Vitalmax 4000 CL Product Specifications

Similar documents
DPM Summary of Features and Benefits

PM-9000 Express Portable Patient Monitor

Product Specifications

1 Appendix - Product Specifications

MRI Patient Monitoring System (Model )

5/ 5i/ 5c/ 7/ 7i/ 7c/ 5 elite/ 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite

Multi-Parameter Patient Monitor

OMNI TOUCH SCREEN PATIENT MONITOR

Omni (K) PATIENT MONITOR

Atlas Monitor Outline Train the Trainer

SPECIFICATION FOR MULTIPARAMETER MONITOR

B40 Patient Monitor. Affordable clinical excellence. Advanced clinical parameters. Ease of use for fast decision-making. Performance and reliability

OMNI Express VS VERSATILITY IN VITAL SIGNS

CLEO VERSATILITY IN VITAL SIGNS

M8000VET /M9000VET Veterinary Monitor User s Manual

PATIENT MONITORING SYSTEMS

Measurements to guide your patient care

PATIENT MONITORING SYSTEMS

Nellcor with OxiMax Technology Product Ordering Guide A Full Line of Oximetry Sensors and Monitors

SureSigns VS4. Vital Signs Monitor Technical Data Sheet

HeartStart XL+ Configuration Worksheet

T1 BEDSIDE MONITOR GUIDELINE

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models:

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor

ncompass 8100H Series Vital Signs Monitor Operator s Manual

Menntor X7 Specification

The information in this document is subject to change without notice.

SERVICE MANUAL YM6000 Patient Monitor

PATIENT MONITORING SYSTEMS

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor

SECTION L MONITORING EQUIPMENT

Operator s Manual. LifeSense LS1-9R. Vital Signs Monitor. Capnography/Pulse Oximeter

Patient Monitoring System YM5500. Service Manual

Meditec M700 Series Patient Monitors

ARGUS VCM Vital Compact Monitor

SAINT JAMES HOSPITAL GROUP PURCHASING DEPARTMENT

T1 Monitor/Module Default Settings

SERVICE MANUAL. Patient Monitor M30. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

DPM4 Bedside Monitor Default Settings

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

Patient Monitor CM400PC Monitoring Module

BEDSIDE MONITORING Dash 3000/4000 V5

VS2000V Veterinary Vital Signs Monitor Operation Manual

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

Infinity M300+ Telemetry

Vital Signs Monitor PC-900. User Manual

L505. Patient monitor user manual. Technology that doesn t miss a beat. Tel:

General Skills. yes. Device controls, buttons, indicators, and connectors

CARDELL Veterinary Vital Signs Monitor

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis)

www. szcomen. com COMEN - SHARE WITH THE WORLD

User s Manual. Multiparametric Vital Sign Monitor

Propaq LT Outline Train the Trainer

VitalCare 506N3 Series Patient Monitor Operator s Manual

Infinity M300 Patient Worn Monitoring

CREATIVE MEDICAL. PC-3000 Multi-Parameter Monitor

Instructions for Use Vista 120

OMNI. Patient Monitor USER S MANUAL

CASMED 740 Vital Signs Monitor

QAM.L L550 series. Patient Monitor Manual QAM.L

Alarm and Measurement Default Settings

Temperature Controllers

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual

GE Healthcare. Dash Variable-Acuity Monitoring

V6 Vital Signs Monitor User s Manual

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1

THE CLINICAL CONNECTION

DINAMAP PRO Series V2 Monitor. Operation Manual

PULSE OXIMETER PACKAGE

INSTRUCTIONS FOR USE. IntelliVue Patient Monitor. MP5 Release G.0 with Software Revision G.0x.xx. Patient Monitoring

Vital Signs Monitor 300 Series

OPERATOR S MANUAL. For YM1000 Vital Signs Monitor. EU Representative TECNOMED 2000 S.L. Valencia, Madrid Spain

Vital Signs Monitor 300 Series

OKM 801INFANT INCUBATOR

Instructions for Use. Infinity Vista. Infinity Patient Monitoring Series

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium.

