BIOVAC 500 SMOKE EVACUATOR

Similar documents
Operator s Manual 2012 Bovie Medical Corporation

Operator s Manual AER DEFENSE TM

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

MODEL 7000 SUCTION UNIT

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

HERCULES (110V) (220V) USER S MANUAL

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

MODEL: 1407, , 1408, & INSTALLATION AND INSTRUCTION MANUAL

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

CONSTANT SUCTION UNIT

POLAR BREEZE Statim Technologies, LLC IMPORTANT SAFETY INSTRUCTIONS AND PRECAUTIONS

Owners Manual Model: XXL

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

AIR PURIFIER CLASSIC INSTRUCTION MANUAL. CAUTION: Read manual carefully for correct procedures and proper operation.

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

OWNER S MANUAL FOR ASPIRATION / FILTRATION VACUUM SYSTEM MODELS: 2511 & 2515

301-P2000. Humidifier Heater, 115V. Instructions for Use. Pegasus Research Corporation, 1518 E Edinger Ave., Unit A, Santa Ana, CA 92705

5580 S. Nogales Highway Tucson, AZ Phone: (800) Fax: (520) ANAPOD Instructions For Use

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

ELITE (110V) (220V)

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

USER INSTRUCTION MANUAL SAPPHIRE SERIES 1100 & 1100EC

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

30 Professional Cooling Unit. Model Number: XC3000. Owner s Manual READ AND SAVE THESE INSTRUCTIONS

USER INSTRUCTION MANUAL SAPPHIRE SERIES 800

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

OWNERS MANUAL. Installation. Operation. Specifications. Features. Maintenance. Warranty REV 02_04 FA SP-ENG OM - 02_04

Cavitator Ultrasonic Cleaners

Operation and Maintenance Manual

RecoverX Oil-Filled Hermetic Refrigerant Recovery System. Operation and Maintenance Manual

Model # VWK-2 INSTRUCTION MANUAL

OWNER S MANUAL FOR DRY VACUUM PUMP MODEL: 2019

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

User Instruction Manual

Eagle 2500 TM OWNER S MANUAL. Electronic Air Purification 110 and 220 volt Systems. EcoQuest International. Application.

SAVE THESE INSTRUCTIONS

USER MANUAL SAVE THESE INSTRUCTIONS

HEALTHCARE SAFETY INSPECTION

VACUUM OVEN. Model: VPX9-2. Installation and Operation Manual

Instruction Cum User Manual For EUROLITE

INSTRUCTIONS FOR OPERATION AND CARE OF

CA3500, BA3500, CH3500 New Comfort

WET & DRY GARAGE VAC 20L CAPACITY WATT 1.5M MOTOR TANK STEEL HOSE KP702 ON BOARD ACCESSORY STORAGE SUCTION STAINLESS

R100 Oil-Less Refrigerant Recovery Unit

Whynter Portable Ice Maker 33 lb capacity - White

MULTI-FUNCTION HAND VACUUM INSTRUCTION MANUAL. For Authentic Monster Replacement Parts Call Model #: H056

ESS-2200 Double Station Scrub Sink. with Two Sensor Operated Gooseneck Spout. ESS-3200 Deluxe Double Station Scrub Sink

Cavitator Ultrasonic Cleaner Instruction Manual

SAFETY ALERT SYMBOLS SAFETY PRECAUTIONS

User s Manual. i-warm Fluid Warmer

Digital Heat Block User Manual

User Instruction Manual

Model: C C Ambassador 20. Automatic Carpet Scrubber OPERATION SERVICE PARTS CARE

Multi-Tech S1000 / S1000-Ultra Air Purifier

18 MaxxAir Portable Evaporative Cooler Owner s Manual This Manual covers the EC18DVS MaxxAir Portable Evaporative Cooler.

