1 G-TECTA Cellaguard Carbon Dioxide and Oxygen Fixed Monitors UK Installation & Calibration Manual
2 2 Contents Safety and Warning information 2 Installation 2 Mounting 3 Cabling Requirements 4 External Alarm Equipment 4 Internal Battery Backup 4 Commissioning 4 LED Status 5 Alarm Conditions 6 Self Test 6 Calibration 7 Calibration Gases 7 Zero Gas Calibration 8 Span Gas Calibration 8 Calibration Unit Disconnection 9 Fuses 9 Whilst The Linde Group has made every effort to ensure that the details and information given in both our printed and online publications are accurate at the time of issue, full technical specifications are not necessarily included. Furthermore The Linde Group policy is one of continuous improvement and the right is reserved to alter details and information as the need arises. Accordingly the Customer should check any details and information they wish to rely on with The Linde Group at the time of purchase. The Linde Group can not accept liability in respect of any errors or omissions herein contained or for any loss or damage malfunction or consequential loss arising from reliance upon our publication. The customer will be responsible for any risk to health or safety from goods in the Customer s possession and/or control. The Customer s attention is drawn to the fact that statutory regulations and recognised codes of practice exist covering the use and handling of some goods (including safety products.) The Customer must ensure that persons who use the goods receive adequate training and safety literature. Safety and Warning information WARNING Ensure that this monitor is installed in accordance with this manual. This product may not be able to warn of gas hazards if installed in any other way. WARNING Do not substitute components. WARNING G-TECTA Cellaguard is not suitable for outdoor or hazardous environments. Installation When carrying out any installation work, ensure that local regulations and site procedures are followed. General G-TECTA Cellaguard is a carbon dioxide (CO 2 ) or a carbon dioxide and oxygen (O 2 ) fixed monitoring system, designed to monitor gas levels in a confined space such as a beer cellar. The Cellaguard system consists of a wall mounted base station, housing the gas monitoring system, and is normally supplied with one repeater unit to be located in a safe position outside the confined space area. G-TECTA Cellaguard features two alarm relays, which can control external visual or audible alarms or ancillary equipment such as a ventilation fan. The Cellaguard system is simple to install using a mains power supply and has been designed for one-man installation. The alarm levels are factory set and the relays are fail-safe to ensure the alarms will activate if the power supply fails. Note: For further information on the operation and use of the G-TECTA Cellaguard gas detection monitor, please see the Operating Manual. This manual contains instructions for installing the Cellaguard base station and repeater unit, for connecting ancillary equipment and for calibrating the gas sensors mounted in the base station.
3 3 Before beginning installation of the G-TECTA Cellaguard please read the mounting, cabling and installation instructions, paying close attention to the location of the base station. Mounting Base Station The Cellaguard base station is designed to be mounted on a wall and should be located near to the valves and manifold of the gas source (within the confined space), approximately 0.5m from the ground (see Figure A). Note: Ensure the mounting position is within 2m of an electricity supply point and is not adjacent to a water or a moisture source. Fixing holes are attached to the base station so that it can be attached to the wall with appropriate screws and wall plugs. Ensure the wall plugs used are rated for a weight of 1.3kg. Repeater(s) The Cellaguard repeater unit(s) should be mounted beside the entry point(s) to the area being protected, (outside of the confined space) and opposite the door hinges, using the fixing brackets provided (see Figure A). The height of the Repeater(s) should be at eye level (approx 1.7m from the ground). Additional Repeater units can be mounted using the same method (maximum 4 per base station). Additional repeater units are supplied with 9m cables that connect in a daisy chain arrangement. Suitable locations are in areas normally occupied by staff to ensure any alarms are immediately noticed. WARNING G-TECTA Cellaguard is not suitable for outdoor or hazardous environments. Carbon Dioxide Warning Signs The Carbon Dioxide warning labels supplied with the Cellaguard should be mounted adjacent to the Repeater unit(s). Additional Repeater Unit Repeater Mains 0.5m Figure A: Where to locate and mount the Cellaguard
4 4 Cabling requirements Cellaguard is fitted with an AC plug for convenience. If, however, permanent installation is required, it should be wired to a dedicated AC spur led from a circuit breaker. The spur or mains plug should be fitted with a 3A fuse. The circuit breaker should be marked as the disconnecting device for the Cellaguard. It must comply with the relevant requirements of IEC and IEC These procedures ensure that the installation complies with the requirements of EN (Low Voltage Directive). Cellaguard Repeater units do not need specialist cabling. Each repeater unit is supplied with a 9m cable as standard. Electrical connections for external alarm equipment The Cellaguard has two alarm relays that can be connected to external visual and audible alarms or other ancillary equipment, such as a fan. An example electrical connection is shown in Figure B. WARNING Before connecting any additional equipment, ensure the mains supply is switched off. Unscrew the top cover to gain access to the screw terminal connectors on the PCB board. The two alarm relay connections are each marked NC (Normally Closed), NO (Normally Open) and Common (C). Use the NO and C screw terminal connectors as shown in Figure B. (L = live, N = neutral) Note: The alarm relays are shown in their normally energised state. When installing the device the normally open (NO) connection will be closed since there is no power. Similarly, the normally closed (NC) connection will be open. The Cellaguard can be connected to one or two pieces of external equipment or to one dual speed fan. Each alarm relay corresponds to the low and high alarms on the Cellaguard. If two pieces of equipment (e.g. two fans) are connected and the low alarm is exceeded the first fan will activate. If gas levels increase further, setting off the high alarm, the second fan will activate to increase ventilation. For dual-speed fans, low speed and high speed will correspond to the low and high alarm relays respectively. Note: It is recommended that external equipment should be fused. FAN Relay 1 Relay 2 NC NO C L N NC NO C WARNING The relay contacts must not be used to switch loads drawing more than 10A Figure B: Alarm relay electrical connections Internal Battery Backup Cellaguard monitors specified with battery backup will be supplied with the batteries disconnected. Connect the battery unit to the connector marked PL1 on the printed circuit board in the Cellaguard unit before switching on. Cellaguard Commissioning The Cellaguard is designed to work out of the box and requires no configuration of the gas sensors or alarm settings. Once the Cellaguard base station and repeater unit(s) are installed, you can switch on the mains supply. The Cellaguard system will become fully operational after a 40-second warm-up period. Note: The LED alarms and sounder will operate momentarily during this period and the display will self-test. L N
