LABORATORY ILLUMINATED FLOTATION BATH

Similar documents
WITH OMRON E5CC CONTROL

REFRIGERATED RECIRCULATORY COOLER

OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN

OPERATING INSTRUCTIONS FOR:

LABORATORY OVENS MODELS INCLUDE: TO-152G & TOH-152G TO-152F & TO-235F DEHYDRATING OVENS MODELS INCLUDE: TD-150F TD-330F TD-500F

for LABORATORY REFRIGERATOR MODEL TELR-145/ L/200L CAPACITY REVISION

for LABORATORY STORAGE FREEZERS MODELS TUF GD/SD, TUF GD/SD, & TUF GD/SD 520, 740, & 950 Litre CAPACITY REVISION

for PHARMACY REFRIGERATOR MODEL TPR SD/GD TPR-750/950 2 SD/GD TPR SD/GD 520, 750, 950, 1150 Ltr CAPACITY REVISION

PHARMACY REFRIGERATOR INSTRUCTION MANUAL

for CLIMATRON GROWTH CABINET MODEL CLIMATRON 520 & 1100-DL/SL-H-CO 2 With Humidity, CO 2 & Touch screen Control REVISION

Mini. Level Control systems For Manual Sump Pumps up to 750W 1 or 2 Pump Systems

Table of Contents. Features Quality Policy Section A - Unit Operation Section B Unit Maintenance Trouble Shooting Guide...

Table of Contents. Forward Quality Policy Design Parameters Section A Unit Operation Section B Unit Maintenance...

PRODUCT MANUAL. TT-MTM2 Hard-Wired Controller with Remote

R A D I A N T O U T D O O R H E A T E R S OPERATION, INSTALLATION AND MAINTENANCE MANUAL. TT-MTM Wall Controller

ETLW7. Product: 7kg Top Load Washing Machine. User Manual

Table of Contents. Forward Quality Policy Design Parameters Section A Unit Operation Setting the Timer...

Drying Cabinet With Timer

OWNER S MANUAL. Convair Tower Heater CTH09. (English) (CTH09)

Installation and Operations Manual

RSA2CL6SS model pictured

PRODUCT MANUAL. The intelligent, energy saving solution

1.7L KETTLE INSTRUCTION MANUAL 220~240V 2200W

Issue 1 ROBINHOOD ISLAND CANOPY 90CM. Product may vary slightly from the picture

Classic Heater Pool Controller

Model pictured: ST6001

Operation Manual Portable Room Air Conditioner Models: RPC26WA / RPC35WA / RPC41WA

Product & Assembly Manual

TBR 40 1 GD Blood Refrigerator Instruction Manual. for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD. (40 x 500ml Blood Bag Capacity)

Operator s Manual. Histology Bath

OPERATION, INSTALLATION AND MAINTENANCE MANUAL. THX Flush Mount Enclosure. Take heating to the Max!

Operation, Maintenance and Installation Manual

MariGlo LED. Inspiring Excellence in Fish Keeping! 45 / 60 / 90 / 120.

OWNER S MANUAL. CoolTower CTF07. (English) (CTF07)

General Care and Safety Guide

Use and Care Manual. Flammable Material Storage Refrigerator LRBFS06W1HC

PRODUCT MANUAL OPERATION, INSTALLATION AND MAINTENANCE MANUAL. Indoor Convective Panel Heater

MD-MINI Dry Bath. Instruction Manual

OPERATING INSTRUCTIONS FOR LEEC AUTOMATIC CO2 INCUBATORS. MODELS GA2000 and GA2010 MODELS GA3000 and GA3010 With Infrared CO2 detectors CONTENTS

EUF92W. Product: Bar Freezer. User Manual

Model pictured: RCB2AH6WH/SS

Operators Manual. Microprocessor Controlled. Water Baths

SFH136-CN TRUSTED FOR OVER 100 YEARS. Sunbeam Heater. Instruction Leaflet PLEASE READ AND SAVE THESE IMPORTANT INSTRUCTIONS

Series: MBC1-TC Mini Benchtop Temperature Controller

IXL Eco Ventflo. User Guide. Model: (200mm) - Extraction Rate: 340m 3 /h Model: (250mm) - Extraction Rate: 490m 3 /h

INSTRUCTION MANUAL PORTABLE FRIDGE/FREEZER. Models: RVX-35 / RVX-45 / RVX-50 RVX-55 / RVX-65 / RVX-75

Digital Heat Block User Manual

USER MANUAL PLEASE READ AND UNDERSTAND THIS MANUAL COMPLETELY BEFORE USE.

