Operating Manual UV Monitor UVT 18

Similar documents
UV - Monitor UVT 16. Use. Description

ZED UV Monitor PRO11DPI-I

User Manual. Dryer Controller M720

EKC 347 Liquid Level Controller REFRIGERATION AND AIR CONDITIONING. Manual

Dryer Controller M720

Insulation fault evaluator EDS460-DG

1.7. Insulation fault evaluators EDS460/490 EDS461/491

PRODUCT BULLETIN. Rev. A Monitoring Station offering PN Rev. B as a direct replacement. OVERVIEW / DESCRIPTION

Two-Channel Gas Controller

1.7. Insulation fault evaluators EDS460/490 EDS461/491

ISOSCAN EDS460-DG. Insulation fault locator for DC IT systems with high system leakage capacitances

THX-DL Data Logger USER & INSTALLATION MANUAL V

Fire Extinguishing Control Panel INSTRUCTION MANUAL. Revision 8/ Instruction Manual Page 1 Revision 8/01.17 of 63

No part of this publication may be reproduced, stored in an automated data file or made public in any form or by any means, whether electronic,

FLOW CONTROLLER TYPE S/601

ISOSCAN EDS460/490 EDS461/491. Insulation fault locators with control and display function for EDS systems (insulation fault location systems)

MC 31 Speed Controller Technical User Guide Code no GB

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

REPEATER FS5200R INSTRUCTION MANUAL

GMH 285 / GMH 285-BNC

Turbimax CUE21 / CUE22

1.7 A-ISOMETER IRDH575. RoHS. Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

RESISTRON. Replacing RES-102/RES-420. Replacement Instructions. Comparison of RES-102 / RES-420. Dimensions. Temperature indication (actual value)

Control of the system, MLP RO pump station

DOL 31 Speed Controller Technical User Guide

Multi-Gas-Controller MGC2

Danfoss gas detection unit Type GD Heavy Duty

Product Data Sheet. Remote Terminals. Features:

Oxygen (O2) Single-Point Gas Detection System

RES-210 (-211, -220, -221) / RES-420

A T L A S - SolarClimatic Test Cabinet SC 2000 MHG

Fire control panels FC360

Fire Control Panel FS5100

ETO 2 - Controller for ice and snow melting

Remote alarm indicator and test combination MK2430

Water temperature LBX 3 LBX 33: recommended range 5-25 C (41 77 F) LBX 90e LBX 1500e: recommended range 5-45 C ( F)

Panel Mounted Fault Annunciator Series

Model: 1 Series 12V. Available in: Sapphire Black and Glacier White

IMPORTANT. PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times!

Danfoss gas detection units

Fire control panel detect 3004 plus and accessories

Gas Controller Module GC-06

NA G µair CONNECT 2

Dryer Moisture Systems Inc. Dryer Master M 2. Product Manual

stationary dose rate measuring systems

EasyTronic III MANUAL SERVICE

T60xDFH-4 and T60xDFH-4+PIR Series Thermostat Controllers with Dehumidification and Occupancy Sensing Capability

Panel Mounted Fault Annunciator Series

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Phone: Fax:

FIL-SPL Flow-through heater for central heating USER MANUAL

Entero Voice Alarm system.

TEC2620 Series Non-Programmable Fan Coil Network Thermostat Controller and Remote I/O Relay Packs

CT224 Setup Tutorial

PROCESS & TEMPERATURE UNIVERSAL INPUT DIGITAL METERS

TYPE CM-2201 NELSON SINGLE POINT CIRCUIT MANAGEMENT SYSTEM

GMA4100-DP. Control Panel. Operation Manual. Worldwide Supplier of Gas Detection Solutions

Chiller Organizer 2/3

Process & TeMPerATUre UniversAl input DigiTAl MeTers

Safety Temperature Limiter STL 50 Certified to DIN EN (replaces DIN 3440)

Smart thermostat with Humidification/De-humidification control

Model: Available in: Sapphire Black and Glacier White. 1 Series

Syncro AS. Analogue Addressable Fire Control Panel. User Manual

AD-1272 Advanced Thermal Dispersion Probe Airflow Measuring System

ECL Comfort 210 controller, Remote Control Units ECA 30 / 31 and Application keys

Operation manual. Rooftop Packaged Unit

Danfoss gas detection Controller unit

ProStar 325 UV-Vis Detector

BSR Addressable Fire Detection control panel. Installation operation manual

TE809-Ats Instructions Manual

ISOMETER IRDH375. Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems)

Application Manual Loop Water Controller

CELLTROL II BIOREACTOR CONTROL SYSTEM OPERATIONS MANUAL

MODEL SF-10 CONTROL OPERATION AND INSTRUCTION MANUAL

Process Alarm Module RTA 421

ZXe INTELLIGENT MULTI PROTOCOL FIRE ALARM CONTROL PANEL

ISOMETER IR425. Insulation monitoring device for unearthed AC/DC control circuits (IT systems)

56W Vapor Proof LED Fixture V Dimmable - DALI Network lms - IP67 - Outdoor Applications

Farm Hand CCU 111 / CCU 10R. Manual for use and maintenance. Farm Hand. Climate Controller. HC/MMA/UmGB /13 rev 1.

