Animal Tissue Culture SQG 3242 Equipment and Lab design. Dr. SITI PAULIENA MOHD BOHARI

Similar documents
Biological Safety Cabinets or Fume Hoods Which one to use?

Biological Safety Cabinets. Biological Safety Cabinets

VP, Facilities and Construction Management

Class II Biosafety Cabinets

PART III. Laboratory equipment

Safe Operating Procedure (Revised 2/10)

Research Document Type: PROCEDURE. Effective Date: January 01,2015 Author Steven Hayes

Environmental Health and Safety Biological Safety Program Guidance Document

Adv. 8.4 Guidelines for Biosafety Cabinets Rev

BIOLOGICAL SAFETY CABINET

Florida International University. Biosafety Cabinet Program

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

Safe Use of a Class II Biological Safety Cabinet

LSU Inter-Institutional Biological and Recombinant DNA Safety Committee (IBRDSC)

Primary Containment for Biohazards: Selection, Installation and Use of Biological Safety Cabinets

Appendix C IBC Safe Use of Biosafety Cabinets SOP

Biological Safety Cabinets. Presented by Mr. Rene Soetens President

GUIDANCE NOTE 3: SAFE USE OF EQUIPMENT

IBC Position on Bunsen Burner Use in Biosafety Cabinets

Clean Air Device. Yale University. Office of Environmental Health & Safety. Primary Containment Device

THE IMPACT OF FACILITY HVAC SYSTEMS ON BIOLOGICAL SAFETY CABINETS

ESCO. A Guide To Biosafety & Biosafety Cabinets

Biosafety Cabinets and Fume Hoods Supplemental Material (OHS_BIO304)

Fume Hood Safety TYPES OF FUME HOODS

CLEAN AIR SYSTEM.

Section 1.0: LAB VENTILATION TABLE OF CONTENTS. C. General Ventilation Considerations 3. D. Pressurization 7. E. Supply Air Arrangements 8

CHEMICAL FUME HOOD PROGRAM

INCREASE YOUR LAB EFFICIENCY

INCREASE YOUR LAB EFFICIENCY

NEW! Thermo Scientific 1300 Series B2 Class II, Type B2 Biological Safety Cabinet

Control of Biological Hazards

6/19/2015 BIOLOGICAL SAFETY CABINET. Types of Classes. Types of BSC.

Biological Safety Cabinet Failure Response Procedure

Appendix A.4.1. Safe Fume Hood User Guide

CBC s Inspection Findings and Possible Remedies Thereto

SCIF SOP for BSL 2 samples on Aria3

Quick Reference Guide to Cabinet types

Ergonomic and Green Class II s

AES Environmental. Operation and maintenance manual. VWS Series. Laminar Vertical Flow clean work station

5.0 Installation Instructions

Laboratory Safety Audit

EMERGENCY RESPONSE GUIDE IMMEDIATE EMERGENCY PROCEDURES

OPERATION AND MAINTENANCE MANUAL

Laboratory Safety Audit

LABORATORY INSPECTION CHECKLIST

Ventilation System Description. September 8, 2000 HETA

Personnel & Product Protection: AGuideTo Biosafety Enclosures

XPert Balance and Bulk Powder Enclosures

Biological Safety Cabinet

University of Maryland Baltimore Environmental Health and Safety updated 1/3/2018 General Lab Inspection Common Findings and Corrective Actions

Emergency Response Guide for Laboratories

LABGARD NU-602 CLASS II TYPE A2 ANIMAL HANDLING BIOLOGICAL SAFETY CABINET

University of California, Davis Laboratory Self-Inspection Checklist

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

LAMINAR FLOW FOR ALL YOUR NEEDS

BARRIER TECHNOLOGY RABS & ISOLATORS FOR ASEPTIC PROCESSING

Factory & Hospital hygiene, Chap 21(7 th edition) Principles of GMP, Chap 23 (8 th edition )

Precision Containment for All Applications

This document provides general instructions on the use of chemical fume hoods to protect lab personnel against exposure to hazardous materials.

Microbiological Safety Cabinets

Commissioning of a Mobile CL3 Laboratory: Lessons Learned

Major Hazards in Laboratory

Laboratory Control and Safety Solutions Application Guide

INCREASE YOUR LAB EFFICIENCY

ENVAIR lab CO2 Incubator

LAB SAFETY FOR CUSTODIANS

Laboratory Fume Cupboards

The Logistics of Filling and Freeze Drying using Multiple Isolators. The use of isolator technology to fill and lyophilize pharmaceuticals

Laboratory Information

BioMAT 2 BioMAT 2-S2. Class 2 Microbiological. Safety Cabinets. Simple. Safe. Quiet. Efficient. To EN 12469:2000

MIT Emergency Response Guide

Safety In Chemistry. Identified Risks and Hazards

Effective Contamination Control with CO 2. Incubators

ORION. Safety Class 2 100% exhaust. ALMOST Rocket Science

BioSafety Cabinet- Class 100

Clean Bench Animal Transfer Stations

Department of Civil and Environmental Engineering University of Waterloo

LABORATORY CABINETS 103

Faster BH-EN and BHG Class II Microbiological Safety Cabinets

A look into our complete product collection.

