ISOMETER isomed427p. Application Medical IT system in accordance with IEC , IEC , IEC and DIN VDE

Similar documents
ISOMETER IR425. Insulation monitoring device for unearthed AC/DC control circuits (IT systems)

ISOMETER IR425. Insulation monitoring device for unearthed AC/DC control circuits (IT systems)

ISOMETER IRDH375. Insulation monitoring device for unearthed AC, AC/DC and DC systems (IT systems)

ISOMETER IRDH275BM-7 with coupling device AGH675-7 and AGH675-7MV15

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

ISOSCAN EDS460/490 EDS461/491. Insulation fault locators with control and display function for EDS systems (insulation fault location systems)

ISOSCAN EDS460-DG. Insulation fault locator for DC IT systems with high system leakage capacitances

ISOMETER IR1575. Insulation monitoring device for unearthed AC/3(N)AC systems up to 480 V and DC systems up to 480 V

LINETRAXX VME420. Multi-functional relay for overvoltage, undervoltage and frequency monitoring in AC/DC systems with external supply voltage

LINETRAXX RCMA420. Residual current monitor for monitoring AC, DC and pulsed DC currents in TN and TT systems

Insulation fault evaluator EDS460-DG

1.7. Insulation fault evaluators EDS460/490 EDS461/491

1.7 A-ISOMETER IRDH575. RoHS. Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Tel.: Fax:

LINETRAXX VMD420. Multi-functional voltage relay for 3(N)AC systems, frequency/overvoltage/undervoltage, phase, phase failure, asymmetry

IT system floor-standing distribution

1.7. Insulation fault evaluators EDS460/490 EDS461/491

isopv425 and AGH420 Ground Fault Detector for Ungrounded Solar Arrays < 100 kw And Isolation Tester Prior to Array Startup (Grounded and Ungrounded)

Remote alarm indicator and test combination MK2430

Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str Grünberg Phone: Fax:

Residual current monitor RCM470LY/RCM475LY. for TN and TT systems (AC and pulsating DC currents)

isopv and AGH-PV Ground Fault Detector for Ungrounded Solar Arrays up to 1100 V And Isolation Tester Prior to Array Startup (Grounded and Ungrounded)

GM420. Digital Ground Continuity Relay For AC Systems. Technical Bulletin NAE /

LINETRAXX RCMS460-D/-L RCMS490-D/-L

Operating Manual RCM420

CMD420 / CMD421 Series

TM800. Remote alarm indicator and operator panels

RCMA423-DM RCMA423-DM1C RCMA423-DM2C RCMA423-DM3C. Operating Manual

Operating Manual VMD420

VME420 Voltage and frequency monitor

VME421H Voltage and frequency monitor

ISOMETER IRDH265 IRDH365

Manual CME420. Multi-functional current relay, AC, overcurrent/undercurrent/ window discriminator function CME420_D00034_02_M_XXEN/05.

Bender GmbH & Co. KG. All rights reserved. Reprinting only with permission of the publisher. Subject to change! Photos: Bender archives

T M LIM2010. Line Isolation Monitor (LIM) Technical Bulletin NAE /

RCMA423. Residual current monitor for monitoring AC-, DC- and pulsed DC currents in TN- and TT systems Software-Version: D330 V1.0x.

BENDER A-ISOMETER IRDH265 IRDH365

Electrical safety costs little. a human life is priceless

RCMA423-DM1C Residual current monitor

LINETRAXX RCMS150. Residual current monitor type B with integrated measuring current transformers for earthed AC/DC systems (TN and TT systems)

MK Series. Remote Indicators For Line Isolation Monitors and Fault Location. Technical Bulletin NAE /

Line Isolation Monitor (LIM) LIM2000plus

Installation / Monitoring Technique

ISOMETER isopv1685rtu isopv1685p isopv1685pfr

Operating Manual MS220DA

A-ISOMETER IRDH575. Operating Manual

ISOMETER IRDH375 IRDH375B

Cat. N. Cat. N (s) 1. DESCRIPTION - USE 3. OVERALL DIMENSIONS. Technology : 4. PREPARATION - CONNECTION 2. PRODUCT RANGE

Parameterisable compact fault annunciator with permanent display

Universal fault annunciator for panel mounting

Operating Manual MS220KA and MSR220KA

Monitoring Relays. 3RN1 Thermistor Motor Protection. For PTC sensors. 7/120 Siemens LV 1 T 2008

Bender GmbH & Co. KG All rights reserved. Reprinting only with permission of the publisher. Subject to change! Photos: Bender archives.

