Formalin Vaporizer The Premium Solution for Biological Safety Cabinet Decontamination. Esco Formalin Vaporizer Model FV-00_

Similar documents
Forced Convection. Laboratory Ovens. Isotherm Forced Convection Laboratory Ovens, 110L, 54L and 32L models.

CelCulture. Cradle for Beautiful Cells. CelCulture Water-Jacketed CO 2 Incubator

Polymerase Chain Reaction Cabinets The Proven Solution for Contaminant-Free PCR

Forced Convection Laboratory Ovens Reliable Performance For Universal Applications

Universal Applications

Forced Convection Laboratory Incubators. Reliable Performance For Universal Applications. Forced Convection Laboratory Incubators

Class III, Biological Safety Cabinets The Premier Solution for High Containment Laboratories

Class III, Biological Safety Cabinets The Premier Solution for High Containment Laboratories

Laminar Flow Clean Benches, Horizontal and Vertical The Leading Solution for Research Laboratories

SubliMate. Bench Top Laboratory Freeze Dryers. Designed and Made in the E.U.

Refrigerated Incubator. Reliable Performance For Universal Applications

Bedding Disposal Workstations The Portable Safety Solution for Animal Research Laboratories. Viva Bedding Disposal Workstation, Model VBD-4A_.

Air Showers The Safety Solution for Animal Research Laboratories. Viva Air Shower, Model - EAS-1B.

Forced Convection Laboratory Incubators. Reliable Performance For Universal Applications. Forced Convection Laboratory Incubators

Class II, Biological Safety Cabinets The Industry s Most Energy Efficient Class II Cabinet

CO2 Incubators Cradle for Beautiful Cells. CelCulture CO2 Incubator

SubliMate. Bench Top Laboratory Freeze Dryers. SubliMate Bench Top Laboratory Freeze Dryer Model FDL-5

Class II Microbiological Safety Cabinet The Industry's Premier Biological Safety Solution from Esco

Ultra-low Temperature Freezer ULT Vault for Your Samples

Ultra-low Temperature Freezer ULT Vault for Your Samples

Low Noise, Class II Microbiological Safety Cabinets The Industry's Most Comfortable, Energy Efficient Cabinet

Polymerase Chain Reaction Cabinets The Proven Solution for Contaminant-Free PCR. Esco PCR Cabinet, Model PCR-4A_. Shown with optional support stand.

Forced Convection Laboratory Incubators. Reliable Performance For Universal Applications. Forced Convection Laboratory Incubators

Class II Biological Safety Cabinets The Safety Solution for Life Science Laboratories

Class II, Microbiological Safety Cabinets The Industry's Premier Energy Efficient Solution from Esco

Class II, Microbiological Safety Cabinets The Industry's Premier Energy Efficient Solution from Esco

Biological Safety Cabinets or Fume Hoods Which one to use?

Ultra-low Temperature Freezer ULT Vault for Your Samples

Low Noise, Class II Biological Safety Cabinets The Safety Solution for Life Science Laboratories

Fume Hood Safety TYPES OF FUME HOODS

Ultra-low Temperature Freezer ULT Vault for Your Samples

Safe Operating Procedure (Revised 2/10)

Class II Biological Safety Cabinets The World's Leading Energy Efficient, Quiet, and Compact Biosafety Cabinet.

CO 2 Incubators Cradle for Beautiful Cells. CelCulture CO 2 Incubator Model CCL-170_-_

Cradle for Beautiful Cells. CelCulture CO 2 Incubators. Designed in the USA

Ultra-low Temperature Freezer ULT Vault for Your Samples. Lexicon Ultra-low Temperature Freezer Model UUS-668A-1

Research Document Type: PROCEDURE. Effective Date: January 01,2015 Author Steven Hayes

Cradle for Beautiful Cells. CelCulture CO 2 Incubators

ENVAIR lab CO2 Incubator

CelCulture CO2 Incubators CelCulture for IVF. Cradle for Beautiful Cells. Cradle for Optimal Embryo Incubation. Designed in the USA