Faster UltraSafe Class II Microbiological Safety Cabinets. Protection, safety, reliability. And more.

DEPARTMENTAL SAFETY ISSUES - SAFETY TECHNOLOGIST / INFECTION CONTROL PRACTITIONER

IVF/ART Series Laminar Flow Cabinets

Models: 1828 and and 1846

Formalin Vaporizer The Premium Solution for Biological Safety Cabinet Decontamination. Esco Formalin Vaporizer Model FV-00_

BIO-MEDICAL WASTE MANAGEMENT

UNC Chapel Hill Design Requirements for Level 3 Biological Safety Laboratories & Animals

ENVAIR lab CO2 Tri-Gas Incubator

Gelaire BH-EN and UltraSafe cabinets

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Application: Response procedures for major and minor spills, including: organic solvents acids bases dry chemicals mercury hydrofluoric acid

WTAMU Academic Research Environmental Health and Safety (AR-EHS) Laboratory Inspection Checklist

LABORATORY EQUIPMENT. A. Section Accessibility for People with Disabilities. B. Section Plumbing Basic Materials and Methods

MCDB4100: LABORATORY SAFETY GUIDELINES

CONSIDERATIONS ON FACILITY AND ENGINEERING CONTROLS IN THE WAKE OF USP <800>

Thermo Scientific 1300 Series A2

SECTION SPECIAL EXHAUST SYSTEMS. A. Laboratory Ventilation and Hazardous Exhaust Systems. 3. Chemical Fume Hoods and Storage Cabinets

Direct Vent Training & Installation Instructions for Venting 40/50 gallon models Telescoping Vent Kit (standard)

DRAFT. Module 4. Decontamination & Environmental Cleaning

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Transcription:

Animal Tissue Culture SQG 3242 Equipment and Lab design Dr. SITI PAULIENA MOHD BOHARI

Sign Symbol? Biohazard sign Means? Hazardous For labeling the biological materials that carry a significant health risk, including viral samples and used Examples? Virus or toxin samples, medical wastes

Laminar flow cabinets Sterile environment Vertical or horizontal air flows Protect samples and/or worker Class 1, 2 or 3 Sample>>>>worker

What are Biological Safety Cabinets (BSC)? Containment and protection devices used in laboratories working with biological agents with a primary purpose of protecting the laboratory worker and the environment from viable organisms.

What is NOT a BSC!!! Chemical Fume Hoods Conventional lab fume hoods should NEVER be used to contain biological hazards. Find a Biological Safety Cabinet!

Class I Biological Safety Cabinets Class I BSC is a ventilated cabinet with an inward airflow and HEPA filters at its outlets. It was previously referred to as the CDC Hood and served a valuable function in its time to protect personnel and the environment. Because it offers no product protection, it has been essentially obsolete for the past several decades.

Class II Biological Safety Cabinets As biomedical researchers began to use sterile animal and cell tissue culture systems, BSC s utility needed to be expanded from simply protecting the operator to protecting the product as well. Class II BSC s are Laminar Flow Biological Safety Cabinets that protect personnel, product, and environment. They provide an inward airflow to protect personnel, a downward flow of HEPA filtered air to the work area to protect the product, and then exhaust HEPA filtered air to protect the environment from particulate and aerosol hazards.

Class II Type A BSC Class II, Type A hoods are used to protect personnel, product and environment from biological aerosols and particulates. These hoods offer personnel protection through negative pressure airflow into the cabinet. To protect the product, the work area in the cabinet is continuously bathed with ultra-clean air provided by the supply HEPA filter. Approximately 70% of the air of each cycle is recirculated through this supply HEPA filter. The remaining air is discharged from the hood through the exhaust HEPA filter, protecting the environment. Although not required, most Class ll hoods have the capability of being vented to the outside.

Class II Type B BSC The Class II, Type B cabinet originated when the National Cancer Institute designed the Type 2 (later-type B) biological safety cabinet. It was created for manipulations of minute quantities of hazardous chemicals such as carcinogens when used in research with in vitro biological systems. Carcinogens used in cell culture or microbial systems require both biological and chemical containment.

Class II BSC Modifications Class II BSCs can be modified to accommodate special tasks: The front sash can be modified by the manufacturer to accommodate a microscope A rigid plate with arm holes can be added if needed. The work surface can be designed to accept a carboy, a centrifuge, or other equipment that requires containment Good cabinet design, microbiological aerosol tracer testing of the modification, and appropriate certification are required to ensure that the basic BSC systems operate properly after modification. Maximum containment potential is achieved only through strict adherence to proper practices and procedures.

