Electron detectors. The most used signals in a scanning electron microscope in standard conditions are the electrons and in particular the SE.
|
|
- Arron Farmer
- 6 years ago
- Views:
Transcription
1 Electron detectors An appropriate detector must be employed to collect and convert the radiation of interest, that leaves a specimen, into an electrical signal for manipulation to create an SEM image. The most used signals in a scanning electron microscope in standard conditions are the electrons and in particular the SE. Considering electron imaging, BSE and SE leave the specimen with drastically different properties in term of energy, fraction relative to the beam, and directionality of emission. Each specific type of radiation potentially carries information on different aspects of the specimen characteristics. This information can be contained in the number of BSE and SE, their energy distribution, and their emission directions, or a combination of all three factors.
2 General characteristics of a detector The main characteristics of an electron detector are: the detector position with respect to the beam and the specimen the dimensions of the detector sensitive area (described by the solid angle Ω) the efficiency in converting a signal incident onto the detector sensitive area in useful signal. In general, the efficiency is not constant, but it depends on the electron energy the amplification properties of the signal These widely different energy characteristics of BSE and SE present a considerable challenge to the design of a detector that can be used for both signals. Alternatively, the differences permit the design of detectors that are selective for one of the signals and thus selective for specific specimen properties.
3 Everhart-Thornley detector The electron detector most commonly used in scanning electron microscopy is the combined secondary/backscattered electron detector named Everhart-Thornley (E-T). It is so popular that it is extremely rare to find a conventional SEM without one. The rise of the SEM as a tool of broad use to the scientific community is due in considerable measure to the performance and utility of the E-T detector (large solid angle of collection/efficiency, high amplifier gain, low noise, low-maintenance performance). Because of its efficient collection of SE, the E-T detector is often mistakenly considered only a SE detector..
4 The E-T detector operates in the following manner: when an energetic electron ( 10 kev energy) strikes the scintillator material (S), light is emitted. Everhart-Thornley detector Light is conducted by total internal reflection in a light guide (LG) (a solid plastic or glass rod) to the photocatode of a photomultiplier (PM). At the photocatode the photons are converted back into electrons which are accelerated onto the successive electrodes of the photomultiplier, producing an ever-increasing cascade of electrons until the final collector is reached. The typical gain of a photomultiplier is
5 A large fraction of the BSE, that originates from incident beam with energies from 10 to 30 kev, carry sufficient energy to directly excite the scintillator, even in the absence of postspecimen acceleration. By applying a large positive potential (10-12 kv) to a thin metal coating on the face of the scintillator (S), the low-energy SE will be accelerated to a sufficient energy to generate light in the scintillator. To protect the beam from unwanted deflection and distortion by this large potential in the specimen chamber, the scintillator is surrounded by a Faraday cage (F) which is insulated from the scintillator bias. To collect the low-energy SE with higher energy efficiency than simply collecting the fraction defined by the line-of-sight solid angle, a sparate bias potential is applied to the Faraday cage, typically in range (-50 V, +250 V). This range from negative to positive provides the possibility of completely rejecting SE (-50 V) or efficiently collecting SE (+250 V).
6 When the Faraday cage bias is positive, the BSE directed towards the detector are detected; but they represent only a small fraction of all the detected electrons. So their contribution is negligible. Thanks to the presence of the Faraday cage, the Everhart- Thornley detector is a device very efficient to detect SE and for flat samples it is possible to detect nearly all the secondary electrons If the Faraday cage bias is negative, the SE are rejected and the detector collect only the BSE. Obviously, without an electric field that directs the electrons towards the scintillator, only the BSE which have the right direction are revealed. So the collecting efficiency for BSE is very low and to detect BSE other detectors are used.
7 Through the lens TTL detector The high-performance field-emission-gun SEM is equipped with a snorkel lens which produces a strong objective lens magnetic field that is projected into the specimen chamber to reach the specimen plane. Magnetic field projected out of lens Snorkel Lens TTL ET This contrasts with the pinhole lens of a conventional SEM in which the lens magnetic field is contained within the bore of the lens so that the specimen resides in a field-free region. One major consequence of the strong magnetic field is to trap with high efficiency those SE emitted from the specimen. The SE spiral along the magnetic field lines and pass up through the lens bore. In this configuration SE are detected by the TTL detector.
