Use of Automation to Achieve

Similar documents
Use of Automation to Achieve High Performance SPE

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

Fundamentals of Solid-Phase Extraction (SPE): Principles and Practical Tips for Developing Sample Preparation Methods

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Supra-Clean and Supra-Poly SPE Solutions. Solid Phase Extraction Columns

300 series HPLC autosamplers

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Spark Holland introduces

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica.

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

On Line SPE LC/MS/MS Applications

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

Reliable elimination of particles and micro organisms PAGE Protein Precipitation Fast and effective sample clean up of biological samples

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

Application Note. Authors. Abstract

An Automated SPE Procedure and Analysis for the Extraction of Chlorpromazine and Thioridazine from Plasma Application Note 213

Sample Preparation Solid Phase Extraction Disks Cartridges. Horizon Technology, Inc. Salem, NH USA

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

MultiPurpose Sampler MPS robotic series

Thermo Scientific Dionex SolEx SPE Cartridges

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

How Important is Performance to your SPE Method Development?

Order all the parts and accessories needed at one time Reduced learning curve for rapid implementation

Direct Injection and Online SPE LC/MS/MS for the Determination of Pharmaceuticals and Personal Care Products (PPCPs) in Water Samples

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

Gagnez avec «David» Mighty MEPS contre les cartouches «Goliath» SPE

Improved HPLC Separations at Elevated Temperatures Using a New Stand-alone Preheater

Methods of solid phase extraction. Methods of solid phase extraction

Tool-free LC connections for HPLC, UHPLC and nano LC systems

Simplicity. Reliability. Performance. Stratum. Automated VOC Sample Prep System

Product Guide for Ludger Clean Post-endoglycosidase and Post-exoglycosidase Clean-up Plate

Chromatography. Research Supplies. 2014/2015 Catalog

detectors UltiMate 3000 Variable Wavelength Detectors

New Designs and Formats in Solid-Phase Extraction Sample Preparation

Pretreatments to LLE

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

IMPROVING THE QUANTITATIVE REPRODUCIBILITY OF AN ATMOSPHERIC PRESSURE CHEMICAL IONISATION SOURCE FOR GC/MS AND GC MS/MS ANALYSIS

An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

HISTORY OF VARIAN CAPILLARY INJECTORS

XENOSEP. SPE Systems for EPA Method 1664A. The Easiest, Most Cost Effective Way to Increase Productivity

Purge and Trap Troubleshooting Made Easy?

Sample Prep Perspectives

Customised HPLC Systems for Highest Performance

DETERMINATION OF PAH In Particulate Matter PM10 With SPE/EVAporation

Thermo Scientific Dionex UltiMate 3000 Fluorescence Detectors

This document is a preview generated by EVS

Agilent InfinityLab Online SPE Solution

Systems. Visit us at

International Journal of Scientific & Engineering Research Volume 8, Issue 11, November-2017 ISSN Volatile Sampling Device Prototype II

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

Simulation Of Pneumatic Drying: Influence Of Particle Diameter And Solid Loading Ratio

Advancing the power of SPE. Intelligent Automated Solid Phase Extraction. SmartPrep Brochure

Agilent 1290 Infinity II Vialsampler Data Sheet

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

FREESTYLE always ready

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

Advancing the power of SPE. Intelligent Automated Solid Phase Extraction. SmartPrep Brochure

Optimizing Chemistry and Coater Settings for Faster Drying

Standard Operating Procedure for Oil & Grease Method EPA1664A

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

CALIDUS tm the Modular micro Gas Chromatograph

Chemistry Instrumental Analysis Lecture 29. Chem 4631

API 2000 LC/MS/MS TurboIonSpray Ion Source Manual

PURITY GRADES OF LAB REAGENTS

SEC-MALS Noise Assessment and Cleaning Guide

Biotage VacMaster 10 & 20 Sample Processing Station. Operation Manual

ISO AIR FILTERS FOR GENERAL VENTILATION: DETERMINING FRACTIONAL EFFICIENCY

Group 2 Herbicide Interactions and Approaches to Detection in Soils

Development Centre for Evaporation and Thermal Separation Technologies

until filters are dry. Once dry, place the filters shiny side up in petrislides. The petrislide do NOT need to be acid washed.

