SIM HYDROGEN SENSOR. Manual. Installation and Calibration of the External Device for Agilent 7890 GC. Scientific Instruments Manufacturer GmbH

Similar documents
USER S GUIDE SPARE PARTS SPARE PARTS FOR SOS-LIFE EARTHQUAKE/EMERGENCY ALARM. SOS-LIFE earthquake/emergency alarm unit

Installing a Fast Oven Upgrade Kit

CP01 Fixed gas detector

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

Indirect Cryogenic Cold Trap System Instructions

Advanced Test Equipment Rentals ATEC (2832) Installation Note

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Keysight 85025/85026 Detector Exchange Program - All Serial Numbers

Protégé Eclipse LED Keypad User Manual PRT-KLES

DH400. Accessories. NOTE: Set up for this sensor should be performed by an AAADM-certified installer.

OWNERS MANUAL MODEL 451 (-101 & -201)

SCHMIDT LED Measured Value Display MD Instructions for Use

Operations Manual TS400. Test Station for G450/G460 Gas Detector

HOBOT. HOBOT-168 Glass Cleaning Robot User s Manual

Technical Specifications and User Guide for the GDCO Detector

User s Manual For Model Aero-551 (Rev. Orig.)

Model 405 AIRGARD Fume Hood Monitor

HAKKO 703B REWORK SYSTEM TROUBLESHOOTING GUIDE

Document No

status AW1 WiFi Alarm System Printed in China PA : AW1-UM-EN-V1.0 User Manual 2016 Chuango. All Rights Reserved.

INSTALLATION AND INSTRUCTION MANUAL

Owner s Manual. Model AC375C Refrigerant Recovery, Recycle, and Recharge Unit

Manual No: Revision: C. Overfill Alarm. Installation Guide

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature

ACCURATE ELECTRONICS INC

Environmental Monitoring SmartSlot Card

External Wireless Sounder

4-Row Active Infrared Presence Detector for Automatic Door Control. Cover

Agilent Electronic Crimping and Decapping Tools Operating Guide. Agilent Technologies

Technical Description: / Rev, Seq: F, 2 / Date: Remarks. At Customer error case. Blinking LED's at end of Flash process

TC100N. Fixed Gas Detector Operation Manual

AFA 500 FUME HOOD ALARMS

User Manual. GPD 3000 e. gas detector

IF79 CAUTION CONTENTS YOUR THERMOSTAT REPLACES 1 PREPARATIONS. Installation Instructions for. Heating & Air Conditioning

DHR3. Accessories. NOTE: Set up for this sensor should be performed by an AAADM-certified installer. INSTALLATION INSTRUCTIONS

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual

! "! # Please read these installation and operating instructions completely and carefully before starting.

SIMRAD GD10PE IR GAS DETECTOR

Intruder alarm panel Terxon MX Operating instructions

ACCURATE ELECTRONICS INC PO BOX SW HALL BLVD BEAVERTON OR USA FAX

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL

CO2 Gas Warning System Assembly and Operating Instructions

Installation and Wiring 1.

Sentry LIQUID LEVEL CONTROLLER MODEL 120 OPERATING MANUAL.

CASS9000 Standalone FingerPrint Door Access System. User Manual

Technical Description: / Rev, Seq: H, 1 / Date: Remarks. At Customer error case. Blinking LED's at end of Flash process

testo 316-EX gas leak detector Instruction manual

UFM-T100H US ES CA Power Alarm Test TM

IMR IX176 Portable Gas Detector User Manual

SIMRAD GD10 IR GAS DETECTOR

Operating Manual. PRO 1000 Series EFS-01. Non-Programmable Thermostat

FPD - Flame Photometric Detector

Environmental Monitoring Card

Buzzer Unit. TAN-5000 Series. Operating Manual

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Digital Heat Block User Manual

2-Port alarmcharge Hub - Manual

MODEL OX-07 Type A INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR PT0E-0860

Operating instructions Page 12

CARSON MANUFACTURING CO., INC NORTH RURAL STREET INDIANAPOLIS, IN (888) TECHNICAL BULLETIN

This instruction manual explains the use of three models listed on the left.

2) This manual covers Fire and General Alarm systems. The differences are described in the appropriate sections.

