Smoke Particulate Monitors: 2006 Update

Similar documents
Three categories of instruments

AXTracker. An Inexpensive Satellite Transmitter for Sensing Changes at Remote Locations. United States Department of Agriculture.

United States Standards for Grades of Beets

Ecological Sites. Topics of Discussion 2/28/2014. Ecological Site Classification. Ecological Site Key. Ecological Site Descriptions

THE SCORE CARD. Observed Soil Properties. Interpretive Soil Properties

Integrating Nephelometer - Visibility and Particulate Monitoring. Summary

The New DUSTTRAK II and DRX Aerosol Monitors

The New DUSTTRAK II and DRX Aerosol Monitors

Seed Vacuum Pickup System: Operator s Manual

Certificate: / 29 April 2014

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

Air Resource Specialists, Inc.

Certificate: _01 / 19 August 2011

Rick Sherman Farm to School / School Garden Coordinator

_03 / 28 July 2015

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Product data sheet Palas Fidas 200 S

Marketing to Architects and Designers

Oregon Farm to School Childhood Obesity Prevention Summit Lebanon, OR - April 20, 2016

Marketing to Builders and Homeowners

NEX-BETA-ABG FACILITY Drinking Water Safety Monitor

CERTIFICATE. about Product Conformity {QAL 1) Number of Certificate: _02

TacBio: A UV-based Biological Aerosol Detector. TacBio Overview Page 1

Marketing to Builders and Homeowners

Marketing to Builders and Homeowners

HART XL High Sensitivity Smoke Detection

icount Oil Sampler (IOS) Portable Condition Monitoring for Hydraulic Oil and Fuel Systems

Finger Lakes National Forest. Backbone Horse Camp Water Hydrant Project

Ventilate Your Swine Nursery Building

IMPROVE Aerosol Monitoring Network

An Inexpensive Dual-Chamber Particle Monitor: Laboratory Characterization

LaserFOCUS VLF-250 Engineering Specification

THEORY OR OPERATION 2 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 OPTIONS 11

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

HART XL High Sensitivity Smoke Detection

The Shortwave Reflectance of the Yankee Environmental Systems Hotplate

APPROVED by U.S. EPA As a Class III Equivalent Method Designation for the measurement of PM 2.5

CHARCOAL CANISTER (Activated Charcoal Adsorption Device) by Proportional Counting

by Richard W. Bukowski Center for Fire Research Program for Fire Detection and Control Systems National Bureau of Standards Washington,

Parker icount Oil Sampler (IOS)

Fundamentals is subdivided into Sources of Radiation, Biological Effects, Mathematics, Chemistry, Physics, and Units and Terminology.

Maximize safety and productivity. Thermo Scientific Nuclear Power Radiation Detection and Monitoring Solutions

HPLC Column Ovens and Temperature Controllers

Instrumentation. Post RDD Event Urban Reoccupancy. James P. Menge PE. The world leader in serving science

SPARTAN Standard Operating Procedures

SMA Microbial Air Sampler System Product Guide

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Audit Report. Global GAW station Pallas, Finland

PL400 Patient Lift Scale Indicator Owner s Manual

Three Rivers Park District Administration Center Rain Garden

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

POLLINATOR HABITAT PLANTING

Simultaneous Real-Time Dust Mass Measurement with the

CERTIFICATE of Product Conformity (QAL1 )

FEATURE ARTICLE. The PLCA-800 Series Inline Particle Sensor for Chemical Solutions. Yoshihito Yuhara. Abstract

Collecting Dormant Hardwood Cuttings for Western Riparian Restoration Projects

Environmental Radiation Monitoring System

Smoke Detection in Low Gravity Results from the Smoke Aerosol Measurement Experiments (SAME) Conducted on the International Space Station

Grove - Laser PM2.5 Sensor (HM3301)

Polarized Light Scattering of Smoke Sources and Cooking Aerosols

Compact Sensor Heads for In-situ and Non-Contact Standoff Gas Sensing using TDLAS in Harsh Environments

Understanding total measurement uncertainty in power meters and detectors

Audit/Twinning Report

Bench Top Cooling Tower H893

The Arizona Wilderness and Urban Visibility Monitoring Networks

HEALTH AND SAFETY MANUAL

Portable Survey Instruments NISP-RP-01

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

A Method for Fire Detecting by Volume and Surface Area Concentration Based on Dual Wavelengths

