DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR
|
|
- Easter Murphy
- 6 years ago
- Views:
Transcription
1 DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR Katsuhito Ito, Ryohei Yanagishima, Daisuke Inui*, Tomoya Minagawa Fuji Electric Systems Co., Ltd. Fuji-machi 1, Hino, Tokyo , Japan Abstract. We developed a large area plastic scintillation detector which fits to the shapes of the measured objects. Effective area of the detector is from about 300 to about 3000 cm 2, and the outer thickness of the detector unit is less than 10 cm for β ray measurements. The detection efficiency is about 25% in 2π geometry with the 60 Co 318 kev β ray source of 100mm x 100mm size set 1cm away from the detector center. The uniformity of detection efficiency over the whole sensitive area is acceptably good within about 20%. KEYWORDS: Plastic scintillation detector, Contamination monitor 1. Introduction The radioactive contamination monitor, such as body surface contamination monitor (portal monitor) and laundry monitor, requires a detector with large sensitive area to monitor the whole objects with various shapes to be measured. The gas flow proportional detector has ever been used usually as the detector for the contamination monitor, however it requires a gas supply system, which has the problem to bring high pressure gas cylinder into a radiation controlled area. We therefore developed a large area plastic scintillation detector without gas supply which fits to the shapes of the measured objects. Effective area of the detector is from about 300 to about 3000 cm 2, and the outer thickness of the detector unit is less than 10 cm for β ray measurements. Table 1 lists the comparison of physical characteristics between gas flow proportional detector and plastic scintillation detector. 2. Detector structure and detection principle We developed a thin (about 0.5 mm thick) and large area (about 300 cm 2 to 3000 cm 2 ) plastic scintillation detector which has a uniform sensitivity for using as body surface contamination monitor (portable monitor), laundry monitor etc. A thin plastic scintillator was used to depress the sensitivity to gamma rays. Figure 1 shows the inner structure of the detector, and Figure 2 shows an example of the outlook of the detector. As shown in Fig.1, the plastic scintillator (Saint-Gobain Industrial Ceramics Inc.) is coupled with air light guide made of aluminum reflector and the scintillation light from β rays is reflected at the light guide like a mirror, and enters into two photomultiplier tubes (ET Enterprises Co. Ltd.) set in parallel. We could realize a thin detector system by using the light guide structure and two photomultipliers setup in parallel to the detector as shown in Fig. 1. In the detector system, the amplifier circuit, high voltage power supply, coincidence circuit and pulse-shaping circuit for photomultipliernoise cut-off are built in, and the pulses of constant pulse heights are outputted without depending on the detector size. Figure 3 shows the block diagram of the detector system. We developed various sizes of plastic scintillators for different purposes. * Presenting author, inui-daisuke@fesys.co.jp 1
2 Table1: Comparison of physical characteristics between gas flow proportional detector and plastic scintillation detector. 1.Background count rate 2.Detection efficiency 3.Measures for the breakage of light protecting film 4.Maintenance 5.Peripheral equipment Gas flow proportional detector Low background counts due to low sensitivity to gamma rays. High efficiency to β rays, but may decrease due to gas leakage through light-protecting film. On the breakage of light protecting film in gas flow counter, the broken film should be replaced by the users as temporary repair. The gas flashing is needed for a few hours after the replacement. Supply of gas is needed. Gas flow is checked in everyday. Gas supply system is required and high voltage power supply is required externally. Plastic scintillation detector Low background counts are realized by taking the coincidence of two output pulses from two photomultipliers. High efficiency to β rays keeps constant for long-term use. On the breakage of light protecting film in plastic scintillator, the broken film should be replaced by the users as temporary repair. It can be used after 30 minutes of the replacement. The routine running cost is not necessary. No gas supply system is needed and high voltage circuit is built in the detector system. Figure 1 : Inner structure of the thin and large area plastic scintillation detector β ray Plastic scintillator Air light guide Aluminum reflector Scintillation light Photomultiplier tube 2 2