6.5L SLOW COOKER INSTRUCTION MANUAL

Instruction Manual. Please read these instructions and the Safety section carefully before using your refrigerator.

HORIZONTAL LAMINAR FLOW CABINET

Zip Hydroboil. Installation and Operating Instructions. Filtered instant boiling water

Corrosion Resistant Fluid and Air Handling Systems.

BX Series. Installation, Operation, Maintenance Instructions & Spare Parts

Pasto Sous vide water bath

Instruction Manual. By Evakool. Please read these instructions and the Safety section carefully before using your refrigerator.

ELECTRIC RADIANT OUTDOOR / INDOOR HEATER

FAN HEATER FH118 User Manual

Convection Panel Heater

THERMOFILM CONVECTIVE PANEL HEATER INSTALLATION, OPERATION AND MAINTENANCE MANUAL. Models: CP 2400 CP 2000 CP 1500 CP 1000 TABLE OF CONTENTS

USER MANUAL RANGE HOOD CG920FGS. User manual_cg920fgs 28/11/2012

Installation and Operating Instructions for Microlene Underbench System

INSTRUCTION MANUAL / INSTALLATION MANUAL. 7kg Vented Dryer MODEL: KA7VENTDRYA

PoolSweepa FloorCova Pool Cleaner. Operating Instructions

Operator s Manual. The Quietek

GG-EM ENTRANCE MONITOR. Installation and Operation Manual

Model pictured: ST7001SLIM/ST7001SLIMA

PERSONAL CUEING DEVICE MODEL NO. BDAAU100 INSTRUCTION MANUAL.

OWNER S MANUAL Compact Portable Refrigerated Air Conditioner CP35CW1

PRIMUS IMPORTANT. User Manual and Product Specifications. Portable Fridge/Freezer. Part No. PRI18LP - 18 Litre. Read instructions before operation.

Reacti-Therm I and III Heating Modules

Model pictured: ST3701

Plant Growth Cabinets

Classic Heater Spa Pool Controller

TRAVELMATE. Instruction Manual

Outdoor Refrigerator USER S MANUAL

DUAL VOLTAGE REFRIGERATORS 220/240 VOLTS AC AND 12/24 VOLTS DC INSTALLATION AND OWNER S MANUAL

INSTRUCTION MANUAL TS21. 2 Heating and 1 Cooling

MaxLite LED MICRO-T PANEL

PoolSweepa Wallclima / Optima Pool Cleaners. Operating Instructions

GG-NH3-2% AMMONIA GAS SENSOR. Installation and Operation Manual

OPERATION & MAINTENANCE INSTRUCTIONS FOR. 230/240 VAC, 1ø, 50Hz

INSTRUCTION MANUAL P Heating and 1 Cooling

WARRANTY BOOKLET. EHPA Electric Heat Pump Water Heaters. Register online at

Zip Econoboil. Installation and Operating Instructions. Budget priced instant boiling water for canteens and kitchens.

Camping Toilet USER MANUAL PLEASE READ AND UNDERSTAND THIS MANUAL COMPLETELY BEFORE OPERATING & USING THE CAMPING TOILET

PoolSweepa FloorCova Pool Cleaner. Operating Instructions

USER MANUAL. Adventure Kings 50L Refrigerator

Professional Warming Drawer and Warming Drawer Cabinet

OWNERS MANUAL MODEL 451 (-101 & -201)

MaxLite LED VAPOR TIGHT LINEAR FIXTURES

Operating instructions Page 12

Dome Kettle. polished stainless steel body BTDK120

ICS PACIFIC. Installation Instructions & User Manual. G135L Medical Refrigerator

VX SERIES Wireless Thermostat with Occupancy Sensor

CD6102 / CD6102 OC / CD6102-2

8 LITRE URN. Instruction Booklet. Please read these instructions carefully and retain for future reference

USER MANUAL LEEC CLASSIC INCUBATOR. C100, C157 and C330

WARRANTY BOOKLET. Commercial Hot Water. Register online at

RHPK3000_IB_FA_ SALT & PEPPER MILLS 2 YEAR WARRANTY RHPK3000

Transcription:

LABORATORY ILLUMINATED FLOTATION BATH MODELS INCLUDE: TFB-1 Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 10-12 Ross Place Wetherill Park NSW 2164. P.O. Box 6862 Wetherill Park DC NSW 1851 Phone: (02) 9604 3911 International: 61 2 9604 3911 Fax: (02) 9725 1706 International: 61 2 9725 1706 Email: thermoline@thermoline.com.au Web: www.thermoline.com.au

2 IMPORTANT INFORMATION This manual contains operating and safety information. The operator must read and understand the contents of this manual prior to using this equipment. Incorrect operation may cause harm or damage the severity is classified by the following Alert Boxes. Ensure that this manual is saved for future reference: Warning alerts apply when there is a Possibility of personal injury. Caution alerts apply when there is a Possibility of damage to the Equipment. To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions. CONTENTS SAFETY INFORMATION 2 ALERT BOXES 2 INTRODUCTION 3 THE MANUAL 3 UNPACKING 3 LOCATION 3 INTENDED USE 3 & 4 FUNCTION 4 PRINCIPLE OF OPERATION 4 PERFORMANCE 4 C-TICK 4 INSTALLATION 4 MAINTENANCE 5 TECHNICAL ASSISTANCE 5 EXPLANATION OF CONTROLS 5-7 OPERATING THE EQUIPMENT 7-8 SENSOR CORRECTION 9 WARRANTY POLICY 10

3 INTRODUCTION Thank you for selecting this product from the large range of products manufactured in Australia by Thermoline Scientific. The materials incorporated in this design have been carefully selected for their durability, technical performance, and practicability, the resultant product is one we are proud of. THE MANUAL Please familiarise yourself with the information contained in this manual before using the equipment. Ensure that the manual is kept safe but accessible for current and future operators to refer to. UNPACKING Remove the equipment from the packing material, check that it is complete. Assemble the components and inspect for damage. Retain the packaging materials until the equipment has been thoroughly tested. Notify the detail of any damage without delay to your supplier, or Thermoline Scientific. LOCATION Place the equipment in a well-ventilated area on a firm surface. Confirm that the power supply is suitable for this equipment. Refer to the equipment rating plate for details. To control temperature effectively, the equipment should be located in a stable ambient. WARNING: Ensure that the electrical outlet is earthed and is of the correct voltage and is capable of carrying the stated current load. INTENDED USE This equipment is used to provide a convenient method of melting embedded media (wax) from tissue samples. CAUTION: The range of equipment described here is not to be used in conjunction with large quantities of water. The introduction of water into the glass dish recess could result in premature failure of components.

4 WARNING: Do not use flammable solvents or combustible materials as fire or explosion may occur. This equipment uses components which may be a source of ignition. FUNCTION To provide an accurate controlled water temperature of ambient +5ºC to +60ºC. PRINCIPLE OF OPERATION The illuminated flotation bath is designed to be light weight and portable. It is ideally suited for use in the Histology laboratory. The heater body and surrounding area is insulated to minimise heat loss. The desired operating temperature is set at the temperature controller. The temperature sensor is located within a spring clip and is designed to rest against the glass insert (dish). The signal from the temperature sensor is monitored by the temperature controller. The resultant calculated pulsed output is directed to a solid state relay that switches power to the heating element located below the base of the well. The warmed base of the well gently and uniformly heats the water within the glass dish. The amount of energy is determined by the temperature sensor (spring loaded against the glass dish). Energy movement is by convection. This limited effect results in an even temperature gradient throughout the bath. The low wattage fluorescent lamp is activated by the press of a button. PERFORMANCE The operating range for this equipment is ambient +5ºC to 60 C with a control accuracy of +/- 0.5ºC MAXIMUM TEMPERATURE For this product is 60 C. C-TICK The equipment referred to here complies with requirement of AS/NZS 1044:1995 electromagnetic interference (EMI). INSTALLATION (Refer to Unpacking Page 3) Place the equipment in a well-ventilated area, on a firm level surface and ensure that a suitable power supply is available. WARNING: Ensure that the electrical outlet is earthed and is of the correct voltage and is capable of carrying the equipment stated current load (AMPS).