Class III BSC or Glovebox The Class III BSC was designed for work with Biosafety Level 4 microbiological agents and provides maximum protection to the environment and the worker. It is a gas-tight enclosure with a non-opening view window. Access for passage of materials into the cabinet is through a dunk tank that is accessible through the cabinet floor or double door pass-through box such as an autoclave that can be decontaminated between uses. Reversing that process allows for safe removal of materials from the Class III BSC. Both supply and exhaust air are HEPA filtered. Exhaust air must pass through two HEPA filters or a HEPA filter and an air incinerator before discharge to the outdoors. Airflow is maintained by a dedicated independent exhaust system exterior to the cabinet which keeps the cabinet under negative pressure

Class III BSC or Glovebox Long, heavy-duty rubber gloves are attached in a gas-tight manner to ports in the cabinet and allow for manipulation of the materials isolated inside. Although these gloves restrict movement, they prevent the user's direct contact with the hazardous materials. The trade-off is clearly on the side of maximizing personal safety. Depending on the design of the cabinet, the supply HEPA filter provides particulate-free, albeit somewhat turbulent, airflow within the work environment.

Safe Work Practices for BSC Use Do not store equipment or supplies inside the cabinet. Do not use the top of the cabinet for storage. The HEPA filter could be damaged and the airflow disrupted. Make sure the cabinet is level. If the cabinet base is uneven, airflow can be affected. Never disengage the alarm. It indicates improper airflow and reduced performance which may endanger the researcher or the experiment. Never completely close the window sash with the motor running as this condition may cause motor burnout. Cabinets should be placed away from doors, windows, vents or high traffic areas to reduce air turbulence.

Safe Work Practices for BSC Use For BSC without fixed exhaust, the cabinet exhaust should have a twelve inch clearance from the ceiling for proper exhaust air flow. Also, allow a twelve inch clearance on both sides of the cabinet for maintenance purposes. Never operate a cabinet while a warning light or alarm is on. The operator should be seated with shoulders level with the bottom of the sash. Perform all work using a limited number of slow movements, as quick movements disrupt the air barrier. Try to minimize entering and exiting your arms from the cabinet, but if you need to, do it directly, straight out and slowly. Keep all materials at least four inches inside the sash opening. To avoid excessive movements in and out of the cabinet, discard pipettes into a tray, container or biohazard bag within the cabinet.

Safe Work Practices for BSC Use If a bunsen burner must be used, place it at the rear of the work area where the air turbulence from the flame will have the least possible effect on the air stream. Often the use of a flame is redundant in what should be a germ free work space. All equipment which has come in contact with the biological agent should be decontaminated. The cabinet should be allowed to run for at least three minutes with no activity so that the airborne contaminants will be purged from the work area before removing equipment. After all items have been removed, wipe the interior surfaces with disinfectant.

Biohazard Spill Control Inside a BSC 1. Keep the BSC on. 2. Put on protective gloves. 3. Spray & wipe walls, work surfaces, and equipment with decontamination solution. 4. Flood tray top, drain pans, and catch basins with decontamination solution. 5. Allow to stand for 20 minutes. 6. Drain excess solution into cabinet base. 7. Lift out tray and any removable exhaust grille work.

Biohazard Spill Control Inside a BSC 8. Clean top and bottom surfaces with sponge/cloth soaked in decontamination solution. 9. Replace tray and grille work. 10. Place everything that is contaminated into autoclave pan. 11. Drain decontamination solution from cabinet base into AUTOCLAVABLE containers. 12. Autoclave. 13. If gaseous decontamination is needed, call EHS at 292-1284.

Biological Safety Cabinet Certification Your cabinet must be certified when first installed and then annually. It must be recertified anytime it is moved even within the same room. Before certification personnel arrive, remove all items from the cabinet and wipe it down with a disinfectant. This will expedite the certification. If you have any questions, or think there may be a problem with your cabinet, do not hesitate to contact EHS (292-1284). Any decontaminations, certifications, repairs or adjustments are to be made by qualified personnel.

And remember.. A laminar flow biological safety cabinet is a valuable supplement to good sterile technique. If the cabinet is not well understood and If the cabinet is not well understood and operated correctly, it will not provide adequate protection for you or the environment.

References Freshney, R.I. (2000) Culture of Animal Cells: A manual of Basic Technique. 5fifth edition. New Jersey: John- Wiley&Sons,Inc. Freshney RI. Culture of Animal Cells: A Manual of Basic Technique. New York: Wiley-Liss, 2005. Bernhard, O., & Bhatia, N. (2004) Tissue Engineering. Pearson Prentice Hall Bioengineering