CALIBRATION HOW IT WORKS

Similar documents
CALIBRATION HOW IT WORKS

CALIBRATION HOW IT WORKS

POWER & ENERGY METERS BEAM PROfILING THZ MEASUREMENT. LASER BEAM MEASUREMENT & BEAM PROfILING

Understanding total measurement uncertainty in power meters and detectors

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

Infrared Thermometer with 12:1 Distance-to-Sight Ratio and Temperature Alarm

COMPACTNESS 34.1 PERFORMANCE ANTICIPATION

COMPACTNESS 34.1 PERFORMANCE ANTICIPATION

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS

2 Ensure that all personnel in the area are wearing the appropriate protective eyewear.

mach 5 100kHz Digital Joulemeter: M5-SJ, M5-IJ and M5-PJ Joulemeter Probes

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

POWER & ENERGY METERS BEAM PROFILING THZ MEASUREMENT COMPANY PROFILE

PRESENTATION OVERVIEW OF THE DIFFERENT MODELS MONITORS

PRESENTATION OVERVIEW OF THE DIFFERENT MODELS MONITORS

C-NRPP Quality Assurance Guidance for Radon Test Devices

POWER & ENERGY METERS BEAM PROFILING THZ MEASUREMENT COMPANY PROFILE

API MANUAL OF PETROLEUM MEASUREMENT STANDARDS

Atomic Absorption Spectrophotometer AAS 2320D EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES

Equipment specifications (D6)

Low-Power Detectors. Key Features. 818 Series. Power levels from pw to 2W. Wavelengths from 190 nm to 1800 nm. Proprietary detector optics

Publication American Institute of Physics. Reprinted with permission.

LaserPoint srl - Via Burona, Vimodrone (Milano) - Italy Phone Telefax

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Construction of a Calibration System for Power Responsivity of Optical detectors Using Different Laser Sources

PRESENTATION OVERVIEW OF THE DIFFERENT MODELS MONITORS

ELD500 PRECISION LEAK DETECTOR

818 Series. Low-Power Detectors ABSOLUTE SPECTRAL RESPONSIVITY (A/W)

In Search of Standard Reference Materials (SRMs) for ASTM E05 Fire Standards by. Norman Alvares & Harry Hasegawa

NTC THERMISTOR SENSOR PERFORMANCE

Laser Analyzers for NH3 Safety Monitoring and Leak Detection

16S-Series 150W. In Vivo SPF Sunscreen Testing Solar Simulator

POWER & ENERGY METERS BEAM PROFILING THZ MEASUREMENT COMPANY PROFILE

2100 Series Laboratory Turbidimeters. Pioneering technology in turbidity monitoring.

300/400 UNITS - MODULAR CONFIGURATION

Ultraviolet (UV) Monitoring System - Open Channel Horizontal Systems Operation and Maintenance Manual

SECTION ULTRASONIC TRANSIT TIME CLAMP-ON FLOW MEASURING SYSTEM

Spectrum Detector, Inc.

Models S310 & S310D Laser Power and Energy Meters Setup and Operating Procedures

LM-80V , h V LM-30V h V

7.3 Design and Development Design and development planning Design and development inputs Design and development outputs 7.3.

Instruction Manual. gray Housed Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

ISO INTERNATIONAL STANDARD. Coating powders Part 13: Particle size analysis by laser diffraction

RaySafe Xi Classic. A complete system for X-ray measurements

Technical Information OUSAF44

818 Series Low-Power Detectors

Paul Kotrappa, Ph.D. and Carolyn Allen. Rad Elec Inc., 5714-C Industry Lane, Frederick, MD USA

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE

PSA Series. Particle Size Analyzer

Instruments Designed for POF & Fibre Optic Cable Testing

Polarized Light Scattering of Smoke Sources and Cooking Aerosols

APPLIED. OMA-406R Rackmount Process Analyzer

Beam Analysis nm and µm - Pyroelectric Array Camera. Pyrocam TM IIIHR & Pyrocam IV Series

Test Report for the determination of Critical Fall Height for Under Playground Surfacing with Plants

Hygrometry. The art or science of humidity observation

Key Features. General Description

An FT-NIR Primer. NR800-A-006

Technical Information OUSAF21

Gas Detection in Process Industry The Practical Approach

FS-200. Model. Instruction Manual. Infrared Food Thermometer. reedinstruments. www. com

Laser Automation &Safety. ME 677: Laser Material Processing Instructor: Ramesh Singh 1

CO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 RECOMMENDATION FOR USE

POWER METER INSTRUCTION MANUAL

Standard ECMA th Edition - December Declared Noise Emission Values of Information Technology and Telecommunications Equipment

Specifications Wavelength Range: 190 nm to 800 nm continuous

Model 6650SP. Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer

LSM Quadrant Source and Measure Unit for LED Production Tests. Two Global Leaders. One Complete Solution.