EARTH FAULT METER MD-08

enproteko.com Specifications Ratings Item voltage to 24 VAC 5/6 Hz 24 VAC 5/6 Hz or 24 VDC Allowable voltage Characteristics 85% to 11% of power suppl

T M. User Manual LIM2010. Line Isolation Monitor

DTA. Motor thermistor relay. Benefits. Description

DUAL MONITORED INPUT/OUTPUT UNIT BN-305-2

PROTECT SRB 200EXi-1A

Analog Annunciator Unit SACO 16 A3. Product Guide

Monitoring Devices 11/1

AO Annunciator Module

Product overview. Alarm indicator, operator and display panels

ACS-30-EU-EMDR- 10-MOD

PRELIMINARY EDITION

GMA4100-DP. Control Panel. Operation Manual. Worldwide Supplier of Gas Detection Solutions

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number

AFMS Audible Flow Monitoring System TECHNICAL SPECIFICATION

DOCUMENTATION IPS640

Instructions for the fan motor control system with integrated wiring terminals SILVER C

DOCUMENTATION IPS640

Residual Current Devices

DEB71. Earth Leakage Monitoring Relay. Benefits. Description

Connection. e. g. Pt100. reference limit. yellow Output I. Composition

HTC-915-CONT. Heat-Trace Control system

Interface and EOL Modules 8

Process Alarm Module RTA 421

C Note! LGI16... Burner Controls. Building Technologies

Two-Channel Gas Controller

Project planning (original) EN. Functional safety. Inverter i550-cabinet kw

PolyGard Single Point Controller SPC-X3-20XX for refrigerant gases and vapours

DPB52. True RMS 3-Phase, Multifunction monitoring relay. Benefits. Description

OilSET Installation and Operating Instructions. Oil Separator Alarm Device

Operating instructions Safety-monitoring module SRB 302X3. 1 About this document

REPEATER FS5200R INSTRUCTION MANUAL

www. ElectricalPartManuals. com Annunciator unit Features Self-contained microprocessor-based annunciator unit with 64 alarm channels Application

DTA04. Motor thermistor relay. Benefits. Description

Annunciator Unit SACO 16 D1. Product Guide

Danfoss gas detection Controller unit

Combustible Gas Detection and Control System

IR5002. Operating Manual. Insulation monitoring device for IT DC systems V ISO Quality System Certified

Fire Control Panel FS5100

Universal controller for panel mounting Models CS6S, CS6H, CS6L

C-Bus Four Channel General Input Unit Installation Instructions

RTA421. Technical Information. Limit alarm switch Alarm switch and power supply for monitoring current or voltage signals

General Routing Interface

Oil Burner Controls LMO24... LMO44...

FX FIRE DETECTION SYSTEM

Oil Burner Controls LMO24... LMO44...

Safety Temperature Limiter STL 50 Certified to DIN EN (replaces DIN 3440)

MLHG6 / MLHG6-AU. Iso-Gard Line Isolation & Overload Monitor (LIOM) and Line Isolation Monitor (LIM) Installation and Reference Guide DANGER

Transcription:

Insulation monitoring device with integrated load and temperature monitoring and locating current injector for insulation fault location systems for medical IT systems _D00043_02_D_XXEN/09.2016

