Rosemount 372. HF Resistant ph sensor. Quick Start Guide LIQ-QSG-372 Rev. C June 2017

Similar documents
Quick Start Guide LIQ-QSG-Hx338, Rev M July Rosemount Hx338. Steam Sterilizable and Autoclavable ph Sensor

Retractable ph/orp Sensor

Rosemount 3900/3900VP

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors

Rosemount 3200HP. ph Sensors. Instruction Manual LIQ-MAN-3200HP Rev. G April 2017

Rosemount 396/396VP/397/398/398VP

ENDURANCE Conductivity Sensors

Durafet II and Meredian II Combination ph/orp Electrodes for Honeywell 7777 Mountings Series 7777 Specifications

PURSense Sanitary Flange Conductivity Sensors

Durafet Non-Glass ph Electrodes Specifications

Dissolved Oxygen Sensors

Model 5190-F Temperature/Humidity Probe

Model Q45P/R. ph/orp Monitor...

Dual Input Toroidal Analyzer

Dual Input ph/conductivity Analyzer

Model 5190-E Temperature/Humidity Probe

HydroMaster USER S GUIDE HM-500

Dual Input ph/orp Analyzer

JENCO ELECTRONICS, LTD.

Instruction Manual LIQ-MAN-FCL-1056, Rev F July FCL with 1056 Transmitter

Hydrazine Analyzer Series 2171 Specifications

Model Q45P/R. ph/orp Monitor...

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

PH/ORP Online Analyzer

203RE ph v 1.0 Instructions

JENCO INSTRUMENTS, INC. JENCO ELECTRONICS, LTD ARJONS DR., SUITE C P.O. BOX LINKOU-117

Signet V(D) to V(D) PVDF Conductivity Electrodes

Duct Mount. Installation Instructions

Signet ph/orp Electrodes

Manual LIQ-MAN-TCL-1056, Rev E July Rosemount TCL. Total Chlorine System with Rosemount 1056 Transmitter

User s Guide. PHE-2724/2726 SERIES ORE 2725 SERIES ph/orp Electrodes

PH / ORP Analyzers MODEL OPM223/253. Wedgewood Analytical Inc. Liquid Analytical Measurement & Control Specialists DESCRIPTION FEATURES APPLICATIONS

ph Measuring System Start-up and maintenance manual e Contents

Rosemount 3051S MultiVariable Transmitter

OWNER S MANUAL & WARRANTY

Free Chlorine Measuring System

Hygro/hygrothermal transducers (hygrometric)

ORP/MILLIVOLT CONTROLLER

Instruction Manual PN 51-FCL/rev.H March Model FCL Free Chlorine Measuring System

Net Safety ST3 XChem Toxic Gas Sensor

245 Sanitary Flow-Through Toroidal Conductivity Sensor. Instruction Manual LIQ_MAN_245 Rev. D January 2014

Instruction Manual CHLORALERT TM Series T17CA4000 Chlorine Gas Leak Detector CAPITAL CONTROLS

Instruction Manual. CAT 100 & CAT 200 Continuous Analyzer Transmitter Addendum for Hazardous Area Applications. 3 rd Edition 05/2007

Black Body Calibration Source User Manual

Magnetically Operated Touch Panel Hints on Operating CAT 100 / BINOS 100 F and CAT 200 / MLT 2 Gas Analyzers

OPERATION MANUAL. THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

ph/orp-7500 ph/orp Transmitting controller Operation manual

SUBMERSIBLE Flush Diaphragm Liquid Level Sensor AST4520

Toroidal (Electrodeless) Conductivity Sensors Series 5000TC Specifications

CLARITY II TURBIDIMETER

HP 101A Series Humidity Temperature Probes INSTRUCTION MANUAL

User Guide. ROSS Process Combination ph Electrode

Xi Advanced Electronics for Zirconium Oxide Oxygen Probes

PERSONAL Ultrasonic Cool Mist Humidifier

SECTION ph/orp Sensors

DIGITAL ULTRASONIC HUMIDIFIER WITH HUMIDISTAT

RK Carbon Monoxide Transmitter Operator s Manual

MEAS CAL ALARM WASH CONF APT4000 PH ENCLOSURE TYPE 2. Description

40/40 Duct Mount P/N Installation Guide. Document Ref: TM777670, Rev. (3) June 2017

Hygro/hygrothermal transducers (hygrometric)

Ultrasonic Humidifier

LAKEWOOD INSTRUMENTS MODEL 353 ORP/MILLIVOLT CONTROLLER

Product waste disposal - Protection of the environment:

FOR THE FOLLOWING MODELS: If you have any questions about the operation of your ultrasonic humidifier, please contact Crane Customer Care.