3 Figure 2: Outlook of the detector Signal output Light protecting film Low Voltage input Figure 3 : Block diagram of the detector Photomultiplier tube Amplifier Discriminator Printed circuit Light shielding film Plastic scintillator Light guide High-voltage power supply Coincidence Photocoupler Connector Photomultiplier tube Amplifier Discriminator 3. Physical characteristics of the detector system To investigate the physical characteristics of the detector system the detection efficiency and its uniformity over the whole detector surface were measured by using β rays of three different energies, 156 kev of 14 C, 318 kev of 60 Co and 709 kev of 36 Cl at Fuji Electric Systems Co. Ltd. These sources are 100 mm x100 mm in size and were set at the center of the large-area detector 3cm away from the detector sensitive surface. For uniformity investigation, a 36 Cl β ray source was collimated into 25-mm diameter with aluminum plate and set 1cm away from the detector surface. The 60 points at each 5-cm interval in the central detector area of 300 mm by 500 mm were selected and the surface distribution of detector counts was measured. As shown in Fig.4, the selected area of 500 x 300mm was divided into 10 (A to J) segments in 500-mm width and into 6 (1 to 6) segments in 300-mm length. A source collimated into 25mm diameter was set at the center of each divided area and the surface distribution of β- ray detection efficiency was measured. 3
4 Figure 4: 36 Cl β ray source setting position A B C D E F G H I J mm φ collimated source position: (x,y)=(j,1) Figure 5 shows an example of measured results of detection efficiencies to the above three β-ray energies for the detector of 520 mm wide and 310 mm long sensitive dimension. The detection efficiency increases with increasing β-ray energy from 3.6% for 156 kev to 32.5% for 709 kev, and is almost equal to that of gas proportional counter. Figure 5: Detection efficiencies to three β-ray energies for the detector of 520mm width and 310mm length 40 β -ray efficiency (%) Co (318keV) 36 Cl(709keV) C(156keV) β- ray energy (kev) Table 2 shows the surface distribution of detection efficiency for 709 kev β rays of 36 Cl as an example for the detector of 520 mm wide and 310 mm long sensitive dimension. The detection efficiency averaged over the whole surface is 33.4 %, and the deviation ratio from the average is 0.78 at minimum and 1.13 at maximum. The uniformity of detection efficiency over the whole sensitive area is acceptably good within about 20 %. 4
5 Table 2 : Surface distribution of detection efficiency for 709 kev β rays of 36 Cl for the detector of 520 mm wide and 310 mm length A B C D E F G H I J % 34.6% 28.5% 27.7% 28.8% 26.0% 28.4% 34.7% 34.1% 33.6% % 33.7% 36.2% 36.4% 36.5% 36.6% 36.6% 36.4% 34.9% 33.8% % 35.2% 35.6% 37.6% 36.1% 36.6% 36.4% 35.3% 35.0% 33.5% % 34.5% 36.3% 35.4% 35.2% 35.2% 35.7% 33.2% 34.3% 33.0% % 32.3% 32.6% 32.8% 32.3% 33.5% 33.8% 34.8% 31.8% 33.6% % 32.0% 29.6% 29.4% 29.4% 30.6% 28.8% 30.6% 31.8% 31.7% Table 3 exemplifies the dimensions and detection efficiencies of five main plastic scintillation detectors for specified applications. The detection efficiencies were obtained with the 60 Co β ray source of 100 mm x 100 mm size set 1 cm away from the detector center. Table 3: Dimensions and detection efficiency data of five main plastic scintillation detectors for specified application Sensitive dimension ( mm ) 700(W) x 350(L) For β -ray 520(W) x 310(L) For β -ray 420(W) x 310(L) For β -ray 420(W) x 310(L) For β -ray+γ-ray Simultaneous measurement 310(W) x 220(L) For β-ray Outer thickness ( mm ) Detection efficiency 60 Co :25.5% Laundry monitor 60 Co :26.0% 60 Co :26.5% 60 Co(β) 60 Co(γ) :23.5% :14.5% 60 Co :21.0% Applications Body surface contamination monitor front and back Small article surface contamination monitor Small article surface contamination monitor Body surface contamination monitor Head, hands and feet Transportable small article surface contamination monitor 4. Application examples (1) Application to the contamination monitor Figure 6 shows two examples of application to the body surface contamination monitor. One is the one-through type to measure the surface contamination on the whole body (head, hands, feet, and front, back and lateral sides of the central body) at one time, which is commonly used in Japan. While in U.S.A, Europe etc., the two-step type to measure the front and the back sides of the body at two times is commonly used. The other is the hand-foot-clothes monitor to measure hands, feet, and clothes. 5
6 Figure 6: Application example for body surface contamination monitor One-through type to measure the surface contamination on the whole body Hand-foot-clothes monitor (2) Application of plastic scintillator to a spot contamination monitoring For the purpose to manufacture a contamination monitor to detect the spot contamination, a large area plastic scintillation detector of 31cm length by 22 to 70 cm width was developed by arraying a couple of detectors with effective areas of 31 cm length by 11 cm width as one unit, as shown at the right side in Fig.7. We aimed to detect the surface contamination radioactivity of 80 Bq 60 Co corresponding to 0.8 Bg/cm 2 radioactivity density for 10x10cm surface source. Although one large area detector shown at the left side in Fig. 6 has ever been used for spot contamination monitoring, this array-type detector using smaller size detectors will be able to have the following better characteristics. Figure 7: Conceptual design of detector array to realize a large area detector for spot contamination monitoring 220~ Large area detector in the past 6
