Customers should provide all necessary operating supplies upon request of the engineer.

Similar documents
Agilent 1290 Infinity II Vialsampler Data Sheet

Agilent 1100 Series Refractive Index Detector. Reference Manual

Agilent InfinityLab Online SPE Solution

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

InfinityLab Quick-Connect Bio-inert Heat Exchanger

Agilent 1200 Infinity Series ELSD

Agilent 1100 Series Variable Wavelength Detector. Reference Manual

Agilent 1290 Infinity II Online SPE Solution

Agilent 1200 Infinity Series ELSD

Agilent 1200 Series Variable Wavelength Detector

Agilent 1290 Infinity II LC

Agilent 1200 Series Diode Array and Multiple Wavelength Detectors

Installation of Hotel Drawers

Agilent 1260 Infinity Multi-Detector Suite

An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

detectors UltiMate 3000 Variable Wavelength Detectors

Nexera-i Specifications Prominence-i Specifications

Agilent 1200 Series Fluorescence Detector

Target Platform. Considerations. it is recommended that the server and bandwidth are optimised for acceptable performance.

Agilent 1100 Series Diode Array and Multiple Wavelength Detectors. Reference Manual

Customised HPLC Systems for Highest Performance

Agilent 1290 Infinity Thermostat

BENEFITS. QUATERNARY HPLC SYSTEMS Quaternary and Other Gradients. The Complete Range of HPLC. Convenience of Maintenance. Systems Configuration

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

PG 990 Atomic Absorption

SEC-MALS Noise Assessment and Cleaning Guide

Agilent 1100 Series Autosampler Thermostat Reference Manual

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Series 5000 Hydrazine Analyzer Model 60002

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu)

SECTION Fluorescent Dissolved Oxygen Measuring System

Precision Temperature Control. Pulse Free Pump. Column Oven. Superb Performance Ion Chromatography. Leak Detector. Modular Flexibility

LABTRON LAAS - Series Atomic Absorption Spectrophotometer

Common Features of DX-600 Modules

EVAPORATIVE LIGHT SCATTERING DETECTORS FOR HPLC

SystemsQT for Waters Alliance HPLC Systems (Revision G)

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Technical specifications of FT-IR Spectrometer

Thermo Scientific Dionex UltiMate 3000 Fluorescence Detectors

Waters 2695 Separations Module

CALIDUS tm the Modular micro Gas Chromatograph

159 UV-VIS Detector. User s Guide. Part Number

SECTION Free/Available Chlorine Analysis System

IC Professional Detector

HPLC Column Ovens and Temperature Controllers

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Waters 2795 Separations Module

Installing the Agilent CE ESI-MS Sprayer Kit (G1607A)

SECTION ph/orp Sensors

Thermo Scientific Dionex Integrion HPIC System

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

Automated Sample Preparation: SPE GPC - EVAporation SUPERIORITY BY ADAPTATION. Visit us at

M2 Series Startup Form

Waters 2695D Separations Module

Refrigerated Incubator Model and Operating Instructions

Online SPE-UHPLC-MS. CHRONOS for Control & System Integration. Flexible autosampler for sample preparation. Superior UHPLC system for analysis

SECTION Process and Immersion Turbidity and Total Suspended Solids Sensor

IndigoVision. GAI-Tronics Integration Module. Administrator's Guide

SECTION INDUCTIVE CONDUCTIVITY ANALYZERS

REFRIGERATED BATH CIRCULATOR. Instruction Manual. Covers model : LRBC-13C / LRBC-22C. Please read this manual carefully before using the instrument

ProStar 325 UV-Vis Detector

On-line Side Stream Oil in Water Analyser - Model OIW-EX 100 or OIW-EX 1000

887 Professional UV/VIS Detector

SimpleVIS Viscometer Reference Guide P

M2 Series Startup Form

Refrigerated Cooling System

CyTOF Autosampler 5 User Manual

CS744 Series: Carbon/Sulfur by Combustion

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC

System Galaxy Quick Guide

UniGas Portable, Flue Gas Analysis Advanced Combustion Flue Gas Analyser

SAMPLE MANUAL - INSTALLATION - FOR REFERENCE ONLY. ChemScan mini Analyzer. Parameter. Method #### Installation, Operation & Maintenance Manual

