Recirculating Cooler SRC14

Similar documents
Water Bath SWB1 SWB2 SWB3 SWB1D SWB2D SWB3D. Instructions for use Version 2.0. water bath SWB2D. mains on on. press to set. set. off.

FGT5 & FGT6 Gelation Timer. Instructions for use

73 Series Spectrophotometer. Accessory fitting and operation manual

Analog Melting Point Apparatus

Hotplate Stirrers. with BioCote antimicrobial protection

Aluminum Heating Mantles

INSTRUCTION MANUAL (English version) Water still

Instruction Manual JEN0002 / Version 1.0. Peltier Accessory

Recirculating Chiller RC350G-RC400G RC1400G-RC3000G. Operating Manual

Assembly and Operating

Instruction Manual ET0002 / Version 1.0. Heating Tape

PARAFFIN SECTION MOUNTING BATH

MICRO-KJELDAHL EQUIPMENT ( MM SERIES ) OPERATING AND SAFETY INSTRUCTIONS

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1

Dry Bath Heating Systems

MD-MINI Dry Bath. Instruction Manual

CAUTION - READ THESE INSTRUCTIONS BEFORE USING THE MACHINE

AVAC9000W Portable Air Conditioner with 24 hour timer Installation Instructions and User Guide

ELECTRIC BUNSEN INSTRUCTION BOOK. BA6101, BA6101/C, BA6101x1, BA6101x3. Page 1 of 20 M6877 Issue 9.6

Autofill counter top water boiler

Operator s Manual. IP-100 Immersion Probe Cooler

Autofill wall mounted water. boiler. Getting the best from your water. boiler. Please read and keep these instructions

Instruction Manual for Bench-Top Cleaning Baths

VR60 GLASS FROSTERS INSTALLATION AND OPERATING MANUAL PLEASE LEAVE WITH OPERATOR. VR60 Glass Froster. Series 1

FWC860 Wine Cooler Installation, Use and Maintenance

ER MANUAL FOR YOUR BAUMATIC

LABORATORY ILLUMINATED FLOTATION BATH

OPERATING INSTRUCTIONS FOR:

Important Notes: Please read carefully before installing your refrigerator.

Operation Manual Portable Room Air Conditioner Models: RPC26WA / RPC35WA / RPC41WA

Whole Room Dryer WRD, WRD/110

OPERATING INSTRUCTIONS FOR: MODEL: TGO-80W, 150W, 240W GLASSWARE DRYING OVEN

Table of Contents. Features Quality Policy Section A - Unit Operation Section B Unit Maintenance Trouble Shooting Guide...

USER S MANUAL INTERSYSTEM PRESSURE TRANDUCER TYPE PM-22

OPERATING MANUAL TORREY PINES SCIENTIFIC, INC. DIGITAL HPLC COLUMN HEATER MODEL CO20 DOCUMENT NUMBER CO20-00

FWC152 / FWC302 / FWC602 / FWC622 Wine Coolers

Heated Display Cabinet Installation & Operating Instructions

INFRARED IP55 HEATER INSTRUCTIONS FOR: MODEL:- QZWP45N 1. SAFETY INSTRUCTIONS

Distinction. D4000 & D4000/EURO Water Still. Assembly and Operating Instructions. 1 Issue 2

INSTRUCTIONS Français 2 Nederlands 15 English 28 Italien

DK-11 Countertop Water System. Owner's Manual

Portable Air Conditioner

OWNER S MANUAL. Vintage Classic HEAT COOL models. Proudly Made in the USA

fwc152 / fwc303 / fwc603 / fwc623 Wine Coolers Manual for Installation, Use and Maintenance

Aquamax KF Oil Evaporator. User Manual

I n s t r u c t i o n m a n u a l f o r b u i l t - i n h o o d. Model code: BORA600

