1 SDAPharma Symonds Davies Associates AUTOCLAVE CHART CHECKING & TESTS
2 AUTOCLAVE CHART CHECKING All charts are usually subject to routine checks and reviews against critical parameters, but tendency to focus on the defined limits for the sterilization phase only for temperature, pressure and absence of alarms. However it is essential to check the complete autoclave cycle profile to determine whether it is typical of the Master Process profile.
3 AUTOCLAVE CHART CHECKING Check time, values and number of negative pulses Atypical traces may indicate leakage, vacuum pump issues
4 AUTOCLAVE CHART CHECKING Check time, values and number of positive pulses Atypical traces could indicate leakage, issues with steam supply, control sensors, valves etc.
5 AUTOCLAVE CHART CHECKING Check for extended heating times Could indicate failure to remove air, vacuum issues, leakage, presence of NCGs in steam
6 AUTOCLAVE CHART CHECKING Check for sterilization temperature, pressure and time. Failure to maintain specified conditions could indicate issues with steam supply, control sensors etc.
7 AUTOCLAVE CHART CHECKING Check vacuum drying phase for atypical values in temperature, time and depth of vacuum Atypical trace could indicate vacuum pump problems, leakage etc.
8 SCHEDULE OF PERIODIC TESTS FOR POROUS LOADS HTM2010, EN285 Daily/Weekly: Bowie-Dick test for steam penetration, Vacuum leak test, Air detector function test, Automatic Control Test Quarterly: Verification of calibration probes, Thermometric test for a small load. Six Monthly: Thermometric tests for full loads, Steam non-condensable gas test, Steam superheat test, Steam Dryness test The schedule is more extensive than those listed above. The daily and weekly tests are extremely valuable together with standard chart checking to detect faults and issues which may otherwise go undetected until full thermometric testing.
9 BOWIE DICK STEAM PENETRATION TEST To steam sterilize items, steam must come into direct contact with every surface of each item to destroy any microbes that may be present. Getting steam to penetrate though packaging, pouches, tubing etc can be difficult as any air present can become trapped within the load and act as an insulating barrier. The Bowie-Dick test consists of a test pack which has a chemical indicator sheet located in the centre. Following a Bowie-Dick sterilization cycle, colour changes on the sheet indicate whether satisfactory air and other non-condensable gases have been removed and steam has penetrated through the test pack. The classic test consisted of freshly laundered towels, but many commercially prepared kits are now available e.g. Steris, 3M etc. The tests may provide important information regarding the ongoing performance of the sterilizer.
10
11 BOWIE DICK STEAM PENETRATION TEST Unused sheet
12 BOWIE DICK STEAM PENETRATION TEST PASS: Cycle Satisfactory Uniform colour change from yellow to blue/purple. Consistently achieving this result demonstrates satisfactory ongoing performance. Note it is important to store and handle sheets correctly to avoid false results.
13 BOWIE DICK STEAM PENETRATION TEST FAIL: Possible Cause Inadequate air removal 1. Faulty vacuum pump incapable of drawing a deep enough vacuum. 2. The water temperature to the vacuum pump is too high preventing a good vacuum in the chamber. 3. A faulty pressure switch/sensor giving an incorrect reading.
14 BOWIE DICK STEAM PENETRATION TEST FAIL: Possible Cause Leak 1. A leak in the chamber door seal owing to wear or damage. 2. A leak in the sensor ports attached to the chamber. 3. A high pressure leak from a pneumatic valve into the steam supply to the chamber. 4. A venturi leak through pipe-work where steam is passing at high velocity.
15 STEAM QUALITY TESTS EN285 These should be undertaken on a six monthly basis or whenever there is an interruption in the steam supply to assure steam quality. To be most effective, the Steam supplied to autoclaves must be dry saturated (composed of water vapour in a state of equilibrium between condensation and evaporation). The release of latent heat as the steam condenses to moisture on the item to be sterilized causes the coagulation of cellular proteins causing the destruction of micro-organisms. Routine Bowie-Dick testing may indicate issues which would otherwise not be detected until the periodic thermometric and biological testing and calibration checks are performed.
16 STEAM QUALITY TESTS EN285 Test for Non-Condensable Gas This includes air and other gases which will not condense to liquid state and release latent heat under sterilization conditions. During sterilization the flow of steam is always towards a component. As any gases are liberated at the point of condensation, NCGs will be forced by the flow of steam towards the product. If the product is hollow or porous, the air will be forced towards the centre of the product. Air is approximately 12,000 times more resistant to heat transfer than copper. Whether present as a film or pocket, it may prevent direct steam contact or insulate the component. Such conditions are identical to inadequate air removal. A sample is taken from the steam supply to the autoclave and condensed. The level of NCG is determined by comparing the amount of water displaced in a burette against the steam condensed. The maximum permitted level is 3.5% gas/condensate.
