AFA 500 FUME HOOD ALARMS

Similar documents
LABCONCO 1000 GUARDIAN DIGITAL MONITOR Operating and Instruction Manual

DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670

AFA1000/1 MK3 Airflow Monitor

Static Pressure Alarm For Hazardous Location Fume Hoods

Installation & Calibration Guide

Model 405 AIRGARD Fume Hood Monitor

AirGard 385 Fume Hood Monitor

FUME HOOD MONITOR ALNOR AIRGARD 200

Laboratory Airflow Monitors & Controls TYPE C STANDARD & NIGHT SETBACK LED & DIGITAL - VAV SYSTEMS

HVAC Controls Laboratory. Model EVERWATCH Face Velocity Monitor. Operation and Service Manual. P/N , Revision E October 2002

Alnor AirGard 410HE Air Flow Monitor for Hazardous Environment Fume Hoods

FUME HOOD MONITOR ALNOR AIRGARD 335

PRESSURA ROOM PRESSURE CONTROLLER MODELS 8630-SC 8630-PC

PRESSURA ROOM PRESSURE MONITOR MODELS 8630-SM 8630-PM

Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL.

SUREFLOW ROOM PRESSURE CONTROLLER MODELS BAC

SUREFLOW ADAPTIVE OFFSET CONTROLLER MODELS BAC

Document No

External Wireless Sounder

GAS MONITOR Model Model

FW-RA-LED Remote Multiplex Annunciator Panels

EVERWATCH FACE VELOCITY MONITOR MODEL 8610

The Icon Series. User Guide. CDV25 Version 1.0. Iconopm1.pm

Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL.

CRITICAL ENVIRONMENT MONITOR ALNOR AIRGARD 350 CEM

OWNERS MANUAL. CARBON MONOXIDE DETECTOR MODELS Panel Mount & Remote Detectors (R) Guardian Avionics 1951 E. AIRPORT DRIVE TUCSON, AZ.

External Wireless Sounder

Corrosion Resistant Fluid and Air Handling Systems.

Critical Environments. Model FHC50/FHM10. Fume Hood Controller/Monitor. Operation and Service Manual. P/N , Revision B October 2010

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

Duct Mount. Installation Instructions

CO Guardian LLC Document: E. AIRPORT DRIVE Date: 11/15/05 OWNERS MANUAL

Model # Airflow Monitor Warranty. Installation, Operation, Maintenance Manual. Analog Airfl ow Monitor Model Number #51403

for LABORATORY REFRIGERATOR MODEL TELR-145/ L/200L CAPACITY REVISION

CIRRUS PRO. User Manual and Responsibilities ASPIRATING FIRE DETECTOR

ProTalk. Expander. Operating Manual Model B1292. July 28, 2009 Rev. 1.01

LV-S-5. Instruction Manual DIRECT CONTACT STRAY VOLTAGE DETECTOR GENERAL DESCRIPTION HOW IT WORKS

Safety Siren Pro Series3 Radon Gas Detector

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model COSTAR 24VC-e Single Relay

Using a Pressure Pan To Diagnose Duct Leakage

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

Model Gas Alarm Panel APPLICABILITY & EFFECTIVITY. This manual provides instructions for the following Sierra Monitor products:

PRESSURA ROOM PRESSURE CONTROLLER MODEL RPC30

Air Cleaning Equipment, Inc. 303 N. Main St. Broadway, NC iers.com

Model OEM-2 INSTRUCTIONS FOR USE

TAG MR VOLTAGE DETECTORS AND ACCESSORIES. Instruction & Operation Manual

Isolation Relay: If Needed. # 9. NO relay for TT To Thermostat NO relay. MZ308 Exquisite Heat Controller

OWNERS MANUAL MODEL 451 (-101 & -201)

GROUP OF COMPANIES. MERLIN STEALTH Electric Fence Energizer. Users Manual

SIARGO MEMS FLOW SENSING PRODUCTS MEMS Mass Flow/Clog Sensors VC Siargo Ltd. ISO 9001 ISO OHSAS 18001

CD6102 / CD6102 OC / CD6102-2

M Series Fan Filter Units

636 and 646 Keypads. User s Manual

installation, operation & Maintenance manual Floor-mounted Induction units acb 30/35