Gas Detection - GD -

2 THERMOSTAT DETAILS 3 REMOVING OLD THERMOSTAT

Fire Command Keypad. XR5 User s Guide

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

HA06. User manual. (SDO Hot Air One) HOT AIR / QUARTZ REFLOW OVEN. Version 2.21

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

CARSON MANUFACTURING CO., INC NORTH RURAL STREET INDIANAPOLIS, IN (888) TECHNICAL BULLETIN

G-TECTA Cellaguard Carbon Dioxide and Oxygen Fixed Monitors

CellarSafe. Carbon Dioxide and Oxygen Safety Monitor

Introduction... 1 System Overview... 1 System Diagram... 2

SIMRAD GD10P - IR GAS DETECTOR OPERATING MANUAL

Setup instructions. Honeywell GAS DETECTOR. Overview. Accessories The instruction x1 Plug dowel x2 Screw x2 Double-sided adhesive tape x2

TECHNICAL INSTRUCTIONS

SA VOLT SIREN AMPLIFIER

LMA LINE MONITOR ALERT USER S MANUAL

Portable type of Carbon Dioxide Detector Model: GT220

SERVICE INSTRUCTIONS FOR. Fire Alarm System MINI-2000

Issue 4 10/10/11. Part Number: 48160

INSTRUCTION MANUAL FOR RIKEN PORTABLE TOXIC GAS MONITOR

Chore-Tronics Load Cell Indicator

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time

NWA300 Field-Based Analog 4-20mA DC Current Meter User's Manual OM

Operating Instructions Version 1.2

Sentry LIQUID LEVEL ALARM MODEL 100 OPERATING MANUAL.

RK Transmitter Technical Notice

MANUAL Oil Level Sensor

NG-9B NATURAL GAS ALARM

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

INSTALLATION AND OPERATION INSTRUCTIONS

OWNERS MANUAL. CARBON MONOXIDE DETECTOR MODELS Panel Mount & Remote Detectors (R) Guardian Avionics 1951 E. AIRPORT DRIVE TUCSON, AZ.

Mobeye CM2410 GSM fire alarm communicator

1F Non-Programmable Electronic Digital Multi-Stage Thermostat INSTALLATION AND OPERATION INSTRUCTIONS

GS100M Gas Detector. Technical Specification

DC1136, DC1131-AA, DC1134-AA, DC1136-AA Input / Output modules

Static Pressure Control

GAS MONITOR Model Model

User s Guide Home Automation Wall Socket for IWATCHALARM (Add-On) Model: SM-001SK

Operations Manual GMA 313 CO 2. Gas Monitor for one detection point

Transcription:

Manual SIM HYDROGEN SENSOR Installation and Calibration of the External Device for Agilent 7890 GC SIM Hydrogen Sensor (V 1.10) Page 1 of 7

Document History Eighth edition 08/2010, Version V1.10 The information contained in this document is subject to change without notice. SIM GmbH makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. SIM GmbH shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. Copyright 2010, SIM GmbH. Printed in Germany. All rights reserved. CONTENT 1 PRINCIPLE OF OPERATION... 3 2 ACCESSORIES KIT... 3 3 INSTALLATION OF THE HYDROGEN SENSOR... 4 3.1 Installation of the gas detector head... 4 3.2 Carrier Gas Connection... 5 4 OPERATING MODE... 6 5 CALIBRATION... 6 5.1 Setting to zero point... 6 5.2 Setting of the alarm value... 6 5.3 Stop calibration mode... 7 6 OPTICAL ERROR MESSAGE... 7 7 SPECIFICATIONS... 7 SIM Hydrogen Sensor (V 1.10) Page 2 of 7

1 PRINCIPLE OF OPERATION The hydrogen sensor constantly detects the hydrogen concentration in the GC oven. In case of a leakage, optical and acoustic alarm will be set off and the carrier gas will automatically be switched to inert gas. One green LED shows the ready status of the hydrogen sensor with no hydrogen in the GC oven. As soon as hydrogen appears, several of the LEDs will light up depending on the gas concentration. If the concentration is above a threshold level of 10 % of the lower explosion limit (LEL), the first red LED will light up and an acoustic signal will be transmitted. At a detected hydrogen level of 25% LEL (equals 1 % by vol. H2), an alarm signal will sound and the carrier gas will automatically be switched to an inert gas. At the same time the GC gets a not ready signal so that the GC run will stop. This prevents chromatographic separations with the wrong carrier gas. When the hydrogen concentration level falls under the set limit of 1% H2 by vol. (25% LEL) the red LEDs will start flashing (inert gas mode) and the remaining LEDs will show the actual hydrogen concentration. After eliminating the leakage, pushing the reset key (on the left beside the LEDs) switches back automatically to carrier gas, the gas chromatograph gets a ready signal and the indicator bar light works normally again. 2 ACCESSORIES KIT sensor box with carrier gas switch gas sensor (for installation into GC) together with fastener, cable and plug connector, 2 tapping screws drilling template for gas detector head 4 drill bits (8 / 6 / 4.3 / 3.8 mm) power supply and power cable Start/Remote cable sensor box SIM Hydrogen Sensor (V 1.10) gas sensor Page 3 of 7