7 By dividing the detector, background count rates can be reduced without decreasing the detection efficiency of spot contamination which is less than 10cm x 10cm. (1) By adding up the counts measured with the two adjacent detectors, the efficiency decrease near the border of these two detectors can be prevented. (2) By putting the detector of 11cm x 31cm together, the area size to detect the spot contamination can be freely changed according to the sizes of measured objects. 5. Conclusion We developed a large area plastic scintillation detector which fits to the shapes of the measured objects. Effective area of the detector is from about 300 to about 3000 cm 2, and the outer thickness of the detector unit is less than 10 cm for β ray measurements. These plastic scintillation detectors will be applied to various contamination monitors. 7
Comprehensive applications of the gas flow proportional counters for Radiological Surveillance
Comprehensive applications of the gas flow proportional counters for Radiological Surveillance D.A.R.Babu, Anand Raman, Ashokkumar P and D.N.Sharma Radiation Safety Systems Division, Bhabha Atomic Research
More informationEnvironmental Radiation Monitoring System
Environmental Radiation Monitoring System Toshihiro Takagi Osamu Kimura Atsushi Minagoshi 1. Introduction In order to ascertain conditions near the perimeter of a supervised area at a nuclear power plant
More informationIdentify the following features and specifications for the ESP1
LEARNING OBJECTIVES: 2.17.01 List the factors which affect an RCT's selection of a portable contamination monitoring instrument, and identify appropriate instruments for contamination monitoring. (Also
More informationCONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:
LEARNING OBJECTIVES: 2.17.01 List the factors which affects an RCT's selection of a portable contamination monitoring instrument. 2.17.02 Describe the following features and specifications for commonly
More informationWM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA
Implementation and Performance Testing of a Sensitive Bulk Laundry Monitor for Use in Nuclear Power Stations -14004 Erik Lindberg*, Richard DelPizzo*, Tom Donohoue*, Ryan Lee**, Marc R. Looman*, E. Ray
More informationRadioactive Contamination Monitor
Radioactive Contamination Monitor Toru Hasegawa Tadao Hashimoto Manabu Hashimoto 1. Introduction At nuclear power plants, surface contamination inspection monitors are installed at the boundaries of radiation
More informationCOUNTING ROOM EQUIPMENT RCT STUDY GUIDE
LEARNING OBJECTIVES: 2.19.01 Describe the features and specifications for commonly used laboratory counters or scalers: a. Detector type b. Detector shielding c. Detector window d. Types of radiation detected
More informationAM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL
AM-806 Waste/Door Radiation Monitor OPERATIONS MANUAL Section Contents 1. SPECIFICATIONS 2. MAIN COMPONENTS 3. THEORY OF OPERATION 4. OPERATING INSTRUCTIONS 5. INSTALLATION Page 2 1.0 SPECIFICATIONS SENSITIVITY
More informationRadiation Control and Monitoring System on the HTTR
Radiation Control and Monitoring System on the HTTR Y. Minowa, T. Nakazawa, H. Kikuchi, T. Nomura and N. Sato Japan Atomic Energy Research Institute Oarai Research Establishment, Oarai-machi, Ibaraki-ken
More informationSuperCDMS SNOLAB: Road to 100 mm Germanium Detectors
SuperCDMS SNOLAB: Road to 100 mm Germanium Detectors H. Chagani University of Minnesota for the SuperCDMS Collaboration 23 July 2012 The SuperCDMS Soudan Detector: izip interleaved Z-sensitive Ionisation
More informationNuclear Associates
Nuclear Associates 05-575 MiniMonitor III Multipurpose X and Gamma Ray Survey Meter Operators Manual February 2005 Manual No. 05-575-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed
More informationTSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110
TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110 APPLICATION NOTE CC-107 Introduction This purpose of this document is to detail the advanced, state of the art features TSI has incorporated in the design
More informationLUDLUM MODEL and PANCAKE CLUSTER DETECTORS. June 2016
LUDLUM MODEL 44-89 and 44-94 PANCAKE CLUSTER DETECTORS June 2016 LUDLUM MODEL 44-89 and 44-94 PANCAKE CLUSTER DETECTORS June 2016 STATEMENT OF WARRANTY Ludlum Measurements, Inc. warrants the products covered
More informationPortable Survey Instruments NISP-RP-01
NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection NISP-RP-01 This is an industry document for standardizing radiation protection processes used by supplemental radiation protection technicians.
More informationLUDLUM MODEL 44-7 Alpha, Beta, Gamma Detector. July 2012 Serial No. PR and succeeding Serial Numbers
LUDLUM MODEL 44-7 Alpha, Beta, Gamma Detector July 2012 Serial No. PR090405 and succeeding Serial Numbers LUDLUM MODEL 44-7 Alpha, Beta, Gamma Detector July 2012 Serial No. PR090405 and succeeding Serial
More informationLARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S)
LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S) Ganibu Dambis 26, P.O.Box 33 Riga, LV-1005, LATVIA Tel. (+371)6738-3947 Fax:(+371)6738-2620 office@bsi.lv www.bsi.lv CONTENTS INTRODUCTION...