Models LBW-420-LEL (24 VDC powered) Ammonia Leak Detector

OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN

PAC-2750 ELECTRIC COOLING FAN CONTROLLER

SPECTRA 1727 USER S GUIDE

Hydronic Manometers. Hydronic Manometer. Model HM670. Owner s Manual

Ultrasonic Reflective Sensor

VX SERIES Wireless Thermostat with Occupancy Sensor

Carbon Monoxide (CO) Detecting Ventilation Fan Controller Model 120VC Single Relay (100/25 PPM) (200/35 PPM)

Reacti-Therm I and III Heating Modules

installation, operation & Maintenance manual Floor-mounted Induction units

PRESSURA ROOM PRESSURE MONITOR MODEL RPM10 AND RPM20

GEYSERWISE INSTRUCTION MANUAL

DH400. Accessories. NOTE: Set up for this sensor should be performed by an AAADM-certified installer.

WATCHMAN Model LBW-WATCHMAN Ammonia Leak Detector

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

CR2003 ROBERTS. 3 band Auto set PLL Synthesised Clock Radio. Sound for Generations. Please read this manual before use

LV-5 Direct Contact Low Voltage Detector and LV-5/K01 Kit including LV-PT Tester, Holster and Available Accessories

TrueClean ToteCleaner

TrueClean ToteCleaner

M113S Users Manual. MERLIN STEALTH M113S Electric Fence Energizer GROUP OF COMPANIES. PDF created with pdffactory trial version

Model 4001 Series Single Channel Controller

Installation and Operations Manual

KP600F/L LED/LCD KEYPADS. Owner's Manual. Use with AV6500 & SP850 Control panels

INSTALLATION AND OPERATING INSTRUCTIONS FOR THE VEHICLE-MOUNTED RADIATION DETECTION SYSTEM

STATUS ALARM HISTORY ALARM RESET

installation, operation & Maintenance manual acb ceiling-mounted cassette models

0$; Installation Instructions. Computerized Control Panel '(6&5,37,21 63(&,),&$7,216 DE5011 1

RR-PM1200 OPERATING AND PROGRAMMING GUIDE

Design Criteria. Installation

PRESSURA ROOM PRESSURE CONTROLLER MODEL RPC30

ELECTRIC FIRE PUMP CONTROLLER

Security Alarm System Owner's Guide 8 sector CDS Dialler 8-24 sector Control Dialler 8-24 sector Control Securitel Version 1.0 lcopm.

WITH OMRON E5CC CONTROL

G-TECTA Cellaguard Carbon Dioxide and Oxygen Fixed Monitors

Portable Induction Ranges

WIRING DIAGRAMS R410A MODELS PAC 2OAC/2OACH CAC OWC PWC

installation & operation manual

ANALOX 101 D2 PORTABLE OXYGEN MONITOR USER MANUAL

Ion Genesis II Pump Controller Digital Level Control with Pump Alternation and High Water Alarm

INSTALLATION INSTRUCTIONS

APM110 LOW AIR VELOCITY ALARM

Controllers. Instruction Manual WARNING

Appearance and sensing characteristic Specification (temperature range) Model 10 to 70 C ES1B. 2 mm 20 mm 40 mm 60 mm. 40 dia. 60 dia.

User s Manual For Model Aero-551 (Rev. Orig.)

Analox 1000 Series. User Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK

GMS-10 RVS Series. Oxygen Sensors Accurate oxygen measurement WARNING WARNING DESCRIPTION TYPICAL APPLICATIONS FEATURES MISUSE OF DOCUMENTATION

Transcription:

AFA 500 FUME HOOD ALARMS Operating and Instruction Manual 19/7/03 Model AFA 500 Built-in or Remote sensor 2 Relay inputs 1 Relay output Com port Used for alarm indication and monitoring on Fume Cupboards 1

OPERATOR DISPLAY PANEL Built-in AIRFLOW SENSOR Not visible when Remote AIRFLOW SENSOR is used LED indicators Function buttons Calibration. ENTER also used as Mute button for audible alarm 2