5 5 WARNING When commissioning the Cellaguard, the customer must be shown where the Cellaguard is installed and instructed that the Cellaguard must NOT be disconnected from the power supply. The LED status, the alarm conditions (shown below) and the Carbon Dioxide warning label should be explained to the customer as part of the installation process. WARNING In the unlikely event that the Cellaguard monitor is installed with no permanent electrical supply available, the Cellaguard should be tested with a temporary power supply to ensure functionality. The Cellaguard should be covered or labelled as NOT FUNCTIONING before leaving the customer s premises. Normal and fault states of LEDs The LEDs on the base station and repeaters indicate conditions as shown in the table below: LED Function Operation Indication Power Normal operation LED flashes (Green LED) Not operating LED extinguished Fault condition LED on continuously Power failure Sounder beeps every 20 seconds Fault System fault LED flashes (Yellow LED) Low Alarm (Red LED)* High Alarm (Red LED)* Low (pre-alarm) alarm level exceeded High (main) alarm level exceeded LED flashes LED flashes
6 6 G-TECTA Cellaguard Alarm conditions for Base Station and Repeaters When the alarm gas level is exceeded the sounder and alarm LEDs will operate as below: Gas Levels LED Indication Sounder Indication Relay Indication Normal No alarm LEDs No sound Not active Carbon Dioxide Alarms Low alarm activates Low alarm LEDs flashing Long beep once per second Pre-alarm relay connection active High alarm activates Low and high alarm LEDs flashing Long beep once per second Pre-alarm and main alarm relay connections active Oxygen Alarms Low alarm activates Low alarm LEDs flashing Long beep once per second followed by one short beep Pre-alarm relay connection active High alarm activates Low and high alarm LEDs flashing Long beep once per second followed by one short beep Pre-alarm and main alarm relay connections active Note: CO 2 alarms are subject to a 25 second time delay to prevent false alarms when exchanging gas cylinders. Oxygen alarms function instantaneously. When the gas level(s) have returned to normal, the alarm LEDs will automatically extinguish and the Cellaguard will return to normal operating mode. Note: The sounder and illuminating LEDs will reactivate if the alarm level(s) are exceeded again. G-TECTA Cellaguard Fault Alarm The system fault alarm will activate if a sensor fault condition occurs. If the system fault alarm is activated the G-TECTA Cellaguard will beep twice every second and the yellow warning LED will flash. G-TECTA Cellaguard Mode Button (Alarm function) The mode button is used to communicate with the Cellaguard system after an alarm has been triggered. Press and hold the mode button for 5 seconds to silence the Cellaguard alarm sounder. The alarm LEDs will continue to flash in an alarm condition (i.e. CO 2 gas is still present or there are low levels of oxygen). G-TECTA Cellaguard Self-Test A self-test routine is provided to test the Cellaguard and any ancillary equipment attached via the alarm relays. To run the self-test routine, press and hold the mode button for 5 seconds. The self-test can be initiated from the base station or any of the repeater units. The base station display will illuminate and test all display segments and the Cellaguard will alarm for a period of 10 seconds; illuminating the alarm LEDS, activating the sounder and activating any additional ancillary alarm equipment connected to the Cellaguard relays.
7 7 Calibration These instructions provide guidelines for the service engineer to calibrate the Cellaguard using the Cellaguard calibration unit (Figure C). The calibration unit is suitable for all models of the Cellaguard gas monitoring system. Calibration requires calibration gases and a pair of bayonet calibration adapters. In calibration mode all alarms and relays are inhibited. Calibration gases Zero calibration is performed using 100% nitrogen. Span calibration is performed using a gas mixture comprising 4.0% carbon dioxide and 20.75% oxygen with a balance of nitrogen. Bayonet calibration adapters A single pipe and bayonet fitting is provided for the Cellaguard carbon dioxide only monitoring system. The adapter for the Cellaguard carbon dioxide and oxygen monitoring system is supplied with a T junction and a single common supply pipe. The flow of the calibration gas should be set to 750 ml/min for the carbon dioxide only monitoring system, and 1.5 l/min for the dual gas monitoring system. Calibrating the Cellaguard gas monitor WARNING Switch off all power and disconnect any batteries BEFORE connecting the Calibration unit. Failure to do so will risk damage to the Cellaguard system. Figure C - Cellaguard Calibration Unit 0 Cal Span Cal G-TECTA CELLAGUARD CAL UNIT Connecting the Calibration Unit 1. Switch off the mains supply to the Cellaguard base station and any ancillary equipment connected to the alarm relays 2. Remove the front cover and disconnect the batteries (if fitted). 3. Plug the lead from the calibration unit into the connector marked PL2 on the printed circuit board 4. Replace the front panel of the Cellaguard base station. Take care not to over tighten the screws 5. Reconnect the mains electrical supply to the Cellaguard base station 6. Allow the instrument to stabilise for a few minutes. It is essential the Cellaguard reaches the appropriate thermal equilibrium for the gas sensors to work correctly and for the calibration to be performed without error. Note: In the unlikely event of a communication problem between the calibration unit and the base station, both LEDs will flicker rapidly to alert the user. If this occurs, the system may be reset by disconnecting the mains electrical supply for a few seconds (the battery should already be disconnected) and re-applying the mains power. Connecting the zero calibration gas WARNING Zero gas calibration MUST BE performed before span gas calibration 1. Insert the bayonet of the calibration adapter into the sensor inlet(s) by gently twisting clockwise until they snap into place 2. Connect the other end of the calibration adapter to the zero calibration gas (100% nitrogen) 3. Without opening the cylinder, adjust the gas flow to 750 ml/min for the carbon dioxide only monitoring system and 1.5 l/min for the dual gas monitoring system. 4. Allow the system to stabilise for several minutes.
