1 Regulation of Alarm Remote Notification Systems Shawn Forrest Center for Devices and Radiological Health Food and Drug Administration Medical Device Alarms Summit October 5, 2011
2 Evolving Clinical Environment Central Alerting Bedside Alerting Remote Alerting 2
3 What are Secondary Alarm Notifications? Intended to supplement Primary notifications User reliance is intended to be on Primary notifications, typically from Bedside monitor Central station 3
4 Related Example Classification Regulation: 21 CFR Cardiac Monitor (including cardiotachometer and rate alarm) Product Code: MSX Physiological Monitors Network and Communication System Classification: Class II (requires 510(k)) No current classification specific to secondary alerts 4
5 FDA Device Review Premarket notification (510(k)) demonstrates substantial equivalence to a predicate device Devices are cleared/approved only for their labeled intended use Secondary alarms are an additional layer of protection Primary alarms are assumed to be functional 5
6 Evolving Clinical Environment Central Alerting Bedside Alerting Intended use of devices should be considered in the monitoring paradigm. Remote Alerting 6
7 Postmarket Experience MAUDE reports provide insights on potential remote notification problems Networking hardware failures Hardware conflicts Spontaneous system reboots Network traffic precludes/delays delivery Receiver loses connection or connects to wrong network (and doesn t notify) Receiver batteries unexpectedly depleted 7
8 Design Considerations Validated network requirement specs Acceptable latency Closed-loop system Backup routing Receiver Ensures constant connection to correct network Ensures battery reliability Logs alarms and connection to assist troubleshooting End-to-end compatibility 8
9 Device Alarm Review Verification and Validation based on level of risk In the case of alarm systems, the risk of missing a critical alarm should guide testing Key aspects: Wireless & EMC validation Software validation Alarm-related standards 9
10 RF Wireless Technology Draft Guidance Document 10
11 Wireless Medical Device Validation Performance Quality of Service Latency Data Integrity Electromagnetic Compatibility Wireless Coexistence Security 11
12 Software Validation Guidance 12
13 Software Validation Guidance 13
14 Software Validation Guidance 14
15 Software Validation for Remote Alarm Notification Controlled development process Hazard analysis for alarms should guide design Requirements and design specifications should adequately address alarm reliability Validation should include comprehensive assessment of potential alarm failure modes 15
16 Related Recognized Standards IEC Ed. 2:2006, General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems AAMI / ANSI HE75:2009, Human factors engineering - Design of medical devices 16
17 Potentially Related Initiative Mobile Medical Applications Draft Guidance Document Public Workshop held 9/12-13, all comments due by 10/19 orkshopsconferences/ucm htm 17
18 Unrelated Initiative Medical Device Data Systems (MDDS) Final rule issued Feb 15, 2011 to reclassify from Class III (default) to Class I Does not include devices intended to be used in connection with active patient monitoring. Includes only systems that transfer, store, convert from one format to another, or display medical device data. 18
19 Conclusion Monitoring systems providing Secondary remote notifications should be implemented based on their Intended Use. Reliance should be on the primary alarms (bedside and central station) 19
June 17, Skanray Technologies Pvt Ltd. c/o Ms. Yolanda Smith Regulatory Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 17, 2015 Skanray
More informationFebruary 23, Skanray Technologies Private Limited Yolanda Smith, Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service February 23, 2016 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Skanray
More informationTOUBRO LIMITED ELECTRICAL & ELECTRONICS DIVISION - ELECTRONIC PRODUCTS 510(K) SUMMARY. (Per section )
LARSEN & TOUBRO LIMITED SEP 2 1 2O Mysore Complex, KIADB Industrial Area, Hebbal -Hootagalli, Mysore -570 018. Tel : +91 (821) 2402561 Fax: +91(821) 2402468 E -Mail: Ref: 30 th March 2007 Page: 01 of 04
More informationLc~rlI. z - 510(k) Summary JUL
JUL - 2 2008 Premarket Notification 510(k) Submission-5 10(k) Summary Report No. Report No. TE20071I01 I Lc~rlI 510(k) Summary z - This summary of 510(k) safety and effectiveness information is being submitted