Connection details :- R1 OUTPUT RELAY Plug in power adapter for mains socket. (To suit local requirements ) Relay output 2 Rated 110V 2A AC1 WARNING REFER TO INSTALLATION INSTRUCTIONS BEFORE CONNECTING Standard options :- 230V 50Hz -- UK 3-pin 220V 50Hz -- Euro 2-pin 120V 60Hz -- USA 2-pin PSU AFA 500 Night set-back Relay input Com port RS232 COMMS PORT INPUT 1 OPTIONAL REMOTE SENSOR SOCKET Sash Proximity switch Bn Bk Bl 15V DC POWER SUPPLY INPUT TEL MADE IN ENGLAND www.tel-uk.com Flying lead with plug 15 V DC Sash Micro Switch Sash High Micro-switch ( Optional ) Flying lead with plug and SM5 airflow sensor 12V DC Sash High Proximity switch ( Optional ) Flying lead from Proxy switch with plug 12V DC Remote SM5 Airflow Sensor ( Optional ) Notes:- 1. The socket for the remote SM5 Airflow Sensor is not provided as standard but is available on request. The unit operates with the in-built sensor OR the remote sensor. 2. The Sash High alarm can operate with the Micro switch OR the Proximity Switch. Both connection points are available as standard. 3

1.1 General Description All systems comprise of the following components :- 1 AFA 500 Alarm unit, 1 AC power supply If the Sash Alarm System option is ordered there will also be a sash micro switch or proximity switch. If the Remote Airflow Sensor is ordered there will be remote sensor provided Operator Features --- the alarm has the following operator features :- Safe LED - Alarm LED - Green LED (Not flashing) will be displayed if the airflow is greater than the Low air alarm point. Red LED (Not flashing) will be displayed if the airflow is lower than the Low air alarm point. Sash High - Red LED (Flashing) will be displayed when the Sash is raised above the max safe working opening. Audible Alarm -- the Audible alarm will sound ( can be muted ) in the Air Fail and the Sash High alarm condition Night Set-back -- when the Night Setback input is activated the Audible alarm will be muted and the Green LED will flash on/off ENTER --- the alarm has an Enter button -- this is multi-functional as follows :- Press Enter momentarily when Low Air alarm is sounding will mute the alarm Press Enter momentarily when Sash High alarm is sounding will mute the alarm and initiate a repeat timer that will re-sound the alarm if the Sash is not lowered to a safe position before the end of the time period Press Enter for 5 secs will gain access to Calibration mode SET -- used during the airflow Calibration of the alarm 4

External Connections -- the alarm unit will have the following connection points :- Input 1 --- volt free relay input ( close contact to activate the input ) This input is configured as :- NIGHT SETBACK Output R1 --- volt free relay output - ( contact closes on activation ) This input is configured as :- LOW AIR ALARM Sash High Input --- a. Connection point for Sash High micro switch. ( Switch contact to close and remain closed in Sash High condition ) b. Connection point for Sash High proximity switch. ( Switch contact to close and remain closed in Sash High condition ) Note:- Use input a. OR input b. for the Sash High alarm Remote Airflow Sensor -- Plug in connection for airflow sensor Note:- The remote Airflow Sensor socket will only be available for units ordered without the built-in Airflow Sensor Com Port --- to enable connection to Laptop or PC for full diagnostics, logging or setting up and for communications to building computer system ( BMS) Power supply --- low voltage DC power supply 15V DC 5

1.2 Alarm Configuration / Calibration The alarm is supplied with a factory configuration. The only part of the configuration that can be changed is the setting for the Sash High repeat timer time delay -- if required this can be changed by connection to a Laptop or PC via the RS232 com port. The alarm has two modes for the calibration of the airflow alarm point. These two modes are selectable via a jumper connection on the back of the alarm PCB. The jumper is labelled as CAL and is situated on the edge of the PCB above the Com port. The two calibration modes are :- a. With the Jumper connected - The airflow is reduced to the alarm point mechanically and this airflow is captured by the alarm. The airflow is then restored to the normal operating value and the Low Air alarm will activate if the airflow subsequently falls to the alarm point b.with the Jumper not connected - This is a two point calibration method. The airflow is set to the normal operating value and this value is captured by the alarm and taken to be 100%. The airflow is then increased to one and a half time the normal operating value and this value is captured by the alarm and taken to be 150%. The Low Air alarm will then activate if the airflow subsequently falls to 80% of the normal operating value. See Quick Start Installation below for details of the Calibration procedures 1.3 Start up When unit is powered up the following sequence of events occur :- 1. The 12V DC power is applied to the airflow sensor and the alarm then performs a self test on the functions, LEDs and audible alarm ( approx 2 sec ) and then initiates a delay timer of 30 secs to allow the airflow sensor to stabilise. 2. During the whole of the 30 sec period all alarms and relay outputs are inhibited and the Red & Green LEDs will be permanently ON. 3. At the end of the 30 sec delay the unit performs one of two options :- a. If the alarm calibration has been previously completed the unit goes to normal operating mode ( Run ) b. If the unit has not been calibrated -- the Red & Green LEDs will flash on/off and the audible alarm will be muted. It is then possible to press the Enter button for 5 secs and go into the calibration mode (See Quick Start Installation below for details of the Calibration procedures) 6

