TES-1339 INSTRUCTION MANUAL

Similar documents
1. INTRODUCTION FEATURES SPECIFICATIONS PANEL DESCRIPTION OPERATING INSTRUCTION MAINTENANCE...6

Safety Lighting. Safe also without electricity. Standards SIST EN 1838: 2011 ISO (2007) CIE S 020 (2007)

4-in-1 Humidity, Temperature Airflow and Light Meter

Class 1 laser beam sensor safe for your eyes

Optical Testing Report for 46 Lamp Sample Sets

Emergency Lighting User's Guide

Operation Guide 3408

Dual Laser InfraRed (IR) Thermometer

USER MANUAL MODEL UV MINDER 3D V2.0. Part Number: Revision Level: A

Class 1 laser beam sensor safe for your eyes

Philips MASTER LEDbulb DIM W & W E27/B K 230 V

Class 1 laser beam sensor safe for your eyes

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com

Dual Laser InfraRed (IR) Thermometer

Optical Measurement Techniques for LED Clusters and LED Safety

CONTENTS USING THE LIGHTS KEY...31 NETWORK ALARMS...32 FAULT AND ERROR MESSAGES...33

USER MANUAL ENVIRONMENTAL METER. Model EN510. Additional User Manual Translations available at

onlinecomponents.com

Pinless Moisture/Humidity Meter with Memory + IR Thermometer

Annexure-I. 1. Goniophotometer and Goniospectroradiometric measurement System

USER GUIDE Heat Index Psychrometer

Operation Guide 5510

Optics & Radiometry Laboratory

RESOLUTION A.752(18) adopted on 4 November 1993 GUIDELINES FOR THE EVALUATION, TESTING AND APPLICATION OF LOW-LOCATION LIGHTING ON PASSENGER SHIPS

Infrared Moisture Balance (FD-660)

Pinless Moisture/Humidity Meter with Memory + IR Thermometer

Philips MASTER LEDbulb 7-40 W and W E27/B K 230 V

COMPASS S3 S2 CALIBRATION

Lighting Guidelines. Planning Services Department Corporation of the Municipality of Clarington

Before using your air conditioner, please read this manual carefully and keep it for future reference. ACDC12B

Mini InfraRed (IR) Thermometer with Type K Input and Laser Pointer

Fire Command Keypad. XR5 User s Guide

Fast. Precise. Versatile. UL Lighting Sciences

Lighting Laboratory. Ensuring the performances and compliance of your lamps and luminaires

Challenges in SSL Design and Photometric testing Gautam Brahmbhatt, Head Technical (Verification Service)

User's Guide. Pinless Moisture/Humidity Meter with Memory + IR Thermometer. Model MO295

FS-200. Model. Instruction Manual. Infrared Food Thermometer. reedinstruments. www. com

High Lux Performance White LED MR16 Lamp

Confined Space Monitoring. Leak Investigation. Functions 4. Bar Hole Sampling. Line Purge Testing. Actual Size GX-2012

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

User s Manual. Electronic Personal Dosimeter (For Gamma(X)-ray) NRF30. Fuji Electric Co., Ltd. TN514629i 1/31

Technical Application Guide for UP-SHINE LED Ceiling light UP-AL W-SS.

Our photo detectors are offered for both power or energy measurements. Measure as low as a few femtojoules in energy or a few picowatts in power.

THL2. Temperature/Humidity USB Datalogger INSTRUCTION MANUAL

LIGHTING DESIGN STANDARDS SECTION City of Maryland Heights ZONING CODE

The two type sensors are innovative and active motion dimmable LED drivers or ballasts.

Medical LED. Superbly balanced and colour corrected LED-modules for medical use.

Air Quality Sensors. Particle Matter, CO 2, Humidity, Temperature and Dust Products. OEM Sensor Components for Air Quality and Environmental Control

BST-MG08/09 Multi-gas Detecting Alarm. Manual Instruction

MicroTool. Not just a simple Measuring Instrument A complete Tool to support your daily activities FLUE GAS ANALYSERS

Movement-based energy control

Table of Contents. Page Quick Set-up Instructions 1. Theory of Operation 2. Specifications 3. TM 1003W Controls and Connections 4

64 Max IR Thermometer

Counting Scales. HC-i Series.

