until filters are dry. Once dry, place the filters shiny side up in petrislides. The petrislide do NOT need to be acid washed.

Similar documents
Cleaning Procedures in ICP-MS Laboratories

Standard Operating Procedure for Oil & Grease Method EPA1664A

UVM Cosmogenic Laboratory - Quartz Purity Testing

Radioactive In Situ Hybridization Protocol (HBMI Core University of Rochester Medical Center)

PRODUCT LINE OVERVIEW CLEAN SAMPLE PREP

RESOLUTION OIV-OENO MONOGRAPH ON TANNINS UPDATE OF THE METHOD FOR DETERMINATION OF POLYPHENOLS

Amerimade RCA Wet Bench SOP Page 1 of 13 Rev /22/2014

Major Hazards in Laboratory

Wet Bench Standard Operating Procedures

Experiment 4: Purification of Liquids by Distillation

Haz Waste Management

Solids, Total Filterable (Total Dissolved)

PDS 2010 Parylene Coater SOP

Rain Forest Quick Guide

GUIDANCE MANUAL. Model 900 BGI 900 HIGH VOLUME CASCADE IMPACTOR

Ion PGM Hi Q View Sequencing Kit

SPIRAL PLATE COUNT METHODS (Raw Commingled Cow Milk) IMS #4. [Unless otherwise stated all tolerances are ±5%] GENERAL REQUIREMENTS

Safety Lab Goggles are required! Apron is optional!

PDMS Casting Station Supplies

CEM Mars 5 Microwave Accelerated Reaction System

wash system V1.3 Important: Up to slides on one injection. Note: Getting need to turn it samples the next day. and put the slides in foil, put it

ChBE Safety Training. July Emily Edwards Lab Coordinator Chemical & Biological Engineering Tufts University 4 Colby St., Room

General Acids/Bases Bench Right and Left SOP

2. What are the advantages and disadvantages of plant tissue culture?

G Trap FliQ FPLC Columns

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX

Safety In Chemistry. Identified Risks and Hazards

PROTOCOL FOR FREE LIPID SOP

Laboratory Glassware Cleaning and Storage

Inherent Factors Affecting Soil ph. Soil ph Management

R I T Title: RCA Bench #2

Simple Chemical Tests of Potting Mixes Used For Container Crops

Title: Glassware Cleaning Copy No: ## SOP No.: 3.02/2.6/S Effective Date: June 18, 2013 Location: ### Glassware Cleaning

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

DISPOSABLE CONTAINERS FOR SUCTION APPLICATIONS FLOVAC CONTAINER WITH MECHANICAL VALVE

THE UNIVERSITY OF NEWCASTLE- SCHOOL OF BIOMEDICAL SCIENCES ASHMAN LABORATORY

Nebraska Soil Science Curriculum

Monitoring Nutrition for Crops

Standard Operating Manual. Ultra Fab Wet Station F Non-standard Processing Station

HOSTA MULTIPLICATION KIT

Test Tube Rack. Test Tube

APPENDIX A: MICROSCALE APPARATUS AND TECHNIQUES A.1 CRAIG TUBES: RECRYSTALLIZATION

Guide for Thin Film Evaporation Edited 7/22/13 by Stephen Albright, BJAM & Steven Palefsky

Transformation Protocol for Tobacco Abbreviated

Inherent Factors Affecting Soil ph. Soil ph Management

Soil Science Curriculum

Product Guide for Ludger Clean Post-endoglycosidase and Post-exoglycosidase Clean-up Plate

NMR SAMPLING ACCESSORIES CLEANING NMR SAMPLE TUBES

CPAP Washbasin Manual

General Guidelines to Assist with Microwave Digestion Method Development

Operator s Manual. Rev 3/14/ O Neil Rd, Macedon NY Telephone: (315) Fax: (315)

Mildew & Rot Proof Standards

AFRICAN VIOLET MULTIPLICATION KIT

Dedicated Analytical Solutions. The Complete Chemical Analysis Collection

DISPOSABLE CONTAINERS FOR SUCTION APPLICATIONS FLOVAC CONTAINER WITH HYDROPHOBIC VALVE

SPIKA NG ICE MAKER MAINTENANCE

R I T Title: Manual Processing Bench #2 Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 08/05/2008