1.4 Events / actions Safe airflow Airflow above alarm level ( eg > 80 fpm ) Green LED on Low airflow Airflow below alarm level for longer than the low air delay time ( 5 secs) Red LED on ( Not flashing ) Audible alarm sounds ( Beep on/off every 1 sec ) -- can be muted via Enter pushbutton Low air relay R1 operates. Reset : -- when airflow rises above Low air level for longer than the low air to safe air delay time ( 2 secs ) the Low air alarm resets automatically Sash High When the input configured as Sash High is activated ( Micro switch or Proximity switch ) Red LED on ( Flashing ) Audible alarm sounds ( Beep on/off every 1 sec ) Audible can be muted via Enter pushbutton -- this silences the alarm and initiates a repeat timer ( factory set to 5 mins ) After the delay time the alarm re-sounds ( and can be re-muted). During this time the Red LED flashes on / off. Reset: -- when Sash lowered to safe position and input de-activated. Night set-back When input configured as Night set-back is activated Green LED on ( Flashing) Audible alarm muted Airflow Sensor Error The connection and each element of the airflow sensor are monitored at all times. In the event of a problem with the sensor the audible alarm will sound using a different and distinctive tone best described as a modulating siren effect The audible alarm can only be silenced by re-connecting the sensor ( if a remote sensor is being used ) or switching the power to the unit off. 7

2.1 Quick Start Installation Follow the instructions below for installing and commissioning the unit. :- Before connecting the unit it is important to decide on the type of calibration that is to be used on the installation. The two methods are described in section 1.2 Alarm Configuration / Calibration above. 1. Fit the alarm to the Fume Hood using the cut-out details provided with the unit --- see page 11 2. If using a remote airflow sensor fit to the Fume Hood using the cut out and installation details provided --- see page 11 & 12 and connect the telephone style airflow sensor plug-in cable to the sensor and the back of the alarm unit --- see typical connection diagram on page 13 & 14 3. Plug in the power adapter to a Mains AC power socket and connect the flying lead to the alarm unit --- see typical connection diagram on page 13 & 14 4. Power up the unit and wait at least 30 secs while the sensor temperature stabilises. During this time the Red & Green LEDs will both be on ( not flashing ) If the unit has not been previously calibrated the Red & Green LEDs will begin to flash on/off at the end of the 30 sec start up time delay but the audible alarm will not sound. If the alarm has been calibrated it will go into normal operation. 5. Calibration :- Alarm Point Capture ( With CAL jumper on PCB connected ) a. Open the Sash to the normal operating height and adjust the face velocity to the Low Air alarm value using a calibrated instrument to check the value. b. Press and hold the Enter button for 5 secs to go into the Calibration mode. This is indicated by both Red and Green LEDS flashing on/off together will the audible alarm sounding ( Beep on/off 4 times every 1 sec ). c. To initiate the alarm point calibration press and hold the ENTER and the SET button at the same time. The unit will then sample the airflow for a 5 sec period during which time the Green LED goes off and the Red LED flashes on/off. The audible alarm continues to sound during this period and if the sampling is successful will give a two tone beep at the end of the period and the unit will then go automatically into the Run mode. If the buttons are released during the sampling period or if the airflow is fluctuating more than the pre-set value the audible alarm will give a lower frequency buzzing sound for a short period and then go back into the calibration mode. If this occurs re-press the ENTER and SET buttons to repeat the airflow sampling. d. When complete re-set the airflow to the normal value and the unit will go to the Safe running condition with the Green LED on. The unit will now function and go into the ALARM condition if the Fume Hood face velocity falls below the alarm value. 8