17 BOWIE DICK TEST STEAM QUALITY FAIL: Possible Cause Non Condensable Gas (NCG) in the steam supply 1. Boiler Feed Water Treatment NCGs are gasses found in steam which do not condense when the steam turns to water. Source is usually at the point of generation or in the distribution system 2. Issues with the Still feed water 3. Leak
18 STEAM QUALITY TESTS EN285 Superheat Test: Excessive amounts of superheat could mean that there is no moisture available to effectively sterilize, leading to dry heat conditions. i.e. steam will not condense and provide moisture until the steam temperature has reduced to the saturation temperature. It therefore acts as hot air or dry heat. The test requires the fitting of an expansion tube onto the pitot tube used in the Dryness Test. A temperature sensor measures the superheat if any as the steam expands to atmospheric pressure. Degrees of superheat measures in free steam at atmospheric pressure shall not exceed 25.
19 BOWIE DICK TEST STEAM QUALITY FAIL: Possible Cause Superheated Steam 1. Steam supply pressure is too high. 2. Chamber jacket is being operated at a temperature higher than the chamber during sterilization. 3. Badly engineered steam supply pipe-work resulting in high steam velocities.
20 STEAM QUALITY TESTS EN285 Dryness Test Excess condensate in the steam can cause can cause an initially dry load to be taken out wet. This can make the load much more vulnerable to recontamination. In addition water can act as an insulator and prevent good heat transfer. The dryness value is determined on a sample of steam which passes through a pitot tube located in the steam supply and into cold water in a flask. The initial and final mass and temperature of the water in the flask is used to derive the value. The value should not be less than 0.90 or 0.95 when sterilizing metal loads
21 BOWIE DICK TEST STEAM QUALITY FAIL: Possible Cause Poor Quality Wet Steam 1. Inadequate baffling in boilers 2. Undersized boiler for application 3. Poorly lagged steam pipe-work 4. No condensate trap near sterilizer 5. Cold chamber jacket due to fault 6. Jacket operating at reduced temperature 7. Test has been carried out in a cold chamber
22 VACUUM LEAK TEST (AKA AIR LEAKAGE TEST) To ensure high standards of performqnce for porous load autoclaves, leakage into the chamber while it is under vacuum, must be controlled. Leakage before and during sterilization will affect sterilization efficacy. Leakage following sterilization may contaminate the load. To assess and monitor leakage EN285 recommends a minimum vacuum of 70mbar absolute is created within the chamber. When the desired vacuum value is attained, the main control values to the chamber are closed and the vacuum pump stops. After a 5 minute stabilization period, the decay in the chamber vacuum is monitored for the following ten minute period. The maximum permitted decay is 1.3mbar per minute over 10 minutes.
23 AIR DETECTOR TEST The Air Detector is a device located within the region of the chamber drain for demonstrating that air in the chamber environment and throughout the load under worst case conditions has been reduced to an acceptable level on each and every cycle. It measures the presence of the last remaining air in the load. There are two types: Temperature Sensing and Pressure Sensing. The devices are set up and calibrated during qualification and are very sensitive to air within the load. They are far superior to the drain probe in detecting the presence of air (from leaks, steam supply or the load itself). A review of log books may reveal periodic, infrequent failures.
24 RECOMMENDED SCHEDULE OF PERIODIC TESTS FOR POROUS LOAD AUTOCLAVES TEST Air Detector Test Air Detector Function Test Leak Rate Test Bowie Dick Test Pack Equilibration Time Vent filter Integrity Test Thermometric Test Thermometric and BI Test FREQUENCY Every Cycle Monthly Daily Weekly 6 months 12 months 6 months 12 months
25 SDAPharma Symonds Davies Associates ANY QUESTIONS?
International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: Vol.9, No.3, pp , 2016
International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: 0974-4304 Vol.9, No.3, pp 374-386, 2016 Sterilization of Porous Loads by Hphv Steam Sterilizer : A Review R.G. Mukunda, N. Vishal
More informationBS En 285:2006+a2:2009 Description: Sterilization.