Optec NGN-2 LED Ambient Integrating Nephelometer

An FT-NIR Primer. NR800-A-006

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS

Air Quality Sensors. Particle Matter, CO 2, Humidity, Temperature and Dust Products. OEM Sensor Components for Air Quality and Environmental Control

Mark 25 Ultrapure Water Conductivity Analyzer

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Davis Soil Moisture and Temperature Station Protocol

A Cost Effective Multi-Spectral Scanner for Natural Gas Detection

Wildland Fire Air Quality Response Program

TAHOE DOUGLAS FIRE PROTECTION DISTRICT JOB DESCRIPTION Fire Inspector Fire Prevention

OPERATING MANUAL. EchoTherm PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS IN35, IN45 and IN55-12VDC. DOCUMENT NUMBER IN35-03 Revised 15 May 2015

The Emergence of Triboelectric Technology

Evaluation of Technology Transfer to Rural Communities for Drying Using Lpg and Solar Energy Cabinet Dryer

ß-Aerosolmonitor FHT 59 C

Investigation of the Potential Use of Blue Light in Forward Scattering Optical Smoke Chambers to Detect all UL217 Fires in the New Standard

SECTION Process and Immersion Turbidity and Total Suspended Solids Sensor

AS950 PORTABLE SAMPLERS

HIGHLY SENSITIVE FOR NUCLEAR POWER PLANTS Real-Time Continuous Water Monitor

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

A Continuous Air Monitoring Sampler for 125 I - RIS-125

Summary. Assessment Methodology

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ELECTRET by Voltametry. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:

OWL. Laser OWL Series. Operations Guide. Singlemode Light Sources. Optical Wavelength Laboratories. Date: October 1, 2015

compressed air or left to consume air even when the associated production machine has been turned off.

Effect of the Electro Breeze Air Filter on Airborne Particles in a Recirculating Air Duct

-Industrial. W hen All Else Fails... Difficult Applications. The Only TRULY Industrial Air Sampling Detector For Hostile and Difficult Environments

Wildland fire phenomenology experiments:

Building and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes

OSU Master Gardener Program

Transcription:

United States Department of Agriculture Forest Service Technology & Development Program Smoke Particulate Monitors: 2006 Update 2500 Watershed, Soil, and Air December 2006 0625 2842 MTDC

Smoke Particulate Monitors: 2006 Update Andy Trent Project Leader USDA Forest Service Technology and Development Program Missoula, MT 6E62F46 Collocation Study of New Smoke Monitors December 2006 The Forest Service, United States Department of Agriculture (USDA), has developed this information for the guidance of its employees, its contractors, and its cooperating Federal and State agencies, and is not responsible for the interpretation or use of this information by anyone except its own employees. The use of trade, firm, or corporation names in this document is for the information and convenience of the reader, and does not constitute an endorsement by the Department of any product or service to the exclusion of others that may be suitable. The U.S. Department of Agriculture (USDA) prohibits discrimination in all its programs and activities on the basis of race, color, national origin, age, disability, and where applicable, sex, marital status, familial status, parental status, religion, sexual orientation, genetic information, political beliefs, reprisal, or because all or part of an individual s income is derived from any public assistance program. (Not all prohibited bases apply to all programs.) Persons with disabilities who require alternative means for communication of program information (Braille, large print, audiotape, etc.) should contact USDA s TARGET Center at (202) 720-2600 (voice and TDD). To file a complaint of discrimination, write to USDA, Director, Office of Civil Rights, 1400 Independence Avenue, S.W., Washington, D.C. 20250-9410, or call (800) 795-3272 (voice) or (202) 720-6382 (TDD). USDA is an equal opportunity provider and employer.