More informationToday s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21
Today s Outline - September 12, 2016 C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016 1 / 21 Today s Outline - September 12, 2016 Detectors C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016
More informationLUDLUM MODEL 44-3 LOW ENERGY GAMMA SCINTILLATOR. May 2018 Serial Number PR and Succeeding Serial Numbers
LUDLUM MODEL 44-3 LOW ENERGY GAMMA SCINTILLATOR May 2018 Serial Number PR134823 and Succeeding Serial Numbers LUDLUM MODEL 44-3 LOW ENERGY GAMMA SCINTILLATOR May 2018 Serial Number PR134823 and Succeeding
More informationSeismic Countermeasures for Environmental Radiation Monitoring System
Seismic Countermeasures for Environmental Radiation Monitoring System Osamu Maekawa Takeshi Miyairi Teruo Ebato ABSTRACT Environmental radiation monitoring system continually measures and monitors the
More informationLUDLUM MODEL ALPHA, BETA, GAMMA HAND DETECTOR. June 2016 Serial Number and Succeeding Serial Numbers
LUDLUM MODEL 44-25 ALPHA, BETA, GAMMA HAND DETECTOR June 2016 Serial Number 155835 and Succeeding Serial Numbers LUDLUM MODEL 44-25 ALPHA, BETA, GAMMA HAND DETECTOR June 2016 Serial Number 155835 and
More informationProvläsningsexemplar / Preview TECHNICAL REPORT. Nuclear medicine instrumentation Routine tests
TECHNICAL REPORT IEC TR 61948-2 First edition 2001-02 Nuclear medicine instrumentation Routine tests Part 2: Scintillation cameras and single photon emission computed tomography imaging Instrumentation
More informationRAM GENE / RAM GENE-1
RAM GENE / RAM GENE-1 Radiation Meters Operating Manual Document #PRIR69N0.DOC Version 6.1 - December 2000 ROTEM INDUSTRIES LTD. ROTEM Industrial Park P.O.Box 9046, Beer-Sheva 84190, ISRAEL Tel: 972-8-6571312/6568652
More informationMission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design and deliver low light detection solutions which solve customer probl
份有限公司 wwwom 先锋科技股份有限公司 份有限公司 wwwom 先锋科技股份有限公司 www Radiation detection solutions for miniaturized pager and handheld markets sensl: light detection and measurement October - 2009 www.sensl.com 1 Mission
More informationGAMMA RAY DETECTION PROBE MODEL SO16
GAMMA RAY DETECTION PROBE MODEL SO16 INSTRUCTION S MANUAL EURORAD 2, rue Ettore Bugatti F67201 ECKBOLSHEIM Tel: +33 (0)3 88 26 81 30 Fax: +33 (0)3 88 28 45 48 E-mail: info@eurorad.com - Web: www.eurorad.com
More informationNUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use
NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection Level 3 Information Use NISP-RP-001 Industry Approval Date: September 14, 2018 This is an industry document for standardizing radiation protection
More informationBattery-operated, hand-held alarm and measuring instruments for radiation protection
Battery-operated, hand-held alarm and measuring instruments for radiation protection The GRAETZ alarm dosemeter ED150 warns the task force when the defined radiation dose for the operation is reached.
More informationResearch Article Development and Comparison of Fiber-Optic Beta Radiation Sensors with Different Diameters of Their Sensing Probes
Hindawi Sensors Volume 2017, Article ID 1452765, 5 pages https://doi.org/10.1155/2017/1452765 Research Article Development and Comparison of Fiber-Optic Beta Radiation Sensors with Different Diameters
More informationElectronic Personal Dosemeter
Electronic Personal Dosemeter Tomoya Nunomia Hideshi Yamauchi Tetsuo Shibata 1. Introduction Since developing an electronic personal dosemeter that uses a semiconductor detector in 1983, Fuji Electric
More informationPORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System
FEATURES: HIGH GAMMA SENSITIVITY MULTI-CHANNEL ANALYZER (MCA) WORKS WITH VARIETY OF PROBES RADIONUCLIDE LIBRARY TAILORED TO CUSTOMER REQUEST SIMULTANEOUSLY MEASURES ; DISPLAYS; ALARMS 1-3 SEPARATE DETECTORS
More informationRAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009
RAM GENE-1 ERK MK-Ι & MK-ΙΙ Contamination & Radiation Meter Operating Manual Document #PRIR90N2.DOC Version 2.2 October 2009 Version / Revision Log: RAM GENE-1 ERK/ Operating Manual Ver./Rev.# Date Revised
More informationdewar or an electrically powered cooler. The sensitive detector surfaces are thus protected from moisture and condensable contaminants.
Germanium Detectors The Company Canberra Industries is the world's largest manufacturer of nuclear radiation detection and analysis systems. In order to better serve our customers worldwide, Canberra operates
More informationREPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES
REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES ABSTRACT TITLE: Inspection of conventional diagnostic
More informationUltraviolet radiation detector to obtain the rate of particles at different heights
Journal of Physics: Conference Series PAPER OPEN ACCESS Ultraviolet radiation detector to obtain the rate of particles at different heights To cite this article: E Ponce et al 2016 J. Phys.: Conf. Ser.
