PURELAB UHQ PS Operator Manual

Size: px
Start display at page:

Download "PURELAB UHQ PS Operator Manual"

Transcription

1 PURELAB UHQ PS Operator Manual

2

3 ELGA PURELAB UHQ PS Operator Manual Copyright Note The information contained in this document is the property of VWS (UK) Ltd. and is supplied without liability for errors or omissions. No part of this document may be reproduced or used except as authorized by contract or other written permission from VWS (UK) Ltd. The copyright and all restrictions on reproduction and use apply to all media in which this information may be placed. VWS (UK) Ltd. pursue a policy of continual product improvement and reserve the right to alter without notice the specification, design, price or conditions of supply of any product or service. VWS (UK) Ltd All rights reserved. Publication ref: MANU36619 Version 11 09/07 ELGA LabWater is a trading name of VWS (UK) Ltd. ELGA and PURELAB are registered trademarks. PURELAB UHQ PS Version 11-09/07 Page i

4 PURELAB UHQ PS Operator Manual ELGA Page ii PURELAB UHQ PS Version 11-09/07

5 ELGA PURELAB UHQ PS Operator Manual TABLE OF CONTENTS 1. HEALTH AND SAFETY NOTES Electricity Pressure Ultra-Violet Light Chemicals Control of Substances Hazardous to Health (COSHH) PRODUCT DESCRIPTION PROCESS DESCRIPTION Technical Specification of PURELAB UHQ PS INSTALLATION INSTRUCTIONS Service Connections Unpacking Installation START UP PROCEDURE FITTING REPLACEMENT CARTRIDGES Identification of Cartridges Replacing the UHQ-1 Adsorption Cartridge Replacing the UHQ-2 Ion Exchange Cartridge Replacing the UHQ-7 Ultra-Microfiltration Cartridge Replacing the UHQ-4 Composite Vent Filter Replacing the UHQ-6 UV Lamp SPECIAL CLEANING FOR CRITICAL APPLICATIONS CONSUMABLE ITEMS Cartridges KEY TO CONTROL PANEL WARRANTY/CONDITIONS OF SALE USEFUL CONTACT DETAILS...26 PURELAB UHQ PS Version 11-09/07 Page iii

6 PURELAB UHQ PS Operator Manual ELGA Page iv PURELAB UHQ PS Version 11-09/07

7 ELGA PURELAB UHQ PS Operator Manual 1. HEALTH AND SAFETY NOTES The PURELAB UHQ PS range of products have been designed with safety in mind, however it is important that personnel working on these units are familiar with any potential dangers. All safety information detailed in this handbook is highlighted as WARNING and CAUTION instructions. These are used as follows: WARNING! WARNINGS ARE GIVEN WHERE FAILING TO OBSERVE THE INSTRUCTION COULD RESULT IN INJURY OR DEATH TO PERSONS. CAUTION! Cautions are given where failure to observe the instruction could result in damage to the equipment, associated equipment and processes. 1.1 Electricity Those products requiring electrical supplies operate at mains supply voltages. For routine operations such as changing the Dl cartridge pack and Pretreatment cartridges it is sufficient to switch the unit off at the power inlet module located at the top rear left hand side. For more detailed work or access to the printed circuit boards the unit must be physically disconnected from the mains supply by removal of the mains feed plug. WARNING! IT IS ESSENTIAL THAT THE ELECTRICAL SUPPLY IS ISOLATED BEFORE ANY WORK IS STARTED. 1.2 Pressure Mains feed water pressure or the internal water pump are used to drive the water through the various stages of purification. This pressure could present a danger if not dissipated prior to changing cartridges etc. WARNING! IT IS ESSENTIAL THAT THE MAINS PRESSURE IS ISOLATED AND RESIDUAL PRESSURE IS ALLOWED TO DISSIPATE PRIOR TO REMOVAL OF ANY CARTRIDGES. Switching off the electricity will isolate the source of pressure on those units requiring an electrical supply. WARNING! ON THOSE UNITS WITHOUT REQUIREMENT FOR ELECTRICITY THE WATER PRESSURE SHOULD BE ISOLATED FROM THE MAINS WATER SUPPLY. PURELAB UHQ PS Version 11-09/07 Page 1

8 PURELAB UHQ PS Operator Manual ELGA 1.3 Ultra-Violet Light On some units short wavelength ultra-violet light is used to destroy micro-organisms. The blue light visible from the front of the unit is quite safe as the harmful component of the ultra-violet is filtered out. WARNING! THE UNIT SHOULD NEVER BE OPERATED WITH THE FRONT COVER REMOVED AS THE OPERATOR COULD BE EXPOSED TO SOME SHORT WAVELENGTH UV LIGHT. 1.4 Chemicals Cleaning using specialist chemicals is recommended on some units. Safety data sheets conforming to COSHH regulations will be supplied with any chemicals and these should be closely followed. WARNING! ENSURE THAT YOU AND ANY OTHER STAFF WORKING ON THESE PRODUCTS ARE FAMILIAR WITH THE POTENTIAL DANGERS AND THAT THE PROPER SAFETY EQUIPMENT IS AVAILABLE. IF IN DOUBT CONTACT YOUR LOCAL DISTRIBUTOR. 1.5 Control of Substances Hazardous to Health (COSHH) Material safety data sheets covering the various replaceable cartridges are available upon request if you contact your local distributor. Page 2 PURELAB UHQ PS Version 11-09/07

9 ELGA PURELAB UHQ PS Operator Manual 2. PRODUCT DESCRIPTION The Ultra High Quality Polishing System unit (PURELAB UHQ PS) has been specifically designed to meet the high purity demands of the modern analytical laboratory which is already served by a purified water supply. Several purification techniques have been packaged into a stylish self contained unit designed to stand alongside modern analytical instruments. Ultra high quality water produced by the unit is intended for use in such applications as atomic absorption spectrophotometry, ion chromatography, high performance liquid chromatography, fluorescence analysis and general microbiology. Fig. 1 illustrates the main external features visible to the user. Situated behind the side covers are the water purification cartridges, and behind the front cover are the electrical circuit boards and components. For normal use, The PURELAB UHQ PS is connected to a pre-purified water supply and electrical supply by the connections and sockets at the rear of the unit. Consistent water quality is assured by the in-built process technology with its recirculation features. Dispensing of the water is by unique pipette type Dispenser Pen, which enables fingertip control of the liquid flow. Briefly, within the PURELAB UHQ PS unit there is an inlet solenoid valve which controls the flow of pre-purified water into the unit reservoir, the contents of the reservoir are then pumped by a recirculation pump through four water purification cartridges. Water quality is continuously monitored by a temperature compensated resistivity meter and the water is irradiated with short-wave radiation in a UV chamber prior to flowing through the Dispense Pen. Recirculated water returns to the reservoir via a non-return valve. The tip of the Dispense Pen remains in a well, bathed in UV radiation, during periods of non-use. The right-hand removable cover needs to be opened every 1-2 months (by simply undoing the ¼ turn fasteners) in order to unclip and replace the exhausted media cartridges with new cartridges. Replacement cartridge packs as well as spare parts, service and/or advice are available from your local Distributor. This handbook describes the unique PURELAB UHQ PS process, how to install the PURELAB UHQ PS unit, and then how to carry out these cartridge changes. By following these simple instructions, you will be able to achieve and maintain the highest quality of water possible from your PURELAB UHQ PS. PURELAB UHQ PS Version 11-09/07 Page 3

10 PURELAB UHQ PS Operator Manual ELGA 3. PROCESS DESCRIPTION The PURELAB UHQ PS process links four purification technologies - organic absorption; de-ionisation; micro-filtration and photo-oxidation (see Fig. 2). Feedwater requirements for the PURELAB UHQ PS are a pre-purified water supply produced by a de-ioniser, reverse osmosis or still that is filtered to 0.2µm. Feedwater is passed via the inlet connection at the rear of the unit into the unit reservoir prior to further purification. A bar scale display on the front facia panel of the PURELAB UHQ PS, as shown in Fig. 3, indicates the volume of water in the reservoir. Level sensors within the reservoir switch off the recirculation pump at low level, and close the inlet solenoid at high level, stopping the feedwater to the reservoir. If required, pre-purified water from a deioniser, reverse osmosis unit or still, can be poured directly into the reservoir after pulling off the manual filling cap on the reservoir. This can speed up the conditioning of new cartridges. WARNING! CARE MUST BE TAKEN WHEN MANUALLY FILLING THE RESERVOIR. MAKE SURE THAT THE RESERVOIR IS NOT OVERFILLED AND THAT WATER IS NOT ACCIDENTALLY SPLASHED INTO THE VENT TUBE WHICH CONNECTS INTO THE NECK OF THE RESERVOIR. IF WATER GETS INTO THE COMPOSITE VENT FILTER IT WILL CAUSE IT TO COMPLETELY BLOCK, AND IT WILL THEN HAVE TO BE REPLACED. In normal use, the make-up to the reservoir is achieved by pressing the REFILL control button on the front facia panel (see Fig. 3), and observing that the red REFILL light is illuminated. This opens the feedwater inlet solenoid valve, which remains open until the reservoir is full, at which stage the inlet solenoid valve closes and the REFILL light switches off. When the PROCESS control button on the front facia panel (see Fig. 3) is pressed, the amber PROCESS light illuminates, indicating that the contents of the reservoir are now being recirculated through each of the various media cartridges (to remove the inorganic and organic impurities), the 0.05 micron filter cartridge (to remove any particulate matter and bacteria), and the short-wave UV lamp chamber (which further protects the now highly purified water from recontamination by micro-organisms and also assists the final removal of trace organics). The contents of the reservoir are processed up to the ultra high quality level during the course of a pre-timed period of continuous recirculation. This period is approximately 10 minutes. The green READY light illuminates after this time and the amber PROCESS light goes off. The unit will continue to operate for a further 10 minutes, after which time the unit will switch off and revert to a standby mode which is indicated by the READY light flashing. After approximately 60 minutes the unit will automatically restart and recirculate for 10 minutes and then switch off for 60 minutes again. Water can be obtained from the Dispense Pen during the off standby mode by depressing the PROCESS control button twice in quick succession, this will start the recirculation pump and enable water to be dispensed. Page 4 PURELAB UHQ PS Version 11-09/07