Instruction Manual PN 51-MCL/rev.C November Model MCL Monochloramine Measuring System

EVAPORATIVE COOL MIST HUMIDIFIER Instruction Manual

ExStik DO600 USER GUIDE. Dissolved Oxygen Meter. 99 Washington Street Melrose, MA Phone Toll Free

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Rosemount 4500 Hygienic Pressure Transmitter Demonstrated best-in-class performance during SIP/CIP for process temperatures up to 400 F (204 C)

Model OEM-2 INSTRUCTIONS FOR USE

Duke Manufacturing Co.

Standard SPECIFICATIONS FEATURES APPLICATIONS

Refrigerated Incubator Model and Operating Instructions

ULTRASONIC HUMIDIFIER With Night Light and Fragrance Diffuser

SECTION Free/Available Chlorine Analysis System

NB: RISER SHELVES. Failure to do so will invalidate your warranty.

ADORABLE Ultrasonic Cool Mist Humidifier

Configuration Data Sheet

Note To The User: This product has updated features not yet mentioned in the product manual.

5000TC Series Toroidal (Electrodeless) Conductivity Sensors. Specification. Overview. Features. Description /00 Page 1 of 8

Quick Start Guide , Rev. BB February Rosemount 1495 Orifice Plate Rosemount 1496 Orifice Flange Union

Intelligent ph Converter PH400G. PH400G Intelligent ph Converter. Bulletin 12B7C1-01E.

Instruction Manual PN OZ/rev.K February Model SOLU COMP Ozone Analyzer

Model gffoz Process Ozone Sensor (In-Line)

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

Emerson Wireless SmartPower Solutions

MEAS CAL ALARM WASH CONF APT4000 PH ENCLOSURE TYPE 2. Description

MS1471 SPECIFICATIONS. PC Board Mountable Pressure Sensor 0-60 mv Output Gage and Absolute Low Cost

Corporate Profile. needs and product demands. CON 08X /03 Rev 0

SuperOX TM Oxygen Sensor U.S. Patent No. 5,635,044

Package Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents

1 GALLON / 3.78 LITER ULTRASONIC HUMIDIFIER

Revolutionary Science RS-IF-202

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

SimpleVIS Viscometer Reference Guide P

Owner s Manual PS-50. Document Number: Revision: Draft

Instruction Manual PN pH/rev.M February Model Solu Comp II. Dual Input ph Analyzer. Model Options and

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

PH/ORP PH/ORP Tester. Instruction for use

PERSONAL COOL MIST ULTRASONIC HUMIDIFIER

Transcription:

Rosemount 372 HF Resistant ph sensor Quick Start Guide LIQ-QSG-372 Rev. C June 2017

hasgkas

Essential Instructions Read this page before proceeding! Emerson designs, manufactures and tests its products to meet many national and international standards. Because these sensors are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life; personal injury; property damage; damage to this sensor; and warranty invalidation. Read all instructions prior to installing, operating, and servicing the product. If you do not understand any of the instructions, contact your Emerson representative for clarification. Follow all warnings, cautions, and instructions marked on and supplied with the product. Inform and educate your personnel in the proper installation, operation, and maintenance of the product. Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources. To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product. When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Third-party substitutions may result in fire, electrical hazards, or improper operation. Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury. The information contained in this document is subject to change without notice. CAUTION Sensor/Process Application Compatibility The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user. CAUTION Before removing the sensor, be absolutely certain the process pressure is reduced to 0 psig and the process temperature is at safe level. CAUTION Special Conditions for Safe Use 1. All ph/orp sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electrostatic charge. 2. All ph/orp sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation.

About This Document This manual contains instructions for installation and operation of the Rosemount 372 HF Resistant ph sensor. The following list provides concerning all revisions of this document. Rev. Level Date Notes C 06/17 Reformatted to reflect the latest Emerson documentation style Specifications and Wiring Diagrams.

Quick Start Guide Table of Contents LIQ-QSG-372 June 2017 Contents Section 1: Specifications 1.1 Specifications...1 1.2 Ordering Information...2 Section 2: Installation 2.1 Unpacking and Inspection...3 2.2 Storage...3 2.3 Electrode Preparation...3 Section 3: Wiring 3.1 Wiring for Rosemount 372...5 Section 4: Calibration and Maintenance 4.1 Two Point Buffer Calibration...9 4.2 Recommended ph Sensor Standardization...9 4.3 Maintenance...9 Table of Contents i