FWC153, FWC304, FWC604 & FWC624

Installation & Operating Instructions

USER GUIDE. TL6015W TL6018W 60cm Larder

Chilller Arctic Instruction Manual

DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4

FWC152, FWC303, FWC603 & FWC623 Wine Coolers Installation, Use and Maintenance

USER MANUAL SILENT12 PORTABLE AIR CONDITIONER

CHEST FREEZER USER GUIDE LEC CF61LW

INSTALLATION MANUAL ESM9+ / ESM11+ / ESM15+

Operating Manual. English. Refrigerators for chemicals KRC 50 KRC 180

Double Door - Dual Zone Wine Cooler

USER MANUAL. Pharmacy Refrigerator Range. Applicable models: PSR353/PGR353 PSR273/PGR273 PSR151/PGR151

G-10s. Instruction Manual. G-Series Cooler UPRIGHT COOLER. Part No.11IPA

Dehumidifier. Instruction Manual. Model DH-320/A

Owners Manual OWNERS MANUAL. Australia s original Outdoor brand. Part No. Comp830Li. SAI Global Cert. SAI AS

10L Dehumidifier. Model Number: LDH V AC 50/60Hz 290W

Operator s Manual. Histology Bath

SH3020 USER INSTRUCTIONS

Swaging Machine Type H50-46ECO

WITH OMRON E5CC CONTROL

THE BOILING WATER DISPENSER INSTALLATION & OPERATING INSTRUCTIONS IMPORTANT: READ AND SAVE THESE INSTRUCTIONS FOR THE BENEFIT OF THE USER

USER MANUAL SILENT16 PORTABLE AIR CONDITIONER

THROUGH-WALL AIR-TO-AIR HEAT PUMP AND AIR CONDITIONER. Instruction Manual. Model AMB-12H

Mobile Air Conditioner Instruction Manual Model TC-N9KM

WINDOW TYPE ROOM AIR CONDITIONER OPERATION AND INSTALLATION MANUAL

Split Portable Air Conditioner

Portable Air Conditioner USER MANUAL

Owner s Manual. Portable Air Conditioner ARP-5008 ARP Royal Sovereign International, Inc.

Optimax Refrigerated Retail Merchandisers

Optimax Refrigerated Retail Merchandisers

Singulier Spout Tube Basin Mixer Installation and User Guide

Thermoelectric Cigar Cooler

1 L Rotary Vacuum Evaporator Instruction Manual

CLEAN WATER PUMP. MODEL No: Hippo4A OPERATION & MAINTENANCE INSTRUCTIONS. Part No:

User and Installation Instructions C6R/75, C6R/100, C6R/105, C6R/125, C6R/130, D6R/100, D6R/125, D5R/100, D5R/125, D3R/100

Bar Fridge USER MANUAL MB46W

USER MANUAL. Laboratory Refrigerator Range. Applicable models: LSR151 LSR288

OWNERS INSTRUCTION MANUAL

CROWN WATER HEATERS CPU10 - CPU15 CPOS10 - CPOS15

Commercial backpack vacuum cleaner

MODULAR REFRIGERATOR. Français p. 16 Español p. 31 ITEM # MODEL #BG179B

Wine Cellar. Instruction Booklet. Model: WC-2038

USERS MANUAL. Desiccant Dehumidifier. Model: ECO DD322FW SIMPLE. Read This Manual

Installation & Operating Instructions

XC Portable Air Conditioning

ARTICA. 8000Btu. Room Air Conditioners Portable Range

The Classeq under counter range

Oblo Basin and Bidet Monobloc Installation and User Guide

PC71 American Style Fridge Freezer

Owner s Manual 33-bottle Under-Counter Wine Cooler Model: WC-3302US

20L Dehumidifier. Model Number: LDH V AC 50/60Hz 440W

User Instruction Manual

PC52 American Style Fridge Freezer

30-pint Dehumidifier User Guide

Transcription:

Recirculating Cooler SRC14 Instructions for use Version 2.0

Tank cover Flow control valve 90 o Elbow Control panel Outlet nozzle Inlet nozzle Cooling coil Circulation tank Maintenance panel Drain with stopper Castor (with Stopper) Fig.1 Recirculating Cooler SRC14 Instructions for use Introduction Electrical Installation THIS EQUIPMENT MUST BE EARTHED Before connection please ensure that the line supply corresponds to that shown on the rating plate. This model requires a supply rated at 220-240V, 50 Hz, ~, single phase. Thank you for purchasing this piece of Stuart equipment. To get the best performance from the equipment and for your own safety, please read these instructions carefully before use. Before discarding the packaging check that all parts are present and correct. This equipment is designed to operate under the following conditions: For indoor use only Use in a well ventilated area Ambient temperature range +5ºC to +35ºC Altitude to 2000m Relative humidity not exceeding 80% Mains supply fluctuation not exceeding 10% Over-voltage category II IEC60364-4-443 Pollution degree 2 If the equipment is not used in the manner described in this manual and with accessories other than those recommended by Stuart, the protection provided may be impaired. Model Supply requirements Power SRC14 230V, 50 Hz, 1.1kW There is an IEC socket at the rear of the instrument for connection to the mains supply. The unit is supplied with two mains leads fitted with IEC plugs for connection to the instrument. One has a UK 3 pin plug and the other has a 2 pin Shuko plug for connection to the mains. Choose the lead appropriate for your electrical installation and discard the other. Should neither lead be suitable, take the lead with the UK plug and replace the plug with a suitable alternative. See the enclosed instruction sheet for advice on how to carry out this procedure. Should the mains lead require replacement a cable of 1mm 2 of harmonised code H05W-F connected to an IEC 320 plug should be used. THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN The appropriate mains lead should be connected BEFORE connection to the mains supply.

General Description The SRC14 accurately controls the temperature of a cooling liquid between -20 to +30ºC. The cooled liquid is then circulated to an external device via the inlet and outlet nozzle. The SRC14 has a capacity of 14 litres. Safety Advice Before Use Beware some cooling mediums are ignitable above ambient temperatures, do not use with hazardous materials or liquids. Never move or carry the unit when in use or connected to the mains electricity supply. Please take care not to touch the cooling elements under the air filter, as these can be sharp. Do not place anything on top of the unit during operation Installation Position your recirculating cooler on a firm level surface. A byproduct of the air cooled refrigeration unit is heat, unnecessary heat build up can lower the cooling performance of the unit. To avoid this please ensure the unit is located in a well ventilated area. The unit should not be exposed to direct sunlight, humidity, dust or vibration. For optimum performance the following clearance should be allowed: 20cm from both sides, 60cm above the unit and 40cm behind. Do not obstruct the front ventilation on either unit. Care should be taken not to tilt the SRC14 by more than 45 degrees when moving. Ensure that the castors on the SRC14 are locked prior to use (only the front 2 castors have stoppers). Operation Connection of hoses: Connect the 90º elbows to the nozzles then connect the elbow adapters including flow control valve (Fig. 2) Connect suitable tubing (ID 9mm) to each nozzle (inlet and outlet) NOTE: Take care not to use excessive force connecting the hoses, use universal clips or similar to secure the tubing if necessary. Always consider the temperature and pressure requirements before selecting your tubing material. Use tubing of the minimum length required, excessively long tubing will reduce the efficiency of the unit. Fig.2 Filling procedure: When choosing your cooling medium you should take care that there is no risk of corrosion of internal parts, these include the following materials; copper, stainless steel, brass, PTFE, polyacetal resin, silicone rubber, polypropylene oxide and ethylene propylene rubber. Distilled water should also be avoided as it could dissolve metal substances in the circulation line. For operation below +7ºC it is recommended that ethylene glycol is used as your cooling medium. Prior to filling the unit make sure that the drain stopper is in place. Pour 14 litres of your cooling medium into the circulation tank, or until the liquid level is approximately 2 cm from the top of the tank, take care not to overfill. Replace the circulation tank cover. Operation Connect the power lead to the IEC connector at the back of the unit. Make sure that the power switch and pump switch are in the off position on the control panel (Fig. 3) then connect the power cable to the mains supply. Turn the power switch on the main control panel to the on position. The display will show the current temperature of the cooling medium. The refrigeration unit will automatically start if the current temperature is 0.5ºC higher than the set temperature (the temperature is set to 20ºC before shipment). The refrigerator must be switched on manually using the switch on the control panel, once on, it will automatically cut in and out to control the temperature of the cooling medium.