8 The upper and lower detectors have a different viewpoint of the specimen and so they see the specimen differently In-lens (TTL) detector gives a shadow free image with ultra-high topographical resolution. Upper SE Detector Lower SE Detector
9 Passive scintillator BSE detectors They operate on the principle that, with an incident beam of 10 kev or more, the majority of the BSE carry sufficient energy to excite the scintillator even in the absence of a postspecimen acceleration. Without such active acceleration, the SE have no effect on the scintillator, so the signal from an unbiased or passive scintillator detector will consist only of contributions from BSE. The elimination of the bias also has the benefit that the detector potential will not disturb the beam, so the detector can be placed close to the specimen for mor efficient collection. By making the scintillator of the same material as the light guide, designs that collect over much larger solid angles are possible. The detector is placed above the specimen and a hole drilled trough the material permits access for the beam. In this configuration, the detector surrounds the specimen nearly simmetrically, so the signal is integrated in all directions, nearly eliminating sensitivity to trajectory effects.
10 Solid state diode detectors They operate on the principle of electron-hole production induced in a semiconductor by energetic electrons. A solid state diode detector (p-n junction) has the form of a flat, thin wafer (typically several millimiters thick) which can be obtained in a variety of shape and size, from small square to large annular detectors. The thinness of the detector permits it to be placed in close proximity to the specimen, which combined with the large area possible, provides a large solid angle for high geometric efficiency. In general, it is mounted under the objective lens, without interference with the normal operation of the instrument. The main drawback is the long response time, with respect to the other detectors, so it doesn t permit high scan rates.
Arquitetura do MEV [6]
[6] http://www4.nau.edu/microanalysis/microprobe-sem/instrumentation.html 1> Elétrons detectores sinais imagem 2> Emissão de elétrons secundários (SE): energia entre 5-50 ev livre caminho médio - L = A
More informationTESCAN MIRA FESEM Presentation Global Supplier of Scanning Electron Microscopes
TESCAN MIRA FESEM Presentation Global Supplier of Scanning Electron Microscopes TESCAN USA 508 Thomson Park Drive Cranberry TWP, PA 16066 Outline About TESCAN Product Portfolio MIRA FESEM Overview Electron
More informationToday s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21
Today s Outline - September 12, 2016 C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016 1 / 21 Today s Outline - September 12, 2016 Detectors C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016
More informationFebruary 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.
PHY357 Lecture 11 Experimental Methods Accelerators Particle Interactions Particle Detectors Full experiment (eg. ATLAS) Introduction! Several different general classes of particle detectors (sensors)!
More informationAsB Detector. Showing Material Contrast and Topographical Information with the AsB Detector. Upgrade Info
AsB Detector Showing Material Contrast and Topographical Information with the AsB Detector Upgrade Info AsB Detector Showing Material Contrast and Topographical Information with the AsB Detector Contact:
More informationULTRA Series. We make it visible. Field Emission - Scanning Electron Microscopes (FE-SEM) from Carl Zeiss
Field Emission - Scanning Electron Microscopes (FE-SEM) from Carl Zeiss ULTRA Series Ultra High Resolution FE-SEM for Nano-scale Compositional Analysis We make it visible. Carl Zeiss Microscopy ULTRA Series
More informationLECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry
LECTURE 11 Dr. Teresa D. Golden University of North Texas Department of Chemistry C. Detectors Properties of Detectors -quantum counting efficiency -linearity -energy proportionality -resolution C. Detectors
More informationExperimental Particle Physics PHYS6011 Joel Goldstein, RAL
Experimental Particle Physics PHYS6011 Joel Goldstein, RAL 1. Introduction & Accelerators 2. Particle Interactions and Detectors (1/2) 3. Collider Experiments 4. Data Analysis Charged Particle Detectors
More informationADVANCED SILICON DETECTORS FOR ELECTRON IMAGING & COUNTING
ADVANCED SILICON DETECTORS FOR ELECTRON IMAGING & COUNTING About PNDetector l PNDetector s Products l Backscattered l STEM Detector Diodes l The pnccd (S)TEM Camera ABOUT PNDetector was founded in 2007
More informationUltraviolet radiation detector to obtain the rate of particles at different heights
Journal of Physics: Conference Series PAPER OPEN ACCESS Ultraviolet radiation detector to obtain the rate of particles at different heights To cite this article: E Ponce et al 2016 J. Phys.: Conf. Ser.