8 8 Note: Once connected, the calibration unit can be switched on or off by pressing and holding the Scroll/ Power button for 3 seconds Zero gas calibration 1. Press the Scroll/Power button to turn on the calibration unit. The Zero Calibration or the Span Calibration LED will illuminate. The base station is now in calibration mode, the Cellaguard display will show alternately the internal temperature of the instrument and the gas reading. 2. If the Zero Calibration LED is not lit, press the Scroll/Power button. 3. Apply zero calibration gas by opening the regulator using the settings above, and wait for the gas reading in the Cellaguard display to stabilise 4. Press the Select button to start the calibration. The Zero Calibration LED will flash to indicate that calibration is in progress. After approximately one minute the Zero Calibration LED will stop flashing and zero calibration is complete Note: The carbon dioxide reading will briefly indicate an error on the operator display panel for a few seconds immediately after a calibration has been performed. This is normal. To abort the calibration process, press the Scroll/ Power button Span gas calibration 1. Connect the calibration adapter to the span calibration gas 2. Without opening the cylinder, adjust the gas flow to 750 ml/min for the carbon dioxide only monitoring system and 1.5 l/min for the dual gas monitoring system. 3. Allow the system to stabilise for several minutes. 4. Press the Scroll/Power button to illuminate the Span calibration LED. The base station is now in calibration mode and the operator display panel will display alternately the internal temperature of the instrument and the gas reading 5. Apply span calibration gas by opening the regulator and wait for the gas reading on the Cellaguard display to stabilise 6. Press the Select button to start the calibration. The Span Calibration LED will flash to indicate that calibration is in progress. After approximately one minute the Span Calibration LED will stop flashing and span calibration is complete Note: The carbon dioxide reading will briefly indicate an error on the operator display panel for a few seconds immediately after a calibration has been performed. This is normal. To abort the span calibration, press the Scroll/Power button. 7. Switch off the calibration unit by pressing the Scroll/Power button once; then press and hold Scroll/Power button 3 seconds. Both LEDs will extinguish
9 9 Calibration Unit disconnection 1. Press and hold Scroll/Power button for 3 seconds to switch off the mains electrical supply to the base station 2. Disconnect the bayonet calibration adapter 3. Remove the front panel of the base station by removing the four screws 4. Unplug the lead from the calibration unit from the printed circuit board 5. If fitted, plug the battery unit into the connector marked PL1 on the printed circuit board 6. Refit the front panel of the base station. Do not over tighten the screws. 7. Reconnect the mains electrical supply and relays (if fitted) to the base station Cellaguard will perform a start up routine and stabilise after 2 minutes. This product has been tested and found to comply with the European Directive 89/336/EEC, the low voltage directive EN61010 and the EMC requirements of EN Fuses There are two fuses fitted in the Cellaguard. One fuse protects the AC inlet; the other 2A (anti-surge) fuse is on the PCB. Either fuse can be replaced by replacing them with a fuse of identical specification. To locate the fuses, disconnect the base station from the mains power supply, remove the four front-cover fixing screws and carefully take off the cover. The fuses are located as above.
10 10 Notes
11 Notes 11
12 Denmark AGA A/S Vermlandsgade København S Tel: Estonia Eesti AGA AS Pärnu mnt Tallinn Tel: Finland Oy AGA Ab Itsehallintokuja Espoo Tel: +358 (0) Iceland ÍSAGA ehf Breidhofda Reykjavik Tel: Latvia AGA SIA Katrīnas iela 5 Rīga 1045 Tel: Lithuania AGA UAB Office Didlaukio st Vilnius Tel: Norway AGA AS Gjerdrumsvei Oslo Tel: Sweden AGA Gas AB Agavägen Lidingö Tel: + 46 (0) V G-TECTA Cellaguard Operating Manual 2010 The Linde Group. All rights reserved. No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above. The Linde Group reserves the right to correct typographical errors. All information is correct at time of going to print.
CellarSafe. Carbon Dioxide and Oxygen Safety Monitor
CellarSafe Carbon Dioxide and Oxygen Safety Monitor Installation and operating instructions for the CellarSafe Monitor and the CellarSafe Calibration Unit M07234 May 2008 Issue 6 Warning Please take care
More informationG-TECTA Cellaguard Carbon Dioxide and Oxygen Fixed Monitors
G-TECTA Cellaguard Carbon Dioxide and Oxygen Fixed Monitors EN Operating Manual 2 3 Contents Safety and Warning information 2 Introduction 3 Operation & Set Up 3 Cellaguard Features 3 Initial Set Up 4
More informationAnalox 1000 Series. User Manual. Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK
Analox 1000 Series User Manual Analox Sensor Technology Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net
More informationANALOX SENSOR TECHNOLOGY LTD ANALOX DD1101 INSTALLATION & OPERATION MANUAL
ANALOX SENSOR TECHNOLOGY LTD ANALOX DD1101 INSTALLATION & OPERATION MANUAL SAFETY WARNING Please Refer to the Safety Warning in Section 7.1 regarding the chemicals in the oxygen sensor Analox Sensor Technology
More informationDoor Release Power Supply
www.protectingpeople.co.uk Door Release Power Supply Engineer / Installation Manual Document: VI55.1 Protecting People Printed : 04/03/2004-1 - Ventcroft Ltd Door Release Power Supply Engineer / Installation
More informationGasflag. Single Channel Control Panel Alarm Only Apparatus. Installation, operating and maintenance instructions, M07225.
