A-dec 300 Support Systems

Similar documents
Instructions for Use. A-dec 461 Cuspidor with Support Center

InstructIons for use A-dec 561 Cuspidor

Instructions for Use A-dec Self-Contained Water System

Performer Cuspidor. Installation Instructions. (For Performer III Chair & International Performer I Chair) You Will Need. About This Install

Preference ICC. Sterilization Center

reference ICC A complete system for efficient sterilization

Instructions for Use A-dec Inspire Dental Furniture

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

Use and Care Guide. Over the Patient Units and Swing Mount Units

Procenter Delivery Systems. [Includes: Delivery Units and Assistant s Units (All Styles)] (no serial numbers on Assistant s Units)

OPERATION AND MAINTENANCE MANUAL

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-12 Task Force Pneumatic Portable Dental System

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-10 AseptiMini Pneumatic Portable Dental System

Regulatory Information and Specifications

PREFERENCE ICC. A Complete System For Efficient Sterilization

Installation Operation Maintenance Troubleshooting Version Dec/17. Mobile Cabinets

PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA USA [215] / FAX [215]

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-17X2 Express Plus II Portable Dental System

OPERATING INSTRUCTIONS

X99 Series. Dental Handpiece. Instruction For Use

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-17A Express Air Portable Dental System

OPERATING INSTRUCTIONS

Dental Unit User's Guide

Operating / User Manual

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

OPERATING INSTRUCTIONS

Table of Contents. English

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

SERRES SUCTION BAG SYSTEM MANUAL

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Cleaning & Sterilization

INSTALLATION INSTRUCTIONS

SAN RAPHAEL PRESSURE LITE VITREOUS CHINA SIPHON JET TOILET

BARRINGTON PRESSURE LITE VITREOUS CHINA SIPHON JET TOILET K-4327

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

KEEP FOR FUTURE REFERENCE 1 SCFM NOMINAL AIRFLOW SINGLE VACUUM SYSTEM DC-VOLTAGE POWER SYSTEM WITH MANUAL VALVE

Syclone Amalgam Separator Installation and operating guide

Instructions for use. 301 plus

Artizan Ortho-Pedo Bench Installation / User Manual

P.N User Manual STAINLESS STEEL ELECTRIC KETTLE 5964/5965 Series MODELS

Installation, Operation and Maintenance Guide AP3900 Series Thermostatic Mixing Valves

MEGATRON MODEL 2, 3, 4, and 5

Hot Chocolate Dispenser

Sloan AQUS Installation Instruction, Operation, and Maintenance Guide

OPERATION MANUAL. LED Coupling OM-TR355E. Please read this operation manual carefully and file for future reference.

P.N ELECTRIC KETTLE User Manual Series MODEL BVSTKT308

Preference Productivity refined

Note: Please provide valve serial number (stamped on cover of valve) when ordering parts. Tempered Water Outlet. Hot Water Inlet.

P.N Rev A ELECTRIC HOT WATER DISPENSER User Manual BVSTWD7915 Series MODEL

INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

ASPIRE Laboratory Aspirator

POWERGROOM. MultiCyclonic 1654 SERIES

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060)

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-425 Transport II Portable Dental System

Note: Please provide valve serial number (stamped on cover of valve) when ordering parts. Thermometer. Hot Water Inlet. Cold Water Inlet G3600LF

A-dec Inspire. Dental Furniture Collection

High Output Vacuum/Pressure Pump

English. Your Model Number is: HD1500 IMPORTANT

Easy Vac BAGLESS VACUUM

MODEL 7000 SUCTION UNIT

DENTAL UNIT. Rod Type OPERATING INSTRUCTIONS

Prime chiller. Inline Chiller. Instructions for Models #2645,2646

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

SeedVac TM REF 90091

User Manual ELECTRIC KETTLE Manual del Usuario HERVIDOR ELECTRICO. MODEL/MODELO BVSTKT7098 Series.