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-12 Task Force Pneumatic Portable Dental System

TFC 400. OPERATOR MANUAL Clarke MODEL /09 Rev C Document Number

AIR SCRUBBER X-3400, X-3500, X-3700

OWNERS MANUAL. Installation. Operation. Specifications. Features. Maintenance. Warranty REV 0905 FAOM15_0905

WHYNTER Eco Pure Air Purifier

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

PORTABLE DEHUMIDIFIER

Compression therapy device & accessories Model 701A User manual

User Manual. Twin Piston Vacuum Pumps: Oil-less WVP Series: WVP12, WVP10, WVP5, WVP4, WVP3, and WVP2

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

Cleanshield HEPA 550 Owner's Manual

Installation & Operating Instructions

Use and Care Manual. Flammable Material Storage Refrigerator LRBFS06W1HC

Easy-Lam School Budget Roll Laminator

PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA USA [215] / FAX [215]

Printed in U.S.A Rev. A 6/2005 0B

Torrent irrigation pump Reorder No Operating and Maintenance Manual

INSTRUCTIONS FOR OPERATION AND CARE OF

Operation & Maintenance Manual MONSOON WET DOWNDRAFT TABLE

48 MaxxAir Portable Evaporative Cooler Owner s Manual

SPECIFICATIONS FEATURES

VACUUM OVENS MODEL: 1410MS,1430MS, 1450MS INSTALLATION AND OPERATIONAL MANUAL

Operating Instructions

BEFORE OPERATING, READ ALL INSTRUCTIONS IN THIS MANUAL CAREFULLY.

OWNER S MANUAL FOR DRY VACUUM PUMPS MODELS: / / / 2047

Cascade Comfort Bathing Systems with Aqua-Aire Safe Operation & Daily Maintenance Instructions

G-7s. Instruction Manual. G-Series Cooler COUNTERTOP COOLER. Part No.11IPA

DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4

Series: MBC1-TC Mini Benchtop Temperature Controller

NOTE: All the illustrations in this manual are for explanation purposes only. Your air conditioner may be slightly different.

OPERATION & INSTALLATION MANUAL AIR PURIFICATION SYSTEM (APS300)

INSTRUCTION MANUAL 4HP WET / DRY VACUUM VC V 60 Hz 7A. Gal

THE MEDIX Actineb. Technical Specifications. % Particles < 5 um 75% * Class II BF. Intermittent (use for 60 Type Rating

The Ottawa Hospital OPERATING ROOM NURSING POLICY, PROCEDURE, PROTOCOL MANUAL

MV 2. Operating Instructions. Before first use of the unit read these operating instructions and act in accordance with them.

Use and Care Manual. For Beverage Cooler RU

INSTALLATION AND OPERATION MANUAL

Dyna-Form Air. Alternating Pressure Relief Overlay System

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

G-10s. Instruction Manual. G-Series Cooler UPRIGHT COOLER. Part No.11IPA

Operator s Manual. 6 Gallon Model No. VQ607SFD FOR YOUR SAFETY. Read and understand this manual before use Keep this manual for future reference

Operation & Maintenance Manual MONSOON WET DOWNDRAFT TABLE. Visit our Website for more information on this product.

MV 4 MV 4 Premium. Operating Instructions. Before first use of the unit read these operating instructions and act in accordance with them.

OPERATIONS MANUAL. Hotplate Stirrer/Hotplate/Magnetic Stirrer Model HTS-1003/HTP-1002/MGS Certified

Transcription:

BIOVAC 500 SMOKE EVACUATOR INSTRUCTION MANUAL REF 909105 and 909105-05 Wallach Surgical Devices 95 Corporate Drive Trumbull, CT 06611 U.S.A. Phone: (203) 799-2000 Fax: (203) 799-2002 IMLP011 Rev. B 2/09 1