Calibration continued:- Two Point Capture ( With CAL jumper on PCB not connected ) a. Open the Sash to the normal operating height and adjust the face velocity to the normal operating value using a calibrated instrument to check the value. b. Press and hold the Enter button for 5 secs to go into the Calibration mode. This is indicated by both Red and Green LEDS flashing on/off together with the audible alarm sounding ( Beep on/off 4 times every 1 sec ). c. To initiate the normal airflow (100%) calibration press and hold the ENTER and the SET button at the same time. The unit will then sample the normal airflow for a 5 sec period during which time the Green LED goes off and the Red LED flashes on/off. The audible alarm continues to sound during this period and if the sampling is successful will give a two tone beep at the end of the period and the unit will then go automatically into the higher airflow calibration mode. If the buttons are released during the sampling period or if the airflow is fluctuating more than the pre-set value the audible alarm will give a lower frequency buzzing sound for a short period and then go back into the calibration mode. If this occurs re-press the ENTER and SET buttons to repeat the airflow sampling. d. The higher airflow calibration mode is indicated by both Red and Green LEDS flashing on/off together with the audible alarm sounding ( Beep on/off for 2 secs every 10 secs ). Close the Sash down until the face velocity rises to one and a half times the normal operating value used in a. above using a calibrated instrument to check the value To initiate the higher airflow (150%) calibration press and hold the ENTER and the SET button at the same time. The unit will then sample the higher airflow for a 5 sec period during which time the Green LED goes off and the Red LED flashes on/off. The audible alarm continues to sound during this period and if the sampling is successful will give a two tone beep at the end of the period and the unit will then go automatically into the Run mode. If the buttons are released during the sampling period or if the airflow is fluctuating more than the pre-set value the audible alarm will give a lower frequency buzzing sound for a short period and then go back into the calibration mode. If this occurs re-press the ENTER and SET buttons to repeat the airflow sampling. The unit will now function and go into the ALARM condition if the Fume Hood face velocity falls below 80% of the normal operating value. 2.2 Calibration Notes :- 1. The two calibration methods are intended to give the installer two options for calibrating the alarm. 2. Using the Alarm Point Capture method it is necessary to adjust the face velocity on the fume hood using a mechanical damper ( or fan speed control if available) to the desired alarm point and this is sampled by the unit. It is then necessary to re-adjust the face velocity back to the normal operating value. This method produces a very accurate alarm point at a fixed value but involves getting access to the ductwork or fan speed controller 9

3. The second method of Two Point Capture is slightly more involved but does not require any access to the ductwork or fan speed controller and is generally a quicker solution. The principle behind this method is as follows :- The output from the airflow sensor is linear over the normal operating range of the face velocities on fume hoods. If we therefore capture the airflow sensor output at two known points we can then calculate the alarm point and give an alarm when the airflow falls below this value. However because the AFA 500 does not have a digital display of the face velocity we can not enter particular values into the alarm. For this reason we take the normal operating value of the face velocity with the sash open to the max safe working opening and call this 100% -- this point is captured during the first sample. Then by closing the sash it is possible to create a face velocity which is one and a half times the normal operating value ie 150% -- this point is captured during the second sample. We then use these two values to calculate the sensor output at 80% and this becomes the Low Air alarm point. The actual value of the normal operating face velocity is not important and in each case the alarm point will be 80% of this value. For example:- A normal face velocity of 100 fpm would give an alarm point of 80fpm A normal face velocity of 80 fpm would give an alarm point of 64 fpm 4. The face velocity readings on the open sash may vary at different points on the measuring grid by up to 20 fpm. This is quite acceptable in terms of the fume cupboard performance so long as no individual point is below the designated Low Air alarm point.the figure entered for the calibration point can be taken as the average value of all the measuring grid readings or could be taken as the individual lowest point on the grid. For most fume hoods this low point is on the bottom row in the centre and is a convenient position to measure and for future reference when checking the alarm during annual maintenance. 5. Take time when measuring the face velocities for the calibration procedure to allow for the velocities across the open sash to stabilise. If the velocities are changing or are turbulent during the sampling period the alarm will detect this and give a low frequency buzz at the end of the sample indicating that the sample must be repeated... 10

