R I T Title: Manual Processing Bench #2 Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 08/05/2008

Similar documents
R I T Title: RCA Bench #2

Title: Nitride and Silicon Wet Etch Semiconductor & Microsystems Fabrication Laboratory Revision: I Rev Date: 01/20/2010

Standard Operating Manual. Ultra Fab Wet Station F Non-standard Processing Station

General Acids/Bases Bench Right and Left SOP

Wet Bench Standard Operating Procedures

Amerimade RCA Wet Bench SOP Page 1 of 13 Rev /22/2014

Standard Operating Manual. PEL Wet Station X Clean Organic Solvent Station

Polypro Acid Wet Bench UCLA Nanolab Yellow Room. Primary contacts: Max Ho, Wilson Lin. PREPARED BY: Steve Franz DATE Rev 2/01/12

GENERAL WET BENCH SAFETY AND USES SOP. October 2013 GENERAL SAFETY RULES

Standard Operating Manual. Ultra Fab Wet Station C Oxide/ Nitride Etch Station

The Cameron Clean Room

CNS Cleanroom Safety Orientation Form

The University of Washington Nanofabrication Facility (WNF) at Fluke Hall 125 Fluke Hall Seattle, WA 98195

Acid / Base Bench Procedures and Operation Manual

Haz Waste Management

HAVE A SAMPLING AND ANALYSIS PLAN (SAP)

Operation Procedure for SMiF Wet Hoods (Solvent, Spin Coat, Acid, and Develop Hoods)

Title: CHA E-Beam Evaporator Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 01/13/2011

Science Department Safety Training Notes

Revised: January 8, Goal: Deposit, expose, and develop standard positive photoresist films as an etch mask, liftoff mask, or sacrificial layer.

PDS 2010 Parylene Coater SOP

The safety refresher are mandatory for ALL users Cleanroom access and instrument usage will be denied for users who:

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Columbia Nano Initiative (CNI) Cleanroom Operations Procedures Rev

Emergency Response Cart Guide

Safety In Chemistry. Identified Risks and Hazards

The Do's and Don ts in the Instrument Sterilisation Process

OPERATION MANUAL Model 800 Nanoimprint Controller

Laboratory Glassware Cleaning and Storage

SAFE WORK PROCEDURE. List the Hazards and risk controls as per risk assessment. assessment reference:

EE-527: MicroFabrication. Wafer Handling and Cleaning

Laser Class 4 Laser Type Nd-YAG Laser Manufacturer New Wave Research Laser Model SoloII-30 PIV Laser Serial / Id # Serial number 16582

NMOS Fabrication Process Description

PERFORM Operating Document

LABORATORY INSPECTION CHECKLIST

LABORATORY CHEMICAL HYGIENE AND SAFETY PLAN

Safety Lab Goggles are required! Apron is optional!

Safety instructions and routines on lab Faculty of Odontology Malmö University

Teaching Lab Headway Spinner SOP

University of California, Davis Laboratory Self-Inspection Checklist

ROTOR COATING WET BENCH OPERATING PROCEDURE

Application: Response procedures for major and minor spills, including: organic solvents acids bases dry chemicals mercury hydrofluoric acid

Manual Parts Washer. Operating Instructions & Warranty Information

DEPARTMENTAL SAFETY ISSUES - SAFETY TECHNOLOGIST / INFECTION CONTROL PRACTITIONER

Chapter 3: Emergency Planning for Laboratories

PDMS Casting Station Supplies

TBJ Model SPCS B Scope Pre-Cleaning Sink. Product Specification. Product Description

Cruise Line Cleaning and Disinfection Protocol for Environmental Services / Housekeeping

SUBFAB DOC PDOC 0152 ROGER SHILE 7/11/97. SCIENCE MISSION SQUID CARRIER SUBSTRATE (Part# ) FABRICATION 1. Document Revision Record

Operation of SemiTool Spin Rinser Dryers

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Chemistry 11 Laboratory Safety Guidelines

Rules for Safe Operation

National Chiao Tung University

1. Researchers are familiar with the location and contents of: The lab s Chemical Hygiene Plan YES NO

Laboratory Safety Audit

Safety Awareness For Researchers in the Giga-to-Nanoelectronics Laboratory. Richard Barber

until filters are dry. Once dry, place the filters shiny side up in petrislides. The petrislide do NOT need to be acid washed.