3 INSTALLATION OF THE HYDROGEN SENSOR 3.1 Installation of the gas detector head Put off the right side cover of the GC. Detach the precut sheet in the middle of the right side of the oven (see fig. below with the mounted gas detector head). Inside of the GC oven with mounting device and gas sensor head (left) and mounting device (above, longish installation sheet is situated in the oven side wall) Prepare the hole for the detector head from outside the GC oven: Carefully detach the back detector board and place the round drilling template onto the round blanking (see fig. below) Detector board (detached) Drilling template Right outside of the GC with the detached detector board and the drilling template onto the round blanking Open up the hole by drilling with a 6 mm drill bit and take an 8 mm drill bit to bore through into the GC oven space. Please note that there should not remain any isolation material in the hole (remove it by blowing or putting the sensor head in and out)! Inside the GC oven: Push the longish installation sheet of the mounting device through the created hole so that the square sheet of the mounting device can be mounted in front of it at the oven inside (see figures above). SIM Hydrogen Sensor (V 1.10) Page 4 of 7

Push the gas detector head from outside through the isolation and the installation sheets as far as it will be possible. Mounting of the gas sensor head at the outside of the GC Outside the GC oven: Drill mounting holes for the detector head (4.3 mm and then 3.8 mm drill bit) and screw the detector tightly with two screws (see fig above). Reassemble the back detector board. The gas sensor cable has to be connected to the circuit board of the hydrogen sensor at the end of the installation (see 3.6). Inside the GC oven: Fasten the detector head mounting device with the two screws of the accessory kit (see first fig of 3.1). Remount the side cover of the GC. 3.2 Carrier Gas Connection NOTE: The pressure of hydrogen and inert gas has to be 7 bar each, otherwise the installed valve does not work correctly! Attach the hydrogen and inert gas pipe of the gas supply to the gas ports situated at the back side of the sensor box (see fig. on the left, hydrogen H 2 in, inert gas inert in). Connect the port in the middle ( out ) with the carrier gas port (EPC) of the GC. Put the gas sensor plug into the port Sensor (second row, left). Connect the Start/RemoteCable with the AUX port of the sensor box and the Remote outlet of the GC (see fig. on the right) Connect the power supply (enclosed) with the port 24 V DC. Sensor box connections (left) and back side of the GC with remote outlet (right) SIM Hydrogen Sensor (V 1.10) Page 5 of 7

4 OPERATING MODE After installing the gas sensor head and connecting carrier and inert gas as current supply, the Hydrogen Sensor is ready for use after pressing the reset key: red LEDs are flashing: green LED on the left is lit: green LED is flashing: inert gas mode, notready signal for GC after first installation of the hydrogen sensor or switchingon the gas chromatograph or previous hydrogen alarm: pressing the reset key (on the left beside the LEDs) switches to carrier gas hydrogen and the ready signal for GC, so that the hydrogen sensor is in the ready mode (one green LED) or rather shows the actual hydrogen concentration in the GC oven ready mode zero point has to be recalibrated red LEDs + acoustic alarm: hydrogen leak, carrier gas is switched to inert gas and not ready signal for GC: eliminate the hydrogen leak and return to the ready mode by by pressing the reset key (automatic switchover from inert gas to hydrogen and ready signal for GC) 5 CALIBRATION 5.1 Setting to zero point (after installation) run the system for 10 15 min press reset button until you can hear a beep (approx. 2 3 s): calibration mode is active shortly press the reset button 2 times: Setting to zero point begins, the green LEDs are flashing after ½ minute, zero point setting is done, a beep resounds automatically, and the sensor skips back to the measuring mode 5.2 Setting of the alarm value (25% of LEL) (factorypreset, anual recalibration by service engineer) attach calibration standard (air with 1% by vol. H 2 ) to the gas sensor so that the standard mixture continuously flows through the sensor head (12 l/min) press reset button until you can hear a beep (approx. 2 3 s): calibration mode is active SIM Hydrogen Sensor (V 1.10) Page 6 of 7

shortly press the reset button 3 times: Setting to alarm value begins, the yellow LEDs are flashing after ½ minute, alarm value setting is done, a beep resounds automatically, and the sensor skips back to the measuring mode detach calibration standard 5.3 Stop calibration mode If you want to drop out of calibration mode: shortly press the reset button once: sensor skips back to the measuring mode 6 OPTICAL ERROR MESSAGE Red LED on the right is flashing: oven deflector fails or oven is opened and the carrier gas has switched automatically to inert gas in case of operating deflector or closed oven, the sensor automatically skips back to the measuring mode 7 SPECIFICATIONS detection range: 0 25 % LEL ( 0 1.0 % by vol. H 2 ) 0.4 % by vol. H 2 : optical alarm: first red LED, acoustic alarm: starts with intermittent beep 1.0 % by vol. H 2 : optical alarm: 4 red LEDs, acoustic alarm continues, carrier gas is switched to inert gas (N 2, He) size (control box): 13 cm (w) x 11,5 cm (d) x 4 cm (h) electrical properties: 100240 V, 50/60 Hz, 1.0 A SIM Hydrogen Sensor (V 1.10) Page 7 of 7