Contents Introduction 1 New Laboratory Evaluation 3 Real-Time Instruments 4 Goals and Methods 8 Results and Discussion 9 Conclusions 14 ii

Introduction The Missoula Technology and Development Center Air Quality Program has been evaluating real-time smoke particulate monitors since 1999. Air quality specialists and fire managers use these commercial monitors to estimate smoke particulate concentrations from prescribed and wildland fires. The center s evaluations include determining the accuracy of the real-time monitors by comparing their results to a gravimetric standard and evaluating the instrument s overall usability and functionality. Usually, evaluations are conducted when new, promising commercial monitors are introduced or when Forest Service personnel are considering using monitors that have not been evaluated. Since the last evaluation in 2003, several developments in smoke monitoring have prompted a new evaluation. Met One Instruments Inc. has a new external ac electrical pump for their E- BAM particulate monitor. This ac pump can be used as a replacement for the short-lived internal dc electrical pump. Additionally, new firmware updates have been developed for the E-BAM to alleviate some technical issues. Also, some Forest Service personnel were using a monitor manufactured by TSI Inc. called the DUSTTRAK. The DUSTTRAK had not been evaluated by MTDC. MTDC completed four evaluations of commercial real-time smoke particulate monitors from 1999 to 2003 (figure 1). Most of the evaluation tests were conducted in controlled laboratory Figure 1 Four reports (9925 2806 MTDC, 0025 2860 MTDC, 0125 2832 MTDC, and 0325 2834 MTDC) detail the evaluations of several real-time smoke particulate monitors in laboratory and field environments.

conditions at the Rocky Mountain Research Station s (RMRS) Fire Sciences Laboratory in Missoula, MT. Other evaluations were conducted by monitoring prescribed and wildfire smoke. The previous reports include: Laboratory Evaluation of Two Optical Instruments for Real-Time Particulate Monitoring of Smoke (http:// www.fs.fed.us/t-d/pubs/htmlpubs/ htm99252806/) Evaluation of Optical Instruments for Real-Time Continuous Monitoring of Smoke Particulates (http://www.fs.fed.us/t-d/pubs/htmlpubs/ htm00252860/) Real-Time Smoke Particulate Sampling Fire Storm 2000 (http:// www.fs.fed.us/t-d/pubs/htmlpubs/ htm01252832/) Laboratory Evaluation of Real- Time Smoke Particulate Monitors (http://www.fs.fed.us/t-d/pubs/htmlpubs/ htm03252834/) This Web site (http://www.fs.fed.us/ t-d/) requires a username (t-d) and password (t-d).

New Laboratory Evaluation This evaluation included two Met One Instruments Inc. E-BAM monitors (one with a dc pump, and the other with an ac pump), two Met One Instruments Inc. E-Samplers, three TSI Inc. DUSTTRAK monitors, four Thermo Fisher Scientific Inc. DataRAM 4 monitors, and three Thermo Fisher Scientific Inc. DataRAM 2000 monitors. The U.S. Environmental Protection Agency's Federal Reference Method (FRM) gravimetric sampler, a BGI Inc. PQ 200, was used as the standard. The FRM gravimetric filters were weighed by the RMRS Fire Sciences Laboratory in Missoula, MT. Previous laboratory evaluations were conducted at the RMRS labora- tory. Because the laboratory was not available for this evaluation, it was conducted in a large enclosed room at MTDC. The room is an industrial paint booth 14 feet wide, 9 feet tall, and 30 feet long. The ends of the booth were blocked off using a large plastic tarp and cardboard to keep the smoke in the room. Monitors were set up side by side inside the chamber (figure 2). Smoke was made by burning small amounts of pine needles on a piece of thick aluminum foil on the floor of the paint booth. A fan mixed the smoke in the room. The accuracy of the real-time monitors was determined by comparing the real-time results to the results from an approved FRM gravimetric sampler. The FRM sampler uses a pump to draw air across a filter that captures the smoke particulate. The filter is weighed before and after the tests to determine the total amount (mass) of particulate that was captured. The sampler calculates the volume of air that is drawn across the filter during the test. The average smoke concentration, represented in micrograms per cubic meter, can be calculated by dividing the mass by the total volume. The average realtime particulate concentration is calculated for the time the FRM sampler was operating. Figure 2 The real-time smoke particulate monitors and the Federal Reference Method sampler shown inside the smoke chamber during testing.