POWER METER INSTRUCTION MANUAL

Surface Mount Ambient Light Sensor PTAS-P1F1708CH6. Features. Descriptions. Applications. Device Selection Guide

HP727S. Single speed swimming pool heat pump controller Operation manual TABLE OF CONTENTS

Thermohygrometer with Single Type K Input and Superheat Function

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Laser Collimated Beam Sensor

EVac II Programmable Digital Vacuum Gauge

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL THL1

IRT2 Dual Laser IR Thermometer - User Guide

USER GUIDE HARDWIRED CONTROL PANELS

Digiplex LED Keypads User s Manual

Indoor/Outdoor Temperature Alert

User Guide InfraRed (IR) Thermometer with Wireless Datalogger, Type K input & Laser Pointer MODEL 42560

MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017

Signature Series 418 Mhz COMMUNICATOR RECEIVER INSTALLATION AND OPERATIONS MANUAL

Specifications Wavelength Range: 190 nm to 800 nm continuous

R2310. Infrared Thermometer. Instruction Manual

MX1A and MX1B: Laser Displacement Detection Sensors. 85% to 110% of the rated voltage MX1A: 15VA, MX1B: 25VA

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

RAM GENE / RAM GENE-1

FR-A180 FR-A360. Occupancy detector. Use

Overview of International Lighting Standards Steve Coyne Director, Light Naturally UNEP Consultant

Manual Vane Anemometer PCE-VA 20

Introduction. OM-THA2

Infrared Moisture Balance (FD-800)

Technical Application Guide for PHILIPS LED Lamps

Technical application guide

How Photometric Design Enhances Energy Savings and Asset Management

USER MANUAL USB Multi-Function Datalogger Model RHT35

CA870 INFRARED THERMOMETER E N G L I S H. User Manual. Auto OFF

Digital Sensors Erythema Sensor PMA2101

LCD and 16-Zone LED Keypads. User s Manual

Humidity Meters. hygrometers & therma-hygrometers. what is a hygrometer? applications. monitor humidity levels for better health. Humidity.

Infrared Thermometer with 12:1 Distance-to-Sight Ratio and Temperature Alarm

EL-USB-CO300 Carbon Monoxide (CO) Data Logger with USB Interface

NATURAL GAS DETECTOR. Confined Space Monitor. Leak Investigation. Line Purge Testing. Bar Hole Sampling

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

Electronic Pellet Burner Controller NPBC-V3M

High Efficiency LED Panel Lights

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Humidity Monitor model 01080

Rugged LED shelter lighting

7XG3130 ReyArc30 Optical Arc Fault Sensor Energy Management

USER MANUAL Digital CO Monitor Model CO30

PRODUCT INFORMATION. Uses tough, high quality PC diffuser, resistant to fracture and to ultraviolet light to limit discolouring.

Transcription:

TES-1339 INSTRUCTION MANUAL COMP AVG D- H REF CONTENTS Title Page 1. INSTRUCTION...1 2. FEATURES...1 3. SPECIFICATIONS...2 4. NAME OF PARTS AND POSITIONS...3 5. OPERATING INSTRUCTIONS......5 6. BATTERY CHECK-UP & REPLACEMENT...13 7. SPECTRAL SENSITIVITY CHARACTERISTIC...13 8. MAINTENANCE...14 9. RECOMMENDED ILLUMINATION...14 1. INSTRUCTION The digital illuminance meter is a precision instrument used to measure illuminance (lux, footcandle)in the field. It meets CIE photopic spectral response, f 1 6%. It is fully cosine corrected for the angular incidence of light. The illuminance meter is compact, tough and easy to handle owing to its construction. The light sensitive component used in the meter is a very stable, long-life silicon phote diode and spectral response filter. U.S. Pat. No. Des. 446,135 U.S. Pat. No. Des. 469,025 2. FEATURES Dual Display, 4 digit LCD reading. Spectral Sensitivity close to CIE photopic Curve.