3.0 Dimensions (1.67" ) (1.67" ) 42.5mm 42.5mm 4mm (0.16" ) 148mm 118mm (4.65" ) 126mm ( 4.96") 88mm Alarm Panel Dimensions 4mm (0.16" ) 85mm (3.35" ) Panel Cutout Dimensions 118mm x 85mm ( 4.65" x 3.35" ) 2 x Fixing Holes 2.5mm dia for 2 x No. 4 Self tapping screws provided 12mm (0.47 ) 12mm (0.47 ) 25mm (0.98 ) 25mm (0.98 ) 2 x Holes for self tapping fixing screw. [ 4mm (0.157 )Fixing Hole provided on SM5 sensor ] 74mm (2.91 ) 17mm 20mm (0.79 ) 30mm Dia (1.18 ) Sensor Hole for Vent Tube 10mm (0.39 ) 10mm (0.39 ) 74mm (2.91 ) SM5 Sensor Dimensions ( Rear view ) 12mm Dia. (0.47 ) for SM1 cable 27mm (1.06 ) SM5 Sensor Panel Cutout Dimensions ( Front view ) NOTE: The optional Remote sensor and connecting socket on the AFA 500 are not included as standard 11

4.0 SM5 Airflow Sensor Installation AFA 500 OR 100mm 100mm SM5 Sensor Normal Sash working height. eg. 500mm 25mm Female adapter 25mm Male bush 25mm Tube AFA 500 or SM5 Sensor Open to the laboratory Fume cupboard inner skin Fume cupboard outer skin It is very important to position the AFA 500 or SM5 airflow sensor in the correct position to give long term stable reading of the face velocity. Please read the INSTALLATION NOTES below and if in doubt contact us for further advice. INSTALLATION NOTES :- 1. The AFA 500 or SM5 sensor must be positioned where it can " see " the room pressure of the laboratory.the back connection spigot of the sensor is designed to accept a 25mm OD tube which should be connected to the inner chamber of the fume hood. (This tube and fittings is known as the " vent kit " ) The ideal position for the end of the 25mm tube for most fume hoods is 100mm back from the sash glass and 100mm higher than the normal sash opening height through the inner side wall. 2. If possible mount the sensor on the front of the fume hood and use a short length of tube. Tube lengths of more than 1metre or smaller diameter will restrict the airflow through the sensor. This will lead to too much sensitivity being required to calibrate the unit which can lead to some instability of the reading or incorrect readings at low velocitiesl. 3. For fume hoods with a single skin side wall or a double skin with a small gap between them it may not be possible to achieve the ideal sensing position using a flexible tube. With a single skin side wall it is possible to fix the sensor on the outside of the fume hood and connect directly to the inner chamber in the ideal position. This method can only be used for up to two fume hoods when they are positioned side by side ( using the two outer walls ). An alternative method is to mount the sensor on the front of the fume hood and connect using a short flexible tube to a rigid wall tube attached to the inner side wall. The open end of this rigid wall tube should be positioned in the ideal position ie. 100mm back from the sash and 100mm higher than the normal sash opening. Fume hoods with a high internal height can present a difficulty because the tube length to reach the ideal position may be longer than 1 metre. In this case it is better to use a tube no longer than 1 metre which may result in a sensing position higher than the ideal. When fitting a sensor to a narrow wall fume hood for the first time it may be necessary to try various combinations of rigid and flexible tube to find the best combination and position. 4. The sensor should not be mounted in a position were it is subject to draughts from the laboratory air input or ventilation system. 12

