Technical Description. Ionisation Chambers
|
|
- Alban Cooper
- 6 years ago
- Views:
Transcription
1 Technical Description Ionisation Chambers
2 Manufacturer: IMS Messsysteme GmbH Dieselstraße Heiligenhaus Tel.: +49 (0) 2056 / Fax: +49 (0) 2056 / info@ims-gmbh.de (c) IMS Messsysteme GmbH. All rights reserved. This documentation is protected in all parts by copyright laws. Distribution and reproduction without the prior written authorisation of IMS Messsysteme GmbH are prohibited. All company and product names used in this documentation may be registered trademarks. 2/20 Vers.: HW_IK_0104_en
3 Contents 1 Application Description Types/ Order Text Types Order Text Technical Data Series KG Series KG Series KB Series KG Series KG Series KG Dimensional Drawings Connection of Ionisation Chambers Ionisation Chambers without RC Filter (High Voltage Filter) Ionisation Chambers with RC Filter (High Voltage Filter) General Information Storage and Transport Mounting Handling of Unusable Ionisation Chambers HW_IK_0104_en Vers.: /20
4 1 Application Ionisation chambers are used in radiometry as detectors to measure radioactive radiation and therefore belong to the most important components of IMS gauging systems in the field of radiometry (e.g. to measure thickness, profile and edge drop). Their output signal is a direct current proportional to the incident radiation. With their technical features, proven operational reliability and practically unlimited lifetimes, they satisfy all the necessary prerequisites for use at points of radiometric gauging systems exposed to radiation. 2 Description In principle the ionisation chamber is constructed like a cylinder or plate capacitor. The signal and voltage electrodes are housed in a closed tubular steel housing, which is filled with inert gas and has a beam entrance window on the front side. The voltage electrode is connected to the positive pole of the high-voltage power supply (generator of the operating voltage = chamber high voltage needed for the operation of ionisation chambers) and the signal electrode to the amplifier input. Special coax cables are used for these connections. Connected to the shielding of the coax cable to the signal electrode, the housing of the ionisation chamber is always connected to earth potential to shield against external parasitic induction. The incident radioactive radiation causes the gas to ionise. Under the influence of the electric field the negative electrons migrate to the positive voltage electrode and the ions to the signal electrode (measuring electrode). The ionisation chambers are operated in so-called saturation range, i.e. small changes in the operating voltage have practically no influence on the ionisation chamber current measurable at the signal electrode. The ionisation chambers come in various versions. The type of construction, gas and operating pressure as well as the thickness and material of the beam entrance window all determine the "measuring" characteristics of the chambers and simultaneously constitute their main distinguishing features. The elementary distinguishing features are summarised in tabular form in the chapter "Technical Data". Ionisation chambers of the same type of construction form a series. The choice of ionisation chamber is application-specific and depends primarily on the type of radiation, which is determined by the particular task in question. If the task allows different versions to be used, other criteria e.g. the mounting situation can influence the choice. 4/20 Vers.: HW_IK_0104_en
5 3 Types/ Order Text 3.1 Types Type Article No. Main Field of Application for Measurement of Series KG 60 KG 60 / 50 (KG 60 / AM) γ - radiation (Am 241) / X-radiation KG 60 / γ - radiation (Am 241) / X-radiation KG 60 / γ - radiation (Am 241) / X-radiation KG 60 / X-radiation KG 60 / 60 (KG 60 / CS 12) γ - radiation (Cs 137) KG 60 / 61 (KG 60 / CS 18) γ - radiation (Cs 137) Series KG 126 KG 126 / 50 (KG 126 / AM) γ - radiation (Am 241) / X-radiation KG 126 / 61 (KG 126 / CS - KR) γ - radiation (Cs 137) KG 126 / 62 (KG 126 / CS - XE) γ - radiation (Cs 137) Series KB 100 KB 100 / 10 (KB 100 / PM) β - radiation (Pm 147) KB 100 / 20 (KB 100 / KR) β - radiation (Kr 85) KB 100 / 30 (KB 100 / SR) β - radiation (Sr 90) Series KG 100 KG 100 / 40 (KG 100 / CM) γ - radiation (Cm 244) / X-radiation KG 100 / 41 (KG 100 / CM - XE) γ - radiation (Cm 244) / X-radiation KG 100 / 50 (KG 100 / AM) γ - radiation (Am 241) / X-radiation Series KG 90 KG 90 / 10 (KG 90 / FE) Secondary radiation KG 90 / 20 (KG 90 / PM) Secondary radiation / X-radiation KG 90 / 30 (KG 90 / KR) Secondary radiation / X-radiation Series KG 20 KG 20 / X-radiation 10/11/12 KG 20 / X-radiation 20/21/22 KG 20 / Secondary radiation / X-radiation KG 20 / Secondary radiation / X-radiation KG 20 / Secondary radiation / X-radiation KG 20 / Secondary radiation / X-radiation Printed in italics = original type designation Article no. = Drawing no. γ - radiation / β - radiation / X-radiation: Measurement - transmission technique Secondary radiation (backscatter radiation, fluorescent radiation): Measurement - backscatter technique HW_IK_0104_en Vers.: /20