Experimental Study to Evaluate Smoke Stratification and Layer Height in Highly Ventilated Compartments

ProStar 325 UV-Vis Detector

Product data sheet Palas Fidas 200 S

APPLIED. OMA-300 Hydrogen Sulfide Analyzer H 2. Ex p. Ex d. The world s safest continuous H 2. S analyzer. The OMA continuously measures H 2

Investigation of the Potential Use of Blue Light in Forward Scattering Optical Smoke Chambers to Detect all UL217 Fires in the New Standard

Customers must fill in and mail the warranty card in order to activate the warranty.

APPLIED. OMA-206P Portable Process Analyzer. A window into your process since 1994.

The J22 and J23 series are high performance InGaAs detectors operating over the spectral range

In-line Oil in Water Analyser - Model OIW-EX100P and OIW-EX1000P

PCT Workshop Before you leave the office

metro B6012 VIBRATIONS SIMPLIFIED... Product brochure Easiest portable vibration analyzer and balancer TECHNOLOGIES PVT. LTD.

OMA-300 Process Analyzer

When Precision is not good enough

Instruction Manual. gray Mounted Detectors. for. greenteg AG Technoparkstrasse Zürich, Switzerland T: F:

Beam Analysis nm and µm - Pyroelectric Array Camera. Pyrocam TM IIIHR & Pyrocam IV Series

Technical Information OUSAF22

More Precision. thicknesssensor // The sensor for precise thickness measurements

UV CURING SYSTEMS OmniCure UV Curing Lamp & LED Systems TO ORDER, PLEASE CALL

On-line Side Stream Oil in Water Analyser - Model OIW-EX 100 or OIW-EX 1000

Lighting Measurement Solutions

Benefits & Technology for Decentralizing a -40 F Dew Point

Return loss measurement of fiber optic components

M15-GSL12P30 M15-GSL12P30S Heat Spy Infrared Thermometer User Manual

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

ISO INTERNATIONAL STANDARD

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

Functional Safety: the Next Edition of IEC 61511

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

The Viva-ProE System

Transcription:

HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS At Gentec-EO, we understand that the essence of our business since 40 years has been delivering accuracy. There are no half measures: it either measures accurately or it doesn t. This is why one of our company s values is rigorousness, because our customers expect nothing less. 6

HOW IT WORKS For each detector that we calibrate, 50 Parameters are collected and logged Our uncertainty values are based on ENERGY DETECTORS We use only GOLD Calibration Standards, guaranteeing our customers the lowest calibration uncertainty possible THE GENTEC-EO ADVANTAGE Proven Statistical Calculation Processes Our Personnal Wavelength CorrectionTM (PWC) data offers you NIST and/or NRC Traceability over the entire range of the detector The calibration reference is checked 2 to 3 Times during EACH calibration process TOTAL ACCURACY of your detector POWER DETECTORS in our quality system Each of these steps contributes to the HIGH POWER SOLUTIONS THE TERMS The error on a measurement is the difference between the measurement result and the true value. REPEATABILITY REPRODUCIBILITY PRECISION The repeatability is the closeness of the agreement between the results of successive measurements under the same conditions of measurements. The reproducibility is the closeness of the agreement between the results of successive measurements under changed conditions of measurements. This is also defined as precision under reproducibility conditions. The precision of a measurement is defined as the closeness of agreement between independent test results obtained under stipulated conditions. Catalogue 2016_V2.0 T 418.651.80 03 1888 5G E N TE C F 418.651.1174 i nfo@gent ec -eo.c om Uncertainty is a measure of the goodness of a result. The definition and concept of uncertainty is a quantitative attribute to the final result of measurement, considering all systematic and random components of all known input quantities. The accuracy of a measurement is defined as the closeness of the agreement between the result of a measurement and the true value. ERROR UNCERTAINTY ACCURACY.com 7

HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS HOW IT WORKS 8 THE PROCESS 1 2 Comparison Process ACCURACY Gold Standard 4 Certification 3 Uncertainty Calculation