5.0 Typical Wiring Diagram Standard Alarm with built-in Airflow Sensor R1 OUTPUT RELAY Plug in power adapter for mains socket. (To suit local requirements ) Relay output 2 Rated 110V 2A AC1 WARNING REFER TO INSTALLATION INSTRUCTIONS BEFORE CONNECTING Standard options :- 230V 50Hz -- UK 3-pin 220V 50Hz -- Euro 2-pin 120V 60Hz -- USA 2-pin PSU AFA 500 Night set-back Relay input Com port RS232 COMMS PORT INPUT 1 OPTIONAL REMOTE SENSOR SOCKET Sash Proximity switch Bn Bk Bl 15V DC POWER SUPPLY INPUT TEL MADE IN ENGLAND www.tel-uk.com Flying lead with plug 15 V DC Sash Micro Switch Sash High Micro-switch ( Optional ) Sash High Proximity switch ( Optional ) Flying lead from Proxy switch with plug 12V DC 13

5.1 Typical Wiring Diagram Alarm with REMOTE Airflow Sensor R1 OUTPUT RELAY Plug in power adapter for mains socket. (To suit local requirements ) Relay output 2 Rated 110V 2A AC1 WARNING REFER TO INSTALLATION INSTRUCTIONS BEFORE CONNECTING Standard options :- 230V 50Hz -- UK 3-pin 220V 50Hz -- Euro 2-pin 120V 60Hz -- USA 2-pin PSU AFA 500 Night set-back Relay input Com port RS232 COMMS PORT INPUT 1 OPTIONAL REMOTE SENSOR SOCKET Sash Proximity switch Bn Bk Bl 15V DC POWER SUPPLY INPUT TEL MADE IN ENGLAND www.tel-uk.com Flying lead with plug 15 V DC Sash Micro Switch Sash High Micro-switch ( Optional ) Flying lead with plug and SM5 airflow sensor 12V DC Sash High Proximity switch ( Optional ) Flying lead from Proxy switch with plug 12V DC Remote SM5 Airflow Sensor 14

6.0 LIMITATION OF WARRANTY AND LIABILITY Seller warrants that this product, under normal use and service as described in the operator s manual shall be free from defects in workmanship and material for a period of twelve (12) months, or the length of time specified in the operator s manual, from the date of shipment to the customer. This limited warranty is subject to the following exclusion :- a. Batteries and certain other components when indicated in specifications are warranted for a period of 90 days from the date of shipment to the customer. b. With respect to any repair services rendered, Seller warrants that the parts repaired or replaced will be free from defects in workmanship and material, under normal use, for a period of 90 days from the date of shipment to the customer c. Seller does not provide any warranty on finished goods manufactured by others. Only the original manufacturer s warranty applies. d. Unless specifically authorised in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, any goods which are incorporated into other products or equipment by the Buyer. All goods returned under warranty shall be at the Buyer s risk of loss, Seller s factory prepaid, and will be returned at Seller s risk of loss, Buyer s factory prepaid. The foregoing is IN LIEU OF all other warranties and is subject to the conditions and LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. THE EXCLUSIVE REMEDY OF THE USER OR PURCHASER, AND THE LIMIT OF LIABILITY OF SELLER FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES IN CONNECTION WITH THIS PRODUCT ( INCLUDING CLAIMS BASED ON CONTRACT NEGLIGENCE, STRICT LIABILITY, OTHER TORT, OR OTHERWISE ) SHALL BE THE RETURN OF THE PRODUCT TO THE FACTORY OR DESIGNATED LOCATION AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF THE SELLER, THE REPAIR OR REPLACEMENT OF THE PRODUCT. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING, REASSEMBLY OR REINSTALLATION COSTS OR CHARGES. NO ACTION, REGARDLESS OF FORM, MAY BE BROUGHT AGAINST THE SELLER MORE THAN ONE YEAR AFTER THE CAUSE OF ACTION HAS ACCRUED. The purchaser and all users are deemed to have accepted the terms of this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended or modified nor may any of its terms be waived except by a writing signed by an authorised representative of the Seller. 7.0 Contact us :- For further information on our range of airflow alarms and controls please contact us at :- HSE HOLLAND SAFETY EQUIPMENT 726 McKinley Ave. Libertyville, Illinois 60048 Phone : 847-680-9930 Fax : 847-680-9938 e-mail: info@hollandsafety.com web site: www.hollandsafety.com TEL Temperature Electronics Ltd Export sales :- Tel : + 44 161 257 2541 Fax : + 44 161 228 8817 e-mail: web site: xportsales@exportsales.com www.tel-uk.com 15