OPERATOR S MANUAL. NELLCOR N-200 Pulse Oximeter

Multiparameter. er Monitor DM

Checkout Procedure Manual

DINAMAP ProCare Monitor. Operation Manual

GIMA VITAL SIGNS MONITORS VITAL PRO-SNET

Vital Signs Monitor 300 Series

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English

SECTION L MONITORING EQUIPMENT

740 SELECT. Multi-Parameter Monitor

HUG.ai 1/24/06 9:33:34 AM HUG.ai 1/24/06 9:33:34 AM C M Y CM MY CY CMY K

Infinity M540 Patient Monitor

LifeSense Vet LS1-10R

Blood Pressure Monitor

Oxylog VE300 Emergency & Transport Ventilation

ARGUS LCM. Patient monitor ARGUS LCM and ARGUS LCM PLUS. Operating Instruction. Art. no.: rev.: d

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual

DGC-500 DIGITAL GENSET CONTROLLER

The Blood Pressure Monitor is digital upper arm monitors intended for use in measuring blood pressure and heartbeat rate with upper arm circumference

ARGUS LCM. Patient monitor ARGUS LCM and ARGUS LCM PLUS. Service Handbook. Art. no.: rev.: d

SurgiVet V6004 NIBP Monitor Operation Manual

MRI Monitor. All-in-one monitor for the MRI environment. Features. Parameters. Accessories

IQvitals Zone TM. For Models: User Guide

Transport Pro Patient Monitor Operator s Manual Software Version 2.1

BM3VET User Manual Code Veterinary Monitor Rev. 2.4

Transcription:

Page 1 of 5 Vitalmax 4000 CL 510 Garden Avenue North, Clearwater, FL 33755 Phone: (727) 442-8118 Fax: (727) 443-7257

A. Mechanical Description Size 10" H x 13" W x 8"D (25 cm x 33 cm x 20 cm) Weight 18 lb 13 oz (8.5 kg) Color Beige and Aqua B. Power Requirements Operation Internal battery Battery operating time Battery charge time AC mains input Fuses C. Performance Specifications 1. ECG AC/DC 12V 7.0Ahr sealed lead-acid 3-3.5 hours full charge 16-20 hours Universal Switching Power Supply, Hospital grade plug, 100-250 V/50-60 Hz 16VDC / 2.8A Output, UL, CSA, GS Two 5.0 A, 250 V, fast blow CMRR 100 db @60 Hz (common mode rejection ratio) Heart rate range 30-240 bpm Heart rate averaging 4 b eat average Heart rate accuracy ±5 bpm or10%, whichever is greater Response time of heart rate meter for a step change in heart rate 80-40 bpm less than 7 seconds. 80-120 bpm less than 11 seconds. Pacemaker pulse rejection Amp: ± 2 mv to ± 700 mv Duration: 0.1 ms to 2.0 ms Input 3-Lead ECG / Respiration patient cable with no series resistance Connector AAMI 6-pin Lead selection I, II, III (Standard Config.) ECG reference signal 1 mv software triggered Patient isolation Breakdown voltage: 5000 VRMS 60 seconds 60 Hz Leakage current a) Internal to ground < 0.01 ma (< 10 ua) b) Enclosure < 0.01 ma (< 10 ua) c) Leads-off sensing current < 10 ua DC current < 0.01 ma (< 10 ua) Frequency response/band width 0.5-40 Hz (±3 db) Patient drive current < 10 ua Maximum T wave rejection capability 120% of QRS Segment Defibrillator & ESIS protected Tested with 5 kv Recovery time following defibrillation Within 8 seconds. Lead fault alarm Audible, visual QRS indicator Audible : user volume control Asystole: audible and visual Gain selection 5 mv,2 mv, 1 mv, 0.5 mv, 0.1 mv Time to alarm tachycardia < 3 seconds Aspect ratio 0.24, 0.4, 0.6, ±0.08 sec/mv Alarm frequency Low alarm 20-24 Hz High alarm 30-34 khz ECG input impedance >2.5 M Ohm @ 10 Hz Page 2 of 5

