Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology

Similar documents
Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

An Automated SPE Procedure and Analysis for the Extraction of Chlorpromazine and Thioridazine from Plasma Application Note 213

How Important is Performance to your SPE Method Development?

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica.

Fundamentals of Solid-Phase Extraction (SPE): Principles and Practical Tips for Developing Sample Preparation Methods

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Supra-Clean and Supra-Poly SPE Solutions. Solid Phase Extraction Columns

Sample Preparation Solid Phase Extraction Disks Cartridges. Horizon Technology, Inc. Salem, NH USA

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

SEC-MALS Noise Assessment and Cleaning Guide

GB National food safety standard Determination of benzoic and sorbic acid in milk and milk products

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

Standard Operating Procedure for Oil & Grease Method EPA1664A

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Spark Holland introduces

300 series HPLC autosamplers

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Chemistry Instrumental Analysis Lecture 29. Chem 4631

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Direct Injection and Online SPE LC/MS/MS for the Determination of Pharmaceuticals and Personal Care Products (PPCPs) in Water Samples

RapidTrace + Automated SPE Workstation. Instruments. Accessories Racks Vial Adapters Test Tubes O-Rings Miscellaneous

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Thermo Scientific Dionex SolEx SPE Cartridges

On Line SPE LC/MS/MS Applications

Order all the parts and accessories needed at one time Reduced learning curve for rapid implementation

Reliable elimination of particles and micro organisms PAGE Protein Precipitation Fast and effective sample clean up of biological samples

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

Application Note. Authors. Abstract

Biotage VacMaster 10 & 20 Sample Processing Station. Operation Manual

Tekmar-DOHRMANN The 3100 Sample Concentrator From ManualLib.com

APPLICATION NOTEBOOK FOR AFFINIMIP SPE

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA

An Automated Method for the Selective Solid Phase Extraction of Ochratoxin A from Wheat Using Molecularly Imprinted Polymers

until filters are dry. Once dry, place the filters shiny side up in petrislides. The petrislide do NOT need to be acid washed.

ProStar 325 UV-Vis Detector

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

APPLICATION NOTE DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. APPLICATION NOTE As of July 2017, Version: 1.

XENOSEP. SPE Systems for EPA Method 1664A. The Easiest, Most Cost Effective Way to Increase Productivity

Agilent 1290 Infinity II Vialsampler Data Sheet

detectors UltiMate 3000 Variable Wavelength Detectors

Without the Dry Down. Successful Bioanalytical Sample Preparation. Microelution 96-Well SPE Plates

October 1, Dear Valued Customer:

Advancing the power of SPE. Intelligent Automated Solid Phase Extraction. SmartPrep Brochure

Pretreatments to LLE

Use of Automation to Achieve

Product Guide for Ludger Clean Post-endoglycosidase and Post-exoglycosidase Clean-up Plate

CENTRIFAN PE. Evaporation the Easy, Environmental Way. Reliable Results for Small Samples. Small Volume Evaporator GREEN.

Methods of solid phase extraction. Methods of solid phase extraction

FREESTYLE always ready

RESOLUTION OIV-OENO MONOGRAPH ON TANNINS UPDATE OF THE METHOD FOR DETERMINATION OF POLYPHENOLS

Energy Conservation with PARAG Energy Efficient Axial Flow FRP Fans

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

Advancing the power of SPE. Intelligent Automated Solid Phase Extraction. SmartPrep Brochure

JSB is an authorised partner of DETERMINATION OF THE HYDROCARBON-INDEX DIN EN ISO :2000. i n s p i r a t i o n m e e t s i n n o v a t i o n!

Lab 12E, 12F, 2E: Acid Rain and Seeds

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

Tubing spares and piping accessories. Gas purification and control 1 Moisture Traps,Carbon traps 2 Gas Purification Panels 3 Gas Control Box

THE PRESSURIZED CONDENSATE RECOVERY SYSTEM:

DETERMINATION OF PAH In Particulate Matter PM 10 with SPE/EVAporation

Purge and Trap Troubleshooting Made Easy?

