Optimizing the Closed Suction Surgical Drainage System

Similar documents
BARD WOUND DRAINAGE. Products to meet your different procedural needs CLOSED WOUND SUCTION GRAVITY DRAINS SUMP DRAINS

Impact Of Clinical Alarms On Patient Safety. American College of Clinical Engineering Healthcare Technology Foundation (AHTF)

Post-operative wound drainage High-vacuum drainage Low-vacuum drainage Gravity drainage

Fluid Management. On-Demand Summary Printouts Set-Up & Non Pulsatile Flow. For Use In GYN, Urology & Arthroscopy. 42 allenmedical.

Hemovac Wound Drainage Family. The Time-Tested System for Reliable Wound Drainage and Auto-Transfusion

CBC II Blood Collection & Reinfusion System. Installation and use in Operating Room

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

Application of two hybrid control methods of expansion valves and vapor injected compression to heat pumps

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

THE LASER COURSE Medical Laser Safety Officer & Surgical Operator/Assistant

Thermal Properties and Temperature

Analysis and Evaluation of a Passive Evaporative Cool Tower in conjunction with a Solar Chimney

Easy Irrigation Tower #O-LPA $3,145

Alarm Fatigue in the ICU. Lynn Maguire MS, RN, PMC, NE-BC

Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency.

Green Infrastructure, Human Health Benefits, & Economic Value: a life course approach

Cryolipolysis fat dissolving device manual

Wound Drainage. Redax quality system is assured by ISO90001:2008 and ISO 13485:2000 certifications audited by TUV SUD (0123).

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

Solution: Problem: How can the air, and surfaces, be cleaned to lower the risk from the usual suspects? OR PACKAGE

Thoracic drainage upgrade:

PERFORMANCE OF VCRS SYSTEM WITH HEAT EXCHANGER AND PHASE CHANGE MATERIAL

Caring for your Drain Tube at Home

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

Healthy Buildings 2017 Europe July 2-5, 2017, Lublin, Poland

Alarm Management at Boston Medical Center A Roadmap to Safe Silence

The Contra Costa EMS System and Fire Station Closures: Impact and Mitigation

Performance Enhancement of Refrigeration Cycle by Employing a Heat Exchanger

UVC in the Operating Room.

Protect the Pump! Worksheet

EFFECT OF COMPACTION ON THE UNSATURATED SHEAR STRENGTH OF A COMPACTED TILL

TRANSMISSIVITY BEHAVIOR OF SHREDDED SCRAP TIRE DRAINAGE LAYER IN LANDFILL COVER SYSTEM *

Oral Syringe Product Stability &

THREE DIMENSIONAL SLOPE STABILITY

Chapter 22. Alternatives to the Proposed Project

The Ottawa Hospital OPERATING ROOM NURSING POLICY, PROCEDURE, PROTOCOL MANUAL

AN ANALYSIS OF THE PERFORMANCE OF RESIDENTIAL SMOKE DETECTION TECHNOLOGIES UTILIZING THE CONCEPT OF RELATIVE TIME

Modeling of Ceiling Fan Based on Velocity Measurement for CFD Simulation of Airflow in Large Room

Medical devices and electromagnetic compatibility. Addressing the requirements of IEC , 4 th Edition

Development of Gas-Liquid Two-Phase Flow Distributor for Improving Energy Efficiency in Air-Conditioners

You finally make it home after a long shift, shower,

LASER SAFETY TRAINING CATALOG LASER LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY.

2. HEAT EXCHANGERS MESA

7/27/2017. Surgical Fires. Prevention, Suppression and Evacuation. Importance of Preparedness. 1. Why is this so important?

ATMOS Surgical Suction

Combination unit to support instruction in Thermodunamics, Fluid Mechanics, and Heat Transfer

USER MANUAL For Clinicians. Rx Only

Surgical Suction SURGICAL SUCTION

1.1. SCOPE OF THE WORK:

Operation of a Two-Phase Reverse Loop Thermosyphon

UNITS /2017-E S-PILOT. The solution for smoke evacuation

Global Journal of Engineering Science and Research Management

Battery Performance Alert

Experimental Study on Match for Indoor and Outdoor Heat Exchanger of Residential Airconditioner

High-efficiency Turbo Chiller (NART Series)

How should electronically tagged patients be handled particularly regarding diathermy?

