STERILIZATION PROTOCOLS

Similar documents
STERILIZATION PROTOCOLS

The Do's and Don ts in the Instrument Sterilisation Process

Purpose. Team Roles and Responsibilities. System Requirements. Instrument Processing Training

Decontamination Process Manual

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Fiona Collins Liliana Donatelli Andrew Smith

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Principles and Techniques of Instrument Processing and Sterilization

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106

Washroom. PURPOSE: To provide clean washrooms for staff, visitors and patients MATERIALS: 19-1

CLEANING INSTRUCTIONS

Table of Contents. English

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

X99 Series. Dental Handpiece. Instruction For Use

Special Precautions Isolation Plan Environmental Services

Harvey Chemiclave 5000 Owner And Operator Guide

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032

PREFERENCE ICC. A Complete System For Efficient Sterilization

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

When you sterilize an instrument set, you know what the standards are.

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental

Jennings Street School

Effective Date: 0 8 / 1 0 / 1 7. Replaces: 08/11/16. Formulated:1997 SPECIAL (MEDICAL) WASTE MANAGEMENT

Cleaning and Sterilization for Instruments & Posts

PERFORM Operating Document

STERITITE INSTRUCTIONS FOR USE

When you successfully complete the test, you will receive a certificate and a registered key ring. We look forward to hearing from you.

Instructions for Use Spinal Punches IFU

Cleaning & Sterilization

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

DRAFT. Module 4. Decontamination & Environmental Cleaning

Operating / User Manual

eware Handling & Cleaning Procedures

R 7420 HANDLING AND DISPOSAL OF BODY WASTES AND FLUIDS

Central Sterile Services Department will receive, decontaminate, process, store and/or dispense sterile and non-sterile to departments requesting.

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

Introduction. Sharps. Working Safely with Sharps at UAB (OHS_HS255) Objectives. Defined

Peninsula Dental Social Enterprise

User Manual J0841X. Piezo Electric Ultrasonic Scaler / Polisher Combo Unit. Jorgensen Laboratories, Inc.

The future is bright. And clean. And fully automated.

A new wave in automated instrument washing. Introducing the new HYDRIM G4 series

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

LAB SAFETY FOR CUSTODIANS

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Aim. Learning outcomes. 13 May Megan Reilly RN BN Cert IV Assessment & Workplace Training MHlthSc (Inf Control) Immunise Cert MACN CICP 1

6 things to consider. when designing your steri-center

CODE OF PRACTICE. No 11 CLEANING OF EQUIPMENT AND PREMISES

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com

Infection Control Policy and Procedures

AUTOMATIC STERILIZERS

Michael Norton 01. How to use this calendar: TIPS CHECKLISTS CONTACT US SUCCESS A LETTER FROM THE DIRECTOR OF EQUIPMENT & TECHNOLOGY,

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Reprocessing Reusable Medical and Dental Equipment: Policy and Protocols First Nations and Inuit Health Alberta Region

Cavitron Systems, Sterilizable Handpieces and Reservoir Bottles INFECTION CONTROL PROCEDURES

PIHA DUTIES HANDBOOK. 64 Glenesk Road, Piha

CLEANING FOR INFECTION CONTROL Ultrasonic Cleaners. Ultrasonic Cleaners. Copyright Quality Medical Innovations P/L & Majac Medical Products P/L

A new wave in automated instrument washing and disinfection Introducing the new HYDRIM G4 WD series

The Hydrim M2 washer disinfector. Perfectly prepares large volumes of instruments for sterilisation.

HYDRIM. Powerful cleaning the first step to effective sterilization. Instrument Washers A HIGHER STANDARD

Cleaning and Sterilisation for Burs, accessories and filling material

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

CLEANING OF INSTRUMENTS

Corporate. Laundry SOP. Document Control Summary

HYGIENE AND SANITATION

The Hydrim C51wd Washer Disinfector. Perfectly prepares instruments for sterilisation in as little as 19 minutes.

Dietary Department, Infection Control

Instrument Processing. Midmark Animal Health Products

Slide 1. Welcome to the on-line training course for Waste Management!

INFECTION PREVENTION AND CONTROL ON LINEN AND LAUNDRY MANAGEMENT. ROSAFE E. PERALTA, RN, MAN Chief Nurse MARYMOUNT HOSPITAL

Processing instructions for the Ligamys suture forceps

reference ICC A complete system for efficient sterilization

OPERATION MANUAL. LED Coupling OM-TR355E. Please read this operation manual carefully and file for future reference.