More informationHitachi Tabletop Microscope TM4000 / TM4000 Plus
Hitachi Tabletop Microscope TM4000 / TM4000 Plus TM4000 Series Notice: For correct operation, follow the instruction manual when using the instrument. Specifications in this catalog are subject to change
More informationCrossBeam Workstations (FIB-SEM) AURIGA Series Information Beyond Resolution. We make it visible.
CrossBeam Workstations (FIB-SEM) AURIGA Series Information Beyond Resolution We make it visible. Carl Zeiss Microscopy Electron and Ion Beam Microscopes More than 160 years of experience in optics has
More informationInstrumental technique (MCP detectors) - Kamalesh ( )
Instrumental technique (MCP detectors) - Kamalesh (02-03-2012) Schematic of a MCP detector MCP as an assembly of millions miniature SEMs Schematic of a MCP detector Microchannel Plates have a combination
More information4-Channel-aSTEM Detector. Detecting Transmitted Electrons in an FESEM with the new Annular STEM Detector. Upgrade Info
4-Channel-aSTEM Detector Detecting Transmitted Electrons in an FESEM with the new Annular STEM Detector Upgrade Info 4-Channel-aSTEM Detector Detecting Transmitted Electrons in an FESEM with the new Annular
More informationBACK SCATTER ELECTRON (BSE) Rev. 0313
Back-scatter Electron Secondary Electron BACK SCATTER ELECTRON (BSE) Rev. 0313 Contents BSE Introduction (3) Mounting the Detector (4) Load the Sample (5) Back-Scatter Settings (6-7) Optimize Image (8)
More informationCustomized MCPs for Analytical Instruments. An MCP for every application
Customized MCPs for Analytical Instruments An MCP for every application What are Microchannel Plates? (MCPs) Microchannel plates (MCPs) are solid state detectors consisting of parallel arrays of tiny single
More informationSILICON DRIFT DETECTORS
SILICON DRIFT DETECTORS PRODUCT BROCHURE 2 JUNE 2018 About PNDetector PNDetector is well-known as a manufacturer of high performance radiation detectors for microanalysis, quality assurance and material
More informationAdvanced Electron Detector and Acquisition System
Combined PNDetector & point electronic Acquisition & Scanning System Advanced Electron Detector and Acquisition System for Electron Microscopy Overview Company Information...3 Collaboration of point electronic
More informationTIMS TIMS. Thermal Ionisation Mass Spectrometry.
TIMS Thermal Ionisation Mass Spectrometry www.nu-ins.com Incorporating our field-proven Zoom lens multi-collector technology, the TIMS improves the versatility and overall performance of this long established
More informationGaseous detectors. 1
Gaseous detectors Very important class of detectors with many applications charged particle, x- & gamma rays, visible light detection applications in many areas of research, & commercial particle & nuclear
More informationChemistry Instrumental Analysis Lecture 4. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 4 Atomic Spectroscopy Atomic Spectroscopy UV to IR Optical spectroscopy are based on: absorption fluorescence phosphorescence scattering emission chemiluminescence
More informationMeasurement system for detection of Radon in building
Measurement system for detection of Radon in building Outline Properties of Radon Detection of ionised particles Measurement system for radon Properties of Radon Radon is a gaseous isotope ( 222 Rn) with
More informationFEATURE ARTICLE. The PLCA-800 Series Inline Particle Sensor for Chemical Solutions. Yoshihito Yuhara. Abstract
FEATURE ARTICLE Technical Reports The PLCA-800 Series Inline Particle Sensor for Chemical Solutions Yoshihito Yuhara Front end process Sensor module Wafer preparation process Film formation Photo lithography
More information高等食品分析 (Advanced Food Analysis) II. INSTRUMENTS FOR OPTICAL SPECTROSCOPY *Instruments for optical spectroscopy: Optical spectroscopic methods are