Real-Time Instruments E-BAM The E-BAM (the E stands for environmentally protected) monitor (figure 3) uses a three-part process to estimate the mass concentration of particulate. First, beta particles from a naturally occurring radioactive isotope are emitted through a clean filter tape and counted. Second, sampled air is passed through the exposed filter tape and particles are deposited. Finally, beta particles are passed through the tape again and recounted. The second count will be lower than the first, because beta particles will have been absorbed by the deposited particulate. The instrument Figure 3 The E-BAM ambient particulate monitor. This instrument collects particulate on a filter tape and uses beta ray attenuation to estimate particulate concentrations. uses an empirically derived algorithm to estimate the total mass of the deposited particulate. The estimated mass is divided by the volume of sampled air to calculate the average concentration in mass per unit volume. The exposed filter tape will advance automatically and begin a new collection after a userspecified time period or when the tape becomes clogged. The E-BAM is a portable instrument that can operate using line (110 volts ac), battery, or solar power in the harsh environments typically found near a wildland fire without the need for an additional enclosure. It can be configured with EPA-approved PM 10 and PM 2.5 particle size separators (cutoff inlets) to remove particulate larger than 10 micrometers (PM 10 ) or 2.5 micrometers (PM 2.5 ) mean diameter, respectively. It uses pressure and ambient temperature sensors to recalculate flow at standard atmospheric conditions. The E-BAM can be mounted on a tripod and configured to accept a variety of meteorological sensors for monitoring ambient windspeed and direction, relative humidity, and temperature. A six-channel data logger stores all the information. The range of measured concentration is 0 to 100 milligrams per cubic meter. The data logger can be programmed to report as often as every minute. Data can be downloaded at any time using an RS 232 serial port. A laptop computer or a modem can be configured to transmit the data using telephone lines, cell phone connections, or satellite telemetry. The E-BAM is 16 inches high by 13½ inches wide by 8 inches deep. It weighs about 28 pounds without the PM 10 or PM 2.5 cutoff inlet. The cutoff inlets combined weigh an additional 7½ pounds and add 30 inches to the total instrument height. The power supply for 110-volt ac line power weighs 14 pounds. The ac pump (figure 4) is 10 by 16 by 18 inches and weighs 30 pounds. For more information, visit the Met One Instruments Inc. Web site at http://www. metone.com or call 541 471 7111. Figure 4 The E-BAM monitor can be used with an external ac electrical pump rather than the internal dc pump, which is more likely to fail after extended use. E-Sampler The E-Sampler (figure 5) is a forward-scattering nephelometer that uses light-scattering principles to estimate mass concentrations of airborne particulate. An internal visible laser diode is directed through the sample of air. Particulate in the sample scatters a portion of the light. The scattered light is collected at a forward angle and focused on a

Figure 5 The E-Sampler uses light scattering to estimate particulate concentrations. photodiode that converts the light to an electrical signal. The resulting signal strength is proportional to the amount of scattered light. An empirically derived algorithm converts the electrical signal to an estimated mass concentration. The E-Sampler is a small, lightweight instrument housed in an environmentally protected enclosure. It can estimate particulate concentrations as high as 100 milligrams per cubic meter. It can be configured with a cutoff inlet to measure total suspended particulate (TSP), PM 10, PM 2.5, or PM 1. The E- Sampler can be mounted on a tripod and operated using line, solar, or battery power. A six-channel data logger accommodates other meteorological sensors that measure windspeed and direction, relative humidity, and ambient temperature. The sampler also has an inlet heater that can be programmed to operate when the inlet airflow s relative humidity reaches a particular level. Humidity can cause the instrument to overestimate particulate concentrations. The heater removes moisture from the air before it is sampled. However, the heater also may remove certain hygroscopic aerosols or convert them to a gas, reducing the estimated particulate concentration. The E-Sampler is considered a dual technology instrument. Besides the realtime optical scattering measurement, the E-Sampler also incorporates a gravimetric filter system using the same 47-millimeter filter as the FRM sampler. While the gravimetric device does not provide results in real time, it can be used to develop correction algorithms to improve the accuracy of the real-time estimations. The E-Sampler is compact just 12 inches high by 10 inches wide by 5 inches deep. It weighs 13 pounds with the PM 2.5 cutoff inlet, heater, and power supply. For more information, visit the Met One Instruments Inc. Web site at http:// www.metone.com or call 541 471 7111. DUSTTRAK The DUSTTRAK (figure 6) is a portable, battery-operated laser photometer that gives a real-time digital readout with a built-in data logger. The DUSTTRAK provides a real-time measurement based on 90- degree light scattering. A pump draws Figure 6 The DUSTTRAK is a lightweight portable aerosol monitor. the sample air through an optics chamber where it is measured. The DUSTTRAK uses a system to provide a sheath of clean air that isolates the aerosol in the chamber, keeping the optics clean for improved reliability and reduced maintenance. The DUSTTRAK has a measurement range of 0.001 to 100 milligrams per cubic meter with a particle size range of 0.1 to 10 micrometers. Flow rate can be adjusted from 1.4 to 2.4 liters per minute. The monitor can be configured to measure TSP, PM 10, PM 2.5, or PM 1. The internal data logger can store 31,000 data points (about 21 days of logging at one data point per minute). The DUSTTRAK can run on ac power or batteries. Four C-size alkaline batteries allow it to operate for 16 hours. The DUSTTRAK is 8.7 inches high by 5.9 inches wide by 3.4 inches deep. It weighs 3.3 pounds with batteries. An environmental enclosure is available to protect the unit when it is used outdoors. For more information on the DUSTTRAK, visit the TSI Inc. Web site at http://www.tsi.com or call 800 874 2811.

