Standard Operating Procedure Gas Chromatograph Mass Spectrometry (GC/MS) - Agilent 7890B GC & 240 Ion Trap MS

Similar documents
Standard Operating Procedure Raman Microscope - Renishaw invia

Agilent 7890 Series Gas Chromatograph

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

SCION 436-GC. Dimensions and Weights

E²M. Innovation with Integrity. Enhanced Environmental Mass Spectrometer. Safety & Security

Perkin Elmer Clarus 500 Gas Chromatograph

Technical specifications of FT-IR Spectrometer

CALIDUS tm the Modular micro Gas Chromatograph

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

Agilent Intuvo 9000 Gas Chromatograph

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Gas Chromatography (GC-342)

POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview

Preparing the 6850 Series II GC for 597x MSD Installation

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

Radiation Safety Guidelines for Non-Users

ANNEX I. Technical specifications for: Supply of various Trace Explosive Detection Systems

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

FIXED GAS ANALYZER (FGA)

Agilent 7890A Network Gas Chromatograph

FPD - Flame Photometric Detector

HFID Heated Total Hydrocarbon Analyzer Module

Cone Calorimeter. (ISO 5660 ASTM E 1354) The most comprehensive bench scale fire test THE BENCHMARK IN FIRE TESTING

Mass Sensors, Inc. Profile. Applications. Technology Baur Blvd, St. Louis, MO (314)

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

Revised: January 8, Goal: Deposit, expose, and develop standard positive photoresist films as an etch mask, liftoff mask, or sacrificial layer.

HISTORY OF VARIAN CAPILLARY INJECTORS

General Procedures: Materials Synthesis Lab 1358 CNAM/Toll

22 The Flame Ionization Detector

ACQUITY UPLC HT Column Heater Instructions

HEALTH AND SAFETY MANUAL

Chemtron Science Laboratories Pvt. Ltd.

Operate the Mutli-Gas Monitor PID

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

Smoke Layer Height and Heat Flow through a Door

TCD-to-FID Series Connection on a 6820 GC, Accessory G4319A

Product Guide for Ludger Clean Post-endoglycosidase and Post-exoglycosidase Clean-up Plate

SAMPLE MANUAL - INSTALLATION - FOR REFERENCE ONLY. ChemScan mini Analyzer. Parameter. Method #### Installation, Operation & Maintenance Manual

Process Gas Chromatograph

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

API 2000 LC/MS/MS TurboIonSpray Ion Source Manual

TIMS TIMS. Thermal Ionisation Mass Spectrometry.

Apex E Quick Installation Guide

Specification Sheet. Equipment Model: BVAC-1200 Vacuum Furnace. Type. Chamber. Maximum Temperature 1250 C. Hot Zone

Lamorinda CERT - Unit 2 09/15/2016

PRODUCT CONFORMITY CERTIFICATE

Glovebox Usage Manual v1.4. When you enter the lab walk straight to the log book. Enter all information including N2 and O2/H2O levels.

Mobile CWA Detection and Identification. Innovation with Integrity. Mobile Mass Spectrometer MM2 CBRNE

Nexis GC-2030 Specifications

XPS Lab Chemical Hygiene Plan

1320 Detectors. 1 of 30. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Heat up ramp rate 100 degrees C per minute - empty chamber.

Analyze IT. Process Gas Chromatograph. PGC2000 Series

TENDER NOTICE NO. 07/2012 INVITATION FOR BIDS / NIT

Installing a Fast Oven Upgrade Kit

An Introduction to Mass Spectrometry

Safety and Emergency procedures in LAMP

HPLC Column Ovens and Temperature Controllers

Indirect Cryogenic Cold Trap System Instructions

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

EMERGENCY RESPONSE GUIDE IMMEDIATE EMERGENCY PROCEDURES

Emergency Action Plan

Safety In Chemistry. Identified Risks and Hazards

Standard Operating Procedures for Process Gas Handling

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Thermal Conductivity Detector on a 6820 GC, Accessory G4313A

Green Chemistry Safety Training PSE/CH 335. Departments of Forest Biomaterials and Chemistry North Carolina State University

Solution. Supply single or multiple GC instruments Continuous Emission and Air monitoring Mud logging application for oil industry

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Safety instructions and routines on lab Faculty of Odontology Malmö University

Installation Instructions

Training Fees 4,000 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch.