*Instruments for optical spectroscopy: Optical spectroscopic methods are based on 6 phenomena: 1) Absorption, 2) Fluorescence, 3) Phosphorescence, 4) Scattering, 5) Emission and 6) Chemiluminescence. Typical
More informationMission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design and deliver low light detection solutions which solve customer probl
份有限公司 wwwom 先锋科技股份有限公司 份有限公司 wwwom 先锋科技股份有限公司 www Radiation detection solutions for miniaturized pager and handheld markets sensl: light detection and measurement October - 2009 www.sensl.com 1 Mission
More informationObservation in UniVac
Advantages! Observation of the sample in natural state! Minimize sample destruction! Controlled evaporation! Minimize charging! LVSTD = Detection of true secondary e- Fig.: Potato starch in a tuber; low
More informationMicromegas in a Bulk
1 Micromegas in a Bulk I. Giomataris 1, R. De Oliveira 2, S. Andriamonje 1, S. Aune 1, G. Charpak 2, P. Colas 1, A. Giganon 1, Ph. Rebourgeard 1, P. Salin 3 1 CEA-DSM/Saclay-France 2 CERN, Geneva, Switzerland
More informationA GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2. Abstract
A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2 1 Dipartimento Interateneo di Fisica \M. Merlin" and Sezione INFN, Bari, Italy 2 CERN and Institute for Chemical
More informationEnvironmental Radiation Monitoring System
Environmental Radiation Monitoring System Toshihiro Takagi Osamu Kimura Atsushi Minagoshi 1. Introduction In order to ascertain conditions near the perimeter of a supervised area at a nuclear power plant
More informationNON-DESTRUCTIVE 2-D BEAM PROFILE MONITOR USING GAS SHEET IN J-PARC LINAC -overview & hardware construction-
NON-DESTRUCTIVE 2-D BEAM PROFILE MONITOR USING GAS SHEET IN J-PARC LINAC -overview & hardware construction- J. Kamiya, N. Ogiwara, Y. Hikichi, Y. Yamada (Vac. Gr.) K. Okabe, A. Miura, K. Moriya (Mon. Gr.)
More informationTHERMAL DIAGNOSIS OF MV SWITCHBOARDS: A COST-EFFECTIVE, DEPENDABLE SOLUTION BASED ON AN OPTICAL SENSOR
THERMAL DIAGNOSIS OF MV SWITCHBOARDS: A COST-EFFECTIVE, DEPENDABLE SOLUTION BASED ON AN OPTICAL SENSOR Christian PETIT Anticipation manager Schneider Electric - Network protection and control unit -Marketing
More informationWarner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax
OW Series, Rev. 120802 OW Series Objective Warmer Warner Instruments 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax OW Series, Rev. 120802 Table of Contents
More informationMission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design and deliver low light detection solutions which solve customer probl
份有限公司 wwwom 先锋科技股份有限公司 www 份有限公司 wwwom 先锋科技股份有限公司 Solutions for Nuclear Medicine sensl: light detection and measurement Mission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design
More informationDiamond Windows & MCP Detectors for Synchrotron Applications
Torr Scientific Ltd. Specialists in UHV, Optical and X-ray Components Diamond Windows & MCP Detectors for Synchrotron Applications www.torrscientific.co.uk Torr Scientific Ltd. Specialists in UHV, Optical
More informationSPECTRAL INSTRUMENTATION
SPECTRAL INSTRUMENTATION The typical setup of an IR experiment includes a light source, a sample, (a microscope), the spectral apparatus, a detector and a computer. IR experiment A light source emits polychromatic
More informationSemiconductor Radiation Detectors
Gerhard Lutz Semiconductor Radiation Detectors Device Physics With 167 Figures and 11 Tables Contents 1. Introduction 1 Part I: Semiconductor Physics 5 2. Semiconductors 7 2.1 Crystal Structure 7 2.2 Energy
More informationdewar or an electrically powered cooler. The sensitive detector surfaces are thus protected from moisture and condensable contaminants.