DataRAM 2000 The DataRAM 2000 (figure 7) is a compact, selfcontained monitor that estimates mass concentration internally from the measured scattering of light. The monitor can measure particulate concentrations from 0.1 to 400,000 micrograms per cubic meter. The monitor continuously displays the current and time-weighted average mass concentration while logging up to 10,000 data points. Data can be downloaded from the monitor through an RS 232 serial data port. The DataRAM can be configured with either a PM 2.5 or PM 10 impactor head to prevent particles larger than 2.5 or 10 micrometers, respectively, from entering the optical chamber. For custom calibrations, or to analyze chemical composition, particulates can be collected on a 37-millimeter filter in the instrument s base. An inline heater also may be installed for monitoring in humid conditions (the manufacturer suggests using the heater when the relative humidity is higher than 70 percent). The monitor s tubular heater is designed to heat the sampled air stream to evaporate liquid water from airborne particles or to eliminate fog droplets. The Data- Figure 7 The DataRAM 2000 is a single-wavelength, light-scattering monitor. Several of these monitors are being used by Forest Service air quality specialists, although they are no longer commercially available. RAM is powered by an internal rechargeable battery or by an external dc or ac power source. The DataRAM has a built-in, internal calibration device. Thermo Fisher Scientific Inc. no longer manufactures the DataRAM 2000. DataRAM 4 The DataRAM 4 (figure 8) is an upgraded version of the DataRAM 2000. The DataRAM 4 uses light-scattering principles to estimate mass concentrations. The DataRAM 4 uses two light sources with different wavelengths (600 and 880 nanometers). In theory, this dual wavelength allows the DataRAM 4 to better estimate mass concentrations using an algorithm that incorporates scattering from each light source. The algorithm uses the mean particle diameter size computed from the light scattering to correct the mass concentration. Some of the new features of the DataRAM 4 include internal temperature and relative humidity sensors, the ability to autocorrect concentrations when measuring air that has high relative humidity, and the ability to measure mean particle diameter size. The monitor can measure particulate concentrations up to 400 milligrams per cubic meter and has a built-in data logger that can store as many as 50,000 data points. Like the E-Sampler, the DataRAM 4 has a built-in gravimetric filter system for custom calibrations, although its 37-millimeter membrane filter is smaller than the 47-millimeter filter used in the FRM sampler.

The DataRAM 4 monitor is 5.3 inches high by 7.3 inches wide by 13.6 inches deep and weighs 11.7 pounds. For information on the DataRAM 4, visit the Thermo Fisher Scientific Inc. Web site at http://www.thermo.com or call 1 800 241 6898. Figure 8 The DataRAM 4 is a dual-wavelength, light-scattering monitor that estimates particulate concentrations.