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

The Complete Guide to Gas Sample Drying for Your Analyzer or Process

VENDED. Speed Queen Stack Tumble Dryers. 30 lb Tumble Dryer 45 lb Tumble Dryer

Development Centre for Evaporation and Thermal Separation Technologies

The Road to Recovery. Mapping Advanced Sample Preparation Techniques for GC, GC-MS and HPLC Analysis

SUMMARY AND CONCLUSION

Improved HPLC Separations at Elevated Temperatures Using a New Stand-alone Preheater

PURITY GRADES OF LAB REAGENTS

Thomas J.M. Weaver PresidentlCEO PROSYS Corporation

Evaporators, Sealers and other Equipment

From Sample to Vial. Automated Sample Prep Solutions. Extraction Cleanup Concentration

Soil characteristics that influence nitrogen and water management

TRACE 700. Slide 1. Hello, my name is Nick Cavitt and I m a marketing engineer in the C.D.S. group at TRANE.

Aqueous Cleaning Design Considerations --us-

Gilson 151/152 UV/VIS Detectors User s Guide

CO2 Incubators The Complete Line

KOCH Curtain Coaters

APPLICATION NOTEBOOK FOR SOLID PHASE EXTRACTION

[ ACQUITY QDa DETECTOR ] SEPARATING BEYOND QUESTION BRINGING THE POWER OF MASS DETECTION TO YOU

Active Dry Air Venting for Water Contamination Control

Choosing the best vacuum concentrator for your laboratory. Tips and best practices from industry pioneers

API 2000 LC/MS/MS TurboIonSpray Ion Source Manual

TECHNICAL DATA 3% AR-AFFF

HPLC Column Ovens and Temperature Controllers

vapic was the first company in the industry to use screw vacuum pumps. We have gained extensive

Soil and Plant Methods List for Automated Ion Analyzers. Flow Injection Analysis

Polaris PHEs vs. Shell-and-Tube Exchangers

SYM BIO INDUSTRIES SDN BHD INTRODUCTION OF SYM BIOGREEN NATURAL PLANT BIO ENHANCER

Storage Tanks and Accessories

Evaporation Technology

FREESTYLE QUATTRO ACCECLEAN ELUVAC GPC COLUMNS IMMUNOAFFINITY COLUMNS SAMPLE SAMPLE PREPARATION PRODUCTS PREP

Transcription:

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology Joan M. Stevens, Ph.D., and Mark Crawford Gilson, Inc., 3000 Parmenter Street., Middleton, WI 53562 www.gilson.com

Introduction: Solid Phase Extraction has become a technique of choice for sample cleanup and trace enrichment Some of the benefits of SPE: Versatility, Selectivity, Speed, and Low Solvent usage There are other methods for sample cleanup: Liquid-Liquid Extraction, Protein Precipitation, Turbulent Flow, and Dialysis Most automated methods find their roots in a manual method Most manual methods don t optimize the extraction mechanism due to the so called time believed to be involved in method development Automation in its most general term removes the repetitive/monotonous nature associated with method development and the drain on human capital 2

Optimization Condition Solvent Volume Load Volume Break-through Wash Volume Wash solvent selection Break-through Elute Volume Elution solvent selection Percent recovery SPE Process Conditioning Loading Sample Washing Eluting Analyte Interferences 3

Let s Face Facts: Time is a limiting factor Although we would appreciate all the steps associated with SPE to be optimized work load doesn t allow us this luxury However if an automated system could accomplish all the variations of solutions, additions and elutions then we could optimize our SPE methods without manual intervention and still reap the benefits 4

Just the Basic Facts: A typical SPE method with disposable cartridges or 96 well plates consists of 4 basic steps: CLWE C = Condition - Let s think about this - What is a SPE cartridge? - Generally speaking its: A dry sorbent/material held in place with a frit on the bottom and one on the top 5

Conditioning: Wets/Activates the packing bed by solvation Very important step often ignored or minimized Major step DO NOT LET THE CARTRIDGE DRY OUT Methanol is usually the initial conditioning step Hydrates the particles, remember the packing material isn t as efficient as the particles found in an HPLC column Need to rely on the basics of science to illicit the required effect 6

Now What? Sample what sample: What do you know about the sample Serum Plasma Urine Why is Serum and Plasma difficult to work with? Because of protein binding A particular drug can be bound to proteins in the biological fluid that will inhibit its interaction with the SPE cartridge, usually greater than 50% bound 7