11 ELGA PURELAB UHQ PS Operator Manual Fig. 1 External Features of the PURELAB UHQ PS PURELAB UHQ PS Version 11-09/07 Page 5

12 PURELAB UHQ PS Operator Manual ELGA Fig. 2 Purification Technologies Within the PURELAB UHQ PS The PURELAB UHQ PS water quality is detected just before the Dispense Pen and is measured in MegaOhm-cm. The quality level is indicated on the right-hand bar scale display (over a 2 to 18 MegaOhm-cm range) on the front facia panel (see Fig. 3). During the PROCESS cycle and subsequent recirculation, the prepurified make-up water is prevented from automatically refilling the reservoir as this would reduce the quality of already processed water. To eliminate re-contamination of the purified water, inert grades of plastic have been selected for all wetted parts. This allows strong cleaning agents and disinfectants to be used when carrying out the special cleaning procedures for critical applications described in Section 6. The PURELAB UHQ PS covers are constructed from robust polyurethane moulded materials and supported on a coated steel chassis. The reservoir is protected from the ingress of airborne microorganisms, carbon dioxide and volatile organics, by the composite vent filter cartridge which filters all air entering the reservoir. For full technical specification of the PURELAB UHQ PS unit see Table 1. Fig. 3 Front Panel Layout Page 6 PURELAB UHQ PS Version 11-09/07

13 ELGA PURELAB UHQ PS Operator Manual 3.1 Technical Specification of PURELAB UHQ PS PURELAB UHQ PS Model No. Serial No: Reservoir Capacity: Process Time: Recirculation Feature: Dispense Rate: Feedwater Feedwater Requirement: Max. Inlet Pressure Min. Inlet Pressure: Feedwater Temperature: Dimensions Height: Width: Depth: Weight: Control Systems Power Requirements: Power Consumption Fuses Environment Noise Level See inside right-hand side cover or rear of unit. 4 litres (nominal), 3.3 litres usable volume. 10 minutes (pre-set continuous recirculation). 10 minutes in every 60 minutes intermittent recirculation after process time. Up to 0.75 litres/min water can be dispensed at any time. Pre-purified by deionisation, reverse osmosis or distilled quality better than 20 µs/cm, filtered 0.2µm. 6 bar (90 psi) bar (1 psi). 4 C 35 C. 400mm 232mm 400mm 10kg (empty) / 14kg (full). 230V ac, 50Hz (Standard). 115V ac, 60Hz (USA Standard). 100V ac, 50/60Hz (Optional). Note: Mains supply voltage fluctuation not to exceed ±10% of the normal voltage. 77VA 2 x 2.0 Amp Clean dry indoor. Altitude up to 2000m, or above 2000m if specified by the manufacturer. Temperature 5-40 C. Maximum Relative Humidity 31 C decreasing linearly to 40 C, non-condensing. Notes: 1) - This equipment is designed to be safe at least under these conditions. If the equipment is used contrary to the instructions the safety of the user may be impaired. 2) - The unit is in Installation Category II, Pollution Degree 2, as per IEC <40dBA Status Displays: LED Display of Refill (red), Process (amber), Ready (green) and Power (green). Tank level Indicator (graduated) Water Quality Indicator (temperature compensated). Safety Features: High Level cut-out in Reservoir. Low Level cut-out in Reservoir. Refill Inhibit during process. Water Purity International Standards: ASTM, CAP, NCCLS Type 1 and BS3978 Grade 1. Inorganics: 18MΩ-cm Resistivity at 25 C. Organics: T.O.C. Bacteria: Particles: <0.0001AU at 254nm. <20ppb <1CFU/ml 0.05µm filtration. PURELAB UHQ PS Version 11-09/07 Page 7

14 PURELAB UHQ PS Operator Manual ELGA 4. INSTALLATION INSTRUCTIONS 4.1 Service Connections Before commencing to install and operate the PURELAB UHQ PS please read and observe the following points: Environment The unit should be installed on a flat, level surface, in a clean, dry environment. The unit is designed to operate safely under the following conditions: Indoor Use. Altitude up to 2000m. Temperature Range 5-40 C. Maximum Relative Humidity 31 C decreasing linearly to 40 C, noncondensing. The unit is in Installation Category II, Pollution Degree 2, as per IEC Electrical Check that your voltage supply is compatible with the PURELAB UHQ PS unit. PURELAB UHQ PS units are built to 230V ac, 50Hz; 115V ac, 60Hz; or 100V ac, 50/60Hz Feedwater Pressure A maximum water pressure of 6 bar (90 psi) and minimum of 0.07 bar (1 psi) should be used to feed the PURELAB UHQ PS. The feedwater pressure of the pre-purified water supply will determine the time taken to fill the PURELAB UHQ PS reservoir. Typical refill times are: Inlet Pressure Refill Time 5 psi 10 minutes 30 psi 4 minutes 60 psi 2.5 minutes 90 psi 2 minutes Feedwater Quality (Standard) The PURELAB UHQ PS unit has been designed to operate with a pre-purified water supply of better than 20µs/cm and filtered to 0.2µm, produced by either de-ionisation, reverse osmosis or distillation. A suitable 0.2µm filter can be provided by the manufacturer for attachment to the incoming purified water supply. The manufacturer s reference for this PURELAB UHQ PS accessory is: Prefilter housing - LA518 Filter element - FILT20081 Page 8 PURELAB UHQ PS Version 11-09/07

15 ELGA PURELAB UHQ PS Operator Manual Alternative Method If a pressurised supply of pre-purified water is for Reservoir Refill not available, or it is not convenient to feed the inlet of the PURELAB UHQ PS unit from a storage tank with a positive head equivalent to 0.07bar (1psi); then direct filling into the PURELAB UHQ PS reservoir is permissible via the manual filler cap, see Fig. 1. Care should be taken when manually filling the reservoir not to overfill and wet the UHQ-4 Composite Vent Filter cartridge. The breather pipe for the UHQ-4 can be clearly seen protruding through the side of the filler neck. Hence: CAUTION! Stop the filling well below this breather pipe and do not fill up to the level of the pipe. If the UHQ-4 cartridge becomes wetted, it does not function and must be changed for a new UHQ-4 cartridge. 4.2 Unpacking 1. Carefully pull out the PURELAB UHQ PS from its carton and remove the top and bottom protective foam packing material. Lift the unit only from underneath, (not inclined) and position on the laboratory bench. 2. Check that the water fittings and electrical mains cable have been supplied. These are contained in a separate pack which is also to be found in the PURELAB UHQ PS carton. 4.3 Installation 1. Position the unit in its desired operating location. 2. Remove the transit plug from the inlet connector at the rear of the unit. To remove transit plug depress dark grey collet and withdraw plug (see Fig. 4). Note: Fig. 5 illustrates the rear of the PURELAB UHQ PS and shows the position of the inlet connections. 3. Insert a length of 8mm (5/16") OD flexible tube into the inlet connector and connect this tube to the feedwater supply. A 0.2µm filter should now be installed at a convenient point in the feedwater supply to the PURELAB UHQ PS unit. 4. Fit a 3 Amp fused mains plug to the electrical cable supplied with the unit. The mains cable supplied has a moulded plug to IEC.320.C14 Standard and should be inserted into the socket at the rear of the unit. The PURELAB UHQ PS is also supplied fitted with 2 x 2 Amp fuses in the power inlet module at the rear of the unit (see Fig. 5) (To check, simply prise open the fuse holder cover and snap-shut afterwards). Fig. 4 Tube Connector details PURELAB UHQ PS Version 11-09/07 Page 9

16 PURELAB UHQ PS Operator Manual ELGA Fig. 5 Rear View of Unit 5. Plug into the mains socket and switch the power on by pushing on the mains switch at the rear of the unit. The green Power light on the front facia panel will illuminate (see Fig. 3). Note: The PURELAB UHQ PS units are factory built to suit 110V ac, 115V ac or 230V ac supplies. Ensure unit is compatible with voltage supply. 6. Remove the right hand side cover of the unit (see Fig. 1) by undoing the four ¼ turn fasteners. Then remove the transit plug from tank breather pipe and connect to the UHQ-4 cartridge as per Fig Slowly turn on the feedwater supply to the PURELAB UHQ PS and check for leaks on the inlet water service connection at the rear of the unit. 8. Press the Refill control button on the facia panel and water will then pass through the inlet solenoid valve into the reservoir. The red Refill light will illuminate as the reservoir is filling up, and the volume bar scale will rise as the reservoir fills. 9. Turn off the feedwater and electric power supplies if the start-up procedure is not to commence at this time. The PURELAB UHQ PS is now ready for the Start-Up Procedure which is described in Section 5. Page 10 PURELAB UHQ PS Version 11-09/07

17 ELGA PURELAB UHQ PS Operator Manual 5. START UP PROCEDURE 1. The PURELAB UHQ PS should firstly be installed correctly on a horizontal surface as described in the instructions in Section Remove the right-hand side cover of the unit (see Fig. 1) by undoing the four visible ¼ turn fasteners and moving the cover sideways until clear of the unit. Then remove the transit plug from the tank breather pipe and connect to the UHQ-4 cartridge as per Fig. 6. Fig. 6 PURELAB UHQ PS with By-Pass Tubes Fitted As Supplied PURELAB UHQ PS Version 11-09/07 Page 11