6 3.2 Order Text The order text comprises the type designation and the article number. 4 Technical Data 4.1 Series KG 60 Type KG 60 / 50 KG 60 / 51 KG 60 / 52 Original type designation (KG 60 / AM) (KG 60 / AM - KR) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Xenon Krypton Xenon Gas pressure [Pa] 7,5x10 5 3,5x ,5x10 5 Window material Stainless steel Stainless steel Stainless steel Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] 30 - Voltage electrode [pf] 74 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 2000 RC filter in connection head No Connection cable (standard length) [m] 6 Weight (without /with connection cable) [kg] 0.9 / 1.3 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing Connection cable for operating voltage (chamber high voltage) marked red 6/20 Vers.: HW_IK_0104_en
7 Series KG 60 (continuation) Type KG 60 / 53 KG 60 / 60 KG 60 / 61 Original type designation (KG 60 / CS - 12) (KG 60 / CS - 18) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] 0,3 0,3 0,5 Radiosensitive volume ca. [l] 0,2 0,2 0,3 Gas Xenon Xenon Xenon Gas pressure [Pa] 14.5x x x10 5 Window material Stainless steel Stainless steel Stainless steel Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] Voltage electrode [pf] Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 2000 RC filter in connection head No Connection cable (standard length) [m] 6 Weight (without / with connection cable) [kg] 0,9 / 1,3 1,1 / 1,6 1,6 / 2 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing Connection cable for operating voltage (chamber high voltage) marked red HW_IK_0104_en Vers.: /20
8 4.2 Series KG 126 Type KG 126 / 50 KG126 / 61 KG 126 / 62 Original type designation (KG 126 / AM) (KG 126 / CS - KR) (KG 126 / CS - XE) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Xenon Krypton Xenon Gas pressure [Pa] 3.5x x x10 5 Window material Stainless steel Stainless steel Stainless steel Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] 75 - Voltage electrode [pf] 160 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 2000 RC filter in connection head 470kΩ/ 10nF/4000V Connection cable No Weight [kg] General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The coax cable (article no ) must be used as connection cable for operating voltage and measuring signal 8/20 Vers.: HW_IK_0104_en
9 4.3 Series KB 100 Type KB 100 / 10 KB 100 / 20 KB 100 / 30 Original type designation (KB 100 / PM) (KB 100 / KR) (KB 100 / SR) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Argon Krypton Xenon Gas pressure [Pa] 1.4x10 5 3x10 5 5x10 5 Window material Titanium Titanium Titanium Window thickness [µm] Window with protective grating Yes Yes Yes Internal capacitance - Signal electrode [pf] Voltage electrode [pf] 560 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 1000 RC filter in connection head No Connection cable No Weight [kg] 4 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The coax cable (article no ) must be used as connection cable for operating voltage and measuring signal HW_IK_0104_en Vers.: /20
10 4.4 Series KG 100 Type KG 100 / 40 KG 100 / 41 KG 100 / 50 Original type designation (KG 100 / CM) (KG 100 / CM - (KG 100 / AM) XE) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Krypton Xenon Xenon Gas pressure [Pa] 1.6x10 5 2x10 5 6x10 5 Window material Titanium Titanium Titanium Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] Voltage electrode [pf] 560 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 1000 RC filter in connection head 4,7 MΩ/ 10nF/4000V Connection cable No Weight [kg] 4 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The coax cable (article no ) must be used as connection cable for operating voltage and measuring signal 10/20 Vers.: HW_IK_0104_en
11 4.5 Series KG 90 Type KG 90 / 10 KG 90 / 20 KG 90 / 30 Original type designation (KG 90 / FE) (KG 90 / PM) (KG 90 / KR) Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Argon Argon Krypton Gas pressure [Pa] 1.1x x x10 5 Window material Beryllium Titanium Titanium Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] 20 - Voltage electrode [pf] 60 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 1000 RC filter in connection head No Connection cable No Weight [kg] 2.7 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The coax cable (article no ) must be used as connection cable for operating voltage and measuring signal HW_IK_0104_en Vers.: /20
12 4.6 Series KG 20 Type KG 20 / 10 KG 20 / 20 KG 20 / 30 Design 0, 1, 2 0, 1, 2 Article No. (Drawing No.) /11/ /21/ Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Xenon Krypton Krypton Gas pressure [Pa] 6.5x x x10 5 Window material Stainless steel Stainless steel Titanium Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] 25 - Voltage electrode [pf] 55 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 1600 RC filter in connection head No Connection cable No Weight (design 0 respectively 1 / 2) [kg] 0.7 / General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The double coax cable (article no ) must be used as connection cable Differences relating to the connecting boxes only are limited to the types KG 20 / 10 and KG 20 / 20 12/20 Vers.: HW_IK_0104_en
13 Series KG 20 (continuation) Type KG 20 / 40 KG 20 / 50 KG 20 / 60 Article No. (Drawing No.) Characteristic Data Chamber volume (gas volume) ca. [l] Radiosensitive volume ca. [l] Gas Argon Argon Argon Gas pressure [Pa] 2x x x10 5 Window material Titanium Titanium Beryllium Window thickness [µm] Window with protective grating No No No Internal capacitance - Signal electrode [pf] 25 - Voltage electrode [pf] 55 Operating voltage (chamber high voltage) - Maximum value U Hmax [V] 1600 RC filter in connection head No Connection cable No Weight (design 0 respectively 1 / 2) [kg] 0.9 General Data Type of construction See dimensional drawing chapter 5 Insulation resistance [Ω] > System of housing protection IP 64 (EN 60259) Climatic conditions - Ambient temperature [ C] Transport and storage temperature [ C] Location of the sensitive chamber volume: see shading respectively hatching in dimensional drawing Measured between the electrode connections in the chamber and housing The double coax cable (article no ) must be used as connection cable HW_IK_0104_en Vers.: /20