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

VAV Thermostat Controller Specification and Installation Instructions. Model TRO24T4XYZ1

SCAN20 Series Gas Detection Controller

CMD420 / CMD421 Series

Window facia fault annunciators

Dryer Master Inc. Dryer Master AM 3. Product Manual

Table of Contents. Product Image Table of Contents What is a Programmable Room Thermostat? Installation Procedure

ECL Comfort 210 controller, Remote Control Units ECA 30 / 31 and Application keys Designed in Denmark

Products no longer available

DIGITAL TEMPERATURE RELAY TR-100

ISOMETER IRDH275BM-7 with coupling device AGH675-7 and AGH675-7MV15

Wallace & Tiernan Controllers & Analysers ChemTrim Disinfection Controller

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

Operating Manual. UV Sensoric Components

Product Manual SZ1145

User Guide

LCF Touch Modbus Datasheet Application Security Advice Caution

S/601 Controller Installation Manual

USER MANUAL DIAGNOSTIC MODULE

Tempered Water Logic Control OPERATION l TROUBLE SHOOTING

Transcription:

Operating Manual UV Monitor UVT 18 Software Version 1.2 stand: 21/11/2016

UV - Monitor UVT 18 Intended use The signal monitor UVT 18 was developed to control small- and medium-sized UV systems and is fully compatible to the rules DVGW W294 and ÖNORM M5873. Next to monitoring the irradiation strength, measured by an UV sensor, it also allows the connection of further sensors reading other important values such as temperature and flow and offers dose calculations. Description The monitor is made panel-mounting. All connection inputs are on the backside of the monitor housing. The front panel is equipped with control keys and a two line alphanumerical display with multi-colored background lighting to monitor measuring values and operation conditions. The values which shall be shown on the display can be preselected in the menu and chosen via the control keys. In addition, the complete adjustment and programming of the monitor is done via these keys. A password protection can limit the operational functions to a reduced menu standard operations, if necessary. Moreover, the monitor UVT 18 does include an optional RS485 interface comtable programming and transferring measuring values to a superior control PLC. Alternatively, this interface is applicable to connect digital sensors variable measuring values. Technical data Mains supply 90-264 V AC 50 to 60 Hz or 100-300 V DC, power consumption 2.5 W Size W x H x D Monitor size 96 x 96 x 64, fixation via screw clamps Installation dimension Cut-out 92 +0,8 x 92 +0.8 acc. to DIN 43700, installation depth minimum 75 mm, Connections Pluggable screw clamps, appropriate wire cross section 0.25-1.5 mm 2 Display Alphanumeric LCD-Display, two lines, colored backlight acc. to alarm status Handling Via 6 keys in front of monitor Temperature Allowed ambient temperature 0 to 40 C Weight Approx. 500 g Inputs - UV diode input direct connection of photodiodes or sensors relative measurements - Input UV sensor with voltage output 0..2/5/10 V DC - Input sensor with voltage output 0..2 V DC ( temperature sensor or flow meter) - Two inputs current loop sensors 0/4-20 ma - Impulse input ( flow meter) - Dry impulse input 115..230 V AC switch-on cycle counter (connection EPS supply) - Switch input (connection dry fault relay contacts of EPS), Ie = 5 ma - Switch input alarm suppression during maintenance/service Outputs - Analog output 0/4-20 ma DC, adjustable via menu, burden max. 100Ω - Output of supply voltage of 24 V DC each two voltage and current sensor - Relay output pre-alarm, min. 5 V DC, 50 ma, max. 250 V AC, 5 A ohmic load - Relay output main-alarm, min. 5 V DC, 50 ma, max. 250 V AC, 5 A ohmic load - Relay output with programmable switch function, (e.g. temperature alarm) Interface (optional) - RS485 interface to import setups and communication with superior control or alternatively connection of digital UV-sensors Functions - Counter total operation hours (not resettable), lamp hours counter and switch-on cycles - Display necessary lamp exchange - Display lamp failure - Display main-alarm, pre-alarm, temperature alarm - Backlight color changes acc. to alarm status - reference sensor check while operation according to DVGW/ÖNORM Possibilities - Private label of front panel modifications - Factory-provided programming specific customized setup (from 10 pieces) - Tailor-made program version (from 100 pieces) - Hardware changes (from 200 pieces) Protection class Front IP65, connection side IP00 Complied standards CE, EMC and Low Voltage Directive fully compatible to rules DVGW W294 and ÖNORM M5873 General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 2 of 8