Mass Sensors, Inc. Profile. Applications. Technology Baur Blvd, St. Louis, MO (314)

Absorbance range Resolution Accuracy Type of microplate Wavelength Optic system Light source Wavelength accuracy. Calculation method

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Specifications Wavelength Range: 190 nm to 800 nm continuous

Atomic Absorption Spectrophotometer AAS 2320D EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Commissioning Methodologies (CM) MUBS-09 Duct Mounted Temperature & Humidity Sensor Testing Method Statement

On Line SPE LC/MS/MS Applications

Model 6650SP. Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer

Date SECTION MEASUREMENT AND CONTROL INSTRUMENTATION Project Name Page 1

FPD - Flame Photometric Detector

HGM300/RDM800 PREVENTATIVE MAINTENANCE AND TECHNICAL SUPPORT. SECTION 1 HGM300 CPU reset, RDM800 CPU reset, WARM UP notes and HGM300 Self Test

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Developing UV LED Curing Processes for Converting Applications

940 Professional IC Vario

ATOMIC ABSORPTION SPECTROMETER AAS 3500 EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES / SERVICES

Perfect Separation Solutions (Bio)SECcurity²

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

TacBio: A UV-based Biological Aerosol Detector. TacBio Overview Page 1

DHR Accessories Requirements. Site Preparation Guide

Temperature + Humidity (%RH) Control & Recording System with 8 / 16 Channel Mapping. Touch Panel PPI

151/152 UV/VIS Detectors User s Guide

Vanquish. Variable Wavelength Detectors VF-D40. Operating Manual

Transcription:

Agilent Installation InfinityLab Checklist LC Series Thank you for purchasing an Agilent instrument solution. This checklist is used by the installing engineer to ensure that the instrument and associated systems are correctly installed, upgraded and functioning as designed in your facility. This checklist will be completed at the end of the service and provided to you as a record of the installation. Customer Information Customers should provide all necessary operating supplies upon request of the engineer. Service Engineer s Responsibilities Only complete/printout sections or pages that relate to the system that has been installed. Complete empty fields with the relevant information. Complete the relevant checkboxes in the checklist using a X or tick mark in the checkbox. Complete Not Applicable check boxes to indicate services not delivered, as needed Complete the Service Review section together with the customer. In case of diagnostic tests, record test results in section service review (test results). System Information Guidance: Check box if instrument configuration report is attached instead of completing the table below. Instrument System Name/I.D: Record the list of system component product numbers below. Instrument Location: List the serial numbers of the components present in the system below. 1. 1. 2. 2. 3. 3. 4. 4. 5. 5. 6. 6. 7. 7. 8. 8. 9. 9. 10. 10.

General Preparation Unpack/verify the condition and completeness of shipment. For discrepancies, use the table below. Product/Part Description Observation Action Discuss any specific questions or issues with the customer prior to starting. Discuss any configuration options with the customer prior to starting Check for required service note applicability and firmware updates if connecting to instruments. When plumbing the customers system, make sure that the relevant capillary kit and instructions are used. If you use the PC or the Instant Pilot for firmware updates, use the correct date/time. Check that customer-supplied PC hardware meets minimum requirements e.g. Site Preparation Tool http://www.agilent.com/en-us/support/software-informatics/swsitepreptool Upgrades only - Ensure with customer that instrument control settings, data, methods etc have been properly saved or archived prior to starting any installation procedures. The installation of ELSD drivers depends upon the customer's choice of software platform. If OpenLAB version A.02.01 is installed then ELSD drivers' version 1.2 is installed as part of the installation. An UPGRADE to 1.3 will be seen during the new install. If a THIRD PARTY software package is used a minimum of ICF A.02.02 is required. Agilent Information Center Install the Agilent Information Center to the PC controlling the system. OR If this is not permitted, install the Agilent Information Center to a different PC. OR If this is also not permitted, use the installation medium ( Agilent InfinityLab LC Series User Documentation (G4800-64600)) to view the required user documentation. Record the location to which the Agilent Information Center has been installed. OR If the Agilent Information Center has not been installed, record that the installation medium has been delivered, and keep it close to the instrumentation for future reference.

G7122A Degasser Set up the degasser(s) solvent inlet/outlet Prime each channel with isopropanol from solvent bottle to pump inlet.