14 5 Dimensional Drawings KG 60 KB 100 and KG ,5 L L3 L E L L L2 L L1 Dim. L [mm] L1 [mm] L2 [mm] LE [mm] Dim. L1 [mm] L2 [mm] L3 [mm] L4 [mm] KG 60 / 50 ca KB 100 / 10 16, ,5 KG 60 / 51 ca KB 100 / 20 16, ,5 KG 60 / 52 ca ca KB 100 / 30 20,2 104,5 151,5 11 KG 60 / 53 ca ca KG 60 / 60 ca KG 100 / 40 16, ,5 KG 60 / 61 ca KG 100 / 41 16, ,5 KG 100 / 50 20,2 104,5 151, /20 Vers.: HW_IK_0104_en
15 KG 126 / 50 KG 126 / 61 and KG 126 / KG HW_IK_0104_en Vers.: /20
16 KG 20 / 10 and KG 20 / 20, each design 0 16/20 Vers.: HW_IK_0104_en
17 KG 20 / 10 and KG 20 / 20, each design 1 KG 20 / 10 and KG 20 / 20, each design 2 HW_IK_0104_en Vers.: /20
18 KG 20 / 30 / 40 / 50 / 60 18/20 Vers.: HW_IK_0104_en
19 6 Connection of Ionisation Chambers 6.1 Ionisation Chambers without RC Filter (High Voltage Filter) Only KG 20 U S 1 S U gr sw 1 S : Signalelektrode / Signal electrode U : Spannungselektrode / Voltage electrode 1 : Schirm / Shielding Rote Markierung Red marking /6 6.2 Ionisation Chambers with RC Filter (High Voltage Filter) Only the connection rooms of the ionisation chambers KG 126 and KG 100 are equipped with RC filters (high voltage filters). The values for R and C are to be found in the technical data (4.2 and 4.4.). U R C 1 S S : Signalelektrode / Signal electrode U : Spannungselektrode / Voltage electrode 1 : Schirm / Shielding Rote Markierung Red marking /6 HW_IK_0104_en Vers.: /20
20 7 General Information See the "Assembly Instructions - Part 1" for detailed information on storage, transport and assembly 7.1 Storage and Transport Some ionisation chambers are equipped with a protective cover to protect the beam entrance window during storage and transport. It is recommended that this protective cover only be removed after mounting of the ionisation chamber. 7.2 Mounting Low-noise, high-impedance coax cables are used as connection cables for the measuring signal and operating voltage (chamber high voltage). They can be bought - as special accessories - under the article numbers given in the table below. Article No. Article Designation Outside Remarks [mm] Coax cable Double coax cable 7.2 Coax 1: black Coax 2: grey A coax cable each is needed for the measuring signal and operating voltage (chamber high voltage). The handling instructions for coax cables in the operating manual should be followed to avoid losses in insulation. For technical measuring reasons, the connections S (signal electrode) and U (operating voltage) in the connection head of the ionisation chambers may not be interchanged. In the case of ionisation chambers delivered ex works with connection cables the coax cable for the operating voltage is marked with a red marking. In double coax cables the - inner - grey coax conductor carries the operating voltage. After mounting the cables the connection-room cover and cable entries must be closed carefully to guarantee permanently the necessary insulation resistance of > 10¹² Ω. 7.3 Handling of Unusable Ionisation Chambers To counteract potential explosion hazards, ionisation chambers that have become unusable should be depressurised by opening the pump connection in the connection room. Special measures must be taken when disposing of ionisation chambers with a beryllium window, e.g. special protective measures. See the chapter "Hazardous Materials" in the document "Safety Information Measuring System" for further information. 20/20 Vers.: HW_IK_0104_en
OilSET-1000 (12 VDC)
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: +358 29 006 260 Fax: +358 29 006 1260 Internet: www.labkotec.fi 20.03.2009 1/11 OilSET-1000 (12 VDC) Oil Separator Alarm Device Copyright 2009
More informationOilSET Installation and Operating Instructions. Oil Separator Alarm Device
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: +358 29 006 260 Fax: +358 29 006 1260 18.11.2010 Internet: www.labkotec.fi 1/10 OilSET-1000 Oil Separator Alarm Device Copyright 2010 Labkotec
More informationOilSET Installation and Operating Instructions. Oil Separator Alarm Device with SET/DM3AL sensor
Labkotec UK Ltd Adminicle House 1 Lumb Lane Audenshaw Manchester M34 5WH GREAT BRITAIN Tel: 0844 3350 477 Fax: 0161 4281 179 E-mail: info@labkotec.co.uk 10.8.2012 Internet: www.labkotec.co.uk 1/13 OilSET-1000
More informationGaseous detectors. 1
Gaseous detectors Very important class of detectors with many applications charged particle, x- & gamma rays, visible light detection applications in many areas of research, & commercial particle & nuclear
More informationSET-2000 Hi Level/Oil
Labkotec Oy Labkotie 1 FI-36240 KANGASALA FINLAND Tel: + 358 29 006 260 Fax: + 358 29 006 1260 13.03.2008 Internet: www.labkotec.fi Alarm Device for Oil Separators Copyright 2008 Labkotec Oy We reserve
More informationSET-2000 Oil/Sludge 12 VDC
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel: + 358 29 006 260 Fax: + 358 29 006 1260 12.2.2015 Internet: www.labkotec.fi 1/14 SET-2000 Oil/Sludge 12 VDC Alarm Device for Oil Separators with
More informationFebruary 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.