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

Experimental Procedure

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Introduction. Sharps. Working Safely with Sharps at UAB (OHS_HS255) Objectives. Defined

Models TX-KE and TX-KP Toxic Gas Transmitters

FERN MULTIPLICATION KIT

Protocol Photolithography

TECHNIQUES IN MOLECULAR BIOLOGY BASIC TECHNIQUES: PIPETING AND STERILE CULTURE

Beakers hold solids or liquids that will not release gases when reacted or are unlikely to splatter if stirred or heated. Beaker

Biotech Staples for Education

Corning Storage Bottles Selection and Use Guide

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

HAVE A SAMPLING AND ANALYSIS PLAN (SAP)

Established in 1968, Pearl Life is one of the most prevalent kitchenware brands in Japan. Peal s wide variety of kitchen gadgets, organizers and

Laboratory filtration systems.

Thank you for choosing the Motif Duo Double Electric Breast Pump. Motif Duo Operation Manual

Operation Procedure for SMiF Wet Hoods (Solvent, Spin Coat, Acid, and Develop Hoods)

GREENHOUSE SCREENING PROTOCOL OF COMMON BEAN FOR NODULATION AND NITROGEN FIXATION

NMOS Fabrication Process Description

Revised: January 8, Goal: Deposit, expose, and develop standard positive photoresist films as an etch mask, liftoff mask, or sacrificial layer.

GREENLINE. Foratia Floor Washing Capsule 2. Foratia Bathroom Washing Capsule 3. Foratia Universal Cleaning Capsule 4. Foratia Degreaser Tablet 5

Beakers hold solids or liquids that will not release gases when reacted or are unlikely to splatter if stirred or heated. Beaker

Methods of solid phase extraction. Methods of solid phase extraction

QAQC LABS ANALYTICAL DRYING

The plating industry

SimpleVIS Viscometer Reference Guide P

GENERAL WET BENCH SAFETY AND USES SOP. October 2013 GENERAL SAFETY RULES

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

CNS Cleanroom Safety Orientation Form

Cold Soak Filtration Test (CSFT) of Biodiesel. ASTM D IP PM EA - CEN N CGSB 3.0 No Range

IFU 005 CLEANING / PREPARATION OF OUTDOOR HEAT EXCHANGE SURFACES & COILS. Date Approved

Procedures for Preparing and Receiving PUF Canisters and Filters

Task #1. Quick Overview: Wipe white residue build- up off one of the show tanks (class tank) Procedure: 10. Wash hands.

GB National food safety standard Determination of benzoic and sorbic acid in milk and milk products

Laboratory Glassware Washer G 7883 CD

Fibretherm. C. Gerhardt - Analytical Systems. Fibrebag-Systems

Mixing Training Course P/N Rev. D

PERFORM Operating Document

Laboratory Glassware Washer G 7883

Arabidopsis Growing Protocol-A General Guide Xiyan Li *

Standard Operating Manual. PEL Wet Station X Clean Organic Solvent Station

Photolithography SOPs

Transcription:

Bottle Washing LDPE bottles must be acid washed prior to use in trace element analyses. The process should take two weeks. Rinse bottles three times with house DI water. One week in 3M HCl (ACS Plus) bath. Rinse three times with Milli-Q. One in 0.5M distilled HCl bath. Rinse three times with Milli-Q then fill with Milli-Q. Acidify to 0.1M HCl with 6M distilled HCl and place in Ziploc bags for storage. Falcon tubes do not need to be cleaned as extensively as they are not used for trace element analyses. Rinse three times with house DI water. One week in 3M HCl (ACS Plus) bath. Rinse three times with house DI water and place in Ziploc bags for storage. Filter Types/Cleaning Methods 1. Polycarbonate/Teflon filters These can be washed in large batches (~100 filters) using the blue lidded plastic bins. Each of the baths should be made to 2 liters total volume. The volumes can be reused several times. 4M HNO 3 (ACS Plus) for 1 week Transfer to 3M HCl (ACS Plus) for 1 week. Repeat batch rinse with Milli-Q. until filters are dry. Once dry, place the filters shiny side up in petrislides. The 2. Polyether sulfone/metricel membrane/pall GN-6 3M HCl (ACS Plus) for 1 week. Repeat 3M HCl (ACS Plus) for 1 week. Repeat batch rinse with Milli-Q. 1