SimpleVIS Viscometer Reference Guide P

Introductions. Safe for cultivation

Effect of the Use Pattern on Performance of Heat Pump Water Heater

Shell-and-Tube Heat Exchanger (Four Passes) - Optional

Evaluation of Sorbents at Elevated Temperatures for Downhole Application

Experimental Investigate on the Performance of High Temperature Heat Pump Using Scroll Compressor

Evaluation of the Development of Capillary Barriers at the Interface between Fine-grained Soils and Nonwoven Geotextiles

Pressure drop analysis of evaporator using refrigerants R-22, R-404A and R-407C

A Building Area: 10,567 SF 20,321 SF total building


Clinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care

Conductive Heating Technologies for Medical Diagnostic Equipment. Russell Strehlow, Manager of Research and Development, Minco

INSTALLATION INSTRUCTIONS HOT GAS BYPASS SYSTEM DESIGN MANUAL

COMPARISON OF THE EXPERIMENTAL PERFORMANCE OF A THERMOELECTRIC REFRIGERATOR WITH A VAPOUR COMPRESSION REFRIGERATOR

Troubleshooting/Access/Alarms Simulation. Wendy McCay, MPH, BSN, OCN Brenda Stofan, RN

ST. JOHN HEALTH SYSTEM PHYSICIAN SAFETY MANUAL

MSU Landscape Architecture Program

Developing a Fire Test Strategy for Storage Protection Under Sloped Ceilings

What extinguishes fires without pressure peaks?

Australian Journal of Basic and Applied Sciences. Leak Detection in MDPE Gas Pipeline using Dual-Tree Complex Wavelet Transform

Experimental Investigation of a Brazed Chevron Type Plate Heat Exchanger

Experimental Study on Performance Parameters for Refrigerants R22, R410a and R404a at Various Air Outlet Temperatures

Development Of A Low-Cost Heat Pump Water Heater For Residential Applications

If These Walls Could Talk What Would They Say?

Healthy Buildings 2017 Europe July 2-5, 2017, Lublin, Poland. Experimental study of a novel dew point indirect evaporative cooler

Comparison of Thermal Comfort by Radiant Heating and Convective Heating

Instructions for Use Spinal Punches IFU

MEDIST s.r.o., Petrusovskeho 4, Humenne, Slovakia Tel.: , Fax: ,

Rocket Quick Guide. Rocket PSU Portable Suction Unit

EXPERIMENTAL PERFORMANCE INVESTIGATION OF SWIRLING FLOW ENHANCEMENT ON FLUIDIZED BED DRYER

EXPERIMENTAL INVESTIGATION OF COMPARISION OF AIR COOLED AND WATER COOLED CONDENSER ATTACHED WITH COOLING TOWER

A Review on Analysis of Vapour Compression Refrigeration System Using Matrix Heat Exchanger

How Much Malpractice Insurance Is Enough? p. 14 Cataract Case Costing to the Penny p. 32 Red Bag Waste Quiz p. 66

Design and Development of Water Cooled Condenser for Domestic Refrigerator

Recent BRANZFIRE enhancements and validation

Columbia St. Mary s Laser Credentialing

Subscripts 1-4 States of the given system Comp Compressor Cond Condenser E Evaporator vol Volumetric G Gas L Liquid

Development of a new X-ray source system using ultraviolet laser for medical treatment

Optimization of Wire Vacuum

Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR DEVICES UNDER BATTERY ADVISORY

CHAPTER I INTRODUCTION. In the modern life, electronic equipments have made their way

PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR

Thoracic drainage. Helping improve patient outcomes

Feasibility of a Liquid Desiccant Application in an Evaporative. Cooling Assisted 100% Outdoor Air System

Transcription:

Reconstructive Department Optimizing the Closed Suction Surgical Drainage System Katherine H. Carruthers, MS Bradley S. Eisemann, MD Susan Lamp, BSN, RN, CPSN Ergun Kocak, MD, MS CE Background: Closed suction drains are indicated in a wide array of postoperative settings, with many distinct drainage systems available to the surgeon. The purpose of this study was to compare the suction gradients Katherine H. Carruthers, MS, is a medical student at the University of Toledo College of Medicine in Toledo, Ohio. She recently earned a master s degree in Biomedical Sciences at the University of Toledo College of Graduate Studies and is currently completing the first year of her medical education. Bradley S. Eisemann, MD, graduated from University of Texas Medical School at Houston and is now a first-year resident in the Department of Surgery, Division of Plastic Surgery at the Baylor College of Medicine in Houston, Texas. Susan Lamp, BSN, RN, CPSN, is a plastic surgery nurse in the Department of Plastic Surgery at The Ohio State University s Wexner Medical Center in Columbus, Ohio. She is the editor of the ASPSNews and serves on the PSN Journal Editorial Board as the chair of the Reconstruction Department. Ergun Kocak, MD, MS, is an Assistant Professor in the Department of Plastic Surgery at The Ohio State University s Wexner Medical Center. He completed his residency in plastic surgery at The Ohio State University and continued his training with a microsurgical fellowship at the M.D. Anderson Cancer Center in Houston, Texas. The authors report no conflicts of interest. Address correspondence to Ergun Kocak, MD, MS, 915 Olentangy River Road, Suite 2100, Columbus, OH 43212 (e-mail: ergun.kocak@osumc.edu). DOI: 10.1097/PSN.0b013e31828425db achieved using 2 different sizes of suction reservoirs and 2 different techniques for generating negative pressure. Materials and Methods: Drainage reservoirs of 100 and 400 ml were chosen to eval uate their ability to achieve suction. Suction was established in both sizes of drains by pressing the sides of the reservoir together or by pushing the bottom of the reservoir toward the top. Negative pressures were recorded with the reservoir empty, and after every 10-ml addition of saline. Averages were graphed to illustrate the applied suction over a range of drain volumes. Results: The 100-ml drainage system reached a peak suction of 117.6 mmhg, while the 400-ml drainage system reached only a peak suction of 71.4 mmhg. Both of the maximum suction readings were achieved using the full-squeeze technique. The bottom-pushed-in technique did not result in any sustained measurable levels of suction using either of the reservoir volumes. Conclusions: Smaller drain reservoirs are more successful in generating a high initial suction than larger reservoirs, especially when the volume of fluid in the drain is relatively low. In all sizes of drains, compressing the sides of the reservoir is a far better technique for establishing negative pressure than pressing the bottom of the drain up toward the top. Closed suction surgical drainage systems are routinely used when there is a postoperative risk of seroma development (Zawaneh & Putnam, 2008). Because of the frequent use of surgical drains, several different variations of drains have been developed, with a range of reservoir volumes and shapes. However, despite some physicians preferences to use larger drainage reservoirs, many patients comment that they find these larger bulbs to be cumbersome and inconvenient during their postoperative recovery, and that they would have preferred smaller reservoirs if they could be confident that they provided comparable drainage. Furthermore, there are several methods of compressing these reservoirs to generate the negative pressure needed to pull collected fluid from the surgical wound (Figure 1). In the plastic surgical setting, drainage systems are especially important when performing certain operations, such as latissimus muscle flaps or breast reconstructions that require elevating abdominal 38 Plastic Surgical Nursing January March 2013 Volume 33 Number 1

Figure 1. Actual photo of a patient with two 100-ml Jackson-Pratt drains placed in the abdomen. It can be observed that the drain with the side-to-side full-squeeze method for obtaining closed suction has begun to collect serous fluid, while the drain with the bottom-pushed-in method still appears to be empty. tissue, as these procedures are known to have especially high rates of seroma formation (Scevola, Youssef, Kroll, & Langstein, 2002). Seromas that are not properly drained can cause delayed wound healing or even flap necrosis. Thus, it is vital to ensure that the surgeon s technique for applying suction and choice of drain reservoir volume maximize the ability of the drainage system to apply negative pressure to the potential space. Because of the wide range of reservoir sizes and methods for establishing negative pressure, it must be wondered whether certain combinations are more successful than others at drawing fluid out of the surgical wound. The ability to effectively remove fluid more completely could potentially reduce complications traditionally associated with flap elevation and even facilitate patient recovery (Durai, Mownah, & Ng, 2009). The goal of this study was to objectively analyze the suction generated by drains with two common reservoir sizes and to compare the negative pressure attained by two frequently observed techniques for achieving closed suction. Plastic Surgical Nursing January March 2013 MATERIALS AND METHODS Air was removed from 100-ml Bard (Bard Medical Division, Covington, GA) and 400-ml Jackson-Pratt (Baxter Healthcare Corp., Deerfield, IL) surgical drain reservoirs by either squeezing the sides together (Figure 2a) or pushing the bottom toward the top (Figure 2b). Of the two ports on the top of each drain reservoir, one port was connected to 80 cm of 19 French reservoir tubing with a three-way stopcock (Bard Medical Division), allowing for saline to be injected into the drain reservoir in known volumes. The other port was connected with similar tubing to a sphygmomanometer (Figure 3). This device measures pressures ranging from 20 to 300 mmhg. Suction readings were repeated five times, and measurements were obtained to the nearest 1.0 mmhg with the drain empty and after each subsequent injection of 10 ml of saline until the reservoir reached capacity or until level of suction dropped below the measurable value of 20 mmhg. Average negative pressures for each size of reservoir and for each method of removing air from the Figure 2. Air was removed from surgical drain reservoirs to generate negative pressure by either (a) squeezing the sides together or (b) pushing the bottom toward the top. Volume 33 Number 1 39 PSN200241.indd 39 13/02/13 2:06 AM