After operation: Switch the circulation pump off, then switch off the power switch and the mains power. Place an appropriate collection vessel for the cooling medium underneath the drain hose and the drain stopper. Once all fluid has been drained replace the drain stopper. Fig.3 Setting the temperature: To set the temperature of the cooling medium press the SET key once, the display will change from the current temperature to the set temperature (blinking) use the up and down keys to reach your desired value, hold down either key to scroll the temperature. After reaching the desired value press the set key again, the display will revert to the current temperature. If the SET key is pressed once and no other key is pressed for five seconds the display will automatically revert to the current temperature. The set temperature is retained in the memory when mains power is switched off. The refrigerator must be switched on manually using the switch on the control panel, once on it will automatically cut in and out to control, the temperature of the cooling medium. Maintenance & Servicing WARNING: Ensure the unit is disconnected from the mains electricity supply before attempting maintenance or servicing. Cleaning the refrigeration air filter: If the refrigeration air filter blocks it can affect the efficiency of the unit. Periodically this filter should be checked and cleaned. Disconnect the unit from the mains power supply. The front panel is removable: press on the two black squares in the top corners of the front panel and the panel will open. Remove the filter (fig 4). Be careful not to touch the cooling elements as these can be sharp. Shake the dust out of the filter and wash with mild detergent if necessary. Allow the filter to air dry before replacement. Air filter Adjustment of flow rate: Use the flow rate valve on the outlet nozzle to adjust the flow rate. Operation of the pump with this valve close could damage the pump. Take care when opening this valve; opening the valve too quickly could cause damage to the tubing or glassware in the circulation line. Calibration of display temperature: For normal operation it is not usually necessary to calibrate the temperature. Measure the temperature of the cooling liquid using a thermometer. Press the set key for more than 5 seconds. Pb will be displayed. Using the up and down key to select the correct value as measured on the thermometer (calibration can only be performed between +9.9ºC and -9.9ºC). After setting the correct value press the set key for more than 5 seconds, the display will return to the current measured temperature. Fig.4 Cleaning of the cooling liquid filter: The cooling liquid filter is located at the bottom of the circulation tank (fig 5). Drain the tank then remove the filter, it should easily pull free. Clean and replace. Cooling liquid filter Fig.5 SRC14 Sensor

General Maintenance Periodically wipe clean the inside of the circulation tank and locate the fuse box at the rear of the unit (Fig. 6). Use mild detergent if necessary. Regularly check connections for leakage. The condition of the o-ring in the inlet and outlet nozzle should be checked periodically and replaced if necessary. At high temperatures or high humidity water can collect in the cooling system of the SRC14, this should be drained periodically via the hose located at the back of the unit. To change the fuse on the SRC14, switch off the mains power. Insert a small screwdriver into the slot and push downwards, as shown in Fig. 7. Fuse box Fig.6 Fig.7 AC outlet Drainage hose Bibby Scientific Ltd. Stone, Staffordshire ST15 0SA United Kingdom Tel: +44 (0) 1785 812121 Fax: +44 (0) 1785 813748 e-mail equipment@bibby-scientific.com www.bibby-scientific.com Warranty Bibby Scientific Ltd warrants this instrument to be free from defects in material and workmanship, when used under normal laboratory conditions, for a period of three (3) years. In the event of a justified claim Bibby Scientific will replace any defective component or replace the unit free of charge. This warranty does NOT apply if damage is caused by fire, accident, misuse, neglect, incorrect adjustment or repair, damage caused by incorrect installation, adaptation, modification, fitting of non approved parts or repair by unauthorised personnel. Technical Specification SRC14 Cooling capacity 1200W at 10ºC Temperature range -20 to +30ºC Control accuracy ±2ºC Bath capacity 14 litres Pump rate 18 litres / minute Dimensions (w x d x h) 354 x 384 x 851mm Net weight 41kg Electrical supply 230V, 50Hz To open fusebox insert a small screwdriver into the slot and push downwards Any repairs or replacement of parts MUST be undertaken by suitably qualified personnel. For a comprehensive list of parts required by service engineers conducting internal repairs and a service manual, please contact the Technical Service Department of Bibby Scientific Ltd. quoting both the model and serial number. Only spare parts supplied or specified by Bibby Scientific Ltd. or its agents should be used. Fitting of non-approved parts may affect the performance and safety features designed into the instrument. If in any doubt, please contact the Technical Service Department of Bibby Scientific Ltd. or the point of sale.

