Oxygen Plant Installation for Paktia Hospital (Gardez City) Technical Specification: 1. Central Supply Oxygen Generator - Application: Medical Generator Specific. - Capacity: 11.2 Nm³/hour - Purity: 95% +/ 1% - Min. inlet pressure: 6 bar(g) - Outlet pressure: 4 bar(g) - Max. oper. pressure: 10 bar(g) - Air consumption: 2.48 m³/min FAD - Hose connection: 3/4" hose - Color: RAL7012 Grey Climatic Conditions - Ambient temp.: 10 C to 35 C - Altitude: less than 1800 meters a.s.l. Compr. Air Spec. - Air delivery: 2.50 m³/min FAD - Air quality spec.: ISO 8573.1:2010.2.4.1 - Dew point: +3 C - Filtration grade: 0.01 micron Display Control DCP2+ 25001 NEW DESIGN Oxygen monitor for oxygen generators (standard)/om1 26101 Range 0,1 100% Includes alarm function (through control) External audio / visual alarm/eav1 26202 Audio and visual alarm in one unit. Can be placed anywhere. V isual alarm is active whenever an alarm is present in the system. Audio will turn on when an alarm appe ars but can be turned off from for control panel Purity control excl. valvekit/26011 Purges product outside valid purity range. Placed after product tank. Valve kit 2/VALVE002 Compressor SK 25 5.5 8.0 barg screw compressor 400V / 3Ph / 50Hz (15.0 kw) SIGMA CONTROL BASIC Filter Package 0210 1 micron and 0,01 micron filter Coal tower 40 1 1/2'' Filter 00210 S 0.01micro 750 L Air tank DN40 (1 1/2'') 11 bar, PED, with handhole 3/4" hose Generator and Other Outlet Connections
470 L Product tank DN20 (3/4'') Filter 00035 A Active carbon Filter Medical Sterile Filter Type A028MS New With Oxygen Manifold Automatic Switch. 2. Air Dryer (Air Dryer for Oxygen Plant) Refrigeration dryer Donaldson Buran DC0225AB Flow 225 m3/h - 01 3. Oxygen Cylinder Filling Station - 01 With High Pressure O2 Compressor Booster pump, Flow 1.7 to 3.2Nm3/h Discharge Pressure 150bar Filling Ramp for 6 Cylinders in one time total 12 Cylinders/day 4. Electric Power Stabilizer 100KVA (SVC) - 01 3Phase, 50/60MHz 1. With Electric Wiring and others from source to Oxygen Room 2. Electric Control Box with all the Electric fuses and others 3. Service & Installation 5. Mechanical room for the Oxygen Plant - 01 Room Thermometer Electric Power Control Box with all the Circuit Breakers & fuses 18000BTU Air conditions will installed. Ventilation Fan will install in the wall 30x40cm Note: wiring system for all electrical equipment s and should be completely installed according to the requirement. 6. HEPA Filters for the Mechanical room - 02 30 X 40cm HEPA Filters should installed in the Mechanical room with all the wiring system 7. Medical Gas Alarm System for Plant - 01 Alarm Switch for Oxygen room 8. Medical Gase Control Box (1Gases ) with Alarm - 02 1. The box is stainless steel 2. The the hose is brass 3. with alarm 4. in the wall 5. GAS: Oxygen & Medical Air
6. Hose size: 19mm 9. BS Standard Medical Oxygen Gases Outlets with regulators - 11 Emergency Rooms: 9 Outlets + Oxygen Regulator Isolation room: 2 Outlet + Oxygen Regulator With the installation of around 110 meters Medical grade copper pipe from source up to each ro om beds, T connectors, Extenders etc. 10. Standard Medical Oxygen Regulators - 22 For the beds with outlets already available and need to supply regulators and connect to the Central supply and check all of the leakages. A.ICU Male: 5 Beds in 5 rooms to be connect with Oxygen Plant it need 50metters Medical grade copper Pipe + Oxygen regulators B.OT room: 6 Beds in 6 rooms to connect Pipe for Manifold almost 50metters Medical grade copper Pipe need to be installed + 6set Oxygen Regulator with hanging type oxygen extender connector C.NICU/Recovery: Incubator 1 Bed + need almost 10 meters copper Pipe need to be install + oxygen regu lator D.Gynecology: 10 beds in 10 separate rooms and in two halls need 80metters Medical grade copper Pipe s need to be installed + oxygen regulators Note: In addition to the above mentioned number and description for oxygen outlets in different wards/sections of the hospital, the supplier/company need to consider additional outlets for oxygen supply in other wards as well, and the total oxygen outlets will be around 90 including the above mentioned ones. 11. One year free Maintenance with Oil changing and filters replacing - 01 (Quarterly service and maintenance of machine)
Medical Air Plant for Paktia Hospital (Gardez City) Technical Specifications 1. MEDICAL AIR PLANT (DUPLEX) Med Air Plant 580 Liters /min Duplex 2 x 5.5kw Oil Lubricated Screw Compressors 1 x 350ltr Vessel 4 bar outlet, 50hz SYSTEM DESCRIPTION & KEY FEATURES One or more vertical Air Receivers fitted with an Automatic and Manual Drain Valve, a Pressure Gauge and a Pressure Relief Valve. Dryer assembly incorporating:- Duplex 1 micron oil/water Separation Filters each fitted with Differential Pressure Gauge and Automatic Float Type Drain Valve. Duplex Desiccant Regenerating Dryers with Pressure Gauges fitted to each column. Duplex activated Carbon filters for odour and dust removal each fitted with Manual Drain Valve. Duplex 0.01 micron Medical Sterile filters each fitted with Differential Pressure Gauge and Manual Drain Valve. Duplex Pressure Reducing Regulator set, and Line Pressure Relief Valve Assemblies. A Dryer Control Unit incorporating Duplex PLC s for Cycle Timing, Economy/Continuous Switches for each dry er, a Duty dryer selector switch), a Dew Point Sensing Circuit to detect moisture after the Dryers and a Line Pres sure Sensors and a pressure switch in each dryer sub assembly. A Plant Control Unit incorporating a Pressure Gauge, a Duty Selector Switch, Receiver pressure transducer and a Plant Emergency Run Pressure Switch. Status Indicators for Normal, Plant Fault, Plant Emergency, Reserve Low syst em fault, Pipeline Pressure Fault. All Alarm Contacts are duplicate allowing complete dual Alarm System or BMS/Alarm Connection. 2. Mechanical room for the Medical Air Plant The plant will be installed within the same room allocated for Oxygen Plant; there is no need for any additional resources like air conditioner, heap filters, control box etc. However, any power supply and Medical Air Plant safety related accessories can be considered in terms of cost estimation. 3. BS Standard Medical Air Outlets with Regulators ( see details above) Medical air outlets will be installed for a total of 90 beds in different wards/service units of the hospital with details given for Oxygen Plant in Section 09 & 10 with special note for your consideration. Special Note: Remember that you can find out details for estimating Medical Air Outlets under the technical specification for Oxygen Plant, and we may let you know that Medical Air Plant will be installed within the same room that has been allocated for Medical Oxygen Plant. 4. One year free Maintenance with Oil changing and filters replacing - 01 (Quarterly service and maintenance of machine)