Sensistor XRS9012. Maintenance manual. Hy d r o g e n Leak Det e c t o r

Similar documents
99 Washington Street Melrose, MA Phone Toll Free Visit us at

Sensistor ISH2000 Hydrogen Leak Detector

LD910. User Manual. Combustible Gas Leak Detector. Detects all Hydrocarbon and other Combustible Gases. Made in U.S.A. MSHA Approved Intrinsic Safety

User Manual for ULTRASCAN High Pressure Leak and Corona Discharge Detector

YELLOW JACKET. H 2 Tracer Gas Leak Detector. Operations Manual. Detects 5% H 2 in 95% Nitrogen Tracer Gas. Model 69341

Gas leak detectors. For a quick overview. testo testo testo testo gas detector testo 316-Ex.

Gas leak detectors. For a quick overview. testo testo testo testo gas detector testo 316-Ex

LS1. Operator Manual. Kane International Ltd

testo 316-EX gas leak detector Instruction manual

CD100A COMBUSTIBLE GAS DETECTOR

We measure it. testo 316-Ex CH 4 testo 316-1

INSTRUCTION MANUAL ROTARY LASER LEVEL GPR-R40H / G40

IRwin. Methane Leak Detector INNOVATIVE METHANE DETECTOR SYSTEM QUALITY ASSURANCE IN THE AUTOMOTIVE INDUSTRY FOR EASY SURVEY OF GAS PIPELINES

LD920. User Manual. H2/N2 Tracer Gas Leak Detector. Detects 5% H2 in 95% Nitrogen Tracer Gas. Made in U.S.A. LeakDetectors-LD920-UserManual.

User manual and installation instruction Control for Fire Ventilation RV / RV 24-32

YELLOW JACKET. ACCUPROBE UV and ACCUPROBE II Refrigerant Leak Detector with Heated Sensor. Instruction and Operation Manual

REFRIGERANT LEAK DETECTOR OPERATIONS MANUAL LD-170

CO warning instruments and gas spillage detectors

FPD - Flame Photometric Detector

Model C-380. Instruction Manual. Refrigerant Leak Detector. reedinstruments

CONTROL DEVICE SLIDETRONIC HD

A3G Green Rotary Laser SPECIAL EDITION INSTRUCTION MANUAL

ATD-3698 Refrigerant Gas Leak Detector. Owner s Manual

Model GD Instruction Manual. Combustible Gas Leak Detector. reedinstruments. www. com

LD-5000 REFRIGERANT LEAK DETECTOR

GAS-Mate. Combustible Gas Leak Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector

INSTRUCTION MANUAL GAS LEAK DETECTOR CD200

Model R9100. REED Instruments Instruction Manual. Ultrasonic Leak Detectortor

actoolsupply.com User Guide Extech RD300 Air Conditioning AC Refrigerant Leak Detector Refrigerant Leakage Detector Model RD300 actoolsupply.

LD 5000 REFRIGERANT LEAK DETECTOR JB INDUSTRIES INC.

AFMS Audible Flow Monitoring System TECHNICAL SPECIFICATION

5, SE GÅNGHESTER,

LD200 / RLD400 LD200-G /RLD400-G

D-TEK Select. Refrigerant Leak Detector

Control box FSTronic DES-FS

Gas leak detectors. We measure it. For a quick overview. testo testo testo testo gas detector testo 316-Ex.

Victoreen Primalert 35 Area Radiation Monitor

Reflective Beam Smoke Detector

HAND HELD METAL DETECTOR

BX444-Mc Gas Detector

CR9976 ROBERTS. Auto Set Dual Alarm CD Clock Radio. Sound for Generations. Please read this manual before use

PVL1300 INSTRUCTIONS FOR USE

TEK-Mate. Refrigerant Leak Detector

Model ATD Refrigerant Gas Leak Detector. User Manual. Detects all CFC, HFC, HCFC, HFO Refrigerants including blends

IRwin. Methane Leak Detector INNOVATIVE METHANE DETECTOR SYSTEM QUALITY ASSURANCE IN THE AUTOMOTIVE INDUSTRY FOR EASY SURVEY OF GAS PIPELINES

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

LEAKSTOPPER MULTIZONE WATER LEAK DETECTION ALARM PANEL INSTALLATION AND OPERATION MANUAL

GasAlertMicroClip X3 Specifications

Operations Manual TS400. Test Station for G450/G460 Gas Detector

SENSIT HXG-2. Instruction Manual For use with combustible gases Gas Leak Detector

Restek s Electronic Leak Detector

H1R Red Rotary Laser INSTRUCTION MANUAL

Model 850 Investigator s Aid Accelerant and Gas Detector Operation and Maintenance

IMR 5000 IMR 400. CEMS Operating Manual IMR USA S/N: Sample IMR Sensors Inside IMR 400. Gas Dryer Inside. IMR Environmental Equipment Inc.