Animal Tissue Culture SQG 3242 Equipment and Lab design. Dr. SITI PAULIENA MOHD BOHARI

Lexicon II. with New Aalto Controller. Lexicon II Ultra-low Temperature Freezers 1. Uncompromised Protection for your Samples. Designed in the USA

ENVAIR lab CO2 Tri-Gas Incubator

Class II, Australian Biological Safety Cabinets The Safety Solution for Life Science Laboratories

Appendix C IBC Safe Use of Biosafety Cabinets SOP

EMERGENCY RESPONSE GUIDE IMMEDIATE EMERGENCY PROCEDURES

Environmental Health and Safety Biological Safety Program Guidance Document

CelMate. CO 2 Incubators. CelMate. Incubators. CelMate CO 2 Incubators. Cradle for Beautiful Cells

Lexicon II Ultra-Low Temperature Freezer 1. Ultra-Low Temperature Freezer. Lexicon II. Uncompromised Protection for your Samples. Designed in the USA

PART III. Laboratory equipment

SCIF SOP for BSL 2 samples on Aria3

Animal Containment Workstation The Portable Safety Solution for Animal Research Laboratories

Biological Safety Cabinet

EMERGENCY RESPONSE PLAN FOR

Safe Use of a Class II Biological Safety Cabinet

CelSafe ESCO GLOBAL NETWORK. The Safest Way To Grow Your Beautiful Cells

Florida International University. Biosafety Cabinet Program

Safety In Chemistry. Identified Risks and Hazards

CelCulture CO2 Incubators CelCulture for IVF. Cradle for Beautiful Cells. Cradle for Optimal Embryo Incubation. Designed in the USA

Class II, Type B2 (Total Exhaust) Biological Safety Cabinets The Safety Solution for Life Science Laboratories

TECHNICAL BULLETIN ACCESSORY INFORMATION NU-984 FORMALIN/AMMONIA VAPORIZER

Biological Safety Cabinets. Biological Safety Cabinets

Application: Response procedures for major and minor spills, including: organic solvents acids bases dry chemicals mercury hydrofluoric acid

SECTION LABORATORY EQUIPMENT CONSTRUCTION STANDARD PART 1: GENERAL Scope of Standard

Appendix A.4.1. Safe Fume Hood User Guide

Automatic Cyanoacrylate Fuming Chamber 24 33T 60T 72XL. Safely Detect Quality Latent Prints and Reproducible Results

Ultra-low Temperature Freezer ULT Vault for Your Samples. Model: UUS-714A. Designed in the USA

Microbiological Safety Cabinets ENVAIR eco safe Basic Plus

EC 160 CO 2. Incubator

BIOLOGICAL SAFETY CABINET

Biological Safety Cabinets. Presented by Mr. Rene Soetens President

Precision Containment for All Applications

Manual v1 Page 1 of 19. Manual. Laboratory autoclave

L a b o r a t o r y E q u i p m e n t C a t a l o g

LSU Inter-Institutional Biological and Recombinant DNA Safety Committee (IBRDSC)

Microbiological Safety Cabinets ENVAIR eco safe Comfort

CLEAN AIR SYSTEM.

Quick Reference Guide to Cabinet types

Microbiological Safety Cabinets ENVAIR eco safe Comfort Plus

Adv. 8.4 Guidelines for Biosafety Cabinets Rev

VP, Facilities and Construction Management

Control of Biological Hazards

Special Precautions Isolation Plan Environmental Services

CelCulture for IVF. Cradle for Optimal Embryo Incubation

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

User Manual. CelCulture

ES-20 Orbital Shaker-Incubator

Esco PCR Cabinet, Model PCR-4A_. Shown with optional support stand. Polymerase Chain Reaction Cabinets The Proven Solution for Contaminant-Free PCR

Biological Safety Cabinet Failure Response Procedure

STERIDENT, STERIMAT, STERIMAT PLUS

Copyright Information

BSC CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET SERIES. Optimum protection of personnel, product and environment. BSC Class