BS En 285:2006+a2:2009 Description: Sterilization. Steam sterilizers. Large sterilizers Steam sterilizers, Sterilizers, Large, Sterilization (hygiene), Medical equipment, Equipment safety, Safety measures,
More informationESTEEM AUTOCLAVE SLIDING SINGLE DOOR:-
ESTEEM AUTOCLAVE SLIDING SINGLE DOOR ESTEEM AUTOCLAVE SLIDING SINGLE DOOR:- ESTA-108A For Sterilization of glass wares / containers, vessels, linen, rubber articles culture media etc. Sliding doors with
More informationSteam Quality Test Kits. Product Code: SQTK A to L. Description: Datasheet : 040
Page 1 Steam Quality Test Kits SQTK A to L The range of Steam Quality Test Kits from Thermal Detection consists of seven kits which cover all of the measurement and reporting requirements in accordance
More informationGeneral Introduction Introduction to sterilization
General Introduction Introduction to sterilization Course content and objectives This course is composed of different modules which may be presented in their entirety or separately with users gaining an
More informationAdvance Autoclave. Designed to give you more for your money with a high capacity for faster throughput: More features and benefits...
Advance Advance Designed to give you more for your money with a high capacity for faster throughput: The Advance has a much bigger capacity than its main competitors and yet has one of the smallest footprints
More informationSBS Design Considerations for Autoclave Sterilization
SBS Design Considerations for Autoclave Sterilization Gregg A. Mosley Biotest Laboratories, Inc. Minneapolis, MN March 2-4, 2010 www.healthpack.net Information and Time How to put 2 days of information
More informationSTEAM STERILIZER SERIE S1000 (RBE)
STEAM STERILIZER SERIE S1000 (RBE) INTRODUCTION: Of Turkish manufacturing, our sterilizers Series S1000 (RBE) are designed for the purpose of inactivating and sterilizing infectious waste generated in
More informationLoad type and process/autoclave selection
Load type and process/autoclave selection PDA Europe Basics and Recent Developments for Autoclaves and Lyophilizers - Tel Aviv/Israel, 13 November 2017 Sara Ferretti Fedegari Group What load can we treat
More informationAdvance Autoclave. Designed to give you more for your money with a high capacity for faster throughput: More features and benefits...
Advance Advance Designed to give you more for your money with a high capacity for faster throughput: The Advance has a much bigger capacity than its main competitors and yet has one of the smallest footprints
More informationCLINOCLAV steam sterilizer AH-21 L
CLINOCLAV steam sterilizer AH-21 L AH-21 L Optional printer Advantages: Agar mode (40-80ºC) Programmable auto-start (24 h) Control software Program for liquids Connection to PC (USB) Heart probe (optional)
More informationSteam Site Guide. Steam Site Guide. Heat Exchange Applications - SI Units
Steam Site Guide Steam Site Guide Heat Exchange Applications - SI Units Steam Site Guide Content Introduction... 03 01. Estimating Steam Loads... 04 02. Steam Table SI... 07 03. Sizing a Steam Pipe...
More informationBest Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices
Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices In BC Health Authorities THIS DOCUMENT IS INTENDED TO DESCRIBE BEST PRACTICES HEALTH
More informationDoes your equipment meet your needs and the operational capacity stated by the manufacturer?
Does your equipment meet your needs and the operational capacity stated by the manufacturer? Richard Bancroft, B.Sc (Hons) Registered Authorising Engineer (Decontamination), UK Convenor, ISO/TC 198 WG
More informationLTE Scientific Ltd. Touchclave-PL Porous Load Steam Sterilizer
LTE Scientific Ltd Touchclave-PL Porous Load Steam Sterilizer Touchclave-PL: the revolutionary new range of small capacity porous load sterilizers from LTE Scientific Developed to satisfy a growing demand
More informationFN 032/055/120 DRY HEAT STERILIZERS / OVENS FN 300/400/500 DRY HEAT STERILIZERS / OVENS KD 200/400/700 OVENS EV 018 VACUUM OVEN
FN 032/055/120 DRY HEAT STERILIZERS / OVENS Three different sizes: 32, 55 and 120 liters. Temperature range: Ambient Temperature +5 C / 250 C. Designed for sterilization, drying and heating purposes. User
More informationFLASH PAK STERILIZATION CONTAINER SYSTEM
INSTRUCTIONS FOR USE FLASH PAK STERILIZATION CONTAINER SYSTEM Description The Flash Pak Sterilization Container System consists of a family of rigid reusable containers that provide an effective sterilization
More informationWHY THE REMOVAL OF NONCONDENSABLE GASES AND AIR IS CRITICAL IN A STEAM SYSTEM
BEST PRACTICE NO. 30 WHY THE REMOVAL OF NONCONDENSABLE GASES AND AIR IS CRITICAL IN A STEAM SYSTEM Air and noncondensable gases are one of the major problems in a steam system that can cause production
More informationSterilizer Model LS2020 Series Beta Star Life Science Products of R-V Industries
SECTION 11630 - STERILIZERS PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1.2 REFERENCES 1. Small Steam Sterilizer A. The following industry, association and government codes and standards are cited
More informationSeries 151 and AV5. A wide range of process thermostatic steam traps is available to match almost any application need.