More informationE. GentleLASE Family of Laser Systems 510k Summary JL1
General Information: E. GentleLASE Family of Laser Systems 510k Summary JL1 This 5 10(k) is to provide notification of substantial equivalence for the Candela GentleLASE Family of Laser Systems, which
More informationRecent Topics in IEC TC 62 (Electrical Equipment in Medical Practice) and its Subcommittees
INTERNATIONAL ELECTROTECHNICAL COMMISSION Recent Topics in IEC TC 62 (Electrical Equipment in Medical Practice) and its Subcommittees Don Heirman presenting for Robert Sitzmann (TC 62) APEMC symposium
More informationB. Company Address: N. Kendall Dr. Miami, Florida 92121
VIII. PREMARKET NOTIFICATION 510K SUMMARY SUBMITTER INFORMATION A. Company Name: Drip Alert, Inc. B. Company Address: 13882 N. Kendall Dr. Miami, Florida 92121 C. Company Phone: 305-385-8000 Company Fax:
More informationTRUSTED PRECISION. A HISTORY OF INNOVATION. Valleylab FT10 Energy Platform Product Information Kit
TRUSTED PRECISION. A HISTORY OF INNOVATION. Valleylab FT10 Energy Platform Product Information Kit A HERITAGE YOU CAN TRUST The Valleylab energy portfolio You can have confidence in almost 50 years of
More informationRecent Topics in IEC TC 62 (Electrical Equipment in Medical Practice) and its Subcommittees
INTERNATIONAL ELECTROTECHNICAL COMMISSION Recent Topics in IEC TC 62 (Electrical Equipment in Medical Practice) and its Subcommittees Don Heirman presenting for Robert Sitzmann (TC 62) APEMC symposium
More informationAutomated Medication Kit
Volume-5, Issue-2, April-2015 International Journal of Engineering and Management Research Page Number: 491-496 Automated Medication Kit V. L. Charishma 1, K. Harika 2, A. Kavya 3, Mr. T. P. Kausalya Nandan
More informationRisk Management of Electromagnetic Disturbances
Ottawa Tutorial FRI AM 2 Introduction to Medical EMC Risk Management of Electromagnetic Disturbances Keith Armstrong, EurIng, CEng, FIET, Senior MIEEE, ACGI phone & fax: +44 (0)1785 660 247 keith.armstrong@cherryclough.com
More informationRFT CARES Android App User Guide
RFT CARES Android App User Guide 2017 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any form or by
More informationNurse Call Systems & Emergency Call Systems. Overview of ANSI/UL 1069 and ANSI/UL 2560
Nurse Call Systems & Emergency Call Systems Overview of ANSI/UL 1069 and ANSI/UL 2560 The Association of Electrical and Overview National Consensus Standards American National Standards Institute (ANSI)
More informationReporting Changes to your Notified Body
Reporting Changes to your Notified Body Ibim Tariah 1) Requirements & Guidance 2) Reporting Changes - What, When, Why & How 3) Some examples 4) Summary & Questions 2 Requirements of the Medical Directives
More informationAlarm Fatigue in the ICU. Lynn Maguire MS, RN, PMC, NE-BC
Alarm Fatigue in the ICU Lynn Maguire MS, RN, PMC, NE-BC Objectives At the completion of this lecture the participant will: Define Alarm Fatigue Discuss the Joint Commission s Patient Safety Goal regarding
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-1-2 Third edition 2007-03 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral standard: Electromagnetic compatibility
More informationBattery Performance Alert
Battery Performance Alert A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR S UNDER BATTERY ADVISORY VERSION 2.0 Abstract BACKGROUND: In October 2016, St. Jude Medical (now Abbott) issued an advisory on a family
More informationAustralian/New Zealand Standard
AS/NZS 3200.2.4:2006 IEC 60601-2-4:2005 AS/NZS 3200.2.4:2006 Australian/New Zealand Standard Medical electrical equipment Part 2.4: Particular requirements for safety Cardiac defibrillators AS/NZS 3200.2.4:2006
More informationTable of Contents for the REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 5 April 2017
Chapter I o Scope and definitions Article 1 Subject matter and scope 13 Article 2 Definitions 15 Article 3 Amendment of certain definitions 20 Article 4 Regulatory status of products 20 Chapter II o Making
More informationImproving Alarm Management Utilizing Technology. NJ/DVHIMSS Annual Fall Event. Empowering the New Healthcare Era
Improving Alarm Management Utilizing Technology NJ/DVHIMSS Annual Fall Event Empowering the New Healthcare Era 1 2 How big a deal is this? Up to 350 alarms per day per patient! Ranks #1 on ECRI s 2014