Device Reprocessing Checklist Scoring Matrix 1=low 5=high

User Manual J841Q. Magnetostrictive Scaler / Polisher Combo Unit. Jorgensen Laboratories Tel: Fax:

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

Safe Use of a Class II Biological Safety Cabinet

PICNet CONFERENCE 2013

FOOD SERVICES DIVISION. November Employee Safety & Sanitation Training For the School Year

Steam Sterilization and Sterility Assurance in Health Care Facilities ANSI/AAMI ST79. Suzanne Latta, CSPDT. Education Materials Coordinator

HydroFlux Welder Torch System T0-102 / T0-103

Student Induction P7-15 Commercial Kitchen

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

Transcription:

Department Lead Brandy Avril, RDH Anna Allen, RDH STERILIZATION PROTOCOLS The Sterilization department is the life-blood of the AKMOM project. The goal is to maximize access to care for patients by allowing Dentists, Assistants and Hygienists to retrieve sterile instruments and dispose of contaminated instruments in the most efficient and safest way possible. In order to achieve this goal the following protocol must be followed: All volunteers scheduled to work in the sterilizing area must have had Hepatitis B Vaccines All volunteers scheduled to work in the sterilizing area MUST wear long pants, close-toed shoes and Personal Protective Equipment (PPE) appropriate for the job they are performing. Clean gloves will be worn when handling sterile instruments. Safety glasses, masks, gloves and gowns will be worn when working with contaminated instruments. Stations that volunteers may be assigned to include: o Contaminated Drop Off area o Ultrasonic Station o Rinsing, Drying and Loading Sterilizer Tray Station o Loading and Unloading Sterilizers o Sorting Sterile Instruments o Assisting Personnel Requesting Restorative and Surgical Instruments o Separating AKMOM Instruments from those used in Dental Hygiene and from the Personal Instruments brought by practitioners. Instruments dropped off or picked up for sterilization brought from private offices must have the owner s name written on the sterilizing pouch and the pouch should be left with the tray of instruments. If more than one pouch is needed, the practitioner s name should be written on each pouch. It is the responsibility of the person dropping off the instruments to write the owner s name on the pouch. Please double check that the name is written on the pouch before they leave. Once sterilized, these instrument pouches are placed in a separate area for retrieval by the practitioner/assistant. 10/2015

All contaminated sharps (including needles, blades etc.) should be removed chair-side prior placing the instruments in the transport tub. All burs should also be removed from handpieces while at chair -side. Drop-off station volunteers should ask the assistant who is bringing the instruments for sterilization to remove any sharps/burs that were not removed chairside. If a runner is used to retrieve instruments, they should enforce removal of sharps in the department prior to transport from the department to the sterilization area. All volunteers in the sterilizing area should familiarize themselves with the needle stick/sharp instrument protocol and report any incident to the department lead immediately. Sterilization Protocol for Portable Dental Clinic ADCF employs 8- uniquely identified MIDMARK M-ll Ultraclave gravity displacement steam sterilizers in support of each free dental clinic. Sterilizers are to be placed as directed by ADCF personnel and according to MIDMARK location requirements. Environmental requirements as identified by MIDMARK are also considered i.e. ambient temperature and area relative cleanliness. Once connected and running, the sterilizers cannot be turned off for the duration of the clinic including in evening and non clinic hours. Electrical supply shall be reviewed with ADCF staff prior to powering up sterilizer units. Sterilizers are then loaded with instruments and a test spore vial identified for the appropriate sterilizer. Test spore vials (Biological Indicators) are manufactured by 3M ESPE. #1261 is a 24 hour test vial, #1262 is a 48 hour test vial. While ADCF provides and prefers the #2161 vials, either is sufficient to prove sterilizer performance. Once the testing cycle is completed, the vials are placed in incubators along with a control vial. The control vial will turn yellow indicating the presence of bacteria. The other individually marked vials will turn purple indicating that no bacteria is present in the vial and proving the operability of the specific sterilizers being tested.