CLASSIC. SIDE CABINET Installation Instructions

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-525 Transport III Portable Dental System

Classic Bubbler Premix Cold Beverage Dispensers Operation Manual

VIVA ace Quick Operation Guide

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

Rocket Quick Guide. Rocket PSU Portable Suction Unit


Operating Manual for Profident / Profident PLUS

WL User Manual. Innovative Technology for a Connected World. global solutions: local support TM

AUTOTROL 363TC AUTOMATIC FILTER VALVE SERVICE MANUAL

ASG EZ-9000GR Tape Dispenser User Manual ASG #66136

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

INSTALLATION INSTRUCTIONS

Corn Flame Energy Corn Stove Model 3000

USER'S GUIDE 4757-E SERIES V

INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE

5000 Series Bottled Water System by.

474A+ INSTRUCTION MANUAL DESOLDERING STATION

INSTRUCTIONS FOR OPERATION AND CARE OF

Short instructions for use ESTETICA E30

Important Safeguards. Save these instructions for future reference. (continued on next page)

Tornado Operations & Maintenance Manual

Model HD 3501 Part No High Pressure Washer Operator Manual

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

Installation Instructions

USER MANUAL. Soho All-In-One Chair Rev A 4/23/18. *shown with optional accessories and/or finishes. Earthlite 2018, All Rights Reserved

PAT LOAD MOMENT INDICATOR DS 150 / 0038

DENTAL VACUUM SYSTEM Part Numbers: VS2O, VS40, VS50, VS50H, VS80 and VS80H RECOMMEND DAILY USE OF MONARCH CLEANSTREAM USER S MANUAL

EDGEMERE DUAL CONTROL WIDESPREAD LAVATORY FAUCET INSTALLATION INSTRUCTIONS

Services Unlimited Inc./Parts Guru. Assembly

Turbo Cyclonic Pro Bagless Cylinder Vacuum Cleaner

Model K 240 plus Part No

Transcription:

Instructions for Use A-dec 300 Support Systems 351, 352, and 353 Assistant s Instrumentation 362 and 363 Support Center and Cuspidor

A-dec 300 Support Systems Instructions for Use Copyright 2017 A-dec Inc. All rights reserved. A-dec Inc. makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. A-dec Inc. shall not be held liable for any errors contained herein or any consequential or other damages concerning the furnishing, performance or use of this material. The information in this document is subject to change without notice. If you find any problems in the documentation, please report them to us in writing. A-dec Inc. does not warrant that this document is error-free. No part of this document may be copied, reproduced, altered, or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or by any information storage and retrieval system, without prior written permission from A-dec Inc. Trademarks and Additional Intellectual Property Rights A-dec, the A-dec logo, A-dec Inspire, Cascade, Century Plus, Continental, Decade, ICX, ICV, Performer, Preference, Preference Collection, Preference ICC, and Radius are trademarks of A-dec Inc. and are registered in the United States and other countries. A-dec 500, A-dec 400, A-dec 300, A-dec 200, and reliablecreativesolutions are also trademarks of A-dec Inc. None of the trademarks or trade names in this document may be reproduced, copied, or manipulated in any manner without the express, written approval of the trademark owner. Certain touchpad symbols are proprietary to A-dec Inc. Any use of these symbols, in whole or in part, without the express written consent of A-dec Inc., is strictly prohibited. Regulatory Information Regulatory information mandated by agency requirements is provided in the Regulatory Information, Specifications, and Warranty document (p/n 86.0221.00), which is available in the Document Library at www.a-dec.com. Product Service Product service is available through your local authorized A-dec dealer. For service information, or to locate an authorized dealer, contact A-dec at 1.800.547.1883 in the USA and Canada or 1.503.538.7478 worldwide, or visit www.a-dec.com. Product Models and Versions Covered in This Document Model Version Description 351 A, B Assistant s Instrumentation 352 A Assistant s Instrumentation 353 A Assistant s Instrumentation 362 A Support Center and Cuspidor 363 A Support Center and Cuspidor

Content Map Bowl Rinse/Cupfill... 6 Touchpad...2 Water Bottle...9 Syringe...3 Cuspidor Bowl...6 Holder Adjustment...4 Solids Collector... 15 HVE/Saliva Ejector...5 Operate / Adjust... 2 Clean / Maintain... 9 Specifications and Warranty... 17 86.0308.00 Rev E 1