SCIF SOP for BSL 2 samples on Aria3

PVA Clean Sponge. for Clean Room, Pharmaceutical. and Laboratory. Applications SUPER PVA SPONGE

Standard Operating Manual

General Procedures: Materials Synthesis Lab 1358 CNAM/Toll

Green Chemistry Safety Training PSE/CH 335. Departments of Forest Biomaterials and Chemistry North Carolina State University

Massachusetts Institute of Technology Kavli Institute for Astrophysics and Space Research (MKI)

1.0 Denton Thermal Evaporator

STUDENT SAFETY RULES

Task #1. Quick Overview: Wipe white residue build- up off one of the show tanks (class tank) Procedure: 10. Wash hands.

Standard Operating Manual. Semitool Spin Rinser Dryer SRD

1. Conduct yourself in a responsible manner at all times in the laboratory.

Application Data Sheet

A. General Safety LAB/ROOM: INSPECTED BY: DATE:

Lab Safety presentation and contract first day of lab know the locations of everything in your lab no student can participate in class without signed

INLINE DI HEATER. Operation / Maintenance Manual

Experiment #6 Photolithography: Microprocessing Technology Fabrication of the microstructures with SPR photoresist

Laboratory Inspection Checklist

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Student Lab Procedures. Lab Safety Orientation

Sethco MODEL P90 SEALLESS DRUM PUMPS. In Use The World Over. Sethco Division DRUM PUMP

panasonic.net/id/pidsx/global Lights up when conditions are normal (and at incipient liquid leak detection) Use as a warning Sensor

Introduction. Sharps. Working Safely with Sharps at UAB (OHS_HS255) Objectives. Defined

Standard Operating Procedure for Oil & Grease Method EPA1664A

Go to the Lab! Lab Safety Begins Before You

BOUMATIC ROBOT AND THE LELY SAMPLE

UVM Cosmogenic Laboratory - Quartz Purity Testing

Cascade Comfort Bathing Systems with Aqua-Aire Safe Operation & Daily Maintenance Instructions

Dianne Harrison, Supervisor QA Lab

ENVIRONMENTAL STANDARD OPERATING PROCEDURE 7 DEGREASING SOLVENT

Chemical Thermo-con. How to Order. Communication A RS-485 B RS-232C. Power supply 2 Single-phase: 180 to 242 VAC 50/60 Hz.

Useful Equations and Conversions

INSTRUCTION MANUAL. Laboratory Equipment Pty Ltd Ph: Fax:

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Workbook Clean floors

Safety in the Malaspina Chemistry Laboratories 1. Emergency shower 2. Eyewash fountain 3. Fire extinguisher 4. Sand bucket 5.

DRAFT. Module 4. Decontamination & Environmental Cleaning

Important Safeguards. Topic 2 SAVE THESE INSTRUCTIONS - FOR HOUSEHOLD USE ONLY SAVE THESE INSTRUCTIONS - FOR HOUSEHOLD USE ONLY 3

UV Lamp and Quartz Sleeve Replacement/Maintenance

NBSeries Installation Operation & Maintenance

Southampton Craft Show Food Vendor Application Form

Laboratory SAFETY RULES

Outlines. 1. Introduction of Safety at Laboratories. 2. Emergency Situations. 3. Basic Safety Rules for All the Labs (Engg.

Transcription:

R I T Title: Manual Processing Bench #2 Approved by: Process Engineer / / / / Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Manual Processing Bench #2 for wet cleaning of silicon wafers. All users are expected to have read and understood this document. It is not a substitute for in-person training on the system and is not sufficient to qualify a user on the system. Failure to follow guidelines in this document may result in loss of privileges. 2 REFERENCE DOCUMENTS o Material Safety Data Sheets for any chemicals that are used. o Safety Training. 3 DEFINITIONS DI H 2 O 2 H 2 SO 4 HCl NH 4 OH SC1 SC2 HF -De-Ionized Water -Hydrogen Peroxide -Sulfuric Acid -Hydrochloric Acid -Ammonium Hydroxide -First clean of the RCA process -Second clean of the RCA process -Hydrofluoric Acid 4 TOOLS AND MATERIALS 4.1 General Description 4.1.1 The Hotpot is a general use tank intended for cleaning wafers in Piranha, SC1 and SC2. Decontamination cleans may be done in this bench for wafers that have been through KOH etching or CMP. The bench is set up with one quartz temperature controlled tank, a dual cascade and a sink. The remaining area is for general use. 4.2 Wafer Boats 4.2.1 Only Teflon wafer boats and handles should be used for processing on this bench. RIT SMFL Page 1 of 6