Figure 3. After removal of the air from the drain reservoir, one of the two ports on the top was connected to tubing with a three-way stopcock, allowing for saline to be injected into the drain reservoir in known volumes. The second port was connected with similar tubing to a sphygmomanometer to measure the suction of the drainage system. reservoir were calculated with the drain empty and after each addition of saline. The resulting data were graphed to illustrate the applied suction achieved over a range of drain collection volumes (Figures 4 and 5). RESULTS Using the side-to-side full-squeeze technique, drains with 100-ml reservoirs reached an average maximum negative pressure of 117.6 mmhg (Table 1a). This value was achieved with the reservoir empty and the suction slowly dropped as the volume of retained saline increased. By the time the reservoir was filled to 60 ml of saline, the suction had dropped below a recordable level. Similar results were obtained using the side-toside full-squeeze technique and the 400-ml reservoir. Using the 400-ml reservoir, the average maximum negative pressure was 71.4 mmhg, again achieved when the drain was empty (Table 2a). A p value of less than.05 was calculated when comparing the maximum suction obtained when using the 100-ml and 400- ml reservoirs and the side-to-side full-squeeze technique. As with Figure 4. Using a 100-ml Bard drainage system, average suction was measured using either the full side-to-side squeeze or the bottom-pushed-in technique of generating negative pressure. Measurements were repeated five times and means were reported the standard deviation. the 100-ml drain, suction slowly decreased as the reservoir was filled; however, the 400-ml reservoir maintained measurable suction until it contained an average of 130 ml of saline. Using the bottom-pushed-in technique on the 100-ml drainage system, no measurable negative pressures were recorded until the suction jumped to an average of 24.0 mmhg when the reservoir contained 70 ml of saline (Table 1b). From this point, suction maintained a relatively stable level until the drain reached capacity. In contrast, the 400-ml drainage system never reached true measurable levels of negative pressure using the bottompushed-in technique at any point while the reservoir was filled (Table 2b). DISCUSSION Because of the wide variety of surgical drainage systems currently available on the market, there is little consensus among surgeons as to which drains and which techniques for generating negative pressure are most effective (Swartz et al., 2012). However, on the basis of our findings, we concluded that smaller drain reservoirs were able to generate more suction than larger reservoirs and that compressing the sides of the drain bulb was the most effective technique for evacuating the air to generate negative pressure. In 2002, Grobmyer and colleagues published a study that attempted to determine the suction gradients of three commonly employed surgical drains (Grobmyer, Graham, Brennan, & Coit, 2002). Among the three types of drains analyzed was the 400-ml Jackson-Pratt closed suction system that was also analyzed in our current study. In accordance with our data, Coit determined that the maximum suction was achieved when the reservoir was empty and that negative pressure 40 Plastic Surgical Nursing January March 2013 Volume 33 Number 1

Figure 5. Using a 400-ml Jackson-Pratt drainage system, average suction was measured using either the full side-to-side squeeze or the bottom-pushed-in technique of generating negative pressure. Measurements were repeated five times and means were reported the standard deviation. decreased as the reservoir filled. While it was not clear what technique was used to eliminate the air from inside the reservoir to generate the reported suction gradient, the authors were able to conclude that different types of closed suction surgical drainage systems do achieve different levels of negative pressure. Whitson and colleagues (2009) further explored the topic of comparing different types of closed suction drains. In this study, the investigators compared not only drains from different manufacturers but also drains with different reservoir volumes (Whitson, Richardson, Iaizzo, & Hess, 2009). They determined that, when empty, 100-ml reservoirs were able to achieve a higher level of suction than 400 ml drain reservoirs. As before, our data verified this conclusion when using the side-to-side full-squeeze technique for generating suction. Although both of these studies did address the questions surrounding the commercially available drain options, neither discussed the variation in negative pressure that could be caused by the technique used when applying suction to the drainage system. In 2009, Halfacree, Wilson, and Baines (2009) published the first study to address how user technique could affect the efficiency of a surgical drainage system. In this report, the investigators compared one-hand and two-hand compression methods for removing the air from the drain reservoir to generate suction. Regardless of the size of the drain reservoir, it was determined that the twohand compression method caused significantly greater negative pressure than the one-hand method. However, the one-hand method did still generate a measurable level of suction. While our study analyzed two different techniques for generating suction, it was concluded that the bottom-pushed-in technique did not generate any measurable level of suction while the side-to-side full-squeeze method demonstrated a much more clinically useful range of negative pressures as the reservoir filled. TABLE 1. Negative Pressures of 100-ml Drainage System Volume of Saline Added (ml) 0 10 20 30 40 50 60 70 80 90 100 A Full squeeze B Average negative 11 7.6 69.0 47.6 34.6 23.2 20.5 20.0 20.0 20.0 20.0 20.0 Standard deviation 16.8 10.9 13.0 8.4 3.9 1.0 0.0 0.0 0.0 0.0 0.0 Bottom pushed in Average negative 20 20 20 20 20 20 20 24.0 28.3 25.7 26.0 Standard deviation 0.0 0.0 0.0 0.0 0.0 0.0 0.0 4.0 5.4 0.6 0.0 TABLE 2. Negative Pressures of 400-ml Drainage System Volume of Saline Added (ml) 0 10 20 30 40 50 60 70 80 90 100 110 120 130 A Full squeeze B Average negative 71.4 58.4 48.8 41.6 36.0 26.6 26.8 26.0 24.6 22.4 21.8 21.0 20.5 20.0 Standard deviation 8.8 8.2 6.5 5.6 4.2 13.5 2.9 1.6 2.1 2.3 1.7 1.4 1.3 0.0 Bottom pushed in Average negative 20 20 20 20 20 20 20 20 20 20 20 20 20.5 20 Standard deviation 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.7 0.0 Plastic Surgical Nursing January March 2013 Volume 33 Number 1 41