INTRODUCTION The Wallach BIOVAC 500 Smoke Evacuator is a simple-to-operate electrical device designed to remove smoke particulate (smoke plume) generated as a by-product of certain medical procedures, such as electrosurgery and electrocautery. It is intended for use in short term procedures performed in a physician s office. As with all such devices, there are instructions and precautions. Please read the manual and become familiar with your equipment prior to use. WARNING READ INSTRUCTIONS FOR USE BEFORE USING THIS PRODUCT. Do not use this device in the presence of potentially explosive anesthetic gases. This device is a BF type device. Type BF equipment is a type B equipment with an F type applied part. Type B is a piece of equipment providing a particular degree of protection against electric shock, particularly regarding: -Allowable leakage current -Reliability of the protective earth connection. F-type isolated (floating) applied part. Applied part isolated from all other parts of the equipment to such a degree that the patient leakage current allowable in single fault condition is not exceeded when a voltage equal to 1.1 times the highest rated mains voltage is applied between the applied part and earth. Always use a three-prong, grounded electrical outlet to prevent the possibility of electrical shock. CAUTION Do not use an electrical extension cord. However, a UL-approved power strip (such as those used on accessory carts) of sufficient current rating is acceptable. The Sterile Disposable Tubing (REF 920002) is for single patient use only. Due to possible cross contamination, it is to be used for one patient, then discarded per local, state, or federal regulations regarding medical waste disposal. The maximum allowable continued use of a single filter assembly (REF 909094) is 30 minutes (see FILTER ASSEMBLY). Filter efficiency and effectiveness are dramatically decreased, rendering it ineffective, if used for a longer period. Only Wallach Surgical Devices is authorized to service or repair the unit. Do not attempt to disassemble or service the device, other than changing a fuse. Unplug the unit before changing the fuse. Do not allow fluids of any kind to enter the device, either from spills on the cabinet, or through the vacuum hose. This may create an electrical hazard. IMLP011 Rev. B 2/09 2

BIOVAC SMOKE EVACUATOR COMPONENTS The Wallach BIOVAC 500 Smoke Evacuator consists of two components: a) the Motor Cabinet, with its electric motor and controls, and b) the Smoke Plume Filter Assembly and a STERILE single use Disposable Smoke Evacuation Tubing. MOTOR CABINET The Motor Cabinet contains a high power motor/blower which is controlled by two switches: a) the Power ON-OFF switch (labeled POWER), and b) the Air Flow Control Switch (labeled SUCTION ADJUST), which adjusts the amount of air being moved. There are no electrical user-serviceable components within the cabinet. Prior to your first use, connect a Filter Assembly and Disposable Smoke Evacuator Tubing, and operate the Smoke Evacuator through the range of suction available. You will notice that the sound level increases as the suction range is increased. Although the cabinet is sound-insulated, it is the nature of high efficiency blower motors to produce elevated sound levels. Some experimentation will lead you to the correct setting. FILTER ASSEMBLY The non-sterile Filter Assembly utilizes two filter devices to clean the collected air. Within the length of the four-foot hose is a smoke scrubbing, odor removing material that will compress as the Smoke Evacuator is operated. This compression will eventually restrict the flow of air in the tube, thus limiting the amount of time that the Smoke Evacuator can be in continuous use (using the same Filter Assembly). The expected maximum operating time limit for the filter is 30 minutes, which far exceeds the typical time needed for one electrocautery or electrosurgery procedure. Should the device continue running beyond this operational range, the unit may become warm to the touch and the filter will be ineffective. The second filter element, a viral, bacterial, particulate filter, is positioned in the filter assembly at the connecting end of the hose (connecting to the Motor Console). It efficiently removes potentially harmful microbes as small as.02 microns. Filter Assembly and Disposable Smoke Evacuator Tubing must be disposed of per your local, state or federal guidelines for the disposal of medical waste. Both are available from your local Wallach distributor. For the name of the nearest distributor, call (203)-799-2000 and request customer service. IMLP011 Rev. B 2/09 3

OPERATING INSTRUCTIONS A powerful motor/blower within the Motor Cabinet creates a vacuum which removes the smoke plume from the operating site via a specially constructed hose and filter assembly. This Filter Assembly traps odorous particulate along its 2.4 m length by means of filter material inserted in the hose passage, and by a final stage micron filter at the connecting end. 1) Set the unit in an upright position on a solid structure with a smooth level surface. 2) Be certain the Wallach BIOVAC 500 Smoke Evacuator is turned OFF. Now plug the cord into a grounded electrical receptacle. 3) Open a fresh Filter Assembly package. Attach the square filter housing of the hose assembly to the smoke collection port in the center of the top cover. Attach the sterilized disposable filter tubing. 4) If tubing is to be connected to a speculum 5) Attach the distal end of the tubing to the speculum. One Filter Assembly has a maximum limit of 30 minutes of continuous operation. 6) Turn the unit ON by depressing the POWER switch, and regulate the SUCTION ADJUST knob control to suit the condition, keeping in mind that the sound level increases as the vacuum increases. Test the unit prior to each patient use to insure that the BIOVAC 500 Smoke Evacuator is drawing air, and to acquaint your patient with the sound of the motor. 7) When the procedure is complete, turn OFF the unit, remove the Disposable Tubing and dispose of as per local, state, and federal regulations for handling medical waste. Note: Do not operate this device without a Filter Assembly in place. IMLP011 Rev. B 2/09 4