ADCF experience over 1000 incubator tests is that the vials begin turning purple after 4 hours, indicating the absence of bacteria in the vial. The vials are read at 12 hours by ADCF staff to verify operability, read again at 18 hours and again at 24 hours for the #1261 vials. Using the #2162 the incubators will continue to be run for the entire 48 hours. Data from each of these tests is to be recorded in the Biological Monitoring Record which is included on each ADCF truck. Once a vial indicates the lack of bacteria by turning purple, sterilizer operability has been validated. Although the vial is referred to as a 24 hour test vial, the vial actually turns either purple of yellow within 12 hours. ADCF has verified with 3M (test manufacturer) that the presence or lack of bacteria can be verified before 24 hours. If a specific vial should turn yellow, a second test will be performed on that sterilizer to verify that human error was not a factor. If a second test is failed, that sterilizer will be pulled from service for that event and declared inoperable. Sterilizers are to be used in Pouches Mode as recommended by MIDMARK for ADCF applications. In this mode, the sterilizer runs at 270 degrees F. for 5 minutes followed by a 30 minute drying time. Due to the extremely large number of instruments required to be repetitively sterilized at an event (over 5000), ADCF recommends removing instruments wet unless adequate time is available to let instruments dry. In order to satisfy concerns with wicking, the process whereby air born particulates penetrate porous materials or wet paper pouches, ADCF uses all nylon pouches which are impervious to contamination from air bourn particulates. ADCF provides nylon pouches donated by Henry Schein Cares. Approximately 12,000 are used at each 100 operatory clinic. Sterilizers are to be loaded, operated and unloaded in accordance with manufacturer s guidelines. ADCF utilizes Zirc cassettes for standard instrument sets- PURPLE for dental hygiene, GREEN for restorative or general dentistry and BLUE for oral surgery. These as well as all other specialized instruments provided by ADCF come to the event pre-sterilized and ready for use. Sterilization volunteers must inspect pouched instruments after set up to find punctured or unsealed pouches. Any compromised pouches must be re-sterilized. Sterilization Set Up ADCF provides: 8 sterilizers 2 power stations 3M Biological Indicator Incubator and test vials Instruments in see through plastic bins (see attached list)

Biological Monitoring Record 2 portable sterilization sinks 6 Ultrasonic cleaners 12,000 Nylon pouches Plywood for sterilization tables States must provide: Table covers, if desired Paper pouches for personal instruments Sharpie markers to mark personal instruments Ultrasonic cleaning solution Distilled water (250-300 gallons total for both sterilization and operatory water tanks) Heavy dishwashing gloves and nylon gloves for volunteers PPE gear Signs indicating dirty instruments and clean instrument areas Towels Infection control protocols Boxes, hot dog paper containers or similar to transport clean and dirty instruments from sterilization to clinic area Hand sanitizer Hot pads for handling sterilization trays Scrubbing brushes for use at the sterilization sinks Small baskets for burs to use in ultrasonic cleaners (small tea strainers work best) The best set up for the sterilization area is to have an inner ring of tables with the 2 power stations inside the ring. Plywood is provided to place on top of the tables which hold the sterilizers. Areas must be set up to receive ADCF dirty instruments, personal dirty instruments, handpiece station, clean ADCF instruments, clean personal instruments, sinks and ultrasonic cleaners. Please see attached diagram for more clarification.

All sterilization volunteers must wear PPE protection garb and must have a Hepatitis B vaccination. No open toe shoes are allowed in the sterilization area. Gloves must be worn by all volunteers who touch dirty instruments and volunteers working in the sink area must wear heavy duty dishwashing gloves. Instruments are all banded to identify each type of instrument. Clean, pouched instruments are placed in bins which are labeled with the instrument name along with an identifying picture. Personal instruments should be kept separate from ADCF instruments so that personal instruments do not get put into ADCF bins. Some volunteers should be designated as dirty instrument volunteers. These volunteers may, clean instruments, sort dirty instruments, check cassettes and bag dirty instruments. These volunteers may also place trays of dirty, bagged instruments into the sterilizers. Clean volunteers should start the sterilizers, remove clean instruments from sterilizers, and sort the instruments into the appropriate bins. Steps for processing instruments: 1) Instruments are dropped off on at the designated dirty or used location. Instruments will then be briefly inspected to make sure there are no foreign objects i.e. teeth, sharps, gauze, cotton rolls or any other items that are not an instrument. 2) Instruments will then go into the ultra-sonic cleaners for the designated amount of time. 3) Upon completion of the ultra-sonic cycle, instruments are then transferred to the sinks for rinsing. At this time instruments will again be inspected and any stubborn blood or saliva can be lightly scrubbed and removed. Instruments should not be scrubbed with disinfecting wipes or gauze pads. 4) After rinsing, instruments are transferred to an area where they can be sorted and dried. This area will also be where instruments are bagged. 5) Instruments should be bagged individually unless they are in a cassette or they are a doctor s personal instruments. An internal indicator strip will also be placed in each bag at this time. Instruments with a hinge should be placed in the sterilization bag with the hinge open. When bagging instruments, you should be aware of the proper manufacturer s instructions which states that bags should be folded and sealed at the appropriate point. Bags folded in half are not considered by the manufacturer to be sterile. Check for a perforated line or a simple dotted line. This is where the bags are intended to be folded over and sealed. 6) After instruments are bagged appropriately and placed on trays they can be transferred to a sterilizer. All instruments will be run on the Pouches cycle. Please note it is very important that the trays are not overloaded trays with instruments. This may cause bags to stick together, cassettes to melt and/or over heating of the sterilizer. If you experience any of these upon the completion of the sterilization cycle please make note and correct those that may be overloading. If the sterilizer over heats, the cycle will not complete.