More informationMedical devices and electromagnetic compatibility. Addressing the requirements of IEC , 4 th Edition
Medical devices and electromagnetic compatibility Addressing the requirements of IEC 60601-1-2, 4 th Edition Executive Summary Risk of electromagnetic interference and immunity have long been important
More informationBattery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR DEVICES UNDER BATTERY ADVISORY
Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR S UNDER BATTERY ADVISORY VERSION 1.0 AUGUST 8, 2017 Abstract: BACKGROUND: In October 2016, St. Jude Medical issued an advisory on a
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SONY GLOBAL MANUFACTURING & OPERATIONS CORPORATION EMC/RF TEST LABORATORY - MAIN LAB 8-4 Shiomi Kisarazu-shi Chiba-ken, 292-0834, Japan Mr. Koichi.Hiyaoka Phone:
More informationACTIVE INFRARED BARRIER
Although PROTECH provides high security indoor intrusion sensors for the military and government markets, our specialty is outdoor protection. Since we first introduced our PIRAMID outdoor dual technology
More informationDEFA HOME. User Manual Base Unit PRO
DEFA HOME User Manual Base Unit PRO Table of Contents Overview 5 Logging into the app 9 Home 10 Calendar 13 Adding an event 14 Editing an event 15 Zones 16 Switch mode 17 Thermostat mode 19 Regulator
More informationAS/NZS IEC :2016
Australian/New Zealand Standard AS/NZS IEC 60601.2.25:2016 (IEC 60601-2-25, Ed. 2.0:2011, IDT) Medical electrical equipment Part 2.25: Particular requirements for the basic safety and essential performance
More informationDevelopments in Laser Safety
DOE LSO Workshop 2018 Developments in Laser Safety ANSI Z136 Standards Standards Regulations Series Update William J. Ertle Barbara Sams & Z136.1 Philosophy Standards Director, IEC/TC 76 Secretary Laser
More informationHospital Ward Alarm Fatigue Reduction Through Integrated Medical Device Instruction and Hospital System Policy
Virginia Commonwealth University VCU Scholars Compass VCU da Vinci Center Student Works VCU da Vinci Center 2014 Hospital Ward Alarm Fatigue Reduction Through Integrated Medical Device Instruction and
More informationUltraview SL2400 With Wireless-LAN-Option
Ultraview SL2400 With Wireless-LAN-Option Spacelabs Healthcare Connecting Innovation with Care Spacelabs Healthcare is fully committed to making strong connections: between clinicians and their patients;
More informationLighting Xpert Insight User Manual
Lighting Xpert Insight User Manual Table of Contents 1 About This Document... 3 1.1 Key Terms... 3 1.2 Related Fifth Light Documentation... 3 2 Lighting Xpert Insight Overview... 4 2.1 Key Features...
More informationTouchpad Exit Controller Administration Guide
Touchpad Exit Controller Administration Guide 2018 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any
More informationSituational Awareness for Life & Fire Safety. October 10, 2013
Situational Awareness for Life & Fire Safety October 10, 2013 New Fire Standards Control and Communication Units For Fire Alarm Control Units Mass notification products shall be constructed to comply with
More informationAlarm Management. Objectives
Alarm Management Michelle Bell, RN, BSN, FISMP, CPPS Kelly Graham, BS, RN Objectives 1. Discuss the importance of appropriate alarm use 2. Describe strategies for improving alarm safety and how to begin
More informationEUROPEAN COMMITTEE FOR ELECTROTECHNICAL STANDARDIZATION (CENELEC) TECHNICAL COMMITTEE No. 62: ELECTRICALEQUIPMENT IN MEDICAL PRACTICE
TC62/SEC1015/INF October 2009 EUROPEAN COMMITTEE FOR ELECTROTECHNICAL STANDARDIZATION (CENELEC) TECHNICAL COMMITTEE No. 62: ELECTRICALEQUIPMENT IN MEDICAL PRACTICE SOFTWARE AND MEDICAL DEVICES WORKING
More informationOperation Manual Fighter ProVision Software. Version: 0.0 Revision: 1
Operation Manual Fighter ProVision Software Version: 0.0 Revision: 1 TABLE OF CONTENTS 1. Introduction 5 2. Software Installation 5 3. PC Users 6 3.1 Introduction 6 3.2 Default Code 6 3.3 Edit PC User
More informationSkyresponse ThingWorx Extension. Version 1.0. User Guide
Skyresponse ThingWorx Extension Version 1.0 User Guide Software Change Log...1 Introduction...1 About Skyresponse...1 Installation...2 Importing the Skyresponse ThingWorx Extension...2 Configuration and
More informationqube mini Overview
D a t a S h e e t qube mini Overview The 91389 qube mini is a compact, robust, and light-weight patient monitor with extensive clinical usability. Key features of the qube mini monitor include: Eye-level