5 SAFETY PRECAUTIONS 5.1 Personal Safety Hazards 5.1.1 The Manual Processing Bench uses several hazardous chemicals including HF, sulfuric acid, hydrogen peroxide, hydrochloric acid and ammonium hydroxide. Users should be aware of the unique hazards of the materials with which they are working. If a chemical is spilled, remove clothing and rinse affected area in safety shower for 15 minutes and inform SMFL staff member or lab instructor. 5.1.2 When working at the Manual Processing Bench, always use appropriate personal protective equipment (PPE) including apron, face shield and heavy rubber gloves. 5.1.3 The PPE should not be worn any place other than the immediate vicinity of the Manual Processing Bench. Do not walk away from bench while wearing PPE. If assistance is needed (i.e. to grab a Clean Wipe or timer, etc.), ask for assistance do not get it while wearing PPE. Do not work at computer or answer the phone while wearing PPE. When finished at the Manual Processing Bench, rinse the PPE, fully dry it and return to hook. 5.1.4 It is imperative that all spills be cleaned up immediately because of the number and variety of materials used at this bench. Please see a staff member for assistance in cleaning up spills. 5.1.5 Keep sash down when processing chemicals. 5.2 Hazards to the Tool 5.2.1 No metals or metal etchants may be processed in the heated tank or rinse tank. 5.2.2 No solvents may be processed in the heated tank. 5.2.3 No HF, buffered oxide etch or pad etch may be used in the heated tank. 5.2.4 Never operate the heated tank without the proper fluid level. 5.2.5 Operate the bench controls with clean gloves only. The chemicals used will cause damage. 5.2.6 The tank is breakable, carefully place wafers in it and never bang a cassette on the side. 5.2.7 Do not drain chemicals that are hotter than 30C. 5.2.8 Chemicals must be drained as soon as they are cool or the o-ring can fail and cause a leak. 5.2.9 Do not spray the bench top with the sprayer hose. RIT SMFL Page 2 of 6

6 INSTRUCTIONS 6.1 Initial State Check 6.1.1 Verify that there are sufficient amounts of the chemicals that you will need in the chemical storage cabinet. Sulfuric Acid is available from staff members. 6.1.2 Make sure that you have reserved the bench with the Tool Reservation System and that you have swiped in your card on the Card Swipe System. 6.2 Resetting the System 6.2.1 The temperature controller may be reset by cycling the power on and off. 6.3 Operating the Hotpot 6.3.1 On the lower right panel, turn on the Bench Power Master Switch. 6.3.2 Make sure that you are wearing safety glasses, a face shield, chemical gloves and a chemical apron. 6.3.3 To drain the old chemicals, wait until temperature is below 30C. Turn on the Hotpot Drain. Press and hold the Water Flow Master Control until you hear the tank start to drain. Rinse out the tanks with the sprayer and then drain the rinse water. Never drain hot chemicals. 6.3.4 Make sure the Hotpot Drain is Off. 6.3.5 Turn on the temperature controller on the lower panel and press Sil to silence the alarm. 6.3.6 Put up a sign describing the chemistry that you will be using. 6.3.7 To operate the cascade, turn the cascade On, press and hold the Water Flow Master Control until water flows. 6.3.8 Changing the temperature set point 6.3.8.1 Press Prog on the controller 6.3.8.2 You will be prompted for a Code. Use the arrow keys to go to 312. 6.3.8.3 Press Prog until PS (process set point) is displayed under the time field. 6.3.8.4 Use the arrow keys to change the temperature set point. 6.3.8.5 Press Save. 6.3.8.6 Press Reset. RIT SMFL Page 3 of 6