1 Measuring Levels Ranging 0.01 to 999900 Lux, 0.001 to 99990 fc, Autoranging 5 step. Accurate and Instant response. Integral illuminance measurement. Luminous intensity measurement. Data Hold function. Data memory and read function. Reference value store for relative or percentage deviation measurement. Ripple measurement for STRAY + LIGHT function. Time-hold function. Point - average function. Comparator function. Auto power off function. 3. SPECIFICATIONS Display:Dual display, 4 digit LCD reading. Measuring Range: 99.99 lux, 999.9 lux, 9999 lux, 99990 lux, (Autoranging 5 step) 999900 lux / 9.999 fc, 99.99 fc, 999.9 fc, 9999 fc, 99990 fc. Note : 1fc =10.76 lux. Overrange Display:LCD will show OL symbol. Spectral Response : CIE Photopic. (CIE human eye response curve). Spectral Accuracy : CIE Vλ function f 1 6% Cosine Response : f 2 2% Accuracy:±3 % rdg ±5digits (calibrated to standard incandescent lamp at color temperature 2856 K). Temperature Characteristics:±0.1%/. Sampling Rate : 5 times/sec. Photo Detector:One silicon photo diode and spectral response filter. 2 Manual Data Memory Capacity:50 sets. Operating Temperature & Humidity: 0 to 40 (32 to 104 )& 0% to 80% RH. Storage Temperature and Humidity: -10 to 50 (14 to 140 )& 0% to 70% RH. Power Source:6 pcs size AAA battery. Battery life ( typical ) :200 hours ( carbon zinc ). Photo detector Lead Length:150 cm ( approx. ). Photo detector Dimensions:92L 60W 29H (mm); Meter Dimensions:150L 72W 35H (mm); Weight:320g. Accessories:Carry case, instruction manual, battery. 4. NAME OF PARTS AND POSITIONS

AVG SEt05 : Comparator High / Low value setting mode. 1 3 2 5 6 8 11 12 COMP D- H REF 4 7 9 10 13 1. LCD Display:4 digit displays with a maximum reading of 999900, and the indicating signs of measured values, unit function symbols, and decimal points etc are display. 3 2. Power Control key: The power switch key turns the illuminance meter ON or OFF. 3. Time key:switch between minute : second (m:s) and Day- Hour (D-H) time display. 4. H Data-Hold key:freeze or unfreeze the displayed readings. 5. AVG key : Active the point-average mode. 6. SET key : Start setting mode. SEt01 : Ripple measurement for STARY + LIGHT mode. SEt02 : Time hold mode. SEt03 : Real time setting mode. SEt04 : Integral illuminance measurement mode. 14 7. Lux/fc key : Illuminance scale selects. 1 footcandle=10.76 lux. 8. / % : Press / % key to enter the Relative mode, zero the display, and store the displayed reading as a reference value. The annunciator is displayed, press / % key again to enter the Relative percentage compute. The annunciator REF % is displayed. Press key exit the Relative mode. 9. key : Enter / Exit a setting mode or store the displayed setting. Exit the point average mode; Relative mode and Luminous intensity mode. 10. cd / COMP key : Enter the Luminous intensity mode or comparator mode. 11. MEM key : Press one time store the one set LCD logging to memory. 12. key : Scroll to a setting mode or increase / decrease the displayed setting. 13. READ key : Enter the memory data reading mode. 14. Photo Detector. 4 5. OPERATING INSTRUCTIONS 5-1 Illuminance Measurement 3. Remove the photo detector cap and face detector to the light source in a horizontal position. 4. Read the illuminance value from the LCD display. 5. Press the H key, if the displayed value needs to be hold. Press H key again to exit the data held mode.

D-H D-H 5-2 Ripple Measurement Use ripple measurement function in the day-time, user can measure light source actual illuminance at night, so user doesn t need to work at night. 3. Turn on the desired measure light source. 4. Remove the photo detector cap and face to light source in a horizontal position. 5 5. Press SET key, the annunciator SEt01 is displayed. 6. Press key, the annunciator STRAY + LIGHT is displayed, the illuminance value from the LCD display is other STRAY light source plus desired measure LIGHT source illuminance value. 7. Press key to store the STRAY + LIGHT illuminance value, the annunciator STRAY is displayed. 8. Turn off the light source. 9. Press key again to store the STRAY illuminance value and display compute actual LIGHT source illuminance value at night, the annunciator LIGHT and D-H is displayed. 10. Press key to exit this mode. 5-3 Time Hold Measurement Time Hold measurement is free from a man s shadow effect the illuminance measurement value. 3. Remove the photo detector cap and face the detector to the light source in a horizontal position 4. Press SET key the annunciator SEt01 is displayed. 5. Press key to change the setting mode to SEt02. 6. Press key to enter to the second setting of the stopwatch. 7. Press keys to set to the desired seconds. 6 8. Press key to enter to minute setting of the stop watch. 9. Press key to set the desired minute. 10. Press key to start count down, you must leave the light source area, free from shadow effect measurement value. 11. When stop watch counts to zero, the meter will auto hold the last measured value, the annunciator TIME - HOLD is displayed. 12. Press key to exit this mode.