TYPES AND APPROPRIATE INJECTORS

Without the Dry Down. Successful Bioanalytical Sample Preparation. Microelution 96-Well SPE Plates

APPLICATION NOTEBOOK FOR AFFINIMIP SPE

WELCOME TO ELLUTIA CHROMATOGRAPHY SOLUTIONS

Injection/Trapping. Analysis

GC Injectors. GC Injectors

SUMMARY AND CONCLUSION

An Automated Method for the Selective Solid Phase Extraction of Ochratoxin A from Wheat Using Molecularly Imprinted Polymers

Nexera-i Specifications Prominence-i Specifications

Exercise 2-4. Heat Exchangers (Optional Exercise) EXERCISE OBJECTIVE DISCUSSION OUTLINE. Description of a brazed plate heat exchanger DISCUSSION

The Complete and Easy Guide to Configuring Your Specific Thermo NanoLC for Mass Spec Analysis

Presenter: Malcolm McLaughlin Date: 04/04/2019. A Review of Medical Device and Optical Critical Cleaning

Perkin Elmer Clarus 500 Gas Chromatograph

Zoom Microplate Washer and Washer-Dispenser. Reliable Solutions for Your Daily Work

The complete approach to the care and management of A/C & refrigeration systems

Transcription:

Use of Automation to Achieve High Performance SPE Mark Hayward, 1 Jonathan Ho, 2 Matthew T. Hardison, 3 Martin Johnson, 3 Tom Moran, 2 and Kim Gamble 1 1 ITSP Solutions Inc., 10 South Carolina St., Hartwell GA 30643 2 Shimadzu Scientific Instruments, 19 Schoolhouse Rd, Suite 107, Somerset NJ 08873 3 Assurance Scientific Laboratories, 2868 Acton Rd, Suite 207, Vestavia, AL 35243 Key take away: Solid phase extraction is liquid chromatography! Operate it as such and get better results!

Chromatography: started with Tswett Monitored pressure to regulate flow Original apparatus: Glass column w/ CaCO 3 particles held in place with screen and filter paper Built fractionation apparatus Extracted plants w/ pet-ether and isolated chlorophyll Controlled flow in real-time with positive pressure While he didn t know why, Tswettknewimportance of regulating flow when performing chromatography 110 years ago Tswett soriginalapparatus is way many perform SPE today: 110 years ago, it was known that it needed to be done better! 110 years ago, it was known that it needed to be done better! Unfortunately, Tswett received little attention for this

Chromatography received little attention until A.J.P. Martin and R.L.M. Synge described partitioning model explaining paper chromatography in 1941 (general mechanistic understanding of chromatography - Nobel) Erika Cremer and Fritz Prior built first GC in 1945 Metal tubes filled with activated carbon used in large volume water sampling (emergence of SPE in 1950: 50 years of SPE, I. Liška, J. Chromatogr., A, 2000, 885, 3 16.) J.J. van Deemter, F.J. Zuiderweg, A. Klinkenberg, Chem. Eng. Sc., 1956, 5, 271 289 allowed understanding of chromatographic flow Then, the basic knowledge was in place for the Then, the basic knowledge was in place for the use of chromatography to grow exponentially

Evolution of Practical LC and SPE Commercialization of LC by J. Waters in the 1960s led to development of pumps, injection systems, & functionalized silica sorbent particles in the 1970s, similar to current use Parallel developments also led to SPE devices (syringe tube) in 1970s (96 well in 1990s, pipette tip in 2000s) still in use Development of HPLC closely followed teachings of van Deemter& Giddings (short diffusion distance & accurate flow) SPE development did not follow these teachings since it is not isocratic (required assumption to derive chromatography equations) Later, lessons of van Deemter& Giddings applied to gradient separations, inc. accurate measurement of plate count (Neue, UD, HPLC Columns: Theory, Technology, and Practice, Wiley, 1997, p77) Until now, single use SPE devices haven t followed these Until now, single use SPE devices haven t followed these teachings, particularly importance of carefully controlled flow