2. ECG/ Thoracic Impedance Respiration Excitation current <100 ua Excitation frequency 65 khz Maximum allowable electrode impedance 2 K ohm Size adjustment User selectable Respiration rate range 4-99 rpm (respirations per minute) Respiration rate accuracy ± 2 rpm Respiration rate averaging 4 breath average Respiration alarm limits Upper 4-99 rpm Lower 4-98 rpm Sensitivity range Adult: 0.3-10 Ohms : 0.1-10 Ohms Apnea Audible and visual alarm in less than 20 seconds. Not recommended for apnea monitoring. Sensing electrodes RA and LA Lead Selection I, II, III 3.END-TIDAL CO2, inco2(min) and RESPIRATION Type Side stream, non-dispersi ve infrared Method CO2 sample line with nasal cannula or endotracheal tube connector CO2 averaging 4 breaths or breath-to-breath CO2 range 0-99 mmhg adult/pediatric (0-13.2 kpa) CO2 accuracy ± 2 mmhg (0-40 mmhg)[± 0.27 kpa (0-5.33 kpa)] ± 5% of reading [(40-99 mmhg)(5.33-13.2 kpa)] Rise time < 300 ms 10% to 90% Co2 CO2 calibration Manual with room air, and 10% CO2 every six months Sample aspiration rate 75 ml/min ±10 ml Sample line purging Automatic Water trap Disposable, Volume 4 cc Respiration range 4-99 rpm (respirations per minute) adult/pediatric Respiration rate accuracy ± 2 rpm Analog output Optional Alarm limits Respiration Upper 4-99 rpm Lower 4-98 rpm ETCO2 Upper 0-99 mmhg (13.2 kpa) Lower 0-98 mmhg InCO2 Upper 0-99 mmhg (0-13.2 kpa) Lower 0-98 mmhg Alarm volume Fixed 4. INVASIVE PRESSURE Channels Two Measurement range -30 mmhg to +300 mmhg (-4.0 kpa to 40.0 kpa) Waveform Span Adjustable to 30, 60, 90, 120, 180 or 300mmHg Pressure offset Adjustable in steps of 10 mmhg Gain accuracy ± 1% or ± 2 mmhg whichever is greater (± 0.27 kpa) Analysis value Maximum, minimum, and mean pressure Pulse rate from arterial waveforms Heart rate range 10-250 bpm Heart rate accuracy ± 1% or ± 5 bpm, whichever is greater Heart rate alarm limits Upper 0-250 bpm Lower 0-249 bpm Transducer sensitivity 5 uv/v/mmhg Transducer excitation 5V DC Transducer zero adjust ± 150 mmhg (± 19.99 kpa) Transducer auto-zeroed on pressing the zero-adjust button Page 3 of 5

