Operation & Maintenance Manual

Similar documents
Operation and Maintenance Manual

Operation and Maintenance Manual

Operation and Maintenance Manual

Operation and Maintenance Manual

Model OI-6000K Sensor Assembly

Model OI-6000K Sensor Assembly

Product Selection Guide Fixed Gas Detection Systems for Personnel & Equipment Safety

GLD-30 Gas Leak Detector

Model D12-IR. Digital Gas Transmitter...

Air Check EX LEL Gas Monitor Instruction Manual

5New features & EXtreme design, now with HART Protocol and DuraSource Technology. Ultima X Series Gas Monitors

SIL 2. Ultima X Series Gas Monitors CERTIFIED

P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification

RK-HC-04 M2A Transmitter Operator s Manual

M2A Transmitter Operator s Manual

SEC 2000 Millenium Infrared Gas Detector

Analog Room Pressure Monitor RPC Series

Carbon Monoxide Transmitter

Toxic Gas Detection. Model F12. Toxic Gas Detection

INSTRUCTION MANUAL BUCKEYE DETECTION SYSTEMS MODEL BFT-48 TRANSMITTER

M2 Transmitter Operator s Manual

OLCT 700 & 710. Certifications FIXED GAS DETECTION. Better detection. Better protection.

RKA-LEL Sample Draw Combustible Gas Detector

SPECIFICATION CHEMGARD INFRARED GAS MONITOR SERIES

FX-SMT Series INSTRUCTION MANUAL P/N THERMO ELECTRON CORPORATION ENVIRONMENTAL INSTRUMENTS 27 FORGE PARKWAY FRANKLIN MASSACHUSETTS 02038

4-20mA CYBER Cyber Transmitter for flammable, toxic and IR gas detection Cyber Head Increased security in ATEX certified head

RK-04 M2 Transmitter Operator s Manual

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

Pump-Down Controller MODEL 4052

Controllers. Instruction Manual WARNING

Pump-Up Controller MODEL 4062

Toxic and Explosive Smart Gas Monitor PN / Installation and User Manual

Inc Series Digital Gas Detector/Controller. Operation Manual

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

WARRANT Y AND LIMITS OF LIABILIT Y Vulcain Inc. warrants to the original purchaser that its product, and the component parts thereof, will be free fro

Model TS4000. Intelligent Sensor for Toxic Gas Detection

Two-Channel Gas Controller

M2 Rig Monitor Operator s Manual

SECTION 1 (Page 2) Introduction, standard accessories and mounting specifications.

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

RK Carbon Monoxide Transmitter Operator s Manual

Instruction Manual WARNING

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

RK-05 Carbon Monoxide Detector Operator s Manual

Danfoss gas detection unit Type GD Heavy Duty

ADDENDUM. xwatch Explosion-Proof Camera with X-Series Flame Detectors DESCRIPTION GENERAL APPLICATION INFORMATION

INSTRUCTION MANUAL MODEL C2 PROTECTOR FOUR CHANNEL CONTROLLER. (Revision 2.2; Beginning Firmware 1.13)

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

Pioneer-R16 Gas Monitor Operator s Manual

Air Check O 2 Sample Draw Monitor

SAFETY MANUAL. Intelligent Sensors for H 2 S Gas Applications

Model IR4000M. Multi-Point Monitor. Instruction Manual 12-10

Industrial Grade Gas Detection Detecting The Difference

ULTIMA XL/XT Series Gas Monitors. [ Ultimate Features Extreme Design ]

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6

SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM. User Instructions For The Model 5200 Gas Monitor System

Air Check Ex O 2 Deficiency Monitor

SAFETY MANUAL. IR5000 Open Path Hydrocarbon Gas Monitoring System

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

Universal Gas Transmitter

HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL

DIGITAL METERS Large Display Temperature Meters Instruction Manual

Warranty Registration

GAS PLUS ALARM SYSTEM

Model TS4000H. Intelligent Sensor for Toxic Gas Detection

TYPE CM-2201 NELSON SINGLE POINT CIRCUIT MANAGEMENT SYSTEM

Models TX-KE and TX-KP Toxic Gas Transmitters

RK Multi Point Detector Operator s Manual

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Nanotechnology Metal Oxide Semiconductor (NTMOS) H 2 S Gas Sensor

Model SM100. Aspirated Sampling Module

Model 1000 H 2 S E-Chem/ CO 2 Process Analyzer

Detectors-Transmitters for Flammable, Toxic Gases and Oxygen. Non-intrusive operation via intrinsically safe infrared remote control.

