Selecting the right generator for your GC system. By Ed Connor, GC Product Specialist, Peak Scientific Instruments Ltd

Similar documents
Gas Chromatograph. Features. Continuous Analysis of Ultra High Purity Argon to ppb levels.

FIXED GAS ANALYZER (FGA)

Chemtron Science Laboratories Pvt. Ltd.

Solution. Supply single or multiple GC instruments Continuous Emission and Air monitoring Mud logging application for oil industry

White paper on the Dielectric Barrier Discharge Detectors. Introduction:

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Series 9000 Total Hydrocarbon Analyzer

SCION 436-GC. Dimensions and Weights

Gas Chromatography (GC-342)

Aromatics and Oxygenates in Gasoline June2018

Perkin Elmer Clarus 500 Gas Chromatograph

Pneumatech: Your guarantee of highquality. air and gas

Key Performance. CompactGC 4.0

Process Gas Chromatograph

PRODUCT CONFORMITY CERTIFICATE

CALIDUS tm the Modular micro Gas Chromatograph

This document is a preview generated by EVS

FPD - Flame Photometric Detector

Process Gas Chromatographs MAXUM edition II

Industrial Oxygen Generation Plant

POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview

Process Gas Chromatograph

Thermo Scientific Model 1160 Principle of Operation

Thermo Scientific. TRACE GC Ultra. Gas Chromatograph Maintenance and Troubleshooting Manual. PN , Revision March 2011

Thermo Scientific TVA2020 Introduction

TENDER NOTICE NO. 07/2012 INVITATION FOR BIDS / NIT

Recent Improvements in Process GC Instrumentation

1320 Detectors. 1 of 30. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

NATURAL GAS CLEANING AND TREATMENT

(12) Patent Application Publication (10) Pub. No.: US 2012/ A1

Datalogging as standard

Chromatography. Research Supplies. 2014/2015 Catalog

GC Injectors. GC Injectors

Process gas chromatography

Process Gas Chromatograph

Process Gas Chromatograph

Gas Detection - GD -

931D Style B Process Gas Chromatograph

intrinsically safe detector for hydrocarbon gases

TECHNICAL BULLETIN SELECTING A COMPRESSED AIR DRYER INTRODUCTION

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

HFID Heated Total Hydrocarbon Analyzer Module

Complete Gas Analysis

GC DETECTORS Reduction Gas Detector - RGD. Overview

T.ex Single Gas Detector VISA

FEEDGAS FOR MODERN HIGH-PERFORMANCE OZONE GENERATORS. Bruce T. Stanley. Ozonia Ltd Duebendorf, Switzerland 1999

Agilent 7890A Network Gas Chromatograph

High quality compressed air from generation to application

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

clean, dry compressed air

Global Leader in Efficiently Treating Compressed Air EXCELLENCE IN DESIGN, MANUFACTURING AND SERVICE

This SOP and the B.C. Field Sampling Manual are available at: www2.gov.bc.ca.

Technical Note # 01.05

Desiccants & Driers. By Norm Christopherson

V A C U U M A N D P R E S S U R E T E C H N O L O G Y

Rosemount 1500XA Process Gas Chromatograph

Rosemount 1500XA Process Gas Chromatograph

pneudri modular dessicant dryer

TYPES AND APPROPRIATE INJECTORS

MICHELL GROUP. Advanced Industrial Hygrometry and Process Instrumentation. Company Profile, 2011

GAS ANALYSIS / DETECTION Gas Detector

Technical specifications of FT-IR Spectrometer

Adsorption Compressed Air Dryer Heatless or heater regeneration

Process Gas Chromatograph

PURIFICATION DRYING SYSTEMS

CS744 Series: Carbon/Sulfur by Combustion

The leader in gas dehydration & filtration

PID Detector. Add-on to any GC For R&D, QC, Industrial Hygiene, and Environmental & Routine ppb Analysis. PID Model C/ PID Model C

ISO INTERNATIONAL STANDARD. Essential oils Analysis by gas chromatography on chiral capillary columns General method

Mass Sensors, Inc. Profile. Applications. Technology Baur Blvd, St. Louis, MO (314)

An ISO 9001 : 2008 Certified Company

Restek s Electronic Leak Detector

HIGH QUALITY COMPRESSED AIR FROM GENERATION TO APPLICATION A GUIDE TO THE ISO 8573 AND ISO SERIES OF COMPRESSED AIR QUALITY STANDARDS

YL6500 Gas Chromatograph

Process Gas Chromatographs

22 The Flame Ionization Detector


Active Reservoir Vent. Fluid Users

Siemens Spares. Maxum II. PD PA AP Maxum Edition II Detectors. Detector Overview 1. Thermal Conductivity Detector 2. Flame Photometric Detector 3

DISSOLVED GAS ANALYZER (DGA)

The Complete Guide to Gas Sample Drying for Your Analyzer or Process

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. TurboMartix Headspace and Headspace Trap Samplers Consumables and Supplies Reference Guide

High Efficiency Compressed Air Dryers. Adsorption Dryers Modular System INTELLIGENT AIR TECHNOLOGY

Cone Calorimeter. (ISO 5660 ASTM E 1354) The most comprehensive bench scale fire test THE BENCHMARK IN FIRE TESTING

CNG. PSB Industries. NG Series - Natural Gas Dryers N G-S R NG-SR D N G - E V N G - SV N G R

Applied Techno Systems

NEW Restek Electronic Leak Detector

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

TCD-to-FID Series Connection on a 6820 GC, Accessory G4319A

BST-MG07 Multi-gas Detecting Alarm Manual Instruction

TECHNICAL NOTE. Response Factors for Flame Ionization Detector Operation FLAME IONIZATION DETECTOR OPERATION BASICS

Dual Cell Microvolume Thermal Conductivity Detector

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

PRODUCT CONFORMITY CERTIFICATE

Quality. Water. Water Quality. Testing. Testing. Toxic Gas Detection. Toxic Gas. Detection. Air Quality. Monitoring. Monitoring. Monitoring.