Insulation monitoring device with integrated load and temperature monitoring and locating current injector for insulation fault location systems for medical IT systems Product description The ISOMETER monitors the insulation resistance of unearthed C circuits (medical IT systems ). t the same time, the load current and temperature of the IT system transformer is monitored. In combination with EDS series insulation fault locators and the appropriate measuring current transformers, the is designed to set up the respective equipment for insulation fault location. pplication Medical IT system in accordance with IEC 60364-7-710, IEC 61557-8, IEC 61557-9 and DIN VDE 0100-710 ISOMETER Device features Insulation monitoring for medical IT systems djustable response value for insulation monitoring Locating current injector for insulation fault location systems Load and temperature monitoring for IT system transformers djustable load current response value Temperature monitoring with PTC thermistor or bimetal switch Self monitoring with automatic alarm PE connection monitoring Internal/external test button LEDs: Power On, larm 1, larm 2 Configurable alarm relay: N/O or N/C operation selectable Compact two-module enclosure (36 mm) MS interface pprovals Function The monitors the insulation resistance as well as the temperature and load current of the IT system transformer in medical IT systems. In addition, the connections to PE, to the measuring current transformer and to the temperature sensor are monitored. The actual measured value is indicated on the LCD. y pressing the or buttons, additonal measured values can be displayed. larms are indicated on the LC display via LEDs and an additional identification. Parameters are assigned to the device via LCD or the function keys on the front of the device. Insulation monitoring The uses the MP measurement method, which is also able to detect DC faults. When the value of the insulation resistance falls below the set response value, the alarm relay K1 switches and the alarm LED L1 lights. When the insulation resistance exceeds the release value (response value plus hysteresis), the alarm relay returns to its initial position and the alarm LED L1 goes out. Insulation fault location is carried out with insulation fault evaluators of the EDS series and the respective measuring current transformers. Once an insulation fault is detected by, the insulation fault location process is started automatically or manually. The generates a test current the amplitude of which is dependent on the existing system voltage and the insulation fault. In the case of insulation faults of low resistance, the locating current is limited to 1 m by the. The locating current pulse flows from the via the live parts to the point of fault. From there, it flows via the insulation fault and the earth conductor (PE) back to the. This locating current pulse is then detected by the measuring current transformers located in the insulation fault path, and is evaluated by the EDS insulation fault locators. When the locating current in the measuring current transformer exceeds the response value, the associated alarm LED at the EDS lights up indicating the faulty sub-circuit. This information is also indicated at the respective MK alarm indicator and test combination. y assigning the measuring current transformers to the respective circuit, the point of fault can easily be detected from a central position. Load current and temperature monitoring The load current is monitored using an STW2 measuring current transformer, temperature is monitored by means of temperature switch or a PTC thermistor in accordance with DIN 44081. When the response value is exceeded, the alarm LED L2 lights up. The required temperature sensors are already incorporated in ender transformers. larm relays The alarm relay switches when an alarm, a device error occurs or in the case of voltage failure. The factory-programmed operating principle can be re-programmed. Standards The ISOMETER of the series complies with the requirements of the device standards: IEC 60364-7-710, IEC 61557-8, IEC 61557-9 and DIN VDE 0100-710. 2 _D00043_02_D_XXEN/09.2016

larm messages LEDs ON L1 L2 Operation System fault* flashing flashing flashing Insulation fault Overcurrent Overtemperature * Detailed alarm information on LCD Test function/connection monitoring self test is carried out once supply voltage is fed and later at hourly intervals. During the self test, the internal device functions, the connections to PE (E/KE) and the connections to the current transformer are monitored for interrruption and short-circuit. In the event of a fault, the alarm relay K1 switches and the LEDs ON/L1/L2 flash. The respective error code appears on the LC display. fter eliminating the fault, the alarm relay automatically switches to its initial position. y pressing the test button, the device functions and also the relay function will be tested. Operating elements 1 2 3 4 5 6 7 1 - Power On LED 2, 3 - larm LEDs L1, L2 4 - LC display 5 - TEST button (>2s): to call up the self test. rrow up button: parameter change to move up in the menu 6 - rrow down button: parameter change to move down in the menu 7 - MENU button (> 2s): to call up the menu system. Enter button: to confirm parameter change Wiring diagram 3 STW2 1 N 2 6 6 R 4 5 J-Y(St)Y 2x2x0,6 on off 1 - Connection to the IT system to be monitored = supply voltage U S via fuse 2 - Temperatur sensor 3 - Measuring current transformer for load current monitoring 4 - Line protection by a fuse in accordance with IEC 60364-4-43 (6 fuse recommended). In case of supply (L1/L2) from an IT system, both lines have to be protected by a fuse. 5 - Serial interface MS _D00043_02_D_XXEN/09.2016 3

Wiring diagram with CMS460-D4 6 Z Z/k I E KE L1 L2 6 LINETRXX CMS460 11 12 14 k1 I k2 k3 I R on off J-Y(St)Y 2x2x0,6 pplication example MS N450 STW2 ES710 Serie EDS151 MK2430 MK2430 4 _D00043_02_D_XXEN/09.2016

