Procedures for Preparing and Receiving PUF Canisters and Filters

Similar documents
Title: Glassware Cleaning Copy No: ## SOP No.: 3.02/2.6/S Effective Date: June 18, 2013 Location: ### Glassware Cleaning

Cleaning Procedures in ICP-MS Laboratories

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ALPHA TRACK. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:

CHARCOAL CANISTER (Activated Charcoal Adsorption Device) by Proportional Counting

COMMERCIAL OFFICE TENANT RECYCLING PROCEDURES MANUAL

JRTC and Fort Polk Recycling Center Accepted Recyclable Commodities and Material Segregation Management

EQUIPMENT CLEANING Indira Pindi Watershed Monitoring September 20, 2017

STANDARD OPERATING PROCEDURE EQUIPMENT DECONTAMINATION PROCEDURES

California State University Bakersfield Radiation Safety Program Revised November 2017

Page 2 of 5. SOP SF SOP for Water Bath Maintenance

Davis Soil Moisture and Temperature Station Protocol

Quality Assurance Plan for Conducting Radon Measurements

Insert Preparation. Fast Forward Middle Station. Refer to the relevant CSL to turn ON equipment. Fast Forward Under Counter Fridge and Coldwell

SAFE WORK PROCEDURE. List the Hazards and risk controls as per risk assessment. assessment reference:

Installation, Operation and Maintenance Instructions

PDS 2010 Parylene Coater SOP

IQ/OQ Protocol Installation Qualification/ Operation Qualification

PERFORM Operating Document

Protocol for National Honey Bee Disease Survey

Standard Operating Procedure for Oil & Grease Method EPA1664A

UVM Cosmogenic Laboratory - Quartz Purity Testing

Pat's Place 7 Park Avenue, Mount Pearl, NL. FOOD PREMISES INSPECTION REPORT for Non-Compliant Critical / Non-Critical Items.

MODEL A5-2 SOLVENT & WATER RECOVERY SYSTEMS (EXPLOSION PROOF UNITS)

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: ELECTRET by Voltametry. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:

Fume Hood Safety TYPES OF FUME HOODS

MSU Extension Publication Archive. Scroll down to view the publication.

MODEL A18 SOLVENT RECOVERY SYSTEMS (EXPLOSION PROOF UNITS)

UNSIGNED HARDCOPY NOT CONTROLLED

THE CITY OF NEW YORK DEPARTMENT OF SANITATION

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

Street Address: Apartment/Unit #: Address: Street Address: Apartment/Unit #: QUALIFIED FOOD OPERATOR (INCLUDE COPY OF CERTIFICATE) YES NO

CUSTODIAL TRAINING GUIDE RESTROOM CARE

ChBE Safety Training. July Emily Edwards Lab Coordinator Chemical & Biological Engineering Tufts University 4 Colby St., Room

HACCP-Based Standard Operating Procedures

General Acids/Bases Bench Right and Left SOP

Rules for Safe Operation

Appendix C IBC Safe Use of Biosafety Cabinets SOP

1350 & AMMCO Brake Washer Systems. Installation and Operating Instructions With Safety and Maintenance Instructions and Parts Identification

This SOP and the B.C. Field Sampling Manual are available at: www2.gov.bc.ca.

ACQUITY UPLC HT Column Heater Instructions

Installation & Operating Instructions

until filters are dry. Once dry, place the filters shiny side up in petrislides. The petrislide do NOT need to be acid washed.

Standard Operating Procedure Glovebox and Thermal Evaporator

X-Ray Protection Program

The RSO will submit a copy of the revised manual to the RSC for approval when the revised manual has substantial changes.

Cavitator Ultrasonic Cleaners

Farmington Valley Health District 95 River Road, Suite C Canton, CT Phone (860) Fax (860)

Reading (E3) ESOL Skills for Life. Reading E3 (Practice Set 1) Practice Set 1 THE TIME ALLOWED FOR THIS TEST IS 1 HOUR 4/4/4 W37564A

PERFORM Operating Document. General maintenance of Stoves, Fridges and Pots and Pans in the Nutrition Suite

Polymer Characterization Facility 165 S. T. Olin Laboratory. Lab Safety Orientation

HPLC Column Ovens and Temperature Controllers

Operator s Manual. Rev 3/14/ O Neil Rd, Macedon NY Telephone: (315) Fax: (315)

SOP SF SOP for LN2 Freezers

IQ/OQ Protocol Installation Qualification/ Operation Qualification

SCIF SOP for BSL 2 samples on Aria3

PDMS Casting Station Supplies

By Silver King Refrigeration, Inc.