PHY357 Lecture 11 Experimental Methods Accelerators Particle Interactions Particle Detectors Full experiment (eg. ATLAS) Introduction! Several different general classes of particle detectors (sensors)!
More informationß-Aerosolmonitor FHT 59 C
Technical Specification ZT-240-1103 E page 1/11 1. FHT 59C Moving Filter Aerosol Monitor 1.1 General The FHT 59 C is a unique moving filter aerosol monitor that can operate for one year or longer without
More informationUV Flame Supervision System
7 783 UV Flame Supervision System DETACTOGYR LFE50 Series 02 ISO 9001 The LFE50 is a self-checking UV flame supervision system designed for use with continuously operating burners or for burners running
More informationOPTIBAR P 1010/2010 C Supplementary instructions
OPTIBAR P 1010/2010 C Supplementary instructions Pressure transmitter Equipment category II 1G / Ga, II 1D / Da in protection type intrinsic safety Exi KROHNE CONTENTS OPTIBAR P 1010/2010 C 1 Safety instructions
More informationExperimental Particle Physics PHYS6011 Joel Goldstein, RAL
Experimental Particle Physics PHYS6011 Joel Goldstein, RAL 1. Introduction & Accelerators 2. Particle Interactions and Detectors (1/2) 3. Collider Experiments 4. Data Analysis Charged Particle Detectors
More informationSET-2000 Oil/Sludge. Alarm Device for Oil Separators. Installation and Operating Instructions
SET-000 Oil/Sludge Alarm Device for Oil Separators Copyright 007 Labkotec Oy We reserve the right for changes without notice TABLE OF CONTENTS GENERAL... INSTALLATION... 4. SET-000 Oil/Sludge Control Unit...
More informationFlame detector. The QRA4.U is for use with the following types of burner control:
7 711 Flame detector QRA4.U UV flame detector for use with burner controls from Siemens for the supervision of gas or oil flames. The QRA4.U and this Data Sheet are intended for use by OEMs which integrate
More informationAND INSTALLATION MANUAL
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel. +358 29 006 260 Fax +358 29 006 1260 Internet: www.labkotec.fi 10.10.2013 SET/DM3AL Level sensor OPERATION AND INSTALLATION MANUAL 1(5) SYMBOLS
More informationFlame Safeguards LFE10. Building Technologies HVAC Products
7 71 Flame Safeguards LAE LFE Flame safeguards for burners with intermittent operation The LAE is used for the supervision and indication of oil flames The LFE is used for the supervision and indication
More informationGMA Panels with double busbar
L1 L2 L3 GAS-INSULATED SWITCHGEAR >> GMA Panels with double busbar Operating Instructions No. AGS 532 521-01 Edition 09/07 (Supplement to Operating Instructions AGS 531 521-01) Technical Instructions AREVA
More informationFlame detector. The QRA4.U is for use with the following types of burner control:
7 711 Flame detector QRA4.U UV flame detector for use with burner controls from Siemens for the supervision of gas or oil flames. The QRA4.U and this Data Sheet are intended for use by OEMs which integrate
More informationIT system floor-standing distribution
IT system floor-standing distribution cabinet series -IPS-F/EDS for supplying power to medical locations in accordance with IEC 60364-7-710 and featuring a built-in insulation fault location system IT
More informationAIR or STACK MONITOR NOBLE GAS, GROSS BETA-GAMMA PARTICULATE, & IODINE. Optional Alpha, Tritium, Carbon-14
FEATURES: THREE to SIX CHANNEL SIMULTANEOUS MEASUREMENT OF: BETA-GAMMA PARTICULATE IODINE NOBLE GAS ALPHA PARTICULATE TRITIUM CARBON-14 DETECTORS MORE SENSITIVE BY FACTORS OF 5-500 THAN REQUIRED BY NRC
More informationLGA... Gas Burner Controls. Siemens Building Technologies Landis & Staefa Division
7 418 ISO 9001 Gas Burner Controls LGA... The LGA... are used for the startup and supervision of atmospheric gas burners of low to medium capacity, without fan assistance, in intermittent operation. The
More informationQRA4.U. Flame Detector. Siemens Building Technologies HVAC Products
7 711 Flame Detector QRA4.U Flame detector for use with burner controls from for the supervision of gas or oil flames. The QRA4.U and this Data Sheet are intended for use by OEMs which integrate the flame
More informationSpecification Document. Color-changing Recovery Type Water Leakage Sensor AD-RS
TS-KS-W11608-1 To: Specification Document Color-changing Recovery Type Water Leakage Sensor AD-RS Date:( ) System Equipment Division Electronic Materials & System Equipment Group TATSUTA
More informationArchive. LC2200 Level Controller. 1. Safety information. 2. General product information. 3. Installation. 4. Setting up the controller
4025250/10 IM-P402-77 AB Issue 10 LC2200 Level Controller Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Setting up the controller 5.