5 The model TFB is suitable for bench top use only. MAINTENANCE The Thermoline Scientific Flotation bath requires no routine maintenance other than normal levels of cleanliness. The glass insert may be removed for cleaning. The external case and the stainless steel surfaces may be wiped clean using a damp, soft cloth. Note: Stainless steel is under most conditions extremely resistant to corrosion. This is in part due to the addition of chromium and nickel to the steel and the formation of a durable chromium oxide at the surface during the manufacturing process. There are several chemicals which will attack the surface of stainless steel. Always have in place a program of regular cleaning using a soft damp cloth with a mild solutions of soap and water and allow to dry. TECHNICAL ASSISTANCE If you require additional operational or technical information regarding this equipment please contact: Thermoline Scientific Customer Service Division Telephone: +61 2 9604 3911 Facsimile: +61 2 9725 1706 Email: thermoline@thermoline.com.au EXPLANATION OF CONTROLS The controls are located in a control enclosure to the right hand side of the bath. MAIN SWITCH: This is a combination power switch and circuit breaker, when pressed this switch allows power to be supplied to the controls. Incorporated in the face of the switch is a MAINS ON light. CAUTION: The equipment is fitted with a combination circuit breaker/switch. If the switch continues to trip, it may indicate an electrical fault exists. Switch off the power on consult a qualified technician. TEMPERATURE CONTROL/SENSOR: This is an Omron E5CC digital control scaled a 0.1ºC increments and has been configured to provide stable control with minimum of overshoot and quick recovery time. Sensor type is a RTD Pt100. Maximum recommended temperature is 55ºC.

6 Page button: Used to view calibration offset parameter. Scroll button: Not used by the operator. Side arrow: Used to move the cursor when changing temperature. Down arrow: Used to decrease parameter setting. Up arrow: Used to increase parameter setting. LIGHT SWITCH: This is a push button, press to turn light on, press again to turn light off. When activated, the fluorescent light will be illuminated.

7 SAFETY THERMOSTAT: Located at the front of the Floatation bath is a Red Cap. This is the reset button for the Safety Thermostat. This will trip if the Heated Well exceeds 80.0degC. This will trip if a failure of the Heater occurs, or if the operator removes the Glass tray for an extended period of time (Approximately 30minutes) without turning the Floatation bath off. To reset simply unscrew the Red cap and press the Red Button. The Thermostat will only reset if the temperature of the Heated Well has dropped below 60.0degC. An indication that this Safety thermostat has tripped is that the controller will turn off but the Mains switch is still illuminated. ELECTRICAL - POWER LEAD (at rear) A removable mains power lead with a three pin plug and right angle female IEC socket is supplied for connection between the male IEC socket located at the side of the bath, and a ten (10 Amp general Purpose Outlet) (G.P.O.) OPERATION WARNING: 1. To avoid injury do not use flammable solvents in this bath, as fire or explosion may result. This bath contains components that may ignite such materials. WARNING: 2. Do not introduce water into the well (it is not a bath) as it may result in component failure. OPERATION: 1. Locate the Bath as previously described in Location, plug the mains lead into the IEC socket at the back of the case and into the power supply. Turn on the power point. CAUTION: Ensure that the ventilation for the rear cooling fan are not obstructed. 2. Place the glass tray into the Heater Well and carefully fill with water. Place the Stainless Steel lid on the Glass tray. Warning: Contact with hot fluid of the bath may result in personal injury. Ensure appropriate protective equipment and procedures are applied at all times. 3. Turn on the main switch on the control panel. The switch is illuminated so you will know when it is on. The temperature will go through a short warm up period where all segments of the display will be on, before indicating the set temperature (SV) on the lower display and incubator temperature (PV) on the top display.