A wide range of process thermostatic steam traps is available to match almost any application need. FEATURES Simple construction Easy to maintain as installed Excellent air handling capability Pressure
More informationEfficient Steam System Design
Efficient Steam System Design The word Efficient is often used to describe the general performance of a system. However it is important to distinguish between efficiency and effectiveness. Efficiency is
More informationTitle: Measurement of dryness fraction by Separating Calorimeter,
Title: Measurement of dryness fraction by Separating Calorimeter, Throttling Calorimeter, Separating and Throttling Calorimeter. Learning objectives: 1. Intellectual skills: a) Measurement of Dryness fraction
More informationAZTECA A LITRE STERILISER
Product Description Azteca A-6615 is a fully automatic high-speed pre & post vacuum steam sterilizer. This model is an electrically heated sterilizer, which operates with saturated steam as a sterilizing
More informationNC SERIES. Horizontal Steam Sterilizers
NC SERIES Horizontal Steam Sterilizers NC 300/430/570/710 HORIZONTAL STEAM STERILIZERS Our continuous investments in research and development have enabled us to develop our horizontal steam sterilizer
More informationRemoval of Non-condensable Gases and Air Is Critical In A Steam System
Removal of Non-condensable Gases and Air Is Critical In A Steam System Air and non-condensable gases are one of the major problems in a steam system. Both can cause production problems for a steam system
More informationA WORLD OF QUALITY, SAFETY AND HYGIENE STERILCLAVE. 18 litres-bhd/b/s
A WORLD OF QUALITY, SAFETY AND HYGIENE STERILCLAVE 18 litres-bhd/b/s STERILCLAVE COMINOX: THE NEW GENERATION Boosting its unique and ergonomic design, Sterilclave Cominox is the la latest result of 20
More informationOpen the Door to Increased Productivity and Cost Efficiency
S t e a m s t e r i l i z e r s Open the Door to Increased Productivity and Cost Efficiency Amsco Century Medium Steam Sterilizers Amsco Century Medium Steam Sterilizers The Family of Sterilizers Designed
More informationYARWAY PROCESS THERMOSTATIC STEAM TRAPS SERIES 151 AND AV5
A wide range of process thermostatic steam traps is available to match almost any application need FEATURES Simple construction Easy to maintain as installed Excellent air handling capability Pressure
More informationChapter 2.3: Steam System
Part I: Objective type Questions and Answers Chapter 2.3: Steam System 1. For industrial process heating, the best quality of steam is: a) dry saturated steam b) superheated steam c) wet steam d) high
More informationLow Temperature Sterilisation ISO 14937
Low Temperature Sterilisation ISO 14937 Validation and Performance Jon Mulhall Senior Validation Engineer ASP Johnson & Johnson Medical 27 th March 2018 What is ISO 14937? A (very) brief history.. In the
More informationSteam Trap Presentation for IDEA Distribution Workshop. Presenter: Joe Radle
Steam Trap Presentation for IDEA Distribution Workshop Presenter: Joe Radle Proper Piping when Reducing Pipe Size Concentric reducer Eccentric reducer Correct Wrong Typical Steam Main Drip Station D Size
More informationPICNet CONFERENCE 2013
PICNet CONFERENCE 2013 SCOPING THE HORIZON Penny Brawn THE ORIGINAL STEAM STERILIZERS ORIGINAL STEAM STERILIZER A LITTLE HISTORY Back in 1450 BC in Moses time, writings were found in books of Leviticus
More informationENDOSCOPE DECONTAMINATION : Where are we now!
ENDOSCOPE DECONTAMINATION : Where are we now! Christina Bradley Laboratory Manager Hospital Infection Research Laboratory City Hospital Dudley Road, Birmingham CHOICE OF DECONTAMINATION METHOD Depends
More informationThe Saturation process
SOUTH METROPOLITAN TAFE WA The Saturation process Dennis Kenworthy 8/5/2016 A student study guide to measuring and interpreting the saturation process of refrigeration and air-conditioning equipment System
More informationA M E M B E R O F T H E K E N D A L L G R O U P
A M E M B E R O F T H E K E N D A L L G R O U P Presented by: Dan Gallagher Energy Opportunities in a Steam System What makes up your Steam System Boiler Room Distribution System Mechanical Rooms Point
More informationCHAPTER 3 BASIC STEAM CYCLE
CHAPTER 3 BASIC STEAM CYCLE To understand steam generation, you must know what happens to the steam after it leaves the boiler. A good way to learn the steam plant on your ship is to trace the path of
More informationAMSCO C Series Small Steam Sterilizers 16 x 16" (406 x 406mm) 20 x 20" (508 x 508 mm)
OPERATOR MANUAL AMSCO C Series Small Steam Sterilizers 16 x 16" (406 x 406mm) 20 x 20" (508 x 508 mm) Prevacuum Steam Flush Pressure Pulse (05/07/12) P129394-176 A WORD FROM STERIS CORPORATION This manual
More informationEconomical and simplified solutions that offer quality and optimum steam trap service.