A-dec 300 Support Systems Instructions for Use Operate / Adjust Operate / Adjust Power On/Off NOTE A-dec recommends that you perform a shock treatment on the dental unit waterlines before you use the system for the first time. For more information, see Waterline Maintenance on page 9. Baseplate Delivery System Master Toggle (A-dec 332 shown) Chair Power Button Master Toggle Use the master toggle to turn on the power, water, and air. The master toggle is normally located on the delivery system, but may instead be located on the support center or floor box on chairs without a delivery system. If there is no power to the system, verify that the chair power button is pushed in. The chair power must be on in order for the master toggle to control system power. When to Turn Off the Power To save energy, turn off the power at the end of the work day and during longer periods of non-use. Status Light Touchpads and the Status Light The A-dec logo on the touchpad illuminates when the system is on and ready for use. If the status light blinks, it may indicate that a safety feature has interrupted chair motion. See Chair Safety Features on page 3 for more information. For complete details on touchpad functionality, please see your delivery system Instructions For Use. 2

Chair Safety Features Chair Stop Plate A-dec dental chairs and systems include several features designed to improve safety. Activated stop switches may halt the chair or prevent it from moving. To help ensure uninterrupted chair motion: remove any potential obstructions under the chair and attached modules avoid pressing the foot control disk or lever keep handpieces properly seated in their holders If the Chair Stops Unexpectedly Check the actions listed above to correct the condition. If the downward movement of the chair stopped because of an obstruction, use the touchpad or footswitch to raise the chair and remove the obstruction. Autoclavable Syringe Water Air 2 Clicks To install the A-dec syringe tip, push the tip in until you feel two clicks. Press both buttons simultaneously for the air/water spray. IMPORTANT For detailed instructions on syringe usage, flow adjustment, and maintenance, see the A-dec Syringes Instructions For Use (p/n 85.0680.00). 86.0308.00 Rev E 3

A-dec 300 Support Systems Instructions for Use Operate / Adjust Instrument Holders The instrument holders independently rotate for customized positioning of each instrument. To reposition, pull the holder slightly away from the adjacent one, rotate it to the desired position, and release. CAUTION Twisting the holder without pulling it away from the adjacent one will damage the mechanism. 1 Pull Out 2 Rotate Auto-Electric Holders Saliva Ejector High Volume Evacuator (HVE) If you have a chair-side vacuum system and auto-electric holders, lift the HVE or saliva ejector from its holder and open the valve to turn on the vacuum. Return the instrument to its holder to turn off the vacuum. Auto-Electric Holder 4

HVE/Saliva Ejector Left/Right Conversion Valve Body To convert the standard HVE and saliva ejector for left-handed operation, push the control valve out of the valve body, rotate the control valve 180, and push it back into place. NOTE Be sure to reinstall the valve on the same side of the valve body. Otherwise, the HVE or saliva ejector will not operate properly. Control Valve CAUTION Do not attempt to remove the HVE valve if you have an Air Vacuum System (AVS). The HVE is permanently attached to the tubing in this configuration and cannot be removed (see below). Air Vacuum System (AVS) HVE (control valve features circular groove) Saliva Ejector The AVS option includes a self-contained air vacuum generator. Also, with this option, the HVE valve body is permanently attached and cannot be removed from the tubing. To activate the saliva ejector vacuum, use the on/off toggle. Saliva Ejector On/Off Toggle 86.0308.00 Rev E 5

A-dec 300 Support Systems Instructions for Use Operate / Adjust Cuspidor Functions Bowl Rinse and Cupfill Cupfill Spout Bowl Rinse Button Cupfill Button Use the buttons under the cupfill spout or on the touchpad to control the bowl rinse and cupfill functions. The buttons support both programmed and manual operation. Button Description and Factory Presets Bowl Rinse Operation: Bowl Rinse Spout Press and release the bowl rinse button for a timed operation. The factory preset is a 30 second rinse. Press and release the bowl rinse button twice in less than two seconds to activate the continuous operation mode. Press and release the button again to end the continuous mode. Bowl Screen Press and hold the bowl rinse button for manual operation. Cupfill Operation: Press and release the cupfill button for a timed operation. The factory preset is a 2.5 second fill. Press and hold the cupfill button for manual operation. Note: Touchpad symbols are proprietary to A-dec Inc. Support Center 6