More informationNEX-BETA-ABG FACILITY Drinking Water Safety Monitor
NEX-BETA-ABG FACILITY Drinking Water Safety Monitor FEATURES: DETECTS ALPHA, BETA AND GAMMA RADIATION AVAILABLE IN PORTABLE OR INSTALLED VERSION MEASURES AT OR BELOW EPA/DHS PAG LEVELS PROTECTIVE ACTION
More informationRadiation Monitoring Systems
Radiation Monitoring Systems Certificate No. 3250 Contents A guide to Southern Scientific radiation monitoring systems Area Gamma & Interlock Systems (page 3) AGM3 AUTOMESS 6150 Contamination Monitors
More informationAlpha Attenuation Due to Dust Loading. Amber E. Dailey Francis Marion University Savannah River Site
Alpha Attenuation Due to Dust Loading Amber E. Dailey Francis Marion University Savannah River Site Purpose The purpose of this experiment was to properly correct for the attenuation of the energy of alpha
More informationHeat Transfer in Evacuated Tubular Solar Collectors
Heat Transfer in Evacuated Tubular Solar Collectors Graham L. Morrison, Indra Budihardjo and Masud Behnia School of Mechanical and Manufacturing Engineering University of New South Wales Sydney 2052 Australia
More informationThe Modular System for Radioactivity Measurements FH 40 G/GL FHT 6020
Earphones The Modular System for Radioactivity Measurements For the connection to the separate earphone output. This enables the simultaneous operation of external FH 40 G probes and earphone. For the
More informationApplication Note. Application Note for BAYEX
Application Note Application Note for BAYEX Preface This application note provides the user a more detailed description of the Bayesian statistical methodology available in Version 8.05 and above, of the
More informationThe Modular System for Radioactivity Measurements FH 40 G/GL FHT 6020
The Modular System for Radioactivity Measurements FH 40 G/GL Digital Survey Meter FHT 6020 Display Unit Rugged and reliable Versatile multipurpose meter and area monitor Data logging Internal detector
More informationLUDLUM MODEL AND LUDLUM MODEL ALPHA, BETA, & GAMMA DETECTORS
LUDLUM MODEL 44-9-18 AND LUDLUM MODEL 44-9-19 ALPHA, BETA, & GAMMA DETECTORS January 2018 Serial Number PR254331 and Succeeding Serial Numbers LUDLUM MODEL 44-9-18 AND LUDLUM MODEL 44-9-19 ALPHA, BETA,
More informationVariable far infrared radiation (VFIR) technique for cubic carrot drying
International Journal of Renewable Energy, Vol. 7, No. 2, July - December 2012 Variable far infrared radiation (VFIR) technique for cubic carrot drying Sirinuch Chindaruksa a *, Pisit Maneechot b, Seishu
More informationFEATURE ARTICLE. The PLCA-800 Series Inline Particle Sensor for Chemical Solutions. Yoshihito Yuhara. Abstract
FEATURE ARTICLE Technical Reports The PLCA-800 Series Inline Particle Sensor for Chemical Solutions Yoshihito Yuhara Front end process Sensor module Wafer preparation process Film formation Photo lithography
More informationLUDLUM MODEL 44-2 GAMMA SCINTILLATOR. February 2018 Serial Number PR and Succeeding Serial Numbers
LUDLUM MODEL 44-2 GAMMA SCINTILLATOR February 2018 Serial Number PR361009 and Succeeding Serial Numbers LUDLUM MODEL 44-2 GAMMA SCINTILLATOR February 2018 Serial Number PR361009 and Succeeding Serial
More informationAdvanced Radiation Measurement Solutions
GE Energy Advanced Radiation Measurement Solutions Reuter Stokes Radiation Measurement Solutions GE Energy, Edison Park in Twinsburg, OH, USA World-class research and development Beginning with the first
More informationDOE 2.17 Contamination Monitoring Instrumentation Study Guide. 00ICP330 Rev. 0 Page 1 of 23
00ICP330 Rev. 0 Page 1 of 23 Course Title: Radiological Control Technician Module Title: Contamination Monitoring Instrumentation Module Number: 2.17 Objectives: 2.17.01 List the factors which affects
More informationA Continuous Air Monitoring Sampler for 125 I - RIS-125
A Continuous Air Monitoring Sampler for 125 I - RIS-125 S.Levinson, I.Belaish, T.Kravchik, U.German, O.Pelled, Y. Laichter, Y.Mazor, U.Wengrowicz, E.Dolev, H.Assido, D.Tirosh Nuclear Research Center Negev,
More informationIAEA-SM-367/14/01 UTILIZATION OF FLUORESCENT URANIUM X-RAYS AS VERIFICATION TOOL FOR IRRADIATED CANDU FUEL BUNDLES
IAEA-SM-367/14/01 UTILIZATION OF FLUORESCENT URANIUM X-RAYS AS VERIFICATION TOOL FOR IRRADIATED CANDU FUEL BUNDLES Iqbal Ahmed 1, Rolf Arlt 2, Victor Ivanov 3 and K G. Qasim 4 [1] Institute for Nuclear
More informationHEAT REMOVAL TESTS FOR THE HIGH PERFORMANCE VAULT STORAGE SYSTEM OF SPENT NUCLEAR FUEL
HEAT REMOVAL TESTS FOR THE HIGH PERFORMANCE VAULT STORAGE SYSTEM OF SPENT NUCLEAR FUEL M. Oda*, M. Shimizu*, N. Kumagai*, H. Kanai**, M. Tanabe** *Nuclear Fuel Cycle Group, Power & Industrial Systems R&D
More informationDomestic Nuclear Detection Office (DNDO)
Domestic Nuclear Detection Office (DNDO) Advancing Technological Capabilities to Prevent Nuclear Terrorism 27 October 2016 Joel Rynes, Ph.D., PMP Assistant Director Transformational and Applied Research