THE TECHNIQUE By definition, calibration is a comparison between measurements, one of a known magnitude or correctness, which is typically called a gold standard, and another measurement comparable to the first one. In the calibration process, there are four critical aspects that need to be controlled precisely: COMPARISON PROCESS The first step in the calibration process is the comparison to a known and traceable standard. At Gentec-EO, we always do this using Gold and not a Silver calibration standards, unlike some of our competitors. This extra carefulness in the comparison process comes from decades of experience in the laser measurement business. CERTIFICATION Gentec-EO has been using its own control quality system for many years and is now accredited ISO 9001:2008. Over and above the quality system certification process, the most important aspect is how rigorously the different steps and parameters are controlled in order to deliver an accurate calibration day after day. ENERGY DETECTORS POWER DETECTORS HIGH POWER SOLUTIONS GOLD STANDARD UNCERTAINTY CALCULATION Gentec-EO s gold laser power detector heads are compared to NIST standard calorimeters at different wavelengths, in accordance to the different lasers used to calibrate your own detector heads. The laser beam has a nominal diameter appropriate for the detector, and is centered on the detector s absorbing surface. The laser energy impinging upon the test instrument is measured concurrently using a NIST standard calorimeter and a calibrated beam splitter. The beam splitter ratio is measured using NIST standard calorimeters. Before the measurements are performed, the test instrument is allowed to reach equilibrium with the laboratory environment. The calibration factor is found by dividing the instrument output reading by the calculated average incident laser power. The calculation is based on the output reading of the NIST standard calorimeters. At Gentec-EO we offer the best uncertainty on the market, which means more than just giving the customer the lowest uncertainty value. These calculations also need to follow recognized statistical calculation standards, including those given in NIST s Technical Note 1927. Another important parameter to verify, and one that is less known, is the confidence level. At Gentec-EO, we use a very high confidence level of 95%. Like every other step in the calibration process, our uncertainty calculations are done rigorously. We don t aim to give you just the lowest number, whatever its meaning, we rather aim to give your the true value, with the highest confidence possible. Catalogue 2016_V2.0 T 418.651.8003 1888 5GENTEC F 418.651.1174 info@gentec-eo.com.com 9

HIGH POWER SOLUTIONS POWER DETECTORS ENERGY DETECTORS ELECTRICAL INSTRUMENTS All of our electrical instruments are calibrated by certified calibration suppliers. They certify that, at the time of calibration, the instruments used for calibration meet or exceed all published specifications and have been calibrated using standards whose calibrations are traceable to the NIST and/or other recognized international standards. The electrical and physical properties of their laboratories meet the highest requirements for ambient temperature, relative humidity and cleanliness. Their equipment is maintained by procedures that meet the requirements of ISO 9001:2000 and ISO 10012:2003. THE FACTS HOW GENTEC-EO CALIBRATES YOUR DETECTOR Every detector is individually calibrated to the best possible accuracy traceable to NIST standards. Stable laser sources at various wavelengths are used in our calibration process. UNCERTAINTY One very common misconception is the absolute value of calibration uncertainty. Be aware that this value is made using a complex statiscal method that takes in account ALL the sources of uncertainty that are present in the process. Figure 1 below shows these steps and their respective contribution to the value of uncertainty. As you can see, the manufacturer itself is only one of these sources. -3% -2.5% -2% Figure 1. Sources of uncertainty in the calibration of a detector -1.5% -1% -0.5% True Value NIST Reference (±0.3% to ±0.9%) Gentec-EO Gold Standard (±1%) (Calibrated by NIST) Your Gentec-EO Detector (±2.5%) (Calibrated by Gentec-EO using a Gold Standard) +0.5% +1% +1.5% +2% +2.5% +3% Gentec-EO Calibration Setup ±1.5% 10

WAVELENGTHS Another misconception is that any wavelength can be NIST calibrated. The NIST only supplies references for distinct wavelengths contained between 157 nm (F 2 excimer lasers) and 10.6 µm (CO 2 lasers). Every other wavelength within this range or out of this range is subject to an additional error. For more information about NIST s calibration wavelengths, please visit their website at: http://www.nist.gov/calibrations/cal-op.cfm PERSONAL WAVELENGTH CORRECTION TM CERTIFICATE To fill the gaps between the NIST references, Gentec-EO offers you the only NIST traceable calibration in nm steps, from 250 nm to 2.5 µm. We achieve this using our proprietary setup that is based on a NIST traceable spectrophotometer. This way, instead of supplying you with typical values, we offer you a NIST traceable calibration. What you get is an overall accuracy that is not more than ±1% away from the original calibration accuracy, in the spectrum of 300 to 2200 nm. Each Gentec-EO detector comes with a Personal Wavelength Correction Certificate. The correction factors are based on measurements that were made with YOUR detector. They are not based on the general curve of the absorbing material or the general response of equivalent products. This means you get the best wavelength correction tool available on the market. This data is stored in the Smart Interface of your Gentec-EO detector, you just have to select the wavelength in your monitor to get the most precise laser measurements on the market. ENERGY DETECTORS POWER DETECTORS HIGH POWER SOLUTIONS Your Detector Model Absorption Curve of your Detector 1 1 2 3 2 3 Wavelength Correction Wavelengths programmed in the EEPROM (nm) (Based on the Absorption Curve of your Detector) Calibrated Wavelength (nm) (Using a Gold Standard) Correction Factors (Multipliers) For more info, see Application Note 202184 - Understanding your Calibration Certificate. Catalogue 2016_V2.0 T 418.651.8003 1888 5GENTEC F 418.651.1174 info@gentec-eo.com.com 11