until filters are dry. Once dry, place the filters shiny side up in petrislides. The 3. Whatmann GFF/Millipore Quartz Place filters on top of a glass Petri dish in furnace Set the temperature to 450 degree for 5 hours Once cooled, place filters in 3M HCl (ACS Plus) for 1 week. until filters are dry. Once dry, place the filters shiny side up in petrislides. The Aerosol Filter Extraction Clean nitrile gloves are worn at all times when handling filters and conducting the filter leaching. Use a pair of clean plastic tweezers to handle the filters. It is best to lay some large Kimwipes down on the counter to provide a clean work surface. All work should be done inside a flow bench or hood. It is best to rinse the 125 ml LDPE sample bottles ahead of time (three times each with Milli-Q water), as this is a time consuming process. Falcon tubes do not require rinsing. Label each bottle and tube with the same information from the petri slide (generally the data and soluble/insoluble). Record the LDPE bottle numbers and their corresponding filter data in your log book. Write the bottle number of the leach on the bottom of the corresponding Petri slide, i.e. DI #1498. The Falcon tubes can be labeled with the collection date. 100 ml Ultrapure (Milli-Q) Water Leaches 18.2MΩ: The Milli-Q should be allowed to equilibrate with the air so that the ph is about 5.6. Fill the labeled 1-L LDPE bottle with DI, leaving some head space. Cap and shake the bottle, open the cap to allow the head space air to exchange, then re-cap the bottle and shake again. Do this five times then check the ph on a 10 ml subsample of the Milli-Q (poured into a clean, rinsed 25mL LDPE bottle) using the ph meter. Repeat the air equilibration until the ph is between 5.5 and 5.7. Rinse the filtration rig 3 times with Milli-Q and shake out the excess water prior to the first use and between each subsequent filter leach. Pour 100mL Milli-Q into a clean plastic graduated cylinder. Place a mixed cellulose ester backing filter in the rig and wet slightly with Milli-Q water. This will aid in laying the polycarbonate sample filter flat in the rig. Tightly seal the filter rig. 2

Connect the length of tubing from the vacuum pump to the Pump port on the shoulder of the bottom reservoir and turn on the pump. Cover the opposite hole in the rig with a clean, gloved finger and gently pour the volume of Milli-Q onto the filter within a few seconds, making sure that the entire filter is covered with solution. Allow the filter to suck dry for 5-10 seconds, then slowly remove the pump tubing from the rig to allow the vacuum to dissipate smoothly. Turn off the pump and carefully pour 25mL of the leach solution out of the filtration rig (through the finger plug hole, not the Pump hole!) and into a Falcon tube. Pour the remainder into a pre-labeled 125 ml LDPE bottle. Freeze the Falcon tube sample. The trace element sample can be acidified at this time by adding 300µL of 6M Teflon distilled HNO 3. Rinse filtration rig. The backing filter can be retained for multiple samples. We have used this same technique for filtering ocean samples with much lower dissolved metal concentrations without any evidence of cross contamination. Teflon Filter Extraction Milli-Q will not flow through a Teflon filter; therefore sonication, following Ying Chen s method, must be used to extract soluble elements. Cut the filter in half. Fill a polypropylene jar with 50mL of Milli-Q, add filter and seal. Seal lid of jar with Parafilm Sonicate for 30 minutes. Filter resulting solution through an acid-washed 0.4µm filter. Subdivide for ICP and IC analyses. Ion Chromatography (consult with Rob Franks regarding IC use) The frozen, unacidified aerosol extracts are analyzed for major anions and oxalate in two separate analytical runs. Major anions are measured using a 25µL sample loop and oxalate with the 200µL sample loop. ICP-MS (consult with Rob Franks regarding ICP use) Samples for ICP analysis are transferred by pipette to acid washed 7mL scintillation vials. Sample volume is 2mL. There is an ICP sequence specifically designed for analysis of soluble and digested aerosols. 1. Standard Preparation (see excel sheet ICPMS Standards 060409 ) There are two primary standards used to calibrate. Each is made through several steps using stock solutions. Standards should be made gravimetrically but the following are in nominal volumes. Standardization is completed using 5 solutions: Acid Blank, Low standard, High standard, 50% Low, and 50% High. The 50% standards are made by dilution with the Acid Blank. These standards should be spiked with the same volume ratio as the samples, e.g. 200µL spike per 2mL. 3