Gasflag Single Channel Control Panel Alarm Only Apparatus Installation, operating and maintenance instructions, M07225 Issue 2 1/07/05 Crowcon Detection Instruments Ltd 2 Blacklands Way Abingdon Business
More information10yr O2NE & Safe-Ox. User Manual
Analox Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net This support line is closed on UK public
More informationInstallation and user manual for the BiWire / Conventional Repeater Panel
Panel EFBWCV-REPEATER Contents Introduction.... 3 Purpose... 3 The Panel.... 3 Indication Equipment (IE).... 3 Power Supply Equipment (PSE).... 4 System Wiring.... 5 Status Indications.... 5 Repeater I/O....
More informationFTEN1, FTEN2, & FTEN4 1, 2 & 4 ZONE FIRE DETECTION AND ALARM CONTROL PANELS
QUICK USER GUIDE FOR, 2 & 4 ZONE PANELS INDICATORS COLOUR INDICATION MON FAULT Yellow On when a fault condition has occurred, or if no other fault Led is on, this indicates 24v auxiliary fuse failure.
More information3500 CONVENTIONAL FIRE ALARM CONTROL PANEL
3500 CONVENTIONAL FIRE ALARM CONTROL PANEL USER MANUAL Protec Fire Detection PLC, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0) 1282 717273 Web:
More informationBeacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08
Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3
More informationOPERATING AND MAINTENANCE MANUAL. for the G30 CARBON DIOXIDE MONITOR
OM117 OPERATING AND MAINTENANCE MANUAL for the G30 CARBON DIOXIDE MONITOR (Part No: G30) (Intentionally Blank) APPROVAL SHEET DIVEX MANUAL NUMBER: ADVITIUM NUMBER: DOCUMENT TITLE: OM117 GA-OM-5715 G30
More informationMODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING
MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING FOR OPERATION FROM 115 AC POWER Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783
More informationANALOX 3000 Carbon Monoxide Monitor
ANALOX 3000 Carbon Monoxide Monitor User Manual Analox Sensor Technology Ltd 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net
More informationUPON COMMENCMENT OF INSTALLATION IT IS OF PARAMOUNT IMPORTANCE THAT THE STEP BY STEP GUIDE (INCLUDED) IS FOLLOWED COMPLETELY
Important UPON COMMENCMENT OF INSTALLATION IT IS OF PARAMOUNT IMPORTANCE THAT THE STEP BY STEP GUIDE (INCLUDED) IS FOLLOWED COMPLETELY UPON COMMENCMENT OF INSTALLATION IT IS OF PARAMOUNT IMPORTANCE THAT
More informationANALOX 5001 Carbon Dioxide Monitor. User Manual ANALOX Analox 5001 Carbon Dioxide Monitor User Manual
ANALOX 5001 ANALOX 5001 Carbon Dioxide Monitor User Manual Analox Sensor Technology Ltd 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT T: +44 (0)1642 711400 F: +44 (0)1642 713900
More informationLC1 & 2. Fire Alarm Panel 6\VWHPLQVWDOODWLRQRSHUDWLQJ PDLQWHQDQFH LQVWUXFWLRQV
LC1 & 2 Fire Alarm Panel 6\VWHPLQVWDOODWLRQRSHUDWLQJ PDLQWHQDQFH LQVWUXFWLRQV ZIRCONLC1 One Zone Conventional Fire Panel ZIRCONLC2 Two Zone Conventional Fire Panel Compliant with EN54-2:1998 & EN54-4:1998
More informationAnalox Safe-Ox/Safe-Ox He
Analox Safe-Ox/Safe-Ox He User Manual Analox Ltd. 15 Ellerbeck Court, Stokesley Business Park North Yorkshire, TS9 5PT, UK T: +44 (0)1642 711400 F: +44 (0)1642 713900 W: www.analox.net E: info@analox.net
More informationMulti Channel Gas Detection System AGDS MC. The Home of Control & Automation in Ireland
Multi Channel Gas Detection System AGDS MC The Home of Control & Automation in Ireland Multi-channel Gas Detection System The AGDS-MC, Gas detection system has been designed for use in boiler houses and
More informationQuantum Production Limited
Quantum Production Limited TELEPHONE NUMBER +44 (0) 1844 339993 FAX NUMBER: +44 (0) 1844 339996 ONE 2 ONE OXYGEN DEFICIENCY MONITOR & REPEATER ALARM Description and Application Installation and Operating
More informationAx60+ Multi-Gas. Service Manual
Analox Limited 15 Ellerbeck Court, Stokesley Business Park, North Yorkshire, TS9 5PT, UK UK/RoW T: +44 (0)1642 711400 F: +44 (0)1642 713900 US T: (714) 891-4478 W: www.analox.net E: info@analox.net Copyright
More informationNoby-448 User Operating Manual & Log Book Page 1 of 12. Fire Control Panel. User Operating Manual & Log Book
Noby-448 User Operating Manual & Log Book Page 1 of 12 Fire Control Panel User Operating Manual & Log Book TABLE OF CONTENTS Page Introduction 1 Keypad Operation 2 LED Indications 3 What to do in the event
More informationMiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING
MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2 TM, and MiniClO 2 TM Responder Detectors Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE
More informationUSER MANUAL COMBINED FIRE & EXTINGUISHING CONTROL PANEL COMPLIES WITH BS EN PART 1 & EN54 PARTS 2 & 4