Economical and simplified solutions that offer quality and optimum steam trap service. It All Comes Down To Protection Regardless of the temperature and pressure characteristics of steam line drips or
More informationREFRIGERANT CHANGEOVER
Date of last update: Dec-17 Ref: CC7.26.5/0117-1217/E Application Engineering Europe REFRIGERANT CHANGEOVER FROM HFC R134a TO HFC/HFO R450A & R513A WARNING Use only Emerson approved refrigerants and lubricants
More informationFOB THIRD-GENERATION BENCH-TOP SATURATED-STEAM STERILIZERS
The key difference between a small vertical or bench-top sterilizer and industrial sterilizer is performance. Small saturated-steam sterilizers with vertical or horizontal chambers were devised many years
More informationlaboratory & sterilization technology
laboratory & sterilization technology NC 23B - NC 23S/32S BENCH TOP STEAM STERILIZERS Small steam sterilizers are widely used for the sterilization of the products used for medical purposes, e.g in general
More informationAIR CONDITIONING. Carrier Corporation 2002 Cat. No
AIR CONDITIONING Carrier Corporation 2002 Cat. No. 020-016 1. This refresher course covers topics contained in the AIR CONDITIONING specialty section of the North American Technician Excellence (NATE)
More informationAccurately With HSE COP PT 1 5
Medical Device Decontamination and Validation (Mr. Hugh O'Connor) Medical Device Decontamination and Validation Learning to Work Efficiently and Accurately With HSE COP PT 1 5 Introduction for HSE COP
More informationB+N 18/23L LED. Ritter CleanTec with LED Display BB 18 + BB 23. Sterilizable Products:
Ritter CleanTec with LED Display BB 18 + BB 23 The CleanTec is available with an 18 to 23 liter chamber size and is a stand-alone autoclave with built-in water storage tank. It works with a fractionated
More informationPRE AND POST VACUUM CLASS B AUTOCLAVE
PRODUCT CATALOG PRE AND POST VACUUM CLASS B AUTOCLAVE www.biolabscientific.com PRE AND POST VACUUM CLASS B AUTOCLAVE Reliable and robust class B autoclaves intended to deliver highest quality of sterilization.
More informationSteam Trap Inspection Methods And Steam Cost Analysis
Steam Trap Inspection Methods And Steam Cost Analysis 14 Hayes Street, Elmsford NY 10523-2536 USA Tel: 914-592-1220 Toll Free USA & Canada 24 Hour Fax Email: Internet 800-223-1325 914-347-2181 info@uesystems.com
More informationStrategies in Optimizing Condensate Return
Strategies in Optimizing Condensate Return By D. Bloom, Sr. Consultant, Ondeo Nalco Company, Naperville, IL Optimizing condensate return for reuse as boiler feedwater is often a viable means of reducing
More informationLSII /51 STERILIZER MACHINE SPECIFICATION A REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH
LSII-262639/51 STERILIZER MACHINE SPECIFICATION A-83-37630 REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH Table of Contents 1.0 PART I GENERAL... 3 1.1 Vessel Volume... 3
More informationOT 300/430/570 STEAM STERILIZERS
OT 300/430/570 STEAM STERILIZERS OT 300/430/570 More than 15 years had passed since the very first steam sterilizer manufactured by Nuve. In course time, we had developed and been manufacturing thirteen
More informationOPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific. (07/16/07) P Rev. 0
OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific (07/16/07) P129373-635 Rev. 0 A WORD FROM STERIS CORPORATION This manual contains important information on proper
More informationYARWAY PROCESS THERMODYNAMIC STEAM TRAPS SERIES 40/40D AND C250/260
SERIES 40/40D AND C250/260 Designed for a variety of high pressure and high capacity applications found in utility, industrial and marine service FEATURES Designed to fail open. Designed for superheat
More informationOT 300/430/570/710/850 STEAM STERILIZERS
OT 300/430/570/710/850 STEAM STERILIZERS OT 300/430/570/710/850 More than 15 years had passed since the very first steam sterilizer manufactured by Nüve. In course time, we had developed and been manufacturing
More informationPOWDER GUIDE. Trouble shooting tips to maximise the benefits of Apcoshield Powder Coatings
POWDER GUIDE From ASIAN PAINTS (I) LTD Trouble shooting tips to maximise the benefits of Apcoshield Powder Coatings Section A : Powder Application Problems Trouble Possible Causes Solutions 1. Poor charging
More informationNUMBER: R&C-ARF-4.0 ISSUED: REV. A: REV. B: REV. C: REV. D: REV. E: 2.1. Frequency of Contact and Non-Contact Bedding Changes:
PAGE 1 OF 6 1. BACKGROUND ARF Personnel are responsible for all routine maintenance of the ARF, which includes cleaning and polishing floors, managing the storage areas, cleaning restrooms and offices,
More informationLaboratory Line. Vertical and Bench-Top Life Science Autoclaves
Laboratory Line Vertical and Bench-Top Life Science Autoclaves The Advanced Laboratory Line The Full Spectrum of Laboratory Sterilizers Sterilization in a laboratory environment has its unique requirements.