Technical data Insulation coordination acc. to IEC 60664-1/IEC 60664-3 Rated insulation voltage 250 V Rated impulse voltage/pollution degree 4 kv/3 Protective separation (reinforced insulation) between (L1, L2, E, KE,,,,, Z, Z/k, l) - (11, 12, 14) Voltage test acc. to IEC 61010-1 2.21 kv Voltage supply Supply voltage U S Power consumption = U n 6,5 V IT system being monitored acc. to IEC 60364-7-710 Nominal system voltage U n C 70 264 V Nominal frequency f n 47 63 Hz Insulation monitoring acc. to IEC 61557-8 Response value R an 50 500 kω (50 kω)* Relative uncertainty ±10 % Hysteresis 25 % Response time t an at R F = 0.5 x R an and C e = 0.5 μf 5 s Response time for PE connection monitoring 1 h Permissible system leakage capacitance C e 5 μf Measuring circuit Measuring voltage U m Measuring current I m (at R F = 0 Ω) Internal DC resistance R i Impedance Z i at 50 Hz Permissible extraneous DC voltage U fg Locating current injector acc. to IEC 61557-9 Locating current Test pulse/break ±12 V 50 μ 240 kω 200 kω DC 300 V 1 m 2/4 s Load current monitoring Response value, adjustable 5 50 (7 )* Relative uncertainty ± 5 % Hysteresis 4 % Nominal frequency f n 47 63 Hz Setting values load current measurement: Transformer 3150 V 4000 V 5000 V 6300 V 8000 V 10000 V Ialarm 1~ 14 18 22 28 35 45 Response time overload, (50 % to 120 %) < 5 s Response time for measuring current transformer monitoring at restart, test or every 1 h Temperature monitoring: Response value (fixed value) 4 kω Rated frequency f n 47 63 Hz Release value (fixed value) 1.6 kω PTC resistors acc. to DIN 44081 max. 6 in series Relative uncertainty ± 10 % Response time overtemperature < 2 s Response time connection fault PTC resistors < 2 s Displays, memory LC display multifunctional, not illuminated Measured value insulation resistance 10 kω 1 MΩ Operating uncertainty ± 10 %, ± 2 kω Measured value load current (as % of the set response value) 10 % 199 % Operating error ± 5 %, ± 0.2 Password on, off/0 999 (off, 0)* Interface Interface/protocol RS-485/MS aud rate 9.6 kbit/s Cable length 0 1200 m Cable (twisted in pairs, one end of shield connected to PE) recommended: J-Y(St)Y min. 2 x 0.8 Terminating resistor 120 (0.25 W), internal, switchable Device address, MS bus 2 90 (3)* Interfaces for measuring current transformer STW2 and temperature sensor Cable lengths: single wire > 0.5 mm² 1 m single wire, twisted > 0.5 mm² 10 m twisted in pairs, shielded > 0.5 mm² 40 m Cable (shield on one side connected to PE) recommended: J-Y(St)Y min. 2 x 0.6 Switching elements Number 1 changeover contact Operating principle N/C operation or N/O operation (N/C operation)* Electrical endurance, number of cycles 10000 Contact data acc. to IEC 60947-5-1 Utilisation category C-13 C-14 DC-12 DC-12 DC-12 Rated operational voltage 230 V 230 V 24 V 110 V 220 V Rated operational current 5 3 1 0.2 0.1 Minimum contact rating 1 m at C/DC 10 V Environment/EMC EMC IEC 61326-2-4 Operating temperature -25 +55 C Classification of climatic conditions acc. to IEC 60721: Stationary use (IEC 60721-3-3) 3K5 (no condensation, no formation of ice) Transport (IEC 60721-3-2) 2K3 Long-term storage (IEC 60721-3-1) 1K4 Classification of mechanical conditions acc. to IEC 60721: Stationary use (IEC 60721-3-3) 3M4 Transport (IEC 60721-3-2) 2M2 Long-term storage (IEC 60721-3-1) 1M3 Connection Connection type Connection properties rigid flexible without ferrule flexible with ferrule Stripping length Opening force Test opening, diameter push-wire terminals 0.2 2.5 mm² (WG 24 14) 0.75 2.5 mm² (WG 19 14) 0.2 1.5 mm² (WG 24 16) 10 mm 50 N 2.1 mm Other Operating mode continuous operation Position of normal use any Degree of protection, internal components (DIN EN 60529) IP30 Degree of protection, terminals (DIN EN 60529) IP20 Enclosure material polycarbonate Flammability class UL94V-0 DIN rail mounting acc. to IEC 60715 Screw mounting 2 x M4 with mounting clip Documentation number D00043 Weight 150 g ( )* = factory setting _D00043_02_D_XXEN/09.2016 5

Ordering information Supply voltage U S= U 1) n C Type rt. No. 70 264 V, 47 63 Hz -2 7207 5301 1) bsolute values of the voltage range Dimension diagram XM420 Dimensions are given in mm Open the front plate cover in direction of arrow! Screw fixing Note: The upper mounting clip must be ordered separately (see ccesories ). ccessories Type designation Mounting clip for screw mounting (1 piece per device) rt. No. 9806 0008 Suitable system components Type designation Type Page Measuring current transformers STW2 942 709 Temperature sensor (PTC) ES0107 924 186 Three phase load monitor CMS460-D4-2 9405 3030 Mounting frame XM420 990 994 ender GmbH & Co. KG P.O. ox 1161 35301 Gruenberg Germany Londorfer Strasse 65 35305 Gruenberg Germany Tel.: +49 6401 807-0 Fax: +49 6401 807-259 E-Mail: info@bender.de www.bender.de ENDER Group _D00043_02_D_XXEN / 09.2016 / pdf / ender GmbH & Co. KG, Germany Subject to change! The specified standards take into account the version that was valid at the time of printing.