Alaris Syringe Module v9

Similar documents
Alaris Syringe Module v9

Alaris Syringe module FAQs

Syringe Pump. User Manual

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE

Administration of AMBISOME Using the GemStar Pump

Directions for Use Alaris System (with PC Unit, Model 8000)

M E D I C A L D E V I C E S

Enteral Pump Instructions for Use To Assemble When the low

Directions for Use. Alaris System (with Alaris PC Unit, Model 8015) September 2009

Mini Rythmic PN+ Set-up Guide

Section. 3: CADD Pump. Updated: June 20, Section Author(s): sstafford & jhval

ALLERGEN MIXING ASSISTANT

Sapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10


Clinician Information

Description of the CADD-Prizm VIP Pump


DS9400 Series. Release Notes for Firmware V2.07. Fire Alarm Control Panel

Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery.

VARIABLE RATE INFUSION PUMP

SINGLE SPEED SYRINGE PUMP

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

Pioneer-R16 Gas Monitor Operator s Manual

Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc CONTENTS. A. About this manual...2

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

bibag System User s Guide

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

Standard Operating Manual

H-10PM. Refrigerant Leak Detector

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Sensor Control Panel

APC BC300 Series 40kW 208/450/480V User Guide

KD-CLN-LP200 Ultrasonic Vinyl Record Cleaner User s Manual English v1.2 Printed in Korea

SCAN200E USER S MANUAL

OPERATOR S MANUAL. The Belmont Rapid Infuser, FMS Boston Road Billerica, MA 01821, USA US/Canada Worldwide

Analog Room Pressure Monitor RPC Series

Syringe Pump. Syringe Pump. Product Model: ZD-50C6 Syringe Pump. ZD-50C6-H Syringe Pump

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications

DS User Manual

Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

SN-50 Serial Syringe Pump User Manual. Version: 2.6

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

MacDonald Steel Limited

Jetstream Series HPLC - PELTIER COLUMN-THERMOSTATS 1. GENERALITIES

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to oven before servicing, unless testing requires it.

MPT Electric Fire Pump Controller

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to oven before servicing, unless testing requires power.

Rocket Quick Guide. Rocket PSU Portable Suction Unit

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

Technical Update TAA.TU rev. 1

MPT Electric Fire Pump Controller

USER MANUAL Digital CO Monitor Model CO30

(Place Logo from Lab here)

Thopaz+ Quick Reference Guide. Precious life Progressive care

MANUAL. Type ETO2 Controller for ice and snow melting. Warmup Inc. The world s best-selling electric floor heating brand

PF1000 CONTROLLER. for the. Heateflex Corporation 405 E. Santa Clara St. Arcadia, CA TEL: (626) ; FAX: (626)

Mark 25 Ultrapure Water Conductivity Analyzer

Laptop / PC Programming Manual

Transmission Cooling System J Operation Manual. TRANSFLOW Transmission Cooling System Service Tool J

one Operating instructions SenTRI ONE panel-based Fire detection and alarm system

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

Table of Contents. kangaroo 924. English. General Information... Operating Procedures... Battery... Beginning Administration...

IMPORTANT. PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times!

Original Perfusor & Infusomat

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

FLT93 Installation, Operation and Troubleshooting Guide

GLD-30 Gas Leak Detector

BeneVision TM80 Telemetry Monitor

Instruction Manual December 2001 ENACOCU02

WARM LINE. IV Fluid Warmer. Product Code: J1448. JORGENSEN LABORATORIES PHONE FAX

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

BeneVision Central Monitoring System. Quick Reference Guide

DEHUMIDIFIER OWNER S MANUAL. Read this manual carefully before operating the appliance and retain it for future reference.