Our study, in conjunction with these previous reports, clearly illustrates the massive variation in suction levels that can occur depending on the size and type of the drainage system, as well as the method used to generate the negative pressure. When the highest levels of suction are needed to prevent seroma development, a surgical drainage system with a small reservoir should be chosen and the surgeon should compress the reservoir in a side-to-side, two-handed manner to evacuate the air. In addition, since it was observed that drain suction decreased as the volume of fluid in the reservoir increased, drains should be emptied as frequently as possible to maintain the highest possible level of suction. At a minimum, since the drains used in this study lost any measurable amount of suction when the reservoir was approximately half full, care providers must empty the drain reservoirs before they reach this point, so that they can continue to pull fluid from the wound bed. If high outputs preclude timely emptying of fluid from the drain reservoirs, larger collection devices, such as the 400-ml container used in this study, should be considered at the small expense of slightly lower suction pressures. Although research is slowly filling in the gaps in our knowledge about the dynamics of surgical drain suction, further studies are still needed. It would be beneficial to address how, or to what degree, the length of the reservoir tubing effects the level of suction in the wound bed and how placement of multiple drains effects the total negative pressure generated in the potential space. In summary, smaller drain reservoirs are more successful in generating a high initial suction than larger reservoirs, especially when the volume of fluid in the drain is relatively low. However, larger bulbs do tolerate higher volumes of fluid collection and allow for longer times to elapse before emptying. In all sizes of drains, compressing the sides of the reservoir is a far better technique for establishing negative pressure than pressing the bottom of the drain up toward the top. REFERENCES Durai, R., Mownah, A., & Ng, P. C. (2009). Use of drains in surgery: a review. Journal of Perioperative Practice, 19(6), 180 186. Grobmyer, S. R., Graham, D., Brennan, M. F., & Coit, D. (2002). Highpressure gradients generated by closed-suction surgical drainage systems. Surgery of Infectious, 3(3), 245 249. Halfacree, Z. J., Wilson, A. M., & Baines, S. J. (2009). Evaluation of in vitro performance of suction drains. American Journal of Veterinary Research, 70(2), 283 289. Scevola, S., Youssef, A., Kroll, S. S., & Langstein, H. (2002). Drains and seromas in TRAM flap breast reconstruction. Annals of Plastic Surgery, 48(5), 511 514. Swartz, A. L., Azuh, O., Obeid, L. V., Munaco, A. J., Toursavadkohi, S., Adams, J., et al. (2012). Developing an experimental model for surgical drainage investigations: An initial report. American Journal of Surgery, 203(3), 388 391; discussion 391. Whitson, B. A., Richardson, E., Iaizzo, P. A., & Hess, D. J. (2009). Not every bulb is a rose: A functional comparison of bulb suction devices. Journal of Surgical Research, 156(2), 270 273. Zawa neh, P. N., & Putnam, D. (2008). Materials in surgery: A review of biomaterials in postsurgical tissue adhesion and seroma prevention. Tissue Engineering Part B Reviews, 14(4), 377 391. 42 Plastic Surgical Nursing January March 2013 Volume 33 Number 1