Purge and Trap Troubleshooting Made Easy?

Similar documents
Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

GC Injectors. GC Injectors

TYPES AND APPROPRIATE INJECTORS

Aromatics and Oxygenates in Gasoline June2018

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

SCION 436-GC. Dimensions and Weights

Perkin Elmer Clarus 500 Gas Chromatograph

FIXED GAS ANALYZER (FGA)

On Line SPE LC/MS/MS Applications

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Velocity XPT Accelerated Purge and Trap Sample Concentrator

INJECTORS Purge & Trap General Operating Procedures continued

300 series HPLC autosamplers

MultiPurpose Sampler MPS robotic series

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

HISTORY OF VARIAN CAPILLARY INJECTORS

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

Introduction to the Fusion TOC Analyzer

Agilent GC Sampler 80/120 SPME Option

Agilent 1290 Infinity II Vialsampler Data Sheet

XENOSEP. SPE Systems for EPA Method 1664A. The Easiest, Most Cost Effective Way to Increase Productivity

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

POPULAR CONFIGURATION GCs Cryosulfur GC. System Overview

Systems. Visit us at

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

GC DETECTORS Reduction Gas Detector - RGD. Overview

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

Standard VistaNET GCC Software: 803V029-(VER)(REV)

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

Recent Improvements in Process GC Instrumentation

CyTOF Autosampler 5 User Manual

Waters 2695D Separations Module

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

Waters 2695 Separations Module

Chemistry Instrumental Analysis Lecture 29. Chem 4631

Spark Holland introduces

FPD - Flame Photometric Detector

Customised HPLC Systems for Highest Performance

Troubleshooting. Edwin A. Burgstaler, MT., HP. (ASCP) ASFA 2015 Annual Meeting May, MFMER slide-1

Genuine Restek Replacement Parts for Agilent GCs

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Agilent 7650A Automatic Liquid Sampler

Nexis GC-2030 Specifications

FACSCalibur Startup Procedures

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous

Methods of solid phase extraction. Methods of solid phase extraction

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica.

Agilent Vac Elut SPS 24 Manifold User Guide

E²M. Innovation with Integrity. Enhanced Environmental Mass Spectrometer. Safety & Security

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. TurboMartix Headspace and Headspace Trap Samplers Consumables and Supplies Reference Guide

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX

Improved Dryer Control

Gas Chromatograph. Features. Continuous Analysis of Ultra High Purity Argon to ppb levels.

On-site Detection of Siloxanes and other unwanted VOCs in Landfill/Sewage/Biogas using Gas Chromatograph-Ion Mobility Spectrometer (GC-IMS)

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

McNeil Group University of Michigan. MBraun Glovebox. Preparing Items To Be Brought Into the Glovebox

Analyze IT. Process Gas Chromatograph. PGC2000 Series

Single Tower Heat Reactivated Desiccant Air Dryer

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products

Agilent InfinityLab Online SPE Solution

Reacti-Therm I and III Heating Modules

SEC-MALS Noise Assessment and Cleaning Guide

HS Dryer. Smart Solutions for Drying Compressed Air. Single Tower Heat Reactivated Desiccant Air Dryer. Engineered Air Products

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

Sample Preparation Solid Phase Extraction Disks Cartridges. Horizon Technology, Inc. Salem, NH USA

NEOTeCHA Reactor Sampling Type PV

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

Polar Air Dryer. Instructions Manual PDRCF PDRCF R E F R I G E R A T E D T Y P E C O M P R E S S E D A I R D R Y E R S

22 The Flame Ionization Detector

(PD[ $LU 'U\HU 5 ( ) 5, * ( 5 $ 7 ( ' 7 < 3 ( & ( 6 6 ( ' $, 5 ' 5 < ( 5 6 ('5&) ದ ('5&) Instructions Manual

wash system V1.3 Important: Up to slides on one injection. Note: Getting need to turn it samples the next day. and put the slides in foil, put it

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

Training Fees 4,000 US$ per participant for Public Training includes Materials/Handouts, tea/coffee breaks, refreshments & Buffet Lunch.

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA

Chromatography. Research Supplies. 2014/2015 Catalog

- Data Brochure Steam Control 279

Application Note. Authors. Abstract

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology

THE RIGHT CONSUMABLES FOR YOUR GC APPLICATIONS. Clarus 590/690 Gas Chromatography Consumables and Supplies Reference Guide

SPME-FFA Field Sampler

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

Air Server & Multi Channel Samplers (MCS06 & MCS08)

Technical specifications of FT-IR Spectrometer

Process Gas Chromatograph

SystemsQT for Waters Alliance HPLC Systems (Revision G)

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

Transcription:

Purge and Trap Troubleshooting Made Easy? Tyler Trent- Application Chemist Preston Hoppes- International Tech Support/FSE Tammy Rellar- Application Sales Specialist Teledyne Tekmar

Purge and Trap Troubleshooting No Response Carryover/Contamination Elution issue- early, middle, late Water interference Internal Standard recovery variability Linearity/Reproducibility issues System leaks

Troubleshooting Tips Take a moment to review manuals Examine the facts and use valid reasoning Identify the root cause of the problem Avoid quick fixes and shotgun approach BE PATIENT!