PREMIER EX8 EXTINGUISHING PANEL USER MANUAL COMBINED FIRE & EXTINGUISHING CONTROL PANEL COMPLIES WITH BS EN 12094 PART 1 & EN54 PARTS 2 & 4 USER MANUAL Approved Document No: GLT.MAN-132 INDEX Extinguishing
More informationOperation and Maintenance Manual
(FKER Series) Repeater Panel (FKER, FKER, FKER8) Operation and Maintenance Manual Man-9EL Issue June Index Section Page. Introduction.... Safety and mounting.... Technical specification.... Control panel
More informationIntelligent Security & Fire Ltd
Sigma CP (K and T Series) and Sigma CP-A Conventional Fire Control Panel (K11020M2, K11040M2, K11080M2) (T11020M2, T11040M2, T11080M2) (KA11020M2, KA11040M2, KA11080M2) User Manual Man-1082 Issue 05 August
More informationUser Operation Manual
APPROVED BY: JBJ PRESCIENT III FIRE ALARM & GAS EXTINGUISHING CONTROL PANEL User Operation Manual PAGE 1 of 6 1 CONTENTS 1. FIRE - GENERAL... 3 2. FIRE - ZONES 1 OR 2 - AUTOMATIC MODE... 3 3. FIRE - ZONES
More informationPAT E N T System 88. COMPLIES WITH Singapore Standard : SS CP10 : 2005 British Standard : BS EN54-2 : 1998 BS EN54-4 : 1998 REV.00
PAT E N T System 88 COMPLIES WITH Singapore Standard : SS CP10 : 2005 British Standard : BS EN54-2 : 1998 BS EN54-4 : 1998 REV.00 Pg 1 of 12 CONTENTS 1. GENERAL 1.1 Introduction 2 1.2 System Description
More informationDMWD 1 to 4 Zone Water Detection Alarm with Distance to Leak Measurement Installation and Operation Manual
CMR Electrical Ltd Bolton House Five Chimneys Lane Hadlow Down East Sussex TN22 4DX Tel: 01825 733600 DMWD 1 to 4 Zone Water Detection with Distance to Leak Measurement Installation and Operation Manual
More informationAT-G-DETECT Gas Sensing System
AT-G-DETECT Gas Sensing System CHECK / CALIBRATION PROCEDURE* *Technician use only This procedure must be carried out by a suitably qualified technician in accordance with these instructions and the standards
More informationINTELLIGENT FIRE TECHNOLOGY. Twinflex and Multipoint V3. User Guide (TO BE RETAINED BY USER) Issue 3
INTELLIGENT FIRE TECHNOLOGY Twinflex and Multipoint V3 User Guide (TO BE RETAINED BY USER) 26-0340 Issue 3 Rafiki Protection Limited Rafiki policy is one of continual improvement and the right to change
More informationBeacon 800 Gas Monitor Operator s Manual
Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be
More informationSmartLINK Module Ei3000MRF for Mains Powered Multi-Sensor Fire / Smoke / Heat / CO Alarms - Ei3000 Series
SmartLINK Module Ei3000MRF for Mains Powered Multi-Sensor Fire / Smoke / Heat / CO Alarms - Ei3000 Series Instruction Manual Read and retain carefully for as long as the product is being used. It contains
More informationUSER MANUAL Digital CO Monitor Model CO30
USER MANUAL Digital CO Monitor Model CO30 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech Instruments Model CO30. The CO30 is a 3 in one
More informationCO2 & Oxygen Monitor Operating Instructions Model: RAD
CO2 & Oxygen Monitor Operating Instructions Model: RAD-0200-2 Table of Contents 1. Overview 2. Package Contents and Description 3. Strobes (Optional) 4. LCD Display Symbols 5. SEU (Main Sensor Unit) 6.
More informationThe most user friendly Security Alarm System L S Section 1 Overview of System Section 2 Planning your Installation
The most user friendly Contents Section 1 Overview of System 1.1 Kit Contents 1.2 Tools Required 1.3 System Features Security Alarm System L S 4 0 0 Section 2 Planning your Installation 2.1 Location of
More information6100 SINGLE LOOP DIGITAL ADDRESSABLE FIRE ALARM CONTROL PANEL
6100 SINGLE LOOP DIGITAL ADDRESSABLE FIRE ALARM CONTROL PANEL USER MANUAL Protec Fire Detection plc, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0)
More informationSigma XT. Extinguishant Control Panel. User Manual Man-1091 Issue 03 October 2009 (K11031M2, K11031F2)
Sigma XT Extinguishant Control Panel (K11031M2, K11031F2) User Manual Man-1091 Issue 03 October 2009 Fire alarm condition General fire indicator lights and zonal fire indicators flash. Buzzer sounds urgently.
More informationFire detection and alarm system Operating instructions
Fire detection and alarm system Operating instructions Operating instructions Table of Contents User responsibility - - - - - - - - - - - - 4 Daily - - - - - - - - - - - - - - - - - - - 4 Weekly - - -
More informationLC3000 Level Controller Installation and Maintenance Instructions
4025550/12 IM-P402-36 AB Issue 12 LC3000 Level Controller Installation and Maintenance Instructions 1. General safety information LC3000 NORM ALARM TEST 2. General product information 3. Installation 4.