6.3.9 Piranha Clean 6.3.9.1 To do a Piranha Clean, fill up the tank with the correct amount of Sulfuric Acid. Next add the Hydrogen Peroxide and allow chemicals to sit 10 minutes before turning on the heater. Since the chemicals will self heat, this will prevent the temperature from overshooting. When the 10 minutes is up, stir well. Finally turn on the heater. 6.3.9.2 For 4 Inch wafers use: 4000mL Sulfuric Acid 1000mL Hydrogen Peroxide 6.3.9.3 For 6 Inch wafers use: 4800mL Sulfuric Acid 1200mL Hydrogen Peroxide 6.3.9.4 A piranha clean is typically done for 15 minutes at 126 C± 10 C and followed by a 5 minute rinse. The set point may be verified by pressing Set point on the temperature controller. 6.3.10 Decontamination Clean following KOH or CMP 6.3.10.1 To do a decontamination clean, fill up the tank with the correct amount of DI water. Next add the Hydrochloric Acid. Finally top off with the Hydrogen Peroxide. Add the chemicals slowly and carefully stir. Be sure to label the chemicals. 6.3.10.2 For 4 Inch wafers use: 3500mL DI water 700mL Hydrochloric Acid 700mL Hydrogen Peroxide 6.3.10.3 For 6 Inch wafers use: 4500mL DI water 900mL Hydrogen Peroxide 900mL Hydrochloric Acid 6.3.10.4 Allow the mixture to heat up to 70C. 6.3.10.5 Soak the wafers to be cleaned for 20 minutes in the solution using a Teflon cassette; do not use one of the dirty KOH cassettes. 6.3.10.6 Rinse in the cascade for 5 minutes and put wafers in the rinser/dryer. 6.3.11 SC1 Clean 6.3.11.1 Mix 4500mL DI water, 300mL Ammonium Hydroxide and 900mL Hydrogen Peroxide. Carefully stir. RIT SMFL Page 4 of 6

6.3.11.2 Allow the mixture to heat up to 75 C. 6.3.11.3 Soak the wafers for 10 minutes. 6.3.11.4 Rinse in the cascade for 5 minutes and put wafers in the rinser/dryer. 6.3.12 SC2 Clean 6.3.12.1 Mix 4500mL DI water, 300mL Hydrochloric Acid and 900mL Hydrogen Peroxide. Carefully stir. 6.3.12.2 Allow the mixture to heat up to 75 C. 6.3.12.3 Soak the wafers for 10 minutes. 6.3.12.4 Rinse in the cascade for 5 minutes and put wafers in the rinser/dryer. 6.4 Shutdown 6.4.1 Turn off the Bench Power Master Switch. The rinse tank is on a timer and will shut off automatically. 6.4.2 Swipe out on the card swipe system. 6.5 Errors during Run 6.5.1 If the controller does not turn on, make sure that the bench power is on and the tool is swiped in. 6.5.2 If the controller alarms or does not heat, an interlock may be tripped. Contact an SMFL staff member for assistance. 6.5.3 If a wafer falls out or breaks, do not attempt to retrieve it. Contact an SMFL staff member. 6.5.4 If the rinse tank will not operate, press the Rinse Tank Enable button again. 7 Appropriate Uses of the Tool 7.1 The heated tank is intended for Piranha cleaning, decontamination cleaning of silicon wafers, SC1 and SC2 cleaning. 7.2 Solvents are not allowed in the heated tank. 7.3 No metal etchants or wafers are allowed in the heated tank. 7.4 No HF or HF mixtures allowed in heated tank. 7.5 Someone must be in lab while the hotpot is in use. 7.6 Only Teflon wafer cassettes are allowed. REVISION RECORD Summary of Changes Originator Rev/Date Original Issue Sean O Brien A-05/19/2005 Modified decontamination clean chemical ratios Sean O'Brien B-02/22/2006 Modified 5.1.4, added 5.2.9, 6.3.8 Sean O'Brien C-08/05/2008 RIT SMFL Page 5 of 6

RIT SMFL Page 6 of 6