IQ/OQ Protocol Installation Qualification/ Operation Qualification SteamScrubber & FlaskScrubber Laboratory Glassware Washers

UNIFIED FACILITIES GUIDE SPECIFICATIONS

FOOD ESTABLISHMENT INSPECTION REPORT

Body Fluid Clean Up Kit: Supplies to ensure personal protection from bloodborne pathogens and bodily fluids.

Physical Resources Standard Operating Procedure

METHOD OF TEST FOR PREPARATION OF MARSHALL SPECIMENS 3. APPARATUS. Canadian Council of Independent Laboratories June 2016 LS-261 R27 ASTM D

Massachusetts Institute of Technology Kavli Institute for Astrophysics and Space Research (MKI)

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX

Special Precautions Isolation Plan Environmental Services

HEALTH AND SAFETY MANUAL

PIFA Heparin Platelet Factor 4 ( PIFA HPF4 ) Rapid Assay IQCP QUIZ. 1. How would a technician verify their PIFA HPF4 device is within expiration?

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

RCRA Emergency/Contingency Plan

The University of Washington Nanofabrication Facility (WNF) at Fluke Hall 125 Fluke Hall Seattle, WA 98195

General. While at work your safety also depends upon your colleagues. If you observe an unsafe situation, point it out to your colleague.

THREE RIVERS HEALTH DISTRICT APPLICATION FOR A TEMPORARY FOOD ESTABLISHMENT PERMIT

Operation Procedure for SMiF Wet Hoods (Solvent, Spin Coat, Acid, and Develop Hoods)

Safe Use of a Class II Biological Safety Cabinet

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

Centrifuge STANDARD OPERATING PROCEDURE

Permitting and Licensing for Hazardous Material Use and Storage

PHONE Web Address

CONVERSION KIT INSTRUCTIONS Commercial Electric Water heaters

Environmental Express, Inc. AutoBlock Plus Operation Manual and Instructions CONTENTS. i. Introduction/General Information 2

FOOD SAFETY TRAINING

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

OHIO UNIVERSITY RADIATION SAFETY REVIEW

Automated Support Removal System

Safety In Chemistry. Identified Risks and Hazards

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

I. C O N T E N T S T A N D A R D S

FOOD SAFETY TRAINING

Standard Operating Procedures (SOPs)

I N ST R UC T I ON MODEL 2612 & 2712 SLICERS MODELS 2612 ML ML FORM (12-98) 2612 SLICER

Osceola County Fair February 9-18, 2018 Gabriela Murza Extension Faculty FCS University of Florida/IFAS Extension - Osceola County

I N ST. MODEL 2912PS AUTOMATIC SLICER... with Portion Scale 701 S. RIDGE AVENUE TROY, OHIO

Polymer Characterization Facility 165 S. T. Olin Laboratory. Lab Safety Orientation

82. Siphon all gas out of the tank. 83. Disconnect the fuel filler and vent tube with a Phillips screwdriver.

MDN Project Specific SOP FGS-MDN-01. MDN Glass Sample Train Collection and Deployment- Trace Metal Clean Sample Handling

Ohio Department of Health, Bureau of Environmental Health and Radiation Protection, Radon Licensing Program. School Radon Testing Checklist

Centrifuge STANDARD OPERATING PROCEDURE

Hardware and Software Requirements

Transcription:

QSM Approval: Procedures for Preparing and Receiving PUF Canisters and 1. Introduction This Standard Operating Procedure provides instructions for preparing PUF canisters and filters for shipment to field sampling sites, and for processing incoming samples. 2. Procedure 2.1 Preparation of Canister and Filter for sampling 2.1.1 Washing Aluminum Foil Clean aluminum foil is used to wrapped filter after it has been sampled. It is to be included in every shipment to the field. When washing aluminum foil, gloves and safety glasses must be worn and washing must be done in the fumehood. Wash one side of the aluminum foil (~ 12 X 12 ) with acetone followed by hexane. Repeat on the other side. Allow to air dry then fold clean foil in four and place in a polyethylene bag. Collect all acetone/hexane waste in a beaker. Empty the solvent waste into the waste receptacle. Refer to SOP 4.11/*.*/S Waste for proper waste disposal. Refill Teflon solvent squirt bottles after use. 2.1.2 Pre-Weighing Wear gloves at all times when handling balance or filters. and its envelope are stamped with a discrete number and weighed before deployment into the field. The stamp was designed to print the same discrete number 3 times: Stamp the filter twice on the opposite edges. Do not stamp on surface where particles will be collected. Stamp the envelop once. Condition all filters to be weighed in the balance room approximately 48 hours prior to weighing. The initial filter weights are recorded in the site specific binder and the filter is placed in its envelope until deployment. 2.1.3 Washing /Assembling/Shipping PUF Canisters 1 of 7