BARRIER TECHNOLOGY RABS & ISOLATORS FOR ASEPTIC PROCESSING

Microbiological Safety Cabinets

DEPARTMENTAL SAFETY ISSUES - SAFETY TECHNOLOGIST / INFECTION CONTROL PRACTITIONER

Manual Page 1 of 8. Manual. Laboratory autoclave

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106

A measured step forward

Thermo Scientific 1300 Series A2

Wet Bench Standard Operating Procedures

Transcription:

Formalin Vaporizer The Premium Solution for Biological Safety Cabinet Decontamination Esco Formalin Vaporizer Model FV-00_

2 Esco Experience As a global leader in premium quality biological safety cabinets, Esco is committed to providing it s customers with a full range of innovative ancillary products to compliment its extensive range of cabinets. The Esco formalin vaporizer is a microprocessor-controlled unit that simplifies and automates the decontamination process for all sizes and makes of biological safety cabinets. This also increases safety for the decontamination operator and laboratory personnel since no manual intervention is required during the entire process. Dependable Construction Durable, heavy-duty full stainless steel construction. Electrical and electronic components are isolated from the heating chamber to prevent exposure to chemical fumes and heat which can reduce component life. Thermostats for the formalin and ammonia chambers prevent overheating. Backup thermostat ensures the temperature of the external casing does not rise above a safe level. Internal heat insulation prevents cross heating of chambers and damage to electrical components. Manual shut-off isolating switch and safety circuit breaker. Innovative Design Full-sized grip handles on both sides for easy handling. Large formalin and ammonia chambers (800 ml maximum capacity each) allows even the largest biological safety cabinets (6 foot console / floor-mounted units) to be decontaminated with the maximum agent volume recommended in EN 12469. Increased diameter liquid chamber inlets to allow the formalin and ammonia solutions to be poured into the chambers more easily without spillage. Decontamination Cycle Operator starts process by pressing a button on the touch pad. Pre-delay timer allows decontamination operator to seal the cabinet before vaporization of formalin commences. Delay period can be set from 1 to 59 minutes. Vaporization of formalin solution commences to release the formaldehyde gas to decontaminate the cabinet. LCD display will show reminder for operator to cycle the cabinet fans to ensure even distribution of the gas. Heater automatically cuts off when formalin solution has completely boiled off. Contact time timer starts to regulate the period for which the formaldehyde gas is exposed to the cabinets internal surfaces for the decontamination to take effect. Neutralization process begins automatically, to vaporize the ammonia solution to neutralize the formaldehyde gas. LCD display will show reminder for operator to cycle the cabinet fans to ensure even distribution of the gas. Neutralization timer starts to regulate the period for which the neutralization reaction takes place. Control System Features Specifically designed for safety cabinet decontamination with automatic control and no manual intervention required through the entire decontamination cycle. In case of power failure the vaporizer will recover automatically once the power resumes; Microprocessor unit will self-reset in case of any processor failure. Pre-delay, contact time, neutralization timers, are adjustable using the easy clean touch control panel with large backlit LCD display on the front of the unit. Audible and visual alarms warn of overheating during a cycle. Password control to prevent decontamination timers from being adjusted by unauthorized personnel. Typical operation of the formalin vaporizer within a biological safety cabinet. Optional Accessories Formalin Vaporizer