Refrigerated Incubator Model and Operating Instructions

Fisher Scientific Model 281A Vacuum Oven. Installation and Operating Manual Rev. 0

Tovatech LLC Phone (973) Rutgers St, Maplewood, NJ 07040

Standard Operating Manual

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

University of California, Davis Laboratory Self-Inspection Checklist

QAQC LABS ANALYTICAL DRYING

Weighing Solutions for Hazardous Areas FULL COMPLIANCE WITH ATEX DIRECTIVE

On-line Side Stream Oil in Water Analyser - Model OIW-EX 100 or OIW-EX 1000

Electron Microscopy Sciences

Process Gas Chromatograph

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

McNeil Group University of Michigan. MBraun Glovebox. Preparing Items To Be Brought Into the Glovebox

Environmental Express, Inc. AutoBlock Plus Operation Manual and Instructions CONTENTS. i. Introduction/General Information 2

CNS Cleanroom Safety Orientation Form

WELCOME TO ELLUTIA CHROMATOGRAPHY SOLUTIONS

Agilent 1290 Infinity II Vialsampler Data Sheet

Amerimade RCA Wet Bench SOP Page 1 of 13 Rev /22/2014

150 & 200 RT MG-SERIES GAS FIRED DOUBLE-EFFECT EFFECT ABSORPTION CHILLER-HEATER

Lab Safety Plan for Optics Room (Rm117)

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Recent Improvements in Process GC Instrumentation

Contingency Plan for Hazardous Materials and Waste Releases

portable gas analyser

Oil-Sealed Rotary Vane Pumps

Hiden HPR-20 QIC EGA for evolved gas analysis in TGA-MS

Transcription:

Standard Operating Procedure Gas Chromatograph Mass Spectrometry (GC/MS) - Agilent 7890B GC & 240 Ion Trap MS The Agilent 7890B is a state-of-art gas chromatograph (GC) that provides superior performance for all GC applications with the use of advanced electronic pneumatic control (EPC) modules and high performance GC oven temperature control. The Agilent 240 Ion Trap mass spectrometer (MS), combined with the Agilent 7890B GC, delivers exceptional performance and unsurpassed flexibility with both MS and GC options. The MS is configured in internal mode and provides both electron ionization (EI) and chemical ionization (CI). The 7693 automatic liquid sampler (ALS) works as the interface on 7890B, providing power and communications for one automatic injector and one automatic sampler tray that are easily installed without the need for alignment. 1.0 Operation Procedure Power outage: If there is a power outage, turn off the GC-MS according to the shutdown procedure to avoid a hazardous situation when power is restored. If there is an emergency, leave the MCL immediately and either return after emergency to shut down the GC-MS or contact the facility manager. When handling/using chemicals for preparation or use within the GC & MS, all applicable local and national laboratory safety practices must be followed. This would include, but is not limited to, correct use of Personal Protective Equipment (PPE), correct use of storage vials, and correct handling of chemicals. There are several important safety notices that you should always keep in mind when using the Agilent 7890 B GC & 240 Ion Trap MS: Many internal parts of the GC & MS carry dangerous voltages If the GC & MS are connected to a power source, even if the power switch is off, potentially dangerous voltages exist on: The wiring between the GC & MS power cords and the AC power supply, the AC power supply itself, and the wiring from the AC power supply to the power switch. With the power switch on, potentially dangerous voltages also exist on: All electronics boards in the instrument. The internal wires and cables connected to these boards. The wires for any heater (oven, detector, inlet, or valve box). Page1