More informationRadiation Safety issues for the PF-AR in KEK
Radiation Safety issues for the PF-AR in KEK H. Nakamura, Y. Namito, K. Iijima,K. Takahashi and S. Ban High Energy Accelerator Research Organization (KEK) Contents Present status of KEK Management of radiation
More informationPRODUCT CONFORMITY CERTIFICATE
PRODUCT CONFORMITY CERTIFICATE This is to certify that the FlowCERT Lite with db3 transducer and remote temperature sensor or db3 transducer with twin Pulsar sun shields using internal temperature compensation
More informationFlame detector FD 2011 / FD 2012
only for existing plants EN 298 / EN 230 2003 19 PCB 3HE / 6TE DIN 41612 2-chanal system continuous operation ionisation with common or separate spark and supervision electrode intermitted UV-mode integrated
More informationKOSPEL S.A. UL. OLCHOWA KOSZALIN
Used product can't be treated as general communal waste. Disassembled appliance has to be delivered to the collection point of electrical and electronic equipment for recycling. Appropriate utilisation
More informationTechnical Manual Ionisation Chamber MUSIC80
Technical Manual Ionisation Chamber MUSIC80 Entwicklungsbüro Dr. R. Schneider Lärchenstraße 9 85640 Putzbrunn Tel: 089 608 533 71 Fax: 089 608 533 72 E-Mail: robert@schneider.net Entwicklungsbüro Andreas
More informationAdditional Operating Instructions SITRANS F. Vortex flowmeters. SITRANS FX330 Ex-i.
Additional Operating Instructions SITRANS F Vortex flowmeters Ex-i Edition 09/2018 CONTENTS 1 Safety instructions 3 1.1 General notes... 3 1.2 EU conformity... 3 1.3 Approval according to the IECEx scheme...
More informationChemistry Instrumental Analysis Lecture 4. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 4 Atomic Spectroscopy Atomic Spectroscopy UV to IR Optical spectroscopy are based on: absorption fluorescence phosphorescence scattering emission chemiluminescence
More informationLFE50. UV Flame Safeguard. Siemens Building Technologies HVAC Products DETACTOGYR. Series 02
7 783 ISO 9001 DETACTOGYR Flame Safeguard Series 02 The... together with the QRA50M / QRA51M form a self-checking flame supervision system (DETACTOGYR ) designed for use with continuously operating oil
More informationBattery-operated, hand-held alarm and measuring instruments for radiation protection
Battery-operated, hand-held alarm and measuring instruments for radiation protection The GRAETZ alarm dosemeter ED150 warns the task force when the defined radiation dose for the operation is reached.
More informationELECTRIC CENTRAL HEATING FLOW BOILER
ELECTRIC CENTRAL HEATING FLOW BOILER EKCO.A1 Used product can t be treated as general communal waste. Disassembled appliance has to be delivered to the collection point of electrical and electronic equipment
More informationTechnical Data. General specifications. Rated operating distance s n 5 mm
0102 Model Number Features 5 mm non-flush Usable up to SIL 2 acc. to IEC 61508 Technical Data specifications Switching function Normally closed (NC) Output type NAMUR Rated operating distance s n 5 mm
More informationEl63 Laser Acceleration at the NLCTA
El63 Radiation Personnel Protection System Design Document El63 Laser Acceleration at the NLCTA October 12,2005 Subject: Personnel Protection System (PPS) Access Control'System for the El63 Experimental
More informationCompetency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery
Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery Trainee: TRAINEE Module: Radiotherapy Physics Competency:
More informationLaser Distance Sensor Type M Analog-Output
Laser Distance Sensor Type M Analog-Output Operating Manual Version 2.3, May 2009 ELAG Elektronik AG l Stegackerstrasse 14 l CH-8409 Winterthur Contents 1. Safety 1.1 Laser safety 1.2 Electrical safety
More informationBe (g) Be + (g) + e - O (g) O + (g) + e -
2.13 Ionisation Energies Definition :First ionisation energy The first ionisation energy is the energy required when one mole of gaseous atoms forms one mole of gaseous ions with a single positive charge
More informationT3/E3 Fiber Optic Line Drivers
MT618A-ST-R3 January 2004 MT619A-ST-R2 MT620AE-R2 T3/E3 Fiber Optic Line Drivers CUSTOMER SUPPORT INFORMATION Order toll-free in the U.S.: Call 877-877-BBOX (outside U.S. call 724-746-5500) FREE technical
More informationWind Alarm Instrument x.x0x
THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA - THE WORLD OF WEATHER DATA Instruction for Use 000/0/0 Wind Alarm Instrument..0x.x0x ADOLF THIES GmbH & Co. KG Hauptstraße 0 Göttingen Germany Box
More information1.7 A-ISOMETER IRDH575. RoHS. Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:
Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 505 Grünberg Tel.: 060 807-0 Fax: 060 807-59 A-ISOMETER IRDH575 Insulation monitoring device for unearthed AC, DC and AC/DC systems (IT systems) with
More informationCapacitive leakage detectors of the Leckmaster range for installation in normally dry rooms
Capacitive leakage detectors of the Leckmaster range for installation in normally dry rooms Jola Spezialschalter GmbH & Co. KG Klostergartenstr. 11 67466 Lambrecht (Germany) Tel. +49 6325 188-01 Fax +49
More informationBeam Loss Position Monitoring with Optical Fibres at DELTA
Beam Loss Position Monitoring with Optical Fibres at DELTA Frank Rüdiger M. Körfer (DESY) W. Göttmann (HMI Berlin) G. Schmidt (DELTA) K. Wille (DELTA) 24. June 2008 Frank Rüdiger 1 Table of Content 1.
More informationEN - english. Instructions for installation and operation. Mobile humidity meter DP 207
EN - english Instructions for installation and operation Mobile humidity meter DP 207 Dear customers, Thank you very much for deciding in favour of the mobile humidity measuring instrument DP 207. Please
More informationSIMRAD GD10PE IR GAS DETECTOR
SIMRAD GD10PE IR GAS DETECTOR Operating Manual P3404E 1st edition Jan. 2004 www.simrad-optronics.com CONTENTS 1. PRODUCT DESCRIPTION... 5 1.1 General description... 5 1.2 Construction... 5 1.3 Application
More informationEngineering Product Guide Ionisation Smoke Detectors Optical Smoke Detectors Resettable Heat Detectors
Engineering Product Guide Ionisation Smoke Detectors Optical Smoke Detectors Resettable Heat Detectors SERIES 60 CONVENTIONAL FIRE DETECTORS Series 60 is a range of conventional fi re detectors, designed
More informationEUROFLAT SERIES TYPES C... S... M...