More informationMyKi Care Thermometer (MCT 1.0) USER MANUAL
Precautions for use Any form of modification to this device is forbidden; Don t use the device together with MRI or CT equipment; Explosion danger: don t use the device in flammable anesthetic gas; Don
More informationMedical electrical equipment
AS IEC 60601.1.2:2017 IEC 60601-1-2:2014 AS IEC 60601.1.2:2017 Medical electrical equipment Part 1.2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic
More informationCommon Customer Challenges
About Cooper-Atkins A leading manufacturer of innovative temperature and environmental solutions Recognized as a leader in wireless technology ISO 9001:2008 registered facility Common Customer Challenges
More informationMaxiCompressor. Limited Warranty. High Performance 50 PSI Compressor 501-S
Limited Warranty Global Medical Holdings (GMH) warrants the MaxiCompressor for 1 year from the date of purchase due to faulty parts or workmanship. This warranty is limited to the original purchaser of
More informationSafety requirement compliance for Lasers
Safety requirement compliance for Lasers in Medical and Beauty Care Applications Building off UL s history of more than 120 years of safety science, UL s Health Sciences division provides comprehensive
More informationAS/NZS IEC :2016
AS/NZS IEC 60601.2.2:2016 (IEC 60601-2-2, Ed.5.0 (2009) IEC 60601-2-2:2009/COR1:2014, IDT) Australian/New Zealand Standard Medical electrical equipment Part 2.2: Particular requirements for the basic safety
More informationAmerican Nuclear Society (ANS)
What s New... American Nuclear Society (ANS) Project Initiation Notification System (PINS) ANS Standards Before work on a draft can begin, a Project Initiation Notification System (PINS) form must be approved
More informationHealthcare PROTEZIONE TOTALE PER I VOSTRI INTERESSI AZIENDALI PRESENTI E FUTURI. When every second counts, Honeywell delivers.
Healthcare PROTEZIONE TOTALE PER I VOSTRI INTERESSI AZIENDALI PRESENTI E FUTURI When every second counts, Honeywell delivers. Your Challenges are Complex. Honeywell Can Help. Evolving regulations, greater
More informationS T A T E O F F L O R I D A D E P A R T M E N T O F J U V E N I L E J U S T I C E BUREAU OF MONITORING AND QUALITY IMPROVEMENT PROGRAM REPORT FOR
S T A T E O F F L O R I D A D E P A R T M E N T O F J U V E N I L E J U S T I C E BUREAU OF MONITORING AND QUALITY IMPROVEMENT PROGRAM REPORT FOR BI, Inc. - Electronic Monitoring (Contract Provider ) 6265
More informationVeriSafe Absence of Voltage Tester The safe way to verify the absence of voltage
VeriSafe Absence of Voltage Tester The safe way to verify the absence of voltage www.panduit.com/verisafe What is an Absence of Voltage Tester? Absence of Voltage Testers (AVTs) are permanently-mounted
More informationSapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design
Sapphire TM Ambulatory Infusion System Bring Home The Future Of Infusion Care > Intuitive, easy-to-use touch screen design > Rugged, durable construction for home care infusion needs > Engineered to provide
More informationREAL TIME LABORATORY MONITORING SYSTEM
REAL TIME LABORATORY MONITORING SYSTEM MONITORING LET XILTRIX WATCH OVER YOUR LABORATORY! 20 years experience and IKS expertise makes XiltriX the total monitoring solution for any laboratory. All day,
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-1-2 2001 AMENDMENT 1 2004-09 Amendment 1 Medical electrical equipment Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility Requirements
More informationCE MARKING PROCESS & NOTIFIED BODIES
CE MARKING PROCESS & NOTIFIED BODIES Ellis Garai Ramin Miri Spring 2011 GLOBAL BIODESIGN STANFORD UNIVERSITY Machinery 9% Others 49% Construction 11% Notified Bodies Industrial and Medical Devices 8% NAWI
More informationCERTIFICATE OF ACCREDITATION
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Washington Laboratories, Ltd. 7560 Lindbergh
More informationDynAMo Alarm & Operations Management
Connected Plant DynAMo Alarm & Operations Management Solution Note Delivering operations integrity through better plant safety, availability and compliance across your entire enterprise Honeywell s DynAMo
More informationNess acomms Install & User Manual
Ness acomms Install & User Manual Phone GUI Tablet GUI 1 P age TABLE OF CONTENTS Copyright Notice... 2 Important Information... 3 Ness D8x / D16x Versions... 3 D8x / D16x Two-Way Communication Settings...