18 PURELAB UHQ PS Operator Manual ELGA 3. Switch on the electrical power at the mains plug and on the PURELAB UHQ PS power inlet module. Turn on the feedwater supply. Only the green POWER light will be illuminated on the front facia panel. 4. Press the REFILL button. The green POWER light and red REFILL lights will both be illuminated. As the reservoir fills, the VOLUME bar scale display will rise. At this stage, ignore the QUALITY bar scale display. 5. Leave the unit until the VOLUME scale indicates 100% showing that the reservoir is full. The red REFILL light will switch off automatically when the reservoir is full and the make-up flow will stop. 6. Once the reservoir is fully filled (100% level on VOLUME scale).look for the tubing that is connected to the white tap near the top of the PURELAB UHQ PS (see Fig. 6), and direct it into a drain, sink or vessel capable of holding 5 litres. At this stage, the two vertical rigid by-pass tubes should be left in place as shown (see Fig. 6). 7. Open the white tap to allow the water to pass to drain and push the PROCESS control button on the facia panel. Water will be pumped from the reservoir to drain. This process will flush the UHQ-7, Ultra Microfiltration cartridge. During the first few minutes periodically bleed the air from the UHQ-7 cartridge by loosening the cap on the air bleed nipple, located on the side of the cartridge (see Fig. 6) and re-tightening when the air has been released. Note: This flush water will contain traces of a bacteriostatic agent which is added to the ultra microfiltration cartridge, both during transit and storage. The pump will automatically switch off after approximately 5 minutes and the flow to drain will cease. 8. When the reservoir is empty, the UHQ-1 and UHQ-2 cartridges can be connected. Firstly, disconnect the two rigid by-pass links by pulling them out from the flexible interconnecting tubing and storing them away safely for future use. 9. Pull off the protective caps from the ends of the two cartridges (UHQ-1 and UHQ-2), and connect them up as shown in Fig. 7, by simply pushing on the flexible silicon interconnecting tubing, ensuring there are no kinks or twists. 10. Refill the reservoir again by pressing the REFILL button. 11. Once the reservoir is fully filled (100% level on volume scale) ensure white tap is open and press the PROCESS button to start pump. This will drain the contents of the reservoir via the UHQ-1 and UHQ-2 cartridges just fitted. Note: It is important that during the first few minutes air is bled from the UHQ-7 cartridge as detailed in step 7. The pump will stop after approximately 5 minutes. Page 12 PURELAB UHQ PS Version 11-09/07

19 ELGA PURELAB UHQ PS Operator Manual 12. Turn off the white tap. Remove the drain tubing and retain for future use. Refit the right-hand cover and lock in place with the four ¼ turn fasteners. 13. The unit is now ready for normal daily use and the operation of the unit is as follows. 14. Push the REFILL button. The red REFILL light will illuminate. Fig. 7 PURELAB UHQ PS With Cartridges Fitted 15. Push the PROCESS button. The amber PROCESS light will illuminate. Note: The PURELAB UHQ PS s purification processing will now start and continue as the reservoir is filling up. It will continue for a full 10 minutes after the reservoir has reached its full (100%) level. 16. The REFILL light will go off when the reservoir is full (100%). 17. The PROCESS light will go off automatically after 10 minutes of recycling the full contents of the reservoir, and the READY light will illuminate to indicate the PURELAB UHQ PS quality levels have been achieved. PURELAB UHQ PS Version 11-09/07 Page 13

20 PURELAB UHQ PS Operator Manual ELGA 18. The PURELAB UHQ PS unit incorporates an intermittent recirculation feature which will allow the unit to operate for a further 10 minutes after the READY light is illuminated, after which time the recirculation pump will switch off and the READY light will be flashing. The unit will stay in this stand-by mode for 60 minutes, after which time the recirculation pump will restart for 10 minutes and then switch off for 60 minutes, this sequence is repeated every hour. 19. To dispense water, remove the Dispense Pen from its well and by depressing the spring loaded valve at the top of the Pen with an index finger or thumb, water can be dispensed at a controlled rate. The dispense valve can be locked on (to maintain the dispense without using a finger or thumb) by rotating the dispense valve clockwise while in the depressed position, to release turn anticlockwise. Note: To dispense water during the off stand-by periods as described in step 18, it will be necessary to depress the PROCESS button twice to start the recirculation pump. Once the pump has started wait until quality reading reaches the desired level, and then dispense water. Depending upon the speed of depressing the PROCESS button twice in succession, the pump will run for either 10 minutes or 20 minutes before switching off i.e. a) Depressing twice quickly - pump will run 10 minutes. b) Depressing twice slowly - pump will run 20 minutes (i.e. PRESS - wait 3 seconds - PRESS). 20. Return the Dispense Pen to its irradiated well when not in use. Page 14 PURELAB UHQ PS Version 11-09/07

21 ELGA PURELAB UHQ PS Operator Manual 6. FITTING REPLACEMENT CARTRIDGES 6.1 Identification of Cartridges There are five types of PURELAB UHQ PS Replacement Cartridges uniquely designed for use in the PURELAB UHQ PS. They are illustrated in Fig. 8 and are called: UHQ-1 - Adsorption Cartridge UHQ-2 - Ion Exchange Cartridge UHQ-7 - Ultra Microfiltration Cartridge UHQ-4 - Composite Vent Filter UHQ-6 - UV Lamp Tube. All cartridges except the UHQ-6 (UV) cartridges are accessible after removing the right-hand side cover by undoing the four ¼ turn fasteners and withdrawing the cover sideways. The UHQ-6 UV lamp tube is under the front cover. Always check that the electrical power supply to the PURELAB UHQ PS and the switch on the power inlet module at the rear of the PURELAB UHQ PS are switched off before attempting to change PURELAB UHQ PS cartridges. WARNING! IF REPLACING UHQ-1, UHQ-2 AND UHQ- 7 CARTRIDGES AT THE SAME TIME, THE UHQ-7 CARTRIDGES SHOULD BE FITTED FIRST AND RINSED AS PER SECTION 6.3 AND THEN THE UHQ-1 AND UHQ-2 CARTRIDGES FITTED AS PER SECTION 6.1 AND 6.2. After replacing the UHQ-1, UHQ-2 and UHQ-7 cartridges it is recommended that the first full reservoir of treated water produced is discarded. Frequency of Cartridge Replacements The following frequency of cartridge replacement is recommended for typical usage when fed with purified water: UHQ-1 and UHQ-2 every three months (maximum) or earlier if indicated on quality display (change both at same time). UHQ-7 and UHQ-4 every four months. UHQ-6 every twelve months. Information on shelf-life and availability of replacement cartridge packs is given in Section 8. PURELAB UHQ PS Version 11-09/07 Page 15

22 PURELAB UHQ PS Operator Manual ELGA Fig. 8 PURELAB UHQ PS Replacement Cartridges 6.2 Replacing the UHQ-1 Adsorption Cartridge 1. Remove the silicon tubing from the top of the exhausted UHQ-1 cartridge (see Fig. 9) mounted in the PURELAB UHQ PS and pull the cartridge away from its securing clips. 2. Place a finger over the top connector of the UHQ-1 cartridge and at the same time remove the silicon tube from the bottom of the cartridge. 3. Discard the exhausted cartridge. 4. Unpack a new UHQ-1 cartridge and remove the two protective caps on the inlet and outlet connections. 5. Locate the new cartridge into its retaining clips in the unit, ensure the cartridge is the correct way up as shown in Fig. 9, and is pushed down against the location tab at the bottom of the cartridge. Page 16 PURELAB UHQ PS Version 11-09/07

23 ELGA PURELAB UHQ PS Operator Manual 6. Refit the silicon tubing to the top and bottom of the cartridge ensuring the tubing is not kinked or twisted. 7. The UHQ-2 cartridge should now also be replaced and both cartridges rinsed, refer to Section Replacing the UHQ-2 Ion Exchange Cartridge 1. Remove the silicon tubing from the top of the exhausted UHQ-2 cartridge (see fig. 9) mounted in the PURELAB UHQ PS and pull the cartridge away from the securing clips. 2. Place a finger over the top connector of the UHQ-2 cartridge and at the same time remove the silicon tube from the bottom of the cartridge. 3. Discard the exhausted cartridge. 4. Unpack a new UHQ-2 cartridge and remove the two protective caps on the inlet and outlet connections. 5. Locate the new cartridge into its retaining clips in the unit, ensure the cartridge is the correct way up as shown in Fig. 9, and is pushed down against the location tab at the bottom of the cartridge. Refit the silicon tubing to the top and bottom of the cartridge ensuring the tubing is not kinked or twisted. 16. Having now fitted new UHQ-1 and UHQ-2 cartridges they should be rinsed as follows: 17. Turn on the water supply to unit and reconnect electrical supply; push the REFILL button and let the reservoir fill up (or manually top up with Dl water). 18. Open the white tap (see Fig. 9) and direct attached tubing to a drain, sink or vessel capable of holding 5 litres. 19. Start the pump by pushing the PROCESS button. During the first few minutes periodically bleed the air from the UHQ-7 cartridge by loosening the cap on the air bleed nipple, located on the side of the cartridge (see Fig. 9). Retighten bleed nipple when the air has been released. 20. The pump will automatically stop when the reservoir is empty, the white tap should be closed. Remove the drain tubing and retain for future use. 21. Refit the right hand cover. 22. The unit is now ready for use. PURELAB UHQ PS Version 11-09/07 Page 17