5-4 Real Time Setting 2. Press SET key, the annunciator SEt01 is displayed. 3. Press key for two times to change the setting mode to SEt03. 4. Press key to enter to the real seconds setting. 5. Press keys to set the real seconds. 6. Repeat procedure 4 and 5 setting the real minute;hour and day setting. 7. Press key finish the real time setting. 8. Press TIME key step through the day-hour (D-H) and minutesecond (m-s) displayed. 7 5-5 Integral Illuminance Measurement Integral illuminance means luminous energy measurement that is a measure of the rate of flow of flux, and so is expressed in lumenseconds. It is also possible to measure any photometric quantity on a time dependent basis. For instance, the illuminance could be integrated over time to yield lux-hour or footcandle hour. 2. Press lux/fc key selection switch to the desired lux or fc unit. 3. Remove the photo detector cap and face the detector to the light source in a horizontal position. 4. Press SET key, the annunciator SEt01 is displayed. 5. Press key for three times to change the setting mode to SEt04. 6. Press key to start the integral illuminance measurement and the meter will set to the fixed measurement range. The annunciator lx-h or fc-h will be displayed and the timer will be started. 7. If measured illuminance value is over the fixed measurement range, the measurement will be stopped and LCD display will be frozen, the annunciator + will be displayed. 8. The maximum indication of integral illuminance is 999999. When this maximum indication value is exceeded, the number of integration can be checked by annunciator CYCLE NO.XX. The maximum number of CYCLE is up to 99. 9. Press key to exit this mode. 8

COMP 5-6 Comparator Setting and Operation 3. Remove the photo detector cap and face the detector to light source in a horizontal position. 4. Press SET key, the annunciator SEt01 is displayed. 5. Press key for four times to change the setting mode to SEt05. 6. Press key to enter to the comparator setting mode. 7. Press key to select the desired lx or fc unit. 8. Press key to enter to setting the measurement range. 9. Press key to select the desired measurement range. 10. Press key to enter to setting the comparator High limit value the annunciator + is displayed. 11. Press keys to adjust the desired High limit value. 12. Press key to enter to setting the comparator Low limit value. 13. Press keys to adjust the desired Low limit value, the annunciator - will be displayed. 14. Press key, the comparator setting mode is finished15. Press cd/comp key for three seconds enter to the comparator function. The annunciator COMP is displayed. 16. If measurement value exceeds the setting value, the annunciator + or - will be displayed and the beeper will sound 17. Press key to exit the comparator mode. 9 5-7 Point Average Measurement Illuminance is a measurement of the amount of visible light incident upon a prescribed surface area. Due to detector doesn t have such large areas, so the area of the detector is multiplied proportionally. You can use point average function to do it. 3. Remove the photo detector cap and face the detector to the light source in a horizontal position. 4. Press AVG key to enter to point illuminance value average function, the maximum measurement is 99 points, the annunciator AVG is displayed. 5. Press MEM key each time, stores one measured point value to memory. The annunciator AVG NO.XX is displayed. 6. Press READ key perform the point average compute, the second annunciator AVG is displayed. 7. Press MEM key to continuously add the illuminance measured value for average, but the LCD will display he average value only, the point measured value will not shows. Press READ key again will change display shown the measured value. 8. Press key to exit the mode. 10