All these parts are shielded by covers. With the covers in place, it should be difficult to accidentally make contact with dangerous voltages. Unless specifically instructed to, never remove a cover unless the detector, inlet, or oven are turned off. Do not use an Uninterruptable Power Supply (UPS) with a GC If the area where the GC is located suddenly loses power, an unsafe condition can result if the GC remains powered on. Do not use the GC with a UPS. Electrostatic discharge is a threat to the GC & MS electronics The printed circuit (PC) boards in the GC & MS can be damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle them, wear a grounded wrist strap and take other antistatic precautions. Wear a grounded wrist strap any time you must remove the GC right side cover. Many parts are dangerously hot Many parts of the GC & MS operate at temperatures high enough to cause serious burns. These parts include but are not limited to: the inlets, the GC oven and its contents, the MS ion trap oven, the MS manifold, the MS transfer line, the GC heated interface, the column nuts attaching the column to an inlet or detector, the valve box, and the MS foreline pump. You should always cool these areas of the GC & MS systems to room temperature before working on them. They will cool faster if you first set the temperature of the heated zone to room temperature. Turn the zone off after it has reached the setpoint. Be careful when working behind the instrument. During cool-down cycles, the GC emits hot exhaust which can cause burns. Oven thermal leaks Objects passing through the oven door seal can cause thermal leaks which create hazardous hot spots which cause burns and melt equipment. Do not allow wiring or temperature probes to pass through the oven door jam. Agilent recommends using one of the access holes. Micro-Electron Capture Detector (μecd) on GC The μecd on GC system contains a cell plated with 63 Ni, a radioactive isotope. The beta particles released at the energy level in the detector have little penetrating power the surface layer of the skin or a few sheets of paper will stop most of them but they may Page2

be hazardous if the isotope is ingested or inhaled. For this reason, handle the cell with care. Cap the detector inlet and outlet fittings when the detector is not in use. Never introduce corrosive chemicals into the detector. Do not use solvents to clean the μecd. Materials that may react with the 63 Ni source, either to form volatile products or to cause physical degradation of the plated film, must be avoided. These materials include oxidizing compounds, acids, wet halogens, wet nitric acid, ammonium hydroxide, hydrogen sulfide, PCPs, and carbon monoxide. This list is not exhaustive but indicates the kinds of compounds that may cause damage to 63 Ni detectors. When handling the μecd on GC system: Never eat, drink, or smoke. Always wear safety glasses when working with or near the open μecd. Wear protective clothing such as laboratory jackets, safety glasses, and gloves, and follow good laboratory practices. Wash hands thoroughly with a mild nonabrasive cleaner after handling the μecd. Cap the inlet and outlet fittings when the μecd is not in use. The oil pan under the standard MS foreline pump can be a fire hazard Oily rags, paper towels, and similar absorbents in the oil pan could ignite and damage the pump and other parts of the MS. Combustible materials (or flammable/nonflammable wicking material) placed under, over, or around the foreline (roughing) pump constitutes a fire hazard. Keep the pan clean, but do not leave absorbent material such as paper towels in it. 1.1 Refill Wash Vial a. Check and refill wash vials on the injector tower (Fig. 1). The volume of solvent should be above the minimum level specified on the wash vial. b. Default wash vials: A: dichloromethane (DCM); B: methyl tert-butyl ether (MTBE). Your method specifies which wash vials will be used. For all ECD methods, use wash vial B. Refill solvent based on your method. NOTE: See separate document for GC-MS method development. Page3

(a) (b) (c) Figure 1 (a) Injector tower; (b) Wash vials; (c) Indicator light 1.2 Switch Injector Tower a. Mount the injector tower over the injection port specified in your method. Default injection ports: Front: mass spectrometry (MS); Back: electron capture detector (ECD). b. Ensure that the Ready light is on after switching the injector tower (Fig. 1-c). c. Restart the GC system. See the Step #9 in Section 5.4. 1.3 Load Samples a. Push the button P (parking) to move the auto-sampler arm to the left (Fig. 2-c). b. Put your samples in the auto-sampler tray and record the sample number (Fig 2-b). c. Push the button P again. Sample injection will not start when the auto-sampler at the parking mode. (a) (b) (c) Figure 2 (a) Auto-sampler tray; (b) Sample number; (c) Button P Page4

1.4 Software Control a. Click System Control (Fig. 3) to open control software (Fig. 4). Figure 3 The button to turn on the control of GC system Figure 4 Instrument control software interface b. Click the minimized window of 7890A GC.44 (Fig. 5) to open the operation window (Fig. 6). Figure 5 Minimized window of 7890A GC.44 Figure 6 GC operation window Page5

c. Load your method by clicking the icon at the right side of method name (Fig. 7). Figure 7 The icon to load your method d. Set up sample list Click File New SampleList to open the new window (Fig. 8). (a) (b) Figure 8 (a) New SampleList and (b) the new window Find the folder with your name. If this is your first time using this GC, add a new folder under the profile of _Users with your name as the folder name. Add a new folder named with the date (for example: 20160531) in your folder. Enter the date as the name of your new sample list, and save it under the folder named with the same date. Click Save to add the new sample list. Enter sample name with vial number (i.e., sample position on the auto-sampler), injection volume, and injector position (Fig. 9). Note that your injection volume should match the value specified in your method. It is recommended to include a solvent blank as your first sample. Page6