24 PURELAB UHQ PS Operator Manual ELGA Fig. 9 Fitting Replacement Cartridges 6.4 Replacing the UHQ-7 Ultra-Microfiltration Cartridge The UHQ-7 Ultra-Microfiltration Cartridge will require replacing about every four months, and should coincide with the replacement of the UHQ-1 and UHQ-2 cartridges. When the time to change comes, the UHQ-7 cartridge should be changed before the new UHQ-1 and UHQ- 2 cartridges are fitted. The UHQ-7 replacement cartridge is supplied containing an aqueous solution of hydrogen peroxide and the following procedure should be rigidly adhered to: 1. Ensure the unit is isolated from the electrical power supply. 2. Remove right hand side cover from unit. Page 18 PURELAB UHQ PS Version 11-09/07

25 ELGA PURELAB UHQ PS Operator Manual 3. Remove the silicon tubing from the bottom of the UHQ- 7 cartridge (see Fig. 9) and then by holding the body of the cartridge unscrew the cartridge anti-clockwise from the fixed top connector. There is no need to remove the bottom connector at this stage. 4. Having removed cartridge from unit, unscrew the bottom connector from the UHQ-7 cartridge body and place connector to one side for later use. The old UHQ- 7 cartridge can be discarded. WARNING! THE WEARING OF RUBBER GLOVES AND GLASSES IS ADVISED. 5. Unpack a new UHQ-7 cartridge from its sealed bag. The cartridge is supplied with an aqueous solution of hydrogen peroxide as a bacteriostat. To speed up the rinsing of the cartridge, both end caps on the new cartridge should be carefully removed and the contents poured to drain. 6. Holding the new cartridge upright so that the label can be read, the bottom connector removed in Step 4 can now be screwed onto the bottom of the UHQ-7 cartridge. Note: Ensure O ring is located inside connector prior to fitting. 7. Refit the new cartridge into the top connector in the PURELAB UHQ unit by screwing clockwise until hand tight. CAUTION! Do not over-tighten and cross thread. Tighten up only until resistance is felt against its sealing in the top container. 8. Refit silicon tubing to the bottom connector. 9. Reconnect electrical supply to unit, and mains water supply. 10. It is necessary to rinse the new UHQ-7 cartridge just fitted by passing one tank full of water to drain via the UHQ-7 cartridge. If tank is empty press REFILL button and let the reservoir refill. 11. Open white tap and place silicon tube attached to tap to convenient sink or drain point. 12. When tank is full press PROCESS button and allow tank contents to pass to drain. During this time the UHQ-7 cartridge should be bled of air by loosening the cap of the bleed nipple (see Fig. 9) until all the air is expelled. The cap should then be retightened. 13. When pump stops at low level the white tap should be closed. 14. The new UHQ-7 cartridge has now been fitted and rinsed. New UHQ-1 and UHQ-2 cartridges should now be fitted as described in Sections 6.1 and Replacing the UHQ-4 Composite Vent Filter 1. Remove the connecting tube from the top of the UHQ-4 Composite Vent Filter (see Fig. 9). 2. Pull the exhausted UHQ-4 filter away from its retaining clips and discard. 3. Fit the new UHQ-4 Composite Vent Filter into its retaining clips and then connect the tube (see Fig. 9). PURELAB UHQ PS Version 11-09/07 Page 19

26 PURELAB UHQ PS Operator Manual ELGA 6.6 Replacing the UHQ-6 UV Lamp 1. Turn off the feedwater and switch off the electrical supply to the PURELAB UHQ PS, and pull the cable and plug out of the PURELAB UHQ PS socket. 2. Remove the left and right side covers from the unit. 3. Remove the 4 fixing screws holding the front cover to the framework. These are accessible at the back corners of the front cover, having removed the two side covers. 4. When standing at the front of the unit, rotate the front panel, about its left-handed, vertical edge to expose the electronics compartment. Care should be taken to ensure no strain is applied to the two ribbon cables connecting the main power circuit board to the display circuit board. 5. The UV housing can now be seen at the bottom of the electronics compartment below the main power circuit board. The UV housing has a white coloured body with two elbow connectors fitted top and bottom, and two grey coloured caps. 6. Remove the old UV lamp by carefully prising off the white coloured end clips (see Fig. 10) and wires situated at either end of the UV housing end caps. Once the end clips have been removed, the lamp can be eased out of the housing. 7. Slide in a new lamp and reconnect the white end clips, ensuring the new lamp is not contaminated with fingermarks, as these will reduce the efficiency of the UV lamp. 8. Refit the PURELAB UHQ PS front and side covers, then plug the cable into the inlet power module, switch on the electrical supply, and then the feedwater supply. Fig. 10 View of UV Housing with Front Cover Removed Page 20 PURELAB UHQ PS Version 11-09/07

27 ELGA PURELAB UHQ PS Operator Manual 7. SPECIAL CLEANING FOR CRITICAL APPLICATIONS The following special cleaning procedures, using hydrogen peroxide (H 2 O 2 ), may be found beneficial for PURELAB UHQ PS users who require to maintain the lowest TOC or bacterial levels. The frequency of such a special clean will depend on many factors and cannot be predicted. As well as prior to critical applications, the special cleans would be of obvious benefit after a period of prolonged storage or non-use of the PURELAB UHQ PS. The procedure can be carried out at any time, by fitting the two vertical by-pass tubes supplied with each PURELAB UHQ PS when new. The procedure involves the recirculation of H 2 O 2 solution and four subsequent rinses using pre-purified water. The procedure for cleaning is as follows: 1. Remove the UHQ-1 and UHQ-2 cartridges, both of which are incompatible with H 2 O 2. The UHQ-7 Ultra Microfiltration cartridge is compatible and can be left in place during recirculation. 2. Fit the two rigid vertical by-pass tubes and push-fit the flexible interconnecting tubes as shown in Fig Check that the PURELAB UHQ PS is connected to the mains power and the POWER light is illuminated. 4. If the PURELAB UHQ PS has not been used for prolonged periods, empty the reservoir as follows: Open the white tap to allow the water to pass to drain and push the PROCESS control button on the facia panel. Water will be pumped from the reservoir to drain. 5. Fill the reservoir with pre-purified water, by pushing the REFILL button or, manually, using good quality deionised or distilled water through the top cap. 6. Add 140 ml of 30% w/v H 2 O 2 (100 volumes) and manually mix thoroughly. 7. Press the PROCESS button (the PROCESS light illuminates) and allow the unit to recirculate for 20 minutes. Ignore the illuminated READY light. During this time, periodically (1 minute approximately every 10 minutes), dispense the H 2 O 2 solution from the Dispense Pen back into the top of the reservoir. This allows the pen tip to be fully disinfected. 8. After 20 minutes drain the reservoir by connecting the tube to the white tap, described previously in Step Refill the reservoir as previously described in Step 5 (automatically, or manually). 10. Push the PROCESS button and recirculate for 5 minutes. 11. Again, drain the reservoir as previously described in Step 4 and flush the pen tip by periodically dispensing water back to the reservoir during each rinse. 12. Repeat steps (9) - (11) a further 3 times. 13. After the fourth rinse disconnect the two rigid vertical by-pass tubes and refit the new replacement UHQ-1 and UHQ-2 cartridges as described in Sections 6.1 and 6.2. PURELAB UHQ PS Version 11-09/07 Page 21

28 PURELAB UHQ PS Operator Manual ELGA 8. CONSUMABLE ITEMS 8.1 Cartridges UHQ-1 and UHQ-2 replacement cartridges are supplied in one box, known as Replacement Cartridge Pack A - comprising 2 each for the UHQ-1 and UHQ-2 cartridges. UHQ-7 and UHQ-4 replacement cartridges are available separately and are individually wrapped in heat sealed plastic containers. It is recommended that a Pack A and UHQ-7 and UHQ-4 cartridges are always held on stock by each PURELAB UHQ user. UHQ-1 Organic Adsorption Cartridge Max. Service Life Max. Shelf-life Pack No. 3 monthly 2 years 2 Cartridges of each in Pack A UHQ-2 Ion Exchange Cartridge 3 monthly 2 years UHQ-7 Ultra Microfiltration 4 monthly 2 years Individual Pack Cartridge UHQ-4 Composite Vent Filter 4 monthly 2 years Individual Pack UHQ-6 UV Lamp 12 monthly Indefinite Individual Pack All Packs and Spare Parts can be obtained from your local Distributor. For further information regarding spare parts references or technical data, please ask for the PURELAB UHQ PS Distributors Service Handbook available from your local Distributor. Note: Care should be taken to order the correct consumable items. Page 22 PURELAB UHQ PS Version 11-09/07

29 ELGA PURELAB UHQ PS Operator Manual 9. KEY TO CONTROL PANEL PURELAB UHQ PS Version 11-09/07 Page 23

30 PURELAB UHQ PS Operator Manual ELGA 10. WARRANTY/CONDITIONS OF SALE ELGA LabWater is a trading name of VWS (UK) Ltd. General Limited Warranty VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products. VWS (UK) LTD. MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED. THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The warranty provided herein and the data, specifications and descriptions of the VWS (UK) Ltd. products appearing in VWS (UK) Ltd. s published catalogues and product literature may not be altered except by express written agreement signed by an officer of VWS (UK) Ltd. Representations, oral or written, which are inconsistent with this warranty or such publications are not authorized and, if given, should not be relied upon. In the event of a breach of the foregoing warranty, VWS (UK) Ltd. sole obligation shall be to repair or replace, at its option, any product or part thereof that proves to be defective in materials or workmanship within the warranty period, provided the customer notifies VWS (UK) Ltd. promptly of any such defect. The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS (UK) Ltd. is willing and able to repair or replace any nonconforming VWS (UK) Ltd. product or part. VWS (UK) Ltd. shall not be liable for consequential, incidental, special or any other indirect damages resulting from economic loss or property damage sustained by any customer from the use of its products. Water System Limited Warranty VWS (UK) Ltd. warrants the water systems manufactured by it, BUT EXCLUDING MEMBRANES AND CARTRIDGES, against defects in materials and workmanship when used in accordance with the applicable instructions and within the operating conditions specified for the systems for a period of one year from the earlier of: a) the date of installation, or b) the 120 th day following the date of shipment. VWS (UK) LTD. MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED. THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The warranty provided herein and the data, specifications and descriptions of the VWS (UK) Ltd. systems appearing in VWS (UK) Ltd. published catalogues and product literature may not be altered except by express written agreement signed by an officer of VWS (UK) Ltd. Representations, oral or written, which are inconsistent with this warranty or such publications are not authorised and, if given, should not be relied upon. Page 24 PURELAB UHQ PS Version 11-09/07