Germanium Detectors The Company Canberra Industries is the world's largest manufacturer of nuclear radiation detection and analysis systems. In order to better serve our customers worldwide, Canberra operates
More informationPyroelectric Infrared Sensor PIR
Pyroelectric Infrared Sensor PIR Infrared radiation is the electromagnetic waves in the wavelength region longer than the visible light wavelengths, lying from 0.75 μm to 1000 μm. The wavelength region
More informationChemistry Instrumental Analysis Lecture 14. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 14 Infrared Spectroscopy IR Spectroscopy FTIR is Fourier-Transform Infrared Spectroscopy It is a chemically specific technique used to identify chemical compounds
More informationSurface Concept Product Catalog 2014
Surface Concept Product Catalog 2014 Delayline Detectors MCP Detectors Time-to-Digital Converters Distributor of Surface Concept Products in the United States is Pulse Processing Electronics 727-532-6144
More informationSOLAR CELL PRODUCTION REQUIRES EFFECTIVE METROLOGY
SOLAR CELL PRODUCTION REQUIRES EFFECTIVE METROLOGY 6/12/2010 Recent IR Thermography Developments Can Help Agenda: Infrared Cameras for Solar Cell Inspection Survey of current solar cell test methods New
More informationParticle Detectors. Summer Student Lectures 2007 Werner Riegler, CERN, History of Instrumentation History of Particle Physics
Particle Detectors Summer Student Lectures 2007 Werner Riegler, CERN, werner.riegler@cern.ch History of Instrumentation History of Particle Physics The Real World of Particles Interaction of Particles
More informationWarner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax
OW Series, Rev. 04.01.27 OW Series Objective Warmers 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax OW Series, Rev. 04.01.27 Table of Contents DESCRIPTION...3
More informationPassive Solar Home Design
Passive Solar Home Design Your home's windows, walls, and floors can be designed to collect, store, and distribute solar energy in the form of heat in the winter and reject solar heat in the summer. This
More informationXGLAB. X and Gamma Ray Electronics. Looking for new frontiers...
XGLAB X and Gamma Ray Electronics Looking for new frontiers... Research And Development XGLab, spin off company of Politecnico di Milano, the leading Italian technical University, holds a leadership and
More informationTM3030Plus August 4, 2014 launched
TM3030Plus Tabletop Microscope More than 2,700 units cumulative shipments! TM3030Plus is now added as the TM3030 series! TM3030Plus August 4, 2014 launched Features 1. No dedicated operators needed 2.
More informationPortable Survey Instruments NISP-RP-01
NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection NISP-RP-01 This is an industry document for standardizing radiation protection processes used by supplemental radiation protection technicians.
More informationNEXTorr Pumps. making innovation happen,together
NEXTorr Pumps group making innovation happen,together NEXTorr NEXTorr: the next step in pumping technology Is there a solution for compact, light and high performing UHV pumps? The quick answer is NEXTorr.
More informationIdentify the following features and specifications for the ESP1
LEARNING OBJECTIVES: 2.17.01 List the factors which affect an RCT's selection of a portable contamination monitoring instrument, and identify appropriate instruments for contamination monitoring. (Also
More informationMeasurement of Light. Photosensors
Measurement of Light Photosensors Detection of light is a basic need for everything from devices to plants and animals. In the case of animals, light detection systems are very highly specialized, and
More informationTable of Contents. Check the LCLS Project website to verify 2 of 12 that this is the correct version prior to use.
Table of Contents Summary...3 Mechanical Design Specifications...3 Vacuum...3 Materials...5 Support Requirements...5 Cabling...6 Light Path...6 BLM Synchrotron Radiation Assembly (BL11)...7 Other Requirements...9
More informationFreiburger Materialforschungszentrum SG Material Characterization & Detector Technology. Albert-Ludwigs-Universität Freiburg
Medipix 2: Processing and measurements of GaAs pixel detectors A. Fauler, A. Zwerger Outline SG Material Characterization & Detector Technology Albert-Ludwigs-Universität Freiburg www.fmf.uni-freiburg.de
More informationThe Growing Viability of UV LED for Wide Web Coating Applications. Tuesday, October 11, 2016
The Growing Viability of UV LED for Wide Web Coating Applications Tuesday, October 11, 2016 Page 2 Phoseon Technology Company Based in Portland, Oregon UV LED Pioneer since 2002 100% Focused on UV LED
More informationCOUNTING ROOM EQUIPMENT RCT STUDY GUIDE
LEARNING OBJECTIVES: 2.19.01 Describe the features and specifications for commonly used laboratory counters or scalers: a. Detector type b. Detector shielding c. Detector window d. Types of radiation detected
More informationMedical Lasers and Laser Safety