Cuspidor Functions (continued) Bowl Rinse Cupfill Program Button Customize Bowl Rinse or Cupfill Timing To reprogram the timed bowl rinse or cupfill functions: 1. Press on the touchpad, or press and hold the cupfill and bowl rinse buttons under the cupfill spout. Release the button(s) when you hear one beep. 2. Within five seconds, press and hold the cupfill or bowl rinse button for the desired time. 3. Release the button. Three beeps confirm that the program has changed. Support Center Cover Bowl Rinse Spout Bowl Rinse Flow Adjustment The bowl rinse function should provide enough water to completely rinse the cuspidor bowl. If the flow is not sufficient, an adjustment can be made inside the support center: 1. Remove the side cover from the support center by pulling the lower edges of the cover out. 2. With the bowl rinse on, squeeze the pinch valve to reduce the flow. Release the valve to increase the flow. 3. For the best rinsing action, adjust the flow pattern by rotating the bowl rinse spout. Bowl Rinse Flow Pinch Valve 86.0308.00 Rev E 7

A-dec 300 Support Systems Instructions for Use Operate / Adjust Telescoping Arm Adjustment To level the telescoping arm: 1. Extend the arm. 2. Locate the adjustment setscrew underneath the arm. 3. Use a 1/8" hex key and turn the setscrew clockwise to tilt the arm up, counterclockwise to tilt the arm down. 8

Clean / Maintain Barrier Protection NOTE For proper use and disposal of barriers, see the instructions provided by the barrier manufacturer. Waterline Maintenance Self-Contained Water Bottle A-dec recommends barrier protection for all applicable touch and transfer surfaces. Touch surfaces are areas that come into contact with hands and become potential cross-contamination points during dental procedures. Transfer surfaces are areas that come into contact with instruments and other inanimate objects. In the USA, barriers must be produced under the Current Good Manufacturing Practice (CGMP) as specified by the U.S. Food and Drug Administration (USFDA). For regions outside the USA, refer to the medical device regulations specific to your location. IMPORTANT For recommendations on cleaning and chemical disinfection of touch and transfer surfaces (where barrier protection is not applicable or when barriers are compromised), please see the A-dec Equipment Asepsis Guide (p/n 85.0696.00). A-dec recommends that you perform a shock treatment on the dental unit waterlines before you use the system for the first time. For on-going waterline maintenance, A-dec recommends a three-part protocol: continuous use of A-dec ICX waterline treatment tablets, regular dental unit water monitoring, and shock treatment. IMPORTANT For complete details on how to maintain good water quality in your system, see the instructions for use provided with ICX and your self-contained water bottle: A-dec ICX... p/n 86.0613.00 A-dec Self-Contained Water System... p/n 86.0609.00 Waterline Maintenance Guide... p/n 85.0983.00 86.0308.00 Rev E 9

A-dec 300 Support Systems Instructions for Use Clean / Maintain Cuspidor and Drain Spouts and Bowl The contoured spouts and smooth bowl of the cuspidor provide for quick and easy cleaning. Remember to empty and clean the bowl screen every time you clean the cuspidor. CAUTION Do not remove the spouts when cleaning the cuspidor. This will help prevent cleaning solutions from damaging the equipment and contaminating the cupfill water. CAUTION Do not empty the solids collector screen or bowl screen into the cuspidor. Doing so could plug the cuspidor drain. Dispose of the contents in an appropriate biohazard container according to the regulations specific to your location. After cleaning, always install the bowl screen in the cuspidor bowl to prevent debris from plugging the drain. Gravity Drain Tubing At the end of each day, flush the cuspidor to remove debris from the flexible drain tubing. If the cuspidor is not flushed regularly, debris may build up and impair draining. To flush the cuspidor, run the bowl rinse for about 60 seconds. Maintenance Parts For replacement cuspidor bowl screens, contact your authorized A-dec dealer and order p/n 75.0035.03 (package of 5). 10

HVE/Saliva Ejector Valve Body Assembly Asepsis NOTE Vacuum lines should be cleaned at the end of each day by evacuating a detergent or water-based detergentdisinfectant through the system. If you use barrier film to protect the main body of the HVE and saliva ejector, replace the barrier film after each patient. Clean and disinfect these instruments at the end of each day or after each patient if the barrier film has been compromised. If you do not use barrier film, or if you perform oral surgery, always clean and disinfect these instruments after each patient. Disinfection To disinfect the HVE and saliva ejector valve body assembly without disconnection from the vacuum system: 1. Remove the vacuum tip from the HVE or saliva ejector. 2. Disinfect the valve body assembly by using an intermediate-level disinfectant according to instructions for the disinfectant; A-dec recommends a phenolic (dual) water-based disinfectant. 3. Follow the manufacturer s instructions for disinfectant contact time. 4. Operate the HVE and the saliva ejector valves several times to verify that they rotate smoothly. 86.0308.00 Rev E 11