Other stock solutions made in 2% HNO 3 were 1000ppb Ti, 10ppm Fe, and 10ppm Al o Low mixed standard contains low concentrations of most of the trace elements and is derived from SPEX Multi-element Solution 2 (10ppm). 100µL SPEX Multi-element Solution 2 Final=10ppb 10mL 2ppm P standard (in Falcon tube) Final=200ppb 10µL 1000ppb Ti Final=1ppb 90mL Milli-Q 2mL 6M distilled HNO 3 o High mixed standard contains several alkalis from SPEX Custom Ion Standard as well as higher concentrations of Fe and Al 1mL SPEX Custom Ion Standard 1mL 10ppm Fe 2mL 10ppm Al 93mL Milli-Q 3mL 6M distilled HNO 3 o 2% HNO 3 /Acid Blank 96.76 ml Milli-Q 3.24mL 6M distilled HNO 3 2. Spike Solution Preparation Samples are spiked with an internal standard to monitor instrument drift during analysis - 200µL spike per 2mL sample o Spike = 100ppb Rh, 1ppm Sc o To make 50mL of spike 5mL Sc stock solution (10µg/mL) 0.2mL of 1ppm Rh 43.69mL Milli-Q 1.44mL distilled HNO 3 Aerosol Sample Digestion Digesting the insoluble aerosol fraction requires the use of concentrated hydrofluoric acid (HF). HF is extremely dangerous and potentially lethal; therefore users should receive specific safety training and follow lab safety guidelines. 1. Microwave Polycarbonate/HPWP This method can be used to dissolve the entire filter or a fraction of the original. Step 1: Microwave program Chen PC o 3mL conc. HCl + 3mL conc. HNO3 + 1mL conc. HF Step 2: Program W41 Step 2 o Add 2mL 5% Boric Acid Decant the digest solution from the digestion vessels into a wide-mouth 60mL LDPE bottle. 4

Rinse the inside of the vessel three times with Milli-Q and pour into the 60mL bottle. Place bottles on ~70 o hotplate with caps off and evaporate to dryness. Rehydrate with 50mL Milli-Q and acidify with 200µL 6M HNO3. 2. Microwave Teflon Digest half the filter. This filter section will not have been extracted and therefore represents the total aerosol fraction (insoluble + soluble). Step 1: Microwave program Teflon o 2mL conc. HNO3 Step 2: Microwave program Teflon St2 o Add 0.1 ml conc. HF Rinse, evaporate, and rehydrate following above. 3. Microwave Vessel Cleaning Rinse vessel five times with Milli-Q. If there is filter residue stuck in vessel, a pipette tip can be used to scrape off as much as possible. Fill vessel with 40-50mL of 1M HNO3. Run program Clean Pour off 1M HNO3 (this can be retained for reuse) and rinse five times with Milli-Q. 4. Hotplate HA/GN-6 Digest half the filter in 15mL Teflon jars. Step 1: Add 8mL Milli-Q, 0.5mL HNO3, 0.4mL HF Place on 80 o hotplate and take to dryness. Step 2: Add 2mL Milli-Q, 0.3mL HNO3, 0.3mL HCl, 0.1mL HF Place on 80 o hotplate and take to dryness. Fill graduated cylinder with 50mL with Milli-Q and use to rinse Teflon jar into narrow-mouth 60mL LDPE bottle. Acidify with 200µL 6M HNO3. To Clean: Rinse jar five times with Milli-Q. Use a Kimwipe to wipe out any residue and fill with 1M HNO3. Seal jar and place on 80 o hotplate overnight. Pour off 1M HNO3 (this can be retained for reuse) and rinse five times with Milli-Q. 5