Common List of Equipments Used Device to measure the formaldehyde concentration. Device to measure the ammonia concentration. Tubes with a measuring range of 0-100 ppm are recommended. Spore strips (optional). Appropriate Protective Clothing A solid front, back closing, full body, long-sleeved disposable lab gown. Disposable shoe covers. Disposable surgical / PVC gloves for hand protection. Gloves should be pulled over the knitted wrists of the gown rather than worn inside. Double gloving should be considered. Gloves should be worn while handling the formalin and ammonia solutions. Nitrile gloves, which exhibit a high resistance to formaldehyde and are less prone to cause allergic skin reactions as compared to latex, should be used. Full-faced mask with a disposable dust and mist respirator. The face shield should be wiped clean with a suitable tissue and water after each use. A solid front, back closing, full body, long-sleeved disposable lab gown. A solid front, back closing, full body, long-sleeved disposable lab gown. Apparatus Beaker Measuring cylinder Tape (3M-3939 is recommended) Air-tight bag that can enclose the entire cabinet (can be ordered from Esco) Measuring tape Biohazard waste bag Hand-drill, screwdriver, and socket key (M4 and M5) General Specifications, Formalin Vaporizer Note to customer: Insert electrical voltage number into last model number digit _ when ordering. Model FV-00_ Dimensions (W x D x H) 350 x 293 x 310 mm 13.8 x 11.5 x 12.2 Vaporizer Construction Stainless steel grade 304 Maximum Capacity Max Cabinet Volume for Decontainmination Electrical Formalin Ammonia For Room with RH <60% For Room with RH >60% 220-240V, AC, 50Hz, 1ø Power Consumption Current 110-130V, AC, 60Hz, 1ø Power Consumption Current 800 ml (27 oz) 4.7 m 3 (166 cu.ft.) 9.4 m 3 (332 cu.ft.) FV-001 300 W 1.25 A FV-002 300 W 2.5 A It may be appropriate to conduct a safety briefing for all personnel in the laboratory where the cabinet is sited prior to carrying out the procedure. Suitable materials may be obtained from Esco. Access to the laboratory should be restricted during the procedure. A warning sign should be posted on the door of the lab to warn all personnel (a suitable label is available from Esco). No personnel should enter the lab until the procedure is complete and the formaldehyde concentration has been verified to be within safe limits. Important Notes Prior to starting the procedure, contingency plans should be made in case there is a breach in the cabinet seal and formaldehyde escapes into the surrounding environment. The room ventilation design should be understood and the appropriate equipment should be available to vent the formaldehyde (for example by means of connecting a flexible hose via a small exhaust fan to vent the vapor into an adjacent fume hood). Special care should be taken in case air from the laboratory is re-circulated to other parts of the building in which case a formaldehyde breach could necessitate the evacuation of other areas in the building. It may be appropriate to conduct a safety briefing for all personnel in the laboratory where the cabinet is sited prior to carrying out the procedure. Suitable materials may be obtained from Esco.