EUROFLAT SERIES TYPES C... S... M... AUTOMATIC CONTROL SYSTEMS FOR GAS BURNERS AND GAS BURNING APPLIANCES WITH OR WITHOUT FAN APPLICATION The electronic systems of this range are suitable to control gas
More informationSeries 65 Product Guide MAN 3036
Series 65 Product Guide MAN 3036 TABLE OF CONTENTS Page No. 1 Ionisation Smoke Detector... 1 1.1 Operating Principles:... 1 1.2 Product Codes... 2 1.3 Integrating Version:... 2 1.4 Options:... 2 1.5 Safety
More informationTIMS TIMS. Thermal Ionisation Mass Spectrometry.
TIMS Thermal Ionisation Mass Spectrometry www.nu-ins.com Incorporating our field-proven Zoom lens multi-collector technology, the TIMS improves the versatility and overall performance of this long established
More informationCERTIFICATE of Product Conformity (QAL1 )
ATÜVF}heinland Precisely Right. CERTIFICATE of Product Conformity (QAL1 ) Certificate No.: 0000059868 AMs designation: Manufacturer: Test Laboratory: FIDAMAT 6 MEASURING SYSTEM fortoc SIEMENS AG Östliche
More informationFlame detector. Building Technologies Division Infrastructure & Cities Sector
7 711 Flame detector QRA4 UV flame detector for use with burner controls from Siemens for the supervision of gas or oil flames. The QRA4... and this Data Sheet are intended for use by OEMs which integrate
More informationOPERATING INSTRUCTIONS MAXIMAT LW CX COMPACT LEAKAGE SENSOR
Safety Precautions Installation, initial start-up and maintenance may only be performed by trained personnel! The device may only be connected to power which complies with the specifications included in
More informationSIMRAD GD10 IR GAS DETECTOR
1 SIMRAD GD10 IR GAS DETECTOR Operating Manual 5th Edition P3359E October 1999 1 CONTENTS Section Page 1. PRODUCT DESCRIPTION 3 1.1 General description 3 1.2 Construction 3 1.3 Application areas 4 1.4
More informationSET Installation and Operating Instructions. Level switch for one sensor
Labkotec Oy Myllyhaantie 6 FI-33960 PIRKKALA FINLAND Tel.: +358 29 006 260 Fax: +358 29 006 1260 7.11.2013 Internet: www.labkotec.fi 1/14 SET-1000 Level switch for one sensor Copyright 2013 Labkotec Oy
More informationThe computer for optimum climate in every pig house
MC 34 H The computer for optimum climate in every pig house 1. Spray cooling channel Because of their modular construction, spray cooling channels are appropriate for nearly every house type with an intermediate
More informationLECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry
LECTURE 11 Dr. Teresa D. Golden University of North Texas Department of Chemistry C. Detectors Properties of Detectors -quantum counting efficiency -linearity -energy proportionality -resolution C. Detectors
More informationInstruction Manual. Alarm Unit For Low Gas Level # Read manual before use! Observe all safety information! Keep manual for future use!
Mess-, Regel- und Überwachungsgeräte für Haustechnik, Industrie und Umweltschutz Lindenstraße 20 74363 Güglingen Telefon +49 7135-102-0 Service +49 7135-102-211 Telefax +49 7135-102-147 info@afriso.de
More informationWrite the full electron configuration of the Al 2+ ion.
Q1.(a) Use your knowledge of electron configuration and ionisation energies to answer this question. The following diagram shows the second ionisation energies of some Period 3 elements. (i) Draw an X
More informationTitle: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection
Procedure No: SOP-020 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, and to
More informationToday s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21
Today s Outline - September 12, 2016 C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016 1 / 21 Today s Outline - September 12, 2016 Detectors C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016
More informationSERIES 65 ENGINEERING PRODUCT GUIDE
SERIES 65 ENGINEERING PRODUCT GUIDE Wide voltage conventional detectors Series 65 incorporates well-proven sensing technologies, including an IC based on that used in XP95 analogue addressable detectors.