SPECIFICATIONS Electrical requirements: 120VAC, 50/60 Hz, 7.9A, 240V AC, 50/60 Hz, 5.0A Electrical Fuses Two TIOAL250V Two T5AL250V Power Cord Length Power Plug Sound Level From One Air Flow Range Filtration Filter Assembly Disposable Tubing & Straw Filter/Tubing/Straw Length: Dimensions Weight Country of Origin: 2.28 meters Hospital Grade Meter: 65 DBA at low speed 3.0 CFM to 6.5 CFM 99.99% Effective to.02 Micron Non-Sterile Sterile 3.35 meters 30.5cm (L) x 25.4cm (W) x 30.5cm H (including handle and feet) 8 kg U.S.A. MAINTENANCE After each use, the unit must be cleaned and disinfected. To sanitize the unit, wipe down with a disinfectant. While the finish on the instrument cabinet will resist scuffing and the chemical attack of most acids and alkalis, any liquids spilled on the cabinet should be wiped up immediately. IMLP011 Rev. B 2/09 5

WARRANTY The Wallach BIOVAC Smoke Evacuator is supported by a one-year warranty from date of purchase covering any failure of the device due to defective workmanship or components, when used in compliance with the product's intended use. The warranty remains valid within the time period, provided only Wallach supplied filter sets and components are used with the device. Only Wallach Surgical Devices is authorized to service or repair this unit. Do not disassemble the device. There are no user-serviceable components within the housing. Two extremely accessible electrical fuses are located on the rear of the Cabinet (see SPECIFICATIONS). SERVICE/REPAIR Only Wallach Surgical Devices is authorized to service or repair this unit. If repair is attempted outside the factory, the warranty will be considered void. Wallach Surgical Devices is not responsible for any injury resulting from repairs made by other individuals or organizations not certified by Wallach Surgical Devices. If a repair is needed, carefully package the Biovac in a protective carton. Equipment must be sanitized before it is returned. Items that are not sanitized will be returned to the customer freight collect. All shipments must be made via prepaid parcel post or U.S. Mail. C.O.D. packages will not be accepted. Return carton to: Wallach Surgical Devices 95 Corporate Drive Trumbull, CT 06611 U.S.A. Phone: (203) 799-2000 Fax: (203) 799-2002 IMLP011 Rev. B 2/09 6

ACCESSORIES REORDER NUMBER DESCRIPTION RECOMMENDED (REF) REPLACEMENT 920002 Sterile Disposable Tubing Every procedure 909094 Single Filter Assembly After 30 minutes of continuous use 109639 Electrical Fuses as needed 109641 Replacement Power Cord as needed TROUBLE SHOOTING/FUNCTIONAL TESTING PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Motor does not run 1. Unit not plugged in 1. Plug in power cord 2. ON switch not activated 2. Activate switch to "1" 3. Fuses blown/defective 3. Replace (be sure unit is 4. Defective motor unplugged) 4. Return to WSD On/off switch is not 1. Unit not plugged in 1. Plug in power cord illuminated 2. Fuses blown/defective 2. Replace be sure unit is unplugged 3. Defective switch 3. Return to WSD MAINTENANCE SCHEDULE Disinfect Unit Check power cord for fray Operate (see instructions) After every use Prior to each use Once yearly IMLP011 Rev. B 2/09 7

EXPLANATION OF SYMBOLS TYPE BF EQUIPMENT = Type of equipment is a type B equipment with an F type applied part. Type B equipment is a piece of equipment providing a particular degree of protection against electric shock, particularly regarding: -Allowable leakage current -Reliability of the protective earth connection. Type F = isolated (floating) applied part Applied part isolated from all other parts of the equipment to such a degree that the patient leakage current allowable in single fault condition is not exceeded when a voltage equal to 1.1 times the highest rated mains voltage is applied between the applied part and earth. Alternating Current Symbol = AC current See Instructions for Use Shock Hazard Warning For Professional Use Only REF SN Reorder Number Serial Number IMLP011 Rev. B 2/09 8