Control box FSTronic DES-FI

Intensity Level. Intensity Control. Signal Type Switch. 12V Power Port. Signal Ports

Analox 1000 Series. User Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK

GfG s EXOTECTOR has been tested for EX-protection by DMT (BVS) Dortmund

BS-316. Gas detection control panel up to 16 inputs. Installation operation manual

TEK-Mate. Refrigerant Leak Detector

LeakSafe. solutions. WaterSwitch2. Installation Instructions

RAGNAR STÅLSKOG AB INSTRUCTION MANUAL HVDC-H HVDC-L VOLTAGE DETECTOR HVDC-H, HVDC-L 15H11

FEC400 Series. Installation Manual

HX-80NRAM. High Mount Outdoor Detector

for Ultima-Tig and for Ultima-Tig-Cut

SprintIR -W CO2 Sensor

INSTRUCTION MANUAL MODEL GX-82A

NGD8800 USER S MANUAL

Data Sheet Fire Safety & Security Products. System Overview. Initiating Circuits SXL-EX FACP 7906

LeakSafe. solutions. WaterSwitch2. Installation Instructions

ANALOX 5001 Carbon Dioxide Monitor. User Manual ANALOX Analox 5001 Carbon Dioxide Monitor User Manual

INSTRUCTION MANUAL CD Fax: (503)

Model C Electronic Nose Accelerant and Gas Detector Operation and Maintenance

Fire Control Panel FS4000

English. Doro CareIP Mobile. User Guide

Danfoss gas detection units

LD900. User Manual. Refrigerant Gas Leak Detector. Detects all HFO, HFC, HCFC, and CFC Refrigerants including blends. Made in U.S.A.

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

USER MANUAL MINITRACER PORTABLE SNIFFER

SERVICE INSTRUCTIONS FOR. Fire Alarm System MINI-2000

KitchenAid Food Stream Solutions Classic and Integrated Series

UniGas Portable, Flue Gas Analysis Advanced Combustion Flue Gas Analyser

6-1 Installing Batteries 6-2 Automatic Circuit/Reset Feature 6-3 Feature Sensitivity Adjustment

Carbon Monoxide (CO) Gas Detector March 15 GCO 300

II 2 G. SENSIT HXG-2d. Ex ib IIB T3 IP20. Suitable for combustible gases such as Methane, Butane, Propane and Natural Gas MADE IN USA

USER MANUAL AND TECHNICAL MANUAL

Electronic Self-Leveling Horizontal Rotary Laser Level Model No Instruction Manual

INSTALLATION MANUAL SOUNDER MODELS RIALTO L LS X LSX TRIADE L LS QL QLS LIDO L X LS LSX

Operate the Mutli-Gas Monitor PID

SENSIT HXG-2d Combustible Gas Detector INSTRUCTION MANUAL Read and understand Instructions before use. PENDING Ex. II3G Ex ic llb T3 IP20

Lite. Installation Manual User Guide

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

H-10PM. Refrigerant Leak Detector

The Windcrest Remote Alarm for Passenger and Goods Lifts with Optional BS EN81-28 & BS EN81-70

Conventional Fire Detection

UltraRAE 3000 Basic Operation Pocket Reference

A measured step forward

Dräger X-am 5600 Personal Monitor Multi-Gas Detection Device

RTD TEMPERATURE SENSING SYSTEM

Transcription:

Sensistor XRS9012 Hy d r o g e n Leak Det e c t o r Publication: INFICON AB - nioa62e1-b (1111) - All information can be modified without prior notice Maintenance manual

2 MAINTANCE MANUAL HYDROG LEAK DETECTOR Sensistor XRS9012 Contents Page 1. Maintenance and repair 3 1.1 Function and controls 3 1.1.2 Battery charger 4 1.2.3 Probe attachment 4 1.2 Changing batteries 5 1.2.1 Removing flex cable 6-7 1.2.2 Removing front cover 8-9 1.2.3 Replacing the circuit board 10 1.3 Product information 11 1.4 Spare parts list 12 1.5 Function check 12 1.6 Sensitivity check 12

MAINTANCE MANUAL HYDROG LEAK DETECTOR Sensistor XRS9012 3 1 Maintenance and repair The Sensistor XRS9012 is maintenance free, except for battery replacement (normally every three years). This section describes the functions of the instrument and gives advice as how to function tests should be carried out. 1.1 Function and controls Following functions are controlled from the panel: - Sensitivity: can be varied in 10 steps, each step doubles the sensitivity. - On/Off: the detector has an automatic switch-off aftr 30 minutes of no activity. - Battery Check: At least 8 of the light emitting diods (LED) will illuminate at full charge. Full charge takes one hour. - Zero: the detector has an automatic zeroing function. The ZERO button allows the operator to force the signal to zero instantaneously. - Sound level: the volume can be changed in 64 steps. - Earphone output - MAX: this is a peak-hold function allowing the operator to compare the maximum gas signal in two different places. - SEARCH (default): this is the normal mode where the signal follows the gas concentration. The detector always starts in this