Finding the Root Cause MS,GC, Concentrator, and Autosampler Split the system into 4 separate entities

Eliminate the GC/MS Not necessary to disconnect Purge & Trap (P&T) Confirm normal operation of the MS Tune parameters/vacuum Clean source Confirm normal operation of GC Do direct inject of same standard used in P&T Poor chromatography/resolution/reduced response Injection liner/septum Clip column Once GC/MS is eliminated as source of issue, proceed to P&T/ASM

Eliminate the Purge and Trap Confirm normal operation of the GC/MS Tune/direct inject Confirm normal operation of the P&T Run diagnostics/leak check Make sure the correct analytical trap is in place Manually load sample into Purge and Trap Eliminates the autosampler

No Response Are the analytes getting to the GC/MS Check the trap heater Reaches the set temperature Check 6-port valve Hear the 6-port actuate 6-port rotor installed correctly Correct installation of the 6-port valve

No Response Are the analytes getting to the trap Bubbles in the sparge tube during purging Correct flow set in method Mass flow controller working System leak tight Gas tank still have pressure

No Response Stratum

No Response Check Autosampler Is there sample in sparge vessel? Pressure problems Sample get pulled Valves working correctly Sample and drain valves

No Response AQUATek100

Carryover/Contamination Carryover for full target list/late eluters and contamination Run GC only Desorb only Check temperature Desorb/Bake Check hot water/meoh rinse

Carryover of Full Target List GC/MS Faulty EPC Purge and Trap Solenoid valve leaking cross port Faulty trap heater Sample not draining entirely Dirty glassware Autosampler Check soil vs. water if available System not rinsing correctly Internal standard valve leaking(allows standard to enter vessel) Not transferring entire aliquot to Purge and Trap

Carryover of Late Eluters GC Inlet too cold Purge and Trap Not enough bake Bad trap Cold spots Transfer line connection to GC inlet Faulty condensate or trap heater Autosampler Check soils vs. waters if available Hot water heater/rinse failure Not enough bake rinses Sample needle not cleaning up Not transferring bake rinse to purge and trap

Carryover of Late Eluters AQUATek100

Contamination Issues? No problem! Able to rinse sample pathway with 90 C water Atomx uses methanol to rinse entire pathway Includes syringe, associated transfer lines and valves, sparger and 3-stage needle. User can vary volume and number of rinse times user settable within the software method

Data 0.60 0.50 0.40 One 90 C Water Sparge Vessel Rinse 0.30 0.20 Two 90 C Water Sparge Vessel Rinses Three 90 C Water Sparge Vessel Rinses 3mL Methanol Sparge Vessel Rinse 0.10 0.00 5mL Methanol Sparge Vessel Rinse 3mL Methanol Sparge Vessel Rinse and 3mL Methanol Needle Rinse

Elution Issues Tune Auto-tune/BFB tune Direct inject If missing peaks could be a bad EPC Column Top of Trap If missing peaks could be sample tubing/transfer line

Early Elution Groups GC Split flow leak Bad column Purge and Trap Small leak Water front too large Too long of dry purge Trap too hot (faulty electronics or heater)

Middle Elution Groups (Very Rare) GC Column phase is bad Purge and Trap Desorb preheat problems Condensate trap has too much water Mass flow controller leaking Bad Trap

Late Elution Groups GC/MS Source maintenance Carrier flow too low Purge and Trap Poor purge efficiency Restrictive sample tubing Dirty glassware Sample not heated properly Plugged vent valve Bad Trap

Water interference MS Vacuum issues GC Split flow too low Bad column Purge and Trap Condensate setting and or failures Dry purge not optimized Sample temperature concerns (too hot) Desorb Time

Water Interference Chromatogram showing the water front for a 2min desorb (white) compared to 4min desorb (green)

Water Interference Chromatogram showing the water front for a 2min desorb at 1ppb (white) compared to 4min desorb at 1ppb (green)

Internal Standard Recoveries GC/MS response increases as Target compound concentration increases Dirty inlet liner or source Validate by preparing 3 increasing target concentrations with Internal Standard concentration fixed in 1mL vials and then direct inject 1uL of the concentrations into the GC If still increasing isolate between source and liner by performing cleaning and or replacement. 6mm draw-out plate

Internal Standard Recoveries Purge and Trap Isolated to an active site Sample tubing/condensate trap/analytical trap Autosampler Hand spike vials first Automated delivery system Tubing restrictions Valve malfunction Pressure problems

Internal Standard Recoveries AQUATek100

Linearity and Reproducibility GC/MS Source maintenance Vacuum problems EPC failures Bad column Method not optimized Oven temperature program Multiplier going bad

Linearity and Reproducibility Purge and Trap Method not optimized Trap is failing Drain valve leaking Excess water in system Temperatures not reaching set-points Too much carry over Sample preparation errors Dirty standard syringe or glassware Too much headspace Analyst variability

Linearity and Reproducibility Autosampler Not pulling same amount of sample Not transferring sample correctly Improper rinsing between samples Compare soils vs. water if possible Determine if IS is varying and causing everything else to look unstable

Purge and Trap System Leaks Perform Leak Checks through Software:

Purge and Trap System Leaks

Purge and Trap System Leaks Since the Atomx is an Autosampler and P&T concentrator combined, the leak check is more complex If fails the overall leak check it will break down into three subsystems

Purge and Trap System Leaks Subsystem A Mass flow controller, Bake valve, Purge valve, Pencil Filter, Purge select valve, Syringe valve, and Pressurize valve

Purge and Trap System Leaks Subsystem B All subsystem A, Syringe,7-port valve, Transfer valve, Sample needle, and Soil valve

Purge and Trap System Leaks Subsystem C All of subsystem A&B, Sample mount, Sparge vessel, Drain valve, and 6-port Valco Notice traps not included so failure of system with A, B, and C passing indicates leak in trap area

Benchmark Make sure system is leak tight first Checks all system controls including solenoid valves, heaters, flows, and 6-port valve Helps isolate a variety of potential issues

Miscellaneous Issues Draining sample -Flag on Hamilton Valve -Drain Valve Communication with GC/MS -Standard 63 -User 31 History Log -To check Purge/Bake pressures -To check Errors and Modes they occurred in

www.teledynetekmar.com Tech Support 1-800-874-2004