Beacon 800 Gas Monitor Operator s Manual

Dräger Polytron Fixed Gas Detector. Dräger Polytron 5700 Infrared DrägerSensor PIR 7000 to detect combustible gases and vapors

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

MYRIAD TRIPLEX PUMP CONTROLLER INSTRUCTION MANUAL

RK-02 Multi Point Detector Operator s Manual

INSTRUCTION MANUAL Detcon Model DM-700

RAM3 Remote Alarm Module

RKA-03 Sample-Draw Detector Head

SenSmart Universal Gas Detector Series

ChemScan mini Analyzer for Low Ammonia Monitoring Installation, Operation and Maintenance Manual

Danfoss gas detection units

D12-17 % LEL Gas Transmiter

RK Hydrogen Sulfide Transmitter Operator s Manual

Fixed Gas Detector User Manual Installation Operation Configuration Troubleshooting

FLT93 Installation, Operation and Troubleshooting Guide

Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual

Model FMD. Field Mounted Display Module

Combustible Gas Sample-Draw Detector Operator s Manual

RKA-06 Sample-Draw Detector

Safe Operation Manual. Honeywell IQ Force Gas Detector

Module Features are-configurable, no module jumpers to set

Air Check O 2 Deficiency Monitor

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

Universal Gas Monitoring System

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

IGD.CO.UK

Improve Performance and Reduce Maintenance Costs. Model Q45D. Dissolved Oxygen Monitor...

Transcription:

Operation & Maintenance Manual GDS-58XP Explosion-Proof Sample Draw System for Toxic & Combustible Gases GDS Corporation 2513 Hwy 646 Santa Fe, Texas 77510 (409) 927-2980 (409) 927-4180 (Fax) www.gdscorp.com

GDS-58XP Sample Draw System Users Manual V2.1 CAUTION: FOR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING OR SERVICING. ATTENTION: POUR DES RAISONS DE SÉCURITÉ, CET ÉQUIPEMENT DOIT ÊTRE UTILISÉ, ENTRETENU ET RÉPARÉ UNIQUEMENT PAR UN PERSONNEL QUALIFIÉ. ÉTUDIER LE MANUE D INSTRUCTIONS EN ENTIER AVANT D UTILISER, D ENTRETENIR OU DE RÉPARER L ÉQUIPEMENT. NOTICE: THIS DOCUMENT CONTAINS LICENSED INFORMATION THAT IS THE INTELLECTUAL PROPERTY OF GDS CORP; ACCORDINGLY, IT IS AVAILABLE ONLY UNDER LICENSE FROM GDS CORP AND MAY NOT BE REPRODUCED, WHOLLY OR IN PART, BY ANY LICENSEE TO ANY OTHER PERSON OR ORGANIZATION WITHOUT THE EXPRESS WRITTEN PERMISSION OF GDS CORP. For further information about GDS Corp, please call (409) 927-2980 or email support@gdscorp.com Copyright 2012 GDS Corp. All Rights Reserved P/N 1200-0638-004 Page 2