A-dec 300 Support Systems Instructions for Use Clean / Maintain HVE/Saliva Ejector Valve Body Assembly Asepsis (continued) NOTE A-dec valve body assemblies are heat tolerant for sterilization. Users may sterilize at their discretion. CAUTION Do not attempt to remove the HVE body or valve if you have an Air Vacuum System. The HVE is permanently attached to the tubing in this configuration and cannot be removed for sterilization. Instead, use barrier film produced under FDA good manufacturing practices for protection on the HVE body, making sure to replace the barrier after each use. Routine Maintenance and Sterilization To routinely maintain and sterilize the HVE and saliva ejector when disconnected from the vacuum system (frequency determined by the user): 1. Turn off the vacuum or open the control valve before disconnecting the HVE or saliva ejector. 2. Remove the vacuum tip from the HVE or saliva ejector. Discard disposable vacuum tips. 3. Remove the HVE or saliva ejector valve body assembly from the vacuum line by pulling it away from the tubing at the tailpiece. 4. Disassemble the valve body assembly by pushing the control valve out of the HVE or saliva ejector valve body. 5. Inspect the O-rings. Remove and discard if damaged. 6. Clean undamaged O-rings with a soft brush and a health authority approved detergent. O-rings may be cleaned in place or removed and cleaned at the discretion of the user. 7. Clean all exterior and interior surfaces of the valve body and control valve using a health authority approved detergent and the brushes provided with the product (or equivalent). Lumens or channels of the valve body must be cleaned with a brush of the proper length and diameter. The brush must be long enough to extend through the lumen. 8. Rinse thoroughly with water. 12

HVE/Saliva Ejector Valve Body Assembly Asepsis (continued) 9. Process the disassembled valve body assembly components through a washer-disinfector. If a washer-disinfector is unavailable, valve body assembly components must be manually disinfected using a health authority approved disinfectant. 10. Allow the components to dry completely. 11. Sterilize the valve body assembly components at 132-134 C (270 273 F) for 4 minutes in a pre-vacuum sterilizer or 6 minutes in a gravity displacement sterilizer. 12. Prior to use, replace O-rings as needed, lubricate O-rings with A dec silicone lubricant, and reassemble the valve body assembly. CAUTION Use only silicone lubricant when lubricating instrumentation O-rings. Petroleum-based products will cause permanent damage to the O-rings. 13. Reinstall the valve body assembly on the tubing tailpiece. 14. Operate the HVE and the saliva ejector control valves several times to verify that they rotate smoothly. 86.0308.00 Rev E 13

A-dec 300 Support Systems Instructions for Use Clean / Maintain HVE/Saliva Ejector Tip Asepsis NOTE Select tips that are compatible with your HVE and saliva ejector openings. See HVE and Saliva Ejector Openings for Tips on page 17 for specifications. Disposable Tips Replace disposable HVE and saliva ejector tips after each use. CAUTION Disposable HVE and saliva ejector tips are not sterilizable and should not be reused. Sterilizable Tips Heat sterilize stainless steel HVE tips after each use. To clean and sterilize a stainless steel HVE tip: 1. Remove the tip from the HVE. 2. Clean and rinse the tip using a mild detergent and water, then allow the tip to dry completely. 3. Sterilize the tip at 132-134 C (270 273 F) for 4 minutes in a prevacuum sterilizer or 6 minutes in a gravity displacement sterilizer. 14

Solids Collector Cap Support Center The solids collector aids in preventing solids from entering the central vacuum system. To ensure proper suction from the central vacuum and maintain proper treatment room asepsis, discard and replace the solids collector screen at least twice a week. Screen Arm-Mounted Systems TIP For solids collector replacement screens, contact your local authorized A-dec dealer and reference Pinnacle p/n 5512 or A-dec p/n 11.1284.01. To replace the solids collector screen: 1. Turn off the vacuum or open the HVE control valve. 2. Remove the solids collector cap and screen. 3. Discard the screen according to your local regulations. DANGER Infectious waste may be present. Follow asepsis protocol to prevent cross contamination. CAUTION Do not empty the screen into your cuspidor. Doing so could plug the drain. 4. Inspect the collector cap O-ring, replace the O-ring as needed, and lubricate with A-dec silicone lubricant as needed. 5. Insert a new screen in the collector and replace the cap. 86.0308.00 Rev E 15