G7120A High Speed Pump, G7104A/C Flexible Pump Set up the pump solvent inlet/outlet tubing waste and seal wash sensor outlet remote control (if used) CAN connect LAN NOTE Agilent LAN connection is made between one (1) of the Agilent Infinity II modules and the Agilent Control Software PC. If a detector is installed connect the LAN to this module. fill each channel with Isopropanol from solvent bottle to degassing unit inlet Prepare the pump prepare solvents turn on module activate Prime configure the purge activate the purge condition the pump, the tuning should be between ±0.2

G7110B Isocratic Pump, G7111A Quaternary Pump VL, G7111B Quaternary Pump, G5654A Bio-inert Pump, G7112B Binary Pump Install optional field upgrade kits for Active Seal Wash (G1398A/G1399A): Install seal wash pump head parts. Install the seal wash pump motor and peristaltic pump. Install tubing connections for the ASW. After completing the pump installation, run a leak rate test. Set up the pump(s) solvent inlet/outlet waste drain remote control (ERI) (if used) seal wash tubing (if installed) CAN connect LAN NOTE Agilent LAN connection is made between one (1) of the Agilent Infinity II modules and the Agilent Control Software PC. If a detector is installed, connect the LAN to this module. Install the solvent cabinet (optional). Priming the pump. Fill each channel with Isopropanol from solvent bottle to degassing unit inlet. Turn on the pump. Purge each channel for 3 min at 3 ml/min. After completing the stack installation, purge the complete system. Monitor the pressure, ripple should be <2 %.

G7129A/B/C Vialsampler G7129A/B/C Vialsampler Set up the sampler(s) solvent inlet multi-draw kit (if ordered) waste drain (must be in a straight line down, no bends, no kinks) - check manual for further info remote control (ERI) (if used) CAN LAN Prepare the sampler and fill the available channel with appropriate wash solvent from the solvent bottles. Turn on the module. Configure Prime Activate Prime ICC heater G7130 Set up the ICC heater G7130 Note the product number / Serial number of the Integrated Column Compartment heater install the solvent inlet/outlet Install the column Perform ICC Heater Test ICC Heater Test Pass/Fail

G7167A/B Multisampler, G5668A Bio-Inert Multisampler Perform an auto referencing under operating conditions. Set up the Multisampler(s) solvent inlet drawer configuration update (if necessary) multi-draw kit (if ordered) waste drain (must be in a straight line down, no bends, no kinks) - check manual for further info remote control (ERI) (if used) CAN LAN Prepare the sampler and fill the available channels with appropriate wash solvent from the solvent bottles Turn on the module. Configure Prime for the available channels Activate Prime for the available channels

G7167-60005 Sample Cooler CAUTION Damage to the multisampler or the sample cooler electronics Do not connect the Sample Cooler cables while the autosampler line power is connected. Wait at least 30 min before you switch on the compressor. Set up the Sample Cooler (the module is integrated in the sampler). Sample cooler power cable Sample cooler signal cable Sample cooler condensed water drain (must be in a straight line down, no bends, no kinks) - check the manual for further info NOTE Check the proper horizontal position NOTE Wait at least for 30 min before swichting on the sample cooler in the cds system Perform Sample Cooler Test. Sample Cooler Test Pass/Fail

G7167-60101 Sample Thermostat Section not applicable WARNING Flammable refrigerant Formation of flammable gas-air mixtures inside the sample thermostat. Keep open fire or sources of ignition away from the device. Ensure a room size of 1 m 3 for every 8 g of r600a refrigerant inside of the sample thermostat. Ensure adequate ventilation: typical air exchange of 25 m 3 /h per m 2 of laboratory floor area. Do not use mechanical devices or other means to accelerate the defrosting process. CAUTION Damage to the multisampler or the sample thermostat electronics Do not connect the Sample Thermostat cables while the autosampler line power is connected. Wait at least 30 min before you switch on the compressor. Set up the Sample Thermostat (the module is integrated in the sampler). Sample thermostat power cable Sample thermostat signal cable Sample thermostat condensed water drain (must be in a straight line down, no bends, no kinks) - check the manual for further info NOTE Check the proper horizontal position NOTE Wait at least 30 min before switching on the sample thermostat in the CDS system. Perform Sample Thermostat Function Test.