More informationSnifter ATEX22 VERSION. User Manual. Distributor
Snifter ATEX22 VERSION User Manual Distributor Version 1.4 09/09/2009 Table of Contents 1. INTRODUCTION... 3 1.1. Safety... 3 1.2. Product overview... 4 1.3. How does it work?... 4 2. INSTALLATION... 5
More informationKM300 Carbon Monoxide Detection System Installation Manual
GE Security KM300 Carbon Monoxide Detection System Installation Manual P/N 1068922 REV 2.0 16SEP09 Copyright Copyright 2009 GE Security, Inc. All rights reserved. This document may not be copied in whole
More informationWARNING FIRE DOOR CLOSING
PRODUCT INSTRUCTIS Stock Code: 088 FCP - Lite Floats Road, Wythenshawe Manchester, M 9WB Tel: + (0) 9 Fax: + (0) 9 Doc No PI - 88 Iss:.0 WARNING FIRE DOOR CLOSING STOP Contents ) Blank ) Features and
More informationControl Panel User Guide
Fire Detection & Alarm System Control Panel V4.14 Control Panel User Guide (TO BE RETAINED BY USER) 26-0397 Issue 6 Fike s policy is one of continual improvement and the right to change a specification
More informationAT-G-SENSE Gas Sensing System
AT-G-SENSE Gas Sensing System INSTALLATION & OPERATING MANUAL INSTALLATION INSTRUCTIONS Open the AT-G-SENSE by undoing the two screws. Remove metal faceplate, turn over and check connection terminals (maximum
More informationScantronic Leading the way in security
USER GUIDE (16 ZONE) HARDWIRED CONTROL PANEL Scantronic Leading the way in security Contents 1. Introduction 3 The 9800+ (16 zone) System... 3 The Keypad... 3 About This Guide... 5 2. Everyday Operation
More informationGasScanner 1C. Single Channel Monitor. Operator s Manual. MINT-0278-XX Rev A
GasScanner 1C Single Channel Monitor Operator s Manual MINT-0278-XX Rev A Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship,
More informationINSTRUCTION MANUAL & 8200 Monitor. Manual Issue 06
INSTRUCTION MANUAL 8100 & 8200 Monitor Manual Issue 06 ** This page is intentionally left blank ** Table of Contents 1 Introduction... 4 1.1 Manual Conventions... 4 1.2 Version 4 PCBA's... 4 1.3 Product
More informationSMART EVO 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS
SMART EVO 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS CONTENTS 1. INTRODUCTION... 5 2. WARNINGS... 6 3. GENERAL DESCRIPTION... 7 4. INSTALLATION... 8 5. LUMINOUS INDICATORS AND COMMANDS... 9 6. DIP-SWITCH
More informationDigital Marine Exhaust Temperature Alarm
Digital Marine Exhaust Temperature Alarm Model: SM007D/S INTRODUCTION COMPONENTS Marine water cooled exhaust systems are designed to withstand temperatures of up to about 120 C. However the exhaust gases
More informationCarbon Monoxide Transmitter
Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or
More informationTS400. Operating Manual. Test Station for Microtector II Series (G450/G460)
Operating Manual TS400 Test Station for Microtector II Series (G450/G460) GfG GESELLSCHAFT FÜR GERÄTEBAU MBH KLÖNNESTRASSE 99 44143 DORTMUND, Germany TEL. +49 / (0)2 31 / 5 64 00 0 FAX +49 / (0)2 31 /
More informationG-TECTA TM. Portable Gas Detection Instruments
G-TECTA TM Portable Gas Detection Instruments 2 Why Linde gas detection? The Linde Group is a global leader in the supply of gases, we provide expertise and product solutions to industry, commerce, science
More informationFour Channel Gas Detection System AGDS 4. The Home of Control & Automation in Ireland
Four Channel Gas Detection System AGDS 4 The Home of Control & Automation in Ireland Four Channel Gas Detection System The AGDS-4, Gas detection system has been designed for use in boiler houses and plant
More informationInstallation and user manual for the CF5000, MF5000 and FXP5000 range of fire panels
CF5000, MF5000 and FXP5000 Installation and user manual for the CF5000, MF5000 and FXP5000 range of fire panels 16 zone panels Contents PANEL INSTALLATION...3 Installation...3 PANEL WIRING... 3 Mains power
More informationBA374C Intrinsically safe field mounting indicating temperature transmitter
BA374C Intrinsically safe field mounting indicating temperature transmitter Issue: 5 3rd December 2003 CONTENTS 1. Description 2. Operation 2.1 Controls 3. Intrinsic Safety Certification 3.1 ATEX certificate
More informationUSER GUIDE. DRENA 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS - WASTE WATER -
USER GUIDE DRENA 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS - WASTE WATER - II CONTENTS 1. SYMBOLS AND WARNINGS... 5 2. GENERAL INFORMATION... 6 3. WARNINGS... 7 4. GENERAL DESCRIPTION... 8 5. INSTALLATION...
More informationABLOY DA460 SWING DOOR OPERATOR. Installation and commissioning manual
ABLOY DA460 SWING DOOR OPERATOR Installation and commissioning manual APPROVALS / STANDARDS 73/23/EEC (Low Voltage directive) 93/68/EEC (Low Voltage directive) 89/336/EEC (EMC directive) Statement of fire
More informationTBR 40 1 GD Blood Refrigerator Instruction Manual. for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD. (40 x 500ml Blood Bag Capacity)
for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD (40 x 500ml Blood Bag Capacity) REVISION 10/10/07 PLEASE READ CAREFULLY BEFORE OPERATION Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 9
More informationINSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system
INSTRUCTION MANUAL 1MN0110 REV. 0 operates with ISO9001 certified quality system TECSYSTEM S.r.l. 20094 Corsico (MI) Tel.: +39-024581861 Fax: +39-0248600783 http://www.tecsystem.it R. 1.1 25/08/16 ENGLISH
More informationIREX Pellistor exchange IR gas detector
IREX Pellistor exchange IR gas detector Manual addendum: Connecting to a Wormald 2003A-1 gas module M07*** Issue 1 January 2009 INTRODUCTION IREX is an Exd Flameproof gas detector, designed for detecting
More informationOPERATING GUIDE FOR YOUR RP1054D KEYPAD
OPERATING GUIDE FOR YOUR RP1054D KEYPAD Napco 1992 DESIGN PATS. PENDING Technical Manuals Online! - http://www.tech-man.com CONGRATULATIONS!...on your purchase of a NAPCO Magnum Alert security system.