31 ELGA PURELAB UHQ PS Operator Manual In the event of a breach of the foregoing warranty, VWS (UK) Ltd. sole obligation shall be to repair or replace, at its option, any product or part thereof that proves to be defective in materials or workmanship within the warranty period, provided the customer notifies VWS (UK) Ltd. promptly of any such defect. The cost of labor for the first ninety (90) days of the above warranty period is included in the warranty; thereafter, labor cost shall be at the customer s expense. The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS (UK) Ltd. is willing and able to repair or replace any nonconforming VWS (UK) Ltd. system or component part. VWS (UK) Ltd. shall not be liable for consequential, incidental, special or any other indirect damages resulting from economic loss or property damage sustained by any customer from the use of its process systems. Products or components manufactured by companies other than VWS (UK) Ltd. or its affiliates ( Non- VWS (UK) Ltd. products ) are covered by the warranty, if any, extended by the Product manufacturer. VWS (UK) Ltd. hereby assigns to the purchaser any such warranty; however, VWS (UK) LTD. EXPRESSLY DISCLAIMS ANY WARRANTY WHETHER EXPRESSED OR IMPLIED, THAT THE NON - VWS (UK) LTD. PRODUCTS ARE MERCHANTABLE OR FIT FOR A PARTICULAR PURPOSE. NOTICE VWS (UK) Ltd. is constantly striving to improve its products and services. Consequently, the information in this document is subject to change without notice and should not be construed as a commitment by VWS (UK) Ltd. Also, VWS (UK) Ltd. assumes no responsibility for any errors that may appear in this document. This manual is believed to be complete and accurate at the time of publication. In no event shall VWS (UK) Ltd. be liable for incidental or consequential damages in connection with or arising from the use of this manual. VWS (UK) Ltd. warrants its products against defects in materials and workmanship as described in the Warranty statement on the preceding pages. PURELAB UHQ PS Version 11-09/07 Page 25

32 PURELAB UHQ PS Operator Manual ELGA 11. USEFUL CONTACT DETAILS ELGA LabWater - Global Operations Centre Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel: +44 (0) Fax: +44 (0) techsupport@elgalabwater.com For the address of your nearest ELGA LabWater Sales and Service office visit the country list on our website or contact ELGA at the number above. ELGA is the global laboratory water brand name of Veolia Water. ELGA, PURELAB, MEDICA and CENTRA are trademarks of ELGA LabWater ELGA LAbWater/VWS(UK) Ltd. All rights reserved. Page 26 PURELAB UHQ PS Version 11-09/07

PURELAB Prima 7/15/30. Operator Manual

PURELAB Prima 7/15/30. Operator Manual PURELAB Prima 7/15/30 Operator Manual Copyright Note The information contained in this document is the property of VWS (UK) Ltd. and is supplied without liability for errors or omissions. No part of this

More information

PURELAB Classic - US. Operator Manual

PURELAB Classic - US. Operator Manual PURELAB Classic - US Operator Manual ELGA PURELAB Classic Operator Manual Copyright Note The information contained in this document is the property of VWS (UK) Ltd. and is supplied without liability for

More information

60 Liter Infinity Storage Reservoir

60 Liter Infinity Storage Reservoir BARNSTEAD THERMOLYNE CORPORATION 60 Liter Infinity Storage Reservoir OPERATION MANUAL AND PARTS LIST Series 902 Model D9021 LT902X1 7/30/97 Serial Number 1 Table of Contents Safety Information... 3 Warnings...

More information

RIOS 3 USER MANUAL MILLIPORE

RIOS 3 USER MANUAL MILLIPORE RIOS 3 USER MANUAL MILLIPORE Notice The information in this document is subject to change without notice and should not be construed as a commitment by Millipore Corporation. Millipore Corporation assumes

More information

Large Direct-Pure EDI Water System. User Manual. RephiLe Bioscience, Ltd.

Large Direct-Pure EDI Water System. User Manual. RephiLe Bioscience, Ltd. Large Direct-Pure EDI Water System User Manual RephiLe Bioscience, Ltd. Table of Contents 1 Introduction 1.1 Product Features........5 1.2 Main Applications....6 1.3 Specifications...7 1.4 Operation......

More information

Direct-Q 3, 5, 8 system Direct-Q 3 UV, 5 UV, 8 UV system User manual

Direct-Q 3, 5, 8 system Direct-Q 3 UV, 5 UV, 8 UV system User manual Direct-Q 3, 5, 8 system Direct-Q 3 UV, 5 UV, 8 UV system User manual Notice The information in this document is subject to change without notice and should not be construed as a commitment by Millipore

More information

STEAM WALLPAPER STRIPPER MODEL HTW5

STEAM WALLPAPER STRIPPER MODEL HTW5 STEAM WALLPAPER STRIPPER MODEL HTW5 From Serial Number 75154 (110 Volt North America only) OWNERS MANUAL & OPERATING INSTRUCTIONS 2016/11 Hiretech Part # 007717 WARNING For safe operation of this machine,

More information

Installation Instructions PP-AS20 Anti-Scale System

Installation Instructions PP-AS20 Anti-Scale System Installation Instructions PP-AS20 Anti-Scale System THIS UNIT MUST BE INSTALLED BY A LICENSED PLUMBER TO VALIDATE THE WARRANTY. COMPONENTS PP-AS20 Filter Housing PP-AS20R Filter Cartridge Installation

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS Please read carefully and retain for future reference. CONTENTS 1. Important Information 2. Included Parts 3. Filter Head Assembly Instructions 4. Installation Instructions 4.1. Auxiliary Tap Connection

More information

Parasitic Cysts - Cryptosporidium and Giardia, Chemicals, Heavy Metals, Dissolved Salts, Offensive Tastes and Odours, Sediment, Dirt,

Parasitic Cysts - Cryptosporidium and Giardia, Chemicals, Heavy Metals, Dissolved Salts, Offensive Tastes and Odours, Sediment, Dirt, Aqua-Pure is a market leader in water filtration. In addition to our comprehensive range of Domestic Water Purifiers, we design and manufacture an extensive commercial and industrial range of purification

More information

NuPure LAB. Ultrapure and RO with PE Tank USER MANUAL

NuPure LAB. Ultrapure and RO with PE Tank USER MANUAL NuPure LAB Ultrapure and RO with PE Tank USER MANUAL This Manual Is Applicable to the Following NuPure-Direct Models NW0P01000S NuPure-Direct Water System, Ultrapure 10 NW0P02000S NuPure-Direct Water System,

More information

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741

More information

MKTF CRO-Series 5-Stage Reverse Osmosis System Installation and Maintenance Manual

MKTF CRO-Series 5-Stage Reverse Osmosis System Installation and Maintenance Manual MKTF - 215 CRO-Series 5-Stage Reverse Osmosis System Installation and Maintenance Manual Please fill out the following information at the time of installation. Save for future reference. Model: Date Code:

More information

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations Model G32-S Model G32-E Disinfection Soak Stations Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482 Website: WWW.CIVCO.COM Copyright

More information

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Product Preservers Anti-Scale System Chemical-Free, Salt-Free Scale Prevention Introduction The Product Preservers Anti-Scale System will condition the tap

More information

Carbon Dioxide Sample-Draw Detector Operator s Manual

Carbon Dioxide Sample-Draw Detector Operator s Manual 35-3001-05-03 Carbon Dioxide Sample-Draw Detector Operator s Manual Part Number: 71-0413 Revision: 0 Released: 8/4/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

ASPIRE Laboratory Aspirator

ASPIRE Laboratory Aspirator ASPIRE Laboratory Aspirator USER MANUAL Rev 2/14/18 Accuris Instruments / Benchmark Scientific Ph: (908) 769-5555 E-mail: info@accuris-usa.com (C) Benchmark Scientific, 2018 THE ACCURIS ASPIRE LABORATORY

More information

Recirculating Chiller RC350G-RC400G RC1400G-RC3000G. Operating Manual

Recirculating Chiller RC350G-RC400G RC1400G-RC3000G. Operating Manual Recirculating Chiller RC350G-RC400G RC1400G-RC3000G Operating Manual Part No 09931/Issue 10 Page 2 April 2010 CONTENTS 1 Safety 4 2 Installation 5 2.1 Connection for closed loop circulation 6 2.2 Connection

More information

MaxLite LED MICRO-T PANEL

MaxLite LED MICRO-T PANEL ` Installation Instructions General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all

More information

MaxLite LED Vapor Tight Linear Fixture

MaxLite LED Vapor Tight Linear Fixture General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all warnings and instructions

More information

MODEL 7000 SUCTION UNIT

MODEL 7000 SUCTION UNIT MODEL 7000 SUCTION UNIT OPERATOR S MANUAL Caution Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices

More information

Laing Thermotech. Autocirc The Instant Hot Water Pump Models E1-BCANCT1W-06 and E1-BCANRT1W. Installation & Operating Manual

Laing Thermotech. Autocirc The Instant Hot Water Pump Models E1-BCANCT1W-06 and E1-BCANRT1W. Installation & Operating Manual Installation & Operating Manual Please read this manual carefully before attempting to install, operate or maintain the product described. Failure to comply with the information provided in this manual

More information

OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN

OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 10-12 Ross Place Wetherill Park NSW 2164. P.O. Box 1851 Wetherill

More information

BARNSTEAD THERMOLYNE CORPORATION. NANOpure Infinity UF OPERATING MANUAL AND PARTS LIST. Series 898 D D LT898X1 5/11/98

BARNSTEAD THERMOLYNE CORPORATION. NANOpure Infinity UF OPERATING MANUAL AND PARTS LIST. Series 898 D D LT898X1 5/11/98 BARNSTEAD THERMOLYNE CORPORATION NANOpure Infinity UF OPERATING MANUAL AND PARTS LIST Series 898 Model # Voltage D8981 120 D8982-33 230 LT898X1 5/11/98 Serial Number 1 Table of Contents Safety Information...