AVG 5-8 Luminous Intensity Measurement Luminous intensity is a light source property. Defined as the quantity of luminous flux emitted uniformly into a solid angle, the basic unit of luminous intensity is the candela, equal to one lumen per steradian. To calculate luminous intensity, the detector s area and its distance from the light source must be known. 3. Remove the photo detector cap and face to light source in a horizontal position. 4. Press cd/comp key to enter to setting of the distance, the annunciator m or ft is displayed. 5. Press key to set the distance between the light center of lamp and measurement base level, the preset maximum distance is 0.01 ~ 9.99m or 0.01 ~ 9.99ft. Measurement should be made in a dark room or equivalent thereof, where the brightness other than light source gives no influence to the meter. The unit shall be set at a position distant from the light source by more than 10 times of the size of such a light source. 6. Press key to perform luminous intensity measurement, the annunciator cd is displayed. 7. Press key to exit the mode. 5-9 Manual Data Memory and Read Mode : 1. Clear the manual memorized data Press key to turn off the meter. Press and hold down "MEM" key, then press key to turn on the meter. When LCD shows "CLr" which means the memorized data is erased. 2. Manual data memory Press "MEM" key each time, one set of reading to will be stored to the memory. At this moment, LCD will show the "M" mark and the memory address number. Total memory size is 50 sets. 3. Manual memory data READ Press "READ" key to enter the READ mode, the LCD will show "R" mark and the memory address number. Press " " or " " key to select the desired memory address number data for display. Press " " enter key to exit this mode. 11 12

5-10 How to Disable Auto Power off Function The meter enters sleep mode if no key pressed occurs for 30 minutes. 1. Press " " key to turn off the meter. 2. Press and hold down "HOLD " key then press " " key to turn on the meter, the auto power off function will be disabled. The time display auto power off mark " " will disappeared. Auto power off mode is enabled each time you turn on the meter and is automatically disabled in the integral illuminance measurement mode. 6. BATTERY CHECK-UP & REPLACEMENT 1. As the battery power is not sufficient, LCD will display " + "; and, replacement of one new battery type 6 1.5V is required. 2. After turning off the meter, press the battery cover and push in the direction of the arrow to open. 3. Disconnect the battery from the instrument and replace it with a standard 6 1.5V battery and go for the cover. 7. SPECTRAL SENSITIVITY CHARACTERISTIC To the detector, the applied photo diode with filters makes the spectral sensitivity characteristic almost meet C.I.E. ( INTERNATIONAL COMMISSION ON ILLUMINATION ) photopic curve V(λ ) as the following chart described. Relative Sensitivity (%) 1.2 1 0.8 0.6 0.4 0.2 0 8. MAINTENANCE 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660 680 700 720 740 760 780 Wavelength in Nonometers 13 1.2 1 0.8 0.6 0.4 0.2 0 Light Meter 1. The white plastic disc on the top of the detector should be cleaned with a damp cloth when necessary. 2. Do not store the instrument where temperature or humidity is excessively high. 3. The reference level, as marker on the faceplate, is the tip of the photo detector globe. 4. The calibration interval for the photo detector will vary according to operational conditions, but generally the sensitivity decreases in direct proportion to the product of luminous intensity by the operational time. In order to maintain the basic accuracy of the instrument, periodic calibration is recommended. 9. RECOMMENDED ILLUMINATION CIE 1fc = 10.76 lux LOCATIONS lux fc OFFICE Conference, Reception room. 200 ~ 750 18 ~ 70 Clerical work 700 ~ 1,500 65 ~ 140

Typing drafting 1000 ~ 2,000 93 ~ 186 FACTORY Visual work at production line 300 ~ 750 28 ~ 70 Inspection work 750 ~ 1,500 70 ~ 140 Electronic parts assembly line 1500 ~ 3,000 140 ~ 279 Packing work, Entrance passage 150 ~ 300 14 ~ 28 Laboratory, Library, Drafting, room 500 ~ 1,500 47 ~ 140 14 HOTEL Public room, Cloakroom 100 ~ 200 9 ~ 18 Reception 200 ~ 500 18 ~ 47 Cashier 750 ~ 1000 70 ~ 93 STORE Indoors Stairs Corridor 150 ~ 200 14 ~ 18 Show window, Packing table 750 ~ 1,500 70~140 Forefront of show window 1500 ~ 3,000 140 ~279 HOSPITAL Sickroom, Warehouse 100 ~ 200 9 ~ 18 Medical Examination room 300 ~ 750 28 ~ 70 Operating room Emergency Treatmet 750 ~ 1,500 70 ~ 140 SCHOOL Auditorium, Indoor Gymnasium 100 ~ 300 9 ~ 28 Class room 200 ~ 750 18 ~ 70