G7116A/B MCT Set up the MCT install the back side of the divider assembly install valve head product (see separate section) Note the product number / part number of the installed valve head (valve head part number or n/a) install the capillary kit, which belongs to the valve head product (if applicable) install the capillaries to heat exchanger install the heat exchanger install the column holders install the low dispersion heat exchangers (if applicable) install the heat exchanger with the InfinityLab Quick Turn Fitting (if applicable) connect the capillaries to the column, using the Infinity Lab Quick Connect or Quick Turn Fittings (if applicable) install the column into the column holders install the column ID Readers (if applicable), module has to be powerless Install the column ID Tag install the waste drain Install the CAN connection install the front cover of the divider assembly Turn on the module Perform Thermostat Test Thermostat Test Pass/Fail

G1170A Valve Drive Install the valve drive(s) Install the leak plane and waste drain Install the CAN connection Install the valve head product Note the product number / part number of the installed valve head (valve head part number or n/a) Install the capillary kit delivered with the valve head

G7114A/B VWD Set up the detector deuterium lamp (if not installed) Turn on the detector and lamp as soon as possible (minimum one hour warm-up time) flow cell solvent inlet/outlet waste drain remote control (if used) analog output (if used) CAN connect LAN (the detector should be always the host of the LAN communication with the PC) Verify the operation of the detector. Perform the Lamp Intensity Test Perform the Wavelength Calibration Test Lamp Intensity WL calibration Pass/Fail Wavelength Deviation of 0-Order Wavelength Deviation of Alpha Line (656 nm)

G7115A DAD, G7165A MWD Set up the detector(s) deuterium lamp (if not installed) Turn on the detector and lamp as soon as possible (minimum one hour warm-up time) solvent inlet/outlet waste drain remote control (if used) analog output (if used) CAN connect LAN (the detector should be always the host of the LAN communication with the PC) Verify the operation of the detector. Perform the Lamp Intensity Test Perform the Wavelength Calibration Test Lamp Intensity WL calibration Pass/Fail D2 Alpha Line Deviation (486 nm) D2 Beta Line Deviation (656 nm)

G7117A DAD FS, G7117B DAD, G7117C DAD HS Set up the detector(s) deuterium lamp (if not installed) Turn on the detector and lamp as soon as possible (minimum one hour warm-up time) solvent inlet/outlet waste drain remote control (if used) analog output (if used) CAN connect LAN (the detector should be always the host of the LAN communication with the PC) Verify the operation of the detector. Perform the Lamp Intensity Test Perform the Wavelength Calibration Test Lamp Intensity WL calibration Pass/Fail D2 Alpha Line Deviation (486 nm) D2 Beta Line Deviation (656 nm)

G7121A FLD, G7121B FLD Spectra Set up the detector solvent inlet/outlet waste drain remote control (if used) analog output (if used) CAN connect LAN (the detector should be always the host of the LAN communication with the PC) Verify the operation of the detector. NOTE New instruments are shipped with a flow cell installed and have a factory wavelength calibration. If a flow cell is used other then the installed/shipped flow cell: verify that the installed flow cell is fixed very tight, perform a wavelength calibration using Glycogen (part of FLD Calibration Kit (G7121-68001)). Perform the Lamp Intensity Test & compare profile Perform the Wavelength test Perform Verification Test using the Raman band. Lamp Intensity WL verification Pass/Fail Excitation Deviation Emission Deviation If test fails perform Wavelength Calibration.

G7162A/B RID Set up the detector NOTE Depending on the detector different inlet tubings are used (from the Accessory Kit): G7162A: Interfacing capillary (G1362-87300) G7162B: Capillary ST 0.075 mm x 220 mm (5067-4784) (Sampler to MCT/TCC) Capillary ST 0.075 mm x 340 mm (5067-4783) (Column to RID) Turn on the detector and heater as soon as possible (minimum one hour warm-up time). solvent inlet/outlet solvent recycle waste drain analog output (if used) CAN connect LAN (the detector should be always the host of the LAN communication with the PC) Verify the operation of the detector Diode Balance Value