Decontamination Process Calculate the amount of 37% formalin, 25% ammonia, and water required, which is 85 ml each, for every 1 cubic meter volume to be decontaminated. Water is poured into the formalin tank to increase the decontamination area humidity to be above 60%. If the room humidity already exceeds 60%, and the formalin tank usage needs to be optimized to decontaminate a large space, such as a small room, then there is no need to add water. Please note that the water shall not be poured into the ammonia tank because it will not assist in the decontamination process. To anticipate a worst case scenario, a relatively large size biological safety cabinet, with the body height of 1.7 m (67 inches) and width of 0.9 m (35 inches), mounted on a tall 90 cm (35 inches) standing height support stand, will have the volume enclosed by the air tight bag, and the corresponding required solution mixture, approximately as follows: Cabinet Width 2ft / 0.6m 3ft / 0.9m 4ft / 1.2m 5ft / 1.5m 6ft / 1.8m Volume (m 3 ) 1 1.5 2 2.5 3 37% Formalin (ml) 85 128 170 213 255 25% Ammonia (ml) 85 128 170 213 255 Water (ml) - optional 85 128 170 213 255 Smaller size cabinets, especially when installed on a 70 cm (28 inches) sitting height support stand, require less formalin, ammonia, and water from the table above. The actual volume to be decontaminated must be measured on site, as it varies from one cabinet to another. The figures on the table above are given for comparative purpose only. Decontamination Steps 1 8 Wear the appropriate protective clothing Thoroughly surface-decontaminate the work surface, side walls, inner back walls, drain pan, and interior surface of the window using 70% ethanol or other disinfectant depending on the user s requirement. Do not use any disinfectant containing chlorine-based substance. You should also surface decontaminate the UV light and electrical outlet when present. Set the needed PREPROC TIME, CONTACT TIME, and NEUTRAL TIME period. Recommended CONTACT TIME and NEUTRAL TIME is 15 hours and 8 hour respectively. 2 Pour formalin into a measuring cylinder to get the correct volume based on the calculation; then pour it into the formalin tank in the vaporizer. 7 Connect the formalin vaporizer to the electrical outlet inside/outside the cabinet work zone. 3 Add an equal volume of water into the formalin tank. OPEN BOTH FORMALIN AND AMMONIA TANK CAPS. If the caps are closed, the vapour can not go anywhere and the overpressure may cause EXPLOSION. When the power supply to the vaporizer is interrupted midway through the process, the vaporizer will automatically resume the decontamination cycle when the power supply is restored again. However, when the vaporizer is connected to the power outlet inside the cabinet work zone, the user should take note that this power source in the cabinet may/may not be restored when the power resumes. Therefore, it is recommended to connect the vaporizer to an external power source, for example a normal wall outlet to which power will be automatically be restored. This guarantees the vaporizer automatically resume the decontamination cycle when the power supply is restored again. 6 4 5 Pour the calculated amount of ammonia into ammonia tank. Put the vaporizer on the table tray of the cabinet. Surface decontaminate the cap of the aerosol sampling port before removing the cap. Remove the cap to ensure formaldehyde penetrates the entire inner surface. Do not open the aerosol sampling cap while the blower is on with the cabinet unsealed. 24 9 Prepare the air-tight plastic bag and tapes that is going to be used to enclose the cabinet before starting the decontamination process. 10 Start the decontamination process by pressing SET from the appropriate menu option. 11 Completely enclose the cabinet using the air-tight plastic bag and ensure air-tightness (as formaldehyde is typically non-penetrating, cabinet does not have to be pressure tight) inside the enclosure by utilizing tape in all appropriate places. Pay special attention to the power supply cables. Remember that the amount of time you have for this procedure is the period of PREPROC TIME that you have specified. 12 Decontamination process start.

After approximately 25, 50, 75, and 100% of the formalin being evaporated, turn on the cabinet blower for 1 minute to circulate the formalin throughout the cabinet. Below is the approximate time interval to turn on the blower, corresponding to the formalin and ammonia table for the worse case scenario as previously discussed: Cabinet Width 2ft / 0.6m 3ft / 0.9m 4ft / 1.2m 5ft / 1.5m 6ft / 1.8m Volume (m3) 1 1.5 2 2.5 3 Blower ON interval 5 min 7.5 min 10 min 12.5 min 15 min Allow the formalin vapor a minimum contact time of 8 hours inside the cabinet, preferably 10 hours, but overnight is better (CONTACT TIME). The ammonia vaporizer will automatically start after the CONTACT TIME is finished. If possible, it s desirable to circulate the ammonia by turning on the cabinet blower at the same interval as the formalin vaporization. However, this is not required, and not as important as circulating the formalin. Allow the ammonia vapor a minimum neutralization time of 2 hours inside the cabinet (NEUTRAL TIME) After the NEUTRAL TIME is finished the time will inform the user to open the plastic bag. Please be aware that the formalin and / or ammonia concentration inside the plastic bag may be above the Short Term Exposure Level (STEL) and full face mask with formalin filter is required when opening the plastic bag. By pressing MENU button after the decontamination process is started, the decontamination / neutralization cycle is interrupted, the following menu options are then displayed. RESUME STEP: This option allows the user to continue with the decontamination process SKIP STEP: This option allows the user to move to the next stage in the decontamination process CANCEL PROCESS: In case of emergency, the user can cancel the decontamination process by choosing this option. 15 Completion Steps 1 6 Check the ammonia concentration in room air. Remove disposable gown and gloves carefully and wash hands using germicidal soap thoroughly. 2 Check the formaldehyde concentration in room air at the end of the decontamination cycle, prior to resuming work on the cabinet. During this procedure, until the area is determined to be safe, a full face gas mask equipped with a formaldehyde canister, should be worn. The filter for the mask must be selected for the appropriate level of formaldehyde exposure. It must also be replaced regularly following the manufacturer s instructions. Validation Of Decontamination Procedure (Optional) Place some Bacillus subtilis var.niger spore strips inside the cabinet prior to decontamination. * work surface/table tray * drain pan 1 * downstream HEPA filter Place similar strips outside the decontamination room for positive control (optional). 4 3 5 Remove the plastic bag that enclosed the cabinet. Remove all the apparatus from inside the cabinet. Check the formaldehyde and ammonia concentration inside the cabinet. Spore strips on work surface 1. Before the filters are removed from the cabinet, their contaminated faces should be taped off using plastic film or cardboard or some other suitable material. This will minimize the risk of personnel in the event the decontamination was not thorough. 2. Used filters should be disposed following local regulations. They may have to be incinerated as medical waste. They should be double bagged and appropriately labeled following removal from the cabinet Remove all the spore strips and place in Trypticase-soy broth and incubate for 2 days at 37 o C. If there is no growth of bacteria inside the broth (the broth remains clear), the decontamination procedure has been successfully verified. Standards Compliance The unit enables decontamination to be accomplished by the standard protocol as recommended in EN 12469 (European standard for microbiological safety cabinets) with formalin and ammonia solutions. ANSI / NSF49 (American standard for Class II cabinets) recommends decontamination using paraformaldehyde and ammonium bicarbonate which is a similar protocol but with agents in the solid (instead of liquid) state, and as such this vaporizer can also be said to be in compliance with the ANSI / NSF49 recommended decontamination protocol.