More informationNorth American Technical Services
WWW.NATS-USA.COM High Resolution Detectors (HPGE) North American Technical Services Bridging Technology Innovative Solutions In in Radiation Detection and & Analysis High Resolution Germanium Detectors
More informationExperimental Particle Physics PHYS6011 Joel Goldstein, RAL
Experimental Particle Physics PHYS6011 Joel Goldstein, RAL 1. Introduction & Accelerators 2. Particle Interactions and Detectors (1/2) 3. Collider Experiments 4. Data Analysis Charged Particle Detectors
More informationDepleted Uranium Waste Assay at AWE. T.J. Miller AWE, Aldermaston, Reading, Berkshire, England, RG7 4PR UK
Depleted Uranium Waste Assay at AWE T.J. Miller AWE, Aldermaston, Reading, Berkshire, England, RG7 4PR UK ABSTRACT The Atomic Weapons Establishment (AWE) at Aldermaston has recently conducted a Best Practical
More informationTexas active target (TexAT) detector part 1: Design and construction progress
Texas active target (TexAT) detector part 1: Design and construction progress E. Koshchiy, G.V. Rogachev, E. Uberseder, and E. Pollacco 1 1 IRFU, CEA Saclay,Gif-Sur-Ivette, France Construction of a general
More informationMission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design and deliver low light detection solutions which solve customer probl
份有限公司 wwwom 先锋科技股份有限公司 www 份有限公司 wwwom 先锋科技股份有限公司 Solutions for Nuclear Medicine sensl: light detection and measurement Mission 份有限公司 wwwom 先锋科技股份有限公司 www Working with our customers we identify, design
More informationConsultation Draft. Issue 2. Good Practice Guide No. 29. The Examination, Testing and Calibration of Installed Radiation Protection Instruments
Good Practice Guide No. 29 The Examination, Testing and Calibration of Installed Radiation Protection Instruments Peter Burgess, Lynsey Keightley, Clare Lee, Max Pottinger, Mike Renouf, David Williams.
More informationß-Aerosolmonitor FHT 59 C
Technical Specification ZT-240-1103 E page 1/11 1. FHT 59C Moving Filter Aerosol Monitor 1.1 General The FHT 59 C is a unique moving filter aerosol monitor that can operate for one year or longer without
More informationMODEL AND MODEL GAS PROPORTIONAL DETECTORS. December 2016 Serial No and Succeeding Serial Numbers
MODEL 43-37 AND MODEL 43-37-1 GAS PROPORTIONAL DETECTORS Serial No. 086152 and Succeeding Serial Numbers MODEL 43-37 AND MODEL 43-37-1 GAS PROPORTIONAL DETECTORS Serial No. 086152 and Succeeding Serial
More informationInstrumentation. Post RDD Event Urban Reoccupancy. James P. Menge PE. The world leader in serving science
The world leader in serving science Instrumentation Post RDD Event Urban Reoccupancy James P. Menge PE Perception In the aftermath of a radiological emergency the public will see radiation and its potential
More informationRemote Gamma Imaging of High Dose Environments
Remote Gamma Imaging of High Dose Environments Paul Griffiths Copyright BIL Solutions Ltd 2006. The copyright in this document is wholly vested in BIL Solutions Ltd who reserve all intellectual property
More informationGaseous detectors. 1
Gaseous detectors Very important class of detectors with many applications charged particle, x- & gamma rays, visible light detection applications in many areas of research, & commercial particle & nuclear
More informationHighlights from a recent test campaign in Finland with scalable neutron detectors
Highlights from a recent test campaign in Finland with scalable neutron detectors Neutron Users Club Dr. Geraint Dermody Introduction Symetrica s activities and products The NKS (Nordic Nuclear Safety
More informationQuality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points
Topic 4 - Audio 46 Key Points : Measures taken to increase confidence in results. Quality Assurance: Program that assures that measures are being done and in a timely manner. Precision: How close measurements
More informationCL4001 HEAT TRANSFER OPERATIONS
CL4001 HEAT TRANSFER OPERATIONS MODULE V Lecture Notes: Debasree Ghosh HEAT EXCHANGER Lecturer, Department of Chemical Engineering, Birla Institute of Technology, Mesra Heat Exchangers: Introduction and
More informationGRAPPLE SCANNER FOR RADIATION IN SCRAP METAL. Model Series ~ GRAPPLE SCAN- GR Models ~ GRAPPLE SCAN- GR-A, -B, -C, & -D
FEATURES: WIRELESS COMMUNICATION BETWEEN DETECTOR & CRANE OPERATOR SURFACE SCAN, LOAD ANALYSIS, SEARCH MODES FAST SCAN, REAL TIME ALARM NO MOVING PARTS QUICK CHANGE DETECTOR BETWEEN GRAPPLE HOUSINGS DETECTS
More informationS.E.A. GmbH - Your Partner for
S.E.A. GmbH - Your Partner for Civil Protection Disaster Relief Fire Departments Technical Support Services Analytical CBRN Dekon-P Task Force A company of Contamination Measurement CoMo-170 ZS CoMo-170
More informationXGLAB. X and Gamma Ray Electronics. Looking for new frontiers...