Medical Lasers and Laser Safety Key Points Lasers & Laser Safety: - Know the three unique characteristics of laser light - Be familiar with the different medical laser technologies - solid, gas, excimer,
More informationDEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR
DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR Katsuhito Ito, Ryohei Yanagishima, Daisuke Inui*, Tomoya Minagawa Fuji Electric Systems Co., Ltd. Fuji-machi
More informationOptical Sensor (2A) Optical Sensor Type Optical Sensor Characteristics. Young Won Lim 9/12/09
Optical Sensor (2A) Optical Sensor Type Optical Sensor Characteristics Copyright (c) 2009 Young W. Lim. Permission is granted to copy, distribute and/or modify this document under the terms of the GNU
More informationSICK AG WHITEPAPER. safehddm TM THE NEW SCANNING TECHNOLOGY FOR SAFETY LASER SCANNERS
SICK AG WHITEPAPER safehddm TM THE NEW SCANNING TECHNOLOGY FOR SAFETY LASER SCANNERS 2016-07 AUTHOR Patrick Hochleitner Product Manager Industrial Safety Systems, Marketing & Sales at SICK AG in Waldkirch/Deutschland
More informationAntoine Drouart DSM/DAPNIA/SPhN. Köln, 5 th of December,
Köln, 5 th of December, 2007 1 Basic CATS/BTD mechanics Köln, 5 th of December, 2007 2 Beam Tracking Today -Detector able to cope with a few 10 5 pps -Spatial resolution (FWHM) : 0.7mm -Time resolution
More informationAPPLICATION FOR UNITED STATES PATENT. Shieldless Detector with Two-Dimensional Directionality INVENTOR:: David E. Newman DATE:: November 9, 2018
APPLICATION FOR UNITED STATES PATENT TITLE:: Shieldless Detector with Two-Dimensional Directionality INVENTOR:: David E. Newman DATE:: November 9, 2018 ABSTRACT Disclosed is a directional gamma ray or
More informationAn Ambient-Light-Ignoring Infrared Active Motion Detector
An Ambient-Light-Ignoring Infrared Active Motion Detector Design Team Members: Namita Ahuja Robert Mitchell Karl Schrader Michael Perry Joe Ha Table of Contents: Project Objectives Background Information
More informationPyroelectric crystal based neutron source and neutron detector
Dipartimento di fisica G. Occhialini Università Degli Studi di Milano-Bicocca Seminar for summer school on Neutron Detectors and Related Applications-NDRA 2016 June 29th July 2nd, 2016, Riva del Garda,
More informationCompetency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery
Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery Trainee: TRAINEE Module: Radiotherapy Physics Competency:
More informationLARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S)
LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S) Ganibu Dambis 26, P.O.Box 33 Riga, LV-1005, LATVIA Tel. (+371)6738-3947 Fax:(+371)6738-2620 office@bsi.lv www.bsi.lv CONTENTS INTRODUCTION...
More informationNUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use
NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection Level 3 Information Use NISP-RP-001 Industry Approval Date: September 14, 2018 This is an industry document for standardizing radiation protection
More informationMOXTEK MULTI-LAYER OPTICS AND X-RAY TECHNOLOGY
MOXTEK MULTI-LAYER OPTICS AND X-RAY TECHNOLOGY Moxtek is a leading supplier of advanced nano-optical and x-ray components used in display electronics, imaging, and analytical instrumentation. Moxtek provides
More informationFireye FLAME SCANNERS
UV7A4 SC-107 APRIL 3, 2013 Fireye SCANNERS UV7SC UV7R4 UV Non Self-Checking Scanner Models: UV7A4, UV7R4 UV Self-Check Scanner Model: UV7SC DESCRIPTION The non self-check UV7A4, UV7R4 and self-check UV7SC
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60462 Edition 2.0 2010-07 Nuclear instrumentation Photomultiplier tubes for scintillation counting Test procedures INTERNATIONAL ELECTROTECHNICAL COMMISSION PRICE CODE T ICS
More informationClass 1 laser beam sensor safe for your eyes
113 Sensor SERIES Related Information General terms and conditions... F-7 About laser beam... P.199~ Sensor selection guide... P.1~ General precautions... P.11 PHOTO PHOTO PARTICUR MEASURE ITY panasonic.net/id/pidsx/global
More informationStatus Report about the TPC Detector and Module at CEPC
Status Report about the TPC Detector and Module at CEPC Huirong QI On behalf of CEPC TPC subgroup, IHEP, CAS 2016.01.19, IAS, HK Content Goals and options Some considerations Preliminary simulation Hybrid
More informationSensing the World. When you have finished this lecture you should be able to:
Free Course: How to build a real robot Lecture 3 Sensing the World MUSES_SECRET: L3, Adding the Brain: ORF-RE 2011-2012 Project - Dr. Alaa PAMI Khamis, Research IEEE Group RAS University Egypt Chapter
More informationAging measurements on triple-gem detectors operated with CF 4 based gas mixtures
IEEE, Roma, October 2004 Aging measurements on triple-gem detectors operated with CF 4 based gas mixtures M. Alfonsi 1, G. Bencivenni 1, W. Bonivento 2,A.Cardini 2, P. de Simone 1, F.Murtas 1, D. Pinci
More informationApplication Note. Which characterization method is best for your application? The Future of Thermal Imaging is Here!!!