Goals and Methods The main goal of this evaluation was to determine the accuracy of the real-time instruments when estimating smoke particulate concentrations. This was accomplished by comparing the real-time results to results from an FRM gravimetric sampler. A correction algorithm can then be established for each type of instrument that allows for better estimation of smoke particulate concentrations in the field. We were also interested in any difference in the results from the E-BAM with the ac pump compared to the dc pump. We tested at least two instruments of each model so we could begin to understand the consistency in performance between identical instruments. All the instruments except the DUSTTRAK were configured with their respective PM 2.5 particle size cutoff device installed. We used the PM 10 cutoff inlet with the DUSTTRAK because Forest Service personnel who use the instrument for smoke monitoring requested that we test the monitor in this configuration. We conducted one test with a DUSTTRAK configured with the PM 2.5 cutoff inlet to determine its effectiveness. We calibrated all the instruments before the tests. Instrument sample airflows were checked. All the real-time instruments were set for an average time of 1 minute and were run continuously throughout the tests. Before testing each day, a self test was conducted on the E-BAM monitors to ensure that they were functioning properly. The E-Sampler, DUSTTRAK, and DataRAM monitors were calibrated and zeroed using procedures recommended by the manufacturer. The DataRAM 4s were programmed with the particle size correction feature set to enable, which allows the monitor to correct the estimated mass concentration based on the mean particle diameter. The FRM sampler was programmed to run for different lengths of time, depending on the smoke particulate concentrations. For these tests, the FRM was programmed to run between 20 and 120 minutes. After the filters were weighed, the mass of the collected particulate was divided by the total volume of air drawn through the filter for the particular time period to obtain the average mass concentration. That value was compared to the average mass concentration estimated by the real-time instruments during the same period.

Results and Discussion A total of 38 tests were conducted. The average particulate concentrations ranged from 21 to 1,706 micrograms per cubic meter as measured by the FRM sampler. Statistical results were based on least squares-linear regression equations and correlation coefficients (R²). Table 1 shows the results for all the instruments. Figures 9, 11, 12, 14, and 15 show the results for each instrument when compared to the FRM sampler. Each figure shows the appropriate data points, the slope equation, and the correlation coefficient for the best-fit line. A line representing a 1:1 relationship also is shown. Figure 10 shows the comparison between the ac pump and dc pump versions of the E-BAM. Figure 13 shows the comparison between a DUSTTRAK with the PM 2.5 cutoff inlet installed and two DUSTTRAKs with PM 10 cutoff inlets installed. Table 1 Evaluation results for each instrument evaluated. The abbreviation for serial number is S/N. Real-time Regression Correlation Particulate Monitor Slope Coefficient E-BAM (S/N C5193 ac pump) 1.01 0.96 E-BAM (S/N E4052 dc pump) 1.01 0.97 E-Sampler (S/N 4886) 1.08 0.96 E-Sampler (S/N 5126) 1.18 0.94 DUSTTRAK (S/N 22347) 3.25 0.96 DUSTTRAK (S/N 22349) 3.18 0.96 DUSTTRAK (S/N 85200005) 3.09 0.96 DataRAM 2000 (S/N 2574) 1.25 0.97 DataRAM 2000 (S/N 2575) 1.30 0.97 DataRAM 2000 (S/N 2576) 0.89 0.97 DataRAM 4 (S/N D041) 2.61 0.95 DataRAM 4 (S/N D042) 2.19 0.96 DataRAM 4 (S/N D043) 2.58 0.95 DataRAM 4 (S/N D044) 2.37 0.96

E-BAM Comparisons between the E-BAM monitors and the FRM sampler are shown in figure 9. Results were essentially identical. Both of the E-BAMs overestimated the smoke particulate concentrations by 1 percent. Correlation coefficients were 0.96 and 0.97. Figure 10 shows the comparison between the ac pump version of the E- BAM and the dc pump version. Results of the two E-BAMs were almost identical. 2000 y = 1.01x R 2 = 0.96 1800 y = 1.01x R 2 = 0.97 1600 1400 1200 1000 800 EBAM S/N C5193 EBAM S/N E4052 600 Linear (EBAM S/N C5193) Linear (EBAM S/N E4052) 400 200 0 0 200 400 600 800 1000 1200 1400 1600 1800 2000 FRM Concentration ( u g/m 3 ) Figure 9 Results for the two E-BAM monitors compared to the BGI PQ-200 Federal Reference Method sampler. 2000 1800 y = 1.00x R 2 = 0.99 1600 1400 1200 1000 dc 800 600 400 200 0 0 200 400 600 800 1000 1200 1400 1600 1800 2000 EBAM S/N C5193 (AC Pump) concentration ( u g/m3) Figure 10 Results comparing an external ac electrical pump to an internal dc pump on an E-BAM monitor. 10