More informationOPERATION & MAINTENANCE MANUAL
OPERATION & MAINTENANCE MANUAL Pre/Post-vacuum Class B Table top Autoclave Model ELARA 11 (CAT Controller) Cat. No. MAN205-0342001EN Rev Q Manufactured by: Tuttnauer Co. Ltd., Har Tuv Industrial zone B
More informationVertical Laboratory Autoclave A series:
Vertical Laboratory Autoclave A series: of Labtron Vertical Laboratory Autoclave LVA-A series is a top loading sterilizer which provides safe, Made up of SUS304 stainless steel Dual scale numerical pressure
More informationREVIEW ON CONDENSER AIR LEAK TEST
Volume 119 No. 16 2018, 3197-3205 ISSN: 1314-3395 (on-line version) url: http://www.acadpubl.eu/hub/ http://www.acadpubl.eu/hub/ REVIEW ON CONDENSER AIR LEAK TEST 1 Neethu Sathyan M, 2 Dr.Golda Dilip,
More informationSTERITITE INSTRUCTIONS FOR USE
STERITITE INSTRUCTIONS FOR USE If a rigid container system is used, the manufacturers instructions regarding set preparation and assembly should be followed. -ANSI/AAMI ST79: 2010 & A1:2010 Notes from
More informationNC SERIES. Bench Top Steam Sterilizers
NC SERIES Bench Top Steam Sterilizers NC 23B - NC 23S/32S BENCH TOP STEAM STERILIZERS Small steam sterilizers are widely used for the sterilization of the products used for medical purposes, e.g in general
More informationMYSTICOOL Max Valve System with Xstream and A.R.M.E.D. Technology Service & Installation Instructions Page 1
Page 1 WHY should I install the MYSTICOOL Max Valve System? XDX is more efficient, saving on power consumption. Use of XDX system decreases defrost cycles. XDX maintains more consistent product temperatures,
More informationWATERLIFE VACUUM VACUUM CLEANING SYSTEM FOR MODIFIED ALCOHOL
WATERLIFE VACUUM VACUUM CLEANING SYSTEM FOR MODIFIED ALCOHOL TYPE OF EQUIPMENT The washing machines for metals Waterlife Series Vacuum machines are the single room type. In a single processing station
More informationOff Design Operation of Hybrid Noncondensable Gas Removal Systems for Flash Steam Cycle Geothermal Power Plants
GRC Transactions, Vol. 40, 2016 Off Design Operation of Hybrid Noncondensable Gas Removal Systems for Flash Steam Cycle Geothermal Power Plants Jason Project Engineer, Vooner FloGard Corporation, Charlotte
More informationKronos Class B Sterilizer
Kronos Class B Sterilizer NOW WITH OUR EXCLUSIVE * PATENTED TECHNOLOGY Meet the NEW Newmed Kronos Class B Sterilizer Good looking and elegant on the outside, but reliable and uncomplicated on the inside,
More information3.1 Deaerator Systems
3.1 Deaerator Systems 3.1.1 Design & Operational Considerations System Overview The primary function of a deaerator system is to protect the boiler and condensate system from corrosion by removing oxygen
More informationPrinciples of Steam-In-Place
Principles of -In-Place ing-in-place and integrity testing of a sterilizing-grade filter assembly ing-in-place (SIP) is a widely adopted method in the pharmaceutical industry for the in-line sterilization
More informationGESTRA Information A 1.7
Issue Date: 9/06 GESTRA Systems GESTRA Information A 1.7 Automatic Deaeration of -Heated Heat Exchangers Preface For optimum efficiency of steam-heated heat exchangers several requirements should be fulfilled,
More informationTABLE TOP - STEAM STERILIZER Dental & Hospital Application