Corpus Christi - Air Monitoring Project

Transcription:

Selecting the right generator for your GC system By Ed Connor, GC Product Specialist, Peak Scientific Instruments Ltd

Selecting the right generator for your GC system Gas generators offer a guaranteed round the clock supply of consistently pure gas, however it is important to choose the correct configuration for your GC system. Generators produce gases at a variety of flow rates and purities, depending on the purification system they employ. Hydrogen generators Precision Range - Hydrogen Gas Generator Hydrogen for detectors. The most commonly used GC detector worldwide is the flame ionisation detector (FID), which uses hydrogen as the fuel gas. Flame photometric detectors (FPD), pulsed flame photometric detectors (PFPD) and nitrogen phosphorous detectors (NPD) also require hydrogen. By checking the flow rate of hydrogen used by each detector, you can easily calculate your total hydrogen requirement for your GC detectors. Hydrogen for carrier gas. If you are considering using hydrogen for carrier gas, you will need to calculate the total flow requirement of your GC method(s) and add this to your detector hydrogen requirement (if applicable). The total flow information can be found in your instrument software, on the GC display panel, or on a method report. Otherwise you can calculate the total flow by adding together the column flow, split flow and septum purge flow. If you are currently using helium carrier gas and are planning to translate your methods to hydrogen carrier gas, you can use method translation software to calculate your hydrogen flow rate for the translated method. When using the same column, with adjusted pressure, the hydrogen requirement is typically around 1.25 x the helium requirement. Which system? Depending on the manufacturer, there is typically a choice between palladium, pressure swing adsorption (PSA) and desiccant hydrogen driers. Each has its relative pros and cons in

terms of purity, longevity and price. For detector supply, all three purification systems are suitable. For carrier gas, people usually opt for palladium or PSA systems since these are completely regenerative driers, compared with desiccant systems, which must be periodically replaced or reconditioned. Nitrogen generators Precision Range Nitrogen Gas Generator Nitrogen for detectors. Nitrogen is commonly used as a make-up gas in GC detectors such as flame ionisation detector (FID), thermal conductivity detectors (TCD), electron capture detectors (ECD), flame photometric detectors (FPD), pulsed flame photometric detctors (PFPD) and nitrogen phosphorous detectors (NPD). By checking the flow rate of nitrogen used by each detector, you can easily calculate your total nitrogen requirement for detectors. Nitrogen for carrier gas. If you want to use nitrogen for carrier gas, you will need to calculate the total flow requirement of your GC method(s) and add this to your detector nitrogen requirement. The total flow information can be found in your instrument software, on the GC display panel, or on a method report. Otherwise you can calculate your total flow by adding together the column flow, split flow and septum purge flow. If you are currently using helium carrier gas and are planning to translate your methods to nitrogen carrier gas, you can use method translation software to calculate your nitrogen flow rate for the translated method. Which system? There are a huge number of nitrogen generators available, from systems producing tens of litres per minute at lower purity, to UHP generators producing nitrogen at 99.9995% purity at lower flow rates. These systems typically purify nitrogen using one of two methods: The first method uses a

membrane, which is selectively permeable to oxygen and hydrocarbons which it removes from compressed air. The other technique is pressure swing adsorption (PSA), where a carbon molecular sieve column removes oxygen and hydrocarbons from compressed air. The PSA system is usually employed in the UHP systems. Additional removal of methane from UHP nitrogen is also available from zero nitrogen generators. These generators produce zero air (methane free air) from which they purify nitrogen. When using nitrogen for carrier gas or in labs with higher methane levels, it is advisable to use zero nitrogen, since external traps will not effectively remove methane, which can affect GC baseline signal. Zero air generators Precision Range Zero Air Generator Zero air for detectors. Zero air (methane-free air) is required as a flame support gas in flame ionisation detectors (FID), flame photometric detectors (FPD), pulsed flame photometric detectors (PFPD) and as the oxidiser in nitrogen phosphorous detectors (NPD). Zero air is compressed air with methane removed typically down to levels below 1.0 ppm. Zero air for carrier gas. For applications requiring zero air for carrier gas, methane removal down to a level below 1.0 ppm is important. Which system? Zero air generators from reputable manufacturers should have a reliable removal of methane down to levels below 1.0 ppm. It is good to check which parts need to be periodically changed/replaced eg. filters, catalyst chamber as this can add considerable cost to the initial capital investment. Ed Connor DR.SC. is GC-MS Application Specialist, Peak Scientific, Inchinnan Business Park, Scotland, UK. Prior to joining Peak in February of this year, Ed completed his Dr.Sc. at ETH Zurich in Switzerland using GC-MS to look at herbivore induced plant volatiles and their interaction with beneficial insects. He then joined the University of Zurich where his work focused primarily on volatile collection methods and analyses using GC-MS and GC-FID. +44 141 812 8100, econnor@peakscientific.com