More information

TrueClean ToteCleaner

TrueClean ToteCleaner TrueClean ToteCleaner Installation, Operation, and Maintenance Manual www.trueclean.us TrueClean Check Valve Table of Contents Introduction.... 1 About... 1 Warranty... 2 Safety... 3 Important Safety Information....

More information

Direct-Pure Water System. User Manual

Direct-Pure Water System. User Manual Direct-Pure Water System EDI with PE Tank User Manual RephiLe Bioscience, Ltd. Table of Contents 1 Introduction 1.1 Product Features........5 1.2 Main Applications....6 1.3 Specifications...7 1.4 Operation......

More information

Humidifier. Owner s Guide. English. Model Form# Hunter Fan Co Hunter Fan Company

Humidifier. Owner s Guide. English. Model Form# Hunter Fan Co Hunter Fan Company Ultrasonic Humidifier Owner s Guide English Model 31206 31207 Form# 44064-01 2011624 2011 Hunter Fan Co. 1 Table of Contents At A Glance Warnings and Introduction... 3 Know Your Humidifier... 4 Using Your

More information

Combustible Gas Sample-Draw Detector Operator s Manual

Combustible Gas Sample-Draw Detector Operator s Manual 35-3001-01 Combustible Gas Sample-Draw Detector Operator s Manual Part Number: 71-0282RK Revision: A Released: 8/3/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

MaxLite LED Round Pendant Highbay Series

MaxLite LED Round Pendant Highbay Series General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts, cuts/abrasions, and other hazards read all warnings and instructions

More information

UFM-T100H US ES CA Power Alarm Test TM

UFM-T100H US ES CA Power Alarm Test TM Alarm Power TM UFM-T100H US ES CA A E F I J K G B C D H L M N O P Q R S T U V W X US 1 3 2 Power Alarm 4 6 5 Power Alarm 7 9 8 Po we Ala r rm Power Alarm 10 11 Po we Po we Ala rm r 12 Pow Ala Ala rm r

More information

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1 Instruction Manual M6627 / Version 7.4 Slide Drying Bench MH6616 MH6616X1 Please take your time to read this Instruction Manual in order to understand the safe and correct use of your new Electrothermal

More information

DRY AIR SYSTEMS, INC Metro Boulevard Maryland Heights, Missouri (314) fax (314)

DRY AIR SYSTEMS, INC Metro Boulevard Maryland Heights, Missouri (314) fax (314) DRY AIR SYSTEMS, INC. 2655 Metro Boulevard Maryland Heights, Missouri 63043 (314) 344-1114 fax (314) 344-0677 HD SERIES DRIERS TABLE OF CONTENTS WHY AN AIR DRYER 3 WHAT IS A DESICCANT AIR DRYER 3 Desiccant

More information

Operator s Manual. Histology Bath

Operator s Manual. Histology Bath Operator s Manual Histology Bath 110-827 05.09.12 Table of Contents Introduction... 2 General Safety Information... 2 Safety Recommendations... 3 Unpacking Your Histology Bath...4 Contents...4 Components

More information

TECHNOSKLO Ltd. Držkov

TECHNOSKLO Ltd. Držkov TECHNOSKLO Ltd. Držkov Distillation apparatus for producing of distilled water TYPE DP 4000 Operating manual TECHNOSKLO Ltd. 468 24 Držkov 135 Czech Republic tel.: 00420 483 346 403 fax: 00420 483 385

More information

Distinction. D4000 & D4000/EURO Water Still. Assembly and Operating Instructions. 1 Issue 2

Distinction. D4000 & D4000/EURO Water Still. Assembly and Operating Instructions. 1 Issue 2 Distinction D4000 & D4000/EURO Water Still Assembly and Operating Instructions 1 Issue 2 Fig. 1 Distinction D4000 & D4000/EURO Water Still 2 Issue 2 Before Use If the equipment is not used in the manner

More information

92831 TEL: (714) FAX:

92831 TEL: (714) FAX: Document No. 1800-11 High-Performance Vacuum/Nitrogen Copyright 2010 Terra Universal Inc. All rights reserved. Revised October 2010 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton,

More information

Operator s Manual. IP-100 Immersion Probe Cooler

Operator s Manual. IP-100 Immersion Probe Cooler Operator s Manual IP-100 Immersion Probe Cooler 110-810 04.27.11 Table of Contents Introduction... 3 General Information... 3 General Safety Information... 3 Safety Recommendations... 4 Unpacking Your

More information

BMW E36 Thermostat Removal And Coolant Flush

BMW E36 Thermostat Removal And Coolant Flush BMW E36 Thermostat Removal And Coolant Flush Disclaimer: The cooling system is critical to the proper operation of your car. Failure to properly install all of the components of the cooling system could

More information

EASYpure RF. Series 1051 OPERATION MANUAL AND PARTS LIST BARNSTEAD THERMOLYNE CORPORATION

EASYpure RF. Series 1051 OPERATION MANUAL AND PARTS LIST BARNSTEAD THERMOLYNE CORPORATION BARNSTEAD THERMOLYNE CORPORATION EASYpure RF OPERATION MANUAL AND PARTS LIST Series 1051 Model D7032 Model D7031 Model D7033 Model D7033-33 100 Volts 120 Volts 240 Volts 230 Volts LT1051X1 8/22/97 1 Table

More information

ispring Whole House Water Filter Systems USER S MANUAL

ispring Whole House Water Filter Systems USER S MANUAL ispring Whole House Water Filter Systems USER S MANUAL Version 2014-5 Introduction Congratulations on your purchase of the ispring Whole House Water Filter system. Featuring a three-stage filtration process,

More information

smardy - pure water Reverse osmosis system YR100-A 75 GPD

smardy - pure water Reverse osmosis system YR100-A 75 GPD smardy - pure water Reverse osmosis system YR100-A 75 GPD Instruction manual [EN] Please read this manual carefully before operating the device and subsequently keep the instruction manual safely and within

More information

Model A Pipe Line Strainer 3, 4, 6, 8 & 10 Inch (DN80, DN100, DN150, DN200 & DN250) 175 psi (12,1 bar) General Description.

Model A Pipe Line Strainer 3, 4, 6, 8 & 10 Inch (DN80, DN100, DN150, DN200 & DN250) 175 psi (12,1 bar) General Description. Technical Services: Tel: (00) 31-9312 / Fax: (00) 91-5500 Customer Service/Sales: Tel: (1) 50-5000 / (00) 55-523 Fax: (1) 50-5010 / (00) -1295 Model A Pipe Line Strainer 3,,, & 10 Inch (DN0, DN100,, DN200

More information

CONGRATULATIONS ON CHOOSING THE WATERLOGIC 4 FIREWALL YOUR WATERLOGIC WATER DISPENSER

CONGRATULATIONS ON CHOOSING THE WATERLOGIC 4 FIREWALL YOUR WATERLOGIC WATER DISPENSER User Manual CONGRATULATIONS ON CHOOSING THE WATERLOGIC 4 FIREWALL Thank you for choosing the Waterlogic 4 Firewall, the advanced water dispenser. All Waterlogic products are designed to deliver pure refreshing

More information

INSTRUCTION MANUAL HS-229G

INSTRUCTION MANUAL HS-229G INSTRUCTION MANUAL HS-229G 510977 STEP 1 - Where to Install the Thermostatic Steam Trap Determine where to install the thermostatic steam trap based on the following information. a. The trap should be

More information

BWT ECO. Fitting and operating instructions. Ultra Violet Water Disinfection System. ECO Range: LC 8 UV LC19 UV LC36 UV LC51 UV

BWT ECO. Fitting and operating instructions. Ultra Violet Water Disinfection System. ECO Range: LC 8 UV LC19 UV LC36 UV LC51 UV Fitting and operating instructions EN BWT ECO Ultra Violet Water Disinfection System ECO Range: LC 8 UV LC19 UV LC36 UV LC51 UV BWT ECO Range /06-2016 BWT UK Limited Changes of technical details are reserved

More information

User, Installation and Servicing Instructions. Silverlink 600 Gas Griddles GS4 and GS7 IS212 ECN3592

User, Installation and Servicing Instructions. Silverlink 600 Gas Griddles GS4 and GS7 IS212 ECN3592 User, Installation and Servicing Instructions Silverlink 600 Gas Griddles GS4 and GS7 IS212 ECN3592 INSTALLATION AND SERVICING INSTRUCTIONS Please read the following carefully before commencing work on

More information

User and Installation Instructions. Wall Mounted Water Boiler WMB3F/B and WMB3F/W IS435 ECN3461

User and Installation Instructions. Wall Mounted Water Boiler WMB3F/B and WMB3F/W IS435 ECN3461 User and Installation Instructions Wall Mounted Water Boiler WMB3F/B and WMB3F/W IS435 ECN3461 Dear Customer Thank you for purchasing this Lincat product. With correct use and careful maintenance as described

More information

JUNEAU JUN. 08/51193/0 Issue 0

JUNEAU JUN. 08/51193/0 Issue 0 JUNEAU JUN 08/51193/0 Issue 0 The product complies with the European Safety Standards EN60335-2-30 and the European Standard Electromagnetic Compatibility (EMC) EN55014, EN60555-2 and EN60555-3 These cover

More information

Eco-Steamer SSC-0312 INSTRUCTION MANUAL CONTENTS

Eco-Steamer SSC-0312 INSTRUCTION MANUAL CONTENTS INSTRUCTION MANUAL Eco-Steamer SSC-0312 CONTENTS Important Safeguards...1 Description of Eco-Steamer...2 Main Features of Eco-Steamer...2 Instructions for Use...3 How to Assemble and Use...3 Accessories...3

More information

HC12W COOL MIST ULTRASONIC HUMIDIFIER OWNER S MANUAL. Read and save these instructions.