TABLE TOP - STEAM STERILIZER Dental & Hospital Application THE CISA GROUP COMPANY COMPANY PROFILE CISA has been manufacturing and selling sterilization systems for over 60 years for both hospitals and
More informationYARWAY NON-REPAIRABLE DRIP AND TRACER STEAM TRAPS SERIES PB, 29, 129Y AND 29S
SERIES PB, 29, 129Y AND 29S Economical and simplified solutions that offer quality and optimum steam trap service FEATURES 29S 29 Thermostatic traps Pressure assisted fail-open design Freeze proof Easy
More informationFS 231: Final Exam (5-6-05) Part A (Closed Book): 60 points
Name: Start time: End time: FS 231: Final Exam (5-6-05) Part A (Closed Book): 60 points 1. What are the units of the following quantities? (10 points) a. Enthalpy of a refrigerant b. Dryness fraction of
More informationLarge Steam Sterilizers for Life Sciences
Large Steam Sterilizers for Life Sciences Large Steam Sterilizers for the Life Sciences Laboratory Line Large Steam Sterilizers for the Life Sciences Sterilization in a laboratory environment has its unique
More informationTMCI Padovan Evaporators
TMCI Padovan Evaporators Evaporators Our range includes 4 types of evaporators: Forced Circulation Falling Film Evaporators Plates Thin Film Concentration systems Evaporator choosing criteria: product
More informationN-Wissen GmbH. Top Loading Autoclaves (95, 120, 135 L) Available top loading models: The Top Loading Autoclave range features:
Top Loading Autoclaves (95, 120, 135 L) The Top Loading Autoclave range features: A choice of 63, 95, 120 and 135 litres models A fully programmable 5,7 Touch Screen Colour Controller Timed/Pulsed Free
More informationPOWER TRANSFORMER OIL PROCESSING SPX TRANSFORMER SOLUTIONS, INC.
Regional Technical Seminar POWER TRANSFORMER OIL PROCESSING SPX TRANSFORMER SOLUTIONS, INC. June 20, 2018 Agenda 1. Oil Processing Theoretical Information Transformer Drying Theory Insulation Media 2.
More informationPERFORM Operating Document
PERFORM Operating Document Use and Maintenance of Tuttnauer Table-Top Autoclave PC-POD-CA-006-v03 Revision History Version Reason for Revision Date 01 New POD 13-Aug-13 02 POD Section 3.1.3 and 3.2 revised
More informationThe Professional s Choice
SureFlow compact range: 1000-C & 1000-M Auto-Fill Catering Water Boiler Installation And User Instructions The Professional s Choice CONTENTS CONTENTS & INTRODUCTION 2 SAFETY & ENVIRONMENTAL INFORMATION
More informationO P E R A T I N G I N S T R U C T I O N S. K S G FLASH - 2 Table top unit
O P E R A T I N G I N S T R U C T I O N S K S G FLASH - 2 Table top unit with microprocessor command Modifications reserved. 01.06.10 Article No.: Aukl 002 96 - left Order-No.: 184763 Serial-No.: 7897
More informationProtection. in un unica autoclave E10
Euronda EN Tutte le qualità Pro System Products, che cercavi, Process, Protection in un unica autoclave E10 Much more than just an autoclave E10 Euronda Pro System Products, Process, Protection Cleaning
More informationHow to Trap: Trap Selection and Safety Factors
How to Trap: Trap Selection and Safety Factors Bringing Energy Down to Earth Say energy. Think environment. And vice versa. Any company that is energy conscious is also environmentally conscious. Less
More informationThe right trap for the right application.
