OWNER S MANUAL FOR WOB L PISTON VACUUM PUMP MODELS: 2561B B C C C C-25

Similar documents
OWNER S MANUAL FOR DRY VACUUM PUMP MODELS: 2567B-01, 2567B-50, 2567C-02, 2567C-50

OWNER S MANUAL FOR DRY VACUUM PUMP MODEL: 2581

OWNER S MANUAL FOR DRY VACUUM PUMP MODELS 2585B-01, 2585B-50, 2585B-55 & 2585Z-50

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

OWNER S MANUAL FOR DRY VACUUM PUMP MODEL: 2019

OWNER S MANUAL FOR ASPIRATION / FILTRATION VACUUM SYSTEM MODELS: 2511 & 2515

OWNER S MANUAL FOR DRY VACUUM PUMPS MODELS: 2014 / 2032 / 2042 / 2034 / 2044 / 2037 / 2047

User Manual. Twin Piston Vacuum Pumps: Oil-less WVP Series: WVP12, WVP10, WVP5, WVP4, WVP3, and WVP2

OWNER S MANUAL FOR CHEMSTAR DRY MODELS: 2070, 2080

OWNER S MANUAL FOR DRY VACUUM PUMPS MODELS: / / / 2047

OWNER S MANUAL FOR DRY VACUUM PUMPS MODELS 2016, 2024, and 8111

OWNER S MANUAL FOR ASPIRATION / FILTRATION VACUUM SYSTEM MODEL: 2511

High Output Vacuum/Pressure Pump

OWNER S MANUAL FOR SELF-CLEANING DRY VACUUM SYSTEM MODELS: 2025, 2026, 2027 & 2028

OWNER S MANUAL FOR SELF-CLEANING DRY VACUUM SYSTEM MODELS: 2025, 2026, 2027 & 2028

USER MANUAL UN-200V / UN-200VH

12 VELOCITY OWNER S MANUAL OPERATING INSTRUCTIONS - MAINTENANCE - SAFETY - TROUBLESHOOTING

SSS Intrepid 16 FMS. Manual

WET/DRY VACUUM. QUEST for Continuous Improvement Windsor s Quality Management System is Certified ISO MODEL: T1. Operating Instructions (ENG)

Xaact Spot. Xaact Hot Spot

CC Series VACUUM PUMPS

SSS BPV- 40 Backpack Vacuum

GULPER 16SV 16 GALLON WET/DRY VACUUM WITH FRONT MOUNT SQUEEGEE

Operating Instructions

RecoverX-CAR Contaminated Automotive Refrigerant Recovery System Operation and Maintenance Manual

INSTRUCTION MANUAL. Full Size Gas Convection Oven

INSTRUCTION MANUAL 4HP WET / DRY VACUUM VC V 60 Hz 7A. Gal

30 Professional Cooling Unit. Model Number: XC3000. Owner s Manual READ AND SAVE THESE INSTRUCTIONS

Operations and Safety Manual

K Specifications. Max. Water Volume

MAXUM 5 5 Gallon Box Extractor

Operator s Manual. Histology Bath

OPERATING INSTRUCTIONS

Indoor/Outdoor Electric Patio Heater Model No: HP Operating Instructions HP Soleus Air International

HC Ceramic Tower Heater Operating Instructions HC Soleus Air International

K Specifications. Max. Water Volume

Operator s Manual. Wet/Dry Vacuum with Detachable Blower 12 Gallon Model No. VBV1210 FOR YOUR SAFETY

Model K 520 M Part No

HOTSPOT. Two Gallon Pump Up Spotter OPERATING & MAINTENANCE

MODEL 7000 SUCTION UNIT

CANISTER VACUUM MODEL SC-19

MN908 No. 282C Replaces 282B LB7013

92831 TEL: (714) FAX:

Fisher Scientific Model 281A Vacuum Oven. Installation and Operating Manual Rev. 0

LARGE AREA VACUUM LAV-30

MIGHTY PRO 1/4 HP CHILLER

HYBRID DRYER. PDH1 - PDH2 Series WARNING: OPERATOR MUST READ AND UNDERSTAND THIS MANUAL COMPLETELY BEFORE OPERATING THIS EQUIPMENT.