More informationWIDE RANGE SPEAKER INSTRUCTION MANUAL TABLE OF CONTENTS
INSTRUCTION MANUAL WIDE RANGE SPEAKER CS-64BS TABLE OF CONTENTS 1. SAFETY PRECAUTIONS... 2 2. GENERAL DESCRIPTION AND FEATURES... 3 3. IMPEDANCE CHANGE... 3 4. WIRING 4.1. When No Bridge Connection is
More information7800 SERIES EC7823/RM7823 Relay Module
7800 SERIES EC7823/RM7823 Relay Module FEATURES SPECIFICATION DATA APPLICATION The Honeywell EC7823/RM7823 is a microprocessor based Flame Detector Relay that can be fitted with any 7800 SERIES amplifier
More informationINSTRUCTION MANUAL TABLE OF CONTENTS
INSTRUCTION MANUAL WIDE RANGE SPEAKER CS-154BS TABLE OF CONTENTS 1. SAFETY PRECAUTIONS... 2 2. GENERAL DESCRIPTION AND FEATURES... 3 3. IMPEDANCE CHANGE... 4 4. WIRING 4.1. When No Bridge Connection is
More informationUV - flame detector dual fuel burners for intermittent burner operations
KLC 11 1(formerly known as KLC 1001) Technical information UV - flame detector KLC 11 for oil -, gas- and dual fuel burners for intermittent burner operations 0085 Description The KLC 11 is a compact UV
More informationAnalogue addressable devices
Apollo XP95 Analogue addressable devices from Channel Safety Systems Apollo XP95 Analogue addressable devices The XP95 range of analogue addressable fire detectors is advanced in design, improved in performance
More informationELRO Cooking Island Range 600
Installation and Initial Operating Instructions ELRO Cooking Island Range 600 Date 10.07.2015 Replaces 03.06.2013 Author PM Responsible Doc. No. VAE Sig VAE X101703E Revised 1 Pages 24 Subject to modification
More informationVERSAFLOW VORTEX Supplementary instructions
VERSAFLOW VORTEX Supplementary instructions Vortex flowmeter Equipment category II 2G CONTENTS VERSAFLOW VORTEX 1 Safety instructions 3 1.1 General notes... 3 1.2 EC conformity... 3 1.3 Approval according
More informationDocumentation. Explosion-Protection. for Smoke Detection Panels SDS-4, SDS-8, SDS-12 and Extension Panels SDS-E8, SDS-E12, SDS-E16, SDS-EC16
safetec Documentation Explosion-Protection for Smoke Detection Panels SDS-4, SDS-8, SDS-12 and Extension Panels SDS-E8, SDS-E12, SDS-E16, SDS-EC16 in Smoke Detection System SDS-48 according IEC 60079-25
More informationSL EX.4.51D
GRUNDFOS DATA BOOKLET SL1.1.1.75.EX.4.51D Grundfos pump 962669 Thank you for your interest in our products. Please contact us for more information, or visit our website https://www.lenntech.com/grundfos/sl1/962669/sl1-1-1-75-ex-4-51d-c.html
More informationApplication Note. Application Note for BAYEX
Application Note Application Note for BAYEX Preface This application note provides the user a more detailed description of the Bayesian statistical methodology available in Version 8.05 and above, of the
More informationUse. Burner controls for use with forced draught gas or gas / oil burners of small to medium capacity in intermittent operation.
7 2 SO 900 Burner Controls Burner controls for use with forced draught gas or gas / oil burners of small to medium capacity in intermittent operation. The burner controls are certified to E 230 and E 298.
More informationSpecification Document. Heat-resisting Type Water Leakage Sensor FR-AD
TS-KS-W1164-8 To: Specification Document Heat-resisting Type Water Leakage Sensor FR-AD Date:( ) System Equipment Division Electronic Materials & System Equipment Group TATSUTA Electric
More informationFlame Monitor for Multiple Burner Furnaces D-LE 603 D-UG 660
Selectivity Adjustable by Means of - High Pass Frequency Filter - Gain - Threshold Different Adjustments for Different Conditions (e.g. According Fuel or Combustion Technology) Possible, to be Activated
More informationElectronic Personal Dosemeter
Electronic Personal Dosemeter Tomoya Nunomia Hideshi Yamauchi Tetsuo Shibata 1. Introduction Since developing an electronic personal dosemeter that uses a semiconductor detector in 1983, Fuji Electric
More informationReplaced by A5367CA. DISCONTINUED PRODUCT SMOKE DETECTOR WITH INTERCONNECT AND TIMER FEATURES
Data Sheet 260.8 TIMER START I / O LOW-V SET TIMER OUT + SUPPLY TIMING RES. FEEDBACK 2 3 4 5 6 7 V DD LOGIC OSC. & TIMING DRIVER 8 V SS 9 ABSOLUTE MAXIMUM RATINGS (Voltages are referenced to V SS ) Supply
More informationSystem Accessories, LCD Annunciators
System Accessories, LCD Annunciators UL, ULC Approved* A4606 Series Color Touchscreen LCD Annunciators for 4007ES Fire Alarm Control Panels Features Fig 1: A4606-9202 LCD Annunciator with Red Trim Fig
More informationPresenceLight 180A-KNX.. PresenceLight 360A-KNX.. PresenceLight 180B-KNX.. PresenceLight 360B-KNX..
Presence detector PresenceLight 180A-KNX.. PresenceLight 360A-KNX.. PresenceLight 180B-KNX.. PresenceLight 360B-KNX.. D F GB E I NL S N FIN DK Bedienungsanleitung 2 Notice d utilisation 32 Operating Manual
More informationLH-20S (RESISTANCE ALARM SYSTEM)
PORTABLE MOISTURE DETECTOR LH-20S (RESISTANCE ALARM SYSTEM) USER MANUAL levr Ver. 2.6 1. Moisture Measurements of Pre-Insulated Heating Pipelines with the Resistance Alarm System For the purpose of measurement,
More informationLAE10 LFE10. Use. Series 02
ISO 00 Flame Safeguards 7 7 LAE L Series 02 Supplementary data sheets 772 and 773 Flame safeguards for burners with intermittent operation. For safety reasons - self-test of flame supervision circuit,
More informationQFM21... Duct sensors. Symaro. for relative humidity and temperature
1 864 1864P01 Symaro Duct sensors for relative humidity and temperature QFM21... Operating voltage AC 24 V / DC 13.5...35 V Signal output DC 0...10 V / 4 20 ma for relative humidity Signal output DC 0...10
More informationUV - flame detector KLC 10 for oil -, gas- and dual fuel burners for intermittent burner operations
KLC 10 Technical information UV - flame detector KLC 10 for oil -, gas- and dual fuel burners for intermittent burner operations Description The KLC 10 is a compact UV flame detector, which has been developed
More informationPressure sensors. Ex ia I / IIC T6 acc. to ATEX. Features. Description. Measuring ranges. Applications
Force Pressure Temperature Switch Pressure sensors Ex ia I / IIC T6 acc. to ATEX with internal diaphragm with front flush diaphragm Accuracy: 0,25% and 0,5 % Standard output: 4...20 ma; 2-wire system Description
More informationUse. The burner controls are certified to EN 230 and EN 298.