Microsanj LLC 3287 Kifer Road Santa Clara, CA 95051 Application Note Comparing Thermoreflectance (TTI), Infrared (IR), Near Infrared Emission (EMMI), and Optical Beam Induced Resistance Change (OBIRCH)
More informationLaser Distance Sensor Type M Analog-Output
Laser Distance Sensor Type M Analog-Output Operating Manual Version 2.3, May 2009 ELAG Elektronik AG l Stegackerstrasse 14 l CH-8409 Winterthur Contents 1. Safety 1.1 Laser safety 1.2 Electrical safety
More informationI Seminario Nazionale Rivelatori Innovativi
I Seminario Nazionale Rivelatori Innovativi Monday 30 November 2009 - Friday 04 December 2009 INFN - Laboratori Nazionali di Frascati Programme I Seminario Nazionale Rivelatori Innovativi / Programme Wednesday
More information4.8 FARADAY CUP ASSEMBLY OPTION
4.8 FARADAY CUP ASSEMBLY OPTION The Faraday cup assembly, mounted on the front end of the Electron Gun, is positioned by either a rotary feedthrough or a pneumatic actuator assembly, located on one of
More informationCCD and CID solid-state detectors
CCD and CID solid-state detectors Technical Overview 5110 ICP-OES Introduction Many of today s ICP-OES instruments have progressed from traditional photomultiplier tube (PMT) detection systems to solid-state
More informationXEUS mirror calibration:
XEUS mirror calibration: The PANTER X-ray Test Facility: accommodation of optics with long focal lengths etc., Wolfgang Burkert, Gisela Hartner, Bernd Budau XEUS Workshop, ESA, ESTEC, Noordwijk, 05-06
More informationThe design of the human body infrared thermometer
The design of the human body infrared thermometer Chenjie Zhang Changchun University of Science and Technology, Changchun 130022, China Abstract In order to overcome the main shortcoming of the traditional
More informationClass 1 laser beam sensor safe for your eyes
13 Sensor SERIES Related Information General terms and conditions... F-17 About laser beam... P.13~ Sensor selection guide... P.967~ General precautions... P.1 PHOTO PHOTO Conforming to EMC Directive Conforming
More informationGEM detectors activity at the Laboratori Nazionali di Frascati of INFN
GEM detectors activity at the Laboratori Nazionali di Frascati of INFN M. Alfonsi', G. Bencivenni, P. de Simone, D. Domenici, F. Murtas, M. Pistilli and M. Poli Lener Laboratori Nazionali di Frascati,
More informationIEC PROJETOS EM VOTAÇÃO SETEMBRO E OUTUBRO. IEC Ed.1: Railway applications - Energy mesurement on board trains - Part 2: Energy measuring
IEC PROJETOS EM VOTAÇÃO SETEMBRO E OUTUBRO IEC/TC 9 - Electrical equipment and systems for railways IEC 62888-2 Ed.1: Railway applications - Energy mesurement on board trains - Part 2: Energy measuring
More informationTSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110
TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110 APPLICATION NOTE CC-107 Introduction This purpose of this document is to detail the advanced, state of the art features TSI has incorporated in the design
More informationBuilding and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes
Building and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes Gabrielle Inglis Advisor: Robert Boni Laboratory for Laser Energetics University of Rochester Summer High School Research Program 200
More informationAn FT-NIR Primer. NR800-A-006
Industry: Product: Oils and Gas NR800 Near Infra-red Spectroscopy Like other forms of spectroscopy, light from a lamp (usually called the source ) is passed through a sample and measured by a detector.
More informationThis slide presentation has been prepared for Temple University Health System and is intended only for use by it s faculty, staff and students.