GDS-58XP Users Manual V2.0 CONTENTS 1 GDS-58XP SAMPLE DRAW SYSTEM...5 1.1 Description...5 1.2 Features...5 2 GDS-58XP SPECIFICATIONS...6 3 HARDWARE INSTALLATION...7 3.1 Installing the GDS-58XP...7 4 INITIAL SETUP...10 4.1 Flow Setup...10 4.2 Adjust Alarm Levels...10 4.3 Verify Remote Readings...10 4.4 Initial Calibration...10 5 OPERATION & MAINTENANCE...11 5.1 Overview...11 5.2 Calibration...11 5.3 Sensor Removal / Replacement...12 6 SPARE PARTS...13 6.1 GDS-58XP Spare Parts...13 6.2 GASMAX Spare Parts...13 7 WARRANTY...13 7.1 Overview...13 8 APPENDIX...14 8.1 MODBUS Registers...14 Page 3

GDS-58XP Sample Draw System Users Manual V2.1 8.2 Moisture Calculations...14 8.3 GASMAX Configuration...15 Page 4

GDS-58XP Users Manual V2.0 GDS-58XP SAMPLE DRAW SYSTEM 1.1 Description Thank you for choosing the GDS Corp GDS-58XP Explosion Proof Sample Draw system. The GDS-58XP provides a safe and reliable way to pull toxic or combustible gas samples from a remote location for measurement by a local GASMAX II gas monitor. The GDS-58XP is recommended for use with any hydrocarbon or hydrogen-based combustible gas or any toxic gas, including reactive components such as chlorine, chlorine dioxide or hydrogen fluoride*. The GDS-58XP features a long life brushless DC pump with low-flow warning switch with 5A local relay contacts for Alarm 1, Alarm 2 and Sensor or Flow fault. The GDS-58XP is designed for use in Class I Division 1 hazardous areas. 1.2 Features Designed for Class I Division 1 Hazardous Areas Long life Brushless DC sample pump with fixed 1.3 liters-per-minute flow rate Integrated flow switch connected to local 5A fault relay Local relay contacts for optional light or warning sounder Standard non-isolated 4-20mA current loop output (source) Optional 110 / 220VAC input (requires additional matching XP enclosure) Standard MODBUS slave interface for remote monitoring of signal and flow status High resolution display shows engineering units and trend data Built-in user-prompted calibration Magnetic interface for non-intrusive operation in hazardous areas GDS-58XP for reactive gases not for use in hazardous areas. Page 5

GDS-58XP Sample Draw System Users Manual V2.1 1 GDS-58XP SPECIFICATIONS Input Power: +24VDC +/- 10%, 12.5W max. Sensor Type: Any GDS Corp Toxic, Combustible or PID Sensor Analog Output: Single 4-20mA calibrated current loop (source), Relay Output: Dual 5A programmable relays for Alarm 1 and Alarm 2; Single 5A relay contacts for Sensor Fault and/or Low Flow Warning Digital Output: RS-485 Two-wire Serial MODBUS Slave Interface Flow rate: 0.5 to 1.0 liters / minute (adjustable) Maximum sample distance: Approximately 100 ft / 33 m Inlet and Outlet Fittings: ¼ Compression Dimensions: 15.25 (W) x 10.0 (H) x 6 (D) Weight 11 pounds / 4.9 KG Page 6

GDS-58XP Users Manual V2.0 2 HARDWARE INSTALLATION 3.1 Installing the GDS-58XP Install the GDS-58XP in a suitable location within 100 ft / 33 m of the intended sample location. The GDS-58XP features four mounting holes that can be used to securely attach the sample draw system to a bulkhead or wall. Make sure there is sufficient clearance to install the sample inlet tubing and sample outlet tubing (if used). Also make sure that the Run / Cal valve can operate freely and that access to the Cal Port is not blocked for any reason. Pump & Flow Switch GASMAX II Monitor Sample Inlet Run / Cal Valve Cal Port Sensor & Flow Cell Flow Meter Sample Outlet Figure 1 GDS-58XP Note: In order for low flow switch to operate properly, the GDS-58XP must be mounted vertically. Page 7