Band width Upper band limit: Factory selectable: 8, 9, 10, 12, 14, 16, 19, 22, 25, 30, 35, 40, Hz Default: 14 Hz Lower band limit: DC Display Digital waveform display with semiautomatic and manual pr essure range scaling Trend display of pressure waveforms Alarms Audio and Visual User adjustable upper and lower alarm limits User selectable pressure channel labels IP1(2), ART1(2), AO1(2), RA1(2), FA 1(2), CVP1(2), PA1(2) Alarm limits Max IP (Systolic ): Upper -30 to 300 mmhg (-4.0 to 40 kpa) Min IP (Diastolic): Upper -30 to 300 mmhg (-4.0 to 40 kpa) Mean IP Upper -30 to 300 mmhg (-4.0 to 40 kpa) 5. PULSE OXIMETRY (SpO2) Saturation range 0-100%, adult/pediatric/neonate Saturation averaging 8 beat average Saturation accuracy ± 2% (70-100%), ±3% (50-69%), (0-49% unspecified) Saturation alarm limits Upper: 0-100% Lower: 0-99% Pulse rate range 30-254 Pulse rate averaging 8 second average Pulse rate accuracy ±2 % @ 30-100 bpm Pulse alarm limits Upper: 0-250 bpm Lower: 0-249 bpm Pulse tone Pitch adjusts to SpO2 value Volume adjustable Pulse rate display Waveform, digital, pulse amplitude Sensor types Finger, Universal "Y", ear lobe clip Disposable and reusable wrap probes 6. NON-INVASIVE BLOOD PRESSURE (NIBP) Method Automatic oscillometric Parameters measured Systolic, diastolic, mean arterial pressure, pulse Scale mmhg or kpa Operating modes Manual, Automatic, Stat Repeat cycles 10-50 seconds; 1-99 minutes Rapid cycle update 1-4 minutes (STAT mode) Measurement range Systolic: Adult/pediatric 30-250 mmhg (4.0-33.3 kpa) 20-160 mmhg (2.7-21.3 kpa) Diastolic: Adult/pediatric 10-180 mmhg (1.3-24.0 kpa) 10-140 mmhg (1.3-18.7 kpa) Measurement time Typical 50 seconds Maximum 120 seconds Typical Stat 30 seconds Cuff inflation rate Not greater than 40-50 mmhg/sec (5.33-6.66 kpa/sec) Cuff inflation pressure 30 mmhg above last systolic (4.0 kpa) Cuff pressure range Adult/pediatric 0-250 mmhg (0-33.3 kpa) 0-140 mmhg (0-18.7 kpa) Initial cuff inflation Adult/pediatric 170 ± 10 mmhg (22.7 ± 1.3 kpa) 120 ± 10 mmhg (16.0 ± 1.3 kpa) Auto deflate pressure Adult/pediatric 280 mmhg; ± 5 mmhg (36.7-38.0 kpa) 235 mmhg; ± 5 mmhg (30.7-32.0 kpa) NIBP display accuracy ± 3 mmhg (0.4 kpa) Page 4 of 5

NIBP alarm limit Systolic: Upper 0-255 mmhg (0-34.0 kpa) Lower 0-254 mmhg (0-33.9 kpa) Diastolic: Upper 0-255 mmhg (0-34.0 kpa) Lower 0-254 mmhg (0-33.9 kpa) Mean: Upper 0-255 mmhg (0-34.0 kpa) Lower 0-254 mmhg (0-33.9 kpa) Pulse rate determinations 30-254 bpm Pulse rate averaging 4 beat average Pulse rate accuracy ±3 bpm @ 40-1 20 bpm ±10 bpm @ 121-200 bpm Pulse rate alarm limits Upper 0-250 bpm Lower 0-254 bpm Cuffs Reusable and disposable cuffs, infant, pediatri c, standard adult, large adult, thigh, with 6 foot air hose 7. TEMPERATURE Temp scale F or C Temp range 82.4-109.8 F (28.0-43.2 C) Temp accuracy ± 0.2 F (± 0.1 C) Temp alarm limits Upper 83.0-110.0 F (28.3-43.3 C) Lower 83.0-109.9 F (28.3-43.2 C) D Temp range 0-27 F (0-15C) D Temp alarm limits Upper 0-27F (0-15C) Lower 0-26.9F (0-14.9C) Temp probes Skin or rectal/esop hageal YSI 400 Series compatible: for use with invasive pressure models only Medtronics / Electromedics 2100 Series compatible: for use with non-invasive pressure models only D. Color LCD Display Matrix Effective display area Signal Indicators E. Printer Type Print mode Waveforms Resolution Waveform annotations F. Environmental Specifications Temperature Relative humidity SVGA: 800 pi xels (Horizontal) x 600 pixels (Vertical) 12.1 inch diagonal Green signal LED Built-in 2-channel thermal array recorder/printer Text or waveform Real t ime or alarm-triggered (10, 15, 20, or 30 seconds) 200 dpi vertical; 400-800 dpi horizontal Time, date, speed, channel identification, waveform lead selection Operating: 66 F - 86 F (19 C - 30 C) Storage 40 F - 1 10 F (4.4 C - 43.3 C) Operating: 20-80% (non-condensing) Storage 10-90% (non-condensing) Connect With Us Page 5 of 5