Interrupting Protein Binding: Prior to loading the sample onto the SPE cartridge protein interactions with the drug need to be interrupted, if not the results will be compromised This is accomplished with the addition of acidic solution or basic solution Optimization of common acidic additions is shown below: Optimum Acid Chlorpromazine Thioridazine Conc. Rec. Rec. CV Acid (%) (%) CV (%) (%) (%) None 0 88 3.0 71 4.0 TFA 0.0033 96 2.0 77 2.3 TFA 0.01 96 1.8 81 2.3 TFA 0.03 92 1.1 79 1.7 Acetic 0.0033 88 0.9 70 1.4 Acetic 0.01 94 2.8 78 3.2 Acetic 0.03 91 2.4 73 2.8 Acetic 0.05 95 1.7 86 1.6 H 3 PO 4 0.0033 89 1.6 74 3.2 H 3 PO 4 0.01 89 1.9 73 1.6 H 3 PO 4 0.03 98 1.7 83 1.8 H 3 PO 4 0.05 90 2.3 76 1.8 H 3 PO 4 0.1 97 1.7 83 1.8 8

Plasma Acid Addition Recoveries: 90.0 Thioridazine Acid Addition Recoveries in Plasma 85.0 Recovery Maximums Recovery (%) 80.0 75.0 70.0 TFA Acetic Acid Phosphoric Acid 65.0 0 0.02 0.04 0.06 0.08 0.1 0.12 Conc (%) Graph 1: As the acids get weaker, the recovery maximums increase, indicating that a weaker acid is more efficient to control plasma matrix effects. 9

Loading the Sample: Load the sample in two steps is possible why? Take a look at the process The media is activated---- expose the media to the sample The longer the sample can interact with the media pushes the attraction of sample with media in the forward direction If the sample can be added in two shots the first interaction with the sample allows the media to relax and therefore be more accepting of the remaining sample Under the same concept the sample must be added slowly to the SPE cartridge usually 1 ml/min ~20 drops /minute Channeling is an issue in SPE cartridges a direct result of introducing the sample much too quickly to the media The sample must become intimate with the material channeling SPE media 10

Relationship Between Sample Recovery and Flow Rate in SPE 11

How Fragile is Fragile: Now what do I do? The sample in now associated with the cartridge material... The sample isn t going anywhere at this point a nice not too aggressive push of positive pressure or vacuum will pull the liquid through the cartridge You will NOT dry the cartridge out, how long does it take to remove a aqueous solution in an evaporator? It will require several minutes of constant gas flow or vacuum to dry the cartridge out 12

Washing the Cartridge: Throwing the baby out with the bath water Optimization of this step is crucial 13

Washing- Before Elution These two steps are not to associated with each other however they are What are my concerns: Adequate Recovery: Will my recovery be adequate????? Reproducibility: Between labs within my lab Confidence Level: Does the assay transfer.how? We need to test what is being eluted As much as inferents always compromise the purity of the effluent there is a fine line between washing the cartridge enough to remove interferents without also decreasing the recovery of the analyte of interest How do we do it..possibly automate the array to find the optimum set conditions Rule of thumb if it elutes with 40% wash use 5/10% of two washes slowly Test the wash solutions for breakthrough of the analyte if mass spec is available choose it as your detection mode of 14

Compound Sorbant Interaction Load Wash Flow = Impurity = Compound Diagram 1: The compound-sorbant interaction is what makes solid phase extraction work. The solid phase interacts with the sample based on chemical structure (polarity, functionalities, size). Reverse phase methodology allows the sample to grip to the sorbant due to hydrophobicity. The compounds therefore is attracted to the sorbant until washed away by an organic solvent. If the solvent is strong enough the sorbant lets go of the compound and elutes. The trick is to be able to wash the impurities that are soluble water and low organic solvents without eluting the compounds of interest. in 15

Optimizing the Wash and Elution Step: Automated There are numerous SPE methods available, the web is a wonderful tool for perusing SPE manufactures sites for methods; textbooks, published references, applications bibliographies Choose a method that extracts a drug/compound with some or all of the same functionality associated with your compound Amines, Carboxylic acids, Hydrophobicity Although the method may not be exact this is a wonderful starting point Automating the method for optimization is now accessible 16