The right trap for the right application. Whenever you re faced with a process application, it s not enough to just select any steam trap to solve your problem. Instead, remember to match your application
More informationHow to Trap: Process Air Heaters
How to Trap: Process Air Heaters 1 Designs, materials, weights and performance ratings are approximate and subject to change without notice. Visit for up-to-date information. Bringing Energy Down to Earth
More informationAn Approach to Energy Efficiency: A Focus on an Increased Operational Efficiency, Natural Resource Conservation and Carbon Footprint Reduction
An Approach to Energy Efficiency: A Focus on an Increased Operational Efficiency, Natural Resource Conservation and Carbon Footprint Reduction A White Paper for the Enercon Venturi Condensate Removal (EVCR)
More informationالمملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية
1. POLICY STATEMENT: Cleaning, decontamination and sterilization method should be done in CSSD service facility not in the end user department side. 2. PURPOSE: To ensure the proper way of collections,
More informationFOF INDUSTRIAL SATURATED STEAM STERILIZERS
PROCESS In the early 1950s, the pharmaceutical industry entered a highly dynamic stage of evolution which required machines and production process that had been unthinkable up to then. Fedegari immediately
More informationSST2200 & Stainless Steel Steam Sterilizers SST1700 &
SST2200 & Stainless Steel Steam Sterilizers SST1700 & Your new assistants Stainless Steel Steam Sterilizers a new complete line of small steam sterilizers. C.B.M. has merged its own experience in steam
More informationBiological Safety Cabinets. Presented by Mr. Rene Soetens President
Biological Safety Cabinets Presented by Mr. Rene Soetens President Biological Safety Cabinets are referred to by several different names, some of the more common are: BSC, Biosafety Cabinet, Bio-Hood and
More informationGas Fired Humidifier. SKG Series Combustion Field Adjustment Instructions. SKG Combustion Field Adjustment Instructions
Fired Humidifier SKG Series SKG - 180301 Standard Combustion Adjustment SKG Fired Humidifier All installation work must be carried out by suitably qualified personnel and must conform to local codes and
More informationBoiler Condensate Return Basics. Presented by Steve Connor February 24, 2016
Boiler Condensate Return Basics Presented by Steve Connor February 24, 2016 What We Are Covering Today? Brief review of the total steam system Why a proper condensate return system is so important Condensate
More informationTechnical Description. for the. Autoclave. Thank you very much for the trust which you have shown by purchasing this autoclave.
Technical Description for the Autoclave Vacuklav 23 B Dear Doctor: Thank you very much for the trust which you have shown by purchasing this autoclave. For 50 years now, MELAG a medium-sized family-owned
More informationCentral Sterile Services Department will receive, decontaminate, process, store and/or dispense sterile and non-sterile to departments requesting.
Title/Description: Central Sterile Services Department: Central Sterile Services Personnel: Central Sterile Services Personnel Effective Date: 1/97 REVISED: 1/97, 11/07, 2/12 I. PURPOSE Central Sterile
More informationOT SERIES. Bench Top Steam Sterilizers
OT SERIES Bench Top Steam Sterilizers OT 18/23B OT 23/32VS OT 23S BENCH TOP STEAM STERILIZERS Small steam sterilizers are widely used for the sterilization of the products used for medical purposes, e.g
More informationIntroducing the latest model in Prestige Medical s first-class range of Advance autoclaves - Advance Pro.
Introducing the latest model in Prestige Medical s first-class range of Advance autoclaves - Advance Pro. Advance Pro is a Class B sterilizer available in 16 or 22 litre capacity units, they offer superior
More informationEliminating Condensation and Resulting Corrosion. Sulphur Tail Gas and De-Gas Lines
Eliminating Condensation and Resulting Corrosion In Sulphur Tail Gas and De-Gas Lines By David R. Hornbaker, Controls Southeast, Inc. and Graham M. MacPherson, P.Eng., Millenia Resource Consulting Abstract
More informationManual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches
Caution Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine content.
More informationPrepared by Technical Team
Prepared by Technical Team Summary Vacuum degassing (VD) and vacuum oxygen de-carburisation (VOD) are the main processes in secondary steel making. The large volumes of dissolved contaminant gases arising
More informationAir Compressor Network compressed air (standard) R Compressor (built-in the equipment) R Compressor (separately)
STEAM STERILIZER HI SPEED II SELECTION STIC Chamber useful dimension (average) B0111-100 - 406 x 406 x 630 mm B0111-210 - 408 x 408 x 1285 mm B0111-300 - 605 x 605 x 880 mm B0111-500 - 605 x 605 x 1285
More informationSteam sterilizer AHS-75-DRY
Steam sterilizer AHS-75-DRY Optional PT-2 probe AHS-75-DRY Optional printer Advantages: Agar mode (40-80º C) Programmable auto-start (24 h) Connection to PC (USB) Control software (optional) Heart probe
More informationDIRECTIVE NO: D-B
DIRECTIVE NO: D-B6 100604 1 LOW PRESSURE THERMAL FLUID PLANT AUTOMATED CONTROL SYSTEMS Date of Issue: June 4, 2010 General Details This directive is being issued to owners, licensed contractors, consulting
More informationLAB ID and/or LABORATORY NAME: ASSESSOR NAME: ELECTRET by Voltametry. Method Number: ELAP method number SOP Number: Revision Number: SOP Date:
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: ELECTRET by Voltametry Method Number: ELAP method number 7034 SOP Number: Revision Number: SOP Date: Personnel / Data Records observed: General LLD:
More informationInstruction Manual DIGITAL MANIFOLD FOR HVAC/R SYSTEMS
English Instruction Manual DIGITAL MANIFOLD FOR HVAC/R SYSTEMS 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Instruction Manual
More information