TFC 400. OPERATOR MANUAL Clarke MODEL /09 Rev C Document Number

READ AND UNDERSTAND THESE INSTRUCTIONS BEFORE OPERATING THE MACHINE

FLOOR MACHINE EDGER MINI SCRUBBER MODEL E-1

WET/DRY VACUUM MODEL SV-12

HYBRID DRYER. PDH1 - PDH2 Series. Save These Instructions

Operation and Maintenance Manual

RecoverX Oil-Filled Hermetic Refrigerant Recovery System. Operation and Maintenance Manual

Oil-Sealed Rotary Vane Pumps

Operator s Manual. IP-100 Immersion Probe Cooler

READ AND SAVE THESE INSTRUCTIONS. Air Boss Model T5200 Custom Packaged Electronic Air Cleaner. TRION

OPERATIONS MANUAL DALPH1002, DALPH1503 PELLET HEATERS PALPH1002, PALPH1503 PLATE HEATERS. Alliance Products, LLC Murfreesboro, TN

Operator s Manual. 6 Gallon Model No. VQ607SFD FOR YOUR SAFETY. Read and understand this manual before use Keep this manual for future reference

R100 Oil-Less Refrigerant Recovery Unit

Quick-Start Operating Guide Document No Acrylic Vacuum Chambers Copyright 2010 Terra Universal Inc. All rights reserved. Revised October 2010

Owner s Manual READ AND SAVE THESE INSTRUCTIONS

MODEL: 1407, , 1408, & INSTALLATION AND INSTRUCTION MANUAL

K 2.90 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS

K 3.97 M. Specifications or visit our website: In USA call: Part No

GENIUS 14 GENIUS 18 OPERATING & MAINTENANCE

Have a Question? Call toll-free: Owners Manual L WARNING. Project Series Paint Sprayers Important Safety Instructions

Installation & Operating Guide

MC MC MC MC MC833130

NEW. APPLIED VACUUM TECHNOLOGY. ASK THE EXPERTS!

Operator s Manual. 2.5 Gallon Model No. VOM205P FOR YOUR SAFETY. Read and understand this manual before use. Keep this manual for future reference.

Dry, oilfree Diaphragm Pumps

DLCLRA. INSTALLATION INSTRUCTIONS Outdoor Unit Single Zone Ductless System Sizes 36 to 58 TABLE OF CONTENTS

Single Phase Simplex SXL21=3, SXL24=3, SXH21=3, and SXH24=3

Appendix 1. Technical Drawings. Models CE5, CE7, CE10, CE14, CEN5, CEN7, CEN10

W Model 1RKU2 w/optional Wall mount. Unit Model Weight (Lbs) Kilowatts Volts- Phase Amps Fan Output Heat Output

OPERATOR S MANUAL. ET D ( ) ET Dx2 ( ) For the dealer nearest you consult our web page at

1 HP Air Mover. Owner s Manual

NEW. APPLIED VACUUM TECHNOLOGY. ASK THE EXPERTS!

Store Name: Location: Date Purchased: Serial Number (located on back of unit):

3-1/2 GALLON PARTS WASHER OPERATING INSTRUCTION AND PARTS MANUAL. Model # 4KTV9

WMHP Series R410a Heat Pump INSTALLATION INSTRUCTIONS

APPLIED VACUUM TECHNOLOGY VACUUM PUMPS & SYSTEMS

Quest Power Electric Heat EHS 62 Pro

Sanitaire. Owner s Guide. 10 GALLON Box Extractor SC6088 Series IMPORTANT

Models Series Series

HR1-08R-21 Reflective Heater Operating Instructions. Model No. HR1-08R Soleus Air International

Reacti-Therm I and III Heating Modules

Thank You & Congratulations On Your Purchase of: ARBE Model # SS-206 Table Top Dust Collector

HM & HM Ultra Thin Micathermic Heater. Model No. HM / HM Soleus Air International

CHAMPION PUMP OWNER S MANUAL

LC Series - Light Commercial Pump Station Installation and Operation Manual

AURA WIDE AREA VACUUM Parts Manual

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Safety, Operation and Maintenance Manual with Parts List

OPERATING INSTRUCTIONS

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

PORTABLE DEHUMIDIFIER

Parts and Instruction Manual ET Extraction Tool

Parts and Service Manual BURNISHER MODEL: MR , MR , MR , MR

Transcription:

WELCH-ILMVAC OWNER S MANUAL FOR WOB L PISTON VACUUM PUMP MODELS: 2561B-50 2563B-24 2561C-50 2563C-24 2561C-56 2563C-25 Warning Not recommended for pumping acid, base or organic vapors or gases. Warning Be sure to properly identify intake and discharge before using pump. See section 2.5 For outside of U.S. and Canada, contact your local Gardner Denver sales offi ce. Welch-Ilmvac 5621 W. Howard Street Niles, IL 60714 Phone: (847) 676-8800 (Technical Support) Fax: (847) 677-8606 E-Mail: gdwelchvacuum@gardnerdenver.com Web-Page: www.welchvacuum.com Part No. 642826 Printed in USA

INSTRUCTION WARNING AND CAUTION PLEASE READ BEFORE OPERATION While reading your manual, please pay close attention to areas labeled: WARNING AND CAUTION. The description of each is found below. WARNING Warnings are given where failure to observe instruction could result in injury or death to people. CAUTION Cautions are found where failure to observe the instruction could result in damage to the equipment, associated equipment and process. These units conform to the SI International system of units of measurement. The following symbols (with recommendation of IEC1010 ) of warning will be found on the pump. Caution - Refer to accompanying documents Caution - Risk of electrical shock Caution - Hot surface WARNING Motor includes a self resetting thermal cutout and the pump could restart without actuation under fault condition. 2

TABLE OF CONTENTS Section 01 - Safety Information 1.10 Warning to prevent injury 1.20 Warning to reduce risk of electrical shock 1.30 Warning to reduce risk of electrocution 1.40 Warning to reduce risk of explosion or fi re Section 02 - Installation 2.10 Introduction 2.20 Unpacking 2.30 Pump mounting 2.40 Pump location / Environmental conditions 2.50 Intake and discharge provisions 2.60 Electrical power 2.70 Vacuum connections 2.80 Vacuum and gauges 2.90 Traps Section 03 - Operation 3.10 Starting procedures 3.20 Leak detection 3.30 Pressure range 3.40 Shutdown procedures Section 04 - Maintenance 4.10 General Maintenance Section 05 - Troubleshooting 5.10 Vacuum problems 5.20 Troubleshooting table Section 06 - Specifi cations 6.10 Specifi cation chart 6.20 Pumping speed curve Section 07 - Warranty Section 08 - Exploded views and parts list 8.10 Dimensional Drawing 8.20 Exploded view and parts list 3

Section 1: SAFETY INFORMATION 1.10 WARNING: TO PREVENT INJURY... 1.11 Never operate this product if it has a damaged cord or plug. If it is not working properly, has been dropped, damaged or has fallen into water, please return the product to a Welch service center for examination and repair. 1.12 Keep the cord away from heated surfaces. All electrical products generate heat. To avoid serious burns never touch unit during or immediately after operation. 1.13 Never block any air openings or place it on a soft surface where the openings may be blocked. The air openings are for ventilation of the motor inside the housing. Keep all air openings free of lint, dirt and other foreign objects. 1.14 Never drop or insert fi ngers or any other object into any openings. 1.15 Do not operate this product where oxygen is being administered. 1.16 This pump is thermally protected and can automatically restart when the protector resets. Always disconnect power source before servicing. 1.17 Wear safety glasses and goggles when operating this product. Never point any air nozzle or air sprayer toward another person or any part of the body. 1.18 Use only in well ventilated areas. The motor on all pumps are totally enclosed fan cooled 1.19 Do not use any tools or attachments without fi rst determining maximum air pressure for that tool or attachment. Be sure to properly identify intake and discharge before using pump. See Section 2.50 1.20 WARNING: TO REDUCE RISK OF ELECTRICAL SHOCK 1.21 Do not disassemble. Disassembly or attempted repairs if accomplished incorrectly can create electrical shock hazard. Refer servicing to qualifi ed service agencies only. 1.22 Unit is supplied with a three pronged plug. Be sure to connect pump to a properly grounded outlet only. 1.30 WARNING: TO REDUCE RISK OF ELECTROCUTION 1.31 Do not use this product in or near area where it can fall or be pulled into water or other liquids. 1.32 Do not reach for this product if it has fallen into liquid. 1.33 Never operate this product outdoors in the rain or in a wet area. 1.40 WARNING: TO REDUCE RISK OF EXPLOSION OR FIRE 1.41 Do not use this pump in or near explosive atmospheres or where aerosol (spray) products are being used. 1.42 Do not pump anything other than atmospheric air. 1.43 Do not pump combustible liquids or vapors with this product or use in or near an area where fl ammable or explosive liquids or vapors may exist. 1.44 Do not use this product near fl ames. WARNING Failure to observe the above safety precautions could result in severe bodily injury, including death in extreme cases. 4