GDS-58XP Sample Draw System Users Manual V2.1 DC Power: Connect a source of +24VDC, +/- 10% power to the power supply wiring terminals (See Fig 2). An external power switch and / or circuit protection device should be provided by the customer s wiring. If the area is considered hazardous then all explosion proof wiring recommendations, guidelines and regulations should be observed. PWR OUT 4-20 ECHEM OUT 4-20 LEL GND 1 2 3 4 C2 C1 GASMAX I/O Power Supply P/N 10-0233 TP1 CATBEAD REF COM ACT +24VDC to TB2 #1 COM to TB2 #4 TB2 C8 C7 S1 TB1 O2/TOXIC: 4-20mA Output on TB2 #2 BRIDGE: 4-20mA Output in TB2 #3 Figure 2 GDS-58XP / GASMAX II Power Supply & Analog Signal Wiring Attach Note: If the optional P/N 1200-0370 Class I Div 1 AC Power Supply is attached, connect the incoming AC wiring directly to the power supply Line, Neutral and Ground terminals. Analog Output: Connect the output 4-20mA current loop wiring to the appropriate controller or DCS input. If a TOXIC sensor is used (Oxygen, Hydrogen Sulfide, etc) then the output is taken from TB2, Pin #2. If a combustible or PID sensor is used, the output is taken from TB2, Pin #3. Alarm Relay Output: Alarm 1 ( Relay 1, TB1 Pins 1,2,3) and Alarm 2 ( Relay 2, TB1 Pins 4,5,6) contacts are located on the Relay / MODBUS board, attached to the back of the GASMAX II display (See Fig 3). These relays can be programmed via the main menu to close at preset signal levels, and can operate as either Alarm Above or Alarm Below contacts. Fault Relay Output: Relay 3 (TB1 Pins 7,8,9) is programmed to indicate a FAULT condition under the following circumstances: 1) Excessive negative sensor drift; 2) Sensor removed or 3) Flow switch dropout (low flow). If only a single wiring pair is available, the 4-20mA output can be routed through the FAULT relay. This will cause the current loop signal to be interrupted in the event of a low flow indication. Page 8

GDS-58XP Users Manual V2.0 Remote Alarm Reset TB3 P1 Relay 1 (K1) Relay 2 (K2) Relay 3 (K3) NO 1 C 2 NC 3 NO 4 C 5 NC 6 NO 7 C 8 NC 9 TB1 Assy: 10-0234 ST-48 Alarm / ModBus Option K1 K2 K3 1 2 3 4 5 A B S A B U1 J1 TB2 RS-485 A & B terminals are connected internally for easier IN / OUT cabling. S is no connect for shield to continue. A (Installs RS-485 Terminating Resistor) B (Omits RS-485 Terminating Resistor) Figure 3 GDS-58XP / GASMAX II Relay & MODBUS Wiring Attach MODBUS Output: The GASMAX II provides a two-wire RS-485 serial data link via TB2 Pins A & B. See the GASMAX II manual for more information on MODBUS registers and operation. Sample Inlet: Attach a length of tubing to the Sample Inlet using the supplied ¼ compression fitting. Note: If the sample location is excessively wet or dusty, GDS Corp recommends the installation of a coalescing or particulate filter on the intake end of the sample draw tubing. This will keep the particulate or moisture from being drawn into the sample tubing, sample draw pump and sensor. In the event the filter becomes clogged, the FLOW OK indication will be replaced by FAULT on the GASMAX II display and the Low Flow / Fault relay will energize. Sample Outlet: Attach a length of tubing to the Sample Outlet using the supplied ¼ compression fitting and route the tubing to a safe area. Note: Do not use more than 25ft of tubing and do not allow back pressure to build up in the sample flow cell. GDS Corp recommends that the outlet point be located no higher than 10 (3m) above the GDS- 58XP. Note: All GDS Corp sensors are designed to operate at ambient atmospheric pressure. Restricting the output in any way may create back pressure in the sensor flow cell that will generate erroneous readings and may damage the sensor if excessive pressure is present for long periods of time. Page 9