Automated Optimization: Simple Liquid Handler and SPE racks The SPE rack allow for multiple elutions or fractionation with various solvents to optimize each and every step of an SPE method without manual intervention 17

Example Method: Human plasma containing Chlorpromazine and Thioridazine at 0.5 mg/ml concentration was acidified with acetic acid 0.05% to interfere with protein binding Basic procedure: Condition:1 ml MeOH, 1 ml Water Load: 1 ml of sample (0.5 mls x 2) Wash: 1 ml Water (x2), 0.5 ml 25% ACN/Water Elute: 0.5 ml (x2) Elution solution (2:2:1, MeOH:ACN:Buffer) Buffer: 4.5 ml Phosphoric Acid, 4.5 ml TEA in 1 liter of water SPE method was downloaded from a web site The procedure is an example, the method is given as a starting point, they are designed to offer all the basic data that you ll need to make the procedure 18

Optimization of Method: As mentioned previously the conditioning steps prior to loading the sample is not as important (excluding drying the cartridge) as how the sample is loaded and the wash solutions/elution solutions used relative to recovery To prove the optimization concept a series of various wash solutions and elution solutions were used in the SPE procedure and all wash/elution solutions were collected and analyzed for breakthrough of the compounds in the wash solutions and recovery in the eluent Wash Solution Used Literature: 25% ACN/Water Solutions: 10% ACN/Water 15% ACN/Water 20% ACN/Water 25% ACN/Water 30% ACN/Water - Elution Solution Used Literature: 2:2:1 MeOH:ACN:Buffer Solutions: 1:3:1 MeOH:ACN:Buffer 3:1:1 MeOH:ACN:Buffer 4:1 ACN:Buffer 19

Optimization Grid: 5 wash solutions and 4 elution solutions Using the basic elution solution 2:2:1 the SPE cartridges were exposed with the various wash solutions repeated 8 times Once the optimum wash solution was determined then the elution solution was varied repeated 8 times Even with this relatively small study close to 100 samples will be generated 20

Multi-Collect to Optimize SPE Procedure: SPE columns Drain Collection tubes 21

Automated Optimization System: Movement and Fractionation on the Automated System, Access to the Individual SPE cartridge 22

Fractionation Task 23

Breakthrough Curve for SPE Device UV Response V B = The volume where analyte first appears (~ 1% maximal value) V R = The retention volume Sample Volume V M = The volume where maximal analyte breakthrough has occurred (~ 99% of maximal value) 24

Data and Results: Breakthrough of the Analyte in the Wash Solution 120000 Area Count at 220 n 100000 80000 60000 40000 20000 0 0 5 10 15 20 25 30 35 Series1 Series2 Linear (Series1) Linear (Series2) Percentage of Methanol Wash Graph 2: Optimization of the wash solution relative to recovery for Chlorpromazine and Thioridazine at various concentrations of methanol used in the wash solution 25

Data: The following data represents the lower levels evaluated 20 ul injection of Spiked Plasma after SPE T C 0.5 ug.ml T/C in plasma 0.1 ug.ml T/C in plasma 0.05 ug/ml T/C in plasma 0.01 ug/ml T/C in plasma Blank Plasma 26

Data and Results: Changing the elution ratio did not have any distinguishable effect on the recovery The recovery of Chlorpromazine and Thioridazine under optimized SPE conditions (15% methanol wash and 2:2:1 elution buffer) yielded 99% and 98%, respectively Although the increase in recovery for Chlorpromazine is negligible the increase in recovery for Thioridazine is significant ~10% A good rule of thumb is that SPE cartridges retain a mass of solute (analyte plus retained contaminants) this is equivalent to 5% of the sorbent mass Therefore using an automated optimization SPE system yielded positive results without manual intervention 27

Flavonoids Found in plant material and products produced from plant material Anthocyanins wide color range Blue to red Produced by plants for environmental protection UV protection in the 280 380 nm range Microorganism infestation protection by phytoalexins Host specific signal compounds 28

Classification of Flavonoids Flavonols Flavonol R1 R2 Quercetin OH H Kaempferol H H Myricetin OH OH Isorhamnetin OMe H 29

Wash Break-Through Wash MeOH Break 25 30 35 40 500 1000 1500 2000 MeOH NH4OAc 25 30 35 40 MeOH HCl 25 30 35 40 2500 X X 3000 X X X 3500 X X 4000 X X X 30