Issues with common single use SPE devices Lack of flow control With vacuum or pneumatic pressure, changing volume (weight) of liquid above sorbent changes flow rate during each SPE step When performed in parallel for high throughput, effect is exacerbated since each SPE device (or well) has different resistance to flow & different variable flow rate Result of flow variability is considerable variability in absolute recovery (50-85% common) & results Internal standards used to achieve meaningful results Overall data evaluated based on worst case scenario (flow far from optimal, low absolute recovery) High absolute recovery against external standards in only solvent, the gold standard in demonstrating absence of matrix effects, isn t used with SPE

Assertion: SPE is LC Fundamentals of achieving high performance operation in LC: well packed sorbents to control variance in diffusion distance, precise flow control to match diffusion velocity/distance, & minimizing dispersion (dilution) History of use to improve column based LC performance is well documented & gains truly significant Despite >40 years performing SPE using LC sorbents, these principles have not been applied to SPE Reality: known principles that apply to gradient Reality: known principles that apply to gradient HPLC, apply to SPE equally

Typical single use SPE device performance flow driven pneumatically or vacuum Flow: slower is better but counter productive Lack of clearly defined Jordan L, LCGC 1993, 11, 634-8 for huge SPE cartridges Commonly heard rationalization: SPE is digital chromatography. Thus, we cannot expect LC like performance despite the fact that we are using LC sorbents. This presentation challenges that claim using packed sorbent & automation to achieve flow control. optimum flow demonstrates lack of flow control or packed chromatographic sorbent, or both Decrea asing Recov very σ 2 = A + B + C Expected chromatographic behavior Optimum Flow Flow

Why do we use SPE? SPE is preferred tool for isolating target analytesfrom complex matrices due to availability of diverse range of chromatographic sorbents enabling targeted approaches based on specific chemistry of analytes& matrices Single use devices help prevent carryover SPE offers the ability to enrich or pre-concentrate analytes, particularly when drying & re-dissolving them afterward Enrichment is valuable allowing one to match analyte concentrations to approach used to measure them Given these unique capabilities, SPE often a first choice Given these unique capabilities, SPE often a first choice in analytical sample preparation

Testing a new micro-spe device Initially, compelling reasons for interest in this new Smart SPE device from ITSP solutions were: Automation using CTC/PAL autosamplerperformed at same cost as other single use SPE devices SPE performed on-line in parallel with LC/MS/MS (or GC/MS/MS) analysis using CTC/PAL autosampler Enrichment of analytewithout need for drying eluantdue to small elution volumes possible with this micro-spe device Testing also showed interesting chromatographic performance not previously seen in other single use SPE devices

Heart of the ITSP form of SPE is patentedsingle use cartridge containing customer-defined packed chromatographic media Crimped-on septum & needle guide (upper 80% of cartridge) enable automation by facilitating accurate cartridge transport on PAL syringe needle The small (16 µl below sorbent) extra-column volume facilitates low volume elution (50-100 µl)

Automated and Higher Quality SPE Using an Autosampler: How it works Automated SPE begins using autosamplersyringe for cartridge conditioning, sample loading & washing over waste receptacle Syringe then used to perform elution over clean vial or well, & used cartridge discarded in a different waste receptacle Automation complete after syringe mixes freshly eluted sample & then injects it into LC/MS/MS or GC/MS/MS used to measure the sample (see next Figure showing CTC/PAL autosampler). SPE performed in parallel after SPE of first sample

Hoses connect the SPE cartridge tray and syringe wash station to an ordinary lab solvent waste container. Used cartridges are typically discarded by the PAL into a box under the wash station and LC valve.

Workflow: minimizing cycle time Parallel PAL operation in the inject ahead mode SPE 1 SPE 2 SPE 3 LC/MS/MS 1 LC/MS/MS 2 4.5 min 4.5 min Total cycle time (SPE + LC/MS/MS) = 4.5 min Method: 71 drugs in urine, RP SPE C18 50 µm particles SPE time = 3.2 min, LC/MS/MS time = 4.5 min

Development of Automated SPE Methods Resulted in Detailed SPE Flow Studies UCT 50 µm C 18 end capped silica sorbent Included in method development was thorough flow optimization study intended to measure cost/benefit in time/recovery Confirmed results not previously expected: The result was a 20 data point U-shaped curve showing that flow of 5 µl/s resulted in 100% absolute recovery Skeptical, the flow study was repeated, then again measuring 94 data points, then again measuring load and elute steps separately while holding the other at 5 µl/s All of these produced the same U-shaped curve and they all looked similar to a van Deemtercurve