GDS-58XP Sample Draw System Users Manual V2.1 3 INITIAL SETUP 4.1 Flow Setup Once hardware installation is completed, apply power to the GDS-58XP and verify that the display is operational. Adjust the Sample Flow Meter to indicate 0.5 liters per minute of flow. If the flow meter does not indicate proper flow, check the inlet tubing for restrictions or blockage. To verify that the pump is running properly, set the Run / Cal valve to the CAL position (pointing to the RIGHT). The pump should be able to pull in excess of 1.0 liters / minute (full scale on the flow meter). If full scale flow cannot be obtained, check the sample outlet tubing for blockage. If normal flow cannot be obtained, contact GDS Corp. To check the Low Flow switch, set the Run / Cal valve to the blocked position (pointing straight up). The flow switch should activate and the FLOW OK message will be replaced by a FAULT indication. 4.2 Adjust Alarm Levels Using a magnetic wand, select the GASMAX Main Menu by placing the magnet next to the EDIT switch located on the lower right side of the GASMAX II display. Select ALARMs to adjust the levels, polarity on/off delays and other alarm-related settings. 4.3 Verify Remote Readings From the Main Menu, select DIAGNOSTICS. The GASMAX II will request an Access Code. Enter the access code by moving the magnetic wand over the UP key four times. The UP key is located on the upper left side of the GASMAX display. Activate the EDIT key to confirm. Select OUTPUT SIMULATION to set the 4-20mA signal to any value between 4ma (Zero) and 20mA (Full Scale). Confirm that the reading on the GASMAX matches the reading shown on any remote display or controller. Select RELAY TEST to activate the internal alarm relays and confirm that any remotely connected devices operate as expected. 4.4 Initial Calibration The GDS-58XP is now ready for initial calibration. See Section 5.2 for detailed instructions. Page 10

GDS-58XP Users Manual V2.0 4 OPERATION & MAINTENANCE 5.1 Overview Standard maintenance for the GDS-58XP consists of periodic checks on flow settings and sensor calibrations. Each time a toxic sensor is calibrated, a Sensor Life reading will appear that gives an approximate indication of the remaining sensitivity. Sensor Life is not necessarily linear and a rapid reduction in the senor life reading can be due to temperature extremes, high levels of target gas, the presence of certain gases that poison toxic sensors and other environmental factors. Sensor life readings also appear for combustible or PID sensors, but as these sensors operate on different principles and/or contain microprocessors that compensate for aging effects, the values shown may not accurately represent sensor life over time. Always check the flow meter for the presence of moisture. In the event that moisture or liquid is drawn in the GDS-58XP, the entire unit should be disassembled and cleaned. In some cases the flow meter or flow switch may need to be replaced. If liquid is drawn into the GDS-58XP, always inspect the sensor for any sign of damage. 5.2 Calibration On average, calibration should be performed approximately every two to three months, depending on target gas and sensor performance. Obtain a cylinder of calibration gas that is approximately ½ scale (For example, for a 0-100 ppm range, obtain 50 ppm gas). If it cannot be guaranteed that the ambient air contains no target gas, also obtain a cylinder of zero air. In addition, a fixed flow regulator (0.5 to 1.0 liters / minute) and connecting tubing are required. Finally, make sure you have a magnetic wand to activate the GASMAX II monitor user interface. Important: Using the magnetic wand, activate the NEXT key until the GASMAX shows the dual display screen (see Fig 4.). If the message FLOW OK message is shown on the TOP display, the unit contains a BRIDGE type sensor and calibration should be done to the BRIDGE/MA Channel ONLY. If the FLOW OK message is shown on the BOTTOM display, the unit contains a TOXIC sensor and calibration should be done on the O2/TOXIC Channel ONLY. Page 11