More informationTouchpad Exit Controller User Guide
Touchpad Exit Controller User Guide 2018 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any form or
More information9/20/2016 2:53 PM. Second Revision No NFPA [ Section No ] Supplemental Information. Submitter Information Verification
Second Revision No. 2011-NFPA 101-2016 [ Section No. 14.2.2.2.4 ] 14.2.2.2.4* Classroom Door Locking to Prevent Unwanted Entry. Classroom doors shall be permitted to be locked to prevent unwanted entry
More informationPATIENT MONITORING SYSTEMS
PATIENT MONITORING SYSTEMS PATIENT MONITORING SYSTEMS The benefits of HEYER Scalis to meet the market situation HEYER Scalis function & features make it fit to match the requirement of standard applications
More informationLaser Safety, Utilization and Credentialing Policy
University at Buffalo Policy Laser Safety, Utilization and Credentialing Policy Category: Date Established: to be determined Responsible Area: Clinical Affairs Date Last Revised: 11/21/2013 Responsible
More informationClinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care
Clinical Alarm Effectiveness: Recognizing & Mitigating Risk to Patient Care Thomas Bauld, PhD, CCE, Biomedical Engineer National Center for Patient Safety (NCPS) Elena Simoncini, Clinical Engineer VA Boston
More informationCode Alert Series 30 Software User Guide
Code Alert Series 30 Software User Guide 2018 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any form
More informationNew requirements for IEC best practice compliance
New requirements for IEC 61511 best practice compliance Proof testing and a SIL study may no longer be enough by Sven Grone Schneider Electric Safety Services Practice Leader Executive summary Since its
More informationCareTemp Touch Free Thermometer. Directions for use
CareTemp Touch Free Thermometer Directions for use 2017 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted
More informationDigijock Home Security
Digijock Home Security User Manual Congratulations on your purchase! The Digijock Home Security is a wireless alarm system that will protect your home or your small business from fire and burglary. Consisting
More informationThe Complexity of Clinical Alarm Systems
HTM NEWS & VIEWS The Complexity of Clinical Alarm Systems Alan Lipschultz About the Author Alan Lipschultz, CCE, PE,CSP, is president of HealthCare Technology Consulting LLC in North Bethesda, MD. E-mail:
More information1 Intended Use of the Breas HA 01 Humidifier
User Manual for Breas HA 01 Humidifier 1 Intended Use of the Breas HA 01 Humidifier Read this manual thoroughly so that you completely understand how the humidifier is operated and maintained before taking
More informationIECEE OPERATIONAL DOCUMENT
IECEE OD-2055 Edition 2.0 2015-06-03 IECEE OPERATIONAL DOCUMENT IEC System of Conformity Assessment Schemes for Electrotechnical Equipment and Components (IECEE System) Guideline Document on Medical Electrical
More informationANSI-ASQ National Accreditation Board/ACLASS
ANSI-ASQ National Accreditation Board/ACLASS SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 Washington Laboratories, Ltd. 1, 2 7560 Lindbergh Drive, Gaithersburg, Maryland 20879 John Repella Phone: 301-216-1500
More information4.2.1 Policies & Procedures. Adds wording that the LSO shall REQUIRE approved written procedures on
ANSI Z136.3, 2018 Sections Relevant to Biomedical Engineering and Service: Professional Medical Education Assn The Laser Training Institute www.lasertraining.org August 2018 DISCLAIMER: Please note that
More informationUntil very recently, it was possible to supply medical devices, equipment or systems to
COVER DISCUSSION Risk Management and the EMC Standard for Medical Devices, Equipment and Systems IEC 60601-1-2 Editions 3 and 4 Keith Armstrong CherryClough Consultants Ltd. Introduction Until very recently,
More informationAchieve Success with European Medical Device Commercialization. By Russ King
Achieve Success with European Medical Device Commercialization By Russ King Achieve Success with European Medical Device Commercialization Given the complexity of the FDA s regulatory pathway, many medical