More informationORP (mv) Controller Model: ORP-XP2. UP DOWN ENTER Time & Date Main Menu Alarm Reset
Instruction Manual ORP (mv) Controller Model: ORP-XP2 Power Control Alarm Comms UP DOWN ENTER Time & Date Main Menu Alarm Reset Supplied by: Convergent Water Controls Pty Ltd 2/4 Huntley Street, PO Box
More informationBX444-Mc Gas Detector
Installation and User Guide Rev.01. BX444-Mc Gas Detector Application Duomo is recognised within the gas industry for providing a comprehensive range of low cost, high reliability gas detection for many
More informationInstallation and user manual for the FX range of fire panels. 1, 2, 4 and 8 zone panels
Installation and user manual for the FX range of fire panels 1, 2, 4 and 8 zone panels Contents Panel installation 3 Panel connections 3 Wiring connection drawings 4 Panel facilities 6 Installation check
More informationIssue 4 10/10/11. Part Number: 48160
User Handbook SHIPSURVEYOR USER HANDBOOK Issue 4 10/10/11 Part Number: 48160 GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our customer publications.
More informationTT /97. Features INSTALLATION INSTRUCTIONS. Horn. Lamps
TT-1023 1/97 INSTALLATION INSTRUCTIONS Original Issue Date: 2/94 Model: 6-2000 kw Market: Industrial Subject: Flush- or Surface-Mount Remote Annunciator Kits PA-293991 and PA-293991-SD The remote annunciator
More information8700 S-linkControl Panel
SIDE-POWER Thruster Systems 8700 S-linkControl Panel EN Installation and user s manual v 1.3 Keep this manual onboard! SLEIPNER MOTOR AS P.O. Box 519 N-1612 Fredrikstad Norway Tel: +47 69 30 00 60 Fax:
More information5000 INSTALLATION GUIDE
5000 INSTALLATION GUIDE Table of Contents Section Page No 1. INTRODUCTION... 3 2. WARNINGS AND CAUTIONS... 3 3. UNPACKING... 3 4. INSTALLATION... 4 5. CONTROL PANEL - REMOVING ITEMS BEFORE MOUNTING...
More informationOPERATION & INSTALLATION MANUAL FOR ALARM PANEL M2AP01
OPERATION & INSTALLATION MANUAL FOR ALARM PANEL M2AP01 Table of Contents Safety Instructions 4 Owner/Operator Responsibility 4 Specifications 5 Introduction 6 Installation Instructions 7 Setup 7 Wiring
More informationDec Wirsbo CoSy Radio. Installation and Operating Instructions
Dec. 2002 Wirsbo CoSy Radio Installation and Operating Instructions Contents 1. Overview... 3 2. Room thermostat... 3 2.1 Thermostat models... 3 2.2 Function... 4 2.3 Positioning... 4 2.4 Setting and limiting
More informationGPPS-evo V3.3. Installation Instructions
GPPS-evo V3.3 Installation Instructions The GPPS-evo is a gas pressure proving system with optional CO2 detection. The system will perform a down stream integrity check ensuring gas tightness before use,
More informationTable of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11
Table of Contents 1. OVERVIEW... 3 2. SYSTEM LAYOUT... 4 3. SPECIFICATIONS... 8 3.1 SYSTEM COMPONENTS...9 3.2 PLC INPUTS AND OUTPUTS...9 3.3 FUNCTION KEYS...10 3.4 DEFAULT SET POINTS AND TIMERS...10 4.
More informationAFA 500 FUME HOOD ALARMS
AFA 500 FUME HOOD ALARMS Operating and Instruction Manual 19/7/03 Model AFA 500 Built-in or Remote sensor 2 Relay inputs 1 Relay output Com port Used for alarm indication and monitoring on Fume Cupboards
More informationInterscan Corporation. Instruction Manual Series Compact Portable Analyzer
Interscan Corporation Instruction Manual 4000 Series Compact Portable Analyzer 1 2 INTERSCAN CORPORATION Simplified 4000 SERIES MANUAL TABLE OF CONTENTS SECTION TITLE PAGE 1.0 Equipment Description 4 2.0
More informationNEVER ACTIVATE THE STROBE WITH THE PCB OUT OF ITS PROTECTIVE CASE. EXTREME CARE IS ADVISED TO REMOVE THE RISK OF ELECTRIC SHOCK.
VI64 Version:1.1 Issued 06/12/2007 INTRODUCTION Ventcroft s external sounder/strobes offer an unsurpassed range of models which provide for virtually any installation need. Most models feature a robust
More informationGARDTEC 580. Installation Instructions
GARDTEC 5 0 0 S E R I E S GARDTEC 580 Installation Instructions Contents Introduction 1 Planning the Control Panel Location 2 Installation 2 Cable Entries 3 Mains Supply Connection 4 Terminal Block Connections
More informationHARDWIRED CONTROL PANELS
USER GUIDE 9651 HARDWIRED CONTROL PANELS Contents 1. Introduction...3 The Alarm System...3 The Keypad...3 About This Guide...5 2. Everyday Operation...6 How Do I Know if the System is Working?...6 Setting
More informationFire Control Panel. Installation & Programming Manual TABLE OF CONTENTS
Fire Control Panel Installation & Programming Manual TBLE OF CONTENTS Page Getting Started 2 Power Supply Unit 3 Inputs 4 Outputs 4 LED Indications 4 User Keypad Functions 5 Engineer Keypad Functions 6
More informationPlease read this sheet as it contains important information
Important Please read this sheet as it contains important information Before commencing installation please familiarise yourself to the equipment by reading the comprehensive installation instructions.