Elution Percent Recovery ACN and MTBE Elute Percent Recovery MeOH Elute Percent Recovery 120 120 100 100 Percent 80 60 40 70% ACN 80% ACN 90% ACN 100% ACN 75% MTBE 25% MeOH Percent 80 60 40 70% MeOH 80% MeOH 90% MeOH 100% MeOH 20 20 0 0 1000 2000 3000 4000 ul Elution Solvent 0 0 1000 2000 3000 4000 ul MeOH 31

Plasma myricetin quercetin kaempferol 32

Optimization of SPE Method: Flavonoids recovery at 99±1% on all analytes versus 80 % for quercetin SPE of all matrices eliminates interfering peaks Optimization of SPE method for multiple matrices allows ease of use when analyzing 6 separate matrices. 33

Elution Volume: Although we ve improved the SPE method the story is not finished In almost every situation the elution volume is too large for consistent detection S/N or the organic concentration is too high for chromatography analysis 34

Strong Sample Solvents H 2 O / CH 3 CN 100 : 0 H 2 O / CH 3 CN 70 : 30 H 2 O / CH 3 CN 50 : 50 H 2 O / CH 3 CN 30 : 70 0 4 8 min 0 4 8 min 0 4 8 min 0 4 8 min The sample should whenever possible be dissolved in the mobile phase If the sample is dissolved in too strong solvent, significant anomalies occur, peak splitting 35

Elution Solutions: Drying down the eluent volume either to the point of dryness or to a consistent volume is a major bottleneck in the SPE process, it takes time to dry the samples down and usually involve solubilizing the sample prior to analysis Degradation of samples via heating is a concern where the compounds may undergo hydrolysis, esters, which could lead to incorrect interpretation of the results 36

Alternative Solution: TEC: trace enrichment cartridges in place of the injection loop The eluent from SPE cleanup is diluted with a consistent amount of aqueous mobile phase significantly decreasing the organic strength of the elution solvent This diluted eluent solution is then exposed to the TEC The TEC will absorb the analyte of interest while the solution will pass though to waste The TEC concentrates the analyte and then the TEC is brought inline to the HPLC system The TEC not only concentrates the analyte of interest but it removes chromatographic anomalies associated with high organic percentages 37

Data and Results for TEC: Graph 3: Overlay of the plasma samples at various concentrations of Chlorpromazine and Thioridazine after SPE but without TEC (trace enrichment). A 20 ul sample was injected onto the analytical column, 0.5 mg/ml-0.005 mg/ml concentrations. 38

Data and Results for TEC: Graph 4: Overlays of the plasma samples at various concentrations of Chlorpromazine and Thioridazine. A 900 ul volume injection was introduced to the TEC (trace enrichment) prior to analysis via the analytical column, 0.5 mg/ml- 0.005 mg/ml concentrations. 39

Data and Results for TEC: Effects of TEC on Signal Chlorpromazine Effects of TEC on Signal Thoridazine 0.6 0.6 0.5 Without TEC 0.5 Without TEC Amount of Drug ug/m 0.4 0.3 0.2 With TEC Amount of Drug ug/ml 0.4 0.3 0.2 With TEC 0.1 0.1 0 0 2000000 4000000 6000000 8000000 Area Count 0 0 100000 0 200000 0 300000 0 400000 0 Area count 500000 0 600000 0 700000 0 Graphs 5 & 6: Represent the substantial increase in signal when TEC was incorporated prior to the chromatographic analysis. A 900 ul sample was loaded onto the TEC, then the TEC was switched in-line to the analytical column, increasing the S/N by 6 fold. 40

Conclusion: SPE is an important cleanup method for biological and environmental samples A good rule of thumb is that SPE cartridges retain a mass of solute (analyte plus retained contaminants) this is equivalent to 5% of the sorbent mass Optimizing SPE methods can be very advantageous however extremely labor intense and time consuming The automated Liquid Handler and SPE racks allow for the optimization process to take place via multiple collection sites and various conditions without manual intervention Incorporating the use of a TEC instead of drying down the eluent from SPE cleanup also offers a very advantageous alternative removing the dry down bottleneck and analyte degradation 41

THANK YOU! 42