7 421 SO 9001 Burner Controls G2... Burner controls for the supervision of single- or two-stage forced draught gas or gas / oil burners of small to medium capacity (typically up to 350 kw) in intermittent
More informationISO 3999 INTERNATIONAL STANDARD. Radiation protection Apparatus for industrial gamma radiography Specifications for performance, design and tests
INTERNATIONAL STANDARD ISO 3999 Second edition 2004-12-15 Radiation protection Apparatus for industrial gamma radiography Specifications for performance, design and tests Radioprotection Appareils pour
More informationISOMETER IRDH375. Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems)
Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems) IRDH375_D00124_01_D_XXEN/07.2018 Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems) Product
More informationCONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:
LEARNING OBJECTIVES: 2.17.01 List the factors which affects an RCT's selection of a portable contamination monitoring instrument. 2.17.02 Describe the following features and specifications for commonly
More informationFlame Safeguards LFE10. Building Technologies HVAC Products
7 71 Flame Safeguards LAE LFE Flame safeguards for burners with intermittent operation The LAE is used for the supervision and indication of oil flames The LFE is used for the supervision and indication
More informationLGA... Gas Burner Controls. Siemens Building Technologies HVAC Products
7 418 Gas Burner Controls LGA... The LGA... are used for the startup and supervision of atmospheric gas burners of small to medium capacity (without fan) in intermittent operation. The LGA... and this
More informationTECHNICAL MANUAL EDITION 1.0 VANDAL RESISTANT AUDIO AND VIDEO DOOR ENTRY SYSTEMS
TECHNICAL MANUAL EDITION 1.0 VANDAL RESISTANT AUDIO AND VIDEO DOOR ENTRY SYSTEMS PAGE 2 of 24 AUDIO/VIDEO VANDAL RESISTANT TECHNICAL MANUAL VER1.0 CONTENTS PAGE Manual introduction 4 System introduction
More informationLC2200 Level Controller Installation and Maintenance Instructions
4025250/9 IM-P402-77 AB Issue 9 LC2200 Level Controller Installation and Maintenance Instructions LC2200 ALARM TEST >50% >100%
More informationNEXT GENERATION DRINKING WATER RADIATION SAFETY MONITOR
FEATURES: WORLD S ONLY PAG-LEVEL WATER MONITOR MEASURES AT OR BELOW EPA/DHS PAG LEVELS Protective Action Guideline Levels and Military Drinking Water Limits REAL TIME, IN-LINE, CONTINUOUS DETECTS ALPHAS,
More information25 TEXAS ADMINISTRATIVE CODE Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines
25 TEXAS ADMINISTRATIVE CODE 289.228 Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines Texas Regulations for Control of Radiation (effective July 1, 2000) 289.228(a)
More informationSAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR
SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations
More informationDSF xx15.xx xhv Ex-Atex Ex-Atex Certified Two wire Hall Effect Speed Sensor
DSF xx15.xx xhv Ex-Atex Product ID Type # Product # Drawing # DSF 1215.01 SHV Ex-atex (5m) 304Z-05197 113216 DSF 1415.01 AHV Ex-atex S148 304Z-05256 113351 DSF 1615.01 SHV Ex-atex (5m) 304Z-05196 113214
More informationISOMETER IRDH275BM-7 with coupling device AGH675-7 and AGH675-7MV15
with coupling device AGH675-7 and AGH675-7MV15 Device combination for insulation monitoring in unearthed AC, AC/DC and DC power systems (IT systems) IRDH275BM-7_D00123_03_D_XXEN/01.2017 with coupling device
More informationCOMPANY PROFILE. DETECTION & MEASUREMENT OF IONIZING RADIATIONS RADIOLOGICAL MONITORING SYSTEMS HEALTH AND ENVIRONMENTAL SAFETY
monitoring for safety COMPANY PROFILE DETECTION & MEASUREMENT OF IONIZING RADIATIONS RADIOLOGICAL MONITORING SYSTEMS ENVIRONMENTAL MONITORING HEALTH AND ENVIRONMENTAL SAFETY www.rtsinstruments.com 2 monitoring
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 62495 Edition 1.0 2011-04 Nuclear instrumentation Portable X-ray fluorescence analysis equipment utilizing a miniature X-ray tube INTERNATIONAL ELECTROTECHNICAL COMMISSION PRICE
More informationDifferential Pressure and Flow Meter Media 5 Indicator 160 PN 40
Differential Pressure and Flow Meter Media 5 Indicator 160 PN 40 Application Instrument designed to measure and indicate differential pressure or measured variables derived from it For gases or liquids
More informationSafety instructions VEGAPULS 64
Safety instructions Intrinsic safety "i" 4 20 ma/hart - two-wire Document ID: 53001 Contents 1 Area of applicability... 3 2 Important specification in the type code... 3 3 Different ignition protection
More information