After the PUFs (i.e. samples) have been removed from the canister, disassemble the canister for cleaning. The stainless steel liners are to be cleaned in the dishwasher in room 167. All other pieces are washed in hot, soapy (Sparkleen) water, rinsed with tap water, followed by de-ionized water, then set to dry on a clean surface lined with toweling. Wear gloves when handling clean canisters or pieces. Fresh PUFs, stored in glass jars, will be placed in liners, and then into canisters as close to actual shipment time as possible. Place stickers on the top and bottom of each canister to indicate to the operators which side should be out when the canister is reassembled after sampling. PUF canisters must be shipped for the sampling date indicated on the sampling schedule. Coordinate PUF canisters shipping with VOC shipping. Ship PUF canisters in coolers when possible. Some sites have their own coolers which should not be used to ship to any other site. Verify with supervisor for these specific sites. Include one piece of washed foil in a plastic bag for each canister being sent. On the return address label, mark the sender as AAQS. 2.2 Canister and Filter Receiving Samples arrive at the 335 River Road Shipping & Receiving loading dock. Personnel are informed of their arrival either verbally or via telephone by the Shipper/Receiver. They may have been shipped in a cooler, cardboard box or case normally used for 6L VOC canisters; therefore, regardless of when VOC canisters are processed, all received containers must be checked for the presence of PUF canisters and filters as quickly as possible. If there will be a delay in the processing of these samples, they must be placed immediately into the freezer in Lab 164. The date and time of receipt as well as the initials of the analyst receiving the sample, must be recorded on the data sheet (Form 7.11S/F1). Most samples will be received within two to four weeks of their sampling date; however, AAQS cannot control the rate of return from the site operators. During this time and while in transit, the samples are at ambient temperature. Thus, any loss of analytes resulting from samples sitting on the loading dock while awaiting processing would be negligible. 2 of 7

2.2.1 Processing PUF Canisters Note: If PAH samples have been placed in the freezer until time permits proper handling, they should be thawed for approximately one hour before being processed. Locate the data sheet. Record on it the date and time received and the initials of the person doing the receiving. Check to ensure that the data sheet and sample canister match. (If they do not match, try to locate the correct data sheet in the PAH Data Sheet binder. If this fails, store the canister in the freezer and contact the site operator to obtain the correct data sheet.) Check sampling time. Sampling time is calculated by subtracting the initial timer reading from the final timer reading. Record sampling time on the data sheet. Samples must have been run for 24 hours ± 1 hour (1440 minutes). Any sample not run for this time is invalid. Check sample volume. Sample volume is calculated by subtracting the initial gas meter volume from the final gas meter volume. Record sample volume on the data sheet. In an optimal condition, samples would have a volume of 720 m 3. There are two samples acceptance limits: Level 1 - Any sample with a volume out of (720 m 3 ± 10%) 648 and 792 m 3 range is flagged but not invalid. If several samples from a site arrive in succession with volume out of this range, notify lab supervisor and NAPS Operation unit manager. The field operators will have to repair/make adjustment to the meters. Level 2 - Any sample with a volume out of the 500 and 1200 m 3 range is invalid. Notify lab supervisor and NAPS Operation unit manager. The field operators would have to repair/make adjustment to the meters. If a sample is invalid, mark it as such on the data sheet and discard the PUF s and the corresponding filters. Locate the filter # in the PAH Filter Weights binder, enter the sample date, put the word VOID in the column for the final weight and initial under Analyst Initials. Any questionable samples should be kept in the freezer until the site operator can be contacted to verify the information on the data sheet. If the sample is valid, record the following information on a new plastic bag: 3 of 7