GDS-58XP Sample Draw System Users Manual V2.1 Set the Run / Cal valve to the CAL position (pointing to the RIGHT). If the area may contain ppm levels of the target gas, attach a cylinder of Zero Air to the cal port using the fixed flow regulator. Open the valve on the regulator and confirm that the flow meter indicates gas flow. Using the magnetic wand, activate the NEXT key until the GASMAX shows the dual display screen (see Fig 4.). Put the GASMAX II in CAL MODE by activating the CAL key (lower left) followed by the EDIT key to confirm. When the SELECT INPUT screen appears, select O2/TOXIC or BRIDGE/MA as necessary (see above). The display will indicate when to apply ZERO gas. Once the reading is stable, activate the EDIT key to confirm. The display will now say to apply SPAN gas. ALM RST UP/ ALM1 ALM2 FAIL NEXT 17.8 % O2 Def. 0 % LEL DOWN/CAL EDIT A1 A2 RXD RS485 TXD Remove the cylinder of zero air and attach the cylinder of calibration gas. Open the regulator and verity flow on the flow meter. Once the reading stabilizes, activate the SPAN key to confirm. If readings are within proper limits, the message CAL SUCCESSFUL will appear. Fig 4 Dual Display Mode. Disconnect the cylinder of calibration gas and set the Run / Cal valve to the Run position (pointing LEFT). 5.3 Sensor Removal / Replacement To remove & replace the GDS-58XP sensor: 1) Loosen and remove the stainless steel tubing that is attached to the sensor flow cell 2) Loosen the set screw located at the top of the sensor flow cell 3) Remote the flow cell by gently pulling down and twisting at the same time. 4) Unscrew the sensor head cover. Use sensor head wrench (p/n 10-0187) if stuck. 5) Remove the sensor by pulling straight down. DO NOT TWIST. 6) Install the new sensor by aligning the arrow on the sensor with the engraved arrow on the sensor head. Press UP to install sensor. DO NOT TWIST. 7) Reinstall the stainless steel sensor head cover. Tighten securely. 8) Reinstall the sensor flow cell. Use approved O-ring grease (p/n 1200-0064) if necessary. 9) Reinstall the stainless steel tubing and tighten securely. 10) Verify operation and calibrate new sensor. Page 12

GDS-58XP Users Manual V2.0 5 SPARE PARTS 6.1 GDS-58XP Spare Parts 1200-0047 Flow switch, non-explosive 1200-0234 Pump, 24VDC brushless 1200-0056 Flow meter, 0-1 liter / minute with adjustment 1200-0034 Flashback arrestors 10-0205 Flow cell 6.2 GASMAX Spare Parts 10-0232 Display assembly 10-0234 Relay / MODBUS interface board 10-0233 Power supply board with ribbon cable and standoffs 6 WARRANTY 7.1 Overview The GDS-58XP electronics is warranted for a period of two years from date of shipment. The Sample Pump and Sensor are warranted for a period of one year from date of shipment. Toxic sensors have a limited shelf life and will lose sensitivity over time, even if not used. Combustible and PID sensors have no such shelf life limitations. Page 13