E-Sampler Figure 11 shows the comparison of the E-Samplers and the FRM sampler. On average, the E- Samplers overestimated the concentration by 13 percent. E-Sampler serial number 4886 overestimated the concentration by 8 percent while E- Sampler serial number 5126 overestimated the concentration by 18 percent. Correlation coefficients for both results were higher than 0.94. 2500 2000 1500 1000 500 y = 1.18x R 2 = 0.94 y = 1.08x R 2 = 0.96 E-Sampler S/N 4886 E-Sampler SN 5126 Linear (E-Sampler SN 5126) Linear (E-Sampler S/N 4886) 0 0 500 1000 1500 2000 2500 FRM Concentration ( u g/m3) Figure 11 Results for the two E-Samplers compared to the BGI PQ-200 Federal Reference Method sampler. DUSTTRAK Results for the three DUSTTRAK monitors evaluated and the FRM sampler are shown in 6000 5000 y = 3.25x R 2 = 0.96 y = 3.18x R 2 = 0.96 y = 3.09x R 2 = 0.96 figure 12. On average, the DUSTTRAK overestimated the smoke particulate 4000 concentration by 217 percent (3.17 times). The difference of the results 3000 among the instruments was +/-2.5 percent. 2000 Dusttrak S/N 22347 Dusttrak S/N 22349 Dusttrak S/N 85200005 1000 Linear (Dusttrak S/N 22347) Linear (Dusttrak S/N 22349) Linear (Dusttrak S/N 85200005) 0 0 1000 2000 3000 4000 5000 6000 FRM Concentration ( u g/m3) Figure 12 Results for the three DUSTTRAK monitors compared to the BGI PQ-200 Federal Reference Method sampler. 11

Figure 13 shows the comparison between the DUSTTRAK equipped with the PM 2.5 cutoff inlet installed and the DUSTTRAKs with the PM 10 cutoff inlet installed. Results (5-minute averages) were essentially identical. Most smoke particles are smaller than 2.5 microns, so neither cutoff inlet would be expected to have much effect. 7000 6000 5000 4000 3000 2000 y = 1.00x R 2 = 1.00 1000 0 0 1000 2000 3000 4000 5000 6000 7000 DustTrak with PM 2.5 Inlet Concentration ( u g/m3) Figure 13 Results showing DUSTTRAK monitors configured with a PM 2.5 or a PM 10 cutoff inlet. DataRAM 2000 Three Data- RAM 2000s were operated during the tests. Figure 14 shows the results from the DataRAM 2000 monitor compared to the FRM sampler. On average, the DataRAM 2000s overestimated particulate concentrations by 15 percent. DataRAM serial number 2575 overestimated the particulate concentrations by 30 percent while DataRAM serial number 2576 underestimated the concentrations by 11 percent. Correlation coefficients were 0.97. 2500 y = 1.30x R 2 = 0.97 y = 1.25x R 2 = 0.97 2000 y = 0.89x R 2 = 0.97 1500 1000 Dataram 2000 S/N 2574 Dataram 2000 S/N 2576 Dataram 2000 S/N 2575 Linear (Dataram 2000 S/N 2575) 500 Linear (Dataram 2000 S/N 2574) Linear (Dataram 2000 S/N 2576) 0 0 500 1000 1500 2000 2500 FRM Concentration ( u g/m3) Figure 14 Results for the three DataRAM 2000 monitors compared to the BGI PQ-200 Federal Reference Method sampler. 12

DataRAM 4 Four DataRAM 4s were evaluated. Figure 15 shows the results of DataRAM 4 monitor compared to the results from the FRM sampler. On average, the DataRAM 4s overestimated the smoke particulate concentration by 144 percent. Data- RAM 4 serial number D041 overestimated concentrations the most (161 percent), while DataRAM 4 serial number D042 overestimated concentrations the least (119 percent). Correlation coefficients were all high, at least 0.95. 5000 4500 4000 3500 3000 2500 2000 1500 1000 500 y = 2.61x R 2 = 0.95 y = 2.58x R 2 = 0.95 y = 2.37x R 2 = 0.96 y = 2.19x R 2 = 0.96 0 0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 FRM Concentration ( u g/m3) Dataram 4 S/N D041 Dataram 4 S/N D042 Dataram 4 S/N D043 Dataram 4 S/N D044 Linear (Dataram 4 S/N D041) Linear (Dataram 4 S/N D043) Linear (Dataram 4 S/N D044) Linear (Dataram 4 S/N D042) Figure 15 Results for the four DataRAM 4 monitors compared to the BGI PQ-200 Federal Reference Method sampler. 13