4 1.1.2 Battery charger The voltage range for charging is 9-15VDC. To operate on mains voltage a voltage converter is connected to the power input at the back of the unit. There is also a possibility to connect the unit to the car cigarettelighter, which enables the unit to be charged in the car. 1.1.3 Probe attachment All probes are attached to the Leak Detector Sensistor XRS9012 via cable C21 to the socket at the back of the unit.

5 1.2 Changing batteries Before reading this section, please see picture on page 11 to get an overview of the assembly. We recommend changing batteries every three years (batteries will normaly last for several more years but operation time is shortened). Follow the procedure described below. The batteries are placed inside the Leak Detector. Remove the four screws holding the back cover Remove the cover carefully Disconnect the flexible cable from the probe connector circuit board (see next page)

6 1.2.1 Removing flex cable Be very careful when removing the flexible cable in order not to demage the rivet, the cable or the circuit board. Read the whole section 1.2.1 before commencing work. - First remove the rivet by carefully pulling it upwards. - Thereafter pull the holder the same way. Rivet Holder Connector Flexible cable Circuit board Figure 1, Rivet and holder in place Figure 1 Figure 2, Rivet is removed holder still in place Figure 2 Figure 3, Rivet and holder removed Next step before pulling out the flexible cable is opening the circuit board connector (see next page). Figure 3

7 To open the connector, pull the locking mechanism away from the connector body Locking mechanism Pull the cable straight out of the connector Closed Open The interior of the Leak Detector is now accessible. It is now possible to loosen the front cover.

8 1.2.2 Removing front cover Remove the four front panel screws from inside and the front cover carefully. NB: the power supply cable from the back cover is still attached to the circuit board on the front cover Diconnect the power supply cable from its connector on the circuit board Front and back covers removed

9 Disconnect the cables attached to the batteries, one battery at the time. Remove the screws holding the battery (see figure). NB: there are two screws holding each battery. Screw holding battery Cables Replace the battery, connect the cables and tighten the screw. Repeat this for all three batteries Battery Re-assemble the Leak Detector The new batteries should be charged for >12 hours to be fully charged. The Sensistor XRS9012 Leak Detector is ready for use.

10 1.2.3 Replacing the circuit board Open the detector following the instruction in section 1.2 Loosen and remove the screws holding the circuit board to the front cover. Disconnect the speaker and earphone connector. Lift the circuit board carefully so that there is access to the connector from the circuit board to the touch pad panel. Loosen the connector carefully. Replace the circuit board. Circuit board front side Front cover without circuit board Circuit board backside

11 1.3 Product information Touch pad keyboard Front cover Rubber seal Loudspeaker Circuit board Battery Battery holder Front panel screw Rubber seal. Back cover Product facts: Serial number label Sensitivity: 0,7ppm H2 in air Response time: < 1sec Start-up time: 6 sec Outputs: - Display 10 level LED bar graph - Speaker 5-1600 Hz - Earphones std earphones, 3,5mm jack, > 8ohm Power Supply: Rechargeble lead batteries Battery capacity: 13h @ 20o C Protection: IP55 Size & weight: 250x120x85 mm, 1,9kg 260x220x95 mm, 2,5 kg (incl. carrying case)

12 1.4 Spare part list Part no. Battery 591-294 Earphones 591-443 Carrying case 591-304 Charger AC (591-300) 591-301 SV, 591-302 USA, 591-303 UK Charger car 591-361 Hand Probe H21 590-200 Cable C21 590-150 Circuit board 591-293 1.5 Function check Action Result Connect charger The instrument starts Disconnect The instruments turns off Change sensitivety The LED s indicates corresponding sensitivity Adjust the volume The volume should change Press Battery Check button The LED should indicate Battery status Disconnect Hand Probe The ON LED should flash red and an audio signal appear 1.6 Sensitivity check Action Result Start XRS9012 The instruments starts and indicates sensitivity 5 Adjust the sensistivity to 10 The no. 10 LED is flashing Apply 10ppm Hydrogen Signal should indicate: minimum LED 1

INFICON AB, Box 76, SE-581 02 Linköping, Sweden Phone: +46 (0) 13 35 59 00 Fax: +46 (0) 13 35 59 01 www.inficon.com E-mail: reach.sweden@inficon.com