XGLAB X and Gamma Ray Electronics Looking for new frontiers... Research And Development XGLab, spin off company of Politecnico di Milano, the leading Italian technical University, holds a leadership and
More informationDevelopment of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment
EFDA JET CP(04)05-07 L.B.C. Worth, R.J.H. Pearce, J. Bruce, J. Banks, S. Scales and JET EFDA Contributors Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use
More informationDetector Configurations for the MAJORANA Demonstrator
Detector Configurations for the MAJORANA Demonstrator Lawrence Livermore National Laboratory UC Berkeley 1 How to determine the best configuration for Majorana? Define a metric reflecting the performance
More informationVictoreen. Operator Manual. Current Mode Beta Scintillation Detector, Model
Victoreen Current Mode Beta Scintillation Detector, Model 943-27 Operator Manual February 2007 Manual No. 943-27-31-1, Rev. 3 2007 Fluke Corporation, All rights reserved. Printed in U.S.A. All product
More informationLUDLUM MODEL 44-9 ALPHA, BETA, GAMMA DETECTOR. February 2010 Serial Number PR and Succeeding Serial Numbers
LUDLUM MODEL 44-9 ALPHA, BETA, GAMMA DETECTOR February 2010 Serial Number PR090405 and Succeeding Serial Numbers LUDLUM MODEL 44-9 ALPHA, BETA, GAMMA DETECTOR February 2010 Serial Number PR090405 and
More informationDEVELOPMENT OF THE INFRARED INSTRUMENT FOR GAS DETECTION
DEVELOPMENT OF THE INFRARED INSTRUMENT FOR GAS DETECTION Ching-Wei Chen 1, Chia-Ray Chen 1 1 National Space Organization, National Applied Research Laboratories ABSTRACT MWIR (Mid-Wave Infrared) spectroscopy
More informationHIGHLY SENSITIVE FOR NUCLEAR POWER PLANTS Real-Time Continuous Water Monitor
FEATURES: HIGHLY SENSITIVE FOR NUCLEAR POWER PLANTS WORLD S ONLY PAG-LEVEL WATER MONITOR MEASURES AT OR BELOW EPA/DHS PAG LEVELS (EPA s PROTECTIVE ACTION GUIDELINE LEVELS) REAL TIME, IN-LINE, CONTINUOUS
More informationMODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017
MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER August 2017 MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER August 2017 ii STATEMENT OF WARRANTY Ludlum Measurements, Inc. warrants the products covered
More informationEnergy and Angular Responses of the Criticality Accident Alarm System Using Current-mode-operated Scintillation Detector
Energy and Angular Responses of the Criticality Accident Alarm System Using Current-mode-operated Scintillation Detector N. Tsujimura 1, T. Yoshida 1 and S. Mikami 1 1 Japan Nuclear Cycle Development Institute,
More informationThe Development of an Integrated Site System for the detection of illicit trafficking of radioactive and nuclear materials
The Development of an Integrated Site System for the detection of illicit trafficking of radioactive and nuclear materials Kevin Whitehouse, Dan Parvin British Nuclear Group, Project Services Limited,
More informationVictoreen Primalert 35 Area Radiation Monitor
Victoreen 05-437 Primalert 35 Area Radiation Monitor Operators Manual March 2005 Manual No. 126011 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationMeasurement Devices. Topic 5 - Audio 56
Topic 5 - Audio 56 Other Classifications Passive devices: Air diffuses in No power required for sampling Active devices: Requires power Pumps air through filter Key Points - Devices What lab sees: Different
More informationInfluence of New Technologies on Technical Reachback
Influence of New Technologies on Technical Reachback Prof. Olof TENGBLAD, IEM CSIC Madrid Spain Dr Sakari Ihantola, Imperial College London 2 Index New detector technologies Advantages of technical reachback
More informationRecent Advances in Mercuric Iodide Detector Fabrication and Instrument Development*
Recent Advances in Mercuric Iodide Detector Fabrication and Instrument Development* Lodewijk van den Berg and Steven W. Pauly Constellation Technology Corp., 7887 Bryan Dairy Road, Suite 100, Largo, FL
More informationCOMPANY PROFILE. DETECTION & MEASUREMENT OF IONIZING RADIATIONS RADIOLOGICAL MONITORING SYSTEMS HEALTH AND ENVIRONMENTAL SAFETY
monitoring for safety COMPANY PROFILE DETECTION & MEASUREMENT OF IONIZING RADIATIONS RADIOLOGICAL MONITORING SYSTEMS ENVIRONMENTAL MONITORING HEALTH AND ENVIRONMENTAL SAFETY www.rtsinstruments.com 2 monitoring
More informationThermal Performance Enhancement of Inclined Rib Roughness Solar Air Heater
Thermal Performance Enhancement of Inclined Rib Roughness Solar Air Heater Gurpreet Singh and Satwant Singh Lala Lajpat Rai Institute of Engineering and Technology, Moga (Punjab) 142001 Abstract : Artificial
More informationDevelopment of high sensitivity radon detectors
Nuclear Instruments and Methods in Physics Research A 421 (1999) 334 341 Development of high sensitivity radon detectors Y. Takeuchi *, K. Okumura, T. Kajita, S. Tasaka, H. Hori, M. Nemoto, H. Okazawa
More informationW.E.T. (Water Extraction Techniques) Lab
SNO-STR-90- /? Report on Activities W.E.T. (Water Extraction Techniques) Lab M.-Q. Liu, H. Lee and A.B. McDonald January 9, 1990 The following is a summary of work to date at Queen s on the Lucas cell
More informationEVALUATION OF A NEUTRON-PHOTON SHIELD FOR TRANSURANIC (TRU) WASTE CONTAINERS
EVALUATION OF A NEUTRON-PHOTON SHIELD FOR TRANSURANIC (TRU) WASTE CONTAINERS M. Gallegos, R. Ruby, E. J. Sullivan, R. J. Wishau Los Alamos National Laboratory ABSTRACT The Los Alamos National Laboratory
More informationDevelopment of the CMS Phase-1 Pixel Online Monitoring System and the Evolution of Pixel Leakage Current
Development of the CMS Phase-1 Pixel Online Monitoring System and the Evolution of Pixel Leakage Current Fengwangdong Zhang On behalf of CMS Pixel Collaboration The 9th International Workshop on Semiconductor
More informationOPERATION OF RADIATION PORTAL MONITORS OF SPECIAL NUCLEAR MATERIALS AT NUCLEAR SITES IN TERMS OF RELIABILITY
Jr. of Industrial Pollution Control 32(2016) pp 442-448 www.icontrolpollution.com Research OPERATION OF RADIATION PORTAL MONITORS OF SPECIAL NUCLEAR MATERIALS AT NUCLEAR SITES IN TERMS OF RELIABILITY E.A.