Section 2: INSTALLATION 2.10 INTRODUCTION 2.11 This manual has been compiled not only for the care and maintenance of the Welch Dry Vacuum Pump now in your possession, but as a helpful reference and guide to prevent many problems which occur if used improperly 2.20 UNPACKING 2.21 Carefully remove the Vacuum Pump from the shipping case and unfasten and remove the wooden skid. Preserve all paperwork for future reference. If damage has occurred from shipment, a claim must be fi led with the carrier immediately; preserve the shipping carton for inspection by the carrier. If you are required to communicate with your dealer or Welch Vacuum Technology, be sure to include your order numbers for quick identifi cation. Do not return the pump to the factory without fi rst calling for a return goods number. 2.30 PUMP MOUNTING 2.31 Rubber feet are attached to the pump. Rubber feet are excellent for applications involving a semi-fl exible surface such as a bench top; they help to isolate noise and eliminate creeping. All Vacuum Pumps should be mounted on a horizontal plane. 2.40 PUMP LOCATION / ENVIRONMENTAL CONDITIONS 2.41 Pump Location: The Vacuum Pump should be located preferably in a clean, dry, and well ventilated area. Please be sure not to block the ventilation holes located on the motor housing. The pump should be placed where the surrounding temperature remains between 10 C and 40 C (50 F and 104 F). Always check to insure the location chosen is protected from direct or indirect moisture contact. Welch recommends that the pump be installed at the highest point within the system to prevent possible water condensate from entering the pump. The pump should be located as closely to its system in order to utilize it most effi ciently. 2.42 Pump Environmental Conditions: The pump is rated for indoor use only. Maximum altitude 2000 meters. Operating temperature range 5 C to 40 C. Maximum relative humidity of 80% for temperatures up to 31 C decreasing to 50% at 40 C. Rated for +/-10% of supply voltage. Pollution Degree 2, Installation Category II. WARNING Don t pump fl ammable or explosive gases or vapors or operate this pump in an atmosphere containing fl ammable or explosive gases or vapors. 2.50 INTAKE AND DISCHARGE PROVISIONS 2.51 The intake fi tting is a NW25-fl ange / 1/4 NPT. A muffl er is supplied for the exhaust. WARNING Bursting Hazard- use on exhaust port the silencer or air handling parts suitable for pressures not less than 100 PSIG Be sure to call Welch technical service prior to start-up at (847) 676-8800 if you have any questions. 5

2.60 ELECTRICAL POWER 2.61 Power Source Review Review the power source and the motor rating to be sure they agree in voltage, phase, and frequency. Serious damage may occur to the motor if it is connected to an improper voltage. All Welch pumps must be grounded. Grounding reduces the risk of electric shock in the event of an electrical short circuit. The plug must be plugged into a properly grounded outlet. Consult your local electrical codes if you have doubts. 2.62 Overload Protection Motor thermal overload protection is made available by the motor manufacturer as an aid to minimize motor failure. Overload protection is a standard feature on both 50 and 60 Hz singlephase-motors. The motors have automatic overload protection. Automatic reset protection is designed to reset itself after a predetermined cooling period. If the fault to the drive remains unaltered, the motor will cycle on and off until the fault is corrected. WARNING The motor is thermally protected and will automatically restart unexpectedly when the overload device resets. Don t pump fl ammable or explosive gases or vapors or operate this pump in an atmosphere containing fl ammable or explosive gases or vapors. 2.63 Identifi cation Symbols: Power Off Power On 2.64 Grounding Instructions This product should be grounded. In the event of an electrical short circuit, grounding reduces the risk of electric shock by providing an escape wire for the electric current. This product is equipped with a cord having a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded in accordance with all local codes and ordinances. DANGER Improper installation of the grounding plug can result in risk of electrical shock - If repair or replacement of the cord or plug is necessary, do not connect the grounding wire to either fl at blade terminal. The wire with insulation having an outer surface that is green with or without yellow stripes is the grounding wire. - Check with a qualifi ed electrician or serviceman if the grounding instructions are not completely understood, or if in doubt as to whether the product is properly grounded. Do not modify the plug provided; if it will not fi t the outlet, have the proper outlet installed by a qualifi ed electrician. - This product is for use on a nominal 120V circuit, and has a grounding plug that looks like the plug illustrated in Sketch A in Figure #1. A temporary adapter, which looks like the adapter illustrated in sketches Band C, may be used to connect this plug to a 2-pole receptacle as shown in sketch B if a properly grounded outlet is not available. - The temporary adapter should be used only until a properly grounded outlet (Sketch A) can be installed by a qualifi ed electrician. The green colored rigid ear, lug, etc. extending from the adapter must be connected to a permanent ground such as a properly grounded outlet box cover. Whenever the adapter is used, it must be held in place by a metal screw. 6