Location Canister ID Sample Date Filter Number Remove the plastic bag containing the filter from the brown envelope. Record the above information on this bag as well. Store this bag in the freezer. All valid samples are bagged and the filters are weighed within a week of reception. See procedures 2.2.2 and 2.2.3 below. Inventory each canister in Excel on the computer atslab in Rm 1701. START-MS EXCEL-C:\EXCEL\PAH. File all data sheets in the PAH Data Sheet binder according to site. The information from the data sheet is then entered onto the spreadsheet to determine the volume and total suspended particulates (TSP) of the sample corrected to standard temperature and pressure. The spreadsheet for each site is located on the share drive at Ncr.int.ec.gc.ca\Lab\Riv\Samplepreplabs. Each sample is then logged into the LIMS system, which will generate a unique lab code. The sample is then labeled with this lab code. 2.2.2 Bagging PUF s Loosen the top of the PUF canister. Lay down a large Kimwipe or piece of WYPALL toweling on the bench top in the fumehood. Do all work on this clean surface, inside the fumehood. Wearing nitrile or Chloroprene gloves, wash a piece of foil (~12 x 14 ) on both sides with acetone followed by hexane and clean a pair of tongs using the same procedure. Remove the top of the canister. Using the tongs, remove both PUFs from the canister, wrap them in cleaned foil making as small a package as possible and place this in the corresponding plastic bag. Seal the plastic bag and place bagged PUFs in the site specific bags in the freezer. Wash the tongs in-between each sample with acetone, followed by hexane. 2.2.3 Weighing When the filter is returned from the field, remove the filter from plastic bags. Unfold the foil covering and allow filter to equilibrate to the controlled room temperature and humidity for a minimum of 4 hours. 4 of 7

Take the binder ( PAH Filter Weights ) containing the filter weight sheets and a pen to the balance room. Calibrate the balance according to the steps outlined in the instrument manual. Remove each filter from its foil wrapping, unfold it and weigh it with the side containing the particulate up. Take care to handle each filter only on its edges, so as not to disturb any particulate on the filter. Record the filter weight on the sheet. Allow the balance to re-zero between each weighing. Tare as necessary. Re-fold and replace filter into its foil wrapper and its corresponding plastic bag. If sub-sampling is required, follow section 2.2.3.1. Re-weigh one filter in duplicate every 10 weighings to ensure the precision of the balance. Seal each plastic bag and store weighed filters in the freezer until they can be submitted for analysis. Calculate the change in mass of the filters by subtracting its initial weight from its final weight. Record this difference on the filter weight sheets and transfer this information to the data sheets. Note: The freezer is equipped with a digital thermometer, and a daily log of temperatures is maintained as per SOP 2.06/*.*/S. The freezer may require servicing if a daily minimum reading is not consistent with historical values. 2.2.3.1 Sub-sampling from filter for special studies Sub-sampling is carried out by punching out two 47 mm diameter discs from the (8 x10 inches, 20.32 x 25.4 cm; sampling area 18 x 23 cm) filter. Before use and in between punching out filters, rinsed the punch with acetone then hexane. Fold the filter half with the side containing the particulate inside. Punch once on the half folded filter using the rinsed 47 mm diameter punch (gives 2 discs). o Each disc has an area of 17.35 cm 2 which correspond to 4% of the sampling area of the filter. Wrap the discs individually in foil wrapper and place them in plastic bag. Label them with the following information: o Site Name o Sampling Date o Filter ID 5 of 7

o Field Sample No. o Total suspended particulate (TSP) mass on the whole filter Log in all the information for these discs in an archived samples inventory list. Seal the plastic bag and store these discs in the freezer until they can be submitted for analysis. 3. Revisions Sep. 1999: Author: Ron Halman; New Document May 2011: Lead Reviewer: Michael Lister May 26 2011 Renumbered to 3.19/1.0/S. Section 2.12-Change to procedures to reflect unit change. November, 2011: Lead Reviewer: Gary Poole Removed references to Air Toxics and replaced with AAQS throughout Section 2.3: added that the time, date and anlyst receiving the sample is recorded onto the datasheet accompanying the sample. Added that the information from the data sheet is entered onto the spreadsheet to determine the STP corrected volumes and TSP values and where the spreadsheet is located. Added to log the sample into the LIMS. Section 2.5: Added a statement on stamping the filters and envelopes with a discrete number and determine the initial weight before deployment to the field. Deleted with heat sealer from Sections 2.4 and 2.5 Section 2.5 Added for a minimum of 4 hours July 2013 Sept 2013 Leader Reviewer: May Siu/Jennifer Verner Deleted revision history 2009 and older Text and numbering revised for clarity Section 2.1.2: Added filter stamping procedure Section 2.2.1: Added samples volume acceptance limit Added Section 2.2.3.1 Sub-Sampling from filter for special studies 6 of 7

Lead Reviewer: Title: Jennifer Verner Technologist, Organic Laboratory Approved By: Title: May Siu Supervisor, Organic Laboratory 7 of 7