More informationRestek s Electronic Leak Detector
Restek s Electronic Leak Detector (cat.# 22839) Instruction Manual 2008 Restek Corporation No part of this manual may be reproduced in any form or by any means without the prior agreement or permission
More informationTS400 & TS410. Intruder Alarm Control Panels. Operating Instructions
TS400 & TS410 Intruder Alarm Control Panels Operating Instructions Zone Location Home Set Chime 1 Armed / Omitted 2 Armed / Omitted 3 Armed / Omitted 4 Armed / Omitted 5 Armed / Omitted F. Exit Always
More informationService Bulletin No. 2930
Service Bulletin No. 2930 MODEL TYPE SECTION/GROUP DATE E / J Series Coaches Service Information 8--Engine Mar. 30, 2006 SUBJECT AMEREX FIRE SUPPRESSION SYSTEM CONDITIONS Service Information Only Description:
More informationGasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08
GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,
More informationOPERATING INSTRUCTIONS FOR LEEC AUTOMATIC CO2 INCUBATORS. MODELS GA2000 and GA2010 MODELS GA3000 and GA3010 With Infrared CO2 detectors CONTENTS
OPERATING INSTRUCTIONS FOR LEEC AUTOMATIC CO2 INCUBATORS MODELS GA2000 and GA2010 MODELS GA3000 and GA3010 With Infrared CO2 detectors CONTENTS 1.0 GENERAL DESCRIPTION 2.0 INSTALLATION AND SETUP 3.0 DESCRIPTION
More informationCO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE
CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE GasAlertMicro 5 IR from Refer to for more details. Manual provided on CD with unit at time of purchase 2 Table of Contents Getting Started.
More informationPA ma Loop Powered Level Preamplifier
4026150/2 IM-P402-139 AB Issue 2 PA420 4-20 ma Loop Powered Level Preamplifier Installation and Maintenance Instructions Safety information General product information 3. Installation 4. Wiring 5. Commissioning
More informationDWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Series 670 Hood Monitor. Bulletin AV-670
Bulletin AV-670 Series 670 Hood Monitor Specifications - Installation and Operating Instructions DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Phone: 219/879-8000 Fax: 219/872-9057
More informationBeacon 110 Gas Monitor Operator s Manual
Beacon 110 Gas Monitor Operator s Manual Part Number: 71-0110RK Revision: H Released: 12/5/17 RKI Instruments, Inc. www.rkiinstruments.com Product Warranty RKI Instruments, Inc., warrants gas alarm equipment
More informationConsultant Specification AGILE Wireless Fire Detection System
Contents Scope of Work... 2 The EN54 Part 2 & 4 Fire System... 2 Standards... 2 Wireless Intelligent Photoelectric Smoke Detector Specification... 4 Wireless Intelligent Thermal Heat Detector Specification...
More informationQUICK USER GUIDE. Alarm Control Panel
Alarm Control Panel Firmware version 2.10 QUICK USER GUIDE The full version of User Manual and other manuals are available on CD included in the control panel delivery set or on the website www.satel.eu
More informationZX1e ZX2e ZX5e. Document No Issue 01 user manual
ZX1e ZX2e ZX5e Document No. 996-130 Issue 01 user manual MORLEY-IAS ZX2E/ZX5E Fire Alarm Control Panels Table of Contents 1 INTRODUCTION... 4 1.1 NOTICE... 4 1.2 WARNINGS AND CAUTIONS... 4 1.3 NATIONAL
More informationWD-CS. Water leak detection cable. Technical Overview. Features. Page 1. Data Sheet Ref: Issue: 5.1.
Data Sheet Ref: 90501085 WD-CS Water leak detection cable Technical Overview The WD-CS range of cable sensors are typically used to detect water leaks usually under floors, and are used in conjunction
More informationSalamander User Guide
SYSTEM USER GUIDE Introduction Salamander User Guide Salamander is a radio-controlled, battery-powered system for holding open fire doors. Before installation, a thorough radio survey will have been conducted
More informationMICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual
MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual Read instructions carefully before operating this device. ❶ This device is not to be used for Human Life Support applications. ❷ To avoid possible
More informationUser Operation Guide Document: VI49.2
www.ventcroft.co.uk Disabled Persons Response Alarm. User Operation Guide Document: VI49.2 Protecting People Ventcroft Ltd Meditell-2 Disabled Person Response (WC) User Guide First Printed: 19/07/05 Document:
More informationInstallation Instructions. The SEC-L is a gas pressure proving system with water and electric isolation, gas detection and ventilation interlock.
V3.3 Installation Instructions The SEC-L is a gas pressure proving system with water and electric isolation, gas detection and ventilation interlock. The system will perform a down stream integrity check
More informationSECURIT 700L USER MANUAL
SECURIT 700L USER MANUAL C & K Systems Ltd 6/12/95 Securit 700L User instructions This manual will give you all the information you need to operate and control the Securit 700L Control Panel. This manual
More informationSigma. K1000 Series 1, 2, 4 & 6 Zone Fire Control Panels. Operation and Maintenance Manual. Man-1048 Issue 05 October 2009
Sigma K1000 Series 1, 2, 4 & 6 Zone Fire Control Panels Operation and Maintenance Manual Man-1048 Issue 05 October 2009 CONTENTS Contents... Page Safety & Installation...2 Installation - continued...3
More informationU ser's Guide PC6010
User's Guide PC6010 Quick Reference Guide This manual is for Basic and Advanced users. Each of these types of user can access a different set of functions. The and symbols next to the title of each procedure
More information