main power switch, located on the right side, to the ON position. Switches on the panel each control two heaters. NOTE: If only one extraction is being made, the adjacent (or companion) heater, operating on the same panel switch, should be loaded with a beaker of solvent or water to load the heater and increase its service life. (10) Pre-heating the heaters for about 8 to 10 minutes prior to starting samples will reduce waiting time for start of refluxing. (11) Water flow rate through the condenser head system can be regulated to provide a wide range of reflux rates, when used alone or in combination with heater control selections. NOTE: If water flow rate is reduced too much, solvent will be lost in the beakers. A condenser without flow rate will permit solvent to escape through the pressure relief valves. (12) When extraction is completed, place heater covers on heaters to prevent solvent from coming into contact with the hot heater. Remove sample tubes, replace with reclaiming tubes and remove heater covers. Heat transfers solvent into the reclaiming tube, leaving oil in the beaker. (13) Drying of the extract is accomplished efficiently with a special swing beaker holder (one holder behind each heater assembly) which inclines the beaker at a slight angle in the warm air above the heater assembly. When all but the last few milliliters of solvent is evaporated, the beaker is placed in a drying oven, weighted and compared to the tare weight of the beaker. Weight difference is the amount of extracted material. Product Service 1-800-522-7658 7

Chapter 3: Normal Operation & Routine Maintenance (14) Heater covers are provided to prevent solvent from coming into contact with the hot heater. Use covers when removing or attaching beakers to condenser heads. (15) The Labconco Fat Extractors are suitable for carrying out the indirect Method of Fat Extraction (also known as Crude Fat Extraction ). This procedure requires the determination of moisture, followed by extraction of the dried material, with the loss reported as the ether extract. Official methods should be referred to for specific test procedure of the indirect method in all cases. Maintenance All of the parts contained in the operating parts package are routine maintenance items. It is suggested that additional supplies of these items be kept in stock. Items such as gaskets or beakers occasionally wear out or break and need to be replaced. You may order these through your local laboratory apparatus dealer or by contacting Labconco Corporation. Be sure to specify the Labconco part number when ordering. If contacting Labconco directly, please furnish the name of the dealer from whom you purchased the equipment. 8 Product Service 1-800-522-7658

Appendix A: Replacement Parts PART # QTY DESCRIPTION 12726 1 Pilot light 13516 1 Valve, needle, water 13159 1 Switch, mercury, toggle single pole, 115 volt, 15 amp 13564 1 Valve, pressure relief 35094 1 Gasket, outer condenser shell 30139 1 Wire, No. 18, stranded nickel with braided glass cover for heater harness and lead, 15' length 35167 Wiring harness 6 unit 35022 1 Heating element, 115 volt, 110 watt 35163 1 Switch, infinite control 35046* Pkg. of 6 Beaker, glass (Pyrex), tall form, 5' high 35053 Pkg. of 6 Beaker, glass (Pyrex), 100 cc, 3 1/4" high 35054 Ring, beaker 35059 Pkg. of 6 Extraction thimble, 22 x 80 mm 35061 Pkg. of 6 Sample tube container (thimble holder), glass (Pyrex) 35062 Pkg. of 6 Reclaiming tube, glass (Pyrex) 35160 Pkg. of 6 Gasket, condenser, upper 35067* 1 Sample tube container (thimble holder), stainless steel 35185 1 Heater cover 35191 6 Req d. Wave washer Beaker seal 35120 Spring, compression heater push rod 35172 Condenser assembly *Not furnished as standard equipment, but offered as an accessory. Note: When ordering replacement parts, specify the Labconco equipment serial number. Product Service 1-800-522-7658 9

Appendix A: Replacement Parts 10 Product Service 1-800-522-7658

Appendix B: Dimensions 35.7" 26.9" 9" Product Service 1-800-522-7658 11

Appendix C: Specifications 6-unit Goldfisch Fat Extractor Model No. 35001 operates on 115 volts, 50/60 Hz, (max.) at 5.2 amps. It is shipped complete with power cord and grounded (3 wire) plug. Standard features are as follows: 1. Main power on-off. 2. Full range variable heat control switch. Each switch controls two heaters. 3. All stainless steel (Type 304) condenser assembly. 4. Pressure relief valves for each condenser. 12 Product Service 1-800-522-7658

Appendix D: Wiring Diagram Product Service 1-800-522-7658 13