More informationCODE ALERT Enterprise Software User Guide
CODE ALERT Enterprise Software User Guide 2018 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any form
More informationWhy Clinical Alarms are a TOP TEN Hazard
1 Why Clinical Alarms are a TOP TEN Hazard James P. Keller, Jr., M.S. Vice President, Health Technology Evaluation and Safety jkeller@ecri.org American Heart Association Conference November 6, 2012 2 An
More informationDocument Preparation Profile (DPP)
1 Document Preparation Profile (DPP) 1. IDENTIFICATION Document Category Working ID: Proposed Title: Proposed Action: Safety Guide DS471 Radiation Safety of X-ray Generators and Radiation Sources Used
More information2017/11/16 P a g e 1
2017/11/16 P a g e 1 Contents What s New... 2 Fixes... 3 Driver Updates... 3 New Sensors... 3 New Features... 3 How To... 4 Security... 4 Auditing... 6 Smart UI/MAPS Title 21 CFR Part 11 Auditing... 6
More informationStandard for the Installation of Carbon Monoxide (CO) Detection and Warning Equipment
NFPA 720 2015 Summary Standard for the Installation of Carbon Monoxide (CO) Detection and Warning Equipment TM, 7Oct14 CODE: Title, Standard, not Code Consistent with recent trend at NFPA, this document
More informationThe Centron Presidio Monitoring System. Centron. Presidio. Rees Scientific. An ISO 9001:2008 Company
The Monitoring System Rees Scientific An ISO 9001:2008 Company About Rees Scientific Rees Scientific specializes in protecting your valuable assets. Since 1982, Rees Scientific has been the industry standard
More informationTHE EDDY IQ METER PRODUCT MANUAL
THE EDDY IQ METER PRODUCT MANUAL THE EDDY IQ METER 2 CONTENTS CERTIFICATION WELCOME PRODUCT INFORMATION EDDY IQ METER FEATURES MY EDDY HOME APP SETTINGS, OPERATION & DEFINITIONS TROUBLESHOOTING CUSTOMER
More informationAn overview of sterilizer equipment standards current and planned. Richard Bancroft Science & Technical Director STERIS
An overview of sterilizer equipment standards current and planned Richard Bancroft Science & Technical Director STERIS #deconevent2018 An Overview of Sterilization & Equipment Standards Current & Planned
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60770-3 First edition 2006-04 Transmitters for use in industrial-process control systems Part 3: Methods for performance evaluation of intelligent transmitters IEC 2006 Copyright
More informationAlarm Management at Boston Medical Center A Roadmap to Safe Silence
Alarm Management at Boston Medical Center A Roadmap to Safe Silence Jim Piepenbrink, Director Department of Clinical Engineering Co-Chair, Clinical Alarm Task Force This presentation is a view into the
More informationAS/NZS IEC :2015
AS/NZS IEC 60947.5.6:2015 (IEC 60947-5-6, Ed. 1.0:1999, IDT) Australian/New Zealand Standard Low-voltage switchgear and Part 5.6: Control circuit devices and switching elements DC interface for proximity
More informationURGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010
. 600 N. Field Drive Lake Forest, IL 60045 (224) 212 2000 www.icumed.com 15 June 2017 URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010 Dear Valued Plum 360 Infusion System
More information--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY
Translated English of Chinese Standard: YY0896-2013 www.chinesestandard.net Sales@ChineseStandard.net NATIONAL STANDARD OF THE GB PEOPLE S REPUBLIC OF CHINA ICS 11.040.50 C 39 Medical electrical equipment
More informationCLETEF-786. Infrared Thermometer Instruction Manual
CLETEF-786 0482 Infrared Thermometer Instruction Manual Introduction Thank you for choosing the Clever Choice Duo TM Ear & Forehead Thermometer, by Simple Diagnostics. The infrared thermometer CLETEF-786
More informationTPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP
TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP The TPCA Ambulatory Patient Controlled (PCA) Syringe Pump with Lockbox has a library of protocols offering MediGuard ; a Dose Error Reduction System (DERS)
More informationX64 Wireless Training
X64 Wireless Training IDS Contents 1 Contents Features 3 Wireless Hardware 4 IDS & Duevi integration PCB 5 LED operation 5 Wireless Device Hardware setup 6 Location 260 7 LED Keypad Instructions 7 Adding