How to plot a van Deemtercurve for SPE Conventional measures of SPE (recovery) differ from conventional measures of GC and LC (retention time & peak width [2σ]) Yet the processes are same (diffusion, adsorption [or not], & desorption [or delayed]), the random walk model applies, & van Deemterequation is a variance (σ 2 )equation [Giddings, JC, Unified Separation Science, Wiley 1991, p 92-101] Use of computerized chromatography data to evaluate separation performance [%RSD method for plate height calculation in Neue, UD, HPLC Columns: Theory, Technology, and Practice, Wiley, 1997, p12-13] offers simple view of relative nature of σ&how to address SPE data %Recovery measures deviation (σ) like LC peak width & thus, (100% -measured %Recovery) 2 is a measure of variance (σ 2 ) Hence, plotting (100% -measured %Recovery) 2 vs. flow should yield typical van Deemtercurve shape if chromatographic processes govern dispersion of molecules in SPE

Flow optimization for ITSP SPE: just like LC column van DeemterCurves for RP SPE using 2 different particle sizes 10 um 50 um 80 UCT 50 µm C 18 Load & elute steps optimized separately, behave identically Methadone % Reco overy vs. exter rnal stds 94 data points per curve measured by LC/MS/MS & fit to van Deemter equation in Excel Supelco10 µm C 18 100 0 5 10 15 20 Flow (µl/s)

Flow Optimization: Outcome and Impact ITSP SPE cartridges behave like LC columns due to: Accurate flow control from PAL autosampler(syringe pump) Sorbent is packed Low extra-column volume Benefits: >99% absolute recovery systematically achieved (within measurement precision of LC/MS/MS ±3%) Same tools used to increase speed & performance for LC can be used for SPE (eg. smaller particles) Thus far, we haven t found an application that can t be done with 10 mg sorbent: high sorbent mass SPE cartridges appear to be band-aid for overcoming low recoveries caused by insufficient flow control

Use of chromatographic SPE knowledge Same SPE & LC/MS/MS method except sample volume 71 drugs LC/MS/MS time = 4.5 min SPE urine:200 µl sample loaded, 50 µm particles, SPE time = 3.2 minutes SPE oral fluid: 1000 µl sample loaded, 10 µm particles, SPE time = 4.5 minutes van DeemterCurves for RP SPE using 2 different particle sizes 10 um 50 um 0 5 10 15 20 Flow (µl/s) SPE gets same benefits of particle size as LC (systematic control of speed) 80 covery ernal stds % Rec vs. exte 100

SPE: Drugs in Urine and Oral Fluids Same C 18 RP method: different sample volumes loaded on SPE cartridge 18 Urine: Enrichment: 3x (200 µl load /75 µl elute MeOH) Cutoffs (all): 1 ng/ml (S/N=20+) 1 mg/day benzos, opioids, & metabolites easily measured (considered challenging) 192 samples/day/lcmsms (50 µm - overnight only typical small to medium lab workflow) Removes: salts (~2%), small organic acids/bases (~1%), sugars (oxidized and intact), amino acids, glucuronidase Maintenance:reagents/solvents & instrument PM / LC column change each 6 months without loss of performance Oral fluid: Enrichment: 13x (1000 µl load /75 µl elute MeOH) Cutoffs (all): 0.2 ng/ml (S/N=20+) 1 mg/day benzos& metabolites easily measured (ordinarily considered not feasible) 192 samples/day/lcmsms (10 µm particles - overnight only) Quantisalsampling/filtering (sample volume +/-10%) Removes: salts (~100 mm), mucopolysacharides, enzymes, glycoproteins Maintenance: reagents/solvents & instrument PM / LC column change each 6 months without loss of performance Validated for 71 drugs used for production >2 yrs

Further Study Rapid progress developing urine & oral fluid methods led to variety of method development efforts expanding the range analytesin more complex matrices (blood/tissues/food) To address the more complex matrices this led to use of more selective sorbents: Cationexchange for drugs (mixed mode RP due to alkyl linker) Anion exchange for lipids (mixed mode RP due to alkyl linker) Chelation for phospholipids and phosphopeptides As might expected, this led to more SPE flow studies