G7102A, G426X A/B ELSD Set up the detector Verify that the instrument is the correct module for the regional voltage (120 V/240 V). Do NOT use instruments which have an incompatible voltage Connect the exhaust line to a fume hood or lab exhaust line Set the correct pressure at the regulator for the pressurized nitrogen line and connect it to the detector Install the capillary connection from detector to HPLC Install the RS-232 connection from detector to ChemStation control PC (if applicable) Install the Aux I/O cable connection from detector to the 1200 stack Connect the power lead Connect the front drain to the waste container/bottle Configure the detector in OpenLAB CDS ChemStation Edition or EZChrom Edition. "OpenLAB CDS ChemStation and EZChrom Edition, A.01.05 Minimum". ELSD driver version 1.3 Installation is essential for data collection G7102A, G426X A/B Checkout Switch on the instrument Prepare the instrument: Set the detector nebulizer and evaporator temperature to 70 C, and gas flow to 1.60 SLM Wait until instrument ready status is displayed on front panel or PC Record 10 min of baseline noise. This is gas only noise and should be typically 0.5 LSU or 0.5 mv for G426X A/B (ASTM baseline noise). Set up a method for the checkout sample as below: Flush the system with Eluent for 15 min AND Verify instrument ready status is still displayed on front panel or PC In the instrument or system, set up a method using the following parameters: Pump: 1.00 ml/min of 100 % water. NOTE Depending on the flow rate setting, the system may overpressure and/or the PEEK fittings may not hold if pressures exceed 350 bar. If this happens, bypass the restriction capillary and guard column (usually for flow rates beyond 2.0 ml/min for UHPLC or 3.0 ml/min for HPLC). Injection Volume: 20.0 µl. NOTE The following parameters work well for the recommended default injection volume of 20.0 µl. However, a scouting run is always recommended because run time varies across systems. Evaluation Std.: 200 µg/ml caffeine standard (part# 5190-0488) Restriction capillary: S/S restriction capillary (Part# 5022-2159)

ELSD Detector: Nebulizer Temperature: 70 C Evaporator Temperature: 70 C Gas Flow: 1.60 SLM Data Rate 10 Hz Smoothing 1 s PMT Gain 1 (No PMT Gain setting for G7102A) LED Intensity 100 %( No Laser setting for the Agilent G7102A or G4261B) Column Compartment: Column Temperature Stability (default is 40 C). If a column compartment is not installed, use ambient temperature. Run Time: 2.0 minutes (adjust if necessary) Table 1 Pass Specs Signal Noise and Drift ASTM baseline noise Slope of regression fit for drift Noise: 2.000 LSU (or 2.000 mv for G4260X A/B) Drift 5.000 LSU/hr (or 5.000 mv/hr for G4260X A/B) Injection Precision: Injection Volume = 20 µl Height RSD 3.00 % Area RSD 3.00 % Signal Height: 300mV LSU (or 300 mv for G426X A/B) Run a test measurement Attach results to installation documentation

G4208A 1200 Instant Pilot Set up the control module Connect via CAN to available instrument module Assure that the latest firmware is installed M8550A Lab Advisor Advanced Section NOT Applicable Install the Lab Advisor software on the controlling PC Open the Lab Advisor software. Navigate to the License tab under Configuration and enter the Lab Advisor Advanced license M8555A Lab Advisor Basic Section NOT Applicable Verify with customers if they would like it installed If applicable install the Lab Advisor software on the controlling PC

Service Review Attach available reports or printouts to this documentation. Record the time and date of installation or upgrade completion in the customer s records or logbook. Complete the Service Engineer Comments section below if there are additional comments. Review the installation or upgrade with the customer. Explain Agilent warranty for instruments. Explain how to use manuals, guides, online help. Explain how to get self-help, FAQs from the web. Explain how to log an instrument service call, support services available. Advise customer of additional instrument training options. If the instrument firmware was updated, record the details of the change in the Service Engineer Comments box below or if necessary, in the customer s IQ records.

Service Engineer Comments (optional) If there are any specific points you wish to note as part of performing the installation or other items of interest for the customer, please write in this box. Firmware was changed (Y/N): Client was consulted and agreed prior to firmware change (Y/N): Service Completion Service Request number. Agilent Signature Total no. of pages for this document: Date service completed Customer Signature.. Other Important Customer Web Links How to get information on your product: Literature Library - http://www.agilent.com/chem/library Need to know more? - www.agilent.com/chem/education Need technical support, FAQs? - www.agilent.com/chem/techsupp Need supplies? - www.agilent.com/chem/supplies