Table of Contents June 2017 Quick Start Guide LIQ-QSG-372 ii Table of Contents

Quick Start Guide Specifications LIQ-QSG-372 June 2017 Section 1: Specifications 1.1 Specifications Table 1-1: Rosemount 372 HF resistant ph sensor specifications Materials of construction Ultem, Viton, glass, Kynar Process connections 3/4 in. forward (insertion) and 1 in. reverse (submersion) MNPT Insertion depth 2.2 in. to 8.3 in. Measured range 0 to 12 ph (0-14 ph if no sodium ions are present) Pressure range -13 to 135 psig (6.9 to 1035 kpa abs) Temperature range -5 to 50 C [23 to 122 F] (100 C [212 F] if no fluoride ions are present) Temperature compensation Pt 100 Weight/shipping weight 0.4 kg/0.9 kg (1 lb/ 2 lb) ph glass type Hemispherical Glass diameter 7.8mm Glass impedance Less than 400 MΩ @ 25 C Acidic error Less than.01 ph in 1.0M HCl @ 0.0pH Reference type Double junction with saturated KCL in cross-linked polymer Specifications 1

Specifications June 2017 Quick Start Guide LIQ-QSG-372 1.2 Ordering Information Table 1-2: Rosemount 372 HF resistant ph sensor ordering information Model Sensor type 372 ph/orp Sensor Preamplifier/cable 01 25 ft. cable with integral preamplifier 02 15 ft. cable without integral preamplifier Typical Model Number: 372-01 2 Specifications

Quick Start Guide Installation LIQ-QSG-372 June 2017 Section 2: Installation 2.1 Unpacking and Inspection Inspect the outside of the carton for any damage. If damage is detected, contact the carrier immediately. Inspect the instrument and hardware. Make sure all items in the packing list are present and in good condition. Notify the factory if any part is missing. 2.2 Storage 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -5 C (23 F). 3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342). 4. For overnight storage, immerse the sensor in tap water or a 4 ph buffer solution. 5. A ph glass electrode has a limited shelf life of one year. Figure 2-1 Sensor dimensions 2.3 Electrode Preparation 1. Remove electrode from shipping container. 2. Remove the protective boot covering the electrode bulb. 3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there. NOTICE Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean before installation. Installation 3

Installation June 2017 Quick Start Guide LIQ-QSG-372 Figure 2-2 Sensor orientation 4 Installation

Quick Start Guide Wiring LIQ-QSG-372 June 2017 Section 3: Wiring 3.1 Wiring for Rosemount 372 For additional wiring information on this product, including sensor combinations not shown here, please refer to the Liquid Transmitter Wiring Diagrams. Figure 3-1: Rosemount 372-01 to Rosemount 1056/56/1057 Transmitter Wiring Wiring 5

Wiring June 2017 Quick Start Guide LIQ-QSG-372 Figure 3-2: Rosemount 372-01 to Rosemount 1066 Transmitter Figure 3-3: Rosemount 372-01 to Rosemount 5081 Transmitter Wiring 6 Wiring

Quick Start Guide Wiring LIQ-QSG-372 June 2017 Figure 3-4: Rosemount 372-02 to Rosemount 1056/56/1057 Transmitter Wiring Figure 3-5: Rosemount 372-02 to Rosemount 1066 Transmitter Wiring Wiring 7

Wiring June 2017 Quick Start Guide LIQ-QSG-372 Figure 3-6: Rosemount 372-02 to Rosemount 5081 Transmitter Wiring 8 Wiring

Quick Start Guide Calibration and Maintenance LIQ-QSG-372 June 2017 Section 4: Calibration and Maintenance 4.1 Two Point Buffer Calibration Select two stable buffer solutions, preferably ph 4.0 and 7.0 (ph buffers other than ph 4.0 and ph 7.0 can be used as long as the ph values are at least two ph units apart). NOTICE A ph 7 buffer solution reads a mv value of approx. zero, and ph buffers read approximately ± 59.1 mv for each ph unit above or below ph 7. Check the ph buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mv/ph value. 1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by transmitter. 2. Once the first buffer has been acknowledged by the transmitter, rinse the buffer solution off of the sensor with distilled or deionized water. 3. Repeat steps 1 and 2 using the second buffer solution. 4. The theoretical slope value, according to the Nernst equation for calculating ph, is approximately 59.17 mv/ph. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mv/ph. 4.2 Recommended ph Sensor Standardization For maximum accuracy, the sensor can be standardized online or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor s slope but will simply adjust the transmitter s reading to match that of the known process ph. 4.3 Maintenance Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. 2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution. 3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% @ 25 C (77 F), then it will be approximately 25% @ 80 C (176 F), and approximately 5% @ 120 C (248 F). Calibration and Maintenance 9

LIQ-QSG-372 Rev. C June 2017 www.emerson.com/rosemountliquidanalysis Youtube.com/user/Rosemount Twitter.com/Rosemount_News Analyticexpert.com facebook.com/rosemount Emerson Automation Solutions 8200 Market Blvd. Chanhassen, MN 55317, USA Tel +1 800 999 9307 Fax +1 952 949 7001 Liquid.CSC@Emerson.com 2017 Emerson Automation Solutions. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.