PT Esco Bintan Indonesia Esco Containment, Clean Air and Laboratory Equipment Products Biological Safety Cabinets, Class II, III Fume Hoods, Conventional, High Performance, Ductless Carbon Filtered Laminar Flow Cabinets, Horizontal, Vertical, PCR Animal Containment Workstations Hospital Pharmacy Isolators, Cytotoxic Safety Cabinets Specialty Workstations: In-Vitro Fertilization, Powder Weighing PCR Thermal Cyclers, Conventional, Real-Time Cleanroom Fan Filter Units, Modular Rooms, Air Showers, Pass Thrus Since 1978, Esco has emerged as a leader in the development of controlled environment, laboratory and cleanroom equipment solutions. Products sold in more than 100 countries include biological safety cabinets, fume hoods, ductless fume hoods, laminar flow clean benches, animal containment workstations, cytotoxic cabinets, hospital pharmacy isolators, and PCR cabinets and instrumentation. With the most extensive product line in the industry, Esco has passed more tests, in more languages, for more certifications, throughout more countries than any biosafety cabinet manufacturer in the world. Esco remains dedicated to delivering innovative solutions for the clinical, life science, research and industrial laboratory community. www.escoglobal.com. NSF / ANSI 49 Biological Safety Cabinets Animal Containment Workstations Fume Hoods Clean Benches Esco Technologies, Inc. 2940 Turnpike Drive, Units 15-16 Hatboro, PA 19040 USA Toll-Free USA and Canada 877-479-ESCO Tel +1 215 441 9661 Fax +1 215 441 9660 us.escoglobal.com usa@escoglobal.com Esco Micro Pte. Ltd. 21 Changi South Street 1 Singapore 486 777 Tel +65 6542 0833 Fax +65 6542 6920 mail@escoglobal.com www.escoglobal.com 2008 Esco Micro Pte. Ltd. Specifications Subject to Change. ES1277 11/07 Esco Global Offices Singapore Philadelphia, USA Leiden, The Netherlands Salisbury, UK Kuala Lumpur, Malaysia Beijing, Shanghai & Guangzhou, China Mumbai, India Manama, Bahrain escoglobal.com LIVE SUPPORT Esco Micro Pte Ltd Cert. No: 651333 Cert. No: 651333/1