The adapter shown in Figure #1 B and C is not acceptable in Canada. 2.65 Extension Cords Motor thermal overload protection is made available by the motor manufacturer as an aid to minimize motor failure. Overload protection is a standard feature on both 50 and 60 Hz single phase-motors. The motors have automatic overload protection. Automatic reset protection is designed to reset itself after a predetermined cooling period. If the fault to the drive remains unaltered, the motor will cycle on and off until the fault is corrected. 2.70 VACUUM CONNECTIONS 2.71 All Dry Vacuum Pumps come with intake and discharge hose barbs which accept 3/8 ID rubber pressure/vacuum hose. Hose clamps should be used to hold the hose in place. Since all four models operate in the viscous fl ow regime, the small diameter of the hose will generate minimal conductance loss. For best results, Welch recommends the length of the tubing between the pump and the chamber be kept as small as possible. 2.80 VACUUM GAUGES 2.81 Typical in the rough vacuum range, a dial vacuum gauge is used to measure pressure in mm Hg or Inches of Hg. The dial vacuum gauge gives negative pressure that is pressure below atmospheric. The reference point for the vacuum gauge is atmospheric pressure. Please keep in mind that atmospheric pressure tends to vary from day to day. As a result of variability, the dial vacuum gauge will indicate slightly different maximum vacuum readings from day to day. 2.90 TRAPS 2.91 The need of a Trap The pumps will handle humid air. All wetted aluminum parts are treated for corrosion protection from moisture. All other wetted parts are stainless steel. If there is a chance liquid may be drawn from the process under evacuation, Welch recommends a liquid trap be placed between the process and the pump. A simple liquid trap is a fi ltering fl ask. When a heavy load of water vapor is evolved from the vacuum process, a cold trap is recommended to help prevent damage to the pump mechanism. The cold trap, immersed in a suitable Dewar fl ask, is installed so that the water vapor may come in contact with the surfaces of the trap and condense. Commonly used refrigerants are liquid nitrogen or dry ice and acetone or alcohol. Dry ice provides suffi cient cooling to freeze out most heavy water vapor loads. A variety of cold traps are available from Welch. 7

The symptom of a high water vapor load is if you have droplets coming out of the exhaust port. If you see droplets of water, Welch recommends you use a cold trap to capture the water before entering the pump or dilute the water vapor stream by adding dry nitrogen to the gas fl ow. Please call our customer service department for additional information at (847) 676-8800. 2.92 The Care of a Trap When using a cold trap the refrigerant should be maintained at a high level in the fl ask to keep the trap at a uniformly low temperature. If the trap is rewarmed it may allow re-evaporation of the condensate. The refrigerant add tube on the liquid nitrogen trap should not be obstructed as the refrigerant boil-off can produce dangerously high pressures. If the trap becomes saturated it should be disconnected from the system, drained, and cleaned. An increase in pressure in the vacuum system will normally indicate that the trap has become saturated. To clean the trap, remove the trap from the system and allow the trap to warm up and rinse off the condensate with a suitable solvent in a fume hood. Thoroughly clean and dry the trap before reinstalling into the system. 8