HC12W COOL MIST ULTRASONIC HUMIDIFIER OWNER S MANUAL. Read and save these instructions. COOL MIST ULTRASONIC HUMIDIFIER HC12W OWNER S MANUAL Read and save these instructions. 2 BECAUSE YOU RE NOT LIKE EVERYONE ELSE And neither are we. Always at the forefront of our industry, our goal is to

More information

OPERATING INSTRUCTIONS FOR:

OPERATING INSTRUCTIONS FOR: OPERATING INSTRUCTIONS FOR: MODEL: BL-30 & BL-130 REFRIGERATED BATH Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 10-12 Ross Place Wetherill Park NSW 2164. P.O. Box 1851 Wetherill Park DC

More information

Pressurized Filter Up to 1250 gallons

Pressurized Filter Up to 1250 gallons Pressurized Filter Up to 1250 gallons REMINDER CALL 1-888-755-6750 BEFORE RETURNING TO STORE. ITEM #FP1250UV Questions, problems, missing parts? Before returning to your retailer, call our customer service

More information

MaxLite LED VAPOR TIGHT LINEAR FIXTURES

MaxLite LED VAPOR TIGHT LINEAR FIXTURES A cost effective LED Vapor Tight Fixture features a full length polycarbonate lens and a one-piece white glass reinforced polyester (GRP) body. Designed to meet or exceed seven to 10 foot-candles at 8

More information

Operating & Safety Instructions

Operating & Safety Instructions EUROPEAN FINISHING EQUIPMENT Corp. 901 PENHORN AVE., UNIT #2 SECAUCUS, NJ 07094 PHONE (201) 210-2247 FAX (201) 210-2549 sales@histeam.com www.histeam.com Operating & Safety Instructions - - - - - - - -

More information

Reacti-Therm I and III Heating Modules

Reacti-Therm I and III Heating Modules INSTRUCTIONS Reacti-Therm I and III Heating Modules TS-18822 TS-18824 Number TS-18822 TS-18824 Description Reacti-Therm I Heating Module (single block) Reacti-Therm III Heating Module (triple block) 2101.1

More information

ELSTON 52 CEILING FAN

ELSTON 52 CEILING FAN ELSTON 52 CEILING FAN MODEL #10290 Español p. 21 Questions, problems, missing parts? Before returning to your retailer, call our customer service department at 1-877-361-3883, Monday - Thursday, 8 am -

More information

DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4

DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4 DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4 Congratulations on your purchase of an Aquaport Executive Filtered Water Cooler This unit is easy to install and provides great tasting chilled

More information

TOUCHDOWN 48 CEILING FAN

TOUCHDOWN 48 CEILING FAN TOUCHDOWN 48 CEILING FAN MODEL #50205 Español p. 20 Questions, problems, missing parts? Before returning to your retailer, call our customer service department at 1-877-361-3883, Monday - Thursday, 8 am

More information

Instruction Manual. Cheese Warmer with Pump

Instruction Manual. Cheese Warmer with Pump Instruction Manual Cheese Warmer with Pump 10700 Medallion Drive, Cincinnati, Ohio 45241-4807 USA 2013 Gold Medal Products Co. Part No. 38298 SAFETY PRECAUTIONS Page 2 INSTALLATION INSTRUCTIONS Inspection

More information

A measured step forward

A measured step forward A measured step forward Operations & Maintenance Manual TM JESCO UV 405-412-440-450-480-550-80 2 LCD LCD PLUS Contents 1. Introduction...3 2. General Principles and Safety Instructions...4 3. Instructions

More information

User Manual. Milli-Q Integral 3/5/10/15 Systems

User Manual. Milli-Q Integral 3/5/10/15 Systems User Manual Milli-Q Integral 3/5/10/15 Systems About this User Manual Purpose This User Manual is intended for use with a Milli-Q Integral Water Purification System. This User Manual is a guide for use

More information

Installation, Operation and Maintenance Manual for Model 8331 Condensate Accumulator

Installation, Operation and Maintenance Manual for Model 8331 Condensate Accumulator Installation, Operation and Maintenance Manual for Model 8331 Condensate Accumulator 2008 Groth Corporation IOM-8331 Ref. ID: 95566 TABLE OF CONTENTS: GENERAL SAFETY INSTRUCTIONS 3 SAFETY WARNINGS 3 INSPECTION

More information

BOILING UNIT REDITAP. Installation and User Guide. IMPORTANT: This booklet should be left with the user after installation and demonstration

BOILING UNIT REDITAP. Installation and User Guide. IMPORTANT: This booklet should be left with the user after installation and demonstration in tap Boiling water to in tap sink Drain Valve (as high as possible) REDITAP CONNECTION SUMMARY Amp mains supply cold mains water into in tap optional filter cold water in hot water BOILING UNIT Installation

More information

Cleanshield HEPA 550 Owner's Manual

Cleanshield HEPA 550 Owner's Manual Cleanshield HEPA 550 Owner's Manual READ AND SAVE THESE INSTRUCTIONS Seaira Global LLC 14021 NC Highway 50 Surf City, NC 28445 910-660-0962 sales@seairaglobal.com WARNING! Please note that the CleanShield

More information

EASYpure LF. OPERATION MANUAL AND PARTS LIST Series 1052 BARNSTEAD THERMOLYNE CORPORATION

EASYpure LF. OPERATION MANUAL AND PARTS LIST Series 1052 BARNSTEAD THERMOLYNE CORPORATION BARNSTEAD THERMOLYNE CORPORATION EASYpure LF OPERATION MANUAL AND PARTS LIST Series 1052 Model D7383 Model D7381 Model D7382-33 100 Volts 120 Volts 230 Volts LT1052X1 3/27/98 1 Table of Contents Safety

More information

Ceiling Mount Air Purification System

Ceiling Mount Air Purification System Ceiling Mount Air Purification System HA-CMP-G2-R/HA-CMP-G2-OV Owner s Manual Table of Contents HealthyAir Series 1219 Filters 1 Important Safety Instructions 2 Technical Specifications 3 Packaging Reference

More information

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU-20B CompriCart TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the Unit...................2 -

More information

The Coffee Professor Reverse Osmosis System Model Genesis Cafe

The Coffee Professor Reverse Osmosis System Model Genesis Cafe The Coffee Professor Reverse Osmosis System Model Genesis Cafe Congratulations on your purchase of the most advanced water purifier system available! Installation and Service Guide PLEASE READ THIS MANUAL

More information

CT-7 Series Toxic Gas Transmitter Operator s Manual

CT-7 Series Toxic Gas Transmitter Operator s Manual 65-2341 CT-7 Series Toxic Gas Transmitter Operator s Manual Part Number: 71-0424 Revision: P1 Released: 6/8/17 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction

More information

Ultrasonic Humidification System Model 5462 & Model 5472 OPERATING MANUAL

Ultrasonic Humidification System Model 5462 & Model 5472 OPERATING MANUAL Ultrasonic Humidification System Model 5462 & Model 5472 OPERATING MANUAL 6/30/06 www.electrotechsystems.com TABLE OF CONTENTS 1.0 General Page 01 2.0 Description 01 3.0 Set-Up 01 4.0 Model 5462 03 4.2

More information

Clean Air Optima HEPA UV ionizer air purifier CA-508

Clean Air Optima HEPA UV ionizer air purifier CA-508 Clean Air Optima HEPA UV ionizer air purifier CA-508 Manual Read these instructions carefully before you start using the air purifier and keep the manual for later usage. Copyright : Clean Air Optima 1

More information

TrueClean ToteCleaner

TrueClean ToteCleaner TrueClean ToteCleaner Installation, Operation, and Maintenance Manual www.trueclean.us TrueClean Check Valve Table of Contents Introduction.... 1 About... 1 Warranty... 2 Safety... 3 Important Safety Information....

More information

INSTALLATION INSTRUCTIONS (please become familiar with these instructions before installing).

INSTALLATION INSTRUCTIONS (please become familiar with these instructions before installing). Customer Support: Tel: 1-347-492-4014 Email: info@pureeffectfilters.com www.pureeffectfilters.com Thank you, and Congratulations on your purchase of The PureEffect TM ULTRA-WHC Whole-House Water Filtration

More information

ES2-HB. Ordering Information. Dimensions. Specifications. Humidity Sensor Temperature/Humidity Sensor

ES2-HB. Ordering Information. Dimensions. Specifications. Humidity Sensor Temperature/Humidity Sensor Humidity Sensor Temperature/Humidity Sensor ES2-THB CSM ES2-THB_DS_E_3_2 Detect Humidity within an Accuracy of ±3% in a Compact Body. Set Provides Superior Humidity Control and Temperature/ Humidity Control.