Mounting The monitor is designated installations in control boards and must be mounted in a cut-out of about W x H 92 +0,8 x 92 +0.8 mm. The required installation depth including space cable connections its rear is at least 75 mm. Fixation is by screw clamps which are in the scope of delivery. The connection side mains supply is executed with a slightly larger plug. Thereby and together with clear descriptions of each and every terminal, maximum installation safety is provided. Connection of terminals 1/2: Operation voltage monitor (90-264 V AC or 100-300 V DC) 3/4: Input operation voltage of electronic ballasts (115-230 V AC) 5/8: Input fault signal contact of electronic ballasts 6/8: Switch input alarm suppression 7/8: Impulse input flow meter with impulse output 9/10: Programmable signaling output e.g. temperature alarm (opener) 11/12: Output UV-pre-alarm (opener) 13/14/15: Output main alarm UV, dose and lamp failure (changeover contact) 16/17: Input UV-sensor (relative), diode, 16-anode, 17-cathode 18/19/20/21: Input UV-sensor with voltage output 2/5/10 V DC 21/22: Input sensor with voltage output 2 V DC (temperature, flow, UV) 21/23: Output operation voltage sensors with voltage output (max. 20 ma) 24/25: Input sensor with current loop output 0/4-20 ma (UV, flow) 26/27: Input Sensor with current loop output 0/4-20 ma (flow, UV) 28/29: RS485 connectivity to superior control, setup-storage/programming 30/31: Output current loop signal 0/4-20 ma, transfer of UV-value to superior control Rear sticker Housing dimensions (rear) 91,0 89,8 91,0 89,8 R3 Please check proper connections bee commissioning! Damages caused by faulty wiring are not covered by warranty! General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 3 of 8

Commissioning After connecting the supply voltage to the monitor, its type and software version are shown in the display. After some seconds the monitor does automatically switch into the measurement mode. The whished adjustments can now be set via the front keys according to the menu (see below). In case that the password protection is active, the menu is just showing reduced options. Hours counter can be reset in this menu after a lamp exchange was made. Details are given in the menu description later. Functionality of keys, navigation in menus In the very beginning, the device has to be adjusted to its individual functions. A keypad with 6 keys on its front helps to navigate through the menu and to adjust/program all values. The menu is intuitive. Pushing the keys shortly is enough. The ESC-key is equipped with a special function (see menu structure). Alternatively, the monitor can be programmed with a computer by using the RS485 interface (option). This procedure is recommended if same configurations must be installed as part of a serial production. If 10 pieces or more are ordered within one batch, monitors can be programmed individually in the factory. Functionality of keys right left up down g enter Esc Cursor one position to the right, scrolling through menu Cursor one position to the left, scrolling through menu Increase digit at cursor position by 1, scrolling through menu Decrease digit at cursor position by 1, scrolling through menu Confirm settings, open up menu/submenu Cancellation of input without changes, return to superior level/ basic display Special function: pushing a long time starts function reference measurement After 2 minutes the monitor does automatically switch back into the basic display if no key is pushed. Display The basic display does always show the alarm status in the first line. Its background lighting refers to the alarm mode and changes color from dark green (normal operation), via lime green (pre-alarm) into red (main alarm). The second line does automatically switch between all values. These can also be changed manually by pushing the arrow keys. During the programming procedure, the first line displays the menu level while the second line shows the chosen menu item. In the setup menu, the first line of the display shows the value which has to be reached and the second line displays the numerical value which must be adjusted. Bringing the cursor to the right position is by pushing the keys and the value is changeable via the keys. General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 4 of 8