This slide presentation has been prepared for Temple University Health System and is intended only for use by it s faculty, staff and students. All questions pertaining to this presentation should be directed
More informationMITSUBISHI ELECTRIC ANNOUNCES DEVELOPMENT OF 5.6 W 213 nm DEEP ULTRAVIOLET SOLID STATE LASER
FOR IMMEDIATE RELEASE No. 2356 Product Inquiries: Media Contact: Yutaka Kamada Travis Woodward Mitsubishi Electric Corporation Public Relations Department Tel: +81-3-3218-2391 Mitsubishi Electric Corporation
More informationCONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:
LEARNING OBJECTIVES: 2.17.01 List the factors which affects an RCT's selection of a portable contamination monitoring instrument. 2.17.02 Describe the following features and specifications for commonly
More informationQuadruple-beam photoelectric detector
New generation 4-channel selectable frequencies abh-50/100/150/200/250m Quadruple-beam photoelectric detector Highly Reliable Photoelectric Detector ABH series quadruple-beam photoelectric detector with
More informationExercise 1. Introduction to Sensors EXERCISE OBJECTIVE
Exercise 1 Introduction to Sensors EXERCISE OBJECTIVE When you have completed this exercise, you will be introduced to photoelectric, capacitive and inductive sensors; You will also be introduced to the
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60825-1 1993 AMENDMENT 2 2001-01 PUBLICATION GROUPÉE DE SÉCURITÉ GROUP SAFETY PUBLICATION Amendment 2 Safety of laser products Part 1: Equipment classification, requirements
More informationEnergy and Angular Responses of the Criticality Accident Alarm System Using Current-mode-operated Scintillation Detector
Energy and Angular Responses of the Criticality Accident Alarm System Using Current-mode-operated Scintillation Detector N. Tsujimura 1, T. Yoshida 1 and S. Mikami 1 1 Japan Nuclear Cycle Development Institute,
More informationTPC BASED IMP. Limin Duan, Gas Detector Group. Institute of Modern Physics, CAS TPC tracker detector technology mini-workshop
TPC BASED GEM @ IMP Limin Duan, Gas Detector Group Institute of Modern Physics, CAS 2016.09.01 Beijing,IHEP TPC tracker detector technology mini-workshop Outline 2 Gas detector group GEM-readout TPC a
More informationRAGNAR STÅLSKOG AB INSTRUCTION MANUAL HVDC-H HVDC-L VOLTAGE DETECTOR HVDC-H, HVDC-L 15H11
RAGNAR STÅLSKOG AB INSTRUCTION MANUAL HVDC-H HVDC-L VOLTAGE DETECTOR HVDC-H, HVDC-L 15H11 Attention: Read this instruction manual carefully before using the detector. 1. FUNCTION / USE This device is designed
More informationDistributed By: M&M Control Service, INC
APPROVED The UV1A3/6, UV2/UV2A6/UV2C, 45UV2, 45UV3, UV90-3/6/9 Scanners are used with the M-Series, M-Series II, MicroM, FlameWorx, MB-2 and D-Series controls as well as the Fireye FLAME-MONITOR TM, BurnerLogix
More informationRonan Model X90 Series Point Level Monitor
Ronan Model X90 Series Point Level Monitor For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing
More informationBeam Loss Position Monitoring with Optical Fibres at DELTA
Beam Loss Position Monitoring with Optical Fibres at DELTA Frank Rüdiger M. Körfer (DESY) W. Göttmann (HMI Berlin) G. Schmidt (DELTA) K. Wille (DELTA) 24. June 2008 Frank Rüdiger 1 Table of Content 1.
More informationVISAGE ENVIRONMENTAL LIGHTING SOLUTION. INDOOR Lighting SERIES
VISAGE ENVIRONMENTAL LIGHTING SOLUTION INDOOR Lighting SERIES LED Lighting Concept LED, known as Light Emitting Diode, uses solid semiconductor chips as luminescent material. When put the forward voltage
More informationComprehensive applications of the gas flow proportional counters for Radiological Surveillance
Comprehensive applications of the gas flow proportional counters for Radiological Surveillance D.A.R.Babu, Anand Raman, Ashokkumar P and D.N.Sharma Radiation Safety Systems Division, Bhabha Atomic Research
More informationUVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation
15 th International Conference on Environmental Science and Technology Rhodes, Greece, 31 August to 2 September 2017 UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation
More information