More information

WITH OMRON E5CC CONTROL

WITH OMRON E5CC CONTROL LABORATORY OVENS, FAN FORCED & GRAVITY CONVECTION. MODELS: TO-152G & TOH-152G TO-152F, TOH-152F & TO-235F TO-330F & TO-500F WITH OMRON E5CC CONTROL Thermoline Scientific Equipment Pty. Ltd. T/A Thermoline

More information

RTE-5/5B & RTE-9/9B Refrigerated Bath Circulators. NESLAB Manual P/N Installation, Operation and Service Manual

RTE-5/5B & RTE-9/9B Refrigerated Bath Circulators. NESLAB Manual P/N Installation, Operation and Service Manual RTE-5/5B & RTE-9/9B Refrigerated Bath Circulators NESLAB Manual P/N 013529 Installation, Operation and Service Manual RTE-5 & RTE-9 Bath and Circulator Instruction and Operation Manual Table of Contents

More information

AUTOMATIC GRANULAR ICE FLAKER

AUTOMATIC GRANULAR ICE FLAKER AUTOMATIC GRANULAR ICE FLAKER INSTRUCTIONS AND WARNINGS 24480 rev. 01 It is strictly forbidden to reproduce this instruction manual or any part thereof. Dear Customer, Congratulations on choosing a

More information

Installation, Operating and Servicing Instructions

Installation, Operating and Servicing Instructions Installation, Operating and Servicing Instructions Wall Mounted Water Boiler WMB3F/B,WMB3F/W Please make a note of your product details for future use: Date Purchased: Model Number: Serial Number: Dealer:

More information

HOT & COLD WATER PURIFIER CLIMA MESA / PISO

HOT & COLD WATER PURIFIER CLIMA MESA / PISO HOT & COLD WATER PURIFIER CLIMA MESA / PISO MESA PISO CONTENTS: Instructions related with power ------------------------------------------------------- 2 Installation Instruction -------------------------------------------------------------------

More information

B 353 COFFEE BREWING MACHINES MACHINES A CAFE KAFFEEBRUEHMASCHINEN KOFFIEZETAPPARATEN

B 353 COFFEE BREWING MACHINES MACHINES A CAFE KAFFEEBRUEHMASCHINEN KOFFIEZETAPPARATEN Chiller A B C1 C2 C3 C4 C D E L H G F J M I K N Fig. 1 Principal components and Control organs E=154 W=78 115 572 755 200 35 482 415 245 E/W = 61 Fig. 2 Dimensions Bravilor Bonamat GB Operating instructions

More information

GENANO AIR PURIFICATION TECHNOLOGY

GENANO AIR PURIFICATION TECHNOLOGY GENANO AIR PURIFICATION TECHNOLOGY GENANO 310 AIR PURIFICATION DEVICE OPERATING INSTRUCTION This manual is for the Genano 310 log. Version. GENANO LTD Kimmeltie 3 FI- 02110 Espoo FINLAND Tel + 358 (0)

More information

EASYCLEAR 3000 YEAR GUARANTEE

EASYCLEAR 3000 YEAR GUARANTEE 2823 3003 EASYCLEAR 3000? Hozelock Ltd. Midpoint Park, Birmingham B76 1AB. England Tel: +44 (0) 121 313 1122 info@hozelock.com www.hozelock.com 33854-005 3 YEAR GUARANTEE 1 G D B C F A E I 2 J 3 K H 4A

More information

Installation & Operating Instructions

Installation & Operating Instructions PREMIUM Installation & Operating Instructions AQP-20SIL Signature Filtered Water Cooler Premium Selection Congratulations on your purchase of a Aquaport Signature Filtered Water Cooler This unit is easy

More information

Models & Replacement Filter Information

Models & Replacement Filter Information HEPA Air Purifier Models 30729 & 30730 Model 30729 30730 Replacement Filter Information Pre-Filter HEPA Filter 30901 30962 30901 30962 41781-01 12/08/2005 2005 Hunter Fan Company 1 Table of Contents Safety

More information

Appearance and sensing characteristic Specification (temperature range) Model 10 to 70 C ES1B. 2 mm 20 mm 40 mm 60 mm. 40 dia. 60 dia.

Appearance and sensing characteristic Specification (temperature range) Model 10 to 70 C ES1B. 2 mm 20 mm 40 mm 60 mm. 40 dia. 60 dia. Infrared Thermosensor CSM DS_E_3_2 Achieve Low-cost Measurements with an Infrared Thermosensor. The has an electromotive output as high as that of a thermocouple, thus connecting directly to the thermocouple

More information

T100 Air Purifier USER MANUAL

T100 Air Purifier USER MANUAL T100 Air Purifier USER MANUAL www.alencorp.com 1-800-630-2396 1 2 Helping People Live Better Table of Contents Air Purifier Safety Using Your Air Purifier Operating Your Unit Identification Of Parts Know

More information

Multi-Tech S1000 / S1000-Ultra Air Purifier

Multi-Tech S1000 / S1000-Ultra Air Purifier Multi-Tech S1000 / S1000-Ultra Air Purifier Instruction Manual (Please read before using air purifier) Surround Air Division Indoor Purification Systems, Inc. 334 Marshall Way, Suite C Layton, Utah 84041

More information

Ultrasonic Cool Mist Humidifier

Ultrasonic Cool Mist Humidifier Model No.: LV450CH Ultrasonic Cool Mist Humidifier Questions or or Concerns? Please contact us Monday Mon-Fri 9:00AM-5:00PM - Friday am - PST 5:00 pm PST at support@levoit.com (888) 726-8520 email or at

More information

AQUA ONE. Reverse Osmosis System Model ARO5000. Installation and Service Guide PLEASE READ THIS MANUAL CAREFULLY BEFORE ATTEMPTING INSTALLATION

AQUA ONE. Reverse Osmosis System Model ARO5000. Installation and Service Guide PLEASE READ THIS MANUAL CAREFULLY BEFORE ATTEMPTING INSTALLATION AQUA ONE Reverse Osmosis System Model ARO5000 Congratulations on your purchase of the most sophisticated water purifier system available! Installation and Service Guide PLEASE READ THIS MANUAL CAREFULLY

More information

aquamanta the experts in aquatic technology

aquamanta the experts in aquatic technology HFX 70/100/150/200 TM aquamanta the experts in aquatic technology Coldwater Tropical Marine 2+1 year guarantee Instruction Manual Instruction version: 05/06/2012 Technical Information MODEL: HFX 70 HFX

More information

3 Scotch Mini-Weld Adhesive System User s Manual

3 Scotch Mini-Weld Adhesive System User s Manual 3 Scotch Mini-Weld Adhesive System User s Manual The Scotch Mini-Weld Adhesive System Includes The items shown below: Contents of User s Manual Ordering Replacement Parts... 1 Important Safety Information...

More information

INSTRUCTION MANUAL (English version) Water still

INSTRUCTION MANUAL (English version) Water still INSTRUCTION MANUAL (English version) Water still Model : Brand : W4000 Merit Merit Water Still W4000 & W4000/EURO Cooling water inlet Condenser Constant level control Boiler Thermostat reset buttons To

More information

PRF-RO Pentair Reverse Osmosis System Installation and Maintenance Manual SAFETY GUIDES THE BASIC REVERSE OSMOSIS SYSTEM. Tools and Materials Required

PRF-RO Pentair Reverse Osmosis System Installation and Maintenance Manual SAFETY GUIDES THE BASIC REVERSE OSMOSIS SYSTEM. Tools and Materials Required PRF-RO Pentair Reverse Osmosis System Installation and Maintenance Manual SAFETY GUIDES Read and follow all steps and guides carefully before installing and using your reverse osmosis system. Do not use

More information

MODULAR AUTOMATIC GRANULAR ICE FLAKER

MODULAR AUTOMATIC GRANULAR ICE FLAKER MODULAR AUTOMATIC GRANULAR ICE FLAKER INSTRUCTIONS AND WARNINGS 24479 rev. 03 It is strictly forbidden to reproduce this instruction manual or any part thereof. Dear Customer, Congratulations on having

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700 INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model ABM - 700 M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: 800-831-1525 Mazeppa, MN 55956 Fax : (507)843-4953

More information

User s Manual. Purity In Each Drop. Latest RO+UV technology Now at your home

User s Manual. Purity In Each Drop. Latest RO+UV technology Now at your home User s Manual Purity In Each Drop Latest RO+UV technology Now at your home Table of Contents 1 Safety Information...1 2 Water Purification Process...2 3 Reverse Osmosis Process...3 4 UV Process...3 5 Specifications:

More information

IMPORTANT! ank You! Model: Twist-In 100 GPD, 3-Stage RO Unit. Installation, Operation & Maintenance Guide

IMPORTANT! ank You! Model: Twist-In 100 GPD, 3-Stage RO Unit. Installation, Operation & Maintenance Guide Twist-In 00 GPD, 3-Stage RO Unit WARNING: Please read carefully before proceeding with installation. Your failure to follow any attached instructions and operating parameters may lead to the product s

More information

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02

RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: RK Revision: P1 Released: 7/18/02 35-3000RK-OXY Oxygen Sample-Draw Detector Operator s Manual Part Number: 71-0071RK Revision: P1 Released: 7/18/02 RKI Instruments Inc. 1855 Whipple Road Hayward, CA 94544 PH: 800-754-5165 Fax: 510-441-5650

More information