Figure 9 The four items need to fill e. Save the data files by clicking Data File. The data files should be saved under the folder named with the date. Do not choose this folder as the default directory. f. Start running samples by clicking Begin and twice OK. In addition, the method should be double-checked in the new window. g. Ensure to fill the log-in sheet to record the start/end time as well as the parameters 1.5 Post-run Cleanup a. Take out your samples from the auto-sampler after detection. Refer to Section 5.3 to move the auto-sampler arm. b. Click the cross icon on the top right corner to shut down the control software. c. Data analysis for your samples will depend on your method. NOTE: See separate document for GC-MS method development. 2.0 Specifications/Features Three compartments: 7890B GC System (G3440B) 240 Ion Trap MS System (G3934A) 7693 Automatic Liquid Sampler (G4514A) 7890B GC System: Retention time repeatability < 0.008% or < 0.0008 min Area repeatability < 1% RSD Pressure setpoint and control precision to 0.001 psi Column oven temperature: ambient +4 C to 450 C Setpoint resolution: 0.1 C Maximum ramp rate: 75 C Cool down (22 C ambient) from 450 C to 50 C in 4 min Reliable, leak-free, in-oven capillary connections by Capillary Flow Technology Micro- Electron Capture Detector (ECD) with EPC module *Radioactive material: use < 15 mci Ni-63 as the electron source Page7

Multimode inlet (MMI) and Split/Splitless capillary inlet (S/SL) with EPC module Carrier gas and auxiliary gas: He Detector gas: argon/methane mixture 240 Ion Trap MS System: Mass range: 10-1000 Da in 0.1 Da steps (time programmable throughout the run) Scan rate: 5000 or 10000 Da/sec Mass axis stability: ±0.1 Da over 72 hr Operation temperature: up to 250 C for mass analyzer, up to 120 C for manifold Electrode surface: SilChrom for maximum inertness Ionization filament current: 10 100 µa Maximum ionization time: 65 msec Internal mode electron ionization (EI) and chemical ionization (CI) Detection system: Detector: off-axis design, ±10 kv HED and electron multiplier Analytical linear dynamic range: 10 3 to 10 4 Vacuum system: 280 L/s turbomolecular pump Bayard-Alpert ion gauge with thoria-coated iridium filaments DS-102 rotary vane foreline pump, > 95 L/min Foreline pump thermal vacuum gauge Control software: MS Workstation (MSWS) NIST 2014 MS Library Bundle including 243,000 spectra with names, chemical structures, and retention indices 3.0 User Requirements The Agilent 7890B GC & 240 Ion Trap MS must be used by authorized personnel only. All authorized users are expected to read and understand this SOP and follow the operation instructions carefully. No unauthorized personnel may use this equipment. All users must wear appropriate personal protective equipment. To become an authorized user, one must: 1. Complete Environment, Health & Safety (EH&S) training 2. Complete initial orientation and training for the Materials Characterization Laboratory 3. Receive training on this piece of equipment from lab personnel 4. Schedule equipment time using the calendar 5. Read and fully understand this SOP Page8

4.0 General Safety 4.1 Required Personal Protective Equipment Users must wear lab coats, safety glasses, and gloves. Shorts, open-toed shoes, high heels, and skirts, are forbidden 4.2 Emergency Procedures and Contacts For non-life threatening emergencies: notify the MCL facility manager and your PI immediately. Facility manager: Zongmin (Shirley) Bei, Ph.D. Office: 109B Furnas Hall, Tel: (716) 645-5165, Cell: (585) 354-5623 Email: zongminb@buffalo.edu or for police / ambulance, call 645-2222 In case of fire or other life threatening emergency: Exit the laboratory through an emergency exit door. Pull one of the fire alarms located in the main hallway. Dial campus police / ambulance at 645-2222. 4.3 University after hours laboratory use policy No working alone, use the buddy system! # Revised by Date Modification 1 Jiale Xu / Ning Dai 06/09/2016 Document initial release 2 3 4 5 Page9