More informationPerformance of Shell and Tube Heat Exchanger under Varied Operating Conditions
Performance of Shell and Tube Heat Exchanger under Varied Operating Conditions N. Prabhu Kishore 1, N. Alekhya 2, J. Ugandhar 3 Asst. Professor, Department of Mechanical Engineering, MLR Institute of Technology,
More informationStatus Report about the TPC Detector and Module at CEPC
Status Report about the TPC Detector and Module at CEPC Huirong QI On behalf of CEPC TPC subgroup, IHEP, CAS 2016.01.19, IAS, HK Content Goals and options Some considerations Preliminary simulation Hybrid
More informationLECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry
LECTURE 11 Dr. Teresa D. Golden University of North Texas Department of Chemistry C. Detectors Properties of Detectors -quantum counting efficiency -linearity -energy proportionality -resolution C. Detectors
More informationContamination Meter Users Manual
RDS-80 Contamination Meter U Users Manual Version 1.01 LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - www.laurussystems.com NOTE This document and the information herewith are copyrighted
More informationDevelopment of Large Size Hybrid Fan
Technical Paper Toshihiko Nishiyama Kengo Koshimizu Keiichi Inaba A lot of small fans with three-dimensional, wide chord length, forward swept figure, and/or shroud ring have been developed and manufactured
More informationA STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE
A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE Satoru Ito 1, Keisuke Tsukada 1, Nobuyoshi Tsuzuki 2, Takao Ishizuka 3 and Hiroshige Kikura 3 1 Department of nuclear Engineering, Graduate
More informationEvaluation of a dynamic model for a cold climate counter flow air to air heat exchanger
Evaluation of a dynamic model for a cold climate counter flow air to air heat exchanger Toke Rammer Nielsen, Associate Professor, Department of Civil Engineering, Technical University of Denmark; trn@byg.dtu.dk
More informationLAUNDRY MONITORING SYSTEM Model # LIM-64S
LAUNDRY MONITORING SYSTEM FEATURES: HIGH SENSITIVITY EVEN TO LOW ENERGY GAMMAS 6 EACH 8 X 8 SCINTILLATORS ABOVE 4 EACH 2 X 12 NaI(Tl) BELOW AUTOMATIC CONVEYOR FOR LAUNDRY AUTOMATICALLY SEPARATES CLOTHES
More information(12) United States Patent (10) Patent No.: US 6,361,301 B1
USOO636.1301B1 (12) United States Patent (10) Patent No.: Scaglotti et al. (45) Date of Patent: Mar. 26, 2002 (54) HEATER ASSEMBLY FOR BLOW MOLDING 5,256,341. 10/1993 Denis et al. PLASTIC PREFORMS 5,549,468
More informationA new fully digital system for RT inspection of metal tube to tube sheet joints of heat exchangers
17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China A new fully digital system for RT inspection of metal tube to tube sheet joints of heat exchangers Uwe ZSCHERPEL 1, Oleksandr
More informationDesign of Microwave Anti-theft Alarm Based on Microcontroller
Design of Microwave Anti-theft Alarm Based on Microcontroller Hanhong Tan Guangdong University of Science & Technology,Dongguan,523083, China Abstract At present, most households use safety gates and install
More informationAPPLICATION FOR UNITED STATES PATENT. Shieldless Detector with Two-Dimensional Directionality INVENTOR:: David E. Newman DATE:: November 9, 2018
APPLICATION FOR UNITED STATES PATENT TITLE:: Shieldless Detector with Two-Dimensional Directionality INVENTOR:: David E. Newman DATE:: November 9, 2018 ABSTRACT Disclosed is a directional gamma ray or
More informationMaximize safety and productivity. Thermo Scientific Nuclear Power Radiation Detection and Monitoring Solutions
Maximize safety and productivity Thermo Scientific Nuclear Power Radiation Detection and Monitoring Solutions Monitoring solutions to optimize safety, efficiency and compliance Our years of expertise and
More information