A-dec 300 Support Systems Instructions for Use Clean / Maintain Utilities and Shutoff Valves Quad Outlet Box Drain Moisture Separator Vacuum Line The utilities for A-dec chair-mounted systems are located in the contoured floor box under your chair. To access, lift the cover up and off the floor box frame. CAUTION When removing or replacing covers, take care not to damage any wiring or tubing. Verify that the covers are secure after replacing them. Water Manual Shutoff Valve Filter Housing O-Ring Pre-Regulator Assembly Filter Beveled Edge Air Manual Shutoff Valve Manual Shutoff Valve CAUTION To ensure proper operation, install the filter with the beveled edge facing the manifold. The manual shutoff valves control the air and water to the system. Leave these valves completely open (turned counterclockwise) during normal use to prevent leaks. From the valves, air and water pass through separate filters before entering the pre-regulators. Replace these filters when they become clogged and restrict flow. To check for a clogged water filter: 1. Turn on the master toggle. 2. While watching the cuspidor, press the bowl rinse button. If the bowl rinse water flow fades or stops, replace the water filter. To replace the filter: 1. Turn off the master toggle and close the shutoff valve (turn clockwise). 2. Bleed the system of air and water pressure by operating the syringe buttons until air and water no longer flow. 3. Using a standard screwdriver, remove the filter housing from the water pre-regulator assembly and remove the filter. 4. Replace the filter if it is clogged or discolored. Install the filter with the beveled edge facing the manifold. 5. Open the shutoff valves, turn on the master toggle, and operate the bowl rinse to remove air from the waterline. 16

A-dec 300 Support Systems Instructions for Use Specifications / Warranty Specifications Warranty Warranty information is provided in the Regulatory Information, Specifications, and Warranty document (p/n 86.0221.00), which is available in the Document Library at www.a-dec.com. Minimum Air, Water, and Vacuum Service Requirements Air: 2.5 scfm (71 sl/min) at 80-125 psi (550-860 kpa) minimum Water: 1.5 gpm (5.7 L/min) at 60±20 psi (410±140 kpa) minimum Vacuum: (wet): 9 scfm (255 sl/min) at 10±2 inches of Hg (34±7 kpa) minimun (dry): 12 scfm (340 sl/min) at 4.5±1 inches of Hg (16±3.5 kpa) minimum HVE and Saliva Ejector Openings for Tips Standard HVE: 11 mm Large Bore HVE: 15 mm Saliva Ejector: 6 mm IMPORTANT For electrical specifications, identification of symbols, and other regulatory requirements, see the Regulatory Information, Specifications, and Warranty document (p/n 86.0221.00), which is available in the Document Library at www.a-dec.com. NOTE Specifications are subject to change without notice. Requirements may vary depending on your location. For more information, contact your authorized A-dec dealer. 86.0308.00 Rev E 17

A-dec Headquarters 2601 Crestview Drive Newberg, Oregon 97132 United States Tel: 1.800.547.1883 within USA/CAN Tel: +1.503.538.7478 outside USA/CAN Fax: 1.503.538.0276 www.a-dec.com A-dec Australia Unit 8 5-9 Ricketty Street Mascot, NSW 2020 Australia Tel: 1.800.225.010 within AUS Tel: +61.(0).2.8332.4000 outside AUS A-dec China A-dec (Hangzhou) Dental Equipment Co., Ltd. 528 Shunfeng Road Qianjiang Economic Development Zone Hangzhou 311106, Zheijiang, China Tel: 400.600.5434 within China Tel: +0571.89026088 outside China A-dec United Kingdom EU Authorized Representative Austin House, 11 Liberty Way Nuneaton, Warwickshire CV11 6RZ England Tel: 0800.ADEC.UK (2332.85) within UK Tel: +44.(0).24.7635.0901 outside UK ÍvÈ.Ç#(È.00SÎ 86.0308.00 Rev E Copyright 2017 A-dec Inc. All rights reserved. IFUcov4