GDS-58XP Sample Draw System Users Manual V2.1 7 APPENDIX 8.1 MODBUS Registers The GASMAX II gas monitor provides more than 100 readable MODBUS variables. Shown below are three that can be used to read system values and alarm status. See the GASMAX manual for more information and a complete listing of MODBUS registers. Register Alias Read Value Notes D2A Raw Ch1 31001 4 200-1000 Output from O2/Toxic type sensors D2A Raw Ch 2 31002 4 200-1000 Output from Bridge type sensors Alarm Status Word 31007 4 Bits See GMII manual for bit definitions Ch1 Sensor Life 31009 4 0-100 Calculated O2/Toxic sensor life % remaining Ch2 Sensor Life 31010 4 0-100 Calculated Bridge sensor life % remaining 8.2 Moisture Calculations When drawing a sample that contains significant moisture, it is important to understand how much moisture may be entrained within the coalescing filter. At 0.5 liters per minute, the GDS-58XP will draw approximately 720 liters (0.72 m 3 ) of sample in a 24 hour period*. Sample Temp Sample RH Water content, g/m 3 Total water content Total coalesced water in 24 hrs* 20 C / 68 F 40% RH 9.4 g/m 3 3.8 g/m 3 2.7 g = 0.09 oz 60% RH 9.4 g/m 3 5.6 g/m 3 4.0 g = 0.13 oz 80% RH 9.4 g/m 3 7.5 g/m 3 5.4 g = 0.18 oz 100% RH 9.4 g/m 3 9.4 g/m 3 6.8 g = 0.23 oz 30 C / 86 F 40% RH 30.4 g/m 3 12.1 g/m 3 8.7 g = 0.29 oz 60% RH 30.4 g/m 3 18.2 g/m 3 13.1 g = 0.44 oz 80% RH 30.4 g/m 3 24.3 g/m 3 17.5 g = 0.59 oz 100% RH 30.4 g/m 3 30.4 g/m 3 21.8 g = 0.73 oz 40 C / 104 F 40% RH 51 g/m 3 20 g/m 3 14.4 g = 0.48 oz 60% RH 51 g/m 3 31 g/m 3 22.3 g = 0.75 oz 80% RH 51 g/m 3 41 g/m 3 29.5 g = 0.99 oz 100% RH 51 g/m 3 51 g/m 3 36.7 g = 1.2 oz 50 C / 122 F 40% RH 83 g/m 3 33 g/m 3 23.8 g = 0.8 oz 60% RH 83 g/m 3 50 g/m 3 36 g = 1.21 oz Page 14

GDS-58XP Users Manual V2.0 80% RH 83 g/m 3 66 g/m 3 47.5 g = 1.6 oz 100% RH 83 g/m 3 83 g/m 3 59.8 g = 2.0 oz For example, the standard filter will hold approximately 3 oz of coalesced water. At 20 C / 40% RH, the filter will need to be drained every 30 days. For a sample at 40 C and 80% RH, the filter may need to be drained every three days. For high moisture applications, GDS Corp recommends the combination coalescing / bypass filter that will remove moisture and particulate in a continuous stream. For ambient sources of sample gas, this may require the installation of an upstream pump to force the sample through the filter and bypass port. 8.3 GASMAX Configuration The GASMAX gas monitor is a two-channel device, and the GDS-58XP relies on the unused GASMAX channel to monitor the low flow switch and generate a relay contact closure (FAULT) if the low flow condition is detected. If a BRIDGE-type sensor is installed (Combustible, Infrared for Photoionization), then the flow switch is monitored by the O2/TOXIC channel (Ch1). If an oxygen deficiency or toxic gas sensor is installed, the flow switch is monitored by the BRIDGE/mA channel (Ch2). To properly configure the GASMAX in the event the display is replaced, or if a Cold Boot is necessary to restore proper functionality, the following settings will need to be made to the default settings: GDS-58XP with Oxygen or Toxic Gas Sensor: 1. Turn on the Bridge/mA Channel by holding the magnet over the NEXT key until the menu shows Enter Key Sequence menu. Enter four UP keys to access the XMITTER CONFIG menu. Select BRIDGE/mA and hit EDIT to turn the channel ON. Hit the NEXT key to exit. The GASMAX will reboot. 2. Setup EUNITS: Once the GASMAX reboots, hit the NEXT key unit the dual display appears. Hit EDIT once and then once again to access the Select Input menu. Select the O2/TOXIC channel and hit EDIT again. Under System Config, change the EUNITS to read FLOW OK. 3. Ensure that the flow switch is connected to the R and C terminals on TB2 on the back of the display module. Polarity is not important. GDS-58XP with Bridge/mA Combustible, Infrared or Photoionization Gas Sensor: 1. Setup EUNITS: Hit the NEXT key unit the dual display appears. Hit EDIT once and then once again to access the Select Input menu. Select the BRIDGE/mA channel and hit EDIT again. Under System Config, change the EUNITS to read FLOW OK. 2. Ensure that the flow switch is connected to the _ and terminals on TB1 on the Power Supply board attached to the back of the explosion proof enclosure. Polarity is not important. Page 15