8 4. The Omron E5CC is a microprocessor based instrument that has been factory configured for range, sensor type, and engineering parameters for optimum control. Limited access to the control parameters is available. The operator can alter the temperature set point and has access to a parameter used for calibration purposes. 5. TO CHANGE THE SET TEMPERATURE: Use the button to move the cursor, the digit will flash indicating that it can be changed. Use the or arrows to change the temperature. When the desired temperature is set leave for a few seconds and the digits will stop flashing to confirm entry. Operating above 55.0degC: When operating at temperatures above 55.0degC it important to Stage the Temperature increase. It is advised to first let the Floatation bath operate at 55.0degC for at least 30minutes before changing to the required value above 55.0degC. This will limit nuisance tripping of the Over Temperature safety Thermostat. 6. OVER TEMP ALARM: The temperature control has been configured with automatic over temperature alarm. In case of a malfunction the alarm will turn off all forms of heating when the bath temperature is more than 2.0ºC above the set temperature. In this instance the controller will display ALM1.

9 SENSOR CORRECTION: There are a number of factors that will affect the accuracy of the temperature displayed in relation to the actual temperature inside the Glass Tray, these include the following: Ambient Temperature Level of water in the Glass tray Location of the sensor in the Glass Tray Whether the Stainless Steel Lid is on or off. Because of the above factors there may be an error between the temperature displayed and the temperature measured at the middle of the bath. For this reason the Omron temperature control has a parameter that can be used to correct the temperature displayed. This sensor correction parameter is displayed as. When checking calibration error in the middle of the Bath it is important that the water level in the Glass Tray is at least 75%full and that the Stainless Steel lid is on the tray. If the Stainless Steel lid is not in place water evaporation occurs which has a cooling effect, this cooling effect can show errors of up to 15.0degC In simple terms this parameter adds or subtracts a correction value to the displayed temperature to make it read the correct temperature. The calibration parameter can be accessed as follows: Prior to any calibration of the temperature display ensure that the equipment used to measure the incubator temperature has a current certificate of calibration to a traceable source, and that you are confident that it is accurate. Press the page button to access the sensor correction parameter. Use the up or down arrow to enter the sensor correction value. Allow the digit to stop flashing and the screen will display the adjusted value. In the example above the incubator temperature displayed (PV) is 37ºC. If a certified reference instrument measured 36ºC we would need to correct the temperature by -1.0ºC (note that the sensor correction parameter is accurate to 1 decimal place).

10 Standard 12 month Warranty Thermoline Scientific Equipment Pty Ltd ABN 80 000 859 129 ( Thermoline ) Thermoline warrants to the original purchaser that this product will perform to its product specification for a period of 12 months from date of purchase, provided that the installation of the product has been carried out in accordance with the latest version of the manufacturer's instructions and further provided that the use of the product complies with that specified in the relevant specification. Thermoline is not responsible for any loss or damage arising from incorrect usage, usage outside the suitability of the product as stipulated in the manufacturer's instruction, damage caused by accident, fire, flood, act of God or failure to properly install, operate or maintain the goods in accordance with the printed instructions provided. The following statement applies only to product sales that fall within the definition of a Consumer Sale set out in the Australian Consumer Law contained within the Competition and Consumer Act (Cth) 2012: Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure. Notwithstanding the preceding clause and to the extent permissible by law, the liability of Thermoline is limited, in relation to the warranted product and at the option of Thermoline to: replacing the product or the supply of equivalent product; the repair of the product; the payment of the cost of replacing the product or of acquiring equivalent product; or the payment of the cost of having the product repaired. To the extent permitted by law, all other warranties whether implied or otherwise, not set out in this Warranty are excluded and Thermoline is not liable in contract, tort (including, without limitation, negligence or breach of statutory duty) or otherwise to compensate the Purchaser for: any increased costs or expenses; calibration/certification services; any loss of profit, revenue, business, contracts or anticipated savings; any loss or expense resulting from a claim by a third party; or any special, indirect or consequential loss or damage of any nature whatsoever caused by Thermoline s failure in complying with its obligations or the purchaser s failure due to accident damage, impact, misuse or negligence. The benefits given to the purchaser in this Warranty are in addition to other rights and remedies under a law in relation to the products or services to which this warranty applies. This warranty applies only to products purchased and installed in Australia and does not cover any consumable items e.g. filters, light globes, ultrasonic nebulizers. The warranty does not extend to labour and freight costs where the warranted product is located outside Australia. To make a warranty claim, contact Thermoline on 02 9604 3911 or service@thermoline.com.au.