More informationThe WAVE Plus Instant Notification System for Schools and Colleges
Instant Notification System for Schools and Colleges Duress alarm has been activated in the School of Engineering! Schools strive to provide a safe environment for students and teachers. In an emergency,
More information2011 National Clinical Alarms Survey: Perceptions, Issues, Improvements and Priorities
2011 National Clinical Alarms Survey: Perceptions, Issues, Improvements and Priorities Healthcare Technology Foundation Clinical Alarms Task Force Presented by Tobey Clark President, Healthcare Technology
More informationNetworX NX-8V2. LED Keypad User Manual
NetworX NX-8V2 LED Keypad User Manual POWER Light is on when AC power is present; flashes to indicate a low battery condition. ARMED Light is on when armed; off when disarmed; flashes to indicate a previous
More informationArchitectural and Engineering Specification for a. Flash / Flare
Architectural and Engineering Specification for a Flash / Flare June 27, 2014 Page 1 of 12 T1DA0115-001 Rev D This document is intended to provide performance specifications and operational requirements
More informationA1UL PERS. Personal Emergency Response System. For Technical Support Please Contact Your Service Provider Or Distributor
A1UL PERS Personal Emergency Response System TABLE OF CONTENTS 1. READ THIS FIRST... 1 2. SYSTEM OVERVIEW.. 1 3. COMPONENTS 2 4. UNIT OPERATION! Standby Mode.. 3! Emergency Activation. 3! Answering Incoming
More informationThe safe way to verify the absence of voltage.
VeriSafe Absence of Voltage Tester The safe way to verify the absence of voltage www.panduit.com/verisafe What is an Absence of Voltage Tester? Absence of Voltage Testers (AVTs) are permanently-mounted
More informationAmbulatory infusion pump for continuous, intermittent, TPN or PCA delivery.
Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery. The CADD-Prizm VIP (Variable Infusion Profile) pump is designed for A NEW CADD PUMP BENCHMARK FOR AMBULATORY INFUSION CARE. We
More informationUltraview SL Command Module Parameter Set ECG NIBP TEMP IP CO SpO2 Options SL Command Module: A B Our goals are simple:
Ultraview SL2700 Spacelabs Healthcare - Connecting Innovation with Care Spacelabs Healthcare is fully committed to making strong connections: between clinicians and their patients; between technological
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SGS-CSTC STANDARDS TECHNICAL SERVICES CO., LTD Shenzhen Branch No.1 Workshop, M-10, Middle Section, Science and Technology Park Shenzhen, Guangdong, 518057,
More informationIntegration Test Plan
Integration Test Plan Terminus Security Prepared by: Kai Chan Stephen Krenzel John O Meara Version: 1.0 1 Contents 1 Introduction 3 1.1 Purpose.................................. 3 1.2 Scope...................................
More informationAS/NZS :2015. Conduit systems for cable management AS/NZS :2015. Part 1: General requirements. Australian/New Zealand Standard
AS/NZS 61386.1:2015 Australian/New Zealand Standard Conduit systems for cable management Part 1: General requirements Superseding AS/NZS 2053.1:2001 AS/NZS 61386.1:2015 AS/NZS 61386.1:2015 This joint Australian/New
More informationRice University Laser Safety Manual
Rice University Laser Safety Manual Environmental Health and Safety MS 123 P.O. Box 1892 Houston, TX 77251-1892 January 2014 RICE UNIVERSITY 1 Introduction The objective of the Rice University Laser Safety
More informationDeltaV Operate. DeltaV Operate. Introduction. DeltaV Product Data Sheet. Robust and secure plant operations
February 2016 Page 1 DeltaV Operate You can easily manage your process with. Robust and secure plant operations Powerful operator interface, optionally up to four monitors Constant visibility and access
More informationManage Alert Rules & Teams
Manage Alert Rules & Teams OneVue Environmental Monitoring Platform OneVue is a trademark of Primex. OneVue is an intelligent environmental monitoring and managed time solution. All other trademarks are
More information