Introduction. OM-THA2

CU-02 with Senaco Plus

GE Healthcare Healthcare Systems 9900 Innovation Drive Wauwatosa, WI USA

SIMPLICITY CO CARBON MONOXIDE DETECTION & VENTILATION PANEL

Portable Air-conditioner

Mitocell: Instructions for Use

Watchguard WGAP864 User Manual

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL

ZSC100 Gas Detection and Alarm System Controller

Troubleshooting your BAXJECT II Needle-less Transfer device

O PE R AT I ON HO851G REVOLVING TRAY OVEN. MODEL HO851G GAS ML Pan ML Pan ML Pan ML Pan

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

USER'S MANUAL PU SENS 01 (A11) PU SENS 01 (A19) Sensor Control Panel

AD-1272 Advanced Thermal Dispersion Probe Airflow Measuring System

Infusion System. Infusion and Syringe Pumps

Thermostats. Digital Programmable. GEAppliances.com

T34TM AMBULATORY SYRINGE PUMP

Pump Station Operator Training Checklist

PumpLab TM Operator s Manual

PRESSURA ROOM PRESSURE CONTROLLER MODELS 8630-SC 8630-PC

Transcription:

Cool Gray 7U Text- PMS 294U + QUICK REFERENCE GUIDE Operator Precautions: To ensure proper operation of the Alaris System (formerly known as Medley System ), user must be familiar with related features, disposables, administration sets, set-up and programming. This guide is not intended to be comprehensive instructions for the setup and operation of the Alaris System. For complete instructions along with Warnings and Cautions, refer to Alaris System Directions for Use (v9). Programming an infusion using Guardrails Suite MX Protection The programming user interface is identical for the Alaris Pump Module and the Alaris Module. Set Priming, Loading and Removal, Manual Priming of Dedicated Sets 1. Ensure set is not connected to patient and that device is level with patient. 2. Hold pressure sensing disc between two fingers and invert disc so patient side is up. 3. Slowly prime set while gently massaging disc to ensure disc does not become over or under filled and all air is expelled. 4. When priming is complete, clamp set. 5. Load syringe into device and install pressure sensing disc (for proper placement information, refer to system Directions for Use). 6. Program infusion. 7. Release clamp and press START. 8. Attach set to patient. Note: Starting infusion before attaching set to patient will minimize any potential bolus that could be released from pressure build up in set due to normal priming and syringe loading. Device Priming 1. Ensure set is not connected to patient and that device is level with patient. 2. Load and select syringe type and size. 3. Program infusion parameters but do not start infusion. 4. Press OPTIONS. 5. Select Prime Set with. 6. Hold pressure sensing disc between two fingers and invert disc so patient side is up. 7. Press and hold PRIME while gently massaging disc to ensure disc does not become over or under filled and all air is expelled. 8. Release PRIME when fluid flow is observed at end of set. 9. Install Pressure Sensing Disc. 10. Press EXIT. 11. Press START. 12. Attach set to patient. Note: Infusions programmed to infuse at low flow rates (<1.0mL/hr) may encounter a delay upon start up. When manual priming is complete, utilize Prime Set with feature to decrease potential start up delays. San Diego, CA 888.876.4287 2003-2010 CareFusion Corporation or one of its subsidiaries. All rights reserved.

Loading: Infusions 1. Open syringe barrel clamp until it clears syringe chamber. 2. Twist gripper control clockwise and raise drive head to fully extended position. 3. Insert syringe barrel flange between barrel flange grippers. 4. Lock syringe in place by closing barrel clamp. 5. Twist gripper control clockwise, lower drive head, lock plunger in place with plunger grippers. 6. Insert pressure disc (if used) with membrane facing device. Apply upward pressure until disc snaps into disc housing. Note: At the start of an infusion program, the system prompts to select and confirm syringe type and size. Ensure syringe manufacturer and size correctly identify the installed syringe. Mismatches may cause under-infusion or over-infusion to the patient. Drive Head Clamp Gripper Control/ Drive Head Release (closed position) Plunger Grippers (closed position) Flange Grippers Flange 2 oz 60 WARNING To prevent free flow, close set tubing clamp before loading or unloading syringe. 5 10 15 20 25 30 35 40 45 50 55 1½ 1 oz ½