Menu structure: BASIC DISPLAY MAIN MENU SUB MENU ADJUST. MENU NOTE STATUS DISPLAY VALUE DISPLAY LANGUAGE CHOOSE LANGUAGE German / English select language COUNTER* LIFETIME LIFETIME reset / return reset counter of operation hours after lamp replacement SWITCHING CYCLES SWITCHING CYCLES reset / return reset counter of ON/OFF cycles after lamp replacement UV SENSOR 1* INPUT SELECT INPUT diode, I in1, I in2, U in1, U in2 select input UV Sensor UV SENSOR 2* as sensor 1 UNIT (not applicable diode input) SELECT UNIT W/m², mw/cm², µw/cm² select the desired unit the value display CALIBRATION CALIBRATION VALUE 1 5000 enter calibration value of UV Sensor and equivalent Iout EQUIVALENT=100% 1 5000 enter value 100% equivalent CALIBRATION DIODE OK / return calibration to start value after lamp replacement CALIBRATION VALUE 25 250 % define a percentage which will be the start value of the lamp OFFSET (not applicable diode input) OFFSET-VALUE -10 0 +10 enter offset value of UV Sensor SENSOR TEMPERATURE on / off activate temperature sensor if present INPUT SELECT UNIT U in1, U in2, I in1, I in2 select input temperature sensor UNIT SELECT INPUT Celsius, Fahrenheit select the desired unit the value display CALIBRATION CALIBRATION VALUE 50 200 C, 122 392 F calibrate to a certain value OFFSET OFFSET-VALUE -100 0 +40 enter offset value of temperature sensor SENSOR FLOW RATE ein / aus activate flow sensor if present INPUT SELECT INPUT fixed val., I in1/2, U in1/2, impulse select input flow sensor UNIT SELECT UNIT m³/h, l/min select the desired unit the value display CALIBRATION CALIBRATION VALUE 0,1 1800 m³/h 1,7 30000 l/min calibrate to a certain value DOSE on / off activate if dose calculation desired CAPACITY CAPACITY 1-1000 Liter enter reactor capacity CORRECTION FACTOR CORRECTION FACTOR 1-100 % enter specific correction factor the reactor type General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 5 of 8

BASIC DISPLAY MAIN MENU SUB MENU ADJUST. MENU NOTE ALARM WARM UP DELAY WARM UP DELAY 0(off) 900 seconds time to reach operation temp. of UV lamps ALARM DELAY ALARM DELAY 1 60 seconds delay time releasing alarm UV MAIN ALARM MAIN ALARM 0(off) 99 % threshold value main alarm UV PRE ALARM PRE ALARM 0(off) 99 % threshold value pre alarm (> main alarm) LAMP ERROR LAMP ERROR ALARM NCC / NOC choose if alarm is at relay position open/closed DOSE ALARM DOSE ALARM 0(off) 99999 J/m² threshold value dose alarm LIFETIME ALARM LIFETIME ALARM 0(off) 30000 h lifetime of lamps according to manufacturer data TEMPERATURE ALARM TEMPERATURE ALARM 0(off) 100 C (212 F) threshold value temperature alarm RELAIS PRE ALARM RELAY PRE ALARM NCC / NOC choose if alarm is at relay position open/closed ALARM RELAY 3 ALARM RELAY 3 NCC / NOC choose if alarm is at relay position open/closed CURRENT OUTPUT MAXIMUM MAXIMUM I out = 10 500 % enter maximal value 20 ma current output MODE MODE 0-20 ma / 4-20 ma set operation mode of current output port CHANGE PASSWORD* PASSWORD 0(off) 30000 enter password access to the complete menu DEFAULT SETTINGS* DEFAULT SETTINGS set / return reset to default settings FACTORY MENU password protected access only manufacturer STATUS DISPLAY VALUE DISPLAY REFERENCE MEASUREMENT* press ESC button 3s 2 min time reference measurement * remaining menu points if the menu is reduced by a password General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 6 of 8

Reference measurement For reference measurements it is essential that the UV system is operating under stable conditions and the display may not swing. Pushing the ESC-key in the basic display about 3 seconds does freeze the actual reading and suppresses all alarm signals (background lighting blue). The field sensor can then be removed without causing alarms and substituted by the reference sensor. The measurement reading of the reference meter can now easily be compared against the frozen value in the monitor display. Since the monitor is giving a time indication, it makes it easy to put the field sensor back into the system on time. After two minutes, the monitor does automatically switch back into its measurement mode. Pushing the ESC-key a longer time allows to return earlier. After that, is has to be checked whether changes of the UV reading occurred. If so, a stable operation condition was not reached yet and the comparison measurement must be repeated. For reference measurements, we recommend our reference radiometer MUV 2.4 WR. All our SUV20 sensors with 40 or 160 opening angle are applicable as field sensor. To use our SUV20 sensors, one of our sensor-ports must be installed in the UV chamber. General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 7 of 8

My remarks: My programming: My connection scheme: My password : My additional notes: General Conditions Sales and Supplies of UV-Technik Speziallampen GmbH are valid. Provision of the manual is mandatory. More at www.uvtechnik.com. page 8 of 8