Flow optimization for Cation Exchange SPE (50 µm particles) 80 % Abs solute Re ecovery SCX %-Recovery in 20% Water SCX %-Recovery in 80% Water SCX %-Recovery in 50% Water Reverse Phase Load and elute steps optimized separately behave identically 100 Oxycodone 0 1 2 3 4 5 6 7 8 n-propyllinked phenyl-so 3 - on silica (UCT) Flow (µl/s) Molecules without cationic sites have optimal flow of 5 µl/s (i.e. RP)

Ionic SPE Flow Optimization: Outcome and Impact Outcomes: Cationexchange SPE has optimum flow of 1.2 µl/s (1/4 that of reverse phase, 50 µm particles) In addition, the relatively narrow acceptable flow range for high recovery widens with additional water content Preliminary results suggest anion exchange and chelationspe have the same van Deemter curves as cation exchange Benefits: 100% recovery is systematically achieved (within the measurement precision of the LC/MS/MS ±3%) [rare with these chemistries] Knowledge of the chemical preference for more water (less solvent) leads to more predictable and consistent outcomes Use silica (not polymer) based particles to minimize solvent levels (n-propyl phenyl linker still gives sufficient mixed mode [RP] behavior) Use strong miscible solvents (IPA/THF) to maximize water content

A possible explanation for water dependence on ionic adsorption / desorption Seems likely ionic sorbent sites have layer around them more concentrated with water (rather than solvent) Likely reason cation, anion, & chelationhave the same optimum flow while having different binding energies (mass transfer regulated by transport through the water layer) Amount of solvent present regulates the thickness of water layer High performance SPE facilitates learning Cation nexcha ange so orbent -SO - 3 Water layer with higher solvent content Water layer with lower solvent content

Use of chromatographic SPE knowledge Cation exchange of basic drugs in blood (always mixed mode, 50 µm particles) Optimum flow for cation exchange load 80 and elute = 1.2 µl/s Solvent wash steps have a different optimum flow! (5 µl/s) Solvent wash step at 1.2 µl/s removes <90% of phospholipids and fatty acids Solvent wash step at 5 µl/s removes >99% of phospholipids and fatty acids % Abs solute Re ecovery 100 Flow chosen depends on chemistry performed! SCX %-Recovery in 20% Water SCX %-Recovery in 80% Water SCX %-Recovery in 50% Water Reverse phase 0 1 2 3 4 5 6 7 8 With 1.2 µl/s solvent wash, phospholipids are observable by ion full scan LC/MS & not observable with 5 µl/s solvent wash

SPE: Basic Drugs in Blood n-propyllinked phenyl-so 3- on silica -50 µm Sample: 100 µl, 2/1 IPA crash, load 150 µl, 100 µl elute H 2 O/IPA/NH 4 OH) Washes: 0.02M ph 6 phosphate buffer, IPA/MeOH, H 2 O Cutoffs (all): 1 ng/ml (S/N =20+), 100x better LogP 4 1 mg/day benzos, opioids, and metabolites readily measured 96 samples/day/lcmsms (50 µm -overnight only typical small to medium lab workflow, more possible) Removes: salts, organic acids, sugars (oxidized and intact), amino acids, LIPIDS >99% absolute recoveries 2D-LC (trap & elute) gives same performance for blood spots Maintenance:reagents/solvents, pre-column filter frit monthly (not needed with column switching), & instrument PM / LC column change each 6 months without loss of performance (forensics) Validated for 63 drugs (inc cationic illicits)

Summary and Conclusions With flow control and packed sorbent, SPE is gradient LC All existing knowledge about gradient LC can be applied to SPE Smaller particles for RP SPE significantly increases speed without impact on performance Applying this knowledge produces significantly better results than alternatives 99+% recovery / matrix removal systematically achieved Adsorption and desorption is a reversible equilibrium Separate measurement of load and elute flow behavior produces the same van Deemter curves (just like isocratic LC) SPE using single use devices with accurate flow control can achieve high quality even in high throughput applications (fast automation) Given the same price point, it s hard to see rationale for continued use of single use SPE devices that utilize loose sorbent and/or vacuum / pneumatic driven flow