Removal: 1. Close tubing clamp. 2. Open plunger grippers and syringe barrel clamp (silence safety alarm). 3. Remove syringe, apply downward pressure to remove disc from housing. Near End of Infusion When infusion is near completion, Near End of Infusion is enabled: Main display alternates between NEAR END and remaining VTBI. Yellow standby status indicator flashes. An audio prompt sounds which requires being silenced just once, and will not reoccur following initial silencing. All Mode enabled If the ALL option is enabled for the VTBI, entire contents of the syringe will be delivered and estimate of syringe volume will be displayed (unless your facility has established a KVO volume). Note: KVO is not available for intermittent infusions. All Mode disabled If the ALL option is disabled for the VTBI, a syringe volume estimate will be displayed after the VTBI soft key is selected. Selecting pressure limits, Pressure Sensing Disc installed 1. Press CHANNEL SELECT key. 2. Press OPTIONS key. 3. Press Pressure Limit soft key. 4. To enter a new pressure limit value, press Change Value soft key and use numeric keys. If Auto Pressure feature is enabled, press Auto Pressure soft key. Note: If Auto Pressure is selected and current pressure is: 100 mmhg or less - system adds 30 mmhg to current pressure, to create a new alarm limit. Greater than 100 mmhg- system adds 30% to current pressure, to create a new alarm limit. 5. Press CONFIRM soft key Pressure Sensing Disc not installed 1. Press CHANNEL SELECT key. 2. Press OPTIONS key. 3. Press Pressure Limit soft key. 4. Press appropriate soft key: Low, Med, High. 5. Press CONFIRM soft key.

Alarms and Troubleshooting: Alarms + Channel Disconnected Check Occlusion Channel(s) disconnected while in operation or there is a communication problem. Plunger grippers opened during infusion and then closed. Infusion stops on barrel clamp opened during infusion and then closed. Infusion stops on plunger not captured while in idle state. System alarms after 30 seconds, to indicate potential siphoning condition. Increased back pressure sensed while infusing. Infusion stops on affected module. To silence alarm and clear message from screen, press CONFIRM soft key. Reattach channel if desired, ensuring it is securely clicked into place at Channel Release Latch. If alarm is still present, replace module with an operational instrument. Securely lock plunger grippers, press CHANNEL SELECT key, and reselect syringe. Securely lock syringe barrel clamp and press RESTART key. Check for potential siphoning. Ensure administration set clamp (roller/slide) is in closed position. Securely lock plunger grippers over syringe plunger. Clear occlusion. Press RESTART key, or press CHANNEL SELECT key and then START soft key. Pressure Disc Installed Pressure sensing disc installed during an infusion. Infusion stops on Press CONFIRM soft key and restart infusion. Pressure Disc Removed Pressure sensing disc removed. Infusion stops on Reinsert pressure sensing disc and press RESTART key.

Alarms and Troubleshooting: Alarms Continued Empty is empty. If syringe is not empty, other possibilities are: Set up new infusion or press CHANNEL OFF key. Pressure sensing disc inappropriate/ defective. Verify appropriate pressure sensing disc is in use and functioning properly. Channel Error Calibration Required Driver Head Error plunger travel impeded. Pressure transducer defective Errors Error detected. Operation stops on Error on infusing module indicating calibration is required. Infusion stops on CALIBRATE scrolls in Message Display. Noninfusing module, with plunger grippers open, senses excessive pressure being applied downward on Drive Head. OCCLUSION scrolls in Message Display. Verify syringe plunger movement is not impeded. If syringe is not empty and above actions do not correct alarm, contact qualified service personnel. operation of unaffected modules, press CONFIRM soft key. Replace module with an operational instrument, as required. Service by qualified personnel is required. operation of unaffected modules, press CONFIRM soft key. Replace instrument with an operational instrument, as required. Service by qualified personnel is required. normal operation, press CONFIRM soft key. 12238765