Instruments should be moved to another sterilizer and the overheated unit shall not be used until ADCF staff has had a chance to examine it. 7) Upon completion of the sterilization cycle, doors to the units will automatically open. Although we do not recommend the full dry cycle, it is recommended that instruments be allowed to dry for a short time before removing them from the sterilizers. This allows the instruments to cool slightly and it also allows the sterilizer to cool. With the frequency of use, if not allowed to cool down slightly between each load the sterilizer may over heat. Instruments and trays will be hot when removed so appropriate heat protection (pot holder) should be used. Any instruments that are sterilized in a half paper, half nylon bag should be allowed to dry for the full drying cycle. If manufacturer s directions are followed and a full drying cycle is not completed, the bags run the risk of tearing due to moisture and you will have to resterilize. Half paper/half nylon pouches should only be used for personal instruments. 8) Instruments are then transferred to an area designated as clean where they can be examined for tears or gaps in the seal one last time. Bags should be checked to determine that the bags are free of tears and instruments should be checked to ensure no debris is present. At this point instruments will be sorted and placed in their corresponding bins to be recirculated back into the clinic for use. 9) Handpieces- When a hand piece is returned for processing it is to be wiped with alcohol to remove any debris. Disinfecting wipes are not to be used to wipe handpieces. Handpieces should not go through the ultra-sonic cleaner or be rinsed under water. They are then lubricated and bagged for sterilization. Handpieces can be processed with the other instruments and steps 5-8 apply when processing hand pieces. 10) If instruments are returned unused and still in their all nylon packaging, they shall be inspected for tears. If the sterilization bag has not been compromised the bag shall be wiped with the appropriate disinfecting wipe and then returned to its bin for redistribution. If the bag is partially made of paper the instrument must be rebagged and resterilized. Instruments supplied on each truck are as follows: Restorative Cassettes 80 Scalpel Handles 100 Extra restorative instruments 23 (Cowhorn) Forcep 30 Explorer 120* 88L Forcep 30 Mirror 250* 88R Forcep 30 Cotton Plier 200* Burnisher 20* Root Tip Elevator 25

Hollenback 20* 53L Forcep 30 Composite Instrument 20* 53R Forcep 30 Condensor 20* Bird Beaks 25 Spoon 20* Anterior Forcep 30 Tanner Carver 20* 17 Forcep 30 Articulating Forcep 20* 151 Forcep 80 Tofflemire 45* 150 Forcep 80 Amalgam Carrier 20* Pedo 151s Forcep 10 Dycal 20 Pedo 150s Forcep 10 Mouth Prop 90 Spatula 20 Hygiene Cassettes 72 Cassettes include one each of the following: Oral Surgery Cassettes 80 Mirror Extra OS instruments Perio Probe Minnesota 20* Explorer 11/12 Periosteal 20* H5/L5 Large Straight Elevator 20* 204 SD Small Straight Elevator 20* Columbia 4L/4R Needle Holder 50 Gracey 13/14 Rigid Bone File 30 Gracey 11/12 Rigid Root Tip Pick 30 Crane Pick Elevator 25 High Speed Handpiece 100 Hemostat 75 Slow Speed Motor 100

Tissue Forcep 50 Nose Cone 100 Spoon Currette 100 Contra Angle 100 Scissors 45 Potts East/West 50 *Denotes one of each in cassette Crown/Bridge Scissor 20 Rongeur 50 Cryers 50 Syringes 350

Sterilization sinks and ultrasonic cleaners Recommended Sterilization Set Up Handpiece station Dirty Personal instrument drop off Sterilizers and power stations Dirty ADCF instrument drop off Personal instrument pick up Clean instruments 1/2014