Section 3: OPERATION 3.10 STARTING PROCEDURES 3.11 Starting A Welch Pressure/Vacuum Pump Before attaching the pump to a system, familiarize yourself with the function and action of the pressure/vacuum pump that you have acquired. Review the power requirements as described in Section 2.60. Welch recommends running the pump for a few minutes to warm it up, before use. The warm-up improves the pumps ability to handle humid air. 3.12 Cleanliness Take every precaution to prevent foreign particulates from entering the pump. Particulates will damage the pump s performance. If you fi nd that particulates come off the process during evacuation, a particulate trap in the fore line will work. A simple, inexpensive trap may be made by placing glass wool in a glass or plastic tube. Screens must be inserted to hold the glass wool in place. WARNING The pump is not recommended for pumping acid, base or organic vapors or gases. Serious damage to the pump will shorten the pump s service life. In addition, pumping fl ammable vapors or gases can lead to serious safety hazard leading to fi re or explosion. 3.20 LEAK DETECTION The importance of eliminating all leaks in a vacuum system is obvious. The pump must remove this added volume of leaked gas to maintain the desired vacuum. Leaks for these pump can be located by slightly pressuring the system and painting the suspected area with a thick soap solution. Escaping air will produce soap bubbles. 3.30 PRESSURE RANGE Vacuum pumps are designed to be run from slightly below atmospheric to their maximum vacuum level on the intake side. The pumps also may be run from atmospheric to their maximum rated pressure rating. Consult the Specifi cation Table (See Section 6, Specifi cations) for the ratings of your specifi c model. 3.40 SHUTDOWN PROCEDURES After use, Welch recommends the pump be run for about 2 minutes disconnected from the vacuum process. The air pumped through the mechanism will purge out water vapor or droplets of water condensate that may have formed on the inside of the pump. This purge of the pump mechanism helps prevent corrosion. Section 4: MAINTENANCE 4.10 GENERAL MAINTENANCE Welch dry vacuum pumps are 100% oil-free. The pump employs a non-lube piston and cylinder. No maintenance is necessary for the bearings. All bearings are sealed and permanently lubricated. Lubrication should not be attempted. The units are built for continuous duty operation with the quietness and durability of a diaphragm, but with piston performance. 9

Section 5: TROUBLESHOOTING 5.10 VACUUM PROBLEMS Welch dry vacuum pumps are 100% oil-free. The pump employs a non-lube piston and cylinder. Leakage, contamination and unusual outgassing are the general causes of problems with poor vacuum. To operate at maximum effi ciency a system must be thoroughly clean. If the system is completely clean and free from leaks, and unwarranted vacuum problems still exist, the pump should be checked. A simple criterion for the condition of the pump is the determination of its maximum vacuum capability. This can be accomplished by blocking of the intake and reading the vacuum level on the gauge (See Section 2.8). 5.20 TROUBLESHOOTING TABLE Poor Pumping Speed Poor Vacuum Loud Unit Possible Cause Corrective Action X X X Damaged valves Replace fl apper valves X X X Debris in valves Remove debris and check for valve damage X X X Damaged gasket Replace gasket X X X Loose head screw Tighten head screw X X X Loose fi tting Tighten fi tting 10

Section 6: SPECIFICATION 6.10 SPECIFICATION CHART 6.20 PUMPING SPEED CURVE Welch Model 2563 Free Air Displacement CFM (L/min.) @ 60Hz 2.12 (60) CFM (L/min.) @ 50Hz 1.77 (50) Ultimate Pressure Torr (mbar) 5 (7) Motor Horsepower Horsepower 1/3 Tubing needed I.D. in Inches (mm) 1/4 (7) Shipping Weight Weight lbs (Kg) 16.5 (7.5) Catalog Number Catalog Number Wired for 115V, 60Hz, 1Ph with 2563B-24 North American Plug Catalog Number Wired for 220V, 50Hz, 1Ph with 2563C-24 European Schuko Plug Catalog Number Wired for 100V, 50/60Hz, 1Ph with 2563C-25 North American Plug 11

Section 7: WARRANTY UNPACKING Inspect the pump carefully. If any damage has occurred, fi le claim with the carrier immediately. Save the shipping container for carrier to inspect. OPERATING PUMP Refer to the enclosed Instruction/Operation Manual for all information to properly operate and maintain the pump. WARRANTY This Welch-Ilmvac Inc. product is warranted to be free from defects in material and workmanship. The liability of Welch-Ilmvac Inc. under this warranty is limited to servicing, adjusting, repairing or replacing any unit or component part which in the judgment of Welch- Ilmvac Inc. has not been misused, abused or altered in any way causing impaired performance or rendering it inoperative. No other warranties are expressed or implied. The method of executing this warranty: servicing, adjusting, repairing or replacing shall be at the discretion of Welch-Ilmvac Inc. Vacuum pumps that have been used for any period, however short, will be repaired under this warranty rather than replaced. The warranty is effective for one year from the date of original purchase when: 1. The warranty card has been completed and returned. 2. The product is returned to the factory or other designated service centers, freight prepaid. 3. The product in our judgment is defective through no action or fault of the user. If the product has become defective through misuse, abuse, or alteration, repairs will be billed regardless of the age of the product. In this event, an estimate of the repair costs will be submitted and authorization of these charges will be required before the product is repaired and returned. To reduce additional charges and delays either within or outside of the warranty period, contact Welch-Ilmvac Inc. @847-676-8800 for a return authorization number. Products without a return authorization number will be refused by our receiving department. Before shipping, properly pack the pump, insure it against loss or damage, and on the outside of the pump packaging and the packing slip write in the return authorization number. Pumps damaged due to improper packaging are the customer s responsibility. 12