Conclusions Of the five instruments we evaluated, the Met One Instrument E-BAM was the monitor that most closely estimated smoke particulate concentrations when compared to the FRM sampler. It overestimated the smoke particulate concentration by 1 percent. The tests also showed that the ac and dc pump versions of the E-BAM produced similar results. The E-BAM loses some of its portability when the bulky ac pump is used. For some users this loss in portability may be a small inconvenience compared to having to replace the dc pump periodically. The E-Sampler was also fairly accurate when estimating smoke particulate concentrations. The two E- Samplers overestimated concentrations on average by 13 percent with a difference of just +/- 4 percent between results of the two instruments and the FRM average. Users should consider correcting the estimates from the E- Sampler by 13 percent (multiply the values by 0.89) when measuring smoke particulate. The DUSTTRAK monitors overestimated the concentration by 217 percent. The difference between results of the three instruments was just +/-2.5 percent. This consistency is the best we have seen in all of our evaluations of these types of instruments. A correction algorithm of 0.32 (multiply the estimated concentration by 0.32) is recommended for those using the DUST- TRAK. The use of the PM 2.5 or PM 10 cutoff inlet does not make much difference when estimating smoke particulate concentrations with the DUSTTRAK. The DataRAM 4 monitors overestimated the smoke particulate concentrations by a large amount, although this was not unexpected based on previous evaluations. On average, the four instruments overestimated concentrations by 144 percent. Previous tests have showed that the DataRAM 4 with the size-correction feature enabled overestimated concentrations even more. Previous recommendations suggested a correction factor of 0.37. Based on these new results, we recommend a slightly higher correction factor of 0.39. On average, the DataRAM 2000 monitors overestimated the mass concentration by 15 percent. This is somewhat of a surprise because previous tests indicated that the DataRAM 2000 normally overestimates the concentration by more than 100 percent. The monitors used in these tests had not been evaluated previously. Based on these results, it may be prudent to develop individual correction algorithms for the DataRAM 2000s. These tests were conducted using smoke generated by burning dry pine needles. Results may differ if the vegetation type or moisture content is significantly different. Smoke from burning buildings or other sources with different fuels may yield dramatically different comparisons between smoke particulate monitors and the FRM sampler. About the Author Andy Trent is a project engineer at the Missoula Technology and Development Center. He received a bachelor s degree in mechanical engineering from Montana State University in 1989. Before coming to MTDC in 1996, Andy worked as a civilian engineer for the U.S. Navy. He works on projects in the nurseries and reforestation, forest health protection, and watershed, soil, and air programs. 14

Library Card Trent, Andy. 2006. Smoke particulate monitors: 2006 update. Tech. Rep. 0625 2842 MTDC. Missoula, MT: U.S. Department of Agriculture Forest Service, Missoula Technology and Development Center. 14 p. This evaluation updates earlier evaluations of real-time monitors. These monitors can be used to evaluate air quality in areas affected by smoke from wildland fires. The evaluation was conducted in a laboratory setting and compared each monitor to at least one more identical monitor, to the other monitors evaluated, and to a Federal Reference Method sampler that was used as the standard. The DUSTTRAK monitor manufactured by TSI Inc. had not been evaluated before, nor had the external ac pump available for the E-BAM monitor manufactured by Met One Instruments Inc. Keywords: air quality, DataRAM, DUSTTRAK, E-BAM, E-Sampler, Federal Reference Method, nephelometers, prescribed fires, real time, wildland fires Single copies of this document may be ordered from: USDA Forest Service, MTDC 5785 Hwy. 10 West Missoula, MT 59808 9361 Phone: 406 329 3978 Fax: 406 329 3719 E-mail: wo_mtdc_pubs@fs.fed.us Electronic copies of MTDC s documents are available on the Internet at: http://www.fs.fed.us/eng/t-d.php For additional technical information, contact Andy Trent at MTDC. Phone: 406 329 3912 E-mail: atrent@fs.fed.us Forest Service and Bureau of Land Management employees can search a more complete collection of MTDC s documents, videos, and CDs on their internal computer networks at: http://fsweb.mtdc.wo.fs.fed.us/search/