Instructions for use Note: Alarm is cleared by turning off the power switch Problem Cause Solution Cooling liquid is not circulated. Flow pressure is reduced. Cooling liquid filter is blocked. Air is trapped in the pump. Tubing is trapped. Pressure loss due to excessive tubing length. The circulator is too low compared to the device it is cooling. Clean the filter See the Maintenance and Servicing section for details. Briefly remove the drain stopper. until all trapped air has been removed. Check the tubing isn t pinched, twisted or trapped Try to reduce the overall length of the tubing. Adjust the position of the recirculating cooler. No refrigeration of cooling medium. Refrigeration unit is broken. The cooling medium isn t sufficiently covering the cooling coil. Overload of refrigeration unit (alarm led illuminates) Contact your local service representative. Top up the cooling medium. Lower ambient temperature below 35ºC, or raise set temperature. Display does not illuminate when power is on. Cooling power is poor. Water freezes in the circulation tank at +7ºC Blinks on the display - - - Blinks on the display, refrigeration stops - F F No mains power. Circuit breaker has tripped. Fuse has blown. Temperature controller is broken. Refrigeration unit does not work. The ambient temperature exceeds 35ºC Exhaust is blocked. Refrigeration air filter is blocked Loss of flow pressure Temperature sensor malfunction Measured temperature is lower than the units lower limit (-50ºC) Temperature sensor short circuit Measured temperature exceeds the upper limit of the unit (more than +40ºC) Check mains power supply. Unscrew maintenance panel and check switchable circuit breaker is in the on position. If problem persists there could be an electrical leakage and you should contact your local service representative. Follow procedure to change fuse in the Maintenance and services section Contact local service representative Contact local service representative. Try to lower the ambient temperature. Check necessary clearance around the unit has been allowed. Follow the procedure detailed in the Maintenance and Servicing section to clean the filter. Check flow pressure, see above Use anti freeze for cooling medium. Contact local service representative. The heating load exceeds capacity, reduce load. Display is abnormal or unstable: E R Electrical noise causes abnormal display on temperature controller Temperature controller is reset by electrical noise Ambient temperature exceeds 35ºC Turn off mains power, wait 5 seconds then switch back on. Move unit to a new location. Lower ambient temperature below 35ºC.

This product meets the applicable EC harmonized standards for radio frequency interference and may be expected not to interfere with, or be affected by, other equipment with similar qualifications. We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice. Where there is a possibility that injury, damage or loss might occur if equipment malfunctions due to radio frequency interference, or for general advice before use, contact the Technical Department of Bibby Scientific Ltd.

INSPECTION REPORT MODEL SRC14 ELECTRICAL SAFETY 1. Earth continuity 2. Insulation 3. Flash test FUNCTIONAL 1. Indicators 2. Speed/temperature 3. Visual acceptance QUALITY CONTROL INSPECTOR Bibby Scientific Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel: +44 (0)1785 812121 Fax: +44 (0)1785 810471 e-mail: info@bibby-scientific.com www.bibby-scientific.com Bibby Scientific France SAS ZI du Rocher Vert - BP 79 77793 Nemours Cedex France Tel: +33 1 64 45 13 13 Fax: +33 1 64 45 13 00 e-mail: bsf@bibby-scientific.fr www.bibby-scientific.com Bibby Scientific US Ltd 3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel: 800-225-9243 Fax: 609-589-2571 www.bibby-scientific.com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel: +39 (0)2 98230679 Fax: +39 (0)2 98230211 e-mail: marketing@bibby-scientific.it www.bibby-scientific.it Bibby Scientific (Asia) Ltd Room 607 Yen Sheng Centre 64 Hoi Yuen Road Kwun Tong Kowloon Hong Kong Tel: +852 3583 1581 Fax: +852 3583 1580 e-mail bibby@bibby-scientificasia.com www.bibby-scientificasia.com