8.10 MODEL 2563 DIMENSIONAL Section 8: PARTS LIST AND EXPLODED VIEWS 13

8.20 MODEL 2563 PARTS LIST AND EXPLODED VIEW Item No. Part No. Description Service Kit 2563K-03 Gasket Kit 2567K-04 Foot Kit 2500K-04 1 2563K-01** Connecting Rod Assembly - - - 2 2567K-02 Valve Plate Assembly 2 - - 3 * Cylinder O-Ring 2 2-4 * Valve Plate Head Gasket 2 2-5 * Cylinder Sleeve 2 - - 6 * Piston Cup - XL9650 2 2-7 * Suction Cup Feet - 4 8 * Retainer Hold Down Screw 2 2 - * These parts are only available in kits. ** These parts should be repaired by qualifi ed service personnel only. 14

Section 9: APPLICATION INFORMATION 9.10 VACUUM SYSTEM FOR VACUUM CHAMBER, GLOVE BOX, ANNULAR SPACE & TRANSFER LINE -2561B-50-2561C-50-2561C-56 9.20 SYSTEM COMPONENTS* Item Catalog No. Description No. 1 - PTFE Coated Aluminum Lid 2 - Vacuum Regulator 3 - PTFE Encapsulated Gasket 4 - Metal Intake Hose Barb for 3/8 ID Hose 5 - Kynar Hose Barb 6 1415B Glass Jar, Inlet Trap 1415C Plastic Jar, Inlet Trap 7 726021 Vacuum Gauge, in. Hg, cm Hg 726024 Vacuum Gauge - mbar/kpa 726025 Pressure Gauge - psi/kgcm 8 2561B-50 Vacuum System 115V 60Hz 1Ph 2561C-50 Vacuum System 230V 50/60Hz 1 Ph 2561C-56 Vacuum System 100V 50/60Hz 1 Ph 9.30 SYSTEM OPERATION Welch s new portable oil-free general utility roughing pumps provide continuous, reliable vacuum that your vacuum chamber, glove box, annular space and transfer line applications require. The pumping mechanism is an economical twin head oil-free Wob-L known for reliability and durability. These rugged pumps handle 20,000 plus roughing cycles from atmospheric to 5 Torr before needing seal maintenance. The pumps come completely equipped and ready to use. An inlet trap to help prevent ingestion of fl uids into the pump, a dial gauge for continuous vacuum level monitoring, and a vacuum regulator to adjust vacuum levels. Just plug it in and you re up and running. Models 2561B-50, 2561C-50 & 2561C-56 are suited for 1.5 cubic foot ovens or smaller. The pumping speed of the 2561 at rough vacuum levels is equal to oil-seal rotary vane pumps of similar size. This makes the 2561 an economical, oil-free solution for replacing oil-seal rotary vane pumps. These pumps are therefore highly recommended for standard duty laboratory applications, general vacuum work and where quick roughing of a chamber is needed. The 2561 Pump Models are for use in standard duty applications. This means that they are not recommended for pumping acidic, basics or organic vapors or gasses. If you are looking for a pump to handles these types of conditions, contact a Welch representative and they will be able to assist you in selecting the correct pump for your application. * System components available as kit Cat. No. 2565K-50 15

Welch-Ilmvac 5621 W. Howard Street Niles, IL 60714 Phone: (847) 676-8800 (Technical Support) Fax: (847) 677-8606 E-Mail: gdwelchvacuum@gardnerdenver.com Web-Page: www.welchvacuum.com Copyright 1997-2007 Gardner Denver Welch Vacuum Technology, Inc. Welch is registered trademarks of Gardner Denver